Bedienungsanleitung Agilent Technologies FS4100

39 Seiten 0.51 mb
Download

Zur Seite of 39

Summary
  • Agilent Technologies FS4100 - page 1

    FuturePlus Systems Corporation USB Analysis Probe Users Manual-FS4100 For Agilent Logic Analyzers Revision 2.7 FuturePlus Systems is a trademark of FuturePlus Systems Corporat ion Copyright 1998 FuturePlus Systems Corporation ...

  • Agilent Technologies FS4100 - page 2

    HOW TO REACH US 4 PRODUCT WARRANTY 5 Limitation of w arranty 5 Exclusive Remedies 6 Assistance 6 INTRODUCTION 7 How to Use This Manual 7 ANALYZING THE UNIVERSAL SERIAL BUS 9 Duplicating the Master Diskette 9 Accessories Supplied 9 Minimum Equipment Required 10 Revisions 10 Signal Naming Conventions 10 The USB Analysis Probe Switches and Diodes 10 T ...

  • Agilent Technologies FS4100 - page 3

    The STAT variable 19 The ADDR and DATA variables 19 The PID variable 19 The CTLCMD variable 21 COMBINATION MODE 22 Bad EOP Detection 22 Low Speed/High Speed detection 23 STATE ANALYSIS 25 Installation Quick Reference 25 Acquiring Data 26 The State Display 26 Analysis Probe related error messages 27 USB CRC and Serial Bit Errors 27 Installation Quic ...

  • Agilent Technologies FS4100 - page 4

    How to reach us For Technical Support: FuturePlus Systems Corporation 36 Olde English Road Bedford NH 03110 TEL: 603-471-2734 FAX: 603-471-2738 On the web http://www.futureplus.com For Sales and Marketing Support: FuturePlus Systems Corporation TEL: 719-278-3540 FAX: 719-278-9586 On the web http://www.futureplus.com FuturePlus Systems has technical ...

  • Agilent Technologies FS4100 - page 5

    Product Warranty This FuturePlus Systems product has a warranty against defects in material and workmanship for a period of 1 year from the date of shipment. During the warranty period, Future Plus Systems will, at its option, either replace or r epair products proven to be defective. For warranty servic e or repair, this product must be returned t ...

  • Agilent Technologies FS4100 - page 6

    product, or improper site pr eparation or maintenance. NO OTHER WARRANTY IS EXPRESSED OR IMPLIED. FUTUREPLUS SYSTEMS SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. THE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND EXCLUSIVE REMEDIES. FUTUREPLUS SYSTEMS SHALL NOT BE LIABLE FOR ANY DIRECT, IND ...

  • Agilent Technologies FS4100 - page 7

    Introduction The USB Analysis Probe module provides a complete interface between any poi nt on the USB and an Agilent Logic Analyzer. T he Analysis Probe interface receives the USB serial bit stream and converts it to an understandable parallel bus that is then acquired by the logic analyzer. The USB Analysis Probe is a passive bus monitor which do ...

  • Agilent Technologies FS4100 - page 8

    • The State Analysis chapter explains how to configure the USB Analysis Probe to perform state analysis on your USB . • The Timing Analysis chapter explains how to configure the USB Analysis Probe to perform timing analysis on your USB . • The General Information chapter provides some general information including the operating characteristic ...

  • Agilent Technologies FS4100 - page 9

    Analyzing the Universal Serial Bus This chapter introduces you to the USB Analysis Probe and lists the minimum equipment required and accessories supplied for Universal Serial Bus analysis. This chapter also contains information that is common to both state and timing analysis. Before you use the USB Analysis Probe software, make a duplicate copy o ...

  • Agilent Technologies FS4100 - page 10

    The minimum equipment required for analysis of a Universal Serial Bus consists of the following equipment: Minimum Equipment Required • A 166x, 167x , 1680/90, 1671x , 1674X, 1655x or 1675x logic analyzer • The USB Analysis Probe Product • A USB target wire Revisions This manual applies to Hardware revisions ACE or later and software revision ...

  • Agilent Technologies FS4100 - page 11

    The following explains how to connect the logic analyzer to the USB Analysis Probe for either state or timing analysis: Connecting to the USB Analysis Probe 1. Remove the probe tip assemblies from the logic analyzer cables. 2. Plug the logic analyzer cables into the USB Analysis Probe cable headers as shown in the appropriate following tables. Logi ...

  • Agilent Technologies FS4100 - page 12

    The logic analyzer can be configured for USB analysis by loading the USB configurati on file. Loading this file will load the Universal Serial Bus inverse assembler and configure your logic analyzer for USB analysis. Setting up the Analyzers installed in the 16500 mainframe. 1. Install the 16500 U SB Analysis Probe Software for the FS4100 flexible ...

  • Agilent Technologies FS4100 - page 13

    configuration file yourself, see table below for a list of analyzers and corresponding configuration files. Note: The Logic Analysis System’s Operating System must be version A.01.40.00 with patches 034 and 035 or higher to be compatible with the software provided on the installation diskette for the 167xx. The 1680/90/900 Analyzer is a PC based ...

  • Agilent Technologies FS4100 - page 14

    the menu bar at the top of the screen and select the decoder from the list. After loading the configuration file of choice, go into the format specification of the configuration by choosing Setup from the menu bar and then selecting Bus/Signal in the drop down menu. When the format specification appears press Define Probes at the bottom of the scre ...

  • Agilent Technologies FS4100 - page 15

    1680/90/900, 167xx Configuration files 167xx Analyzer 169xx Analyzer File Name 16550 / 16710/1/2 CUSB41_1 16555/6/7 CUSB41_2 16715/6/7, 1674x, 1675x CUSB41_3 1680/90 (State) 1680/90, 16750/1/2, 1691x (State) CUSB41_6 1680/90 (Timing) 1680/90, 16750/1/2, 1691x (Timing) CUSB41_7 Offline Analysis Data that is saved on a 167xx analyzer in fast binary f ...

  • Agilent Technologies FS4100 - page 16

    will be a series of questions, answer the first question asking which startup option to use, select Continue Offline. On the analyzer type question, select cancel. When the application comes all the way up you should have a blank screen with a menu bar and tool bar at the top. For data from a 1680/90/900 analyzer, open the .ala file using the File, ...

  • Agilent Technologies FS4100 - page 17

    picture; please choose the appropriate decoder for the trace you are working with. After the decoder has loaded, select Preferences if required, from the overview screen and set the preferences to their correct value in order to decode the trace properly. This is a general requirement, some decoders do not have preferences, if this is the case then ...

  • Agilent Technologies FS4100 - page 18

    The Format Menu The USB Analysis Probe diske tte sets up the format menu as shown in the followi ng table. This format is the same for both Timing and State Analysis. Pods 1 and 2 are for STATE analysis and POD 3 is useful for Timing analysis. Label Pod 4 Pod 3 Pod 2 Pod 1 STAT 15:11 16 ADDR 10:0 DATA 15:0 15:0 ADR 10:4 ENDPNT 3:0 PID 15:11 16 CTLC ...

  • Agilent Technologies FS4100 - page 19

    The STAT variable is used by the USB inverse assembler to decode USB transactions. It should not be changed or deleted from the format m enu. The STAT variable The ADDR and DATA variables The ADDR variable is the address and endpoint as decoded from the USB serial bit stream. The DATA variable is the Data associated with a Dat a0 or Data1 packet. F ...

  • Agilent Technologies FS4100 - page 20

    Symbol Binary Code Comment RESET END 111001 RESET DEASSERTS INVALID 011101 INVALID PID RECEIVED BAD PID 011110 BAD PID RECEIVED SUSPEND 111110 SUSPEND CONDITION DETECTED RESUME 111101 RESUME CONDITION DETECTED KEEP ALIVE 111011 KEEP ALIVE SETUP SBS ERROR 001001 SETUP PACKET WITH SERIAL BIT STUFFED ERROR DETECTED IN SBS ERROR 001010 IN PACKET WITH S ...

  • Agilent Technologies FS4100 - page 21

    Symbol Binary Code Comment ERROR DETECTED SOF CRC ERROR 010100 START OF FRAME PACKET WITH CRC ERROR DETECTED OUT CRC ERROR 010011 OUT PACKET WITH CRC ERROR DETECTED CRC DATA CRC ERR 010111 CRC DATA WITH CRC ERROR DETECTED ANY ERROR 111000 ERROR SUMMARY BIT SLOW SOF 011111 SOF TOKEN DETECTED AT SLOW SPEED RESUME SE0 111100 SE0 SENT AT END OF RESUME ...

  • Agilent Technologies FS4100 - page 22

    Standard Device Requests GET_CONFIGURATION SET_CONFIGURATION GET_INTERFACE SET_INTERFACE SYNCH_FRAME COMBINATION MODE The default mode set up by the software shipped with the Analysis Probe is the following configuration. Bad EOP Detection If the FS4100 operating in full speed mode, detects a valid SE0 for at least 60ns but a valid J_STATE does not ...

  • Agilent Technologies FS4100 - page 23

    3. Pass a “POSSIBLE BAD EOP” status to the Analyzer. A status of “POSSIBLE BAD EOP” should be treated as cautionary and the user would be advised to view the received signals in timing mode by doing the following. 1. Attach POD 3 of the analyzer to POD 3 of the USB Analysis Probe. 2. Attach a flying lead set to stake pins RCV, VP and VM. 3. ...

  • Agilent Technologies FS4100 - page 24

    Sequence 2 1. Detection of a PREAMBLE - switch to low speed 2. Look for IN packet and a following DATA packet and then switch back to high speed. If a DATA packet is not received wit hin the time-out period (18 bit times), the Analysis Probe will switch back to high speed. 3. In high speed look for PREAMBLE and handshake If the Analysis Probe is pl ...

  • Agilent Technologies FS4100 - page 25

    State Analysis This chapter explains how to configure the USB Analysis Probe to perform state analysis on the Universal Serial Bus. T he configuration software on the flexible diskette sets up the format specification menu of the logic analyzer for compatibility with the USB Analysis Probe. The nex t chapter explains how to configure the USB Analys ...

  • Agilent Technologies FS4100 - page 26

    Probe OUT connector can be connected down stream to either a Hub or device or it can be left unconnected. 4. Turn on the logic anal yzer so that the VCC and READY LEDs on the USB Analysis Probe are lit. NOTE: Connect the USB Analysis Probe to the USB wire as described in step 2 prior to powering up the logic analyzer. 5. Load the USB Analysis Probe ...

  • Agilent Technologies FS4100 - page 27

    Analysis Probe related error messages The following Analysis Probe related error messages are reported by the USB inverse assembler. FAILED TO IDENTIFY PID This error will be reported by the inverse assembler if the USB Analysis Probe hardware reports a status code that is undefined. If this error occurs please contact the FuturePlus Systems factor ...

  • Agilent Technologies FS4100 - page 28

    The USB Analysis Probe will also detect invalid PIDS, bad PIDS and SOF tokens s ent at slow speed. The Analysis Probe implements an error summary status called ANY ERROR, this st atus will be sent to the Analyzer if any of the above error conditions is detected by the Analysis Probe. 28 ...

  • Agilent Technologies FS4100 - page 29

    Timing Analysis Pod 3 of the USB Analysis Probe contains signals that describe in more detail the actual state of the USB wire. Installation Quick Reference The following procedure describes the major steps required to perform timing analysis measurements with the USB Analysis Probe module. 1. After removing the probe tip assemblies, plug logic ana ...

  • Agilent Technologies FS4100 - page 30

    Acquiring Data Touch RUN and the logic analyzer will begin to acquire data. The analyzer will continue to acquire data and will display the data when the analyzer memory is full, the trigger specification is TRUE or when you touch STOP. The logic analyzer will flash “Waiting for Trigger” or “occurrences remaining in level x” where x is numb ...

  • Agilent Technologies FS4100 - page 31

    POD 3 channel Signal Name Description 14 SOFTIC Start of Frame. 1 millisecond timer generated from recovered start of frame 13:11 EOP2_0 End of Packet state machine 10:7 LBC3_0 Load Byte State Machine 6:4 RSRVD Reserved 3 FEOPR End of packet 2 FEOSYN End of Sync 1 LSDET Low Speed Detect 0 unused The end of packet state machine state encodings are a ...

  • Agilent Technologies FS4100 - page 32

    State Name State Encoding Description EOPR_STATE 001 J state transition detected. Valid end of packet SLOW_SEO_FI RST 101 First slow speed single ended zero condition detected SLOW_SEO_S ECOND 011 First slow speed single ended zero condition detected BAD_EOPR_ST ATE 111 Incorrect transition for single ended zero detected. NOTE: FEOPR will be genera ...

  • Agilent Technologies FS4100 - page 33

    DELAY_1 0010 State delayed by one clock tic DELAY_2 0011 State delayed by two clock tics SOF_OR_ADR 0100 Receiving address for token packet or Frame number SOF_OR_EP 0101 Receiving endpoint for token packet or the remainder of the frame number for start of frame CRC 0110 Receiving CRC data PRE_DATA 0111 About to receive data DATA 1000 Receiving dat ...

  • Agilent Technologies FS4100 - page 34

    POD 4 Description the transceiver is generating in response to the USB traffic. POD 4 channel Signal Name Description 0 RCV Differential data signal from the USB transceiver 1 VM VM Signal from the USB transceiver 2 VP VP Signal from the USB transceiver. The Waveform Display Captured data is displayed as shown in the following figure. 34 ...

  • Agilent Technologies FS4100 - page 35

    General Information This chapter provides additional reference information including the characteristics and signal connections for the USB Analysis Probe module. Characteristics The following operating characteristics are not specifications, but are typi cal operating characteristics for the USB Analysis Probe. Standards Supported The Universal Se ...

  • Agilent Technologies FS4100 - page 36

    Environmental Temperature Operating: 0 to 55 degrees C (+32 to +131 degrees F) Non operating:-40 to +75 degrees C (-40 to +167 degrees F) Altitude Operating: 4,6000m (15,000 ft) Non operating: 15,3000m (50,000 ft) Up to 90% non condensing. Avoid sudden, extreme temperature changes which would cause condensation on the Analysis Probe module. Humidit ...

  • Agilent Technologies FS4100 - page 37

    Analysis Probe Cable Header and Pin number Logic Analyzer channel number USB Analysis Probe Signal name Header 3 pin 3 CLK/16 CLK12 5 no connect 7 15 MDATA 9 14 SOFTIC 11 13 EOP<2> 13 12 EOP<1> 15 11 EOP<0> 17 10 LBC<3> 19 9 LBC<2> 21 8 LBC<1> 23 7 LBC<0> 25 6 Reserved 27 5 Reserved 29 4 Reserved 31 3 FEOPR ...

  • Agilent Technologies FS4100 - page 38

    Analysis Probe Cable Header and Pin number Logic Analyzer channel number USB Analysis Probe Signal name Header 2 pin 3 CLK/16 MCLK 5 no connect 7 15 STAT4 9 14 STAT3 11 13 STAT2 13 12 STAT1 15 11 STAT0 17 10 ADR06 19 9 ADR05 21 8 ADR04 23 7 ADR03 25 6 ADR02 27 5 ADR01 29 4 ADR00 31 3 ENDPNT03 33 2 ENDPNT02 35 1 ENDPNT01 37 0 ENDPNT00 38 ...

  • Agilent Technologies FS4100 - page 39

    Analysis Probe Cable Header and Pin number Logic Analyzer channel number USB Signal name Header 1 pin 3 CLK/16 STAT5 5 no connect 7 15 D15 9 14 D14 11 13 D13 13 12 D12 15 11 D11 17 10 D10 19 9 D09 21 8 D08 23 7 D07 25 6 D06 27 5 D05 29 4 D04 31 3 D03 33 2 D02 35 1 D01 37 0 D00 39 ...

Produzent Agilent Technologies Kategorie Computer Drive

Dokumente, die wir vom Produzenten des Geräts Agilent Technologies FS4100 erhalten, können wir in mehrere Gruppen teilen. Unteranderem in:
- technische Zeichnungen Agilent Technologies
- Bedienungsanleitungen FS4100
- Produktkarten Agilent Technologies
- Informationsbroschüren
- oder Energieetiketten Agilent Technologies FS4100
Jede von ihnen ist wichtig, jedoch finden wir die wichtigsten Informationen für den Nutzer des Geräts in der Bedienungsanleitung Agilent Technologies FS4100.

Die Dokumentengruppe, die als Bedienungsanleitungen bezeichnet wird, wird ebenfalls in detaillierte Arten geteilt, solche wie: Montageanleitungen Agilent Technologies FS4100, Wartungsanleitungen, Kurzanleitungen oder Benutzeranleitungen Agilent Technologies FS4100. Abhängig vom Bedarf, sollten Sie das Dokument finden, das Sie brauchen. In unserem Service können Sie sich die populärste Bedienungsanleitung des Produkts Agilent Technologies FS4100 ansehen.

Die komplette Bedienungsanleitung des Geräts Agilent Technologies FS4100, wie sollte sie aussehen?
Die Bedienungsanleitung, auch bezeichnet als Benutzerhandbuch, oder einfach nur „Anleitung”, ist ein technisches Dokument, das dem Benutzer bei der Nutzung von Agilent Technologies FS4100 hilfreich sein soll. Die Bedienungsanleitungen werden in der Regel von technischen Schriftstellern geschrieben, aber in einer Sprache, die für alle Nutzer von Agilent Technologies FS4100 verständlich ist.

Eine gänzliche Bedienungsanleitung von Agilent Technologies sollte einige Grundelemente enthalten. Ein Teil von ihnen ist nicht so wichtig, wie z.B.: die Titelseite oder Autorenseiten. Die restlichen von ihnen jedoch, sollten Informationen liefern, die für den Nutzer von enormer Wichtigkeit sind.

1. Einführung und Hinweise, wie man sich in einer Bedienungsanleitung von Agilent Technologies FS4100 bewegt - Am Anfang jeder Bedienungsanleitung sollten wir Hinweise bezüglich der Nutzungsart eines bestimmten Ratgebers finden. In ihr sollten sich Informationen über die Lokalisierung des Inhaltsverzeichnisses von Agilent Technologies FS4100 befinden, FAQ oder über oft auftretende Probleme – also Stellen, die von den Benutzern in jeder Bedienungsanleitung am meisten gesucht werden
2. Inhaltsverzeichnis - Index aller Ratschläge bezüglich Agilent Technologies FS4100, die wir im aktuellen Dokument finden
3. Ratschläge zur Nutzung der Grundfunktionen des Geräts Agilent Technologies FS4100 - die uns die ersten Schritte während der Nutzung von Agilent Technologies FS4100 erleichtern sollten
4. Troubleshooting - geordneter Tätigkeitslauf, der uns bei der Diagnose und als nächstes bei der Lösung wichtiger Probleme mit Agilent Technologies FS4100 hilft
5. FAQ - häufig gestellte Fragen
6. Kontaktdaten Informationen darüber, wo man Kontakt zum Produzenten / Service von Agilent Technologies FS4100 im bestimmten Land suchen kann, wenn es nicht gelingt, das Problem selbst zu lösen.

Haben Sie eine Frage bezüglich Agilent Technologies FS4100?

Nutzen Sie das untere Formular

Wenn Sie mit Hilfe der gefundenen Bedienungsanleitung Ihr Problem mit Agilent Technologies FS4100 nicht gelöst haben, stellen Sie eine Frage, indem Sie das untere Formular nutzen. Wenn einer der Nutzer ein ähnliches Problem mit Agilent Technologies FS4100 hatte, ist es möglich, dass er mit Ihnen die Lösung teilen möchte.

Text vom Bild übertragen

Kommentare (0)