    VP AP ™ III & III S T User Manual English 2481 27/1 05 07 Global leaders in sleep and respiratory medicine www .r esmed.com 248127/1 VP AP III & III ST User Manual ENG ...

  • ResMed 248127 - page 2

    VP AP ™ III & III S T Illustrations Component of 2481 27/1 HUMIDIFIERS A B 12 4 Remo vable front cover SETTING UP THE VP AP D 1 2 4 LCD screen Handle Control panel Air outlet Air filter cover Serial communications port (9-pin) Power switc h Power soc k et © 2005 ResMed Ltd. © 2005 ResMed Ltd. © 2005 ResMed Ltd. © 2005 ResMed Ltd. O THER A ...

  • ResMed 248127 - page 3

    A TT ACHING A P ASSO VER HUMIDIFIER H 4 © 2005 ResMed Ltd. 3 A TT ACHING A HUMIDAIRE HUMIDIFIER G © 2005 ResMed Ltd. 12 4 3 12 (2 m/3 m) (52 cm) (2 m/3 m) (52 cm) REPLACING THE AIR FIL TER I © 2005 ResMed Ltd. 1 2 3 248127r1 illos.qxd 20/07/2005 3:28 PM Page 2 ...

  • ResMed 248127 - page 4

    VP AP ™ III & III S T User Manual English ...

  • ResMed 248127 - page 5

    Protected by patents: A U 697 65 2, A U 699726, AU 71 3679, EP 0661 071, EP 0858352, US 494431 0, US 5199424, US 5522382, US 62131 19, US 6240921, US 670531 5. Other patents pending. Protected by design registrations: A U 1 47283, AU 1 47335, A U 1 47336, CH 1 28.709, CH 128.71 0, CH 128.71 1, CH 1 28.71 2, DE 4020 1 723, DE 40202007 , DE 40202008, ...

  • ResMed 248127 - page 6

    iii Contents Cont ents Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 User/Owner R esponsibilit y 1 Medical Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 What the VP AP III and VP AP III ST are Inten ded for 1 Contraindications 1 W arnings 2 Cautions ...

  • ResMed 248127 - page 7

    iv Limited W ar ranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43 Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45 ...

  • ResMed 248127 - page 8

    1 Introduction Introduction The VPAP™ III and VPAP III ST are bilevel pressure support ventilators specifically designed for non-invasive mask ventilation. This user manual contains the information you need for the correct use of your VP AP . User/Owner Responsibility The user or owner of this system shall have sole responsibility and liability f ...

  • ResMed 248127 - page 9

    2 • conditions predisposing to a risk of aspiration of gastric contents • impaired abilit y to clear secretions • hypotension or significant intra vascular v olume depletion • pneumothorax or pneumomediastinum • recent cranial trauma or surger y . Warnings • The entire manual should be read bef ore using the VP AP . • Advice contained ...

  • ResMed 248127 - page 10

    3 Medical Information Note : The above are ge neral warnings and cautions. Further specific warnings, cautions and notes appear next to the relevant instructions in the manual. Adverse Effects You should report unusual chest pain, severe headache or increased breathlessness to your phy sician. An acute upper respiratory tract infection may require ...

  • ResMed 248127 - page 11

    4 ...

  • ResMed 248127 - page 12

    5 The VP AP Sy stem The VPAP System Please ref er to the illustrations in section A of the illustration sheet. Please identify and familiarise yourself with the following components of the VP AP unit: • VP AP front vie w ( A- 1 ) • VP AP rear view ( A- 2 ) • P ower cord ( A- 3 ) • Carr y bag ( A- 4 ) • 2 m air tubing ( A- 5 ). ! WARNING ? ...

  • ResMed 248127 - page 13

    6 Notes: • ResMed VP AP devices hav e been designed and manufact ured to provide optimum performance using R esMed vented mask sy stems. While other vented mask sy stems ma y be used, performance and data outputs may be af f ected. T o select an appropriate set ting for another mask sy stem, find the closest matc h to a ResMed mask in T able 3 on ...

  • ResMed 248127 - page 14

    7 Setting up the VP AP Sy stem Setting up the VPAP System Please ref er to the illustrations in section D of the illustration sheet. Setting Up The VPAP 1 Place the VP AP unit on a flat surface near the head of your bed. If the unit is placed on the floor, ensure that the area is free from dust and clear of bedding, clothes or any other objects tha ...

  • ResMed 248127 - page 15

    8 Attaching a HumidAire 2i Humidifier Please ref er to the illustrations in section E of the illustration sheet. The HumidAire 2i attaches to the front of the VP AP to provide heated humidification. No other accessories are required for its use. The VP AP automatically detects the presence of the HumidAire 2i . No menu changes are required. Please ...

  • ResMed 248127 - page 16

    9 Setting up the VP AP Sy stem The VP AP is now ready for use with the HumidAire . To start treatment, see “Starting Treatment” on page 13. Attaching a Passover Humidifier Please ref er to the illustrations in section H of the illustration sheet. Medium size (52 cm) air tubing is a necessary accessory for connecting the VP AP unit to the ResMed ...

  • ResMed 248127 - page 17

    10 Using the LCD Screen and Keypad The control panel of the VP AP includes an LCD screen and keypad. LCD Screen The LCD screen displays the menus and treatment screens. To assist you in adjusting the VP AP, the keypad and LCD are equipped with a backlight. The LCD backlight comes on when the unit is turned on or when you press a key, and turns off ...

  • ResMed 248127 - page 18

    11 Setting up the VP AP Sy stem The keypad backlight is on at all times when the VP AP is powered. Left (gre en) • P erforms the function indicated b y the guidi ng text display ed abo ve it on the LCD screen. Guiding text includes menu , enter , ch an ge and apply . Right (red) • P erforms the function indicated b y the guidi ng text display e ...

  • ResMed 248127 - page 19

    12 ...

  • ResMed 248127 - page 20

    13 Starting T reatment Starting Treatment The VP AP unit should be assembled beside your bed with the air tubing and mask system connected. See “Setting Up The VPAP” on page 7. 1 Turn the main power switch at the back of the unit to on ( I ) . When the VP AP is turned on, a welcome message is displayed on the LCD screen. The VP AP (or Ramp) scr ...

  • ResMed 248127 - page 21

    14 Stopping Treatment To stop treatment at any time, remove your mask and press the Fr ont key or if your clinician has enabled the SmartStart function, simply remove your mask and treatment will end (SmartStop is not applicable with the "Mir Full" mask setting). Using the HumidAire 2i Warm-Up Feature If using a HumidAire 2i with the VPAP ...

  • ResMed 248127 - page 22

    15 Starting T reatment Once you ha ve a secure and comf ortable fit, c hec k y our Mask-Fit st ar rating on the LCD screen. Definitions of the Mask-Fit star ratings are presented in T able 1. Note: If there is another person nearby to check your Mask-Fit star rating, you can adjust your mask, mask cushion and headgear while lying down. 4. After thr ...

  • ResMed 248127 - page 23

    16 ...

  • ResMed 248127 - page 24

    17 Cleaning and Maintenance Cleaning and Maintenance You should regularly carry out the cleaning and maintenance described in this section. Daily Cleaning Mask Clean the mask according to the instructions supplied with the mask. Air tubing Disconnect the air tubing from the VPAP unit (and humidifier, if used) and hang the tubing and mask in a clean ...

  • ResMed 248127 - page 25

    18 ! WARNING Beware of electric shock. Do not immerse the flow g enerator or power cord in water. Always unp lug the flow generator befo re cleaning and be sure that it is dry before reconnecting. ! CAUTION Do not attempt to open the VP AP . There are no user serviceable parts inside. Repairs and internal servicing should only be performed by an au ...

  • ResMed 248127 - page 26

    19 How to Use the VP AP Menus How to Use the VPAP Menus The VP AP unit provides a set of functions arranged in menus and submenus. Via the LCD screen, the menus and submenus allow you to view and change the settings for a particular function. You can access the menus regardless of whether the VP AP is in standby mode or delivering therapy. Your cli ...

  • ResMed 248127 - page 27

    20 Figure 1: VPAP Standard Menu Series • T o access the VP AP menus: Press the Lef t k ey (menu) while the VP AP (or Ramp) screen is displa yed. • T o scroll through items within the menu: Press the Up/Do wn key • T o exit out of the menu: Press the Ri ght ke y (exit) How to Use the Detailed Menu (if enabled by your clinician) The detailed me ...

  • ResMed 248127 - page 28

    21 How to Use the VP AP Menus Figure 2: VPAP Detailed Menu Series (if enabled by your clinician ) SETTINGS OPTIONS enter exit enter exit SER VICING enter exit 456789 exit SN:123456789123 MASK: UL TRA change exit TUBE LENGTH: 2M change exit TUBE LENGTH: 2m HUMID: NONE change exit SMARTST ART : OFF change exit MASK ALARM: OFF change exit SW: SX365020 ...

  • ResMed 248127 - page 29

    22 • T o access the VP AP menus: Press the Lef t k ey (menu) while the VP AP (or Ramp) screen is displa yed. • T o scroll through items within a menu or submenu: Press the Up/Do wn key • T o enter a submenu: Press the Lef t k ey (enter). • T o c hange a set ting option for a function: 1 . Press the Lef t k ey (c hange) 2. Press the Up/Do wn ...

  • ResMed 248127 - page 30

    23 How to Use the VP AP Menus SmartStart™ The VP AP has a function called SmartStart which can be enabled by your clinician. If SmartStart is enabled, VP AP will start automatically when you breathe into the mask and will stop automatically when you take your mask off. This means you do not have to press the Fron t key to begin or end treatment. ...

  • ResMed 248127 - page 31

    24 Results Menu Note: This menu appears only if at least one Smart Data™ option has been enabled by the clinician. See the VPAP III Smart Data Diary for further details. T able 4: Results Menu ST AND ARD Mirage Ac tiva Nasal Ma sk Mirage Vi sta Nasal Mask Modular Nasal Mask MIR FULL Mirage F ull Face Mask Mirage F u ll Face Mask Series 2 Ultra Mi ...

  • ResMed 248127 - page 32

    25 How to Use the VP AP Menus Options Menu T able 5: Options Menu Servicing Menu T able 6: Servicin g Menu * This it em also appears in the Standard Menu. Function Default F unction Description Set ting Op tions Smart Data –Auto Appear OFF The Smart Data menu is displa yed only if one or more of the options ha ve been set to ON by the clinician. ...

  • ResMed 248127 - page 33

    26 Treatment Screens After starting treatment, you can display one of the treatment screens below. Press the Up/Down key to switch between views. Figure 3: Treatment screen 1 Figure 4: Treatment screen 2 Figure 5: Treatment screen 3 The treatment screens contain the following information: T r eatment mode: Mode of treatment set by your clinician. O ...

  • ResMed 248127 - page 34

    27 How to Use the VP AP Menus Minute v entilation: Volume of air inhaled per minute (units: litres per minute). It is the product of respiratory rate and tidal volume. Tidal volume: Volume of air inhaled per breath (units: millilitres per breath). Cycle indicat or : How the VP AP changes the press ure when you are exhaling. "C" indicates ...

  • ResMed 248127 - page 35

    28 ...

  • ResMed 248127 - page 36

    29 Helpful Hints Helpful Hints Starting out Mouth Leaks If using a nasal mask, try to keep your mouth closed during treatment. Air leaks from your mouth can decrease the effectiveness of your treatment. If mouth leaks are a problem, a full face mask or chin strap may help. Contact your clinician or equipment supplier for further details. Mask Fitti ...

  • ResMed 248127 - page 37

    30 Note : When using the VPAP with a HumidAire 2i, use a pure sine wave inverter not a modified sine wave inverter. Please refer to the VPAP III battery power guide for battery capacities and further details. ...

  • ResMed 248127 - page 38

    31 T roubleshooting Troubleshooting If there is a problem, try the following suggestions. If the problem cannot be solved, contact your equipment supplier or ResMed. Do not attempt to open the unit. Pro bl em Possible Cause Sol ut i on No display . P o wer not connected or switc h at bac k is not on. Ensure the pow er cable is connected and th at t ...

  • ResMed 248127 - page 39

    32 Breath is not deep enough to trigger SmartSt art. T ak e a deep breath in and out through the ma sk. There is e xcessi ve leak. Adjus t position of mask and headgear . Plugs ma y be missing from ports on mask. Repla ce them. Air tubi ng not connected properly . Connect firmly at both ends. Air tubi ng kinked or punct ured. Straighten o r replace ...

  • ResMed 248127 - page 40

    33 T roubleshooting Excessiv e motor noise. Component fai lure. Ret urn your unit f or ser vicing. Displa y error message: High leak in last session . Y ou have experienced ex cessiv ely high leak lev els during the night. Chec k that your air t ubing is connected properly and that your mask does not leak ex cessiv ely . Use the mask-fit ting f eat ...

  • ResMed 248127 - page 41

    34 ...

  • ResMed 248127 - page 42

    35 Sy stem Specifications System Specifications Dynamic pr essure c haract er istics IPAP: 2 cm H 2 O to 25 cm H 2 O (measured at the end of standard 2 m air tubing) EPAP: 2 cm H 2 O to 25 cm H 2 O (measured at the end of standard 2 m air tubing) CPAP: 4 to 20 cm H 2 O (measured at the end of standard 2 m air tubing) Maximum single f ault pressur e ...

  • ResMed 248127 - page 43

    36 Air T ubing: Flexible plastic, 2 m or 3 m length IEC 6060 1 -1 Classifications Class II (double insulation) Type CF T able 7: Displayed values * Results may be inaccurate in the presence of leaks. ** The display ed values are estimates. T hey are pro vided for trending pur poses only . Va l u e Ra n ge Accuracy Display Re so l u t i on Pressure ...

  • ResMed 248127 - page 44

    37 Sy stem Specifications Pressure Variation Pressure Volume curve Note: The manufacturer reserves the right to change these specifications without notice. 0.00 5.00 10.00 15.00 20.00 25.00 0 1 2 3 Breaths Pressure (cm H 2 O) 10 BPM 15 BPM 20 BPM Max Pressure 2/3 Max Pressure 1/3 Max Pressure Max Pressure 0 100 200 300 400 500 600 19.7 19.8 19.9 20 ...

  • ResMed 248127 - page 45

    38 Symbols which appear on the product Attention, consult accompanying documents Class II equipment Type CF equipment Drip Proof Start/Stop or Mask-Fit Envir onmental infor mation WEEE 2002/96/EC is a European Directive that requires the proper disposal of electrical and electronic equipment. This device should be disposed of separately, not as uns ...

  • ResMed 248127 - page 46

    39 Sy stem Specifications Guidance and Manufacturer’s Declaration - Electromagnetic Emissions and Immunity Guidance and manufact urer’ s declaration – electromagnetic emissions The VP AP is intended for use in the electromagnet ic environment specified below . The customer or the user of the VP AP should assure that it is used in suc h an env ...

  • ResMed 248127 - page 47

    40 (Continued next page) Guidance and manufact urer’ s de claration – electr omagnetic immunity The VP AP is intended for use in the electromagnetic en vironment specified below . The customer or the user of the VP AP should assure that it is used in suc h an en vironment. Immunity t est IEC6060 1 -1 -2 test lev el Compliance level Electromagne ...

  • ResMed 248127 - page 48

    41 Sy stem Specifications Guidance and manufact urer’ s declaration – electromagnetic immunity (Continued) The VP AP is intended f or use in th e electromagnetic en vironment specified belo w . The custome r or the user of the VP AP should assure that it is used in suc h an environment. Immunity tes t I E C 6 0 6 01- 1- 2 test level Compliance ...

  • ResMed 248127 - page 49

    42 Recommended separation distances between portable and mobile RF communications equipment and the VP AP The VP AP is intended f or use in an e nvironment in whic h radiated RF dist urbances are controlled. T he customer or the u ser of the VP AP can help prev ent electromagnetic interf erence b y maintaining a minimum distance between portable an ...

  • ResMed 248127 - page 50

    43 Limited W arranty Limited Warranty ResMed warrants that your ResMed product shall be free from defects in material and workmanship for the period specified below from the date of purchase by the initial consumer. This warranty is not transferable. Note: Some models are no t available in all regions. If the product fails under conditions of norma ...

  • ResMed 248127 - page 51

    44 ...

  • ResMed 248127 - page 52

    45 Index Index A Accessories 6 Air filter 18 , 35 air tubing 36 cleaning 17 B Bac k-light, LCD and Keypad 10 C Cautions 2 cleaning periodic 17 weekly 17 Cleaning and Maintenance 17 cleaning mask 17 Cleaning P eriodically 17 D Daily Cleaning 17 definitions 1 Detail ed Menu 20 , 21 Dimensions 35 Dr yness 29 Dynamic pressure c haracteristics 35 E Elec ...

  • ResMed 248127 - page 53

    46 S Ser vicing 18 Ser vicing Me nu 25 set ting up 7 Set tings Menu 22 SmartStart 23 Stand ard Menu 20 Starting T r eatment 13 Stopping T reatment 14 Sy stem Specifications 35 T T r ave ll i ng 29 treatment screen 26 T roubleshooting 31 U Up/Down K ey 10 Used (time) screen 19 User/Owner R esponsibility 1 Using the Mask-Fit ting f eat ure 14 Using t ...

Produzent ResMed Kategorie Network Router

Dokumente, die wir vom Produzenten des Geräts ResMed 248127 erhalten, können wir in mehrere Gruppen teilen. Unteranderem in:
- technische Zeichnungen ResMed
- Bedienungsanleitungen 248127
- Produktkarten ResMed
- Informationsbroschüren
- oder Energieetiketten ResMed 248127
Jede von ihnen ist wichtig, jedoch finden wir die wichtigsten Informationen für den Nutzer des Geräts in der Bedienungsanleitung ResMed 248127.

Die Dokumentengruppe, die als Bedienungsanleitungen bezeichnet wird, wird ebenfalls in detaillierte Arten geteilt, solche wie: Montageanleitungen ResMed 248127, Wartungsanleitungen, Kurzanleitungen oder Benutzeranleitungen ResMed 248127. Abhängig vom Bedarf, sollten Sie das Dokument finden, das Sie brauchen. In unserem Service können Sie sich die populärste Bedienungsanleitung des Produkts ResMed 248127 ansehen.

Ähnliche Bedienungsanleitungen

Die komplette Bedienungsanleitung des Geräts ResMed 248127, wie sollte sie aussehen?
Die Bedienungsanleitung, auch bezeichnet als Benutzerhandbuch, oder einfach nur „Anleitung”, ist ein technisches Dokument, das dem Benutzer bei der Nutzung von ResMed 248127 hilfreich sein soll. Die Bedienungsanleitungen werden in der Regel von technischen Schriftstellern geschrieben, aber in einer Sprache, die für alle Nutzer von ResMed 248127 verständlich ist.

Eine gänzliche Bedienungsanleitung von ResMed sollte einige Grundelemente enthalten. Ein Teil von ihnen ist nicht so wichtig, wie z.B.: die Titelseite oder Autorenseiten. Die restlichen von ihnen jedoch, sollten Informationen liefern, die für den Nutzer von enormer Wichtigkeit sind.

1. Einführung und Hinweise, wie man sich in einer Bedienungsanleitung von ResMed 248127 bewegt - Am Anfang jeder Bedienungsanleitung sollten wir Hinweise bezüglich der Nutzungsart eines bestimmten Ratgebers finden. In ihr sollten sich Informationen über die Lokalisierung des Inhaltsverzeichnisses von ResMed 248127 befinden, FAQ oder über oft auftretende Probleme – also Stellen, die von den Benutzern in jeder Bedienungsanleitung am meisten gesucht werden
2. Inhaltsverzeichnis - Index aller Ratschläge bezüglich ResMed 248127, die wir im aktuellen Dokument finden
3. Ratschläge zur Nutzung der Grundfunktionen des Geräts ResMed 248127 - die uns die ersten Schritte während der Nutzung von ResMed 248127 erleichtern sollten
4. Troubleshooting - geordneter Tätigkeitslauf, der uns bei der Diagnose und als nächstes bei der Lösung wichtiger Probleme mit ResMed 248127 hilft
5. FAQ - häufig gestellte Fragen
6. Kontaktdaten Informationen darüber, wo man Kontakt zum Produzenten / Service von ResMed 248127 im bestimmten Land suchen kann, wenn es nicht gelingt, das Problem selbst zu lösen.

Haben Sie eine Frage bezüglich ResMed 248127?

Nutzen Sie das untere Formular

Wenn Sie mit Hilfe der gefundenen Bedienungsanleitung Ihr Problem mit ResMed 248127 nicht gelöst haben, stellen Sie eine Frage, indem Sie das untere Formular nutzen. Wenn einer der Nutzer ein ähnliches Problem mit ResMed 248127 hatte, ist es möglich, dass er mit Ihnen die Lösung teilen möchte.

Text vom Bild übertragen

Kommentare (0)