    Leica DB LB Research mi croscope and Stu do Lite Imaging so ftware Roo m B523 User Guide Molecular I maging Unit University of Helsinki www .m iu.helsin ki.fi 9.4.2008 ...

  • Leica B523 - page 2

    1 GENERA L USER INFORMATION .............................................................................. 1 2 SETTINGS FOR TRAN SMITTED LIGHT M ICROSCOPY...................................... 1 2.1 Turning on the microscope .................................................................................................... 1 2.2 XY-stage.......... ...

  • Leica B523 - page 3

    1 1 General user information Leica DM LB research microscope is onl y for th e Molecular Canc er Biology (MCBP) and Genome-scale Biology (GSB ) r esearch p rograms internal us e. It i s a transmitted light microscope that cannot be used for fluorescence microscop y . Bright-field, polarization contrast, and low-quality d ark-field i llumination are ...

  • Leica B523 - page 4

    2 movement in y -direction (back and forth), lower in x -direction (side to side). If you need to rotate the stage to get the image in any wanted angle, open the rotation lock by unscrewing it and th en rotating the stage by ha nd (circle in Figure 2). Re-screw the lock after use! Figure 2. Adjustment wheels for stag e control (arrows) and ro tatio ...

  • Leica B523 - page 5

    3 Objective Numerical Aperture (NA) Color ring Coverg lass (c .g.) thickness Position for auxili ary condensor lens N Plan 2.5x 0.07 Brown Any/without c.g. off N Plan 5x 0.12 Red Any/without c.g . off N Plan 10x 0.25 Yellow Any/without c.g. on N Plan 20x 0.4 Green 0.17 mm on N Plan 40x 0.65 Blue 0.17 mm on C Plan 63x 0.75 Black 0.17 mm on Table 1. ...

  • Leica B523 - page 6

    4 The fine focus has two modes: when the left focus wheel is pulled out, each l ine on the wheel equals 1 µ m of movement. When it is pushed in, each line equals 4 µ m of movement. Figure 5. Focusing wheels 2.6 Eye pieces (oculars) Adjust the folding bridge o f the ocul ars until you see, using bot h e y es, onl y one circular field of view throu ...

  • Leica B523 - page 7

    5 Figure 7. Dual viewing a ttachment Figure 8. Power source for the light pointer There is a light poin ter i n the viewing att achment that uses an external power sourc e. Thus, y ou need to connect the transformer into a power outlet b efore using the pointer (Figure 8). After use, disconnect the transformer. Notice that y ou can us e the monitor ...

  • Leica B523 - page 8

    6 Figure 9. Light pointer controls 2.8 Koehler illumination Koehler ill umination needs to be adjusted in the beginning of ever y imagi ng session for optimal illumination. Proceed as follows: Instruction View in eyepiece 1. Use 10x objective and front lens. Focus on the sample. 2. Close the luminous-field diaphragm by turning t he adjustment wheel ...

  • Leica B523 - page 9

    7 Instruction View in eyepiece 4. Center the diaphragm i n the field of view by r otating the condenser a djusti ng screws. In other words, you adjust the illuminating path so that it is pr operly aligned along the optical axis of the microscope. 5. Open t he l uminous- field diaphragm by turning t he adjustment whee l counter- clockwise until the ...

  • Leica B523 - page 10

    8 2.9 Polarization contr ast Figure 10 . Analyzer for polarization contra st imaging Figure 11 . Polarizer For polarization contrast microscop y , an an alyzer (a polarizi ng filter) is inserted in the light path b y pus hing it all the wa y in (Fi gure 10), and another polarizing filter (polarizer) is placed labeled side upwards i n a holder below ...

  • Leica B523 - page 11

    9 • Grun : a green panchromatic filter that was used for contrast enchantment for B/W photography These filters were used for fil m photography and are not useful for digital photography. However, when usin g low intensit y light (yellowish), DLF-filter adjusts the light back to the direction of blue, making the image more pleasant to view. Figur ...

  • Leica B523 - page 12

    10 3 Image acquisition 3.1 Accessing the com puter If the computer is turned off, push the power b utton to turn it on. Whe n the lo gin screen appe ars, lo g in using y our hyad lo gin a ccount. C heck that the domain is set to LTDK. 3.2 View Finder applic ation A 10-bit Ol y mpus DP50 color CC D camera is attached on the t op of the microscope. I ...

  • Leica B523 - page 13

    11 simply m ultiplied values, thus the same effect can be added later on in image editing software. Therefore, changing the value from 100 is not recommended. 3.3.3 Exposure Mode In the Exposure Mode box , select either Automatic or Manual : • Automatic ( AE) : The software mea sures the optim al expos ure ti me automatically. Change Spot and Exp ...

  • Leica B523 - page 14

    12 Figure 16 . Histogram level adj ustment. Red square: levels butt on 3.3.6 Focus tool Focusing tool assi sts you i n focusing. If the focusing toolbar is not visible, enable it in the Options menu. Once activated, two bars appear (Figure 17). The bl ue one shows the curre nt focus v alue, and the red one a history of the best level reached since ...

  • Leica B523 - page 15

    13 15). Make sure the stage do es not move at all between the images or image qualit y decreases. 3.4 Capturing the im age Press the Camera button next to the Live button to take an image. After acquisition, the image will appear in Studio Lite. 3.5 Viewing images To zoom in or out, select the zoom tool in the toolbar at the right edge of the scree ...

  • Leica B523 - page 16

    14 3.6.2 Scale bars The Studio L ite software does not provide means to insert scale bars. They can be added later on in an y im age editing software such as Adobe Photoshop. R eady -made scale bars for the full resolution images are avail able at http://www.miu.helsinki.fi/instruments/LeicaDMLB/scale_pictures.htm (in P hotoshop format). The page a ...

  • Leica B523 - page 17

    15 • Cover the microscope with the red hood 6 T roubleshooting 6.1 There is no light • Microscope is not turned on (see Section 2.1) • Light intensity set too low (see 2.1 on page 1) • Do you see any light in t he lamp chamber? If not, the halogen bulb m ay n eed to be changed - contact MIU 6.2 Light is too dark Check: • whether the polar ...

  • Leica B523 - page 18

    16 6.7 Computer login fa ils Check that: • domain is LTDK • use are usi ng y our h yad account user ID and p assword - the same ones you are using for accessing your university e-mail account ...

Produzent Leica Kategorie Microscope & Magnifier

Dokumente, die wir vom Produzenten des Geräts Leica B523 erhalten, können wir in mehrere Gruppen teilen. Unteranderem in:
- technische Zeichnungen Leica
- Bedienungsanleitungen B523
- Produktkarten Leica
- Informationsbroschüren
- oder Energieetiketten Leica B523
Jede von ihnen ist wichtig, jedoch finden wir die wichtigsten Informationen für den Nutzer des Geräts in der Bedienungsanleitung Leica B523.

Die Dokumentengruppe, die als Bedienungsanleitungen bezeichnet wird, wird ebenfalls in detaillierte Arten geteilt, solche wie: Montageanleitungen Leica B523, Wartungsanleitungen, Kurzanleitungen oder Benutzeranleitungen Leica B523. Abhängig vom Bedarf, sollten Sie das Dokument finden, das Sie brauchen. In unserem Service können Sie sich die populärste Bedienungsanleitung des Produkts Leica B523 ansehen.

Ähnliche Bedienungsanleitungen

Die komplette Bedienungsanleitung des Geräts Leica B523, wie sollte sie aussehen?
Die Bedienungsanleitung, auch bezeichnet als Benutzerhandbuch, oder einfach nur „Anleitung”, ist ein technisches Dokument, das dem Benutzer bei der Nutzung von Leica B523 hilfreich sein soll. Die Bedienungsanleitungen werden in der Regel von technischen Schriftstellern geschrieben, aber in einer Sprache, die für alle Nutzer von Leica B523 verständlich ist.

Eine gänzliche Bedienungsanleitung von Leica sollte einige Grundelemente enthalten. Ein Teil von ihnen ist nicht so wichtig, wie z.B.: die Titelseite oder Autorenseiten. Die restlichen von ihnen jedoch, sollten Informationen liefern, die für den Nutzer von enormer Wichtigkeit sind.

1. Einführung und Hinweise, wie man sich in einer Bedienungsanleitung von Leica B523 bewegt - Am Anfang jeder Bedienungsanleitung sollten wir Hinweise bezüglich der Nutzungsart eines bestimmten Ratgebers finden. In ihr sollten sich Informationen über die Lokalisierung des Inhaltsverzeichnisses von Leica B523 befinden, FAQ oder über oft auftretende Probleme – also Stellen, die von den Benutzern in jeder Bedienungsanleitung am meisten gesucht werden
2. Inhaltsverzeichnis - Index aller Ratschläge bezüglich Leica B523, die wir im aktuellen Dokument finden
3. Ratschläge zur Nutzung der Grundfunktionen des Geräts Leica B523 - die uns die ersten Schritte während der Nutzung von Leica B523 erleichtern sollten
4. Troubleshooting - geordneter Tätigkeitslauf, der uns bei der Diagnose und als nächstes bei der Lösung wichtiger Probleme mit Leica B523 hilft
5. FAQ - häufig gestellte Fragen
6. Kontaktdaten Informationen darüber, wo man Kontakt zum Produzenten / Service von Leica B523 im bestimmten Land suchen kann, wenn es nicht gelingt, das Problem selbst zu lösen.

Haben Sie eine Frage bezüglich Leica B523?

Nutzen Sie das untere Formular

Wenn Sie mit Hilfe der gefundenen Bedienungsanleitung Ihr Problem mit Leica B523 nicht gelöst haben, stellen Sie eine Frage, indem Sie das untere Formular nutzen. Wenn einer der Nutzer ein ähnliches Problem mit Leica B523 hatte, ist es möglich, dass er mit Ihnen die Lösung teilen möchte.

Text vom Bild übertragen

Kommentare (0)