Manual de instrucciones Leica DM4000M

69 de páginas 0.87 mb
Descargar

Pasar a la página of 69

Summary
  • Leica DM4000M - page 1

    1 Leica DM4000 B Leica DM4000 M Leica DM5000 B Operating Manual ...

  • Leica DM4000M - page 2

    2 Published 2003 by: Leica Microsystems Wetzlar GmbH Ernst-Leitz-Straße D-35578 Wetzlar (Germany) Responsible for contents: Katja Peter, Karin Schwab Marketing CM, Compound Microscopy, Product Management Phone +49(0)6441-292261 Fax +49(0)6441-292255 In case of questions, please contact: E-mail: MQM-Hotline@leica-microsystems.com ...

  • Leica DM4000M - page 3

    3 Leica DM4000 B Leica DM4000 M Leica DM5000 B Operating Manual ...

  • Leica DM4000M - page 4

    4 Copyrights Copyrights All rights to this documentation are held by Leica Microsystems Wetzlar GmbH. Reproduction of text or illustrations (in whole or in part) by print, photocopy, microfilm or other methods (includ- ing electronic systems) is not allowed without express written permission from Leica Microsystems Wetzlar GmbH. The term "Wind ...

  • Leica DM4000M - page 5

    5 Contents 6. Startup ........................................................ 31 6.1 Functional Principle .................................. 31 6.2 Switching on the Microscope ................ 34 6.3 The Display (Leica DM4000 B/DM4000 M) ................. 35 6.4 The Function Keys .................................... 36 6.5 Köhler Illumination ... ...

  • Leica DM4000M - page 6

    6 Contents 8.2 Fluorescence ............................................. 57 9. Imaging Procedure for Leica DM4000 M ....................................... 58 9.1 Incident Light ............................................. 58 9.1.1 Bright Field ...................................... 58 9.1.2 Dark Field ......................................... 58 ...

  • Leica DM4000M - page 7

    7 1. Important Notes about this Manual (1.2) → p. 20 ! * Numbers in parentheses, such as "(1.2)", corre- spond to illustrations (in the example, Figure 1, Item 2). Numbers with pointer arrows (for example → p.20), point to a certain page of this manual. Special safety instructions are indicated with the triangle symbol shown here, and ...

  • Leica DM4000M - page 8

    8 2. Safety Notes 2. Safety Notes 2.2 Electrical Safety General specifications Leica CTR5000 electronics box (for DM5000 B) For indoor use only. Supply voltage: Frequency: Power input: Fuses: Ambient temperature: Relative humidity: Overvoltage category: Pollution degree: Microscope For indoor use only. Supply voltage: Frequency: Power input: DM4000 ...

  • Leica DM4000M - page 9

    9 2. Safety Notes Caution! Never use any fuses as replacements other than those of the types and the current rat- ings listed here. Using patched fuses or bridging the fuse holder is not permitted. Caution! The microscope’s electrical accessory com- ponents are not protected against water. Water can cause electric shock. Caution! Protect the micr ...

  • Leica DM4000M - page 10

    10 3. Overview of the Instrument Specification Imaging Procedure Transmitted Light Axis Incident Light Axis Z Pinion Objective nosepiece X/Y Stage Tube 3. Overview of the Instrument Leica DM4000 B / DM5000 B • transmitted light: BF, DF, PH, Pol (DM5000 B also ICT) • incident light: fluorescence • integrated into the stand • motorized 5x fil ...

  • Leica DM4000M - page 11

    11 3. Overview of the Instrument Specification Condenser Magnification Changer Control Panels Computer Interface Software Tools CTR5000 Electronics Box Leica DM4000 B / DM5000 B • manual • absolute coded • 1x; 1.25x; 1.6x For Leica DM5000 B only: Separate control unit with power supply for 100W halogen lamp see p. 8 (electrical safety) Leica ...

  • Leica DM4000M - page 12

    12 3. Overview of the Instrument 1 Eyepiece 2 Eyepiece tube 3 Tube 4 Objective nosepiece with objectives 5 Specimen stage with specimen holder 6 Condenser 7 LC display 8 Function keys field diaphragm 9 Transmitted light/incident light switch 10 Function keys aperture diaphragm 11 Function keys: Light intensity 12 Focus dial with coarse and fine adj ...

  • Leica DM4000M - page 13

    13 3. Overview of the Instrument 15 Lamp housing for incident light 16 Lamp housing for transmitted light 17 Transmitted light filter, optional 18 Transmitted light filter, optional 19 Variable function keys (factory pre-assigned) 20 X/Y coaxial drive, height adjustable 21 Focus fine adjustment 22 Motorized filter cube exchanger Fig. 2 Leica DM4000 ...

  • Leica DM4000M - page 14

    14 4. Unpacking the Microscope The device is delivered in two boxes. The stand box contains the following compo- nents: • Stand with integrated incident light axis and objective nosepiece • Specimen stage with stage bracket • Power cable and PC connecting cable • CD with Leica software package • Instructions and list of microscope default ...

  • Leica DM4000M - page 15

    15 4. Unpacking the Microscope Installation location Work with the microscope should be performed in a dust-free room, which is free of oil vapors and other chemical vapors, as well as extreme humidity. At the workplace, large temperature fluctuations, direct sunlight and vibrations should be avoided. These conditions can distort measurements and m ...

  • Leica DM4000M - page 16

    16 5. Assembly 5. Assembling the Microscope T he microscope components are logically as- sembled in this order: • Stage • Condenser • Tube • Eyepieces • Objectives • Light sources • Filter cubes/reflectors* Only a few commonly used screwdrivers and keys are necessary for assembly, which are in- cluded in the delivery package. When usi ...

  • Leica DM4000M - page 17

    17 5. Assembly • From above, set the stage clamp onto the dovetail guide (4.2) and push the stage down- wards until the upper end of the dovetail guide is tightly fastened to the upper end of the stage clamp. • Firmly tighten the stage clamp (4.1). Note: For thicker specimens (Leica DM4000 M) the stage can be set to a correspondingly lower leve ...

  • Leica DM4000M - page 18

    18 5. Assembly 5.2 Condenser • Using the condenser height adjuster (5.4), turn the condenser holder (5.1) completely down- wards. • Unscrew the clamping screw for the con- denser (5.3) far enough so that the condenser can be inserted from the front. • From the front, insert the condenser into the condenser holder as far as it will go. On the ...

  • Leica DM4000M - page 19

    19 5. Assembly 5.3 Tube and Eyepieces The tube is mounted to the stand either directly or with the use of intermediate modules. It is fastened in place with the side clamping screw (9.1). • Loosen the clamping screw (9.1). • Insert the tube in the circular receptacle (dovetail ring). • Retighten the clamping screw (9.1). • Only for the MBDT ...

  • Leica DM4000M - page 20

    20 5. Assembly 5.5 Light Sources for the Transmitted Light Axis Caution: Be sure that the lamp housing is discon- nected from the power supply. Unplug the power plug and the power supply during as- sembly. 107/2 Lamp Housing This lamp housing is used with a 12V 100W halo- gen lamp, which is already mounted. In case the lamp has to be removed: • R ...

  • Leica DM4000M - page 21

    21 5. Assembly 5.6 Light Sources for the Incident Light Axis Caution: During assembly, always unplug the power supply unit of the 106 z lamp housing from its socket. Never touch the glass parts of the burner with bare hands. Never look directly into the beam path (blind- ing hazard). During assembly work on xenon burners, al- ways wear the supplied ...

  • Leica DM4000M - page 22

    22 5. Assembly Fig. 17 Lamp mount with 12 V 100 W halogen lamp • Insert the lamp with the dust cover straight into the socket until it stops. • Remove the dust cover. • Reinsert the lamp mount and retighten the fas- tening screw (16.8). Caution: Do not remove the lamp’s dust cover until af- ter you have installed the lamp. Be certain to avo ...

  • Leica DM4000M - page 23

    23 5. Assembly Inserting the gas discharge lamps (Hg and Xe) into the 106z lamp housing Hg and Xe lamps are powered by the separate ebq 100 supply unit. Read the separate instruction manual provided with this supply unit. The following gas discharge lamps may be used and require different lamp mounts (Fig. 19): Type Typical bulb life* 50 W high-pre ...

  • Leica DM4000M - page 24

    24 5. Assembly Caution: Hg 50 burner: After installation, the labeling must be upright . If a glass melt nipple is present (19a.4), posi- tion it by turning the burner so that the nipple does not come in the way of the beam path later, but instead is positioned sideways. Xe 75 burner: Remove the burner’s dust cover (19b.5) after you have installe ...

  • Leica DM4000M - page 25

    25 5. Assembly Fig. 22 Rear side of the ebq 100 supply unit 1 Lamp connection • Insert the lamp mount, with the burner in- stalled, into the lamp housing and tighten it with the screws (20.8). • Put the lid down again. Plug in the contact plug as far as it goes and retighten the screws. • Place the lamp housing in the incident light lamp hous ...

  • Leica DM4000M - page 26

    26 5. Assembly Fig. 25 Removing the front panel 1 Filter receptacle 2 Retention pin 3 Front panel Fig. 26 Inserting the filter or reflector cubes 1 Mounting 1 2 Fig. 23 Filter cube front side 5.7 Equipping the Incident Light filter turret The receptacles on the turret are numbered. According to your equipment, the individual filter and/or reflector ...

  • Leica DM4000M - page 27

    27 5. Assembly ICT/P transmitted light polarizer • Using the left clamping screw, fasten the ICT/P transmitted light polarizer to the underside of the condenser holder (Fig. 27). • Make sure that the red index point on the front of the polarizer is aligned with 0. • If necessary, insert the compensators ( λ - and λ /4 plates) into the polar ...

  • Leica DM4000M - page 28

    28 5. Assembly • Remove the plug cap on the left side of the stand. • Insert the polarizer into the receptacle until it latches in place (Fig. 30). Motorized analyzer • Insert the analyzer cube as described in sec- tion 5.7 "Equipping the Incident Light filter turret" → p. 26, in the corresponding position on the filter turret. Se ...

  • Leica DM4000M - page 29

    29 5. Assembly Ergomodule For raising the eye level of the tube opening, the ergomodule may be used. It is fastened in place with the side clamping screw. Mirror Housing • Place the mirror housing directly onto the lamp housing receptacle on the back of the stand and attach it using the side clamping screw. • Place the lamp housing onto the mir ...

  • Leica DM4000M - page 30

    30 5. Assembly Fig. 34 Rear side of stand Leica DM4000 B/M 1 Power switch 2 Power supply 1 2 After completing the assembly work, connect the stand to the power supply using the power cable (Fig. 34.2). 5.12 Connection to the CTR5000 Electronics Box Only for the Leica DM5000 B: • Connect the microscope (36.1) to the "Microscope" jack (35 ...

  • Leica DM4000M - page 31

    31 6. Startup 6. Startup 6.1 Functional Principle The microsc ope’s most important functions may be easily accessed using function keys. • The microscope may be switched between various contrast processes by pressing a single button. • The micro scope recognizes the objective chosen and the respective contrast process. There- fore, the values ...

  • Leica DM4000M - page 32

    32 6. Startup Possible Assignments for the Function Keys For Leica DM4000 B/DM5000 B: Function key Meaning BF Bright field (Transmitted light) PH Phase contrast (Transmitted light) ICT Interference contrast (Transmitted light) DF Dark field (Transmitted light) POL Polarization (Transmitted light) CHANGE TL Switch through all transmitted light proce ...

  • Leica DM4000M - page 33

    33 6. Startup For Leica DM4000 M: Function key Meaning BF Bright field (Incident light) ICR Interference contrast (Incident light) DF Dark field (Incident light) POL Polarization (Incident light) CHANGE RL Switch through all incident light processes INT ↑ Increase brightness ( incident light) INT ↓ Reduce brightness (incident light) AP ↑ Open ...

  • Leica DM4000M - page 34

    34 6. Startup Fig. 37 Display after initialization • Switch-on the microscope at the power switch (34.1,36.1). All motorized microscope components first undergo an initialization phase. After initialization is complete, the display on the stand shows the current microscope setting (Fig. 37). The microscopic components such as dia- phragms, conden ...

  • Leica DM4000M - page 35

    35 6. Startup The display shows the current microscope set- tings. The display depends on the microscope’s configuration. In the first column, corresponding pictograms indicate the type of information: con- trast method, magnification, light intensity, dia- phragms, light splitting for photo tubes. Please see the abbreviation index for a list of ...

  • Leica DM4000M - page 36

    36 6. Startup Variable function keys: A factory preset is performed which fits your mi croscope configuration. The function keys are labeled accordingly, and a separate description of the key occupation accompanies the microscope (“Identification Sheet”). Abbreations are listed on p.32f. 6.5 Köhler Illumination For each objective, optimal valu ...

  • Leica DM4000M - page 37

    37 6. Startup However, depending on how the condenser is disassembled and reassembled, it may be nec- essary to re-adjust the condenser in some cases. Therefore, check the condenser centering. The following procedure is provided for the transmitted light-bright field illumination. • Select an objective with moderate magnification (10x-40x). • A ...

  • Leica DM4000M - page 38

    38 6. Startup • Open the field diaphragm just enough for it to disappear from the field of view (41d). Caution: Do not adjust the aperture diaphragm. The aper- ture diaphragm is already set optimally for each objective. 6.6. Checking Phase Contrast Rings A B CD Fig. 41 Köhler Illumination a Field diaphragm not focused, not centered b Field diaph ...

  • Leica DM4000M - page 39

    39 6. Startup If your microscope is equipped for the use of phase contrast, the light rings that fit the objec- tives are built into the condenser. The light rings are already leveled in the factory. However, the leveling should be rechecked. Note: Every objective is assigned its own light ring in the condenser disc. Therefore, a check must be perf ...

  • Leica DM4000M - page 40

    40 6. Startup • Insert the centering key through the corre- sponding openings (44.1) in the condenser holder. • Turn the centering screws until the dark ring (phase ring in the objective) is congruent with the slightly narrower bright ring (light ring in condenser) (43 c). • Repeat the process for all other phase con- trast objectives. • Re ...

  • Leica DM4000M - page 41

    41 6. Startup The 107/2 lamp housing with 12 V 100 W halogen lamp has a defined presetting. The lamp need not to be centered. Incident light axis (IL) with 106 z lamp housing • When a supply unit is used, it is turned on first. • Activate the incident light axis using the TL/IL function key. FLUO (Leica DM4000 B/ DM5000 B) or IL (Leica DM4000 M ...

  • Leica DM4000M - page 42

    42 6. Startup • In the adjustment window, you see the direct filament image and the mirror image, which in most cases are shifted together. • Focus the direct filament image with the col- lector (46.6). • Use the adjusting buttons on the rear side of the lamp housing (46.2, 46.4) to pivot the lamp filament’s mirror image to the side or com- ...

  • Leica DM4000M - page 43

    43 6. Startup • Focus the direct image with the collector (46.6). • Use the adjusting buttons on the rear side of the lamp housing (46.2,46.4) to pivot the arc’s mirror image to the side or completely out of the beam path. The lamp filament’s focused image remains visible (Fig. 50). • Use the adjusting buttons (46.1) and (46.5) to place t ...

  • Leica DM4000M - page 44

    44 6. Startup Fig. 55 Direct arc image and mirror image in target position (in reality, the image is less focused) • Focus the direct image with the collector (46.6). • Use the adjusting buttons to pivot the arc’s mirror image on the rear side of the lamp housing (46.2,46.4) to the side or completely out of the beam path. The arc’s focused ...

  • Leica DM4000M - page 45

    45 6. Startup In older lamps, the structure of the arc is no longer clearly recognizable. The image is then more like that of a HG 50 lamp. The im- age and mirror image can no longer be su- perimposed exactly. In this case, align both images. • Using the collector, defocus the image with the knob (46.6) until the arc image and mirror image are no ...

  • Leica DM4000M - page 46

    46 7. Operation Fig. 58 Revolving object stage 1 Object motion (Y direction) 2 Object motion (X direction) 3 Torque adjustment (Y direction) 4 Torque adjustment (X direction) 5 Focus dial for fine focusing 7. Operation 7.1 Switching on the Microscope When using a gas discharge lamp, the ebq 100 external supply unit must be turned on separately (56. ...

  • Leica DM4000M - page 47

    47 7. Operation Rotating the Stage The swiveling range of the rotating stages is 0°- 110°. • In order to revolve the stage, loosen the fas- tening screw (59.1). • Bring the table into the desired position. • Retighten the fastening screw. Fig. 59 Revolving object stage 1 Clamping screw 2 Fine focusing 3 Coarse focusing 1 7.3 Focusing There ...

  • Leica DM4000M - page 48

    48 7. Operation 7.4 Tubes Note: Close any unused tube openings, as otherwise stray light can interfere with observation . Note: Make sure that the connector cable is plugged in on the MBDT25+ motorized tube (60.1). Adjusting the Viewing Distance • Adjust the viewing distance of the eye- pieces so that a congruent total image is seen (Fig. 60). Fi ...

  • Leica DM4000M - page 49

    49 7. Operation Beam Splitting in Photo Tubes EDT22 tube: The beam splitting between the observation and documentation outputs has a definite presetting (50:50). BDT25+ tube: The beam splitting is set manually by pulling out a control bar. Control Bar Observation Photo VI S 100 % 0 % 50/50 1 50 % 50 % PHOTO 11 0 % 100 % MBDT25+ tube: This tube is s ...

  • Leica DM4000M - page 50

    50 7. Operation 7.6 Objectives The objective must be moved manually into the light path. Be sure that the nosepiece turret locks into place. The objective’s position in the turret is factory- set and must be adhered to while screwing in the objectives (see Objective Assembly → p. 19) When you rotate the objective into position, the microscope a ...

  • Leica DM4000M - page 51

    51 7. Operation 7.7 Magnification Changer Optionally, a coded magnification changer can be used, which is manually operated. On the knurled ring, the following magnification factors can be set: B Stand M Stand 1x 1x 1.25x 1.5x 1.6x 2x The selected factor is indicated in the display and included in the total magnification. Fig. 63 Immersion objectiv ...

  • Leica DM4000M - page 52

    52 7. Operation 7.8 Light Sources • The brightness is set using the function keys (65.5). Then, the INT function keys are as- signed to the currently active axis for trans- mitted light (TL) or incident light (IL). • For TL and IL: Settings can be made either in large or small increments. Pushing both INT buttons simultaneously switches between ...

  • Leica DM4000M - page 53

    53 8. Imaging Procedure for Leica DM4000 B/DM5000 B 8.1 Transmitted Light 8.1.1 Bright Field (TL) • Switch to the transmitted light axis (TL) by pushing the TL/IL button. • Select the BF (bright field) contrast method. Do so by pressing the BF variable key. Alternatively: Press the CHANGE TL variable key. (For key occupation please see “Ident ...

  • Leica DM4000M - page 54

    54 8. Imaging Procedure for Leica DM4000 B/DM5000 B Notes: • The microscope automatically selects the correct light ring in the condenser. • When selecting the phase contrast method, the aperture diaphragm is opened completely and may not be adjusted. To avoid errors in operation, the function keys for setting the ap- erture diaphragm (AP) are ...

  • Leica DM4000M - page 55

    55 8. Imaging Procedure for Leica DM4000 B/DM5000 B 8.1.4 Polarization (TL) • Switch to the transmitted light axis (TL) by pushing the TL/IL button. • Select the POL (polarization) contrast method. Do so by pressing the POL variable key. Alternatively: Press the CHANGE TL variable key. (For key occupation please see “Identification Sheet”.) ...

  • Leica DM4000M - page 56

    56 8. Imaging Procedure for Leica DM4000 B/DM5000 B 8.1.5 Differential Interference Contrast (TL) (only for DM5000 B) • Switch to the transmitted light axis (TL) by pushing the TL/IL button. • Insert a specimen and rotate a suitable objec- tive into place. • Select the DIC contrast method. Do so by pressing the DIC variable key. Alternatively ...

  • Leica DM4000M - page 57

    57 8. Imaging Procedure for Leica DM4000 B/DM5000 B 8.2 Fluorescence • Switch to the fluorescent light axis (FLUO) by pushing the TL/IL button. • Insert a specimen and rotate a suitable objec- tive into place. • The current fluorescence cube is indicated on the display. • Closing the incident light shutter protects your specimen from fading ...

  • Leica DM4000M - page 58

    58 9. Imaging Procedure for Leica DM4000 M 9.1 Incident Light 9.1.1 Bright Field • Switch to the incident light axis (IL) by pushing the TL/IL button. • Select the BF (bright field) contrast method. Do so by pressing the BF variable key. Alternatively: Press the CHANGE RL variable key. (For key occupation please see “Identification Sheet”.) ...

  • Leica DM4000M - page 59

    59 9. Imaging Procedure for Leica DM4000 M 9.1.3 Polarization • Switch to the incident light axis (IL) by pushing the TL/IL button. • Select the POL (polarization) contrast method. Do so by pressing the POL variable key. Alternatively: Press the CHANGE RL variable key. (For key occupation please see “Identification Sheet”.) The display indi ...

  • Leica DM4000M - page 60

    60 9. Imaging Procedure for Leica DM4000 M 9.1.4 Interference Contrast • Switch to the incident light axis (IL) by pushing the TL/IL button. • Insert a specimen and rotate a suitable objec- tive into place. • Select the DIC contrast method. Do so by pressing the DIC variable key. Alternatively: Press the CHANGE RL variable key. (For key occup ...

  • Leica DM4000M - page 61

    61 10. Trouble Shooting 10. Trouble Shooting Problem Stand The microscope does not respond. Illumination The image is completely dark. The image is unevenly or not uniformly illumi- nated. The illumination "flickers." The lamp does not illuminate immediately upon being switched on. Cause/Remedy  Make sure that voltage is impressed.  ...

  • Leica DM4000M - page 62

    62 10. Trouble Shooting Problem Bright Field The specimen can not be brought into focus. Dark Field No definite DF contrast is possible. The image is unevenly or not uniformly illumi- nated. Undesirable stray light Phase contrast No phase contrast is possible. Cause/Remedy  Use the correct immersion medium.  Lay the specimen with the cover gl ...

  • Leica DM4000M - page 63

    63 10. Trouble Shooting Problem Polarization No polarization contrast is possible. Fluorescence The image is completely dark (no fluorescence). The fluorescence is too weak. Display The display flashes. FAIL! appears. Cause/Remedy  Bring the polarizer and analyzer into cross po- sition until they reach maximum darkness (without specimen) ( → p ...

  • Leica DM4000M - page 64

    64 11. Care of the Microscope Caution! Unplug the power supply before performing cleaning and maintenance work! Protect electrical components from moisture! Microscopes in warm and warm-damp climatic zones require special care in order to prevent fungus contamination. The microscope should be cleaned after each use, and the microscope optics should ...

  • Leica DM4000M - page 65

    65 11. Care of the Microscope Cleaning Glass Surfaces Remove dust on glass surfaces with a fine, dry and fat-free hair brush, by blowing with a blow bag or vacuum suction. Carefully remove stubborn dirt on glass surfaces with a clean cloth moistened with distilled water. If the dirt still can not be removed, use pure alcohol, chloroform or benzine. ...

  • Leica DM4000M - page 66

    66 12. Wear and Spare Parts Order No. Material No. Name Used for Replacement Lamp 500 974 Halogen lamp 12 V 100 W 107/2 lamp housing 500 137 High-pressure mercury burner 50 W 106 z lamp housing 500 138 High-pressure mercury burner 100 W 106 z lamp housing 500 321 High-pressure mercury burner 100 W 106 z lamp housing (103 W/2) 500 139 High-pressure ...

  • Leica DM4000M - page 67

    67 13. Abbreviations and Pictograms 13. Abbreviations and Pictograms Contrast method Magnification Light intensity/diaphragms Beam splitting Transmitted light shutter open Transmitted light shutter closed Incident light shutter open Incident light shutter closed ↓ ↓ ↑ ↑ +   AET Advanced Ergo Tube AP Aperture diaphragm BF Brifght field ...

  • Leica DM4000M - page 68

    68 14. Index A djusting the light sources 41 Allowable ambient conditions 15 Ambient temperature 8, 9 Ambient conditions 15 Analyzer 28, 55, 56, 59 Analyzer cube 55, 56 Aperture diaphragm 12, 35, 38, 52 B eam splitting 49 Bright field 53, 58 Booster Lens 29, 5 7 C leaning 64 Cleaning objectives 65 Coaxial pinion 46 Condenser connector 18 Condenser ...

  • Leica DM4000M - page 69

    69 15. EU Declaration of Conformity 15. EU Declaration of Conformity ...

Fabricante Leica Categoría Microscope & Magnifier

Los documentos del dispositivo Leica DM4000M que obtenemos del fabricante se pueden dividir en varios grupos. Entre ellos están::
- dibujos técnicos Leica
- manuales de instrucciones DM4000M
- hojas de producto Leica
- folletos informativos
- o etiquetas energéticas Leica DM4000M
Todos son importantes, pero lo más importante desde el punto de vista del usuario de un dispositivo lo encontraremos en el manual de instrucciones Leica DM4000M.

Un conjunto de documentos determinado como manual de instrucciones se divide también en tipos más detallados, tales como: instrucciones de montaje Leica DM4000M, instrucciones de servicio, instrucciones cortas o instrucciones de usuario Leica DM4000M. Dependiendo de la situación debes buscar el documento que necesitas. En nuestra página puedes consultar el manual de instrucciones más popular del producto Leica DM4000M.

Manuales de instrucciones parecidos

El manual de instrucciones del dispositivo Leica DM4000M, ¿cómo debería ser?
El manual de instrucciones, también determinado como el manual de usuario o simplemente instrucciones, es un documento técnico que tiene como objetivo ayudar a los usuarios a utilizar Leica DM4000M. Las instrucciones normalmente las realiza un escritor técnico pero en un lenguaje comprensible para todos los usuarios de Leica DM4000M.

El manual de instrucciones completo de Leica debe contener unos elementos básicos. Una parte de ellos es menos importante, como por ejemplo: la portada / la página principal o las páginas de autor. Sin embargo, lo demás debe facilitarnos la información más importante desde el punto de vista del usuario.

1. Introducción y pistas sobre cómo orientarse con el manual Leica DM4000M - Al principio de cada manual se deben encontrar las indicaciones acerca de la manera de usar un manual de instrucciones. Debe contener información sobre dónde encontrar el índice Leica DM4000M, preguntas frecuentes o problemas más comunes – son los apartados más buscados por los usuarios de cada manual de instrucciones.
2. Índice - listados de todos los consejos acerca de Leica DM4000M que encontraremos en el documento
3. Consejos de uso de las funciones básicas del dispositivo Leica DM4000M - que deben facilitarnos los primeros pasos durante el uso de Leica DM4000M
4. Troubleshooting - una secuencia sistematizada de acciones que nos ayudará a diagnosticar los problemas más importantes con Leica DM4000M
5. FAQ - las preguntas frecuentes
6. Datos de contacto Información acerca de cómo encontrar los datos de contacto del fabricante / servicio de Leica DM4000M en cada país si no somos capaces de solucionar el problema por nuestra cuenta.

¿Tienes una pregunta acerca de Leica DM4000M?

Utiliza el formulario que se encuentra abajo

Si con la ayuda del manual de instrucciones encontrado no has podido solucionar tu problema con Leica DM4000M, haz una pregunta utilizando el formulario que se encuentra abajo. Si alguno de los usuarios ha tenido un problema parecido con Leica DM4000M probablemente compartirá una forma de solucionarlo.

Copia el texto de la imagen

Comentarios (0)