    Page 1 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 Leica HER2 FISH System - 30 T est Instructions For Use For use on Leica Biosystems’ BOND-MAX and BOND-III System. T A9217 is a uorescence in situ hybridization product designed to stain 30 tests (30 slides stained with LS ...

  • Leica TA9217 - page 2

    English Page 2 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 Contents Intended Use ............................................................................................................................................................................... 3 For in vitro diagn ...

  • Leica TA9217 - page 3

    Page 3 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 English Intended Use For in vitro diagnostic use The Leica HER2 FISH System - 30 T est is designed to detect amplication of the HER2/neu gene via uorescence in situ hybridization (FISH) in formalin-xed, parafn-embe ...

  • Leica TA9217 - page 4

    English Page 4 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 The characteristic morphology of the tissue and the spatial distribution of oncogene copies in individual uncultured primary breast carcinomas are retained. Aberrations in chromosome 17 copy number (aneusomy) are also ...

  • Leica TA9217 - page 5

    Page 5 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 English The Leica HER2 FISH System - 30 T est is for use only on the Leica BOND-MAX and BOND-III System. Components Provided The materials listed below (T able 1) are sufcient to stain 30 tests (30 slides stained with LSI H ...

  • Leica TA9217 - page 6

    English Page 6 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 T issue sections mounted on charged slides (Leica BOND Plus Slides S21.21 13) can be held for up to 12 months at 2–8 °C before staining. Following sectioning, it is recommended that slides are incubated at 60 °C fo ...

  • Leica TA9217 - page 7

    Page 7 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 English bandpass uorescence microscope lter sets optimized for use with the Leica HER2 FISH System - 30 T est are available for most microscope models. The recommended lter sets for the Leica HER2 FISH System - 30 T e ...

  • Leica TA9217 - page 8

    English Page 8 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 5. T urn on the PC attached to the Leica BOND-MAX and BOND-III System. 6. Open the Leica BOND software. 7. For a new Leica HER2 FISH System - 30 T est kit, scan the reagent tray barcode with the handheld scanner to ent ...

  • Leica TA9217 - page 9

    Page 9 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 English Signal Assessment and Enumeration T o assess the signal quality and enumerate the HER2 and CEP17 signals follow the process below: 1. Assessment of Slide Quality Evaluate slide quality using the following criteria: • ...

  • Leica TA9217 - page 10

    English Page 10 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 Recommended Method for LSI HER2 to CEP17 Ratio Determination T o determine the LSI HER2 to CEP17 ratio, use the following method: 1. Record and determine the number of LSI HER2 and CEP17 signals in 20 nuclei (see Figu ...

  • Leica TA9217 - page 11

    Page 1 1 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 English Figure 1: Interpretation Guide Leica HER2 FISH S ystem - 30 T est Interpretation Guide Count as 2 orange signals and 1 green signal Do not count. Nuclei with no signals or with signals of only 1 color should not be s ...

  • Leica TA9217 - page 12

    English Page 12 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 Leica HER2 FISH S ystem - 30 T est Score Sheet 20 Nuclei Signal Count Nucleus # HER2 Copy Number CEP17 Copy Number Nucleus # HER2 Copy Number CEP17 Copy Number 1 11 2 12 3 13 4 14 5 15 6 16 7 17 8 18 9 19 10 20 T otal ...

  • Leica TA9217 - page 13

    Page 13 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 English Quality Control Use of Control Slides It is recommended that a Leica HER2 FISH Control Slide is included in each test run to monitor assay performance and to assess the accuracy of signal enumeration. Control slides s ...

  • Leica TA9217 - page 14

    English Page 14 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 Unexpected FISH staining, or variations in the staining, may be a result of alterations in the expression levels of the encoding genes. Any change in expected staining patterns should be interpreted in association wit ...

  • Leica TA9217 - page 15

    Page 15 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 English 2x2 Concordance Results Leica BOND-MAX System Data was grouped as negative (<2.00) or positive (≥2.00) for a 2x2 analysis The observed agreement for 300 samples between the two tests in a 2x2 analysis show a conc ...

  • Leica TA9217 - page 16

    English Page 16 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 2x2 Concordance Results Leica BOND-III System Data was grouped as negative (<2.00) or positive (≥2.00) for a 2x2 analysis The observed agreement for 300 samples between the two tests in a 2x2 analysis show a conc ...

  • Leica TA9217 - page 17

    Page 17 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 English Precision T esting – Leica BOND-MAX System A. Within Run Precision Study The within run precision study was performed in a randomized and blinded fashion. Within run precision testing of the Leica HER2 FISH System - ...

  • Leica TA9217 - page 18

    English Page 18 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 F . Lot-to-Lot Precision Study The lot-to-lot precision study was performed in a randomized and blinded fashion. Lot-to-Lot precision was determined on three independently manufactured lots of the Leica HER2 FISH Syst ...

  • Leica TA9217 - page 19

    Page 19 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 English K. Between Observer Precision Study The between observer precision study was performed in a randomized and blinded fashion. Between observer reproducibility testing of the Leica HER2 FISH System - 30 T est was evaluat ...

  • Leica TA9217 - page 20

    English Page 20 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 denaturation time is recommended. • No difference in amplication status was observed when the hybridization time was performed for 9 hours and 15 hours when compared to the default Leica HER2 FISH System - 30 T e ...

  • Leica TA9217 - page 21

    Page 21 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 English T roubleshooting Problem Probable C ause Remedial Action No or weak uorescent signal/staining Inappropriate xation or processing of test specimen Ensure a formalin-based xativ e is used and that processing sc ...

  • Leica TA9217 - page 22

    English Page 22 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 Problem Probable C ause Remedial Action Nonspecic background uorescent signal/staining Inadequate post hybridization washing Increase post hybridization wash incubation time. Inappropriate bulk reagents dispense ...

  • Leica TA9217 - page 23

    Page 23 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 English References 1. Bargmann CI, Hung MC, Weinberg RA. The neu oncogene encodes an epidermal growth factor related protein. Nature 1986;319:226–30. 2. Lonardo F , Di Marco E, King CR, Pierce JH, Segatto O, Aaronson SA, et ...

  • Leica TA9217 - page 24

    Page 24 of 24 English Leica Biosystems Leica HER2 FISH System - 30 T est Instructions for Use T A9217 EN-CE-Rev_D 08/04/2013 License Agreement This product contains PathV ysion FISH probes supplied by Abbott Molecular Inc. PathV ysion, LSI and CEP are a trademark of Abbott Molecular Inc. All Rights Reserved. Used under License. Amendments to Previo ...

Fabbricante Leica Categoria Oxygen Equipment

I documenti che otteniamo dal fabbricante del dispositivoLeica TA9217 possono essere suddivisi in diversi gruppi. Questi sono tra gli altri:
- disegni tecnici Leica
- manuali per l’uso TA9217
- schede prodotto Leica
- opuscoli
- o etichette energetiche Leica TA9217
Tutti sono importanti, ma le informazioni più importanti dal punto di vista di utilizzo del dispositivo possono essere trovate nel manuale per l’uso Leica TA9217.

Un gruppo di documenti denominato manuali per l’uso, è anche suddiviso in tipi più specifici, come: Manuali di installazione Leica TA9217, manuali di manutenzione, brevi manuali o manuali utente Leica TA9217. A seconda delle esigenze, si dovrebbe cercare il documento necessario. Nel nostro sito web, puoi visualizzare il manuale più popolare per l'uso del prodotto Leica TA9217.

Manuali per l’uso simili

Manuale completo del dispositivo Leica TA9217, come deve essere?
Il manuale per l’uso, denominato anche istruzioni per l’uso, o semplicemente il manuale è un documento tecnico progettato per aiutare ad utilizzare Leica TA9217 dagli utenti. I manuali sono di solito scritti da uno scrittore tecnico, ma in un linguaggio accessibile a tutti gli utenti Leica TA9217.

Il manuale per l’uso completo Leica, dovrebbe includere vari elementi di base. Alcuni di loro sono meno importanti, come ad esempio: copertina / pagina del titolo o pagina di autore. Tuttavia, la parte rimanente, dovrebbe fornire informazioni importanti dal punto di vista dell'utente.

1. Introduzione e la guida su come utilizzare il manuale Leica TA9217 - All'inizio di ogni manuale, dovremmo trovare indizi su come utilizzare il documento. Dovrebbe contenere informazioni sulla posizione dell’indice Leica TA9217, FAQ o i problemi più comuni : i punti che sono più spesso cercati dagli utenti di ogni manuale
2. Indice - elenco di tutti i suggerimenti per Leica TA9217 che possono essere trovati nel documento corrente
3. Suggerimenti su come utilizzare le funzioni di base del dispositivo Leica TA9217 - che ci dovrebbe aiutare nei primi passi durante l'uso Leica TA9217
4. Troubleshooting - sequenza sistematica delle attività che vi aiuterà a diagnosticare e, successivamente, risolvere i problemi più importanti del Leica TA9217
5. FAQ - domande più frequenti
6. Dettagli di contatto Informazioni su dove cercare il contatto con il fabbricante / servizio Leica TA9217 in un dato paese, se il problema non può essere risolto da noi stessi.

Hai una domanda su Leica TA9217?

Utilizza il modulo sottostante

Se non hai risolto il problema con Leica TA9217 con l'aiuto del manuale che hai trovato, fai una domanda utilizzando il modulo sottostante. Se un utente ha avuto un problema simile con Leica TA9217 è probabile che voglia condividere il modo di risolverlo.

Riscrivi il testo dall’immagine

Commenti (0)