Manuale per l’uso Agilent Technologies E6380-90073

352 pagine 3.24 mb
Scarica

Vai alla pagina of 352

Summary
  • Agilent Technologies E6380-90073 - page 1

    Agilent T echnologies 89 35 Seri es E6380A CDMA Cellular/PCS Base Station T est Set GPIB Syntax Reference Guide Firmware V ersion: B.02.00 and above Agilent Part Number E6380-90073 Revision E Printed in UK September 2000 ...

  • Agilent Technologies E6380-90073 - page 2

    2 Notice Information co ntained in t his document is subject to change witho ut notice . All Rights Reserved. Reproduct ion, adaptation, or tr anslation without prior wri tten permission is p rohibited, exce pt as allowed under the cop yr igh t la w s. This material may be repro duced by or f or the U .S. Government pursuant to the Copyright Lice n ...

  • Agilent Technologies E6380-90073 - page 3

    Contents 3 Manufacturer' s Declaration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 2 Safety Consi derations for this Instr ument . . . . . . . . . . . . . . . . . . . . . ...

  • Agilent Technologies E6380-90073 - page 4

    4 Contents :NOTCh:FREQuency? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61 :RANGing <stri ng> :RANGing? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62 :SMPoint <string> :SMPoint? . . . . . . . . . . . . . . . . . . . . . . . . . . ...

  • Agilent Technologies E6380-90073 - page 5

    Contents 5 :POLarit y? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73 :SEND:MODE <string > :SEND:MODE? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73 :SEND . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...

  • Agilent Technologies E6380-90073 - page 6

    6 Contents :DTMF:OFF Time <real number> :DTMF:OFF Time? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84 :DTMF:ON Time <real number> :DTMF:ON Time? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85 :DTMF:SEQuence <str ing> :DTMF:SEQuen ...

  • Agilent Technologies E6380-90073 - page 7

    Contents 7 :NAMPs|NTACs:FVC:MESS < stri ng> :NAMPs|NTACs:FVC:ME SS? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95 :NAMPs|NTACs:FVC:RATE <r eal number> :NAMPs|NTACs:FVC:RATE? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96 :NAMPs|NTACs:FVC:AM <real number > :NAMPs|NTAC ...

  • Agilent Technologies E6380-90073 - page 8

    8 Contents :PNINcrement <real number> :PNINcrement? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107 :PNMode <string> :PNMode? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108 :PNOF fset <real number> :PNOFfset? . . . ...

  • Agilent Technologies E6380-90073 - page 9

    Contents 9 :EVENt:QUALifier <str ing> :EVENt:QUALifier ? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119 :EVENt:TRIGger <st ring> :EVENt:TRIGger? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120 :EVENt:TRIGger:D ELay <real number> :EVE ...

  • Agilent Technologies E6380-90073 - page 10

    10 Contents :THRShld <real number> :THRShld? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130 :IS2000:C ONTrols < string> :IS2000:CON Trols? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 131 :IS2000:G AI N:MOD E <strin g> :IS2000 ...

  • Agilent Technologies E6380-90073 - page 11

    Contents 11 :IS2000:P O Wer: UN IT <s tr ing> :IS2000:P OWer:UNIT? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141 :IS2000:THReshold <real numb er> :IS2000:T HReshold? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 142 :IS2000:T RIGger:ARM . . . . . . . ...

  • Agilent Technologies E6380-90073 - page 12

    12 Contents :NOTChmode <string > :NOTChmode? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 153 :OFLevel:M ODE <string> :OFLevel:M ODE? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 153 :OFLevel:ANTenna <real number> :OFLevel:ANT enn ...

  • Agilent Technologies E6380-90073 - page 13

    Contents 13 :RFIMped <s tr ing> :RFIMped? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 164 :SPOR9|SP ORT9|SB 9|SP9:BAU D <stri ng> :SPOR9|SP ORT9|SB9|SP9:BAU D? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 165 :SPOR9|SP ORT9|SB 9|SP9:PAR ity <string > ...

  • Agilent Technologies E6380-90073 - page 14

    14 Contents :SPOR9|SPORT9| SB9|SP9:MODem:SECurity: PASSword :SPOR9|SP ORT9|SB9|SP9:MO Dem:SECurity:PASSwor d? . . . . . . . . . . . . 174 :SPOR9|SP ORT9|SB 9|SP9:SIN |SINP <stri ng> :SPOR9|SP ORT9|SB9|SP9:SIN |SINP? . . . . . . . . . . . . . . . . . . . . . . . . . . . 174 :SPOR9|SP ORT9|SB9|SP9:STATus:LI NE? . . . . . . . . . . . . . . . . . ...

  • Agilent Technologies E6380-90073 - page 15

    Contents 15 :SPOR10| SPORT10| SB10|SP10:M ODem:SEC urity:STA Te :SPOR10|SPORT1 0|SB10|SP1 0:MODem:SECurity:ST ATe? . . . . . . . . . . . . 182 :SPOR10| SPORT10| SB10|SP10:M ODem:SEC urity:PAS Sword :SPOR10|SPORT1 0|SB10|SP1 0:MODem:SECurity:P ASSword? . . . . . . . . 183 :SPOR11| SPORT11| SB11|SP11:B AUD <str ing> :SPOR11|SPORT1 1|SB11|SP1 1: ...

  • Agilent Technologies E6380-90073 - page 16

    16 Contents :SPOR11| SPORT11| SB11|SP11:M ODem:SEC urity:PAS Sword :SPOR11|SPORT1 1|SB11|SP1 1:MODem:SECurity:P ASSword? . . . . . . . . 192 :SPSelect <string> :SPSelect? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 192 :SRLocation <string > :SRLocation? . . . . . . . . . . . ...

  • Agilent Technologies E6380-90073 - page 17

    Contents 17 :NAMPs| NTACs:STANdard <st ring> :NAMPs|NTACs:STANda rd? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204 :NAMPs|NTACs:TRIGger:PATTern < string> :NAMPs|NTACs:TRIG ger:PATTern? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 04 DISPlay subsystem . . . . . . . . . . . . . . . . . ...

  • Agilent Technologies E6380-90073 - page 18

    18 Contents :AFRequency:FREQuency <meas cmnd> :AFRequency:FREQuency ? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 215 :AFRequency:SELect <meas cmnd> :AFRequency:SELe ct? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 216 :AFRequency:SINAD <meas cmnd> :AFReque ...

  • Agilent Technologies E6380-90073 - page 19

    Contents 19 :CDANalyzer: ADC <meas cmnd> :CDANalyzer: ADC? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 226 :CDANalyzer:CARRier:FE EDthrough <meas cmnd> :CDANalyzer:CARRier:FEEDthrough? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 226 :CDANalyzer: CPOWer <meas cmnd> :CD ...

  • Agilent Technologies E6380-90073 - page 20

    20 Contents :CDANalyzer: IS2000:PN umber:OFFSet? :CDANalyzer: IS2000:PN UMber:OFFSet? . . . . . . . . . . . . . . . . . . . . . . . . . . . 240 :CDANalyzer: IS2000:RHO [:ESTimated ]? . . . . . . . . . . . . . . . . . . . . . . . . . . . 241 :CDANalyzer: IS2000:TIME:OF FSet? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 241 :CDANalyz ...

  • Agilent Technologies E6380-90073 - page 21

    Contents 21 :CDANalyzer:IS2000:TRACe:POWer:NOISe[:VALue]? . . . . . . . . . . . . . . . . . 257 :CDANalyzer:IS2000:TRACe:POWer:POWer: ACTive? . . . . . . . . . . . . . . . . . . 258 :CDANalyzer:IS2000:TRACe:POWer:POWer: ALL? . . . . . . . . . . . . . . . . . . . . 258 :CDANalyzer:IS2000:TRACe:POWer:POWer[:VALue]? . . . . . . . . . . . . . . . . . 2 ...

  • Agilent Technologies E6380-90073 - page 22

    22 Contents :DEC:NAMP:DTMF:HI GH:DISPlay <stri ng> :DEC:NAMP:DTM F:HIGH:DISPlay? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 269 :DEC:NAMP:DTM F:TIME:ON <mea s cmnd> :DEC:NAMP:DTM F:TIME:ON? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 270 :DEC:NA MP:DTMF :TIME:OF F <meas cmnd> :DEC:NAMP: ...

  • Agilent Technologies E6380-90073 - page 23

    Contents 23 :MARKer:N PEak . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 285 :MARKer:P PEak . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 286 :MARKer:POSition <real number> :MARKer:P OSition? . . . . . . . . . . . . . . . . . . . . . . . . . . ...

  • Agilent Technologies E6380-90073 - page 24

    24 Contents [:SELected]:WAIT [:SELected]:WAIT ? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 297 RAM Usage Infor mation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 298 SPEC:RAMDISKALLOC? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...

  • Agilent Technologies E6380-90073 - page 25

    Contents 25 :CHANnel <string> :CHANnel? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 309 :FM:CO UPling < string> :FM:COUPling? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 309 :FM:DCZero . . . . . . . . . . . . . . . . . . . ...

  • Agilent Technologies E6380-90073 - page 26

    26 Contents :MARKer[:N ORMal]:NPLe vel <real number > :MARKer[:N ORMal]:NPLe vel? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 319 :MARKer[:N ORMal]:PEAK . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 319 :MARKer[:N ORMal]:POSit ion <real number > :MARKer[:N ORMal]:POSit ion? . . ...

  • Agilent Technologies E6380-90073 - page 27

    Contents 27 :MASK:TYPE :MASK:TYPE? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 329 :RFGenerator <string> :RFGenerator? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 330 :RLEVel <real number> :RLEVel? . . . . . . . . . . . . . . ...

  • Agilent Technologies E6380-90073 - page 28

    28 Contents :HARD1[:EVENt?]:N TRansition :HARD1[:EVENt?]:NTRansition? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 338 :HARD1[:EVENt?]:PTRans ition :HARD1[:EVENt?]:PTRansition? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 339 :HARD2:CONDition :HARD2:CONDition? . . . . . . . . . . . . . . . . . . . . . . ...

  • Agilent Technologies E6380-90073 - page 29

    Contents 29 :CDMA1:ENABle :CDMA1:ENABle? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 343 :CDMA1[:EVENt?] :NTRansition :CDMA1[:EVENt?]:NT Ransition? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 343 :CDMA1[:EVENt?] :PTRansition :CDMA1[:EVENt?]:PTRa nsition? . . . . . . . . . . . ...

  • Agilent Technologies E6380-90073 - page 30

    30 ...

  • Agilent Technologies E6380-90073 - page 31

    31 Regulato ry Infor mation Manuf acture r's D eclarat ion This statement is p rovided to comply with th e requirements of the German Sound Emission Directive, from 18 January 1991. This product has the following sound pressure emission specification: • sound pressure Lp <70 dB(A) • at t he operator position • under normal operat ion ...

  • Agilent Technologies E6380-90073 - page 32

    32 Safety GENERAL This product and related do cumentation must be reviewed for familiarization with sa fet y markings and instructions before operation. This product has been designed and tested in acc ordance with IEC Publicati on 1010, " Safety Requireme nts for Electroni c Measuring Apparatus," and has been supplied in a saf e conditio ...

  • Agilent Technologies E6380-90073 - page 33

    33 Safety Consider ations for this Instrument SAFETY EARTH GROUND A uninterruptibl e safety earth ground must be provided from the main power so urce to t he product input wiring terminals, power cord, or supplied power cord set. CHASSIS GROUND TERMIN AL T o prevent a potent ial shock hazard, always connect the rear -panel chassis ground ter minal ...

  • Agilent Technologies E6380-90073 - page 34

    34 W ARNING Whenever it is likely that th e protection has been impaire d, the instrument must be made inoperative and be secured against any unintended operation. If this instrument is to be energized via an autotransformer (for voltage reduction), make sure the common terminal is connected to the ear th terminal of the power source. If this produ ...

  • Agilent Technologies E6380-90073 - page 35

    35 CA UTION Always use the three-p rong ac powe r cord supplied with this product. Failure to ensure ad equat e earth grounding by not using th is c ord may cause personal injury and/or pr oduct damage. This product is designed for use in Installation Category II and Pollution Degree 2 per IEC 1010 and IEC 664 respectively . For indo or use only . ...

  • Agilent Technologies E6380-90073 - page 36

    36 Certification Agilent T echnologies certifi es that t his product met its published specifi cations at th e time of shipment from the f actory . Agilent further certifie s that its calibration measuremen ts are trac eable to the United States National Inst itute of Standards and T echnology , to the extent allowed by the Institute ’ s calibrat ...

  • Agilent Technologies E6380-90073 - page 37

    37 Agilent T ech nologies W arranty Statement for Commercial Produ cts Agilent T echnol ogies 8935 Series E6380A CDMA Cellular/PCS Base Station T est Set Duration of W arranty: 1 Y ear 1. Agilent T echnologies warrants Agilent hardware, accessories and supplies against defects in mate rials and workmanship f or the period spec ified above. If Agil ...

  • Agilent Technologies E6380-90073 - page 38

    38 7. TO THE E XTENT ALLOWED BY LOCAL L A W , THE ABOVE W ARRANTIES ARE EXCLUSIVE AND NO OTHER W ARRANTYOR CONDITION, WHETHER WRITTEN OR ORAL I S EXPRESSED OR IMPLIED AND AGILENT TECHN OLOG IES SPECIFICALL Y DI SCLAIMS ANY IM PLIED W ARRANTIES OR CONDITIONS OR MERCHANT ABILITY , SA TISF ACTOR Y QUALITY , AND FITNESS FOR A P ARTICULAR PURPOSE. 8. Ag ...

  • Agilent Technologies E6380-90073 - page 39

    39 Assist ance Maintenance Agreements Pro duct maintenance agr eements and othe r customer assistance agreements are available for Agi l ent T echnologies produc ts. For any assistance, contact your nearest Agilent T echnologies Sales and Ser vice Office. Regional Sales and Service Offices T able 1 Regional Sales and Service Offices United States o ...

  • Agilent Technologies E6380-90073 - page 40

    40 Power Cables T able 2 Power Cables Plug T ype Plug Descriptions male/female Agilent Part # (cable & plug) Cable Descriptions Straight/Straight Straight/90 ° 8120- 1689 8120-1692 79 inches, mint gray 79 inches, mint gray Used in the f ollowing loca tions Afghanist an, Albania, Algeria, Angola, Armenia, Austri a, Azerbaijan, Azores Bangladesh ...

  • Agilent Technologies E6380-90073 - page 41

    41 USSR, Ukraine, Uzbekist an W estern Africa, W estern Sahara Y ugo slavia Zaire T able 3 Power Cables Plug T yp e Plug Descriptions male/female Agilent Part # (cable & plug) Cable Descr ip ti ons Straight/Stra ight 8120-06 98 90 inches, bl ack Used in t he follow ing locations Peru T able 4 Power Cables Plug T yp e Plug Descri ptions male/fem ...

  • Agilent Technologies E6380-90073 - page 42

    42 T able 5 Power Cables Plug T ype Plug Descri ptions male/female Agilent Part # (cable & plug ) Cable Descriptions Straight/S traight Straigh t/90 Straight/S traight 8120-137 8 8120-152 1 8120-175 1 90 inches, jade gray 90 inches, jade gray 90 inches, jade gray Used in th e following lo cations American Samoa Bahamas, Barb ados, Beliz e, Berm ...

  • Agilent Technologies E6380-90073 - page 43

    43 T able 6 Power Cables Plug T ype Plug Descriptions male/female Agilent Part # (cable & plug) Cabl e Descript ions Straight/ Straight Straight/ 90 ° 8120-4 753 8120-4 754 90 inches, dark gray 90 inches, dark gray Used in the f ollowing loca tions Japan T able 7 Power Cables Plug T ype Plug Descriptions male/female Agilent Part # (cable & ...

  • Agilent Technologies E6380-90073 - page 44

    44 Leso tho Malawi South-W e st Africa (Nami bia), Swaziland Zambia, Zimbabwe T able 9 Power Cables Plug T ype (Male) Plug Descriptions male/female Agilen t Part # (cable & plug) Cable Descriptions 90 ° /Straig ht 90 ° /90 ° 8120-1 351 8120-1 703 90 inches, mint gray 90 inches, mint g ray Used in the following locations Bahrain, British Indi ...

  • Agilent Technologies E6380-90073 - page 45

    45 T able 10 Power Cabl es Plug T yp e Plug Descr ipti ons mal e/f ema le Agilent Part # (cable & plug) Cable Descri ptions Straight/S traight Straight /90 ° 8120-1369 8120-06 96 79 inches, gray 80 inches, gray Used in th e following lo cations Argentina, Australia China (People’s Republic) New Zeal and Papua New Guinea Uruguay We s t e r n ...

  • Agilent Technologies E6380-90073 - page 46

    46 Attention ATTENTION Static S ensitive Devices This instrument was constructed in an ESD (elect ro-static dischar ge) protected environment. This is because most of the semiconductor devices used in this instrument are suscept ible to damage by static discharge. Depending on the magnitude of the char ge, device substrate s can be punc ture d or d ...

  • Agilent Technologies E6380-90073 - page 47

    47 Conventions Used i n This Manual The following conventions are used througho ut this manual to help clarify instr uctions and reduce u nnecessary text: • “T est Set” refers to the Agilent T echnologi es 8935 CDMA Cellular/PCS Ba se St ation T est Set. • T est Set keys are ind icat ed like this: Preset • T est Set scree n information, s ...

  • Agilent Technologies E6380-90073 - page 48

    48 T rademark Ac knowledgments Hewlett-Packard  and HP  are registered trademarks of Hewlett-Packard Company . Microsoft ‚ W indows  , and MS-DOS  ‚ are regist ered tradem a rk s of Micro s oft Co rp orat i o n. ProComm‚ is a registered trade mark of Data Storm T echnolo gies, Inc. HyperT erminal  is a re gistered t rademark of ...

  • Agilent Technologies E6380-90073 - page 49

    49 Which Do cumen ts are Re quired The fol l o w i ng do cu m ents a re pa r t o f th e T e st Se t’s docum e nt se t. U se the table to help yo u decide whi ch document you need. T able 12 Document Navigation Document Part Number Usage CDMA Application Guide E6380-90016 Use this manual f or basic CDMA meas ureme nts and for ge tting st arted wit ...

  • Agilent Technologies E6380-90073 - page 50

    50 ...

  • Agilent Technologies E6380-90073 - page 51

    51 1 GPIB Command Dictionary This is the GPIB command dictionary . It contains syntax and guidelin es for the GPIB commands available in the T est Set. ...

  • Agilent Technologies E6380-90073 - page 52

    52 Chapter 1 GPIB Command Dictionar y Notice Notice Permission to use, copy , and dist ribu te this te mplate is here by granted, provided that the above cop yright notice appears in all copie s and that both that copyright notice and this permission notice appear in supporting har dcopy and online documentation . All o ther rights res erved . The ...

  • Agilent Technologies E6380-90073 - page 53

    Chapter 1 53 GPIB Command Dictionar y Using This Dict ionary Using Thi s Dictionary This dictionary is arranged according to GPIB subsystem. Each subsystem contains syntax for a specific set of functions within the T est Set. NO TE For all int ents and pur poses, GPIB and HP-IB may be considered one and the same protocol thro ughout this manual. Wh ...

  • Agilent Technologies E6380-90073 - page 54

    54 Chapter 1 GPIB Command Dictionar y AF ANalyzer subsystem AF ANalyz er subsystem :AIN <string> :AIN? These commands set/query t he input s tate of the AUDIO IN LO connector . Syntax AFAN:AIN ‘Gnd’ ‘Float’ ‘600 to Hi’ AFAN:AIN? Screen/field equivalent AF AN:AIN sets the Au dio In Lo field on the AF ANAL YZER screen. :DEMPhasis &l ...

  • Agilent Technologies E6380-90073 - page 55

    Chapter 1 55 GPIB Command Dictionar y AF AN alyzer subsys tem :DEMPhasis:GAIN <string> :DEMPhasis:GAIN? These commands set/quer y the AF analyzer ’ s amplifier gain. This command is only effective when autoranging is turned of f using the AF AN:RANG ‘Hold’ command. Gain is automatically set when the AF A N:RANG ‘Auto’ command is use ...

  • Agilent Technologies E6380-90073 - page 56

    56 Chapter 1 GPIB Command Dictionar y AF ANalyzer subsystem :DET ector <string> :DET ector? These commands set/quer y the type of detector used for AF signals. Syntax AFAN:DET ‘RMS’ 'RMS*SQRT2' 'PK+' 'PK-' 'PK+-/2' 'PK+-MAX' 'PK+ HOLD' 'PK- HOLD' 'PK+-/2 Hd' & ...

  • Agilent Technologies E6380-90073 - page 57

    Chapter 1 57 GPIB Command Dictionar y AF AN alyzer subsys tem :DET ector:SET Tling <string> :DET ector:SET Tling? These commands set/quer y the settling time for audio measurements. (Use Fast when measuring signals greater than 2 00 Hz). Syntax AFAN:DET:SETT ‘Fast’ ‘Slow’ AFAN:DET:SETT? Screen/field equivalent AF AN:DET SETT controls ...

  • Agilent Technologies E6380-90073 - page 58

    58 Chapter 1 GPIB Command Dictionar y AF ANalyzer subsystem :FIL T1|FIL T ER1 <string> :FIL T1?|FIL TER1? These commands set/quer y the audio filter 1 selection. Syntax AFAN:FILT1 ‘20Hz HPF’ ‘50HZ HPF’ ‘300Hz HPF’ ‘C MESSAGE’ AFAN:FILT1? Screen/field equivalent AF AN:FIL T1 controls the Filter 1 field on the AF ANAL YZER scree ...

  • Agilent Technologies E6380-90073 - page 59

    Chapter 1 59 GPIB Command Dictionar y AF AN alyzer subsys tem :GTIMe <real number> :GTIMe? This command sets the gate time (how long the AF counter samples the input si gnal) for t he audio f requency count er . This command utilizes the “Real N umber Setting Syntax” on page 299, but does no t use the :ST A T e command. Syntax AFAN:GTIM & ...

  • Agilent Technologies E6380-90073 - page 60

    60 Chapter 1 GPIB Command Dictionar y AF ANalyzer subsystem :INPut:GAIN <string> :INPut:GAIN? These commands set/query the input gain setti ng fo r the audi o analyzer . This co mmand is only effe ctive when auto ranging is tur ned off using t he AF AN:RANG ‘Hold’ co mmand. Gain is automatically set when the AF A N:RANG ‘Auto’ command ...

  • Agilent Technologies E6380-90073 - page 61

    Chapter 1 61 GPIB Command Dictionar y AF AN alyzer subsys tem :NOTCh:GAIN <string> :NOTCh:GAIN? These commands set/quer y the gain of t he AF analyzer ’ s notch f ilter amplifier . This co mmand is only effe ctive when autoranging is tur ned off using t he AF AN:RANG ‘Hold’ co mmand. Gain is automatically set when the AF A N:RANG ‘Aut ...

  • Agilent Technologies E6380-90073 - page 62

    62 Chapter 1 GPIB Command Dictionar y AF ANalyzer subsystem :RANGing <string> :RANGing? These commands set/quer y the gain co ntrol mode of the AF analyzer . Syntax AFAN:RANG ‘Auto’ ‘Hold’ AFAN:RANG? Screen/field equivalent AF AN:RANG controls the Gain Cntl field on the AF ANAL YZER screen. :SMPoint <string> :SMPoint? These comm ...

  • Agilent Technologies E6380-90073 - page 63

    Chapter 1 63 GPIB Command Dictionar y AF AN alyzer subsys tem :SPEaker:MODE <str ing> :SPEaker:MODE? These commands set/quer y the automatic le vel control (ALC) function for the instrument’ s internal speaker . Syntax AFAN:SPE:MODE ‘On’ ‘Off’ AFAN:SPE:MODE? Screen/field equivalent AF AN:SPE:MODE controls the Speaker AL C fiel d on ...

  • Agilent Technologies E6380-90073 - page 64

    64 Chapter 1 GPIB Command Dictionar y AFG1|AFGENERA T OR1 subsystem AFG1|AFGENERA TOR1 subsystem :AM <real number> :AM? This command sets AM m odulat ion dep th when the AFG1:DEST ‘AM’ is used first. This command utilizes the “Real Number Setting Syntax” on page 299. Syntax AFG1:AM <real number> AFG1:AM? !returns the value of th ...

  • Agilent Technologies E6380-90073 - page 65

    Chapter 1 65 GPIB Command Dictionar y AFG1|AFGENERA TOR1 subsys tem :FM <real number> :FM? This command sets FM modulation devia tion when the AFG1:DEST ‘FM’ command is used. This command utilizes the “Real N umber Setting Syntax” on page 299. Syntax AFG1:FM <real number> AFG1:FM? Screen/field equivalent AFG1:FM controls the AFG ...

  • Agilent Technologies E6380-90073 - page 66

    66 Chapter 1 GPIB Command Dictionar y AFG1|AFGENERA T OR1 subsystem :OUTPut <real numbe r> :OUTPut? This command sets the amplitude of the audio signal (volts rms) at the AUDIO OUT connector if the AFG1:DEST is ‘Audio Out’. This command utilizes the “Real Number Setting Syntax” on page 299 Syntax AFG1:OUTP <real number> AFG1:OUT ...

  • Agilent Technologies E6380-90073 - page 67

    Chapter 1 67 GPIB Command Dictionar y AFG2|ENCoder subsys tem AFG2|ENCoder s ubsystem :AM <real number> :AM? This command sets AM modulation dept h if the command AFG2:DEST ‘AM’ is used first. This command utilizes the “Real N umber Setting Syntax” on page 299. Syntax AFG2:AM <real number> AFG2:AM? !returns the value of the ampl ...

  • Agilent Technologies E6380-90073 - page 68

    68 Chapter 1 GPIB Command Dictionar y AFG2|ENCoder subsystem :DESTination <string> :DESTination? :DEST sets/queries the destinati on of the AF generato r 2’ s signal. Syntax AFG2:DEST ‘AM’ ‘FM’ ‘Audio Out’ AFG2:DEST? Screen/field equivalent AFG2:DEST control s the AFGen2 To field (uppe r su bfie ld) o n the RF GENERA TOR screen. ...

  • Agilent Technologies E6380-90073 - page 69

    Chapter 1 69 GPIB Command Dictionar y AFG2|ENCoder subsys tem :FIL T er <string> :FIL T er? T o improve performance, one of four premodulatio n filters is automati cally selected fo r each encoder mode. The automatic ally selected filter can be changed only by using GPIB commands. W e recommend that you leave this setting at its automatica ll ...

  • Agilent Technologies E6380-90073 - page 70

    70 Chapter 1 GPIB Command Dictionar y AFG2|ENCoder subsystem :FIL T er:MODE <string> :FIL T er:MODE? These commands set/quer y the premod ulation filter ’ s control. W e r ec omme n d that y ou le av e this settin g at it s default se tti n g (filter mode OFF is the power up default state .) See the :FIL T er command on page 69 for more inf ...

  • Agilent Technologies E6380-90073 - page 71

    Chapter 1 71 GPIB Command Dictionar y AFG2|ENCoder subsys tem :FREQuency <real number> :FREQuency? This command set/que ries the outp ut frequency of AF generator 2. This command utilizes the “Real N umber Setting Syntax” on page 299, but does no t use the :ST A T e command. Syntax AFG2:FREQ <real number> AFG2:FREQ? Screen/field equ ...

  • Agilent Technologies E6380-90073 - page 72

    72 Chapter 1 GPIB Command Dictionar y AFG2|ENCoder subsystem :OUTPut <real numbe r> :OUTPut? This command sets the amplitude of the audio signal (volts rms) at the AUDIO OUT connector if the AFG2:DEST is ‘Audio Out’. This command utilizes the “Real Number Setting Syntax” on page 299. Syntax AFG2:OUTP <real number> AFG2:OUTP? Scr ...

  • Agilent Technologies E6380-90073 - page 73

    Chapter 1 73 GPIB Command Dictionar y AFG2|ENCoder subsys tem :POLarity <string> :POLarity? This command causes the digitally modulat ed signals to be sent with a normal or inve rted response to the di gital da ta. When I nverted, a digital 1 pr odu ces a f req uency shi ft in an FM c arrier op posite t o normal operation. Syntax AFG2:POL ‘ ...

  • Agilent Technologies E6380-90073 - page 74

    74 Chapter 1 GPIB Command Dictionar y AFG2|ENCoder subsystem :SEND This com mand sends the encode r data s tream. Syntax AFG2:SEND Screen/field equivalent AFG2 : SEND co n t ro ls the Send field on the SIGNALING ENCODER screen when the Mode fi eld is s et to AMPS-TA CS , NAMP-NTAC , o r DTMF . :STOP This command stops the data stream that is being ...

  • Agilent Technologies E6380-90073 - page 75

    Chapter 1 75 GPIB Command Dictionar y AFG2|ENCoder subsys tem :AMPS|T ACS:BUSY <string> :AMPS|T ACS:BUSY? These commands set/query the busy/idle status information included in the signaling sequenc e. • ‘Idle’ sets the busy/idle bits of t he forward control cha nnel information to indi cate an idle state. • ‘Busy’ sets the busy/id ...

  • Agilent Technologies E6380-90073 - page 76

    76 Chapter 1 GPIB Command Dictionar y AFG2|ENCoder subsystem :AMPS|T ACS:BUSY :DELay <integer > These commands set/query the number of bits that are counted before a busy/idle bit c hanges from the i dle state t o the busy state. This f unction is used i n conjunction with t he WS Delay and 1 stBitDly settings of the AFG2:AMPS|T ACS:BUSY comm ...

  • Agilent Technologies E6380-90073 - page 77

    Chapter 1 77 GPIB Command Dictionar y AFG2|ENCoder subsys tem :AMPS|T ACS:DA T A: AM <real number> :AMPS|T ACS:DA T A: AM? These commands set/query the data level when the AFG2:DEST ‘AM’ command is used f irst. The d ata level unit s are %. This command utilize s the “Rea l Number Setting S yntax” on page 299. Syntax AFG2:AMPS:DATA:AM ...

  • Agilent Technologies E6380-90073 - page 78

    78 Chapter 1 GPIB Command Dictionar y AFG2|ENCoder subsystem :AMPS|T ACS:DA T A: LEV el <real number> :AMPS|T ACS:DA T A: LEV el? These commands set/query the dat a level when the AFG2:DEST ‘Aud io Out’ command is used first. The data level units are mV . This command utilizes the “Real Number Setting Sy ntax” on page 299. Syntax AFG2 ...

  • Agilent Technologies E6380-90073 - page 79

    Chapter 1 79 GPIB Command Dictionar y AFG2|ENCoder subsys tem :AMPS|T ACS:FILLer:DA T A1 <string> :AMPS|T ACS:FILLer:DA T A2 <string> :AMPS|T ACS:FILLer:DA T A1? :AMPS|T ACS:FILLer:DA T A2? These commands set/quer y FOCC filler data. Each filler contains 7 hexadecimal charac ters representing the 2 type bits and 26 infor mation bits of ...

  • Agilent Technologies E6380-90073 - page 80

    80 Chapter 1 GPIB Command Dictionar y AFG2|ENCoder subsystem :AMPS|T ACS:FILLer:STOP This command stops t he output of filler data. T o use this command, AFG2:AMPS:CHAN must be set to ‘Cntl’. Syntax AFG2:AMPS:FILL:STOP Screen/field equivalent AFG2:AMPS:FILL:STOP controls t he Stop Filler field on the SIGNALING ENCODER s creen when the Mo de fie ...

  • Agilent Technologies E6380-90073 - page 81

    Chapter 1 81 GPIB Command Dictionar y AFG2|ENCoder subsys tem :AMPS|T ACS:MESSage:DA T A1 <string> :AMPS|T ACS:MESSage:DA T A2 <string> :AMPS|T ACS:MESSage:DA T A1? :AMPS|T ACS:MESSage:DA T A2? These commands set/quer y FOCC Stream A and S tream B message data. Like the filler data, the message i nfor mation can only be input in full (s ...

  • Agilent Technologies E6380-90073 - page 82

    82 Chapter 1 GPIB Command Dictionar y AFG2|ENCoder subsystem :AMPS|T ACS:SA T :F M <real number> :AMPS|T AC S:SA T :FM? These commands set/quer y the SA T level when the AFG2:DEST ‘FM’ and AFG2:AMPS:CHAN ‘ V oice’ commands are used. The SA T level’ s units are kHz. This command utilizes the “Real Number Setting Sy ntax” on page ...

  • Agilent Technologies E6380-90073 - page 83

    Chapter 1 83 GPIB Command Dictionar y AFG2|ENCoder subsys tem :AMPS|T ACS:SA T :F REQuency <real numbe r> These commands set/query the super visory audio tone’ s frequency . This signal is sent continuously whenever AFG2:AMPS:CHAN ‘V oice’ is used, but is interrupte d when a FVC message is sent. This command utilize s the “Rea l Numbe ...

  • Agilent Technologies E6380-90073 - page 84

    84 Chapter 1 GPIB Command Dictionar y AFG2|ENCoder subsystem :DTMF:FREQuency:COLum n <multiple real numb er> :DTMF:FREQuency:COLum n? These commands set/quer y the freq uencies for t he verti cal columns i n the symbol fre quencies tabl e of the DTMF encod er . Syntax AFG2:DTMF:FREQ:COL <integer value>,<real number> AFG2:DTMF:FREQ ...

  • Agilent Technologies E6380-90073 - page 85

    Chapter 1 85 GPIB Command Dictionar y AFG2|ENCoder subsys tem :DTMF:ONTime <real number> :DTMF:ONTime? These commands set/query the length of time eac h DTMF tone is on during the sequence . This command utilizes the “Real Number Setting Syntax” on page 299, but does no t use the :ST A T e command. Syntax AFG2:DTMF:ONT <real number> ...

  • Agilent Technologies E6380-90073 - page 86

    86 Chapter 1 GPIB Command Dictionar y AFG2|ENCoder subsystem :DTMF:ST ANdard <string> :DTMF:ST ANdard? These com mands set/quer y the stan dard applied to the e ncoded dat a. The standard affects t he types of frames that c an be sent, t he range of valid channel numbers, the e ncoding of th e frame da ta, and th e interpret atio n of receive ...

  • Agilent Technologies E6380-90073 - page 87

    Chapter 1 87 GPIB Command Dictionar y AFG2|ENCoder subsys tem :FGENerator:SUNits <string> :FGENerator:SUNits? These commands set/query whether the signal’ s output is in units of rms or peak. The AFG2:DEST ‘Audio Out’ command must be used with this command. Syntax AFG2:FGEN:SUN ‘RMS’ ‘Peak’ AFG2:FGEN:SUN? Screen/field equivalent ...

  • Agilent Technologies E6380-90073 - page 88

    88 Chapter 1 GPIB Command Dictionar y AFG2|ENCoder subsystem :NAMPs|NT ACs:BUSY <string> :NAMPs|NT ACs:BUSY ? These commands set/query the busy/idle status information included in the signaling sequenc e. • ‘Idle’ sets the busy/idle bits of t he forward control cha nnel information to indi cate an idle state. • ‘Busy’ sets the bus ...

  • Agilent Technologies E6380-90073 - page 89

    Chapter 1 89 GPIB Command Dictionar y AFG2|ENCoder subsys tem :NAMPs|NT ACs:BUSY :DELay <integer> This command sets the number of bits that are counted before a busy/idle bit c hanges form the i dle state to the busy st ate. This functi on is used i n conjunction with t he WS Delay and 1 stBitDly settings of the :NAMPs|NT ACs:BUSY comm and. T ...

  • Agilent Technologies E6380-90073 - page 90

    90 Chapter 1 GPIB Command Dictionar y AFG2|ENCoder subsystem :NAMPs|NT ACs:DSA T :MESS <string> :NAMPs|NT ACs:DSA T :MESS? These commands set/quer y the 24-bit d igital superviso ry audio tone (DSA T) sequence . The seve n standard sequences a re: 25 56CB, 255B2B, 256A9B, 25AD4D, 26AB2B, 26B2AD, and 2969AB. (These c odes are the inverse of th ...

  • Agilent Technologies E6380-90073 - page 91

    Chapter 1 91 GPIB Command Dictionar y AFG2|ENCoder subsys tem :NAMPs|NT ACs:DSA T :STOP This command stops the DS A T if it is being sent. See “:NAMPs |NT ACs:DSA T :SEND <s tr ing> :NAMPs|NT ACs:DSA T :SEND?” on page 90. Syntax AFG2:NAMP:DSAT:STOP Screen/Field Equivalent AFG2:NAMP:D SA T :STOP controls the Stop DSAT fie ld o n the SIGNAL ...

  • Agilent Technologies E6380-90073 - page 92

    92 Chapter 1 GPIB Command Dictionar y AFG2|ENCoder subsystem :NAMPs|NT ACs[:FOC C]:FM <real number> :NAMPs|NT ACs[:FOC C]:FM? This command sets/queries the data level when the AFG2:DEST ‘FM’ command is used. The data le vel units ar e kHz. This command utilizes the “Real Number Setting Sy ntax” on page 299. Syntax AFG2:NAMP:FOCC:FM AF ...

  • Agilent Technologies E6380-90073 - page 93

    Chapter 1 93 GPIB Command Dictionar y AFG2|ENCoder subsys tem :NAMPs|NT ACs[:FOC C]:RA TE <real number> :NAMPs|NT ACs[:FOC C]:RA TE ? These commands set/query the data rate of the encode d message. This command utilize s the “Rea l Number Setting S yntax” on page 299 but does no t use the :ST A T e command. Syntax AFG2:NAMP:FOCC:RATE AFG2 ...

  • Agilent Technologies E6380-90073 - page 94

    94 Chapter 1 GPIB Command Dictionar y AFG2|ENCoder subsystem :NAMPs|NT ACs[:FOC C]:FILLer:SEND This command sends the filler inf ormation that is entered with the AFG2:FILL:DA T A1 and AFG2:FILL:DA T A2 commands. T o use this command, AFG2:NAMP:CHAN must be set to ‘Cntl’. Syntax AFG2:NAMP[:FOCC]:FILL:SEND Screen/field equivalent AFG2:NAMP[:FOCC ...

  • Agilent Technologies E6380-90073 - page 95

    Chapter 1 95 GPIB Command Dictionar y AFG2|ENCoder subsys tem :NAMPS|NT ACS[:FOCC]:MES Sage:DA T A1 <string> :NAMPS|NT ACS[:FOCC]:MES Sage:DA T A2 <string> :NAMPS|NT ACS[:FOCC]:MES Sage:DA T A1? :NAMPS|NT ACS[:FOCC]:MES Sage:DA T A2? These commands set/quer y FOCC Stream A and S tream B message data. Like the filler data, the message i ...

  • Agilent Technologies E6380-90073 - page 96

    96 Chapter 1 GPIB Command Dictionar y AFG2|ENCoder subsystem :NAMPs|NT ACs:FVC:R A TE <real nu mber> :NAMPs|NT ACs:FVC:R A TE? These commands set/query the data rate of the encode d message. This command utilizes the “Real Number Setting Sy ntax” on page 299 but does no t use the :ST A T e command. Syntax AFG2:NAMP:FVC:RATE <real numbe ...

  • Agilent Technologies E6380-90073 - page 97

    Chapter 1 97 GPIB Command Dictionar y AFG2|ENCoder subsys tem :NAMPs|NT ACs:FVC:F M <real number> :NAMPs|NT ACs:FVC:F M? This command sets/queries the data level when the AFG2:DEST ‘FM’ command is used. The data le vel units ar e kHz. This command utilize s the “Rea l Number Setting S yntax” on page 299. Syntax AFG2:NAMP:FVC:FM AFG2:N ...

  • Agilent Technologies E6380-90073 - page 98

    98 Chapter 1 GPIB Command Dictionar y AFG2|ENCoder subsystem :NAMPs|NT ACs:SEND :NAMPs|NT ACs:SEND ? This co mma nd sets/ queri es the typ e of data sent w hen the A FG2:SE ND command is used. • Selecting ‘ Message’ se nds the cont ents of the forward voic e message (AFG2:NAMP:FVC:MESS <string>). • Selecting ‘ DST’ causes the digi ...

  • Agilent Technologies E6380-90073 - page 99

    Chapter 1 99 GPIB Command Dictionar y CANalyzer subsystem CANalyzer subsystem The CANalyzer subsystem controls func tio ns unique to the CDMA ANAL YZER screen. :ACP:POW er:CALibrate This command initiat es calibration of the ad jacent channel power measurement for the current RF analyzer frequen cy at the selected frequency o ffset and filter band ...

  • Agilent Technologies E6380-90073 - page 100

    100 Chapter 1 GPIB Command Dictionar y CANalyzer subsystem :ACP:POW er:OFFSet <real num ber> :ACP:POW er:OFFSet? This command sets the frequency offset used when making adjacent channel power measurements. The setting d esignates the fr equency offset f rom the center frequency at which the measurement is made. V alid values are 100 kHz to 3 ...

  • Agilent Technologies E6380-90073 - page 101

    Chapter 1 101 GPIB Command Dictionar y CANalyzer subsystem :AUTO:GAIN? This command returns the gain setting. The value and mode of the gain are set by the commands :P A TH :GAIN and :P A TH:G AIN:MODE. This query is valid only while measuring rho, EVM or channel power . Syntax CAN:AUTO:GAIN? Screen/field equivalent CAN :AU TO:G A IN ? q u e ri e s ...

  • Agilent Technologies E6380-90073 - page 102

    102 Chapter 1 GPIB Command Dictionar y CANalyzer subsystem :CHANnel:POW er:FIL T er :CHANnel:POW er:FIL T er? These commands set/quer y the filt er for t he channel p ower measurement. • ‘1.23M’ me asures the absolu te channel p ower level of the RF signa l in a 1.23 MHz bandwidth centered around the RF cha nnel or tune frequency setting. Thi ...

  • Agilent Technologies E6380-90073 - page 103

    Chapter 1 103 GPIB Command Dictionar y CANalyzer subsystem :DIRection <string> :DIRection? These commands set/quer y the directio n of the CDMA analyzer . This command is valid only when frequency tuning mode is selected (CONF:RFD ‘Freq’). :DIR is not used while measuring average power . Syntax CAN:DIR ‘Fwd’ ‘Rev’ CAN:DIR? Screen/ ...

  • Agilent Technologies E6380-90073 - page 104

    104 Chapter 1 GPIB Command Dictionar y CANalyzer subsystem :EVENt:TRIGger <str ing> :EVENt:TRIGger? These co mma nd s set/qu e ry th e tr i gge r se t tin g of the CD MA ANAL YZER screen. Syntax CAN:EVEN:TRIG ‘27 ms’ ‘20 ms’ ‘80 ms’ ‘2 s’ ‘Delay’ ‘Immed’ CAN:EVEN:TRIG? Example CAN:EVEN:TRIG ‘Delay’ !sets the trigge ...

  • Agilent Technologies E6380-90073 - page 105

    Chapter 1 105 GPIB Command Dictionar y CANalyzer subsystem :MODE <string> :MODE? These commands set/quer y the measurement d isplayed in t he upper left co rner of t he displa y . Syntax CAN:MODE ‘Avg Pwr’ ‘ACP’ ‘Rho’ ‘EVM’ ‘Chan Pwr’ CAN:MODE? Screen/field equivalent CAN:MODE selects the measurement to be displayed in the ...

  • Agilent Technologies E6380-90073 - page 106

    106 Chapter 1 GPIB Command Dictionar y CANalyzer subsystem :P A TH:GAIN <string> :P A TH:GAIN? These commands set/quer y the gain of t he CDMA analyzer . This command will only be enabled when autoranging is turned off (CAN: P A TH: MODE ‘Ho ld’). This command may conf lict with the c ode domain ana lyzer ’ s settings CDAN:P A TH:GAIN a ...

  • Agilent Technologies E6380-90073 - page 107

    Chapter 1 107 GPIB Command Dictionar y CANalyzer subsystem :P A TH:GAIN:MODE <st ring> :P A TH:GAIN:MODE? These commands set/query t he mode of the Ga in field, allowi ng you to choose between an autoranging gain (‘Auto’ ) or a fixed gain (‘Hold’). This command may conf lict with the c ode domain analyzer ’ s settings CDAN:P A TH:GA ...

  • Agilent Technologies E6380-90073 - page 108

    108 Chapter 1 GPIB Command Dictionar y CANalyzer subsystem :PNMode <string> :PNMode? These commands set/query t he PN offset mod e. • ‘Auto’ is used to calcula te the system’ s PN of fset if it is unknown. When ‘Auto’ is selected, the CAN:PNIN command can be used to enter a PN incre ment and speed the PN offset c alculation. • ? ...

  • Agilent Technologies E6380-90073 - page 109

    Chapter 1 109 GPIB Command Dictionar y CANalyzer subsystem :POW er:GAIN <string> :POW er:GAIN? These commands set/quer y the gain of t he CDMA analyzer . These commands are valid only when an average power (A vg P wr) measurement is selecte d. For c hannel power , EVM, or r ho measurements, use the CAN:P A TH:GAIN command. Syntax CAN:POW:GAIN ...

  • Agilent Technologies E6380-90073 - page 110

    110 Chapter 1 GPIB Command Dictionar y CANalyzer subsystem :POW er:GAIN:MODE <s tring> :POW er:GAIN:MODE? These commands set/query t he mode of the Pw r Gain field, allo wing you to choose between auto matic ga in (Auto) or fixed gain (Hold).This command is val id only when aver age power is selec ted. For channel power , EV M and rho me asur ...

  • Agilent Technologies E6380-90073 - page 111

    Chapter 1 111 GPIB Command Dictionar y CANalyzer subsystem :POW er:ZERO This command zeroes the power met er for an average power measurement. Syntax CAN:POW:ZERO Screen/field equivalent CAN:POW :ZERO controls the Pwr Zero field on the CDMA ANAL YZER screen. This f ield is only available when the average power measurement is selecte d. :POW er:ZERO ...

  • Agilent Technologies E6380-90073 - page 112

    112 Chapter 1 GPIB Command Dictionar y CANalyzer subsystem :SAMPle:TIME <real number> :SAMPle:TIME? These commands set/quer y the amount o f time that rho or EVM is measured. This c ommand utilizes the “Real Numbe r Setting Syntax” on page 299, but does not use the :ST A T e command. Syntax CAN:SAMP:TIME <real number> !values from 0 ...

  • Agilent Technologies E6380-90073 - page 113

    Chapter 1 113 GPIB Command Dictionar y CANalyzer subsystem :TRIGger:ST A T e <string> :TRIGger:ST A T e? These commands set/query t he state of the trigger . CAN:TRIG:ST A T can be o verridden wit h the TRIG:MODE:RETR REP and TRIG:MODE:RETR SING tri gger commands. When the trigge r commands are used, t he state of the f ield on the display i ...

  • Agilent Technologies E6380-90073 - page 114

    114 Chapter 1 GPIB Command Dictionar y CBUFfer subsystem CBU Ffer subs ystem CBUF controls the CDMA data buffe r . This system allows you to send CDMA data through the CDMA generator . Y ou can send all zeroes, exte r n a l da t a v ia th e DA T A I N co n n e ct or , rand om data, or data supplied via GPIB. :DA T A <48 or 72 character he x stri ...

  • Agilent Technologies E6380-90073 - page 115

    Chapter 1 115 GPIB Command Dictionar y CB UFf er s ubsy s tem :FRAMe:ST ARt <integer>,<rea l number> :FRAMe:ST ARt? These commands set/quer y the first frame of data to be sent. This command utilizes the “Multiple Real Number Set ting Syntax” on page 282, but does not use the :DUNits and :UNIT s commands. Syntax CBUF:FRAM:STAR <i ...

  • Agilent Technologies E6380-90073 - page 116

    116 Chapter 1 GPIB Command Dictionar y CBUFfer subsystem :ST A T e <string> :ST A T e? These commands set/query the mode of the data buffer . Syntax CBUF:STAT ‘Idle’ ‘Send’ CBUF:STAT? Screen/field equivalent CBUF:ST A T controls Idle/Sen d subfield of the Data Source field on the CDMA GENERA TOR screen. The Data Sou rce field mu st be ...

  • Agilent Technologies E6380-90073 - page 117

    Chapter 1 117 GPIB Command Dictionar y CCOMmon subsys tem CCOMmon subsystem :P A TH <string> :P A TH? These commands set/query RF path of t he CDMA generator . Syntax CCOM:PATH ‘Bypass’ ‘IQ’ CCOM:PATH? Screen/field equivalent CCOM:P A TH controls the CW RF P ath field of the CDMA GENERA TOR screen. ...

  • Agilent Technologies E6380-90073 - page 118

    118 Chapter 1 GPIB Command Dictionar y CD ANalyzer subsystem CDANalyz er subsystem CDAN controls the code domain analyzer . :CONT rol <string> :CONT rol? These commands set/quer y the control menu displayed on the code domain analyzer , when the CDMA standard is set to “IS-95 Only”. Syntax CDAN:CONT ‘Main’ ‘Marker’ ‘Trigger’ ? ...

  • Agilent Technologies E6380-90073 - page 119

    Chapter 1 119 GPIB Command Dictionar y CD ANalyzer subsystem :CPOW er:CALibrate This command initiat es the channel power calibr ation. When this command is rece ived, an internal ly generated calibr ation signal is measured using the average power technique. Channel power is also measured and a correcti on factor is generated. This correction fact ...

  • Agilent Technologies E6380-90073 - page 120

    120 Chapter 1 GPIB Command Dictionar y CD ANalyzer subsystem :EVENt:TRIGger <str ing> :EVENt:TRIGger? These co mma nd s set/qu e ry th e tr i gge r se t tin g of the CO D E DO M screen. Syntax CDAN:EVEN:TRIG ‘27 ms’ ‘20 ms’ ‘80 ms’ ‘2 s’ ‘Delay’ ‘Immed’ CDAN:EVEN:TRIG? Screen/field equivalent CDAN:EVEN:TRIG sets the T ...

  • Agilent Technologies E6380-90073 - page 121

    Chapter 1 121 GPIB Command Dictionar y CD ANalyzer subsystem :FPOW er:NA VG <integer> :FPOW er:NA VG? These commands set/quer y the number o f averages made by the fast power measurement. Syntax CDAN:FPOW:NAVG <integer> CDAN:FPOW:NAVG? Screen/field equivalent CDAN:FPOW :NA VG contro ls the Num Avgs fiel d on the FP Setu p men u of the C ...

  • Agilent Technologies E6380-90073 - page 122

    122 Chapter 1 GPIB Command Dictionar y CD ANalyzer subsystem :GAIN <string> :GAIN? These commands set/quer y the gain of t he code domain analyzer . This command will only be enabled when autoranging is turned off (CDAN: GAIN:M ODE ‘ Ho ld ’). This command may conf lict with the CDMA analyzer ’ s settings CAN:P A TH:GAIN and CA N:P A TH ...

  • Agilent Technologies E6380-90073 - page 123

    Chapter 1 123 GPIB Command Dictionar y CD ANalyzer subsystem :GAIN:MODE <string> :GAIN:MODE? These commands set/query t he mode of the Ga in field, allowi ng you to choose between an autoranging gain (‘Auto’ ) or a fixed gain (‘Hold’). This command may conf lict with the CDMA analyzer ’ s settings CAN:P A TH:GAIN and CA N:P A TH:MOD ...

  • Agilent Technologies E6380-90073 - page 124

    124 Chapter 1 GPIB Command Dictionar y CD ANalyzer subsystem :MARKer:POSition <r eal number> :MARKer:POSition? These commands set/query the posi tion of t he marker . Syntax CDAN:MARK:POS <real number> !values from 0 to 63 CDAN:MARK:POS? Screen/field equivalent CDAN:MARK:POS controls the Walsh Chan field on the Marker menu of the CODE D ...

  • Agilent Technologies E6380-90073 - page 125

    Chapter 1 125 GPIB Command Dictionar y CD ANalyzer subsystem :MODE <string> :MODE? These commands set/quer y the measurement stat e of the code do main analyzer . CDAN:MODE can be overrid den with the TRIG:MODE:RETR REP and TRIG:MODE:RETR SI NG trigger commands. When the trigger commands are used, t he state of the f ield on the display i s n ...

  • Agilent Technologies E6380-90073 - page 126

    126 Chapter 1 GPIB Command Dictionar y CD ANalyzer subsystem :PNINcrement <real number> :PNINcrement? These commands set/quer y the PN increment . If you know the PN incrment, en ter it with this command to spee d the PN offset se arch. If you do not know the incr ement, enter a 1. This command utilizes the :INCR command of the “Real Number ...

  • Agilent Technologies E6380-90073 - page 127

    Chapter 1 127 GPIB Command Dictionar y CD ANalyzer subsystem :PNOFfset <real numb er> :PNOFfset? These commands set/query the PN offset of the system. This command utilizes the :INCR command of the “Real Number Setting Synt ax” on page 299. Syntax CDAN:PNOF <real number> !values from 0.0 to 511.984375 CDAN:PNOF? !returns the value o ...

  • Agilent Technologies E6380-90073 - page 128

    128 Chapter 1 GPIB Command Dictionar y CD ANalyzer subsystem :PUNit <string> :PUNit? These commands set/quer y whether power measure ments are absolute power measurements or relative power measurements. • ‘Abs’ - Absolute c ode domain power d isplays the po wer in eac h of the 64 W alsh channels, relative to the total power inside a 1.2 ...

  • Agilent Technologies E6380-90073 - page 129

    Chapter 1 129 GPIB Command Dictionar y CD ANalyzer subsystem :SCALe:PHASe <string> :SCALe:PHASe? These commands set/query the scale that is used when measuring code domain phase when using the code d omain analyzer . Syntax CDAN:SCAL:PHAS ‘1 mRad’ ‘2 mRad’ ‘5 mRad’ ‘10 mRad’ ‘20 mRad’ ‘50 mRad’ CDAN:SCAL:PHAS? Screen/f ...

  • Agilent Technologies E6380-90073 - page 130

    130 Chapter 1 GPIB Command Dictionar y CD ANalyzer subsystem :SCALe:TIME <string> :SCALe:TIME? These commands set/query t he time per division porti on of the scale that is used when measuring code domain ti ming. Syntax CDAN:SCAL:TIME ‘1 ns’ ‘2 ns’ ‘5 ns’ ‘10 ns’ ‘20 ns’ ‘50 ns’ CDAN:SCAL:TIME? Screen/field equivalent ...

  • Agilent Technologies E6380-90073 - page 131

    Chapter 1 131 GPIB Command Dictionar y CD ANalyzer subsystem :IS2000:CONT rols <string> :IS2000:CONT rols? These commands set/quer y the contr ols menu displayed on the Code Domain analyzer screen when the CDMA standard is set to “IS-2000”. Syntax CDAN:IS2000:CONT ‘Main’ ‘Marker’ ‘Trigger’ ‘Aux’ ‘Gain’ ‘Reference’ ...

  • Agilent Technologies E6380-90073 - page 132

    132 Chapter 1 GPIB Command Dictionar y CD ANalyzer subsystem :IS2000:GAIN:MODE <s tring> :IS2000:GAIN:MODE? For IS-2000 only . These commands set/query the mode of the Gain fi eld, allowing you to choose between an autor anging gain (‘Auto’ ) or a fixed gain (‘Hold’ ). This command may conf lict with the CDMA analyzer ’ s settings C ...

  • Agilent Technologies E6380-90073 - page 133

    Chapter 1 133 GPIB Command Dictionar y CD ANalyzer subsystem :IS2000:GAIN[:SETTing] <string> :IS2000:GAIN[:SETTing]? For IS-2000 only . These commands set/query t he gain of t he Code Domain analyzer . This comm and will only be enabled when autoranging is t urned off ( CDAN:IS2000:GAIN:MODE ‘Hold’). This command may conf lict with the CD ...

  • Agilent Technologies E6380-90073 - page 134

    134 Chapter 1 GPIB Command Dictionar y CD ANalyzer subsystem :IS2000:GAIN:V AL ue? IS-2000 only . This query returns the gain value, as a string, that is shown in th e gain field of the cod e domain screen. This command will only be enabled wh en autoranging is t urned on (CDAN:IS2000:GAIN:MODE ‘Auto’) . Syntax CDAN:IS2000:GAIN:VAL? !returns 0 ...

  • Agilent Technologies E6380-90073 - page 135

    Chapter 1 135 GPIB Command Dictionar y CD ANalyzer subsystem :IS2000:MARKer:POSition <integer> :IS2000:MARKer:POSition ? IS-2000 only . These commands set/query t he integer that corresponds to the current po sition of the marker in the COD E DOMAIN screen when the CDMA standard is set to “IS-2000”. Syntax CDAN:IS2000:MARK:POS <integer ...

  • Agilent Technologies E6380-90073 - page 136

    136 Chapter 1 GPIB Command Dictionar y CD ANalyzer subsystem :IS2000:MARKer:SCALe <strin g> :IS2000:MARKer:SCALe? IS-2000 only . These commands set/query t he scale that is used when measuring code domain power in dB/div units in the code domain analyzer . Syntax CDAN:IS2000:MARK:SCAL ‘1 dB/div’ ‘2 dB/div’ ‘5 dB/div’ CDAN:IS2000:M ...

  • Agilent Technologies E6380-90073 - page 137

    Chapter 1 137 GPIB Command Dictionar y CD ANalyzer subsystem :IS2000:MEASurement:MO DE <string> :IS2000:MEASurement:MO DE? IS-2000 only . These commands set/query the me asurement state for the code domai n anal yze r . This mode can be overridden with the TRIG:MODE:RETR REP and TRIG:MODE:RETR SING trigger commands. When the trigger commands ...

  • Agilent Technologies E6380-90073 - page 138

    138 Chapter 1 GPIB Command Dictionar y CD ANalyzer subsystem :IS2000:MEASurement[:T YPE] <string> :IS2000:MEASurement[:T YPE]? IS-2000 only . These commands set/query t he type of measurement performed in the CODE DOMAIN screen. Syntax CDAN:IS2000:MEAS[:TYPE] ‘Power’ ‘Fast Pwr’ ‘Power&Noise’ ‘Complex Pwr’ ‘FstPwr Sync’ ...

  • Agilent Technologies E6380-90073 - page 139

    Chapter 1 139 GPIB Command Dictionar y CD ANalyzer subsystem :IS2000:PNUMber:MODE <string > :IS2000:PNUMber:MODE? :IS2000:PNumber:MODE <string > :IS2000:PNumber:MODE? IS-2000 only . These commands set/query t he PN setup. • ‘Auto’ is used to calcula te the system’ s PN of fset if it is unknown. When ‘Au to’ is selected, the ...

  • Agilent Technologies E6380-90073 - page 140

    140 Chapter 1 GPIB Command Dictionar y CD ANalyzer subsystem :IS2000:PNUMber:OFFSet <inte ger> :IS2000:PNUMber:OFFSet? :IS2000:PNumber:OFFset <integ er> :IS2000:PNumber:OFFset? IS-2000 only . These commands set/query the conten ts of the PN offset field in the Code Domain analyzer . This command utilizes the :INCR command of the “Real ...

  • Agilent Technologies E6380-90073 - page 141

    Chapter 1 141 GPIB Command Dictionar y CD ANalyzer subsystem :IS2000:POW er:F AST :NA VG <integer> :IS2000:POW er:F AST :NA VG? IS-2000 only . These commands set/query t he number of averages mad e during fast power measurements. Syntax CDAN:IS2000:POW:FAST:NAVG <integer> CDAN:IS2000:POW:FAST:NAVG? Screen/field equivalent CDAN:IS2000:PO ...

  • Agilent Technologies E6380-90073 - page 142

    142 Chapter 1 GPIB Command Dictionar y CD ANalyzer subsystem :IS2000:THReshold <r eal number> :IS2000:THReshold? IS-2000 Only . These commands set/query the minimum display threshold for code domain measurements. W alsh channels that have power levels below the threshold are not displayed as active channels. Syntax CDAN:IS2000:THR <real nu ...

  • Agilent Technologies E6380-90073 - page 143

    Chapter 1 143 GPIB Command Dictionar y CD ANalyzer subsystem :IS2000:TRIGger:ARM IS-2000 only . This command arms code domain me asur ements. This command does not affect co ntinuo us measurement and will always be overrid den by GPIB trig g er ing commands. T o effectively put the co de domain analyzer in single triggering mode, use the TRIG:MODE: ...

  • Agilent Technologies E6380-90073 - page 144

    144 Chapter 1 GPIB Command Dictionar y CD ANalyzer subsystem :IS2000:TRIGger:DELay <real n umber> :IS2000:TRIGger:DELay? IS-2000 only . This command/query sets/re turns the real number that matches the Trig De lay field in t he code d omain measurements. The contents are measured in microseconds. Syntax CDAN:IS2000:TRIG:DEL <real number> ...

  • Agilent Technologies E6380-90073 - page 145

    Chapter 1 145 GPIB Command Dictionar y CD ANalyzer subsystem :IS2000:TRIGger:QUALifier <string> :IS2000:TRIGger:QUALifier ? IS-2000 only . This command returns the string that matches the Qual Event field in the code domain measurements. Syntax CDAN:IS2000:TRIG:QUAL ‘None’ ‘27 ms’ ‘20 ms’ ‘80 ms’ ‘2 s’ ‘Ampl Lo’ ‘Amp ...

  • Agilent Technologies E6380-90073 - page 146

    146 Chapter 1 GPIB Command Dictionar y CGENerator subsystem CGENerat or subsyst em :CONT rol <string> :CONT rol? These commands set/quer y the CDMA generator ’ s mode. It defines the type of data being sent. Syntax CGEN:CONT ‘Data’ ‘Eb/No’ ‘Noise’ CGEN:CONT? Screen/field equivalent CGEN:CONT controls th e Gen Mode field of th e ...

  • Agilent Technologies E6380-90073 - page 147

    Chapter 1 147 GPIB Command Dictionar y CGENerator subsys tem :DA T A :RA TE <string > :DA T A :RA TE? :DA T A:RA TE sets the dat a transfer rat e for data fr om the :DA T A:SOUR command. Syntax CGEN:DATA:RATE ‘14.4 Kbps’ ‘9.6 Kbps’ CGEN:DATA:RATE? Screen/field equivalent CGEN:DA T A:RA TE controls the Data Rate field of the CDMA GENER ...

  • Agilent Technologies E6380-90073 - page 148

    148 Chapter 1 GPIB Command Dictionar y CGENerator subsystem :EBNO:LEV el <real n umber> :EBNO:LEV el? These commands set/quer y the level of the generator when it is i n Eb/No mode. This command u tilizes the “Real Number S etting Syntax” on pa ge 299, but d oes not use the :ST A T e command. Syntax CGEN:EBNO:LEV <real number> !valu ...

  • Agilent Technologies E6380-90073 - page 149

    Chapter 1 149 GPIB Command Dictionar y CGENerator subsys tem :EVENsec <string> :EVENsec? These commands set/quer y the even -second input ’ s setting. When enabled, signals applied to the EVEN SECOND SYNC IN port synchronize the T est Set’ s timing circuits. Syntax CGEN:EVEN ‘Enable’ ‘Not’ CGEN:EVEN? Screen/field equivalent CGEN : ...

  • Agilent Technologies E6380-90073 - page 150

    150 Chapter 1 GPIB Command Dictionar y CONFigure subsystem CONFigure subs ys tem The CONFigure subsyste m contains commands that control several differe nt screens: I/O CONFIGURE, INSTRUMENT CONFIGURE, and PRINTER CONFIGURE. :BADDress <integer> :BADDress? These commands set/quer y the GPIB address for the T est Set. This command utilizes the ...

  • Agilent Technologies E6380-90073 - page 151

    Chapter 1 151 GPIB Command Dictionar y CONFigure subsys tem :BMODe <string> :BMODe? These co mma nd s set/qu e ry th e GP IB co nt rol mode . Syntax CONF:BMOD ‘Control’ ‘Talk&Lstn’ CONF:BMOD? Screen/field equivalent CON F :B M OD con t rols th e Mode fi eld of t he I/O CONFIGURE s creen. :CDMA:MODE <string> :CDMA:MODE? These ...

  • Agilent Technologies E6380-90073 - page 152

    152 Chapter 1 GPIB Command Dictionar y CONFigure subsystem :DISPlay:MESSages :DISPlay:MESSages? These commands set/quer y the reporti ng of messages duri ng normal operation. A ‘Y es’ setting means that all messages are displayed as they occur . Syntax CONF:DISP:MESS ‘Yes’ ‘No’ CONF:DISP:MESS? Screen/field equivalent CONF:DISP:MESS cont ...

  • Agilent Technologies E6380-90073 - page 153

    Chapter 1 153 GPIB Command Dictionar y CONFigure subsys tem :NOTChmode <string> :NOTChmode? These commands set/quer y the coupl ing between the AF Generator 1 and a notch filter . The frequency and gain of the notch filter is set by the AF AN :NOTC:GAIN and AF AN:NO TC:FREQ commands. Syntax CONF:NOTC ‘AFGen1’ !notch filter switched in ‘ ...

  • Agilent Technologies E6380-90073 - page 154

    154 Chapter 1 GPIB Command Dictionar y CONFigure subsystem :OFLevel:ANT enn a <real number> :OFLevel:ANT enn a? These commands set/query the path los s from the device-under -test to the T est Set’ s ANT IN port. This correcti on is applied when the CONF:OFL:MODE ‘On’ command i s used. This command utilizes the “Real Number Setting Sy ...

  • Agilent Technologies E6380-90073 - page 155

    Chapter 1 155 GPIB Command Dictionar y CONFigure subsys tem :OFLevel:RFINout <re al number> :OFLevel:RFINout? These commands set/query the path los s from the RF IN/OUT connector to th e device-under -test. This corr ection is applied when the CONF:OFL:MODE ‘On’ command i s used. This command utilizes the “Real Number Setting Syntax” ...

  • Agilent Technologies E6380-90073 - page 156

    156 Chapter 1 GPIB Command Dictionar y CONFigure subsystem :OMODe <string> :OMODe? These commands set/query the state of the RF offset def ined in the CONF:OFR command. Syntax CONF:OMOD ‘On’ ‘Off’ CONF:OMOD? Screen/field equivalent CONF:OMOD controls the RF Offset field of th e INS TRUME NT CONFIGURE screen. :OPERation:AUTO :OPERation ...

  • Agilent Technologies E6380-90073 - page 157

    Chapter 1 157 GPIB Command Dictionar y CONFigure subsys tem :PCMCia:CARD:TYPE? This command queries the type of PC card inserted in the T est Set’ s front-panel car d slot (such as, RAM or ROM). Syntax CONF:PCMC:CARD:TYPE? Screen/Field Equivalent CONF:PCMC:CARD:TYPE? queries the Card St atus field on the I/O CONFIGURE screen. :PCMCia:CARD:SIZE? T ...

  • Agilent Technologies E6380-90073 - page 158

    158 Chapter 1 GPIB Command Dictionar y CONFigure subsystem :PRINt:LINes|LINE <integer> :PRINt:LINes|LINE? :PRIN:LIN sets/queries the number of li nes to be printed p er page. This command utilizes the “Integer Number Set ting Syntax” on page 206. Syntax CONF:PRIN:LIN <integer> !values: 5 to 120 CONF:PRIN:LIN? Screen/field equivalent ...

  • Agilent Technologies E6380-90073 - page 159

    Chapter 1 159 GPIB Command Dictionar y CONFigure subsys tem :PRINt:FFST a rt <string> :PRINt:FFST a rt? These commands set/quer y a form f eed (blank p age) at the start of printing. Syntax CONF:PRIN:FFST ‘Yes’ ‘No’ CONF:PRIN:FFST? Screen/field equivalent CONF:PRIN:FFST control s the FF at Start field of the PRINTER CONFIGURE screen. ...

  • Agilent Technologies E6380-90073 - page 160

    160 Chapter 1 GPIB Command Dictionar y CONFigure subsystem :PRINt:TITLe <string> :PRINt:TITLe? These commands set/query t he title of the pr int output. A vailable character set: ABCDEFGHIJKLMN OPQRSTUVWXYZ abcdefghijklmnopqrstuvwxyz _0123456789 !@#$%^&*()- +=<>?[ ]{}|’;:”,./ Syntax CONF:PRIN:TITL ‘<string>’ !up to 50 ...

  • Agilent Technologies E6380-90073 - page 161

    Chapter 1 161 GPIB Command Dictionar y CONFigure subsys tem :REFerence:INPut:EXT ernal <string> :REFerence:INPut:EXT ernal? These commands set/quer y the externa l referenc e frequency . The external re ference is sele cted using the :REFerence:INPut :SELect command. Syntax CONF:REF:INP:EXT ‘1 MHz’ ‘2 MHz’ ‘5 MHz’ ‘10 MHz’ ‘ ...

  • Agilent Technologies E6380-90073 - page 162

    162 Chapter 1 GPIB Command Dictionar y CONFigure subsystem :REFerence:INPut:SELec t <string> :REFerence:INPut:SELec t? These commands set/quer y the refe rence input selection. ‘Internal’ and ‘External ’ define an int ernal or e xternal refe rence. ‘Aut o’ will switch between external and inte rnal inputs, based on the signal’ s ...

  • Agilent Technologies E6380-90073 - page 163

    Chapter 1 163 GPIB Command Dictionar y CONFigure subsys tem :RFCStandard <str ing> :RFCStandard? These commands set/quer y the RF channe l standard. This command is used in conjunc tion with t he CONF:RFD ‘Chan’ c ommand. Syntax CONF:RFCS ‘N AMER PCS’ ‘KOR PCS 0’ ‘KOR PCS 1’ ‘MS AMPS’ ‘LS AMPS’ ‘MSL NAMPS’ ‘MSM N ...

  • Agilent Technologies E6380-90073 - page 164

    164 Chapter 1 GPIB Command Dictionar y CONFigure subsystem :RFDisplay <string> :RFDisplay? These commands set/query t he RF display mode . See the Refe rence Guide for more inf ormation ab out frequency and channel tunin g. Syntax CONF:RFD ‘Freq’ !sets RF display to frequency tuning mode ‘Chan’ !sets RF display to channel tuning mode ...

  • Agilent Technologies E6380-90073 - page 165

    Chapter 1 165 GPIB Command Dictionar y CONFigure subsys tem :SPOR9|SPORT9|SB9|SP9:BA UD <string> :SPOR9|SPORT9|SB9|SP9:BA UD? These commands set/query the baud rate of the SERIAL 9 port. Syntax CONF:SPOR9:BAUD ‘150’ ‘300’ ‘600‘ ‘1200‘ ‘2400‘ ‘4800‘ ‘9600‘ ‘19200‘ CONF:SPOR9:BAUD? Screen/field equivalent CONF:SP ...

  • Agilent Technologies E6380-90073 - page 166

    166 Chapter 1 GPIB Command Dictionar y CONFigure subsystem :SPOR9|SPORT9|SB9|SP9:DA T A <string> :SPOR9|SPORT9|SB9|SP9:DA T A? These commands set/quer y the number of data bit s per data word sent over the SERIAL 9 port. Syntax CONF:SPOR9:DATA ‘7 bits’ ‘8 bits’ CONF:SPOR9:DATA? Screen/field equivalent CONF:SPOR9:DA T A controls the Da ...

  • Agilent Technologies E6380-90073 - page 167

    Chapter 1 167 GPIB Command Dictionar y CONFigure subsys tem :SPOR9|SPORT9|SB9|SP9:FC ONtrol|FLOW <stri ng> :SPOR9|SPORT9|SB9|SP9:FC ONtrol|FLOW? These commands set /query the flow control to set the pace of serial communication s for the SERIAL 9 port. Syntax CONF:SPOR9:FCON ‘Xon/Xoff’ ‘None’ CONF:SPOR9:FCON? Screen/field equivalent C ...

  • Agilent Technologies E6380-90073 - page 168

    168 Chapter 1 GPIB Command Dictionar y CONFigure subsystem :SPOR9|SPORT9|SB9|SP9:IEC Ho <string> :SPOR9|SPORT9|SB9|SP9:IEC Ho? These commands set/quer y the character and screen echo ing (instrument echo) when using an external ASCII RS-232 terminal or com pu t e r to en t e r or ed it the IB A S IC pr o gra m . Syntax CONF:SPOR9:IECH ‘On? ...

  • Agilent Technologies E6380-90073 - page 169

    Chapter 1 169 GPIB Command Dictionar y CONFigure subsys tem :SPOR9|SPORT9|SB9|SP9:MO Dem:CALL:DISConnect This command causes the modem to disconnect f rom the call in progress. It also cancels any pending call originati on retries (CONF:SPOR9:MOD:CONN:RETR y). This field i s inactive when CONF:SPOR9:MOD:MODE ‘Disable’ command is used. Syntax CO ...

  • Agilent Technologies E6380-90073 - page 170

    170 Chapter 1 GPIB Command Dictionar y CONFigure subsystem :SPOR9|SPORT9|SB9|SP9:MO Dem:CONNect:TIM eout :SPOR9|SPORT9|SB9|SP9:MO Dem:CONNect:TIM eout? This command sets/ queries the number of seconds to wait befo re aborting a connec tio n attempt. This applies to both answer and originate ac tivities. Syntax CONF:SPOR9:MOD:CONN:TIM <integer> ...

  • Agilent Technologies E6380-90073 - page 171

    Chapter 1 171 GPIB Command Dictionar y CONFigure subsys tem :SPOR9|SPORT9|SB9|SP9:MO Dem:CONNect:RE TR y :SPOR9|SPORT9|SB9|SP9:MO Dem:CONNect:RE TR y? This command sets/queries the number of calls t hat will be tr ied before an ori gination attempt is terminated. The total number of tries will be the number entered in this field, plus one for the o ...

  • Agilent Technologies E6380-90073 - page 172

    172 Chapter 1 GPIB Command Dictionar y CONFigure subsystem :SPOR9|SPORT9|SB9|SP9:MO Dem:CONFigure:S TRing1 :SPOR9|SPORT9|SB9|SP9:MO Dem:CONFigure:S TRing1? :SPOR9|SPORT9|SB9|SP9:MO Dem:CONFigure:S TRing2 :SPOR9|SPORT9|SB9|SP9:MO Dem:CONFigure:S TRing2? :SPOR9|SPORT9|SB9|SP9:MO Dem:CONFigure:S TRing3 :SPOR9|SPORT9|SB9|SP9:MO Dem:CONFigure:S TRing3? ...

  • Agilent Technologies E6380-90073 - page 173

    Chapter 1 173 GPIB Command Dictionar y CONFigure subsys tem :SPOR9|SPORT9|SB9|SP9:MO Dem:CONFigure:E CEQivalent? This command repo rts the dec imal equivalent of the escape character entered with the CONF:SPOR9:MOD:CONF:ECH. Syntax CONF:SPOR9:MOD:CONF:ECEQ? Screen/field equivalent CONF:SPOR9:MOD:CONF:ECEQ? queries the Decimal Equivalent field on th ...

  • Agilent Technologies E6380-90073 - page 174

    174 Chapter 1 GPIB Command Dictionar y CONFigure subsystem :SPOR9|SPORT9|SB9|SP9:MO Dem:SECurity:P A SSword :SPOR9|SPORT9|SB9|SP9:MO Dem:SECurity:P A SSword? This command sets/ queries the p assword required by a security challenge when the CONF:SPOR9:MOD:SEC:ST A T ‘On’ command is used. Syntax CONF:SPOR9:MOD:SEC:PASS <string> CONF:SPOR9: ...

  • Agilent Technologies E6380-90073 - page 175

    Chapter 1 175 GPIB Command Dictionar y CONFigure subsys tem :SPOR9|SPORT9|SB9|SP9:ST A T us:MODem? This command querie s the modem st atus register . Syntax CONF:SPOR9:STAT:MOD? Screen/field equivalent No screen/ field equivalent. :SPOR10|SPORT10|SB10 |SP10:BAUD <string> :SPOR10|SPORT10|SB10 |SP10:BAUD? These commands set/query the baud rate ...

  • Agilent Technologies E6380-90073 - page 176

    176 Chapter 1 GPIB Command Dictionar y CONFigure subsystem :SPOR10|SPORT10|SB10 |SP10:DA T A <string> :SPOR10|SPORT10|SB10 |SP10:DA T A? These commands set/quer y the number of data bit s per data word sent over the SERIAL 10 por t. Syntax CONF:SPOR10:DATA ‘7 bits’ ‘8 bits’ CONF:SPOR10:DATA? Screen/field equivalent CONF:SPOR10:DA T A ...

  • Agilent Technologies E6380-90073 - page 177

    Chapter 1 177 GPIB Command Dictionar y CONFigure subsys tem :SPOR10|SPORT10|SB10 |SP10:FCONtrol|F LOW <string> :SPOR10|SPORT10|SB10 |SP10:FCONtrol|F LOW? These commands set /query the flow control to set the pace of serial communication s for the SERIAL 10 port. Syntax CONF:SPOR10:FCON ‘Xon/Xoff’ ‘None’ CONF:SPOR10:FCON? Screen/field ...

  • Agilent Technologies E6380-90073 - page 178

    178 Chapter 1 GPIB Command Dictionar y CONFigure subsystem :SPOR10|SPORT10|SB10 |SP10:MODem:MODE <string> :SPOR10|SPORT10|SB10 |SP10:MODem:MODE? These commands set/query the modem mode. Syntax CONF:SPOR10:MOD:MODE ‘Disable’ ‘Ignore’ ‘Answer’ ‘Dial Back’ CONF:SPOR10:MOD:MODE? !returns quoted string Screen/field equivalent CONF: ...

  • Agilent Technologies E6380-90073 - page 179

    Chapter 1 179 GPIB Command Dictionar y CONFigure subsys tem :SPOR10|SPORT10|SB10 |SP10:MODem:CALL:ST A T us? This command querie s the status of the mode m. Syntax CONF:SPOR10:MOD:CALL:STAT? Screen/field equivalent CONF:SPOR10:MOD:CALL:ST A T? queries the Status field on the MODEM CONFIGURE screen when the S erial Port fiel d is s et to Port 10 . : ...

  • Agilent Technologies E6380-90073 - page 180

    180 Chapter 1 GPIB Command Dictionar y CONFigure subsystem :SPOR10|SPORT10|SB10 |SP10:MODem:CONNect :HDELay :SPOR10|SPORT10|SB10 |SP10:MODem:CONNect :HDELay? This command sets/ queries the number of seconds to wait befo re attempting an originatio n. This applies to the delay pri or to attempting a dialback as well as the delay prior to r etrying a ...

  • Agilent Technologies E6380-90073 - page 181

    Chapter 1 181 GPIB Command Dictionar y CONFigure subsys tem :SPOR10|SPORT10|SB10 |SP10:MODem:CONFigur e:STRing1 :SPOR10|SPORT10|SB10 |SP10:MODem:CONFigur e:STRing1? :SPOR10|SPORT10|SB10 |SP10:MODem:CONFigur e:STRing2 :SPOR10|SPORT10|SB10 |SP10:MODem:CONFigur e:STRing2? :SPOR10|SPORT10|SB10 |SP10:MODem:CONFigur e:STRing3 :SPOR10|SPORT10|SB10 |SP10:M ...

  • Agilent Technologies E6380-90073 - page 182

    182 Chapter 1 GPIB Command Dictionar y CONFigure subsystem :SPOR10|SPORT10|SB10 |SP10:MODem:CONFigur e:ECEQivalent? This command repo rts the dec imal equivalent of the escape character entered with the CONF:SPOR10:MOD:CONF:ECH. Syntax CONF:SPOR10:MOD:CONF:ECEQ? Screen/field equivalent CONF:SPOR10:MOD:CONF:ECEQ? queries the Decimal Equ ivalent fiel ...

  • Agilent Technologies E6380-90073 - page 183

    Chapter 1 183 GPIB Command Dictionar y CONFigure subsys tem :SPOR10|SPORT10|SB10 |SP10:MODem:SECurit y:P ASSword :SPOR10|SPORT10|SB10 |SP10:MODem:SECurit y:P ASSword? This command sets/quer ies the password required by a security challenge when the CONF:SPOR10:MOD:SEC:ST A T ‘On’ command is used. Syntax CONF:SPOR10:MOD:SEC:PASS <string> C ...

  • Agilent Technologies E6380-90073 - page 184

    184 Chapter 1 GPIB Command Dictionar y CONFigure subsystem :SPOR1 1|SPORT1 1|SB1 1|SP1 1:P ARity <str ing> :SPOR1 1|SPORT1 1|SB1 1|SP1 1:P ARity? These commands set/query the parit y of the SERIAL 1 1 port. Syntax CONF:SPOR11:PAR ‘None’ ‘Odd’ ‘Even’ ‘Always 1’ ‘Always 0’ CONF:SPOR11:PAR? Screen/field equivalent CONF:SPOR1 ...

  • Agilent Technologies E6380-90073 - page 185

    Chapter 1 185 GPIB Command Dictionar y CONFigure subsys tem :SPOR1 1|SPORT1 1|SB1 1|SP1 1:STOP <str ing> :SPOR1 1|SPORT1 1|SB1 1|SP1 1:STOP? These commands set/quer y the number o f stop bits used for serial communication over the S ERIAL 1 1 port. Syntax CONF:SPOR11:STOP ‘1 bit’ ‘2 bits’ CONF:SPOR11:STOP? Screen/field equivalent CONF ...

  • Agilent Technologies E6380-90073 - page 186

    186 Chapter 1 GPIB Command Dictionar y CONFigure subsystem :SPOR1 1|SPORT1 1|SB1 1|SP1 1:ST A T us:MODem? This command querie s the modem status register . Syntax CONF:SPOR11:STAT:MOD? Screen/field equivalent No screen/ field equivalent. :SPOR1 1|SPORT1 1|SB1 1|SP1 1:MODem:MODE <strin g> :SPOR1 1|SPORT1 1|SB1 1|SP1 1:MODem:MODE? These command ...

  • Agilent Technologies E6380-90073 - page 187

    Chapter 1 187 GPIB Command Dictionar y CONFigure subsys tem :SPOR1 1|SPORT1 1|SB1 1|SP1 1:MODem:CALL:D ISConnect This command causes the modem to disconnect f rom the call in progress. It also cancels any pending call originati on retries (CONF:SPOR1 1:MOD:CONN:RETRy). This field is inactive when CONF:SPOR1 1:MOD:MODE ‘Disable’ command is used. ...

  • Agilent Technologies E6380-90073 - page 188

    188 Chapter 1 GPIB Command Dictionar y CONFigure subsystem :SPOR1 1|SPORT1 1|SB1 1|SP1 1:MODem:CONNec t:TIMeout :SPOR1 1|SPORT1 1|SB1 1|SP1 1:MODem:CONNec t:TIMeout? This command sets/ queries the number of seconds to wait befo re aborting a connec tio n attempt. This applies to both answer and originate ac tivities. Syntax CONF:SPOR11:MOD:CONN:TIM ...

  • Agilent Technologies E6380-90073 - page 189

    Chapter 1 189 GPIB Command Dictionar y CONFigure subsys tem :SPOR1 1|SPORT1 1|SB1 1|SP1 1:MODem:CONNe ct:RETRy :SPOR1 1|SPORT1 1|SB1 1|SP1 1:MODem:CONNe ct:RETRy? This command sets/queries the number of calls t hat will be tr ied before an ori gination attempt is terminated. The total number of tries will be the number entered in this field, plus o ...

  • Agilent Technologies E6380-90073 - page 190

    190 Chapter 1 GPIB Command Dictionar y CONFigure subsystem :SPOR1 1|SPORT1 1|SB1 1|SP1 1:MODem:CONFig ure:STRing1 :SPOR1 1|SPORT1 1|SB1 1|SP1 1:MODem:CONFig ure:STRing1? :SPOR1 1|SPORT1 1|SB1 1|SP1 1:MODem:CONFig ure:STRing2 :SPOR1 1|SPORT1 1|SB1 1|SP1 1:MODem:CONFig ure:STRing2? :SPOR1 1|SPORT1 1|SB1 1|SP1 1:MODem:CONFig ure:STRing3 :SPOR1 1|SPORT ...

  • Agilent Technologies E6380-90073 - page 191

    Chapter 1 191 GPIB Command Dictionar y CONFigure subsys tem :SPOR1 1|SPORT1 1|SB1 1|SP1 1:MODem:CONFig ure:ECEQivalent? This command repo rts the dec imal equivalent of the escape character entered with the CONF:SPOR1 1:MOD:CONF:ECH. Syntax CONF:SPOR11:MOD:CONF:ECEQ? Screen/field equivalent CONF:SPOR1 1:MOD:CONF:ECEQ? queries the Decimal Equ ivalen ...

  • Agilent Technologies E6380-90073 - page 192

    192 Chapter 1 GPIB Command Dictionar y CONFigure subsystem :SPOR1 1|SPORT1 1|SB1 1|SP1 1:MODem:SECur ity:P ASSword :SPOR1 1|SPORT1 1|SB1 1|SP1 1:MODem:SECur ity:P ASSword? This command sets/ queries the p assword required by a security challenge when the CONF:SPOR1 1:MOD:SEC:ST A T ‘On’ command is used. Syntax CONF:SPOR11:MOD:SEC:PASS <strin ...

  • Agilent Technologies E6380-90073 - page 193

    Chapter 1 193 GPIB Command Dictionar y CONFigure subsys tem :SRLocation <string> :SRLocation? These commands set/query the save and recall locations. Syntax CONF:SRL ‘INTERNAL’ ‘CARD’ ‘RAM’ CONF:SRL? Screen/field equivalent CON F :S R L co n t rols th e S ave/Recall field of the I/O CONFIGURE screen. :TIME <real number> :TIM ...

  • Agilent Technologies E6380-90073 - page 194

    194 Chapter 1 GPIB Command Dictionar y DECoder subsystem DECoder s ubs ystem :ARM:MODE <string> :ARM:MODE? These commands set/quer y the measurement armi ng mode. This command specifi es how long you want the analyzer to dec ode inc oming signals. • ‘Single’ t ells the analyzer to di splay the infor mation received during one gate t ime ...

  • Agilent Technologies E6380-90073 - page 195

    Chapter 1 195 GPIB Command Dictionar y DECoder subsys tem :LEV el:AM <real num ber> :LEV el:AM? These commands set/quer y the expect ed data signal level (aft er de-emp hasis, if u sed). For mor e infor mation about t he effect s of de-emphasi s, see the “De-emphasis and Input Level” de sc rip tion in the Referenc e Guide . The higher the ...

  • Agilent Technologies E6380-90073 - page 196

    196 Chapter 1 GPIB Command Dictionar y DECoder subsystem :LEV el:FM <real num ber> :LEV el:FM? These commands set/quer y the expect ed data signal level (aft er de-emp hasis, if u sed). For mor e infor mation about t he effect s of de-emphasi s, see the “De-emphasis and Input Level” in the Re feren ce Guide . The higher the level of signa ...

  • Agilent Technologies E6380-90073 - page 197

    Chapter 1 197 GPIB Command Dictionar y DECoder subsys tem :LEV el:VOL T s <real number> :LEV el:VOL T s ? These commands set/quer y the expect ed data signal level (aft er de-emp hasis, if u sed). For mor e infor mation about t he effect s of de-emphasi s, see the “De-emphasis and Input Level” in the Re feren ce Guide . The higher the lev ...

  • Agilent Technologies E6380-90073 - page 198

    198 Chapter 1 GPIB Command Dictionar y DECoder subsystem :POLarity <string> :POLarity? These commands match/qu ery the pol arity of the encod ed signal be ing analyzed. This f unct ion is helpful for restoring the proper d ata polarity when the transmitter , repeater , or receiver in you communicati ons system has an odd number of inversions. ...

  • Agilent Technologies E6380-90073 - page 199

    Chapter 1 199 GPIB Command Dictionar y DECoder subsys tem :AMPS|T ACS:GA TE <real number> :AMPS|T ACS:GA TE? This command specifies ho w long the de coder analyzes a signal after it has been trigger ed. Up to 6.55 seconds of gate time can be spe cified. The minimum gate time must be long enough to allow the preamble and all necessary data bit ...

  • Agilent Technologies E6380-90073 - page 200

    200 Chapter 1 GPIB Command Dictionar y DECoder subsystem :AMPS|T ACS:ST ANdard <string> :AMPS|T ACS:ST ANdard? These commands set/quer y the expect ed frame struct ure and channe l range of the decode r ’ s incoming sign al. T rying to r un a test with the wrong standard selected will result in incorrectly decod ed data, or will result in a ...

  • Agilent Technologies E6380-90073 - page 201

    Chapter 1 201 GPIB Command Dictionar y DECoder subsys tem :DTMF:GA TE <real number> :DTMF:GA TE? This command specifies ho w long the de coder analyzes a signal after it has been triggered. Up to 6.55 seconds of gate time can be specified. If the gat e time is too long, the dec oder ’ s buffer b ecomes full. A message is dis played inst ruc ...

  • Agilent Technologies E6380-90073 - page 202

    202 Chapter 1 GPIB Command Dictionar y DECoder subsystem :NAMPs|NT ACs:CHAN nel <string> :NAMPs|NT ACs:CHAN nel? These comman ds set/ query the typ e of da ta to d ecode. • ‘Cntl’ select s reverse control channel (RECC) data. • ‘V oice’ selects reverse voice channel (R VC) data. Syntax DEC:NAMP:CHAN ‘Cntl’ ‘Voice’ DEC:NAMP ...

  • Agilent Technologies E6380-90073 - page 203

    Chapter 1 203 GPIB Command Dictionar y DECoder subsys tem :NAMPs|NT ACs:DTMF :GA TE <real n umber> :NAMPs|NT ACs:DTMF :GA TE? These commands set/q uery how long the decod er analyzes a signal after it has been triggered . Up to 6.55 seconds of gate time can be specified. The minimum gate time must be long enough to allow the preamble and all ...

  • Agilent Technologies E6380-90073 - page 204

    204 Chapter 1 GPIB Command Dictionar y DECoder subsystem :NAMPs|NT ACs:ST ANdard <string > :NAMPs|NT ACs:ST ANdard? These commands set/quer y the expect ed frame struct ure and channe l range of the decode r ’ s incoming sign al. T rying to r un a test with the wrong standard selected will result in incorrectly decod ed data, or will result ...

  • Agilent Technologies E6380-90073 - page 205

    Chapter 1 205 GPIB Command Dictionar y DISPlay subsys tem DISPlay subs ystem The DISPlay subsystem cont rols the display of scre ens. A scree n should be displayed before a measurement from that screen is made. DISPlay DISPlay? DISP sets/queries which screen is displayed. Y ou use the DISP REM com ma n d to l oc k th e T e st Se t’s displa y . Sy ...

  • Agilent Technologies E6380-90073 - page 206

    206 Chapter 1 GPIB Command Dictionar y Integer Number Setting Syntax Integer Number Setting Syntax This syntax is for setting values of commands which require integer values. It is to be used with commands whi ch call for t he Integer Number Sett ing Syntax. An example of a command that requires the Integer Sett ing Syntax is the RF ANalyzer:RFCHan ...

  • Agilent Technologies E6380-90073 - page 207

    Chapter 1 207 GPIB Command Dictionar y I/Q Plot Subsystem I/Q Plot Subsystem CONT rols <string> :CONT rols? These commands set/query the string that matches the current field cont ents . Syntax IQPL:CONT ‘Main’ ‘Trigger’ IQPL:CONT? Screen/field equivalent IQPL:CONT? read s curre nt content s of Controls f ield on the I Q PLOT screen. ...

  • Agilent Technologies E6380-90073 - page 208

    208 Chapter 1 GPIB Command Dictionar y I/Q Plot Subsystem :MEASurement:MODE <string> :MEASurement:MODE? Thes e co mma n d s se t/ q u e ry th e cu r re n t measu re m ent mod e fo r the I/Q plot screen . Syntax IQPL:MEAS:MODE ‘Single’ ‘Cont’ IQPL:MEAS:MODE? Screen/field equivalent IQPL:MEAS:MODE reads the Measur ement fi eld of t he M ...

  • Agilent Technologies E6380-90073 - page 209

    Chapter 1 209 GPIB Command Dictionar y I/Q Plot Subsystem :TRIGger:DARM This command disarms the IQ Plot measurement when it has been put into single t riggering mode using t he TRIG:MODE:RETR SING. See “TRIGger subsystem” on page 346 for a more complete discussion of GPIB triggering. Syntax IQPL:TRIG:DARM !disarms the IQ Plot measurements. Scr ...

  • Agilent Technologies E6380-90073 - page 210

    210 Chapter 1 GPIB Command Dictionar y I/Q Plot Subsystem :TRIGger:QUALifier <s tring> :TRIGger:QUALifier? These commands set/quer y the qualifying event for the trigger in the IQ plot screen. The qualif ying event is the event that must occur before a trigger is accepted. The trigger event is set by the IQPL:TRIG:EVEN command. Syntax IQPL:TR ...

  • Agilent Technologies E6380-90073 - page 211

    Chapter 1 211 GPIB Command Dictionar y MEASure subsys tem MEASure subsystem The MEAS subsystem has a set of unique commands. These are the measurement command < meas cmnd> set. They contro l features of the T es t Set such as setting measurement limits, units and the meters. See “Number Measurement Syntax” on page 283 for more details abo ...

  • Agilent Technologies E6380-90073 - page 212

    212 Chapter 1 GPIB Command Dictionar y MEASure subsyste m :AFRequency:AM <me as cmnd> :AFRequency:AM? These commands set/quer y the AM depth measurement. T o use this measurement you must selec t eithe r the AM Mod or AM Demod audio source ( AF AN :INP). Syntax MEAS:AFR:AM <meas cmd> !See “Number Measurement Syntax” on page 283. MEA ...

  • Agilent Technologies E6380-90073 - page 213

    Chapter 1 213 GPIB Command Dictionar y MEASure subsys tem :AFRequency:DCFM <meas cm nd> :AFRequency:DCFM? These commands set/quer y the DC FM measurement. T o use this measurement, you must have the DC level meas urement select ed (see MEAS:AFR:SEL on page 216), and you must be using the FM Demod or FM Mod a udio s ource (s ee AF AN:I NP on p ...

  • Agilent Technologies E6380-90073 - page 214

    214 Chapter 1 GPIB Command Dictionar y MEASure subsyste m :AFRequency:DCV olts <meas cmnd> :AFRequency:DCV olts? These commands set/query t he DC voltmeter . T o use this measurement, you must have the DC leve l measurement selecte d (see MEAS:AFR:SEL on page 216), and have select ed one of the following audio sourc es: SSB Demod, Audio In, E ...

  • Agilent Technologies E6380-90073 - page 215

    Chapter 1 215 GPIB Command Dictionar y MEASure subsys tem :AFRequency:FM <me as cmnd> :AFRequency:FM? These commands set/quer y the FM deviation measurement. This measurement is active when FM Mod or FM Demod is selected by using the AF AN:INP command ( see page 59) and SNR is not selec ted by the MEAS:AFR:SEL command (see p age 216). Syntax ...

  • Agilent Technologies E6380-90073 - page 216

    216 Chapter 1 GPIB Command Dictionar y MEASure subsyste m :AFRequency:SELect <meas c mnd> :AFRequency:SELect? These commands set/quer y which audio fr equency measurement is displayed on the AF ANAL YZER screen.These measurements are also displayed on the RF GENERA TOR and RF ANAL YZER screens. Syntax MEAS:AFR:SEL ‘SINAD’ ‘Distn’ ‘S ...

  • Agilent Technologies E6380-90073 - page 217

    Chapter 1 217 GPIB Command Dictionar y MEASure subsys tem :AFRequency:SNR <m eas cmnd> :AFRequency:SNR? These commands set/quer y the SNR measurement. SNR must first be selected using the MEAS:AFR:SEL ‘SNR’ command (see page 216). Syntax MEAS:AFR:SNR <meas cmd> !controls the SNR command MEAS:AFR:SNR? !returns the measurement’s val ...

  • Agilent Technologies E6380-90073 - page 218

    218 Chapter 1 GPIB Command Dictionar y MEASure subsyste m :CANalyzer:ADC:EVMadc :LEV el <meas cmnd> :CANalyzer:ADC:EVMadc :LEV el? These commands set/query t he ADC for t he EVM measurement. The ADC is a measure of how close the actual inp ut signal is to the maximum input l evel. This measurement util izes the Number Measurement Synt ax, but ...

  • Agilent Technologies E6380-90073 - page 219

    Chapter 1 219 GPIB Command Dictionar y MEASure subsys tem :CANalyzer:ADC:TPOW eradc:LEV el <m eas cmnd> :CANalyzer:ADC:TPOW eradc:LEV el? These commands set/quer y the ADC for the channel power measurement. The ADC is a me asure of how close the actual input signal is to the maximum input level. This measurement utili zes the Number Measureme ...

  • Agilent Technologies E6380-90073 - page 220

    220 Chapter 1 GPIB Command Dictionar y MEASure subsyste m :CANalyzer:ERRor:FREQ uency <meas cmnd> :CANalyzer:ERRor:FREQ uency? These commands set/query the frequency err or measur ement. This measurement util izes the Number Measurement Synt ax, but do es not use the :MET er command. Syntax MEAS:CAN:ERR:FREQ <meas cmd> !See “Number Me ...

  • Agilent Technologies E6380-90073 - page 221

    Chapter 1 221 GPIB Command Dictionar y MEASure subsys tem :CANalyzer:ERRor:PHAS e <meas cmnd> :CANalyzer:ERRor:PHAS e? These commands set/quer y the phase erro r measurement. This measurement utili zes the Number Measurement Synt ax, but do es not use the :MET er command. Syntax MEAS:CAN:ERR:PHAS <meas cmd> !See “Number Measurement Sy ...

  • Agilent Technologies E6380-90073 - page 222

    222 Chapter 1 GPIB Command Dictionar y MEASure subsyste m :CANalyzer:PNOFfset? This command queries the PN offset measur ement. Syntax MEAS:CAN:PNOF? !returns the measurement Screen/field equivalent MEAS:CAN:PNOF? reads the PN Offset measurement fie ld on the CDMA ANAL YZER screen. This me asurement is also displayed on the CDMA GENERA TOR s cree n ...

  • Agilent Technologies E6380-90073 - page 223

    Chapter 1 223 GPIB Command Dictionar y MEASure subsys tem :CANalyzer:POW er:ACP:URA Tio <meas cmnd> :CANalyzer:POW er:ACP:URA Tio? These commands set/query up per ratio measurement of t he adjacent channel power mea s urement. This ratio is the rati o of the total power in the specif ied bandwidth at the specif ied offset above the cente r fr ...

  • Agilent Technologies E6380-90073 - page 224

    224 Chapter 1 GPIB Command Dictionar y MEASure subsyste m :CANalyzer:POW er:A VG <mea s cmnd> :CANalyzer:POW er:A VG? These commands set/quer y the average power measurement. This measurement util izes the Number Measurement Synt ax, but do es not use the :MET er command. Syntax MEAS:CAN:POW:AVG <meas cmd> !See “Number Measurement Syn ...

  • Agilent Technologies E6380-90073 - page 225

    Chapter 1 225 GPIB Command Dictionar y MEASure subsys tem :CANalyzer:RHO <me as cmnd> :CANalyzer:RHO? These commands set/query t he rho measur ement. This measurement utili zes the Number Measurement Synt ax, but do es not use the :MET er command. Syntax MEAS:CAN:RHO <meas cmd> !See “Number Measurement Syntax” on page 283. MEAS:CAN: ...

  • Agilent Technologies E6380-90073 - page 226

    226 Chapter 1 GPIB Command Dictionar y MEASure subsyste m :CDANalyzer:ADC <m eas cmnd> :CDANalyzer:ADC? IS-95 Only mode. These commands set/query the ADC measurement in the code domain analyzer . This measurement util izes the Number Measurement Synt ax, but do es not use the :MET er command. Syntax MEAS:CDAN:ADC <meas cmd> !See “Numb ...

  • Agilent Technologies E6380-90073 - page 227

    Chapter 1 227 GPIB Command Dictionar y MEASure subsys tem :CDANalyzer:CPOW er <meas c mnd> :CDANanlyzer:CPOW er? IS-95 Only mode. These command set/query the channel power measurement. This measurement utili zes the Number Measurement Synt ax, but do es not use the :MET er command. Syntax MEAS:CDAN:CPOW <meas cmd> !See “Number Measure ...

  • Agilent Technologies E6380-90073 - page 228

    228 Chapter 1 GPIB Command Dictionar y MEASure subsyste m :CDANalyzer:ERHo? IS-95 Only mode. This command queries the estimat ed rho measurement. Estimated rho is an approximation of rho that is calculated without taking the system out of service. Syntax MEAS:CDAN:ERH? Screen/field equivalent MEAS:CDAN:ERH? queries the Est Rho field CODE DOM screen ...

  • Agilent Technologies E6380-90073 - page 229

    Chapter 1 229 GPIB Command Dictionar y MEASure subsys tem :CDANalyzer:MARKer:LE V el <meas cmnd> :CDANalyzer:MARKer:LE V el? IS-95 Only mode. These commands set/query the level at th e marker on the code domain analyzer ’ s trace during power or fast power measurements (see :CDAN:MEAS on page 124). This m easurement is made relative to the ...

  • Agilent Technologies E6380-90073 - page 230

    230 Chapter 1 GPIB Command Dictionar y MEASure subsyste m :CDANalyzer:MARKer:PH ASe <meas cmnd> :CDANalyzer:MARKer:PH ASe? IS-95 Only mode. These commands set/query the phase at the marker on the code domai n analyze r ’ s trace during phase measurements (see :CDAN:MEAS on page 124).This measuremen t is made rel ative to the value at W alsh ...

  • Agilent Technologies E6380-90073 - page 231

    Chapter 1 231 GPIB Command Dictionar y MEASure subsys tem :CDANalyzer:MARKer:TI ME <meas cmnd> :CDANalyzer:MARKer:TI ME? IS-95 Only mode. Thes e commands set/query the time at the mark er on the code domain analyze’ s trace during a timing measurement (see CDAN :MEAS on p age 124) . Th is meas urem ent is made relat ive to the value at W al ...

  • Agilent Technologies E6380-90073 - page 232

    232 Chapter 1 GPIB Command Dictionar y MEASure subsyste m :CDANalyzer:TIME:OFFSe t <meas cmnd> :CDANalyzer:TIME:OFFSe t? IS-95 Only mode. These commands set/query the time offset measurement of t he code domain analyzer . This measurement util izes the Number Measurement Synt ax, but do es not use the :MET er command. Syntax MEAS:CDAN:TIME:OF ...

  • Agilent Technologies E6380-90073 - page 233

    Chapter 1 233 GPIB Command Dictionar y MEASure subsys tem :CDANalyzer:TRACe:TOF Fset? IS-95 Only mode. This command queries the timin g in the 64 trace points of the co de domain analyzer . The measurement returns 64 values (in seconds), one for each point . Syntax MEAS:CDAN:TRAC:TOFF? Screen/field equivalent MEAS:CDAN:TRAC:TOFF? does not correspon ...

  • Agilent Technologies E6380-90073 - page 234

    234 Chapter 1 GPIB Command Dictionar y MEASure subsyste m :CDANalyzer:IS2000:ADC <meas cmds> :CDANalyzer:IS2000:ADC ? IS-2000 only . These commands set/query t he ADC measurement in t he Code Domain analyzer . This measurement util izes the Number Measurement Synt ax, but do es not use the :MET er command. Syntax MEAS:CDAN:IS2000:ADC <meas ...

  • Agilent Technologies E6380-90073 - page 235

    Chapter 1 235 GPIB Command Dictionar y MEASure subsys tem :CDANalyzer:IS2000:CHA Nnel[:POW er]? IS-2000 only . This query returns the c hannel power value (when active ) with th e units d Bm. Syntax MEAS:CDAN:IS2000:CHAN[:POW]? Screen/field equivalent MEAS:CDAN:IS2000:CHAN[:POW]? reads the C han Power measurement f ield from the Reference menu when ...

  • Agilent Technologies E6380-90073 - page 236

    236 Chapter 1 GPIB Command Dictionar y MEASure subsyste m :CDANalyzer:IS2000:MAR Ker:CHANnel:RA TE:RC3? IS-2000 only . This query returns t he RC-3 data rate value for the channel select ed by the Marke r ’ s current position. The retur n value is an integer . Syntax MEAS:CDAN:IS2000:MARK:CHAN:RATE:RC3? Screen/field equivalent MEAS:CDAN:IS2000:MA ...

  • Agilent Technologies E6380-90073 - page 237

    Chapter 1 237 GPIB Command Dictionar y MEASure subsys tem :CDANalyzer:IS2000:MAR Ker:CHANnel:RA TE:RC5? IS-2000 only . This query returns t he RC-5 data rate value for the channel select ed by the Marke r ’ s current position. The retur n value is an integer . Syntax MEAS:CDAN:IS2000:MARK:CHAN:RATE:RC5? Screen/field equivalent MEAS:CDAN:IS2000:MA ...

  • Agilent Technologies E6380-90073 - page 238

    238 Chapter 1 GPIB Command Dictionar y MEASure subsyste m :CDANalyzer:IS2000:MAR Ker:CHANnel[:W A LSh]:NUMBer? IS-2000 only . This query returns the parenthetical W alsh channel number for the the channel select ed by the marker ’ s cur rent position. The return value is an integer . Syntax MEAS:CDAN:IS2000:MARK:CHAN:WALS:ORD? Screen/field equiva ...

  • Agilent Technologies E6380-90073 - page 239

    Chapter 1 239 GPIB Command Dictionar y MEASure subsys tem :CDANalyzer:IS2000:MAR Ker:COMPlex:I[:LEV el]? IS-2000 only . This query returns t he Marker Level f or the I channel (T op bar), when the Complex Power Measurement is active. The return is a floating point number . Syntax MEAS:CDAN:IS2000:MARK:COMP:I[:LEV]? Screen/field equivalent MEAS:CDAN ...

  • Agilent Technologies E6380-90073 - page 240

    240 Chapter 1 GPIB Command Dictionar y MEASure subsyste m :CDANalyzer:IS2000:MAR Ker:NOISe? IS-2000 only . This query returns the Marker Level for the Noise, when the Power & Noise measur ement is active. The r eturn will be a flo ating point number . Syntax MEAS:CDAN:IS2000:MARK:NOIS? Screen/field equivalent MEAS:CDAN:IS2000:MARK:NOIS? reads t ...

  • Agilent Technologies E6380-90073 - page 241

    Chapter 1 241 GPIB Command Dictionar y MEASure subsys tem :CDANalyzer:IS2000:RHO [:EST im ated]? IS-2000 only . This query returns t he Estimated Rho whe n present (not available in F AST Power Measurement). The returned value is a floating po int nu mber . Estimated rh o is an app roximation of rho that is calculated when there are more signals th ...

  • Agilent Technologies E6380-90073 - page 242

    242 Chapter 1 GPIB Command Dictionar y MEASure subsyste m :CDANalyzer:IS2000:TRA Ce:ABSolute:POW er[:BASi c][:V ALue]? IS-2000 only . This query returns an array of 1 28 values represent ing the true (absolute) power of t he W alsh channel on which the Marker is curre ntly pos itioned. NO TE The value return ed represents the absolute po wer of the ...

  • Agilent Technologies E6380-90073 - page 243

    Chapter 1 243 GPIB Command Dictionar y MEASure subsys tem :CDANalyzer:IS2000:TRA Ce:ABSolute:POW er[:BASi c]:ALL? IS-2000 only . This query returns the information necessary to rec reate the Code Domain trace when making a Power , Fast Power , or Fast Power Synchroni ze measurement. The returned e ntity con sists of 128 four -value reco rds for eac ...

  • Agilent Technologies E6380-90073 - page 244

    244 Chapter 1 GPIB Command Dictionar y MEASure subsyste m : CDANalyzer :IS2000:TRACe:A BSolute:POW er:COMPlex:I:AC Ti ve? :CDANal yzer:IS2000:TRACe:ABSolute: POW er:COMPlex:COMBined:A CTi ve? :CDANal yzer:IS2000:TRACe:ABSolute: POW er:COMPlex:QACT ive? :CDANal yzer:IS2000:TRACe:ABSolute: POW er:COMPlex:IS2000: ACT ive? IS-2000 only . These queries ...

  • Agilent Technologies E6380-90073 - page 245

    Chapter 1 245 GPIB Command Dictionar y MEASure subsys tem :CDANalyzer:IS 2000:TRACe:ABS olute:POW er:COMPlex:I :ALL? :CDANalyze r:IS2000:TRACe: ABSolute:POW er:COMPlex:COMBined:ALL? :CDANalyzer:IS2000:TRACe:ABSolute:POW er:COMPlex:Q:ALL? :CDANalyzer:IS 2000:TRACe:ABS olute:POW er:COMPlex:I S2000:ALL? IS-2000 only . These queries return the infor ma ...

  • Agilent Technologies E6380-90073 - page 246

    246 Chapter 1 GPIB Command Dictionar y MEASure subsyste m :CDANal yzer:IS2000:TRACe:ABSolute: POW er:COMPlex:I[:V ALue]? :CDANal yzer:IS2000:TRACe:ABSolute: POW er:COMPlex:Q[: V ALue]? :CDANal yzer:IS2000:TRACe:ABSolute: POW er:COMPlex:COMBined[:V ALue]? :CDANal yzer:IS2000:TRACe:ABSolute: POW er:COMPlex:IS2000[:V ALue]? IS-2000 only . These querie ...

  • Agilent Technologies E6380-90073 - page 247

    Chapter 1 247 GPIB Command Dictionar y MEASure subsys tem :CDANalyzer:IS2000:TRA Ce:ABSolute:POW er:NOISe:ACTive? IS-2000 only . This query returns active channel information for the Code Domain trace when making a Power and Noise measurement. The return enti ty cons ists of an integer repre senting the numb er of active channels, fo llowed by a f ...

  • Agilent Technologies E6380-90073 - page 248

    248 Chapter 1 GPIB Command Dictionar y MEASure subsyste m :CDANalyzer:IS2000:TRA Ce:ABSolute:POW er:NOISe:ALL? IS-2000 only . This query returns the information necessary to rec reate the Code Domain trace when making a Power and No ise measurement. The return entity consists of 128 four value rec ord s for each bar in the trace. The four values fo ...

  • Agilent Technologies E6380-90073 - page 249

    Chapter 1 249 GPIB Command Dictionar y MEASure subsys tem :CDANalyzer:IS2000:TRA Ce:ABSolute:POW er:POW er:ACTive? IS-2000 only . This query returns active channel information for the Code Domain trace when making a Power and Noise measurement. The return enti ty cons ists of an integer repre senting the numb er of active channels followed by a fou ...

  • Agilent Technologies E6380-90073 - page 250

    250 Chapter 1 GPIB Command Dictionar y MEASure subsyste m :CDANalyzer:IS2000:TRA Ce:ABSolute:POW er:POW er:ALL? IS-2000 only . This query returns the information necessary to rec reate the Code Domain trace when making a Power and No ise measurement. The return entity consists of 128 four value rec ord s for each bar in the trace. The four values f ...

  • Agilent Technologies E6380-90073 - page 251

    Chapter 1 251 GPIB Command Dictionar y MEASure subsys tem :CDANalyzer:IS2000:TRA Ce:POW er[:BASic]:ACTive? IS-2000 only . This query retu rns the ac tive channel informati on for the Code Domain trace when making a Power , Fast Power or Fast Power Synchroni ze measurement. The ret urn entity consists of an intege r represen ting the number of activ ...

  • Agilent Technologies E6380-90073 - page 252

    252 Chapter 1 GPIB Command Dictionar y MEASure subsyste m :CDANalyzer:IS2000:TRA Ce:POW er[:BASic]:ALL? IS-2000 only . This query returns the information necessary to rec reate the Code Domain trace when making a Po wer , Fast Power or Fast Power Synchronize measure ment. The return consi sts of 128 four value records for each bar in the trace . Th ...

  • Agilent Technologies E6380-90073 - page 253

    Chapter 1 253 GPIB Command Dictionar y MEASure subsys tem :CDANalyzer:IS2000:TRA Ce:POW er:COMPlex:I:ACTive? :CDANalyzer:IS2000:TRA Ce:POW er:COMPlex:COMBi ned:ACT ive? :CDANalyzer:IS2000:TRA Ce:POW er:COMPlex:QACTive? :CDANalyzer:IS2000:TRA Ce:POW er:COMPlex:IS2000A CT ive? IS-2000 only . These queries return the active channe l information for th ...

  • Agilent Technologies E6380-90073 - page 254

    254 Chapter 1 GPIB Command Dictionar y MEASure subsyste m :CDANalyzer:IS2000:TRA Ce:POW er:COMPlex:I:ALL? :CDANalyzer:IS2000:TRA Ce:POW er:COMPlex:COMBi ned:ALL? :CDANalyzer:IS2000:TRA Ce:POW er:COMPlex:Q:ALL? :CDANalyzer:IS2000:TRA Ce:POW er:COMPlex:IS2000:ALL? IS-2000 only . These queries return the infor mat i on necessary to recreate the Code D ...

  • Agilent Technologies E6380-90073 - page 255

    Chapter 1 255 GPIB Command Dictionar y MEASure subsys tem :CDANalyzer:IS2000:TRA Ce:POW er:COMPlex:I[:V ALu e]? :CDANalyzer:IS2000:TRA Ce:POW er:COMPlex:COMBi ned[:V ALue]? IS-2000 only . This query returns an array of 128 valu es that r epresents the Spreade r I Input Power of both IS-95 and IS-2000 channe ls relative to the select ed reference po ...

  • Agilent Technologies E6380-90073 - page 256

    256 Chapter 1 GPIB Command Dictionar y MEASure subsyste m :CDANalyzer:IS2000:TRA Ce:POW er:COMPlex:Q[:V AL ue]? :CDANalyzer:IS2000:TRA Ce:POW er:COMPlex:IS2000? IS-2000 only . These commands query and return an array of 128 values that repr esents Sp reader Inpu t Q P ower rela tive to th e se lected reference channel power . NO TE The values retur ...

  • Agilent Technologies E6380-90073 - page 257

    Chapter 1 257 GPIB Command Dictionar y MEASure subsys tem :CDANalyzer:IS2000:TRA Ce:POW er:NOISe:ALL? IS-2000 only . This query returns the information necessary to rec reate the Code Domain trace when making a Power and No ise measurement. The return consists of 128 four value records for each bar in the trace. The four values for each recor d are ...

  • Agilent Technologies E6380-90073 - page 258

    258 Chapter 1 GPIB Command Dictionar y MEASure subsyste m :CDANalyzer:IS2000:TRA Ce:POW er:POW er:ACT ive? IS-2000 only . This query returns the active channel info rmation for theCode Domain trace when making a Power and No ise measurement. The return value consists of an integer repr ese nting the number of active channels fo llowed by a four val ...

  • Agilent Technologies E6380-90073 - page 259

    Chapter 1 259 GPIB Command Dictionar y MEASure subsys tem :CDANalyzer:IS2000:TRA Ce:POW er:POW er[:V ALue]? IS-2000 only . This query returns an array of 1 28 values that r epresent the channel po wer relati ve to t he selected ref ere nce channel power in a Power & Noise measurement . NO TE The values returned repre se nt the relative power of ...

  • Agilent Technologies E6380-90073 - page 260

    260 Chapter 1 GPIB Command Dictionar y MEASure subsyste m :DEC:AMPS|T ACS:CDA T a? This command queries the decod ed data for the control channel . Syntax MEAS:DEC:AMPS:CDAT? Example MEAS:DEC:AMPS:CDAT? !returns a quoted string reporting the decoded data obtained from the control channel measured by the signaling unit. For example, the Signaling De ...

  • Agilent Technologies E6380-90073 - page 261

    Chapter 1 261 GPIB Command Dictionar y MEASure subsys tem :DEC:AMPS|T ACS:DA T A? This command queries the decod ed dat a for the voice channel. Syntax MEAS:DEC:AMPS:DATA? Example MEAS:DEC:AMPS:DATA? !returns a quoted string reporting the decoded data obtained from the voice channel measured by the signaling unit. For example, the Signaling Decoder ...

  • Agilent Technologies E6380-90073 - page 262

    262 Chapter 1 GPIB Command Dictionar y MEASure subsyste m :DEC:DTMF:LOW :FREQuency:ERRo r <meas cmnd> :DEC:DTMF:LOW :FREQuency:ERRo r? These commands set/quer y the freq uency error of the lo w tone in th e DTMF pair . The MEAS:DEC:DTMF:LOW: FREQ:DISP ‘Freq Err ’ command must be used to di splay the fr equency error of the low tone. Synta ...

  • Agilent Technologies E6380-90073 - page 263

    Chapter 1 263 GPIB Command Dictionar y MEASure subsys tem :DEC:DTMF:HIGH :FREQuency:ABSolute <meas cmnd> :DEC:DTMF:HIGH :FREQuency:ABSolute? These command set/query the fr equency of the high tone in the DTMF pair . The MEAS:DEC:DTMF:HIGH:FREQ:DISP ‘Freq’ command m ust be used to display the absolute frequen cy of the high tone. Syntax ME ...

  • Agilent Technologies E6380-90073 - page 264

    264 Chapter 1 GPIB Command Dictionar y MEASure subsyste m :DEC:DTMF:HIGH :FREQuency:DISPlay <string> :DEC:DTMF:HIGH :FREQuency:DISPlay? These command set/query the displ ay mode of t he high tone in the DTMF pa ir . Syntax MEAS:DEC:DTMF:HIGH:FREQ:DISP ‘Freq’ ‘Frq Err’ MEAS:DEC:DTMF:HIGH:FREQ:ERR? Screen/field equivalent MEAS:DEC:DTMF: ...

  • Agilent Technologies E6380-90073 - page 265

    Chapter 1 265 GPIB Command Dictionar y MEASure subsys tem :DEC:DTMF:SYM Bol? These command set/q uery the symbol number assigned by the encoder for each of the DTMF tones. As each tone is analyzed, the symbol that represen ts each tone is returned as a quoted st ring. Syntax MEAS:DEC:DTMF:SYMB? !returns a quoted string Screen/field equivalent MEAS: ...

  • Agilent Technologies E6380-90073 - page 266

    266 Chapter 1 GPIB Command Dictionar y MEASure subsyste m :DEC:NAMPs|NT ACs:RECC:DA T A? This command reads the decoded RECC data, serially , as it received. Syntax MEAS:DEC:NAMP:RECC:DATA? !returns a quoted string Screen/field equivalent MEAS:DEC:NAMP:CDA T? reads the RECC Data (hex) measurement field on the SIGNALING DECODER screen when the Mode ...

  • Agilent Technologies E6380-90073 - page 267

    Chapter 1 267 GPIB Command Dictionar y MEASure subsys tem :DEC:NAMP:DTM F:LOW :FREQuency: ABSolute <meas cmnd> :DEC:NAMP:DTM F:LOW :FREQuency: ABSolute? These commands set/query the frequency of the low tone in the DTMF pair . The MEAS:DEC:NAMP:DTMF:LOW :DISP ‘Freq’ command must be used to display the absolute frequen cy of the low tone. ...

  • Agilent Technologies E6380-90073 - page 268

    268 Chapter 1 GPIB Command Dictionar y MEASure subsyste m :DEC:NAMP:DTMF:LOW:DISPlay <string> :DEC:NAMP:DTMF:LOW:DISPlay? These commands set/quer y the display mode of the low to ne in the DTMF pa ir . Syntax MEAS:DEC:NAMP:DTMF:LOW:DISP ‘Freq’ ‘Frq Err’ MEAS:DEC:NAMP:DTMF:LOW:ERR? Screen/field equivalent MEAS:DEC:NAMP:DTMF:LOW :DISP c ...

  • Agilent Technologies E6380-90073 - page 269

    Chapter 1 269 GPIB Command Dictionar y MEASure subsys tem :DEC:NAMP:DTM F:HIGH:FREQuency:ERRor <meas cmnd> :DEC:NAMP:DTM F:HIGH:FREQuency:ERRor? These commands set/query the frequency err or of the high tone in the DTMF pair . The MEAS:DEC:NAMP:DTMF :HIGH:DISP ‘Frq Err ’ command must be used to di splay the fr equency error of the high to ...

  • Agilent Technologies E6380-90073 - page 270

    270 Chapter 1 GPIB Command Dictionar y MEASure subsyste m :DEC:NAMP:DTMF:TIME:ON <meas cmnd> :DEC:NAMP:DTMF:TIME:ON ? These commands set/quer y the timing of t he DTMF frequencies. Syntax MEAS:DEC:NAMP:DTMF:TIME:ON !See “Multiple Real Number Setting Syntax” on page 282 MEAS:DEC:NAMP:DTMF:TIME:ON? !returns up to 19 real values Screen/field ...

  • Agilent Technologies E6380-90073 - page 271

    Chapter 1 271 GPIB Command Dictionar y MEASure subsys tem :DEC:NAMP:DTM F:SYMBol? This command querie s the symbol number assigned by the enc oder for each of the DTMF tone s. As each tone is analyzed, t he symbol that represen ts each tone is returned as a quoted st ring. Syntax MEAS:DEC:NAMP:DTMF:SYMB? !returns a quoted string Screen/field equiva ...

  • Agilent Technologies E6380-90073 - page 272

    272 Chapter 1 GPIB Command Dictionar y MEASure subsyste m :OSCilloscope:MARKer: LEV el:AM <me as cmnd> :OSCilloscope:MARKer: LEV el:AM? These commands set/query the AM depth measurement at the oscilloscope’ s marker positi on. For this measurement to be valid, you must ha ve AM Mod or AM Dem od chosen as the a udio input (see AF AN:INP on p ...

  • Agilent Technologies E6380-90073 - page 273

    Chapter 1 273 GPIB Command Dictionar y MEASure subsys tem :OSCilloscope:MARKer: LEV el:FM <me as cmnd> :OSCilloscope:MARKer: LEV el:FM? These commands set/query the FM deviation measurement at the oscilloscope’ s marker positi on. For this measurement to be valid, you must have FM Mod or FM Demod cho sen as the audio input (see AF AN:INP on ...

  • Agilent Technologies E6380-90073 - page 274

    274 Chapter 1 GPIB Command Dictionar y MEASure subsyste m :OSCilloscope:MARKer: LEV el:VOL T s <meas c mnd> :OSCilloscope:MARKer: LEV el:VOL T s? These commands set/query the voltage measurement at the oscilloscope’ s marker positi on. For this measurement to be valid, you must have Au dio In , Audio Out, E xt Mo d, or S SB Demo d chos en a ...

  • Agilent Technologies E6380-90073 - page 275

    Chapter 1 275 GPIB Command Dictionar y MEASure subsys tem :OSCilloscope:TRACe? This command retur ns an arra y of 417 re al values, corr esponding to points on the os cil loscope’ s displa y . 0 corresponds to the fi rst valu e (lef t side of t race display) and 416 is the last value ( right side of the tra ce display). Syntax MEAS:OSC:TRAC? Scre ...

  • Agilent Technologies E6380-90073 - page 276

    276 Chapter 1 GPIB Command Dictionar y MEASure subsyste m :RFRequency:FREQuency :ABSolute <meas cmn d> :RFRequency:FREQuency :ABSolute? These commands set/quer y the RF frequenc y measurement when the :RFR:SEL ‘Frequency’ co mmand is used. Syntax :RFR:FREQ:ABS <meas syntax> !See “Number Measurement Syntax” on page 283. :RFR:FREQ ...

  • Agilent Technologies E6380-90073 - page 277

    Chapter 1 277 GPIB Command Dictionar y MEASure subsys tem :RFRequency:POW er <meas c mnd> :RFRequency:POW er? These commands set/quer y the transmitter power measurement. The measurement is eithe r peak or sampled, as determined by the RF AN:PME:DET command on page 305. Syntax MEAS:RFR:POW <meas cmd> !See “Number Measurement Syntax” ...

  • Agilent Technologies E6380-90073 - page 278

    278 Chapter 1 GPIB Command Dictionar y MEASure subsyste m :SANalyzer:MARKer:DEL T a:FREQuency <meas cmnd> :SANalyzer:MARKer:DEL T a:FREQuency ? These commands set/quer y the delta marker frequency measurement. This measurement is the frequency of the del ta marker minus the frequency of the normal marker . If the delta marker is to the right ...

  • Agilent Technologies E6380-90073 - page 279

    Chapter 1 279 GPIB Command Dictionar y MEASure subsys tem :SANalyzer:MARKer:DEL T a:LEV el <meas cm nd> :SANalyzer:MARKer:DEL T a:LEV el? These commands set/quer y the delta marker level measurement. This measurement is the level of t he normal marker minus t he frequency of the delta marker in dBm. If the delta marker is higher than normal m ...

  • Agilent Technologies E6380-90073 - page 280

    280 Chapter 1 GPIB Command Dictionar y MEASure subsyste m :SANalyzer:MARKer[:NO RMal]:FREQuency <me as cmnd> :SANalyzer:MARKer[:NO RMal]:FREQuency? These commands set/query the freq uency at the marker on the spectrum analyzer ’ s trace. This measurement util izes the Number Measurement Synt ax, but do es not use the :MET er command. Syntax ...

  • Agilent Technologies E6380-90073 - page 281

    Chapter 1 281 GPIB Command Dictionar y MEASure subsys tem :SANalyzer:TRACe? This command retur ns an arra y of 417 re al values, corr esponding to points on the spectrum analyzer ’ s display . 0 corr esponds to the first value (left side of trace displa y) and 416 is the last value (right side of the trace d isplay). This measurement utili zes th ...

  • Agilent Technologies E6380-90073 - page 282

    282 Chapter 1 GPIB Command Dictionar y Multiple Real Number Setting S yntax Multiple Rea l Number Setting Sy ntax This section defines the syntax to be used with commands that require the Real N umber Setting Syntax, but allow multi ple numbers to be set in one command. This format is typically used for entering values that are grouped in rows or c ...

  • Agilent Technologies E6380-90073 - page 283

    Chapter 1 283 GPIB Command Dictionar y Number Measurement Syntax Number Measurement Syntax This syntax is used with measurement commands. It applies to both real and integer values. This syntax is typically used with subsystems like the MEASure sub syste m/ Syntax :Previous Syntax:AUNits <GPIB units> ! sets the GPIB units :Previous Syntax:AUN ...

  • Agilent Technologies E6380-90073 - page 284

    284 Chapter 1 GPIB Command Dictionar y Number Measuremen t Syntax :Previous Syntax:LLIMit:STATe 1|ON !se ts state of the limit :Previous Syntax:LLIMit:STATe 0|OFF !s ets state of the limit :Previous Syntax:LLIMit:STATe? !return s 1 or 0 :Previous Syntax:METer[:STATe] 1|ON :Previous Syntax:METer[:STATe] 0|OFF :Previous Syntax:METer[:STATe]? !retur n ...

  • Agilent Technologies E6380-90073 - page 285

    Chapter 1 285 GPIB Command Dictionar y OSCilloscope subsystem OSCilloscope subsystem :CONT rol :CONT rol? These commands set/quer y the oscilloscope’ s contro l menus. Syntax OSC:CONT ‘Main’ ‘Trigger’ ‘Marker’ OSC:CONT? Example OSC:CONT ‘Trigger’ !displays the Trigger menu of the oscilloscope Screen/field equivalent OSC :C O NT co ...

  • Agilent Technologies E6380-90073 - page 286

    286 Chapter 1 GPIB Command Dictionar y OSCilloscope subsystem :MARKer:PPEak This command moves the marker to the maximum value of the a verage level o f the displ ay . This command has no query . Syntax OSC:MARK:PPE !moves the marker to the maximum Screen/field equivalent OSC:MARK:PPE controls the Marker To Peak+ fie ld on the Marker menu of the SC ...

  • Agilent Technologies E6380-90073 - page 287

    Chapter 1 287 GPIB Command Dictionar y OSCilloscope subsystem :SCALe:TIME <string> :SCALe:TIME? These commands set/quer y the horizont al sweep t ime per di vision. Syntax OSC:SCAL:TIME ‘200 ms’ ‘100 ms’ ‘50 ms’ ‘20 ms’ ‘10 ms’ ‘5 ms’ ‘2 ms’ ‘1 ms’ ‘500 us’ ‘200 us’ ‘100 us’ ‘50 us’ ‘20 us’ ...

  • Agilent Technologies E6380-90073 - page 288

    288 Chapter 1 GPIB Command Dictionar y OSCilloscope subsystem :SCALe:VERTical:AM <string> :SCALe:VERTical:AM? These commands set/query the vertical axis amplitude per division when AM Mod or AM Demod is selected as the audio source (see AF AN:INP on pa ge 59). Syntax OSC:SCAL:VERT:AM ‘50%’ ‘20%’ ‘10%’ ‘5%’ ‘2%’ ‘1%’ ? ...

  • Agilent Technologies E6380-90073 - page 289

    Chapter 1 289 GPIB Command Dictionar y OSCilloscope subsystem :SCALe:VERTical:FM <string> :SCALe:VERTical:FM? These commands set/query the vertical axis amplitude per division when FM Mod or FM Demod is selected as the audio source (see AF AN:INP on pa ge 59). Syntax OSC:SCAL:VERT:FM ‘50 kHz’ ‘20 kHz’ ‘10 kHz’ ‘5 kHz’ ‘2 kHz ...

  • Agilent Technologies E6380-90073 - page 290

    290 Chapter 1 GPIB Command Dictionar y OSCilloscope subsystem :SCALe:VERTical:OFFSet <real number> :SCALe:VERTical:OFFSet? These commands set/query the vertical scale (DC) off set of the vertical axis of the oscilloscope display . This moves the s igna l up to four divisions up or down wit h respect t o the oscillo scope’ s fixed center lin ...

  • Agilent Technologies E6380-90073 - page 291

    Chapter 1 291 GPIB Command Dictionar y OSCilloscope subsystem :SCALe:VERTical:VOL T s <string> :SCALe:VERTical:VOL T s? These commands set/q uery the vertica l scale of the oscilloscop e display . This command is valid when the AF AN:INP command specifies one of the follow ing so urces : Audi o Out, Audio In, E xt Mod , or S SB D emod. Syntax ...

  • Agilent Technologies E6380-90073 - page 292

    292 Chapter 1 GPIB Command Dictionar y OSCilloscope subsystem :TRIGger:LEV el <real number> :TRIGger:LEV el ? These commands set/query the level of the os cilloscop e’ s trigger . This command utilizes the “Real Number Setting Syntax” on page 299, but do es not use t he :ST A T , :DUN, :UN IT , :I NCR:MO DE, :I NCR:DUN, commands. Syntax ...

  • Agilent Technologies E6380-90073 - page 293

    Chapter 1 293 GPIB Command Dictionar y OSCilloscope subsystem :TRIGger:DELay <rea l number> :TRIGger:DELay? These commands set/quer y the trigger d elay . Positive values del ay the trigger , negative values apply a pre-trigger funct i on to eac h measurement. This command utilizes the “Real Number Setting Syntax” on page 299, but do es n ...

  • Agilent Technologies E6380-90073 - page 294

    294 Chapter 1 GPIB Command Dictionar y OSCilloscope subsystem :TRIGger:RESet This command triggers an oscilloscope measurement. Syntax OSC:TRIG:RES !triggers the oscilloscope Screen/field equivalent OSC:TRIG:RESet con trols the Reset field on th e Trigger menu of the SCOPE screen. :TRIGger:SENSe <str ing> :TRIGger:SENSe? These comman ds set/ ...

  • Agilent Technologies E6380-90073 - page 295

    Chapter 1 295 GPIB Command Dictionar y OSCilloscope subsystem :TRIGger:TYPE <string> :TRIGger:TYPE? These commands set/query how the tr igger level is set . ‘Auto’ triggers a measurement if a triggering signal is not detect ed withi n approximately 50 ms of the last trigger . ‘Norm’ re quires a spe cific triggering signal before trigg ...

  • Agilent Technologies E6380-90073 - page 296

    296 Chapter 1 GPIB Command Dictionar y PR OGram subsystem PROGram subsystem The Program subsystem provides a set of commands which allow an external co ntroller to generate and control an IBASIC p rogram within the T est Set. These commands have no equivalent fields or screens. [:SELected]:DEFine <program data> [:SELected]:DEFine? PROG:SEL:DE ...

  • Agilent Technologies E6380-90073 - page 297

    Chapter 1 297 GPIB Command Dictionar y PROGram subsys tem [:SELected]:ST A T e <variable> [:SELected]:ST A T e? These commands set/query (from an external IBASIC controller) the execution st ate of the I BASI C pr ogram currently load ed in the T est Set. Syntax PROG:SEL:STAT PROG:SEL:STAT? PROG:STAT !equivalent command [:SELected]:NUMBer < ...

  • Agilent Technologies E6380-90073 - page 298

    298 Chapter 1 GPIB Command Dictionar y RAM Usage Information RAM Usage Information The number of bytes report ed to the remote interface is diff erent than that report ed on the scr een when using the IB_U TILS program. This is because to calculate kilobyt es, the total is divided by 1024 before it is displayed on the screen. SPEC:RAMDISKALLOC? SPE ...

  • Agilent Technologies E6380-90073 - page 299

    Chapter 1 299 GPIB Command Dictionar y Real Number Setting Syntax Real Nu mber Se tting S yntax This syntax is for setting rea l values in commands. It is to be used with commands that r equire the Real N umber Setting Syntax. An example of a command that requires the Real Number Setti ng Syntax is the RF ANalyzer:FREQuency command. (See “RF ANal ...

  • Agilent Technologies E6380-90073 - page 300

    300 Chapter 1 GPIB Command Dictionar y [REGister] subsystem [REGister] subsystem Register Subsystem cont ains the save/recall commands used in the Save/Recall register s. :CLEar <integer or s tring> This command clears the register defined by the inte ger value or stri ng ar gumen t. Syntax :CLE ‘<string>’ !clears the register named ...

  • Agilent Technologies E6380-90073 - page 301

    Chapter 1 301 GPIB Command Dictionar y [REGister] subsys tem :RECall <integer or s tring> This command recal ls the register def ined by the integer value or string ar gumen t Syntax :REC ‘<string>’ !recalls the register named ‘string’ :REC ‘<integer value>’ !recalls the numbered register REG:REC ‘<string>’ ! ...

  • Agilent Technologies E6380-90073 - page 302

    302 Chapter 1 GPIB Command Dictionar y [REGister] subsystem :SA VE:LIST? :SA VE:LIST? lists the save/recall registers that have been named. Syntax REG:SAVE:LIST? :SAVE:LIST? !equivalent command Screen/field equivalent :SA VE:LIST? has no e quivalent field. ...

  • Agilent Technologies E6380-90073 - page 303

    Chapter 1 303 GPIB Command Dictionar y RF AN alyzer subsys tem RF ANalyz er subsystem :A TT enuator <string> :A TT enuator? These commands set/query the amount of input attenuati on in the path of the select ed input port. CONF: A TT :MODE ‘Auto’ overrides this command. Syntax RFAN:ATT ‘0 dB’ ‘5 dB’ ‘10 dB’ ‘15 dB’ ‘20 d ...

  • Agilent Technologies E6380-90073 - page 304

    304 Chapter 1 GPIB Command Dictionar y RF ANalyzer subsystem :FREQuency <real number> :FREQuency? These commands set/quer y the tune frequency for the RF anal y zer . This command re quires that t he CONF:RFD ‘Freq’ command is used. This command utilizes the “Real Number Setting Syntax” on page 299, but does no t use the :ST A T e com ...

  • Agilent Technologies E6380-90073 - page 305

    Chapter 1 305 GPIB Command Dictionar y RF AN alyzer subsys tem :IFBW <string> :IFBW? These commands set/quer y the IF filter bandwidth. Syntax RFAN:IFBW ‘15 kHz’ ‘230 kHz’ RFAN:IFBW? Screen/field equivalent RF AN:IFBW controls the IF Filter field on the RF ANAL YZER screen. :INPut <string> :INPut? These commands set/quer y the R ...

  • Agilent Technologies E6380-90073 - page 306

    306 Chapter 1 GPIB Command Dictionar y RF ANalyzer subsystem :PMEasurement:ZERO This command zeroes the TX power measurement. Syntax RFAN:PME:ZERO Screen/field equivalent RF AN:PME:ZERO controls the TX Pwr Zero field on the RF ANAL YZER screen. :PMEasurement:ZERO:MO DE :PMEasurement:ZERO:MO DE? This command sets/queries the zeroing mode of the TX p ...

  • Agilent Technologies E6380-90073 - page 307

    Chapter 1 307 GPIB Command Dictionar y RF AN alyzer subsys tem :SQUelch <string> :SQUelch? :SQU sets/queries the sque lch cont rol setting. Syntax RFAN:SQU ‘Pot’ ‘Open’ ‘Fixed’ RFAN:SQU? Screen/field equivalent RF AN:SQU controls the Squelch field on the RF ANAL YZER screen. ...

  • Agilent Technologies E6380-90073 - page 308

    308 Chapter 1 GPIB Command Dictionar y RFGenerator subsyst em RFGenerator subsystem The RF generato r subsystem controls th e functions of the RF GENERA TOR screen. :AMPLitude <real number> This command sets/ queries the ampl itude of the RF generator . This fiel d also contr ols the RF generat or viewed in the spectru m analyzer . This comma ...

  • Agilent Technologies E6380-90073 - page 309

    Chapter 1 309 GPIB Command Dictionar y RFGen erato r subs ystem :CHANnel <string> :CHANnel? These commands set/quer y the RF channe l number . Channel tuning must be en abled (CONF:RFD ‘ Chan’). Y ou must al so choose the c orrect channel standard (CONF: RFCS <string>). Syntax RFG:CHAN <integer> RFG:CHAN? Screen/Field Equivale ...

  • Agilent Technologies E6380-90073 - page 310

    310 Chapter 1 GPIB Command Dictionar y RFGenerator subsyst em :FREQuency <real number> :FREQuency? These commands set/quer y the freque ncy of the RF gener ator . The T est Set must first be in frequency tuning mod e. (See CONF:RFD ‘Freq’ o n page 164). This command utilizes the “Real Number Setting Syntax” on page 299, but does no t ...

  • Agilent Technologies E6380-90073 - page 311

    Chapter 1 311 GPIB Command Dictionar y RFGen erato r subs ystem :MODulation:EXT ernal:AM <rea l number> :MODulation:EXT ernal:AM? These commands set/quer y the AM sensitivity o f the RF generato r when AM is applied t hrough the modulation input port. The modulation input must be set to AM. This command utilize s the “Rea l Number Setting S ...

  • Agilent Technologies E6380-90073 - page 312

    312 Chapter 1 GPIB Command Dictionar y RFGenerator subsyst em :MODulation:EXT ernal:FM <rea l number> :MODulation:EXT ernal:FM? These commands set/query the FM deviat ion of the RF generator when FM is applied through the modulation input port. The modulation input must be set to FM. This command utilizes the “Real Number Setting Sy ntax” ...

  • Agilent Technologies E6380-90073 - page 313

    Chapter 1 313 GPIB Command Dictionar y SANalyzer subsys tem SANalyzer subsystem :A TT enuator <string> :A TT enuator? These commands set/quer y the atten uator setting f or the input port. Syntax SAN:ATT ‘0 dB’ ‘5 dB’ ‘10 dB’ ‘15 dB’ ‘20 dB’ ‘25 dB’ ‘30 dB’ ‘35 dB’ SAN:ATT? Screen/field equivalent SAN:A TT cont ...

  • Agilent Technologies E6380-90073 - page 314

    314 Chapter 1 GPIB Command Dictionar y SANalyzer subsystem :CFRequency <real number> :CFRequency? These commands set/quer y the cente r frequency of the spe ctrum analyzer ’ s display . This command utilizes the “Real Number Setting Syntax” on page 299, but does no t use the :ST A T e command. Syntax SAN:CFR <real number> SAN:CFR? ...

  • Agilent Technologies E6380-90073 - page 315

    Chapter 1 315 GPIB Command Dictionar y SANalyzer subsys tem :DISPlay:SCALe <str ing> :DISPlay:SCALe? These commands set/query the vertical scale of the spectrum analyzer display . Syntax SAN:DISP:SCAL ‘1 dB/div’ ‘2 dB/div’ ‘10 dB/div’ SAN:DISP:SCAL? Screen/field equivalent SAN:DISP:SCAL controls the l ower subfield o f the Sensiti ...

  • Agilent Technologies E6380-90073 - page 316

    316 Chapter 1 GPIB Command Dictionar y SANalyzer subsystem :MARKer:DEL T a:NPEak This command sets the delta marker at the next peak of the spect rum analyzer display . Syntax SAN:MARK:DELT:NPE Screen/field equivalent SAN:MARK:DEL T : NPE contro ls the Marker To, Next P eak field o n the Mark er menu of the SPEC ANL screen when the Norm/Delt a fiel ...

  • Agilent Technologies E6380-90073 - page 317

    Chapter 1 317 GPIB Command Dictionar y SANalyzer subsys tem :MARKer:DEL T a:POSition <real number> :MARKer:DEL T a:POSition? Thes e co mma n d s se t/ q u e ry th e po si t ion of th e de lt a mark er on the spectrum analyzer . This command utilizes the “Real Number Setting Syntax” on page 299, but does no t use the :ST A T e command. Syn ...

  • Agilent Technologies E6380-90073 - page 318

    318 Chapter 1 GPIB Command Dictionar y SANalyzer subsystem :MARKer:MODE <string> :MARKer:MODE? These commands set/quer y the whether the marker is i n normal mode or delt a mode. In normal mode the spectrum analyzer measures the freque ncy and level at t he marker ’ s position. In delt a mode the sp ectrum analyzer displays a second marker ...

  • Agilent Technologies E6380-90073 - page 319

    Chapter 1 319 GPIB Command Dictionar y SANalyzer subsys tem :MARKer[:NORMal]:NPEa k This command moves the marker to the next peak of the spectr um analyzer ’ s measurement tr ace. Syntax SAN:MARK:NORM:NPE Screen/field equivalent SAN:MARK:NORM:NPE contro ls the Marker To, Next P eak fie l d on the Marker menu of the SPEC ANL scree n when N orm/De ...

  • Agilent Technologies E6380-90073 - page 320

    320 Chapter 1 GPIB Command Dictionar y SANalyzer subsystem :MARKer[:NORMal]:POSi tion <real number> :MARKer[:NORMal]:POSi tion? These commands set/query the posi tion of t he marker on the sp ectrum analyzer . This command utilizes the “Real Number Setting Syntax” on page 299, but does no t use the :ST A T e command. Syntax SAN:MARK:NORM: ...

  • Agilent Technologies E6380-90073 - page 321

    Chapter 1 321 GPIB Command Dictionar y SANalyzer subsys tem :MASK:BEEP :MASK:BEEP? This command turns the failure indicato r beeper off or on. Syntax SAN:MASK:BEEP ‘On’ ‘Off’ SAN:MASK:BEEP? Screen/field equivalent SAN:MASK:BEEP controls the Mask Beep fi eld on the Mask menu of the SPEC ANL screen. :MASK:DISPlay :MASK:DISPlay? This command s ...

  • Agilent Technologies E6380-90073 - page 322

    322 Chapter 1 GPIB Command Dictionar y SANalyzer subsystem :MASK:FIXed:LOW er:POIN ts:EDIT This command selects/queries the po int number to ed it on the lower spectrum analyzer mask. The maximum number i s 15. Syntax SAN:MASK:FIX:LOW:POIN:EDIT <integer 1-15> SAN:MASK:FIX:LOW:POIN:EDIT? Screen/field equivalent SAN: MA S K: F IX: L OW:POIN: ED ...

  • Agilent Technologies E6380-90073 - page 323

    Chapter 1 323 GPIB Command Dictionar y SANalyzer subsys tem :MASK:FIXed:LOW er:POIN ts:FREQuency1 (through FR EQ15) :MASK:FIXed:LOW er:POIN ts:FREQuency1? (through F REQ15) These commands set/quer y the freq uency (Hz) of the defined points on the lower spectrum analyzer mask. The units for the mask are GHz, MH z, kHz, and Hz. Syntax T o define the ...

  • Agilent Technologies E6380-90073 - page 324

    324 Chapter 1 GPIB Command Dictionar y SANalyzer subsystem :MASK:FIXed:UPPer:POIN ts:EDIT :MASK:FIXed:UPPer:POIN ts:EDIT? This command selects/queries the po int number to ed it on the upper spectrum analyzer mask. The maximum number i s 15. Syntax SAN:MASK:FIX:UPP:POIN:EDIT <integer 1-15> SAN:MASK:FIX:UPP:POIN:EDIT? Screen/field equivalent S ...

  • Agilent Technologies E6380-90073 - page 325

    Chapter 1 325 GPIB Command Dictionar y SANalyzer subsys tem :MASK:FIXed:UPPer:POIN ts:FREQuency1 (through FR EQ15) :MASK:FIXed:UPPer:POIN ts:FREQuency1? (through F REQ15) These commands set/quer y the freq uency (Hz) of the defined points on the uppe r spectrum analyzer mask. The units for the mask are GHz, MH z, kHz, and Hz. Syntax T o define the ...

  • Agilent Technologies E6380-90073 - page 326

    326 Chapter 1 GPIB Command Dictionar y SANalyzer subsystem :MASK:RELative:LOW er:P OINts:EDIT :MASK:RELative:LOW er:P OINts:EDIT? This command selects/queries the po int number to ed it on the lower spectrum analyzer mask. The maximum number i s 15. Syntax SAN:MASK:REL:LOW:POIN:EDIT <integer 1-15> SAN:MASK:REL:LOW:POIN:EDIT? Screen/field equi ...

  • Agilent Technologies E6380-90073 - page 327

    Chapter 1 327 GPIB Command Dictionar y SANalyzer subsys tem :MASK:RELative:LOW er:P OINts:FREQuency1 (through FREQ15 ) :MASK:RELative:LOW er:P OINts:FREQuency1? (through FREQ1 5) These commands set/quer y the freq uency (Hz) of the defined points on the lower spectrum analyze r mask (relat ive to the ce nter line of the display). The units for the ...

  • Agilent Technologies E6380-90073 - page 328

    328 Chapter 1 GPIB Command Dictionar y SANalyzer subsystem :MASK:RELative:UPPer:P OINts:EDIT This command selects/queries the po int number to ed it on the upper spectrum analyzer mask. The maximum number i s 15. Syntax SAN:MASK:REL:UPP:POIN:EDIT <integer 1-15> SAN:MASK:REL:UPP:POIN:EDIT? Screen/field equivalent SAN:MASK:REL:UPP:POIN:EDIT con ...

  • Agilent Technologies E6380-90073 - page 329

    Chapter 1 329 GPIB Command Dictionar y SANalyzer subsys tem :MASK:RELative:UPPer:P OINts:FREQuency1 (through FREQ15) :MASK:RELative:UPPer:P OINts:FREQuency1? (throug h FREQ15) These commands set/quer y the freq uency (Hz) of the defined points on the uppe r spectrum analyzer mask. The units for the mask are GHz, MH z, kHz, and Hz. Syntax T o define ...

  • Agilent Technologies E6380-90073 - page 330

    330 Chapter 1 GPIB Command Dictionar y SANalyzer subsystem :RFGenerator <string> :RFGenerator? This command selects between the fixed frequen c y generator and the tracking generat or . The query form r eturns the mode setting of the generato r . Syntax SAN:RFG ‘Track’ ‘Fixed’ SAN:RFG? Screen/field equivalent SAN: R F G co n t rols th ...

  • Agilent Technologies E6380-90073 - page 331

    Chapter 1 331 GPIB Command Dictionar y SANalyzer subsys tem :SP AN <real number> :SP AN? These commands set/quer y the span of the sp ectrum analyzer display . This command utilizes the “Real Number Setting Syntax” on page 299, but does no t use the :ST A T e command. Syntax SAN:SPAN <real number> SAN:SPAN? Screen/field equivalent S ...

  • Agilent Technologies E6380-90073 - page 332

    332 Chapter 1 GPIB Command Dictionar y SANalyzer subsystem :TGENerator:DESTination <string> :TGENerator:DESTination? These commands set/quer y the outp ut port f or the tracking generator . T o sele ct the tr acking generat or use the SAN: RFG ‘T rack’ command. Syntax SAN:TGEN:DEST ‘RF Out’ ‘Dupl’ SAN:TGEN:DEST? Screen/field equiv ...

  • Agilent Technologies E6380-90073 - page 333

    Chapter 1 333 GPIB Command Dictionar y SANalyzer subsys tem :TGENerator:SWEep <string> :TGENerator:SWEep? These commands set/query the sweep type of the tracking generator . T o select the tr acking generator use the SAN:RFG ‘T rack’ command. ‘Norm’ specifies a sweep from low to high frequency . ‘Invert’ specifies a sweep from hig ...

  • Agilent Technologies E6380-90073 - page 334

    334 Chapter 1 GPIB Command Dictionar y SANalyzer subsystem :TRACe:MHOLd <string> :TRACe:MHOLd? These commands set/quer y the type of averaging used in the spectrum analyzer ’ s display . Syntax SAN:TRAC:MHOL ‘No Pk/Avg’ ‘Pk Hold’ ‘Avg 1’ ‘Avg 2’ ‘Avg 3’ ‘Avg 4’ ‘Avg 5’ ‘Avg 10’ ‘Avg 20’ ‘Avg 50’ ‘Av ...

  • Agilent Technologies E6380-90073 - page 335

    Chapter 1 335 GPIB Command Dictionar y SANalyzer subsys tem :TRACe:NORMalize <s tring> :TRACe:NORMalize? These commands set/ query the type of t race di splay . ‘A Only’ provides a continuousl y upd ate d display (normal operatio n). ‘A-B’ displays the differe nce between the trace saved using SAN:TRAC:SA VE and the currently displaye ...

  • Agilent Technologies E6380-90073 - page 336

    336 Chapter 1 GPIB Command Dictionar y ST A T u s ST A T us The ST A T us sub s ystem repo rts many of the modes of the T est S et, including the states. These command s do not corre spond to fields in specific screens. For d etailed inf ormation about status registers, see “Status Reporting” in the Advance d Operations chapter o f the T est Se ...

  • Agilent Technologies E6380-90073 - page 337

    Chapter 1 337 GPIB Command Dictionar y ST A T us :CALibration:NTRansition :CALibration:NTRansition? These commands set/quer y the state of the negative t ransition of the calibration. This query returns an integer value. Syntax STAT:CAL:NTR STAT:CAL:NTR? :CALibration[:EVENt?]:PT Ransition :CALibration[:EVENt?]:PT Ransition? These commands set/query ...

  • Agilent Technologies E6380-90073 - page 338

    338 Chapter 1 GPIB Command Dictionar y ST A T u s :COMMunicate[:E VENt?]:NTRansition :COMMunicate[:E VENt?]:NTRansition? This query returns an integer value. Syntax STAT:COMM:NTR STAT:COMM:NTR? :COMMunicate[:E VENt?]:PTRansition :COMMunicate[:E VENt?]:PTRansition? This query returns an integer value. Syntax STAT:COMM:PTR STAT:COMM:PTR? :HARD1:CONDi ...

  • Agilent Technologies E6380-90073 - page 339

    Chapter 1 339 GPIB Command Dictionar y ST A T us :HARD1[:EVENt?]:PTRans ition :HARD1[:EVENt?]:PTRans ition? This query returns an integer value. Syntax STAT:HARD1:PTR STAT:HARD1:PTR? :HARD2:CONDition :HARD2:CONDition? This query returns an integer value. Syntax STAT:HARD2:COND STAT:HARD2:COND? :HARD2:ENABle :HARD2:ENABle? This query returns an inte ...

  • Agilent Technologies E6380-90073 - page 340

    340 Chapter 1 GPIB Command Dictionar y ST A T u s :OPERation:CONDition? This query returns an integer value. Syntax STAT:OPER:COND? :OPERation:ENABle :OPERation:ENABle? This query returns an integer value. Syntax STAT:OPER:ENAB STAT:OPER:ENAB? :OPERation[:EVENt?]:NTR ansition :OPERation[:EVENt?]:NTR ansition? This query returns an integer value. Sy ...

  • Agilent Technologies E6380-90073 - page 341

    Chapter 1 341 GPIB Command Dictionar y ST A T us :CALibrating:ENABle :CALibrating:ENABle? This query returns an integer value. Syntax STAT:CAL:ENAB STAT:CAL:ENAB? :CALibrating[:EVENt?]:NT Ransition :CALibrating[:EVENt?]:NT Ransition? This query returns an integer value. Syntax STAT:CAL:NTR STAT:CAL:NTR? :CALibrating[:EVENt?]:PT Ransition :CALibrati ...

  • Agilent Technologies E6380-90073 - page 342

    342 Chapter 1 GPIB Command Dictionar y ST A T u s :QUEStionable[:EVENt?]:N TRansition :QUEStionable[:EVENt?]:N TRansition? This query returns an integer value. Syntax STAT:QUES:NTR STAT:QUES:NTR? :QUEStionable[:EVENt?]:P TRansition :QUEStionable[:EVENt?]:P TRansition? This query returns an integer value. Syntax STAT:QUES:PTR STAT:QUES:PTR? :MEASuri ...

  • Agilent Technologies E6380-90073 - page 343

    Chapter 1 343 GPIB Command Dictionar y ST A T us :MEASuring[:EVENt?]:PTR ansition :MEASuring[:EVENt?]:PTR ansition? This query returns an integer value. Syntax STAT:MEAS:PTR STAT:MEAS:PTR? :CDMA1:CONDition? This query returns an integer value. Syntax STAT:CDMA1:COND? :CDMA1:ENABle :CDMA1:ENABle? This query returns an integer value. Syntax STAT:CDMA ...

  • Agilent Technologies E6380-90073 - page 344

    344 Chapter 1 GPIB Command Dictionar y ST A T u s :IBASic:CONDition? This query returns an integer value. Syntax STAT:IBAS:COND? :IBASic:ENABle :IBASic:ENABle? This query returns an integer value. Syntax STAT:IBAS:ENAB STAT:IBAS:ENAB? :IBASic[:EVENt?]:NTRans ition :IBASic[:EVENt?]:NTRans ition? This query returns an integer value. Syntax STAT:IBAS: ...

  • Agilent Technologies E6380-90073 - page 345

    Chapter 1 345 GPIB Command Dictionar y SYST em:[ERRor ?] SYST em:[ERRor?] This command returns an integer foll owed by quoted string. Syntax SYST:ERR? ...

  • Agilent Technologies E6380-90073 - page 346

    346 Chapter 1 GPIB Command Dictionar y TRIGger subsystem TRIGger su bsystem The T r igger subsystem contains commands for triggering measurements. There are no sc reen/field equivalents for the trigger subsystem commands. These commands will override any triggering control fiel ds on individual screens. :ABORt This command en ds a measurement cycl ...

  • Agilent Technologies E6380-90073 - page 347

    Index Index 347 Symbols # o f Fram es , 11 4 #Pt s , 321 , 32 3 , 325 , 327 , 329 Numeric s 0 dB ref , 127 A A Only/A-B , 335 abo r t measu reme nt , 346 AC Level , 21 1 ACP , 105 ACP Cal CDMA an alyzer , 99 ACP Fltr BW CDMA an alyzer , 99 ACP Offset CDMA an alyzer , 100 ADC FS aver age pow er , 217 channe l po wer , 219 EVM , 218 rho , 218 ADCfs , ...

  • Agilent Technologies E6380-90073 - page 348

    348 Index Index Control , 151 , 152 Controls code domain analyzer , 11 8 , 123 osc illosco pe , 285 spectrum analyzer , 314 CW RF Path , 11 7 D Data Buff , 146 data buffer , 114 Data Length Seri al 10 Port , 176 Serial 1 1 Port , 184 Seri al 9 Port , 166 Data Level (AMPS-T ACS) AM , 77 FM , 77 mV , 78 Data Level (NAMP-NT AC) FOCC AM , 91 FM , 92 mV ...

  • Agilent Technologies E6380-90073 - page 349

    Index Index 349 hexadec im al values , 206 Hi Limit key , 283 Hi T one Freq , 263 , 268 Frq Err , 263 , 269 Holdoff Delay Seri al 9 Port , 170 HP-IB Adrs , 150 HP-IB, printer port , 158 I I/O CONFIGURE scre en , 205 IBASIC Echo Seri al 9 Port , 167 Idle/S end , 11 6 IF Filte r , 305 Incomi n g Call Securi ty Seri al 9 Port , 173 Incr S et key , 299 ...

  • Agilent Technologies E6380-90073 - page 350

    350 Index Index oscillo scope , 273 , 274 frequency at spectrum ana ly zer , 280 level at code domain analyzer , 22 8 , 229 , 235 , 236 , 237 , 238 , 239 , 240 phase at code domain analyzer , 23 0 , 240 RF level at spectrum ana ly zer , 279 , 280 time at code domain analyzer , 23 1 oscillo scope , 27 4 Mask T ype Fix , 321 , 32 2 , 323 , 324 , 325 ...

  • Agilent Technologies E6380-90073 - page 351

    Index Index 351 Phase/div , 129 Phse , 230 PK- , 56 Pk Det T o , 56 PK- HOLD , 56 Pk Hold , 334 PK+ , 56 PK+ HOLD , 56 PK+-/2 , 56 PK+-/2 Hd , 56 PK+-MAX , 56 PK+-MX Hd , 56 PN Increme nt , 107 , 126 PN Offset , 108 , 12 7 , 222 PN Ofs , 231 Polarity , 73 decoder , 198 Port/Sweep Norm/Inve rt , 333 RF Out/Dupl , 332 Pos/Neg , 294 Position , 286 spe ...

  • Agilent Technologies E6380-90073 - page 352

    352 Index Index questi onable , 34 1 sta tus r egiste r group s , 336 See also Staus Reporting in the Progra mmer ’ s Guide Stop enco der , 74 Stop F ill er AMPS-T ACS , 80 NAMP-NT AC FOCC , 94 Stop Length Seri al 10 Port , 176 Serial 1 1 Port , 185 Seri al 9 Port , 166 Stop Me as , 198 sweep , osci lloscope , 287 Sym , 265 Symb ol Frequ encie (H ...

Fabbricante Agilent Technologies Categoria Cell Phone Accessories

I documenti che otteniamo dal fabbricante del dispositivoAgilent Technologies E6380-90073 possono essere suddivisi in diversi gruppi. Questi sono tra gli altri:
- disegni tecnici Agilent Technologies
- manuali per l’uso E6380-90073
- schede prodotto Agilent Technologies
- opuscoli
- o etichette energetiche Agilent Technologies E6380-90073
Tutti sono importanti, ma le informazioni più importanti dal punto di vista di utilizzo del dispositivo possono essere trovate nel manuale per l’uso Agilent Technologies E6380-90073.

Un gruppo di documenti denominato manuali per l’uso, è anche suddiviso in tipi più specifici, come: Manuali di installazione Agilent Technologies E6380-90073, manuali di manutenzione, brevi manuali o manuali utente Agilent Technologies E6380-90073. A seconda delle esigenze, si dovrebbe cercare il documento necessario. Nel nostro sito web, puoi visualizzare il manuale più popolare per l'uso del prodotto Agilent Technologies E6380-90073.

Manuale completo del dispositivo Agilent Technologies E6380-90073, come deve essere?
Il manuale per l’uso, denominato anche istruzioni per l’uso, o semplicemente il manuale è un documento tecnico progettato per aiutare ad utilizzare Agilent Technologies E6380-90073 dagli utenti. I manuali sono di solito scritti da uno scrittore tecnico, ma in un linguaggio accessibile a tutti gli utenti Agilent Technologies E6380-90073.

Il manuale per l’uso completo Agilent Technologies, dovrebbe includere vari elementi di base. Alcuni di loro sono meno importanti, come ad esempio: copertina / pagina del titolo o pagina di autore. Tuttavia, la parte rimanente, dovrebbe fornire informazioni importanti dal punto di vista dell'utente.

1. Introduzione e la guida su come utilizzare il manuale Agilent Technologies E6380-90073 - All'inizio di ogni manuale, dovremmo trovare indizi su come utilizzare il documento. Dovrebbe contenere informazioni sulla posizione dell’indice Agilent Technologies E6380-90073, FAQ o i problemi più comuni : i punti che sono più spesso cercati dagli utenti di ogni manuale
2. Indice - elenco di tutti i suggerimenti per Agilent Technologies E6380-90073 che possono essere trovati nel documento corrente
3. Suggerimenti su come utilizzare le funzioni di base del dispositivo Agilent Technologies E6380-90073 - che ci dovrebbe aiutare nei primi passi durante l'uso Agilent Technologies E6380-90073
4. Troubleshooting - sequenza sistematica delle attività che vi aiuterà a diagnosticare e, successivamente, risolvere i problemi più importanti del Agilent Technologies E6380-90073
5. FAQ - domande più frequenti