    USER’S GUIDE A GILENT 6010A , 6011A , 6012B, 6015A , 6023A and 6028A DC A UTORA NGING POWER SUPPLY Agilent Part N umber 5964-8120 Printed in Malay sia Microfiche Part No. 5964-8121 September, 2004 ...

  • Agilent Technologies 6010A - page 2

    2 CERTIFICA TION Agilent Technologies c ertifies that this product met its published specifications at time of shipment from the factory. Agilent Technologies fur ther certifies that its calibr ation measurements ar e traceable to the United States National Bureau of Standards, to the ex tent allowed by the Bureau’s calibration facility, and to t ...

  • Agilent Technologies 6010A - page 3

    3 Safety Summary The f ollowin g g eneral sa f et y p recautions must be observed duri n g all p hases o f o p eration o f this instrument. Fa ilure to com p l y with these p recautions or with s p eci f ic warnin g s elsew here in this manual violat es sa f et y standards o f desi g n, manu f acture, and int ended use o f the instrument. A g ilent ...

  • Agilent Technologies 6010A - page 4

    4 SAFETY SYMBOLS Direct current Alternating current Both direct and alternating current Three-phase alternatin g current Earth (grou nd) terminal Protective earth (ground ) terminal Frame or ch assis terminal Terminal is at earth potential. Used for measurement and control circuits designed to be o p erated with one term inal at earth potential. Te ...

  • Agilent Technologies 6010A - page 5

    5 DECLARATION OF CONFORMITY According to ISO/IEC Guide 22 and CEN/CENELEC EN 4501 4 Revision: B.00.00 Issue Date: Created on 11/24/2003 3:35 PM Document No. 60xyA.11.24.doc Manufacturer’s Name and Address Responsible Party Alternate Manufacturing Site Agilent Technologies, Inc. Agilent Technologies (Malaysia) Sdn. Bhd 550 Clark Drive, Suite 101 B ...

  • Agilent Technologies 6010A - page 6

    6 A coustic Noise Statement Herstellerbescheinigung Diese I nformatio n ste ht im Zusamm enhang mit de n Anfo rd e rungen d e r Maschine nlärminfor matio nsve ror dnung vo m 18 Janua r 1991. * S challdruck pegel Lp < 70 dB(A ) * Am Arbeitsplatz * Normal er Betrieb * Nach DIN 45635 T. 19 ( T ypp rüfung) Manufacturer’ s Declaratio n This state ...

  • Agilent Technologies 6010A - page 7

    Table Of Contents 1. General Information Introduction .................................................................................................................................................... 9 Description ..................................................................................................................................... ...

  • Agilent Technologies 6010A - page 8

    Multiple-Su pply Operation...................................................................................................... ...................... 39 Auto- Parallel Operation ...................................................................................................... ........................ 39 Series Operation ...................... ...

  • Agilent Technologies 6010A - page 9

    General Information 9 1 General Information Introduction This manual contains specifications, installation instructions, and operating instructio ns for DC P ower Supply Models: Agi lent6010A, 6011A , 6012B, 6015A, 6023A and 6028A. Ref er to "Related Docum ents" for oth er inform ation con cerning these products. Description These power s ...

  • Agilent Technologies 6010A - page 10

    10 General Information Option Description 100 Input pow er: 100 Vac + 6%, -10%; 48- 63 Hz single ph ase. 120 Input pow er: 120 Vac +6%, -13%. 48-63 Hz s ingle ph ase. 220 Input Pow er: 220 Vac +6%, -13%; 48- 63 Hz, single ph ase. 240 Input pow er: 240 Vac +6%, -13%; 48- 63 Hz, single ph ase. 800 908 Rack mounting kit 909 Flanges with Handles A cces ...

  • Agilent Technologies 6010A - page 11

    General Information 11 PSI board s to lie on table outside unit, and co ntro l board test co nnecto r. 5060-0138 GP-IB conn ector non- metr ic to met ric convers ion kit . 5060-2860 FET service kit, inclu des FETs and all compon ents that sh ould be replaced with FETs. 59510A Relay Accessory 59511A Relay Accessory (Polarity Revers ing) Instrument a ...

  • Agilent Technologies 6010A - page 12

    12 General Information Table 1-1 Perf ormance Specifications Agilent Technologies Model 6010A 6011A DC Output: V oltage, c urrent and power spans indi c at e range Volts 0-200 V 0-20 V over which output m ay be varied using f ront panel cont rols. A mps 0-17 A 0-120 A Maximum Power 1000-1200 W 840-1072 W Load Effect (Load Regulat ion) Vol t age loa ...

  • Agilent Technologies 6010A - page 13

    General Information 13 6012B 6015A 6023A 6028A NOT ES. 0-60 V 0-500 V 0-20 V 0-60 V 0-50 A 0-5 A 0-30 A 0-10 A 1000-1200 W 1000-1050 W 200-242 W 200-242 W 0.01% + 5 m V 0.01% + 40 m V 0. 01% + 2 m V 0.01% + 3 mV 0.01% + 10 m A 0.03% + 34 m A 0. 01% + 9 m A 0 01% + 5 m A 0.01% + 3 m V 0.01% + 13 m V 0.01% + 1 m V 0. 01% + 2 m V 0.01% + 10 m A 0.03% ...

  • Agilent Technologies 6010A - page 14

    14 General Information Table 1-2 Supplemental Characteristics (continued) DC Floating Voltage : Either output term inal may be floated up to the following voltage (including the output vo ltage) from earth gr ound : ± 240 Vdc on Models 6011A , 6012B, 6023A, an d 6028A ± 550 Vdc on Models 6010A and 6015A Exceeding these voltages can result in dam ...

  • Agilent Technologies 6010A - page 15

    General Information 15 Model Agilent 6010A Agilent 6011A Agilent 6012B Agilent 6015A Agilent 6023A Agilent 6028A Vp1 200 V 20 V 60 V 500 V 20 V 60 V Ip1 5 A 50 A 17.5 A 2 A 10 A 3.3 A Vp2 120 V 14 V 40 V 350 V 14 V 40 V Ip2 10 A 76 A 30 A 3 A 17.2 A 6 A Vp3 60 V 7 V 20 V 200 V 6.7 V 20 V Ip3 17 A 120 A 50 A 5 A 30 A 10 A Figure 1-1. Output Charact ...

  • Agilent Technologies 6010A - page 16

    16 General Information Figure 1-2. Output Impedance ...

  • Agilent Technologies 6010A - page 17

    Installation 17 2 Installation Introduction This section contains instructions for ch ecking and repackag ing the supply ; bench or rack mountin g, connecting th e supply to ac input power, and converting the supply from one line voltage to another if required.. Note: All power supplies generate magnetic f ields that may af fect the operatio n o f ...

  • Agilent Technologies 6010A - page 18

    18 Installation Outline Diagrams Figure 2-1 illustrates the outline shape and d imensions of the cabinet. 6023A ON LY 6028A ONLY Figure 2-1. Outline Diagram Bench Operation The supply cabinet h as plastic feet, wh ich are shaped to ensure self- aligning w hen stacked w ith other Agilent Technologies System II cabinets. Rack Mounting The supply can ...

  • Agilent Technologies 6010A - page 19

    Installation 19 Pow e r Connection Caution: Conn ection of th is supply to an ac pow er source sh ould only be perform ed by an el ectrician or oth er qualified person . Before connecting the su pply to the ac pow er source, check th e label on the rear panel to ensure that th e supply is set for the correct ac v oltage to be used. If necessary , c ...

  • Agilent Technologies 6010A - page 20

    20 Installation The offset pin on the standard pow er cable three-prong connector is the groun d connection. If a tw o-contact receptacle is encount ered, it must be replaced w ith a properly g rounded three-con tact receptacle in accordance w ith the National Electrical Code and any local codes and ordinances. The w ork should be don e by a qualif ...

  • Agilent Technologies 6010A - page 21

    Installation 21 Figure 2-2. Power-Cord Plug Configurations A gilent M odels 6023A and 6028A . Line voltag e conversion is accomplish ed via three compon ents; a two-section line select switch , line voltage jumper, and a rear panel fu se. T o convert the su pply from one voltage to anoth er, proceed as follows: a. Unplug th e power supply and w ait ...

  • Agilent Technologies 6010A - page 22

    22 Installation Agilent 6010A, 6011A, 6012B, 6015A Agilent 6023A, 6028A Figure 2-3. Line Voltage Conversion Components ...

  • Agilent Technologies 6010A - page 23

    Installation 23 AC Line Impedance Check The power s upply is design ed for proper operation w ith l ine im pedances ty pically f ound in ac power lin es. Howev er, if th e supply is connected to an ac pow er line havin g a high im pedance com bined with lin e voltage near the m inim um specif ied value, (e.g., 104 Vac f or nominal 120 Vac), th e un ...

  • Agilent Technologies 6010A - page 24

    24 Installation Rear Panel Screw Sizes and Part Numbers Refer to the f ollowing list if you n eed to replace any of the rear panel connection hardware: A gilent M odels 6010A, 6011A , 6012B and 6015A Item Description Agilent Part number ac inpu t cover 5060-3237 ac input cover screw s M4 X 0.7 X 60 mm (qt y 4) 0515-0156 ac input barrier block 3-ter ...

  • Agilent Technologies 6010A - page 25

    Operating Instructions 25 3 Operating Instructions Introduction This section describes the operating controls and indicators, turn -on checkout procedu res, and operating procedures and considerations for the p o wer supply. Front-panel operation and remote resistance/voltage pr o gramm ing is described in this section. T he front-pan el controls a ...

  • Agilent Technologies 6010A - page 26

    26 Operating Instructions Table 3-1. Controls and Indicators Number Controls/Indicators Description Page 1 L INE Switch Pressing at the top o f the sw itch app lies ac mains voltage to the units bias and pow er circuits. Unit is operational approx imat ely 3 secon ds after pow er on. 27 2 VOL T AGE CONTROL Clockwise rotation increases the output vo ...

  • Agilent Technologies 6010A - page 27

    Operating Instructions 27 a. Check th at the rear-panel mode sw itches are set as show n in Figu re 3-3. b. Check th at + lead is connected to +S and the – lead is con nected to –S and tighten ed securely. T he + Sense lead is connected to + Outpu t and the − sense lead is conn ected to − Output lead at the f actory. c. Check that the rear ...

  • Agilent Technologies 6010A - page 28

    28 Operating Instructions Initial Setup and Interconnections WA RNING: Turn off input ac po w er before changing a ny rea r-panel connection and make certain all wires and straps are properly connected and term inal block screws are securely tightened before reapplying power. Be certain to replace both terminal block covers before reapplying pow er ...

  • Agilent Technologies 6010A - page 29

    Operating Instructions 29 Table 3-2. M aximum Wire Lengths To Limit Voltage Drops Wire Size Resistivity Maximum Length In Meters (Feet)To Lim it Voltage Drop To 0.5V Or L ess AWG Cross-section (m m 2 ) Ω /kft Ω /k m 5 A 10A 17 A 30A 50A 120A 2 2 1 6 . 1 5 ( 6 . 1 9 ) ***** 0 , 5 4 0 . 1 2 . 5 ***** 2 0 1 0 . 1 6 ( 9 . 8 ) ***** 0 , 7 5 2 6 . 7 ...

  • Agilent Technologies 6010A - page 30

    30 Operating Instructions Either positive or negative vo ltages can be ob tained from the supply by grounding one of the output terminals. It is best to avoid grou nding the output at any poi nt other th an the pow er supply output termin als to avoid n oise problem s caused by common-mode curr e nt flowing through the lo a d lead s to gro und. Alw ...

  • Agilent Technologies 6010A - page 31

    Operating Instructions 31 circuits wit hin the s upply the pow er supply output drops t o zero and the UNREGUL ATED indicator tu rns on. The pow er supply can be disabled by ov ervoltage, overtem perature, or by low or high ac line (main s) voltage. Overvoltage . If th e voltage across the pow er supply ou tput terminals rises abov e a preset level ...

  • Agilent Technologies 6010A - page 32

    32 Operating Instructions crosses the operating lo cus at point 1. P oint 1 is o n the p a rt o f the operating locus d e fined b y the voltage setting, so the pow er supply operates in CV mode. Similarly , the line representing load resistance C, the low est load resistance shown on the graph, crosses the operatin g locus at point 3. Poi nt 3 is o ...

  • Agilent Technologies 6010A - page 33

    Operating Instructions 33 Figure 3-5. Overrange O peration Constant Voltage Operation By pressing the DISPLAY SETT INGS pushbutton switch you can observe the setting (limits) of both the output voltage and the output current, rather than the actual output values. T his you set the current lim it when the power supply is operating in CV mode, or set ...

  • Agilent Technologies 6010A - page 34

    34 Operating Instructions Constant Current Operation To set up the p ower supp ly for consta nt curre nt o per ation: a. With power supply turned off, connect the load to the output terminals. b. Turn on the power supply. Hold in DISPLAY OVP pushbutton switch and set OVP ADJUST po tentiometer for the desired OVP trip voltage. In CC mode the voltage ...

  • Agilent Technologies 6010A - page 35

    Operating Instructions 35 Figure 3-6. Remote Voltage Sensing Because the sensing leads carry only a f ew m illiamperes, the w ires used for sensin g can be much lighter than the load leads. Each sense lead sh ould have no m ore than 0.2 ohm s resistance. Use the resistivity colum ns in Table 3-2 to determin e the min imum w ire size for the length ...

  • Agilent Technologies 6010A - page 36

    36 Operating Instructions Analog Programming These instrum ents can obtain their ou tput voltage and cu rrent program ming in formation from three distin ct sources: 1) locally from the front panel, 2) remotely from an external isolated, voltage source or, 3) remotely from an external isolated, resistance. Mod e switches B 1 , B2 , .thro ugh B6, lo ...

  • Agilent Technologies 6010A - page 37

    Operating Instructions 37 Note: If the program min g terminals (VP to P) become open ci rcuited durin g resistan ce programm ing, the output voltage will rise above the power supply rating. The supply will not be damaged if this occurs, but the overv oltage trip poi nt shou ld be properly adju sted to protect the u ser’ s load. Figure 3-7. Resist ...

  • Agilent Technologies 6010A - page 38

    38 Operating Instructions Figure 3-9. Optional Voltage Divider for Program S our ce Constant Current Output, Resistance Control . The setup s hown in Fig ure 3-10 allow s the ou tput curren t to be varied by using an extern al resistor to prog ram the s upply . A program min g resistor v ariable from 0 to 4000 ohm s produces a proportion al output ...

  • Agilent Technologies 6010A - page 39

    Operating Instructions 39 Constant Current Output, Voltage Control . The setup sho wn in Figure 3-11 allows the output current to be varied by using an external voltage to pro gram the supply. A voltag e source variabl e from 0 t o + 5 volts produces a proport ional outpu t current f rom zero to f ull scale. The st atic load on the programming volt ...

  • Agilent Technologies 6010A - page 40

    40 Operating Instr uctions Figure 3-12. A uto-Parallel Operation Setting Voltage and Current . P rogr am the slave unit's out put volt age ab ove the master 's to avoid inter ferenc e with m ast er- unit CV control. The slave unit's mode sw itches disable the slave unit's digital current setting from having any eff ect in auto- ...

  • Agilent Technologies 6010A - page 41

    Operating Instructions 41 Caution: When two supplies are operated in series, they should be pro gramm ed to the same voltage to prevent possible d a mage to the lower voltage supp ly during short circuit co nd itions. Contact the factory if this is not poss ible. Figure 3-13. Series Operation Monitor Signals Am plified and buf fered voltage and cur ...

  • Agilent Technologies 6010A - page 42

    ...

  • Agilent Technologies 6010A - page 43

    Appendix A 43 A 100 VAC Input Power Option 100 General Information Description Option 100 is a modif ication of t he power s upply t hat inv olves chang ing th e values of resistors l ocated in the Ov ervoltage Protectio n and Po wer Limit Circuits. It also entails recalibrating the unit and changing the Front Panel. T hese changes allow the un it ...

  • Agilent Technologies 6010A - page 44

    44 Appendix A Section 3 Manual Changes On page 30 u nder “Ove rvoltage Protect ion”, chang e 107% to 90%. On page 36 un der “Analog Programmi ng”, in the 3rd parag raph, change the secon d sentence to read: “A resistance of 0 to 3.33 K ohm s program s the out put voltag e from 0 t o full scal e and a resistan ce of 0 to 4 K ohm s program ...

  • Agilent Technologies 6010A - page 45

    Index 45 Index A airflow........................................................................................................................ .................................................. 17 ampacity - w ire ...................................................................................................................................... ...

  • Agilent Technologies 6010A - page 46

    Index 46 R remote se nsing ....................................................................................................................................................... 34, 40 S +S input (see sense leads) -S inpu t (see sense leads) safety class .............................................................................................. ...

  • Agilent Technologies 6010A - page 47

    Agilent Sales and Support Of fice 47 A gilent Sales and Support Office For more inf ormation abou t Agilent Technologies test an d measurem ent products, applications , services, and for a current sales office listing, visit our w eb site: h t t p :/ / www. t m. ag i l e nt . c o m/ You can also cont act one of the follow ing cen ters and ask for a ...

  • Agilent Technologies 6010A - page 48

    48 Manual Updates The following updates have been made to this manu al since the print revision indicated on the title page. 2/01/00 All references to HP have been ch anged to Agilent. All referen ces to HP-IB have been ch anged to GPIB. 9/20/04 The Declaration of Con formity has been u pdated. ...

メーカー Agilent Technologies カテゴリー Power Supply

Agilent Technologies 6010Aのメーカーから受け取ることができるドキュメントは、いくつかのグループに分けられます。その一部は次の通りです:
- #BRANDの図面#
- 6010Aの取扱説明書
- Agilent Technologiesの製品カード
- パンフレット
- またはAgilent Technologies 6010Aの消費電力シール
それらは全部重要ですが、デバイス使用の観点から最も重要な情報は、Agilent Technologies 6010Aの取扱説明書に含まれています。

取扱説明書と呼ばれる文書のグループは、Agilent Technologies 6010Aの取り付け説明書、サービスマニュアル、簡易説明書、またはAgilent Technologies 6010Aのユーザーマニュアル等、より具体的なカテゴリーに分類されます。ご必要に応じてドキュメントを検索しましょう。私たちのウェブサイトでは、Agilent Technologies 6010Aの製品を使用するにあたって最も人気のある説明書を閲覧できます。

関連する取扱説明書

Agilent Technologies 6010Aデバイスの取扱説明書はどのようなものですか?
取扱説明書は、ユーザーマニュアル又は単に「マニュアル」とも呼ばれ、ユーザーがAgilent Technologies 6010Aを使用するのを助ける技術的文書のことです。説明書は通常、全てのAgilent Technologies 6010Aユーザーが容易に理解できる文章にて書かれており、その作成者はその分野の専門家です。

Agilent Technologiesの取扱説明書には、基本的な要素が記載されているはずです。その一部は、カバー/タイトルページ、著作権ページ等、比較的重要度の低いものです。ですが、その他の部分には、ユーザーにとって重要な情報が記載されているはずです。

1. Agilent Technologies 6010Aの説明書の概要と使用方法。説明書にはまず、その閲覧方法に関する手引きが書かれているはずです。そこにははAgilent Technologies 6010Aの目次に関する情報やよくある質問、最も一般的な問題に関する情報を見つけられるはずです。つまり、それらはユーザーが取扱説明書に最も期待する情報なのです。
2. 目次。Agilent Technologies 6010Aに関してこのドキュメントで見つけることができる全てのヒントの目次
3. Agilent Technologies 6010Aデバイスの基本機能を使うにあたってのヒント。 Agilent Technologies 6010Aのユーザーが使い始めるのを助けてくれるはずです。
4. トラブルシューティング。Agilent Technologies 6010Aに関する最も重要な問題を診断し、解決するために役立つ体系化された手続き
5. FAQ。よくある質問
6. 連絡先。一人では問題を解決できない場合に、その国におけるAgilent Technologies 6010Aのメーカー/サービスへの連絡先に関する情報。