    Miniature Multi-Channel Analyzer MCA166-USB User’s Manual Version 3.0 ...

  • Braun MCA166-USB - page 2

    MCA 166- USB Exclusion of liability The GBS Elektronik GmbH is not liable for errors and does not guarantee the specific utility of the MCA166-USB software or firmware. In particular, the GBS Elektronik GmbH is not liable for indirect or subsequent damages due to errors of the MCA166-USB software or firmware. The information in this manual has been ...

  • Braun MCA166-USB - page 3

    Table of Contents 1 Introduction....................................................................................................................... 5 2 Hardware.......................................................................................................................... 6 2.1 Users safety information................................. ...

  • Braun MCA166-USB - page 4

    MCA 166- USB 6.1.4 Preamplifier Voltages..................................................................................... 18 6.1.5 High Voltage.................................................................................................. 18 6.1.6 Battery........................................................................................ ...

  • Braun MCA166-USB - page 5

    1 Introduction The MiniMCA MCA166-USB is a battery powered high performance 4K Multi-Channel Analyzer/Multi-Channel Scaler mo dule comparable in its performance with laborator y grade MCA. High voltage supply for detector and preamplifier power suppl y are integrated as well as an internal amplifier and pulse shaping net work. Together with a small ...

  • Braun MCA166-USB - page 6

    MCA 166- USB 2 Hardware 2.1 Users safety information Read all these instructions first! Save these instructions for later use. Do Not Remove Connectors To avoid personal injury or damage of equipment, do not remove the connectors for the High Voltage supply, preamplifier supply, and the input connector until the High Voltage is shut down and the de ...

  • Braun MCA166-USB - page 7

    2.2.1 Power management The MCA166-USB works with a built in rechargeable Li-Ion battery, which has no memory effect and is deep discharge and short circuit protected. If the device is working, the remaining battery life time is permanently checked. W hen the battery voltage has dropped down to 6.8 Volts, while a data collection is in progress, the ...

  • Braun MCA166-USB - page 8

    MCA 166- USB 2.2.3 Charging the MC A 166- USB Charging the MCAs battery is usually done by connecting the wall adapter to the ”Charger In” connector and power it from the mains supply. The wall adapter is able to deliver enough power to operate the MCA166-USB stationary under a maximum load condition (HV on, detector draws up to 2W preamplifier ...

  • Braun MCA166-USB - page 9

    2.2.5 Connection of detectors To connect a detector switch the device off . Before connecting the high voltage connector of the detector make sure that the built in High Voltage supply has the correct polarity and that the power consumption of the detector does not exceed the maximum value ( 0.5m A or 0.25 Watt). On the right side of the MCA166-USB ...

  • Braun MCA166-USB - page 10

    MCA 166- USB Typical errors which may be observed with the detector preamp signal are: ● Rise time is too slow (>0.5 µs). This may cause that the pile up rejector does not work correctly. There ma y be even a large amount of regular pulses rejected and especially the high energetic part of the spectrum seriously affected. When using such dete ...

  • Braun MCA166-USB - page 11

    3.2 SPEC (MS-DOS), WinSPEC (Windows) These programs are the default programs for measuring spectra. It supports semi- automated measurements of gamma ray spectra with the MCA166-USB. The spectra are stored on mass storage media on a PC. There are t wo variants of W inSPEC. WinSPEC-I (for Inspectors) was designed for standard use. W inSP EC-A (for A ...

  • Braun MCA166-USB - page 12

    MCA 166- USB 3.7 RATE This program supports measurements of count rates with an y radiation detector (HPGe, NaI, CdTe, CdZnTe and neutron counters). In the most convenient mode of operation only the menu button "measurement" needs to be "pressed" to start a measurement and showing the count rates in form of a LCD display and a b ...

  • Braun MCA166-USB - page 13

    3.10 MCAPlot and MCAPrint MCAPlot and MCAPrint are programs to view and print spectra recorded with the MCA166-USB. Up to 32 spectra can be viewed at the same time and be printed on one page. Also some evaluation functions are available. MCAPlot is not included in the MCA price. MCAPrint is a free downgraded version of MCAPlot. 3.11 MMCAEVAL MMCAEV ...

  • Braun MCA166-USB - page 14

    MCA 166- USB 3.13 Miscellaneous For communication and evaluation some libraries for MS Windows and Linux are available. This may help the make own application programs or to integrate the MCA166-USB in an existing system. 4 MCA166-USB parameters for use w ith different detectors This table shall just give an overview of possible settings useful wit ...

  • Braun MCA166-USB - page 15

    5 Some of the most important photon energies Table 3: Photon energies for different isotopes Isotope Half-life Energy (keV) Branching ratio (%) Am241 432.2y 26.34/59.54 2.4/36 Cd109 462.6d 88.03 3.63 Bi214 19.9min, daughter of Ra226 609.31 46.09 Ce139 137.64d 165.86 79.87 Co60 5.271y 1173.23/1332.49 99.85/99.98 Co57 271.8d 14.4/122.06/136.47 9.16/8 ...

  • Braun MCA166-USB - page 16

    MCA 166- USB 6 Technical Data 6.1 MCA166-USB Hardware Specifications 6.1.1 A m plifier 1. Type ● Gaussian programmable shaping amplifier with pile-up rejector and gated base line restorer 2. Gain ● From 1 ... 1000 ● Minimum gain step 0.5 channel for 2K scale ● Gain tuning under visual control of spectrum ● Gain adjustment by peak stabiliz ...

  • Braun MCA166-USB - page 17

    6.1.2 A D C 1. Successive approximation ADC with sliding scale linearisation ● Channel splitting 128, 256, 512, 1K, 2K, or 4K ● Full input voltage range 3V, positive and negative (10V with external attenuator) 2. ADC Dead time (including time nee ded to write into memory) ● < 8 µsec 3. Nonlinearities ● Differential nonlinearity (over 95 ...

  • Braun MCA166-USB - page 18

    MCA 166- USB 6.1.3 Power supply ● Power suppl y adequate for commonly used NaI, HPGe, CdTe and proportional counters ● Current limited mode with indication of overload condition through software; short circuit protected and automated recovery after overload or shortening 6.1.4 Preamplifier Voltages ● ± 12V; maximum current 60mA for each of t ...

  • Braun MCA166-USB - page 19

    6.1.7 Computer Interface ● RS-232C serial port ● USB 1.1; virtual COM port is used ● 38.400Baud and 307.200Baud; both baud rates available on both interfaces ● proprietary software protocol 6.1.8 Mechanical specification ● Size: 155mm x 95mm x 45mm (footprint of palmtop HP200LX) ● Weight: app. 700g (incl. battery ) 6.1.9 Environmental r ...

  • Braun MCA166-USB - page 20

    MCA 166- USB 6.2 Diagrams and pin connections Figure 3: Block Diagram of the MCA166-USB 20 ...

  • Braun MCA166-USB - page 21

    Table 4: Wiring Diagram of the Charger Input Connector (female Lemo 00-302) Contact Signal 1 + charge voltage 2 - charge voltage Table 5: Wiring Diagram of the PC Interface Connector (female Lemo 0S-306) Contact Signal 1 5V USB 2 D- USB 3 D+ USB 4 GND 5 RxD RS232 (Input) 6 TxD RS232 (Output) Table 6: Preamplifier Power Supply Connector (D9 female) ...

  • Braun MCA166-USB - page 22

    MCA 166- USB 7 Troubleshooting General Remark: Before changing anything concerning the hardware, plugging or pulling cables etc. shut down the high voltage and switch off the MCA. Soldering or manipulating with the SMD boards is strongly not recommended for users. We have specialists for that. If y ou have a problem which cannot be solved by the ta ...

  • Braun MCA166-USB - page 23

    5. The threshold of the spectrum seems to be much higher than expected. ● The input polarity may be wrong. Does the threshold depend on radioactive source and count rate? Try some other input polarity. ● P/Z may not be set correctly. Please check if P/Z is correctly adjusted. ● If the lower cutoff rises b y itself, this is most probable cause ...

  • Braun MCA166-USB - page 24

    MCA 166- USB 12. Bad resolution in a HPGe spectrum or from a test generator, specially at high amplification factors ● Check electrical noise and EMC compatibility of surroundings. Some laptops and also some switch mode power supply battery chargers may disturb. Remove all connections from the MCA to ground or to mains voltages. Try a different c ...

  • Braun MCA166-USB - page 25

    A PPENDIX 1 - MCA 166- USB spectral data format The MiniMCA spectral data format is in compliance with the IAEA SPE Spectral Data Format (see SPEDAC PRO User’s Manual Rel. 1.0, IAEA, Feb 1994). All information is stored in a block structured ASCII BSA file so that its content can be viewed with any text editor and the file content can be printed ...

  • Braun MCA166-USB - page 26

    MCA 166- USB Setup Information FILE CONTENT EXPLANATION $ADC: ADC resolution 4096 Channels 0 LLD (lower level discriminator channel) 3967 ULD (upper level discriminator channel) $PRESETS: MCA Presets Integral Live Time, Real Time, Integral (sum of all counts within a ROI) Area (sum of all net counts within a ROI) 10000 Value 1 ROI number 266 332 RO ...

  • Braun MCA166-USB - page 27

    Setup information - continued $MCS_CHANNELS: Multi Channel Scaler (MCS) Setup 4096 Channels $MCS_INPUT: MCS input Extern TTL Extern TTL (external TTL signal), Input Rate (count rate from the internal amplifier / fast discriminator) $MCS_TIME: MCS time per channel 10 Time in msec , 0.01 – 655.35 sec $MCS_SWEEPS: Number of sweeps 0 0 .... 65535 0 M ...

  • Braun MCA166-USB - page 28

    MCA 166- USB Setup information’s – continued FILE CONTENT EXPLANATION $REC_COUNTER: Counter of received commands 2823 $REC_ERROR_COUNTER: Counter of received commands with errors 2 $SPEC_INTEGRAL: Counts in the spectrum 4098917 $ROI_INFO: ROI information 1 266 332 299.74 24.19 1233477 1142868 2066 ROI# Begin End Centroid FWHM Integral Area Area ...

  • Braun MCA166-USB - page 29

    WinSPEC (inspection information if CsRation is attached) FILE CONTENT EXPLANATION $WINSPEC_INFO: WinSPEC inspection information ANALYSIS:CSRATIO Extern analysis application attached with WinSPEC 1999/001 Inspection number 1969/12/31 Inspection date Inspector1/Inspector2 Inspector names DIV1/DIV2 Division or section of the inspectors FACC Facility c ...

  • Braun MCA166-USB - page 30

    MCA 166- USB WinSCAN (inspection information) FILE CONTENT EXPLANATION $WINSCAN_INFO: WinSCAN inspection information 1999/001 Inspection number 1999/11/01 Inspection date Inspector1/Inspector2 Inspector names DIV1/DIV2 Division or section of the inspectors Facility name Facility name FACC Facility code MBA MBA ItemID Item ID Stratum Stratum Item de ...

  • Braun MCA166-USB - page 31

    UF6 / WinUF6 (inspection information) - continued FILE CONTENT EXPLANATION ICOD Instrument code 5 0156 MMCA ID 8289/025 Computer ID NAJ Detector type 9475/029 Detector ID Collimator Collimator Filter Filter U235 / WinU235 (inspection information) FILE CONTENT EXPLANATION $INSP_INFO: U235 / WinU235 inspection information 123 Inspection number 199910 ...

  • Braun MCA166-USB - page 32

    MCA 166- USB A PPENDIX 2 - MCA166-USB algorithm, formulas This appendix lists the algorithms used by the MiniMCA for the following functions: ● Area, Area uncertainty ● Centroid ● FWHM ● Smooth ● Strip ● Energy Calibration The Area and the Area Uncertainty Function: The area algorithm calculates the number of counts above the background ...

  • Braun MCA166-USB - page 33

    Centroid: The peak centroid is the sum of the channel contents 7 times the channel number divided by the sum of the channel contents in the range of the channels above the half maximum: Centroid = i Spectrum Spectrum i i l h i i l h * = = ∑ ∑ where, i = channel number Spectrum i = net contents of channel i FWHM: The Full Width at Half Maximum ( ...

  • Braun MCA166-USB - page 34

    MCA 166- USB Strip: Strip subtracts a specified fraction of the spectrum. The Strip factor can be “positive“ or “negative“ : Spectrum = Spectrum - F * Spectrum y ' y x where Spectrum y ' - Count content of channel i in the resulting spectrum Spectrum y - Count content of channel i in the original spectrum F - Strip factor Spectrum ...

  • Braun MCA166-USB - page 35

    A PPENDIX 3 - Description of the communication DLLs 8 A3.1 MS-Windows DLL for communication with one MCA166-USB The Dynamic Link Library exists as a 16-bit variant (mca_comm.dll) and as a 32-bit variant (mca32com.dll). The DLL must reside in the system directory of Windows or in the same directory as the calling application. In order to include the ...

  • Braun MCA166-USB - page 36

    MCA 166- USB The first argument is a string with the name of th e command. In the following table all available commands (use of capital and small letters is possible), the corresponding firmware specific names and the number of the required parameters are listed. Name of the command Firmware specific name Number of the parameters INIT CMD_INIT 1 S ...

  • Braun MCA166-USB - page 37

    The other arguments are the parameters of the firmware command. These arguments are optional since not all commands use the same number of parameters. Unnecessary arguments you can leave out. Information about the parameters you can find in appendix 4. You can ignore the parameters ”preamble”, ”command” und ”end flag” listed in this doc ...

  • Braun MCA166-USB - page 38

    MCA 166- USB A3.2 Command set MCA Reset Command: Name MMCA_RESET Description All MCA Parameters will be reset to their initial state and the spectra are cleared. The measurement is aborted, but the states of the preamplifier power and the value of the high voltages are unchanged! Parameter - Results - Return Code ERROR_OK = 0 (command accepted succ ...

  • Braun MCA166-USB - page 39

    MCA Mode Commands: Name MMCA_SET_MODE_MCA Description The command sets the acquire mode to Multi-Channel Analyser Parameter - Results - Return Code see MMCA_RESET command Name MMCA_SET_MODE_MCS Description The command sets the acquire mode to Multi-Channel Scaler Parameter - Results - Return Code see MMCA_RESET command MCA Clear Commands: Name MMCA ...

  • Braun MCA166-USB - page 40

    MCA 166- USB MCA Measurement Setup Commands: Name MMCA_SET_PRESET_NONE Description None preset Parameter - Results - Return Code see MMCA_RESET command Name MMCA_SET_PRESET_LIVE_TIME Description The command sets the time for measurement (dead time corrected). 1. Parameter Live time integer value 1 ... 65535 sec Results - Return Code see MMCA_RESET ...

  • Braun MCA166-USB - page 41

    Name MMCA_SET_ADC_RES_DISCR Description The command sets the ADC resolution and the software discriminator range. 1. Parameter Channels integer value 128, 256, 512, 1024, 2048 or 4096 2. Parameter LLD integer value 0 <= LLD < ULD 3. Parameter ULD integer value LLD < ULD <(Channels-(Channels>>5)) Results - Return Code see MMCA_RESE ...

  • Braun MCA166-USB - page 42

    MCA 166- USB Name MMCA_SET_MCS_INPUT_ICR Description The command selects the count rate signal as MCS input. Parameter - Results - Return Code see MMCA_RESET command Name MMCA_SET_MCS_INPUT_TTL Description The command selects the external TTL signals as MCS input. Parameter - Results - Return Code see MMCA_RESET command Name MMCA_SET_MCS_INPUT_DISC ...

  • Braun MCA166-USB - page 43

    MCA Hardware Setup Commands: Name MMCA_SET_GAIN Description The command sets the amplifier coarse and fine gain. 1. Parameter Coarse gain integer value 2, 5, 10, 20, 50, 100, 200, 500 or 1000 2. Parameter Fine gain 15 integer value 5000 ... 15000 (10000 if Coarse Gain=1000) Results - Return Code see MMCA_RESET command Name MMCA_SET_MCA_INPUT_AMPLIF ...

  • Braun MCA166-USB - page 44

    MCA 166- USB Name MMCA_SET_THRESHOLD Description The command sets the analog threshold 1. Parameter Threshold integer value 2 ... 60 Results - Return Code see MMCA_RESET command Name MMCA_SET_SHAPING_TIME_LOW Description The command sets the amplifier shaping time (1µs). 16 Parameter - Results - Return Code see MMCA_RESET command Name MMCA_SET_SHA ...

  • Braun MCA166-USB - page 45

    Name MMCA_SET_FAST Description The command sets the fast discriminator threshold. 1. Parameter Threshold integer value 0 ... 2499, default 400 Results - Return Code see MMCA_RESET command Name MMCA_SET_SLOW Description The command sets the slow discriminator threshold. 1. Parameter Threshold integer value 0 ... 2499, default 400 Results - Return Co ...

  • Braun MCA166-USB - page 46

    MCA 166- USB MCA Stabilisation Commands: Name MMCA_SET_STABILISATION Description The command sets the peak stabilisation. 1. Parameter Stabilisation flag Integer value 0 turns stabilisation off 1 turns stabilisation on (keep peak at current position) >1 stabilisation channel (move peak to the stabilisation channel and hold peak at the stabilisat ...

  • Braun MCA166-USB - page 47

    MCA Data Setup Commands: Name MMCA_SET_TDF Description The command sets the dead time correction parameter. 1. Parameter Tdf integer value 100 ... 3000, default 800 Results - Return Code see MMCA_RESET command Name MMCA_SET_USER_DATA Description The command stores a 32 bit value in the MCA parameter memory. 1. Parameter Number integer value 0 ... 6 ...

  • Braun MCA166-USB - page 48

    MCA 166- USB MCA Query Commands: Name MMCA_QUERY_USER_DATA Description The command reads 32 user data (32 bit values) from the MCA parameter memory. 1. Parameter Number n integer value 0 ... 63 Result array User_data[n], ... long or float value 0 ... Return Code see MMCA_RESET command Contents of the MMCA User Date memory: 21 Byte Offset 0 22 U235: ...

  • Braun MCA166-USB - page 49

    Byte Offset 64 1. ROI pair (ROI End) 25 Byte Offset 66 1. ROI pair (ROI Begin) Byte Offset 68 2. ... ... Byte Offset 120 15. Byte Offset 124 16. ROI pair Byte Offset 128 26 U235/UF6: I nspection description WinMCS: 16 ROI pairs of the MCA spectrum ( similar the other ROIs ) 25 In applications with MCS spectrum the ROIs of the MCS spectrum are saved ...

  • Braun MCA166-USB - page 50

    MCA 166- USB Name MMCA_QUERY_POWER Description The command reads the MMCA power state. Parameter - Result array Battery current [mA] long value Byte Offset 4 HV primary current [mA] long value Byte Offset 8 +12 primary current [mA] long value Byte Offset 12 -12 primary current [mA] long value Byte Offset 16 +24 primary current [mA] long value Byte ...

  • Braun MCA166-USB - page 51

    Name MMCA_QUERY_STATE Description The command reads the MMCA state. Parameter - Results array MCA acquire mode MCA_MODE MODE_MCA, MODE_MCS Byte Offset 2 MCA preset MCA_PRESETS PRESET_NONE PRESET_REAL PRESET_LIVE PRESET_INT PRESET_AREA Byte Offset 4 Preset value long Byte Offset 8 Elapsed preset long Byte Offset 12 Repeat value integer value Byte Of ...

  • Braun MCA166-USB - page 52

    MCA 166- USB Name MMCA_QUERY_STATE STAB_IDLE, STAB_ERROR Byte Offset 72 Stab. ROI begin integer value Byte Offset 74 Stab. ROI end integer value Byte Offset 76 ADC input MCA_INPUT_ADC ADC_AMPLIFIER, ADC_ DIRECT_POS, ADC_DIRECT_NEG Byte Offset 78 ADC input polarity MCA_INPUT_POL INPUT_POLARITY_POSITIVE, INPUT_POLARITY_NEGATIVE Byte Offset 80 DTC int ...

  • Braun MCA166-USB - page 53

    Name MMCA_QUERY_SYSTEM_DATA Description The command reads the MMCA system data. Parameter - Results array 0 integer Byte Offset 2 Peak detect counter 48 Bit integer Byte Offset 8 0 integer Byte Offset 10 Fast detect counter 48 Bit integer Byte Offset 16 0 integer Byte Offset 18 Peak detect counter at time-1 48 Bit integer Byte Offset 24 0 integer B ...

  • Braun MCA166-USB - page 54

    MCA 166- USB Byte Offset 116 Stabilisation area preset long Byte Offset 120 Stabilisation time preset integer v alue Byte Offset 122 Low shaping time 34 byte value Byte Offset 123 High shaping time 35 byte value Byte Offset 124 0 integer value Byte Offset 128 MMCA state STATE_POWER_ON, STATE_READY, STATE_RUN, STATE_SUSPEND, STATE_FINISH, STATE_ERRO ...

  • Braun MCA166-USB - page 55

    Name MMCA_QUERY_SPECTRA Description The command read the MMCA spectrum data. 1. Parameter No. of 1. Channel and read out control 37 integer value 2. Parameter Compress factor (1 ... 32) long value Result array Channel contents N long value Byte Offset 4 Channel contents N+1 long value ... Byte Offset 124 Channel contents N+31 long value Byte Offset ...

  • Braun MCA166-USB - page 56

    MCA 166- USB Name MMCA_QUERY_UF6_INFO Description The command reads current dead time, real time the integral, begin and end of the UF6 ROIs Parameter - Result array Real time long value Byte Offset 4 Dead time [msec] long value Byte Offset 8 Integral ROI 1 long value Byte Offset 12 Integral ROI 2 long value Byte Offset 16 Integral ROI 3 long value ...

  • Braun MCA166-USB - page 57

    Name MMCA_QUERY_VOLTAGE_CURRENT Description The command reads the voltages and currents from the MMCA. Parameter - Results array Charger current [mA] long Byte Offset 4 HV primary current [mA] long Byte Offset 8 Battery current [mA] long Byte Offset 12 Battery voltage [mV] long Byte Offset 16 HV reference voltage [V] long Byte Offset 20 HV control ...

  • Braun MCA166-USB - page 58

    MCA 166- USB A PPENDIX 4 - Description of the MC A 16 6-USB firmware functions MCA Reset Command: Command name CMD_INIT Description All MCA Parameters will be reset to their initial state and the spectra are cleared. The measurement is aborted, but the states of the preamplifier power and the value of the high voltage are unchanged! Command syntax ...

  • Braun MCA166-USB - page 59

    MCA Mode Command: Command name CMD_SET_MODE Description Depending on the value of the first parameter (mod): MODE_MCA = 0 MODE_MCS = 1 the command set MMCA to the required mode by activating the previous specified setup parameter. The command will be ignored and responded with a error messages, if a measurement is still running. Command syntax Form ...

  • Braun MCA166-USB - page 60

    MCA 166- USB MCA Measurement Setup Commands: Command name CMD_SET_ADC_RES_DISCR Description The command set the ADC resolution(res: 128, 256 … 4096), the LLD(Low level discriminator) and the ULD(Upper level discriminator) to the values of its three parameter. The command will be ignored and responded with a error messages, if one ore more paramet ...

  • Braun MCA166-USB - page 61

    Command name CMD_SET_REPEAT Description The command sets the number(rep: 0 ... 65535) of sweeps for repetitive measurement. Command syntax Format integer integer integer long integer Parameter preamble command rep 0 end flag Byte String (HEX) A5 5A 4A 00 rep l rep h 00 00 00 00 B9 9B Used setup value None Affected setup value Repeat MCA state STATE ...

  • Braun MCA166-USB - page 62

    MCA 166- USB MCA Hardware Setup Commands: Command name CMD_SET_GAIN Description The command sets the amplifier coarse(cg) and fine gain(fg). cg: 2, 5, 10, 20, 50, 100, 200, 500 or 1000 fg: 5000 ... 15000 (for cg = 1000 fg: 5000 ... 10000) Command syntax Format integer integer integer long integer Parameter preamble command cg fg end flag Byte Strin ...

  • Braun MCA166-USB - page 63

    Command name CMD_SET_MCS_INPUT Description The command sets the input for the Multi-Channel Scaler. ip = 0 external TTL signals ip = 1 internal count rate signal ip = 2 counts only pulses with amplitude between LLD and ULD threshold Command syntax Format integer integer integer long integer Parameter preamble command ip 0 end flag Byte String (HEX) ...

  • Braun MCA166-USB - page 64

    MCA 166- USB Command name CMD_SET_PUR Description The command sets the pile up rejection. Command syntax pur = 0 t urns PUR off pur ≠ 0 turns PUR on Format integer integer integer long integer Parameter preamble command pur 0 end flag Byte String (HEX) A5 5A 53 00 pur 00 00 00 00 00 B9 9B Used setup value None Affected setup value mca_pur MCA sta ...

  • Braun MCA166-USB - page 65

    Command name CMD_SET_PZC Description The command performs the Pole Zero Cancellation. pv: 0 ... 2499 PZC control voltage pm = 0 set PCZ only pm > 0 set PZC and measure PZC offset Command syntax Format integer integer integer long integer Parameter preamble command pm Pv end flag Byte String (HEX) A5 5A 58 00 pm l pm h pv l pv h 00 00 B9 9B Used ...

  • Braun MCA166-USB - page 66

    MCA 166- USB MCA Stabilisation Commands: Command name CMD_SET_STABILISATION Description The command sets the peak stabilisation. fl: 0 turns stabilisation off 1 turns stabilisation on rb+3<fl<re-3 stabilisation to channel fl rb: Peak ROI begin (LLD<= begin < end) re: Peak ROI end (begin < end <= ULD, (end-begin) < 250) Command ...

  • Braun MCA166-USB - page 67

    MCA Detector Power Commands: Command name CMD_SET_PREAMPLIFIER_POWER Description The command sets the preamplifier power(pp). 0X80 -24V on 0X40 +24V on 0X20 -12V on 0X10 +12V on 0XF0 all on 0X00 all off Command syntax Format integer integer integer long integer Parameter preamble command pp 0 end flag Byte String (HEX) A5 5A 4E 00 pp 00 00 00 00 00 ...

  • Braun MCA166-USB - page 68

    MCA 166- USB MCA Data Setup Commands: Command name CMD_SET_TDF Description The command sets the TDF parameter. tdf: TDF(100 ... 3000, default 800) Command syntax Format integer integer integer long integer Parameter preamble command tdf 0 end flag Byte String (HEX) A5 5A 61 00 tdf l tdf h 0 0 0 0 B9 9B Used setup value None Affected setup value tdf ...

  • Braun MCA166-USB - page 69

    MCA Query Commands: Command name CMD_QUERY_POWER Description The command reads the power state. Command syntax Format integer integer integer long integer Parameter preamble command 0 0 end flag Byte String (HEX) A5 5A 59 00 00 00 00 00 00 00 B9 9B Results see MMCA_QUERY_POWER function in DLL description. Command name CMD_QUERY_STATE Description Th ...

  • Braun MCA166-USB - page 70

    MCA 166- USB Command name CMD_QUERY_USER_DATA Description The command reads a 32 bit value from the MCA parameter memory. e: 0 ... 63 parameter number Command syntax Format integer integer integer long integer Parameter preamble command e 0 end flag Byte String (HEX) A5 5A 5E 00 e 00 00 00 00 00 B9 9B Used setup value None Result see MMCA_QUERY_USE ...

  • Braun MCA166-USB - page 71

    Command name CMD_QUERY_ENRICHMENT Description The command read the integral of two ROIs and the real and dead time. pb: Peak 1 begin bb: Peak 2 begin pl: Peak 1 length (8 ... 250 channels) bl: Peak 2 length (8 ... 250 channels) Command syntax Format integer integer integer long integer Parameter Preamble command end flag Byte String (HEX) A5 5A 5D ...

  • Braun MCA166-USB - page 72

    MCA 166- USB A PPENDIX 5 - Changing the MC A1 66- USB Firmware A t tention: The order of changing the switches must be followed exactly, otherwise the µ P w il l be destroyed! The programming voltage (+12V) shall only be applied when the working voltage (+5 V) is swi tched on !! Each MMCA must be programmed wi th it’s corresponding HEX-file (e.g ...

メーカー Braun カテゴリー Network Card

Braun MCA166-USBのメーカーから受け取ることができるドキュメントは、いくつかのグループに分けられます。その一部は次の通りです:
- #BRANDの図面#
- MCA166-USBの取扱説明書
- Braunの製品カード
- パンフレット
- またはBraun MCA166-USBの消費電力シール
それらは全部重要ですが、デバイス使用の観点から最も重要な情報は、Braun MCA166-USBの取扱説明書に含まれています。

取扱説明書と呼ばれる文書のグループは、Braun MCA166-USBの取り付け説明書、サービスマニュアル、簡易説明書、またはBraun MCA166-USBのユーザーマニュアル等、より具体的なカテゴリーに分類されます。ご必要に応じてドキュメントを検索しましょう。私たちのウェブサイトでは、Braun MCA166-USBの製品を使用するにあたって最も人気のある説明書を閲覧できます。

関連する取扱説明書

Braun MCA166-USBデバイスの取扱説明書はどのようなものですか?
取扱説明書は、ユーザーマニュアル又は単に「マニュアル」とも呼ばれ、ユーザーがBraun MCA166-USBを使用するのを助ける技術的文書のことです。説明書は通常、全てのBraun MCA166-USBユーザーが容易に理解できる文章にて書かれており、その作成者はその分野の専門家です。

Braunの取扱説明書には、基本的な要素が記載されているはずです。その一部は、カバー/タイトルページ、著作権ページ等、比較的重要度の低いものです。ですが、その他の部分には、ユーザーにとって重要な情報が記載されているはずです。

1. Braun MCA166-USBの説明書の概要と使用方法。説明書にはまず、その閲覧方法に関する手引きが書かれているはずです。そこにははBraun MCA166-USBの目次に関する情報やよくある質問、最も一般的な問題に関する情報を見つけられるはずです。つまり、それらはユーザーが取扱説明書に最も期待する情報なのです。
2. 目次。Braun MCA166-USBに関してこのドキュメントで見つけることができる全てのヒントの目次
3. Braun MCA166-USBデバイスの基本機能を使うにあたってのヒント。 Braun MCA166-USBのユーザーが使い始めるのを助けてくれるはずです。
4. トラブルシューティング。Braun MCA166-USBに関する最も重要な問題を診断し、解決するために役立つ体系化された手続き
5. FAQ。よくある質問
6. 連絡先。一人では問題を解決できない場合に、その国におけるBraun MCA166-USBのメーカー/サービスへの連絡先に関する情報。

Braun MCA166-USBについてご質問がありますか?

次のフォームを使用してください

見つけた説明書を読んでもBraun MCA166-USBの問題を解決できない場合、下記のフォームを使用して質問をしましょう。ユーザーのどなたかがBraun MCA166-USBで同様の問題を抱えていた場合、その解決方法を共有したいと考えるかもしれません。

画像のテキストを入力してください

コメント (0)