Invacare IOC100P - page 1
Dealer: This manual MUST be given to the end user. User: Before using this product, read this manual and save for future ref erence. For more information regarding Invacare products, parts, and services, please visit www.invacare.com Operator’s Manual Invacare® Pneumatic Oxygen Conserver IOC100P ...
-
Invacare IOC100P - page 2
WARNING Pneumatic Oxygen Conserver 2 Part No 1160840 WARNING DO NOT use this product or any available optional equipment without first completely reading and understanding these instruct ions and any additional instructional material such as owner’s manuals, service manuals or instruction sheets supplied with this product or optional equipmen ...
-
Invacare IOC100P - page 3
TABLE OF CONTENTS TABLE OF CONTENTS Part No 1160840 3 Pneumatic Oxygen Conserver REGISTER YOUR PRODUCT .............................. 4 SPECIAL NOTES ............................................... 5 Receiving Inspection ................... ........................................................ 5 Intended Use .............. ...................... ...
-
Invacare IOC100P - page 4
REGISTER YOUR PRO DUCT Pneumatic Oxygen Conserver 4 Part No 1160840 REGISTER YOUR PRODUCT The benefits of registering include: 1. Safeguarding your investment. 2. Ensuring long-term maintenance and servicing of your product. 3. Receiving updates wi th product information, maintenance tips and industry news. Register ONLINE at warranty.invacare.com ...
-
Invacare IOC100P - page 5
SPECIAL NOTES Part No 1160840 5 Pneumatic Oxygen C onserver SPECIAL NOTES Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage. Refer to the following table ...
-
Invacare IOC100P - page 6
SPECIAL NOTES Pneumatic Oxygen Conserver 6 Part No 1160840 Disposal Instructions This device and its packaging contain no hazardous materials. No special precautions need to be taken when disposing the device and/or it’ s packaging. Principles of Operation The ...
-
Invacare IOC100P - page 7
SPECIAL NOTES Part No 1160840 7 Pneumatic Oxygen C onserver DANGER DO NOT SMOKE while using this device. Keep all matches, lit cigarettes or oth er sources of ignition out of the room in which this product is located and away from ...
-
Invacare IOC100P - page 8
INTRODUCTION Pneumatic Oxygen Conserver 8 Part No 1160840 INTRODUCTION The oxygen conserving regulator includes a combination of low ‐ pressure regulator and an oxygen conserver, designed for use with a cylinder as an ambulatory oxygen system. It ? ...
-
Invacare IOC100P - page 9
SECTION 1—GENERAL GUIDELINES Part No 1160840 9 Pneumatic Oxygen C onserver SECTION 1—GENERAL GUIDELINES When using the oxygen conserving regulator, basic precautions should AL W A YS be followed, including the following: WARNING SECTION 1 ‐ GENERAL GUIDELINES contai ...
-
Invacare IOC100P - page 10
SECTION 1—GENERAL GUIDELINES Pneumatic Oxygen Conserver 10 Part No 1160840 DO NOT attempt to repair product. DO NOT immerse product in any kind of liquid. Store product in a clean area when not in use. NEVER smoke in an area where ? ...
-
Invacare IOC100P - page 11
SECTION 1—GENERAL GUIDELINES Part No 1160840 11 Pneumatic Oxygen Conserve r DO NOT allow cylinders to tip or fall. Secure gas cylinders so they cannot fall. Ensure cylinder is always in upright position. DO NOT allow oil from your hands ...
-
Invacare IOC100P - page 12
SECTION 1—GENERAL GUIDELINES Pneumatic Oxygen Conserver 12 Part No 1160840 DO NOT place the unit under such items as blankets, bed coverings, chair cushions, clothing and away from heated or hot surfaces, including space heaters, stoves and ...
-
Invacare IOC100P - page 13
SECTION 2—FEATURES Part No 1160840 13 Pneumatic Oxygen Conserve r SECTION 2—FEATURES DANGER This product is not intended as a life ‐ sustaining or life ‐ supporting device. CAUTION Missing or illegible labels MUST be replaced . Contact customer service at ? ...
-
Invacare IOC100P - page 14
SECTION 2—FEATURES Pneumatic Oxygen Conserver 14 Part No 1160840 Oxygen Pressure Gauge: Enables the user to monitor the contents of the compressed oxygen cylinder and is protected by a rubber guard. Selector Switch: Enables the user to sele ...
-
Invacare IOC100P - page 15
SECTION 2—FEATURES Part No 1160840 15 Pneumatic Oxygen Conserve r FIGURE 2.2 Connection view of the IOC100P with CGA870 Connection ...
-
Invacare IOC100P - page 16
SECTION 3—INSTALLING THE CONSERVING REGULATOR Pneumatic Oxygen Conserver 16 Part No 1160840 SECTION 3—INSTALLING THE CONSERVING REGULATOR WARNING DO NOT allow oil from your hands or other source s to come into contact with the regulator gasket or cylinder post ? ...
-
Invacare IOC100P - page 17
SECTION 3—INSTALLING THE CONSERVING REGULATOR Part No 1160840 17 Pneumatic Oxygen Conserve r NOTE: For information on how to operate the cylinder, cons ult your equipment pr ovider. NOTE: For this procedure, refe r to FIGURE 3.1. 1. Loosen the cylinder post ? ...
-
Invacare IOC100P - page 18
SECTION 4—OPERATING INSTRUCTIONS Pneumatic Oxygen Conserver 18 Part No 1160840 SECTION 4—OPERATING INSTRUCTIONS DANGER NEVER smoke in an area where oxygen is being administered. NEVER use near any type of flame or flammable/explosive substances, vapors or at ...
-
Invacare IOC100P - page 19
SECTION 4—OPERATING INSTRUCTIONS Part No 1160840 19 Pneumatic Oxygen Conserve r 4. Listen for leaks. If a leak is present, close the cylinder valve, check the CGA870 seal washer, and reinstall. If the leak persists, DO NOT use the equipment ...
-
Invacare IOC100P - page 20
SECTION 4—OPERATING INSTRUCTIONS Pneumatic Oxygen Conserver 20 Part No 1160840 WARNING NO OXYGEN is delivered in between settings. To obtain desired flow, the indicating pointer MUST align with a specific number on the dial. To avoid injury to ...
-
Invacare IOC100P - page 21
SECTION 4—OPERATING INSTRUCTIONS Part No 1160840 21 Pneumatic Oxygen Conserve r 12. When finished using the system, close th e oxygen supply cylinder valve and continue breathing through the nasa l cannula until no further oxygen flow is detected. 13. ...
-
Invacare IOC100P - page 22
SECTION 5—REMOVING THE CONSERVING REGULATOR Pneumatic Oxygen Conserver 22 Part No 1160840 SECTION 5—REMOVING THE CONSERVING REGULATOR DANGER NEVER attempt to remove the oxygen conserving regulator from an oxygen cylinder unless the cylinder valve is completely closed . ...
-
Invacare IOC100P - page 23
SECTION 6—TYPICAL PRODUCT PARAMETERS Part No 1160840 23 Pneumatic Oxygen Conserve r SECTION 6—TYPICAL PRODUCT PARAMETERS OXYGEN DELIVERY: SELECTOR SWITCH POSITION LITER FLOW EQUIVALEN CY 0O F F 11 22 33 44 55 66 CONTINUOUS FLOW EMERGENCY BYPASS SETTING: Factory Preset at 2 ± .5 L/min. REQUIRED OPERATING PRESSURE: 200 psi to 3000 psi (1 3.8 bar ...
-
Invacare IOC100P - page 24
SECTION 6—TYPICAL PRODUCT PARAMETERS Pneumatic Oxygen Conserver 24 Part No 1160840 STORAGE/ TRANSPORTATION: Maximum -40°F (-40°C), 1% RH Maximum 145°F (63°C), 44% RH SHOCK VIBRATION: Not to exceed IEC 601-1 requirements Not to exceed IEC 68-2-6, IEC 68-2-34 No smoking or open flames Consult accompanying documents ...
-
Invacare IOC100P - page 25
SECTION 6—TYPICAL PRODUCT PARAMETERS Part No 1160840 25 Pneumatic Oxygen Conserve r Oxygen Cylinder Duration The total delivery of o xygen via the conserver is related to the breathing rates, it is user adaptive in that the total oxygen delivered per ? ...
-
Invacare IOC100P - page 26
SECTION 7—MAINTENANCE Pneumatic Oxygen Conserver 26 Part No 1160840 SECTION 7—MAINTENANCE CAUTION DO NOT use cleaning solutions. DO NOT immerse product in any kind of liquid. All repairs MUST be done by Invacare Corporation. 1. Disconnect all connections ...
-
Invacare IOC100P - page 27
SECTION 8—TROUBLESHOOTING Part No 1160840 27 Pneumatic Oxygen Conserve r SECTION 8—TROUBLESHOOTING If oxygen conserving regulator fails to function, consult the T roubleshooting guide. If problem cannot be corrected, consult your Dealer/Home Care equipment provi ...
-
Invacare IOC100P - page 28
NOTES Pneumatic Oxygen Conserver 28 Part No 1160840 NOTES ...
-
Invacare IOC100P - page 29
NOTES Pneumatic Oxygen Conserver 29 Part No 1160840 NOTES ...
-
Invacare IOC100P - page 30
LIMITED WARRANTY Pneumatic Oxygen Conserver 30 Part No 1160840 LIMITED WARRANTY PLEASE NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4, 1975. This warranty is extended only to the original purchaser/user of our products. This warranty gives you specif ic legal rights and you may ...
-
Invacare IOC100P - page 31
LIMITED WARRANTY Part No 1160840 31 Pneumatic Oxygen Conserve r LIMITATIONS AND EXCLUSIONS : THE FOREGOING WARRANTY SHALL NOT APPL Y TO SERIAL NUMBERED PRODUCTS IF THE SERIAL NUMBER HAS BEEN REMOVED OR DEFACED; PRODUCTS SUBJE CTED TO NEGLIGENCE, ACCIDENT, IMPROPER OPERATION, MAINTENANCE, OR STORAGE; PRODUCTS MODIFIED WITHOUT INVACARE’S EXPRESS WR ...
-
Invacare IOC100P - page 32
Invacare Corporation www.invacare.com USA One Invacare Way Elyria, Ohio USA 44036-2125 800-333-6900 Technical Services 800-832-4707 Distributed by Invacare All rights reserved. Trademarks are identified by ™ and ®. All trademarks are owned by or licensed to Invacare Corporation or its subsidiaries unle ss otherwise noted. WD-40 is a registered t ...
Do you have a question concerning Invacare IOC100P?
Use the form below
If you did not solve your problem by using a manual Invacare IOC100P, ask a question using the form below. If a user had a similar problem with Invacare IOC100P it is likely that he will want to share the way to solve it.