    Weighing Indicator for Check Weighing Scales 1WMPD4002313 ...

  • A&D 1WMPD4002313 - page 2

    This Manual and Marks All safety messages are identified by the following, “W ARNI NG” or “CAUTION”, of ANSI Z535.4 (American National S tandar d Institute: Product Safety Signs and Labels). The meanings are as follows: WA R N I N G A potentially hazardous situatio n which, if not avoided, could result in death or serious injury . CAUTION A ...

  • A&D 1WMPD4002313 - page 3

    1 Content s 1. COMPLI ANCE ................................................................................................. 2 2. INTRO DUCTION .............................................................................................. 3 3. UNP ACKING ............................................................................................... ...

  • A&D 1WMPD4002313 - page 4

    2 1. COMPLIANCE Compliance with FCC rules  Please note that this equipment generat es, uses and can radiate radio frequency energy . This equipment has been tested and has been found to comply with the limits of a Class A computing de vice pursuant to Subpart J of Part 15 of FCC rules. These rules are designed to provide r easonable protection a ...

  • A&D 1WMPD4002313 - page 5

    3 2. INTRODUCTION This manual describes how this product wor ks and how to get the most out of it in terms of performance. The FS - D weighing indicator is used to bu ild high performance check weighing scales. It has similar specifications and functions as an indicator of the FS - i series. Please refer to the instruction manual for the FS - i sup ...

  • A&D 1WMPD4002313 - page 6

    4 3. UNP ACKING Unpack the indicator carefully and keep the p acking material if you are likely to transport the indicator again in the future. When unpacking, check whether all of the following items are included: FS- i series INSTRUCTION MANUAL Instruction manual FS- i series FS-D SET UP INSTRUCTION Set Up Instruction (this manual) Connector plug ...

  • A&D 1WMPD4002313 - page 7

    5 4. INST ALLA TION and PRECAUTIONS 4-1. Precautions for Installing the Indicator  Ground the indicator so that the user will not be subjected an electric shock.  Do not handle the main power cord with wet hands.  The AC plug is not water-resist ant. Install it in an area where it does not get w et.  Do not install the indicator where t ...

  • A&D 1WMPD4002313 - page 8

    6 Setscrew B Setscrew A 4. Solder the wires to the pin contact s of the connector .  A 5 m or less load cell cable is recommended as the FS - D uses a 4 wire system (no remote sensing). 5. Reassemble the connector plug and tighten the setscrew A. 6. Move the load cell cable back and forth and around, and tighten each part s again. Finally , tigh ...

  • A&D 1WMPD4002313 - page 9

    7 4-4. Setting Up the Indicator  Connect the main power cord to an outlet that has an earth ground terminal. Y ou may use the earth terminal on the rear si de of the display to ground the indicator .  If necessary , adjust the viewing angle of the display by loosening the 2 clamps, changing the angle and re-tightening the clamps.  The Disp ...

  • A&D 1WMPD4002313 - page 10

    8 5. NAMES AND FUNCTIONS No. NAMES FUNCTIONS / OPERA TION WEIGHT DA T A DISPLA Y Indicates the weight data. WEIGHING UNIT INDICA T OR Indicates the weighing unit in use. ANALOG WEIGHT DISPLA Y Indicates zero to full scale or comparison limit s and results. OVER RANGE INDICA TORS T urns ON when the weight is outside the range of the analog sweep dis ...

  • A&D 1WMPD4002313 - page 11

    9 No. NAMES FUNCTIONS / OPERA TION COMP ARISON INDICA T ORS Indicates the results of the weight comparison LO (RED), OK (GREEN) and HI (YELLOW). ST ABLE ANNUNCIA TOR T urns ON when the weight reading is ST ABLE. ZERO ANNUNCIA TOR T urns ON when the weight is in the center of ZERO range. NET ANNUNCIA T OR T urns ON when the display shows the NET wei ...

  • A&D 1WMPD4002313 - page 12

    10 6. FUNCTION P ARAMETER SETTINGS T o design a platform scale, se t the cap acity , weighing unit and other functions according to the specifications of the scale. Normally , end users do not have to set these function parameters and these setting can be sealed together with the calibration switch. 6-1. The procedure for setting parameters  Do ...

  • A&D 1WMPD4002313 - page 13

    11 6-2. Function list Item Function number Description  C01- 0 CAL switch enabled, 0 key enabled C01- 1 CAL switch disabled, 0 key disabled Calibration C01- 2 CAL switch enabled, 0 key disabled Do not set “ C01-1 ”  C02- 0 6 kg (15 lb) C02- 1 15 kg (35 lb) C02- 2 30 kg (70 lb) C02- 3 60 kg (150 lb) C02- 4 150 kg (350 lb) C02- 5 300 kg (70 ...

  • A&D 1WMPD4002313 - page 14

    12 7. CALIBRA TION This function adjusts the scale (i ndicator) for accurate weighing. Calibration must be done when the FS - D is initially connected to a load cell (base unit). Calibration may also be required acco rding the changes in the environment.  When the scale (indicator) has been moved.  When the ambient environment has significant ...

  • A&D 1WMPD4002313 - page 15

    13  If you need ZERO calibration but do not need SP AN calibration, turn the power off to exit from the calibration procedure. 7. T o calibrate with a different weight, change the displayed value using the 10-key pad.  If you en ter the wrong number , press the C key . The value returns to the capacity and enter the number again. 8. Place a c ...

  • A&D 1WMPD4002313 - page 16

    14 8. AUT O-T ARE FUNCTION The FS - D has an auto-tare function to be used with the comp arator function enabled. If the weight value is within the OK r ange and stable for a preset time, then the indicator will automatically tare the weight and show zero. The Function “ f22 ” designates the timing to tare automatically . See “ 9. ADDITIONAL ...

  • A&D 1WMPD4002313 - page 17

    15 9. ADDITIONAL INFORMA T ION This section describes the additional functions and corrections to the instruction manual of FS - i series (WM+PD4001332 and 1WMPD4001368) which are common to the FS - D. 9-1. F-Function list The following function settings are added (“ 1 1-2. Function list ” in WM+PD4001332). Item Function number Description f06- ...

  • A&D 1WMPD4002313 - page 18

    16 The frame ground and the shie ld are connected internally and there is no difference between them in use. Data form at  Example of data Out of range “kg” (+) O L , +99999 9 9 9 _ k g C R L F Out of range “kg” (+) O L , +9999 . 9 9 9 _ k g C R L F The position of decimal point is diffe rent according to the model and/or “ f02 ”. Ex ...

  • A&D 1WMPD4002313 - page 19

    17 Command H I , + 0 0 1 0 0 C R L F A s su m e 2 d e cimal place number . Reply H I , + 0 0 1 0 0 C R L F 1.00% will be set as LO limit %. Command L O , + 0 0 1 0 0 C R L F A s su m e 2 d e cimal place number . Reply L O , + 0 0 1 0 0 C R L F 1.00% will be set as LO limit %.  Store the comparator limits into the specified memory number. (N o r ...

  • A&D 1WMPD4002313 - page 20

    18 10. ERROR MESSAGES Overload error W arning to indicate that an object beyond the weighing capacity has been placed on the weighing p an. Remov e the object from the weighing pan. Power-on error W arning to indicate that an object beyond the power-on zero range (50% of the weighing capacity or 10%) i s placed on the weighing platfo rm. Remove the ...

  • A&D 1WMPD4002313 - page 21

    19 1 1. SPECIFICA TIONS 1 1-1. Capacity and minimum display C02 - setting 0 * 1 2 3 4 5 6 7 Capacity 6 15 30 60 150 300 600 1500 0.002 * 0.005 0.01 0.02 0.05 0.1 0.2 0.5 0.001 0.002 0.005 0.01 0.02 0.05 0.1 0.2 kg Min. display 0.0005 0.001 0.002 0.005 0. 01 0.02 0.05 0.1 Capacity 6000 15000 30000 2 * 5 10 1 2 5 g Min. display 0.5 1 2 NA NA NA NA NA ...

  • A&D 1WMPD4002313 - page 22

    20 1 1-3. Options OP - 02 (HC - 02 i ) SLA Battery (Y uasa Battery NP4-6 recommended.) OP - 03 (FS - 03 i ) RS - 232C + Comp arator Rela y Output (See note.) OP - 04 (FS - 04 i ) RS - 422/485 + Comp arator Re lay Output (See note.) Note OP - 03 and OP - 04 cannot coexist. 1 1-4. Dimensions Unit: mm ...

  • A&D 1WMPD4002313 - page 23

    ...

  • A&D 1WMPD4002313 - page 24

    3-23-14 Higashi-Ikebukuro, Toshima-ku, Tokyo 170-0013 JAPAN Telephone: [81] (3) 5391-6132 Fax: [81] (3) 5391-6148 A&D ENGINEERING, INC. 1756 Automation Parkway, San Jose, California 95 131 U.S.A. Telephone: [1] (408) 263-5333 Fax: [1] (408)263-0119 A&D INSTRUMENTS LIMITED <UK Office> Unit 24/26 Blacklands Way, Abingdon Busine ss Park, ...

Manufacturer A&D Category Scale

Documents that we receive from a manufacturer of a A&D 1WMPD4002313 can be divided into several groups. They are, among others:
- A&D technical drawings
- 1WMPD4002313 manuals
- A&D product data sheets
- information booklets
- or energy labels A&D 1WMPD4002313
All of them are important, but the most important information from the point of view of use of the device are in the user manual A&D 1WMPD4002313.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals A&D 1WMPD4002313, service manual, brief instructions and user manuals A&D 1WMPD4002313. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product A&D 1WMPD4002313.

Similar manuals

A complete manual for the device A&D 1WMPD4002313, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use A&D 1WMPD4002313 by users. Manuals are usually written by a technical writer, but in a language understandable to all users of A&D 1WMPD4002313.

A complete A&D manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual A&D 1WMPD4002313 - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the A&D 1WMPD4002313, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the A&D 1WMPD4002313, that we can find in the current document
3. Tips how to use the basic functions of the device A&D 1WMPD4002313 - which should help us in our first steps of using A&D 1WMPD4002313
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with A&D 1WMPD4002313
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of A&D 1WMPD4002313 in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning A&D 1WMPD4002313?

Use the form below

If you did not solve your problem by using a manual A&D 1WMPD4002313, ask a question using the form below. If a user had a similar problem with A&D 1WMPD4002313 it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (0)