    15 Liberty St., Metuchen NJ 08840 1-800-LAB-SPEX (1-800-522-7739) ext 465 6870 FREEZER/MILL ® OPERATING MANUAL This manual is part number 85018 ...

  • Allied Telesis 6870 - page 2

    Page 2 of 25 SPEX SamplePrep LLC 9.24.07 TABLE OF CONTENTS SECTION DESCRIPTION PAGE 1.0 Introd uction .................................................................................................................................. 3 2.0 Specif icati ons .............................................................................................. ...

  • Allied Telesis 6870 - page 3

    Page 3 of 25 SPEX SamplePrep LLC 9.24.07 SECTION 1.0: INTRODUCTION SPEX SamplePrep Freezer/ Mills are cryogenic laboratory mills that cool samples to cryogenic temperatures, and pulverize them by m agnetically shuttling a steel impactor back and forth against two stationary end plugs. Since the vial is closed, the integrity o f its contents is main ...

  • Allied Telesis 6870 - page 4

    Page 4 of 25 SPEX SamplePrep LLC 9.24.07 SECTION 2.0: 6870 FREEZER/MILL SPECIFICATIONS Type of mill: cryogenic impact mill Grinding mechanism: steel im pactor driven by dual electromagnets Coolant: liquid nitrogen Weight (empty, without vial or coolant): a pprox. 44 lbs. (20 Kg) Grinding Vials: one 6801 Large Vial, one Mid-Size Vial (6871, 6881, 68 ...

  • Allied Telesis 6870 - page 5

    Page 5 of 25 SPEX SamplePrep LLC 9.24.07 SECTION 3.0: UNPACKING Inspect the exterior of the packing box and repor t any visible damage to the carrier. Remove all packing documents from the exterior of the box, a nd save them for your records. Open the top of the shipping box. Remove the packing material and accessories, and gen tly remove the Freez ...

  • Allied Telesis 6870 - page 6

    Page 6 of 25 SPEX SamplePrep LLC 9.24.07 6870 FREEZER/MILL, BACK VIEW SECTION 4: SETTING UP The 6870 Freezer/Mill weighs 44 pounds (20 Kg) empty. The lid opens fro m the front and the lid latch is in the center. The AC Input module is on the back of the cab inet, and includes the power switch and fuse tray. The fuse tray can be opened with small sc ...

  • Allied Telesis 6870 - page 7

    Page 7 of 25 SPEX SamplePrep LLC 9.24.07 The control panel is a touch-screen, activated by a stylus. The Freezer/Mill is progra mmed and operated through this control panel, whose successive screens are described and illustrated in Section 5, starting on page 8. Programmable parame ters include pre-cooling time, grinding tim e, cooling time between ...

  • Allied Telesis 6870 - page 8

    Page 8 of 25 SPEX SamplePrep LLC 9.24.07 4.4 Grinding Vial Sizes, Adapters, and Vial Openers. Large vials (6801, 6871) are run one at a time in the 6870 Freezer/Mill, w ithout any adapter. Large vials are opened by the 6804 Extractor/Vial Opener, which is p art of the 6870L Accessory Package. Mid-Size vials (6881, 6883, 6885) ar e run one at a time ...

  • Allied Telesis 6870 - page 9

    Page 9 of 25 SPEX SamplePrep LLC 9.24.07 SECTION 5.0: TOUCHSCREEN DISPLAYS The 6870 Freezer/Mill is programmed and operated through a series of touchscreen displays. Transition between screens, a nd all programming and operating commands, are done by touching the screen with a stylus. D o not use a sharp point as this can damage or deface the scree ...

  • Allied Telesis 6870 - page 10

    Page 10 of 25 SPEX SamplePrep LLC 9.24.07 6) Third cycle: the coil is activated for the final 2-minute cycle. When it is done, the program is over and the sample has been ground for a total of 6 m inutes. 5.3 STATUS SCREEN To run the program displayed on the CURRENT SETTINGS screen, touch “Start SCRN.” This transfers you to the STATUS screen, w ...

  • Allied Telesis 6870 - page 11

    Page 11 of 25 SPEX SamplePrep LLC 9.24.07 During the PRECOOL stage the TIME REMAINING line co unts down the time for that stage in 5-second increments: 4:45, 4:40, 4:35, etc. After the PRECOOL stage comes the first GRINDI NG stage, and the screen displays STATUS: GRINDING. The TIME REMAINING line will displa y the time left in that grinding stage. ...

  • Allied Telesis 6870 - page 12

    Page 12 of 25 SPEX SamplePrep LLC 9.24.07 impactor strikes both ends of the vial, a rate of 10 represents 20 impacts per second. T he factory default setting for the rate is 10, and any saved program w ill revert to a rate of 10. To stop a program, touch STOP. The program will end. If RUN is pressed the entire program will start over. To pause duri ...

  • Allied Telesis 6870 - page 13

    Page 13 of 25 SPEX SamplePrep LLC 9.24.07 To store the new program on the CURRENT SETTINGS screen, or to recall a stored program to override the new program, touch STO/RCL to take you to the first of two STORE/RECALL screens: 5.5 STORE/RECALL SCREENS Rcl Sto PRESET O: FISH SCALES Lbl Rcl Sto PRESET 1: Lbl Rcl Sto PRESET 2: Lbl Rcl Sto PRESET 3: Lbl ...

  • Allied Telesis 6870 - page 14

    Page 14 of 25 SPEX SamplePrep LLC 9.24.07 As you touch the letter/number keys, what you ha ve entered appears above the keyboard in a mini-screen with a centered cursor. The SHIFT key lets you shift back and forth from upper case to lower case. Touching the SPACE key advances the cursor one space. Touching the BACK key sends th e cursor back one sp ...

  • Allied Telesis 6870 - page 15

    Page 15 of 25 SPEX SamplePrep LLC 9.24.07 Note that there are three different sizes of vial that can be ru n in the 6870 Freeze r/Mill: Large, Mid-Size, and Small. Mid-Size vials must be run with a 6886 Adapter in the coil, and Sm all vials must be run with a 6806 Adapter in the coil. Large vials do not need an Adapter. See Section 4.4 for details. ...

  • Allied Telesis 6870 - page 16

    Page 16 of 25 SPEX SamplePrep LLC 9.24.07 procedure for your own materials should be dete rm ined by trials; in practice almost no two materials grind alike. Avoid overfilling vials as this greatly reduces grinding efficiency. If at first the sample doesn’t gri nd as intended, try these strategies: • Reduce the overall weight of the sample. (So ...

  • Allied Telesis 6870 - page 17

    Page 17 of 25 SPEX SamplePrep LLC 9.24.07 Accessory Packs can be placed in an insulated container to hold the v ials upright during chilling. At the same time, chill an im pactor separately. As soon as the submerged end of the vial is chilled, drop in the cold sam ple and then the chilled impactor. (Always wear insulated gloves when handling chille ...

  • Allied Telesis 6870 - page 18

    Page 18 of 25 SPEX SamplePrep LLC 9.24.07 6.32 Opening the Vials To open a vial, slip the open end of the appropr iate Extracto r/Vial Opener over a threaded end- plug, align the pegs in the end-plug with the sl ots in the Extractor, and turn the knob clockwise until the end-plug is drawn out. If the end-plug is drawn out within a m illimeter or tw ...

  • Allied Telesis 6870 - page 19

    Page 19 of 25 SPEX SamplePrep LLC 9.24.07 The impactor will have to be separated from the sample and can be handled with gloved fingers, a strong magnet, tongs, or pliers. Tools without a good gripping surface are li kely to slip off the impactor. If some condensation on a cold sam ple is unaccept able, you can either wait for the sample vial to wa ...

  • Allied Telesis 6870 - page 20

    Page 20 of 25 SPEX SamplePrep LLC 9.24.07 6.4 Checking The Liquid Nitrogen Level Every time you remove or insert a vial, note the le vel of liquid nitrogen (LN) in the tub. You should top off the LN when needed as a matter of routine. If your programmed grinding cycle includes more than twenty or th irty minutes of actual grinding time, you should ...

  • Allied Telesis 6870 - page 21

    Page 21 of 25 SPEX SamplePrep LLC 9.24.07 Operation : when the 6870 Freezer/Mill is set up for the firs t run of the day, with a loaded vial in the coil, make sure the LN valve on the supply tank or line is open. Go to the STATUS screen and touch RUN. The Auto-Fill valve will open and let LN into the Free zer/Mill tub. The valve may close and open ...

  • Allied Telesis 6870 - page 22

    Page 22 of 25 SPEX SamplePrep LLC 9.24.07 temperature before attempting to clean it. Once the unit has attained room tem perature, wipe down the unit as indicated previously. Af ter every period of use, the m ill is going to be wet with condensation since the Freezer/Mill will pi ck up moisture as long as it is cold. To maintain the exterior of the ...

  • Allied Telesis 6870 - page 23

    Page 23 of 25 SPEX SamplePrep LLC 9.24.07 7.04 Contacting SPEX SamplePrep LLC Within the United States, write, telephone, or e-mail us at: SPEX SamplePrep LLC 15 Liberty St. Metuchen, New Jersey 08840 Tel.: 732-549-7144 or 800-522-7739 Fax: 732-906-2492 Website: www.spexcsp.com E-mail: sampleprep@spexcsp.com Outside the United States, contact the S ...

  • Allied Telesis 6870 - page 24

    Page 24 of 25 SPEX SamplePrep LLC 9.24.07 SECTION 8.0: TROUBLE-SHOOTIN G GUIDE, PAGE 1 OF 2 PROBLEM CAUSE SOLUTION 1 The unit does not turn on No power Ma ke sure power cord is plugged into outlet and inlet 2 The unit does not turn on No power Check outlet for power and correct as required 3 The unit does not turn on Power switch not turned on Swit ...

  • Allied Telesis 6870 - page 25

    Page 25 of 25 SPEX SamplePrep LLC 9.24.07 SECTION 9.0: TROUBLE-SHOOTIN G GUIDE, PAGE 2 OF 2 PROBLEM CAUSE SOLUTION 14 Blunt end-plug doesn’t come off room-temperature vial Hard to grip Warm vial under hot water. Use wide-jaw pliers if necessary. 15 Extractor jams Water on extractor screws or bell Dry extractor before each use 16 Extractor jams Ic ...

Manufacturer Allied Telesis Category Freezer

Documents that we receive from a manufacturer of a Allied Telesis 6870 can be divided into several groups. They are, among others:
- Allied Telesis technical drawings
- 6870 manuals
- Allied Telesis product data sheets
- information booklets
- or energy labels Allied Telesis 6870
All of them are important, but the most important information from the point of view of use of the device are in the user manual Allied Telesis 6870.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals Allied Telesis 6870, service manual, brief instructions and user manuals Allied Telesis 6870. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product Allied Telesis 6870.

A complete manual for the device Allied Telesis 6870, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use Allied Telesis 6870 by users. Manuals are usually written by a technical writer, but in a language understandable to all users of Allied Telesis 6870.

A complete Allied Telesis manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual Allied Telesis 6870 - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the Allied Telesis 6870, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the Allied Telesis 6870, that we can find in the current document
3. Tips how to use the basic functions of the device Allied Telesis 6870 - which should help us in our first steps of using Allied Telesis 6870
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with Allied Telesis 6870
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of Allied Telesis 6870 in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning Allied Telesis 6870?

Use the form below

If you did not solve your problem by using a manual Allied Telesis 6870, ask a question using the form below. If a user had a similar problem with Allied Telesis 6870 it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (0)