    Models 396 and 398 pH/ORP Sensors Instruction Sheet PN 51A-396/rev .F January 201 1 SPECIFICA TIONS pH Range*: AccuGlass: 0-14 pH Wetted Materials: Model 396: polypropylene, 316 stainless steel, glass, EP Model 398: T efzel; titanium; glass; and choice of Kalrez, Viton or EPDM o-rings Interconnecting Cable: 15 ft (4.6 m), for use with remote preamp ...

  • Emerson 398 - page 2

    STORAGE 1. It is recommended that electrodes be stored in their original shipping containers until needed. 2. Do not store at temperatures below -5°C (23°F). 3. Electrodes should be stored with a protective cap containing KCl solution (PN 9210342). 4. For overnight storage, immerse the sensor in tap water or 4 pH buffer solution. 5. A pH glass el ...

  • Emerson 398 - page 3

    3 ORP CALIBRA TION 1. After making an electrical connection between the sensor and the instrument, obtain a st andard solu- tion of saturated quinhydrone. This can also be made quite simply by adding a few crystals of quin- hydrone to either pH 4 or pH 7 buffer . Quinhydrone is only slightly soluble, therefore only a few crystals will be required. ...

  • Emerson 398 - page 4

    4 FIGURE 1. Sensor Dimensions MILLIMETER INCH ...

  • Emerson 398 - page 5

    5 FIGURE 2. Flow Through and Insertion Inst allation For Models 396 and 398 1-1/2 inch Pipe T ee (PN 200201 1) with 1 inch threaded connections. NOTE: The sensor must be mounted at least 10° above the horizon. ...

  • Emerson 398 - page 6

    6 FIGURE 3. Submersion Inst allations For Models 396 and 398 MILLIMETER INCH ...

  • Emerson 398 - page 7

    7 FIGURE 4. Models 396 and 398 St andard Wiring Option used with Models 1054A, 1054B, 2054, and Preamplifiers PN 22698-02 and 22698-03 FIGURE 5. Models 396 and 398 Option -62 for Wiring to Models 1054, 1 181, 2081, 54e, 81, 3081, 4081, 5081-P , Xmt, SCL-P/Q, and 2700 Preamplifier (PN 23054-03). Also for use with remote Junction Box PN 23555-00. Fig ...

  • Emerson 398 - page 8

    8 MILLIMETER INCH SENSOR MODELS 54, 54e, 3081, 81, 4081, 5081-P pH/ORP FIGURE 7. Wiring to Junction Box (PN 23555-00) and Remote Preamplifier . ...

  • Emerson 398 - page 9

    9 FIGURE 10. Wiring to Model 1055-22-32. FIGURE 8. Wiring to Model Xmt FIGURE 9. Wiring to Models 3081, 4081, and 5081-P ...

  • Emerson 398 - page 10

    10 FIGURE 1 1. Wiring to Model 1055-22-32 through a Remote Junction Box FIGURE 12. Wiring to Model 1055-01-10-22-32 ...

  • Emerson 398 - page 11

    11 FIGURE 13. Prep aring Coax Cable with BNC Connector for Wiring to Models 54, 54e, 81, 2081, 3081, 4081, 5081-P , and Xmt-P FIGURE 14. Prep aring Coax Cable for Wiring to Model 1 181 ...

  • Emerson 398 - page 12

    For more detailed instructions, see electronic file 51-396to398 on “Instruction Manuals” CD-rom. FIGURE 17. Wiring to Model 2081 FIGURE 18. Wiring to Models 1054A/B and 2054 SENSOR FIGURE 16. Wiring to Model 1 181 FIGURE 15. Wiring to Models 54e and 81 Credit Cards for U.S. Purchases Only . The right people, the right answers, right now . ON-LI ...

Manufacturer Emerson Category Photo Scanner

Documents that we receive from a manufacturer of a Emerson 398 can be divided into several groups. They are, among others:
- Emerson technical drawings
- 398 manuals
- Emerson product data sheets
- information booklets
- or energy labels Emerson 398
All of them are important, but the most important information from the point of view of use of the device are in the user manual Emerson 398.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals Emerson 398, service manual, brief instructions and user manuals Emerson 398. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product Emerson 398.

Similar manuals

A complete manual for the device Emerson 398, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use Emerson 398 by users. Manuals are usually written by a technical writer, but in a language understandable to all users of Emerson 398.

A complete Emerson manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual Emerson 398 - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the Emerson 398, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the Emerson 398, that we can find in the current document
3. Tips how to use the basic functions of the device Emerson 398 - which should help us in our first steps of using Emerson 398
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with Emerson 398
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of Emerson 398 in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning Emerson 398?

Use the form below

If you did not solve your problem by using a manual Emerson 398, ask a question using the form below. If a user had a similar problem with Emerson 398 it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (0)