    SULLIV AN ® Comfort  Bi-level System Operating Manual ...

  • ResMed Comfort - page 2

    Please read and understand this manual before using the system. Record of information for future reference Complete the following details when you receive your SULLIVAN ® Comfort. Unit prescribed by (physician) _______________________________ Sleep clinic _______________________________ Date prescribed _______________________________ Prescribed pr ...

  • ResMed Comfort - page 3

     Contents 3 Content s 1 Introduction ...................................................................................................... 4 How ...

  • ResMed Comfort - page 4

    SULLIVAN Comfort Operating Manual  4 1 Introduction How bi-level treatment works Obstructive sleep apnea or OSA occurs when the tongue and soft palate relax during sleep, blocking the airway and cutting off the b ...

  • ResMed Comfort - page 5

    Section 1: Introduction 5 Features of SULLIV A N ® Comfort Delay timer If you find the initial sensation of pressure uncomfortable while trying to fall asleep you may use the delay timer. This allows you to select the time it will take (5, 10 or 20 minutes) for the air pressure in your mask to build up gradually to your prescribed pressures. Yo ...

  • ResMed Comfort - page 6

    SULLIVAN Comfort Operating Manual  6 SULLIVAN Comfort bi-level system 2 Component s of SULLIV A N ® Comfort You should take time to become familiar with the components of your SULLIVAN ® Comfort bi-level sys ...

  • ResMed Comfort - page 7

    Section 3: Preparing for use 7 Prep aring for use Assembling the mask system The mask frame has two access ports which should be plugged and are not used except for initial pressure setting by a technican. It has a swivelling mask elbow which allows the air tubing to be positioned without kinks. The air outlet vent allows exhaled air to escape quie ...

  • ResMed Comfort - page 8

    SULLIVAN Comfort Operating Manual  8 4 Fit the headgear to the mask frame as shown below. • Thread the headgear straps through the slots in the mask frame – the lower straps should also pass through the slots ...

  • ResMed Comfort - page 9

    Section 3: Preparing for use 9 2 Switch the main switch at the rear of the flow generator on ( I ). Press the START button on the front panel (see figure below). Allow the unit to run for a few seconds, then press the STOP button. STOP button Front panel 3 Positioning the flow generator • Place the flow generator on a table close to the head of y ...

  • ResMed Comfort - page 10

    SULLIVAN Comfort Operating Manual  10 Connect the air tubing to the flow generator and mask IMPORTANT:  Make sure the power cord and plug are in good condition and the equipment is not damaged.  Only use th ...

  • ResMed Comfort - page 11

    Section 4: Fitting the mask 11 Fitting the mask The mask frame and headgear should be assembled as described in section 3, Preparing for use . To fit the mask to your face: 1 Loosen the headgear straps and pull the headgear over your head. Make sure the straps are sufficiently loose so that the nasal cushion is not pulled against your nose. 2 Posit ...

  • ResMed Comfort - page 12

    SULLIVAN Comfort Operating Manual  12 5 Adjust the air tubing – try both ways and select whichever is more comfortable Using the START button Start the flow generator by pressing the START button. The pressure ...

  • ResMed Comfort - page 13

    Section 5: Daily operation 13 * The pressure strip light display gives an approximate indication of the pressure you are receiving. You will notice that the indicator lights fluctuate as the unit switches between your IPAP and EPAP pressures. Using the delay timer If you prefer to have the air pressure build up gradually while you fall asleep, sele ...

  • ResMed Comfort - page 14

    SULLIVAN Comfort Operating Manual  14 S topping 1 Press the STOP button to stop the flow generator. To remove the mask, stretch the headgear slightly and pull it over your head. Or release one of the bottom headg ...

  • ResMed Comfort - page 15

    Section 6: Helpful hints 15 Help ful hint s • When you use your SULLIVAN ® Comfort™ bi-level system for the first time you may notice an unusual sense of pressure when breathing. Even though a reduced pressure is delivered when you breathe out, you may still need to consciously push out when breathing out. This is normal and will occur automat ...

  • ResMed Comfort - page 16

    SULLIVAN Comfort Operating Manual  16 Use in different countries • Your SULLIVAN ® Comfort™ has a switch mode power supply which means it will operate on the different power supplies in other countries. The ...

  • ResMed Comfort - page 17

    Section 8: Warnings related to treatment 17 W arnings related to treatment The SULLIVAN ® Comfort™ bi-level system is designed to help you have a good night’s sleep. However, you need to be aware of possible problems that may arise during the course of your treatment. If you have concerns about any of these problems consult your physician imme ...

  • ResMed Comfort - page 18

    SULLIVAN Comfort Operating Manual  18 9 Cleaning and maintenance You should regularly carry out the cleaning and maintenance described in this section. These procedures do not require a specialist service organiz ...

  • ResMed Comfort - page 19

    Section 9: Cleaning and maintenance 19 Remove air filter box at rear of flow generator Periodically All items of the mask and air tubing system are subject to normal wear and tear and may eventually need to be replaced. Check periodically for excessive wear in these items. 1 When dirty, remove the headgear and wash it in warm water using a mild det ...

  • ResMed Comfort - page 20

    SULLIVAN Comfort Operating Manual  20  Do not wash the filter strips as they will be damaged and will not function effectively.  Do not attempt to open the flow generator unit. Repairs and internal servicin ...

  • ResMed Comfort - page 21

    Section 10: Servicing 21 10 Servicing Your SULLIVAN ® Comfort™ is designed to give years of trouble-free operation. It is not necessary to send your unit away for regular servicing provided you follow the recommendations in section 9, Cleaning and maintenance . Should you feel that your unit is not performing properly, check section 11, Troubles ...

  • ResMed Comfort - page 22

    SULLIVAN Comfort Operating Manual  22 T roubleshooting If you have any problems with your SULLIVAN ® Comfort™, try the following suggestions. If it is still not working properly, consult your equipment supplie ...

  • ResMed Comfort - page 23

    Section 12: Accessories and spare parts 23 Accessories and sp are p art s Accessories A range of accessories is available to adapt your SULLIVAN ® Comfort™ to your needs. Contact your sleep clinic or equipment supplier for further information on the following items. Modular Mask Frame Kit Adapts to a wide range of face sizes and features a T-bar ...

  • ResMed Comfort - page 24

    SULLIVAN Comfort Operating Manual  24 SULLIVAN ® HCH Disk – a foam insert which fits inside your mask frame HC100 Heated Humidifier Passover Humidifier Humidifiers Bi-level treatment sometimes results in a dry ...

  • ResMed Comfort - page 25

    Section 13: System specifications 25 13 System specifications SULLIV A N ® Comfort bi-level system Flow generator unit PERFORMANCE Operating pressure range: 4-18 cm H 2 O Maximum IPAP time range: 0.2-10.8 seconds DIMENSIONS (H x W x D) 145 x 240 x 350 mm (5.7 x 9.5 x 13.8 in) WEIGHT 4.1 kg (9.1 lb) POWER SUPPLY Input range 110-240 V, 50-60 Hz, ...

  • ResMed Comfort - page 26

    21802/3 00 08 • ResMed Malaysia Sdn Bhd Unit 4.04, Wisma BICMA Jalan 243/51A 46100 Petaling Jaya Selangor Malaysia Tel: +60 3 7877 9099 Fax: +60 3 7877 6099 Email: reception@resmed.com.my • ResMed NZ Ltd PO Box 51-048 Pakuranga Auckland New Zealand Tel: +64 9530 9049 or +0800 737 633 (NZ toll free) Fax: +64 95309048 Email: reception@resmed.com. ...

Manufacturer ResMed Category Home Care Product

Documents that we receive from a manufacturer of a ResMed Comfort can be divided into several groups. They are, among others:
- ResMed technical drawings
- Comfort manuals
- ResMed product data sheets
- information booklets
- or energy labels ResMed Comfort
All of them are important, but the most important information from the point of view of use of the device are in the user manual ResMed Comfort.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals ResMed Comfort, service manual, brief instructions and user manuals ResMed Comfort. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product ResMed Comfort.

A complete manual for the device ResMed Comfort, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use ResMed Comfort by users. Manuals are usually written by a technical writer, but in a language understandable to all users of ResMed Comfort.

A complete ResMed manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual ResMed Comfort - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the ResMed Comfort, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the ResMed Comfort, that we can find in the current document
3. Tips how to use the basic functions of the device ResMed Comfort - which should help us in our first steps of using ResMed Comfort
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with ResMed Comfort
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of ResMed Comfort in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning ResMed Comfort?

Use the form below

If you did not solve your problem by using a manual ResMed Comfort, ask a question using the form below. If a user had a similar problem with ResMed Comfort it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (0)