    1 Rev 05 Operator's Manual Model HNC-63-INT Neuro v ascular Arr a y Coil f or the Philips Intera 1.5T MRI System Part Number 500082 ...

  • SMC Networks 500082 - page 2

    2 Rev 05 February 2002 ©2002 Inviv o Cor poration All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any language in an y form by an y means without the wr itten per mission of In vivo Cor poration. Licenses and T rademarks The Inviv o Logo is a registered ...

  • SMC Networks 500082 - page 3

    3 Rev 05 Introduction This manual describes the safety precautions, features, use and care, of the Invivo HNC-63-INT Neurovascular Array Coil, compatible with the Philips Intera 1.5T MRI System. Please review this manual thoroughly before using the device. If you have any questions or comments on this manual, or need any assistance with the use of ...

  • SMC Networks 500082 - page 4

    4 Rev 05 HNC Neuro v ascular Arr a y Coil Your HNC-63-INT Neurovascular Array Coil package consists of the following parts. Please inspect the package upon receipt to make sure all parts are present and in good order. Use this guide to refer to part names thoughout this manual. The HNC-63-INT consists of two housings to facilitate lifting and posit ...

  • SMC Networks 500082 - page 5

    5 Rev 05 T ab le of Contents HNC-63-INT Neurovascular Array Coil ........................................................ 4 Chapter 1 - Pati ent Safety ............................................................................ 6 Training ........................................................................................................ 6 Qua ...

  • SMC Networks 500082 - page 6

    6 Rev 05 Chapter 1 - P atient Saf ety Cautions The following general warning statements apply to scanning with a magnetic resonance system. For further details, review the warnings in your MR system Operators Manual. Indications Contra- indications Precautions This manual contains detailed information on the setup, positioning and use of the Invivo ...

  • SMC Networks 500082 - page 7

    7 Rev 05 If the patient complains of warming, tingling, stinging, or similar sensations, promptly stop the scan procedure, examine the patient, and contact the responsible physician before continuing the procedure. Pay special attention to very young, sedated, or other compromised patients who may not be able to communicate effectively. Patients wi ...

  • SMC Networks 500082 - page 8

    8 Rev 05 The coil should only be used with the accessories specified in the operator’s manual. The use of accessories other than those specified in the operator’s manual may result in decreased ESD immunity of the coil or MR system, causing damage to the coil and/or system. The equipment should not be used with other coils or equipment present ...

  • SMC Networks 500082 - page 9

    9 Rev 05 Chapter 2 - Using the HNC-63-INT Arr a y Coil P ositioning the HNC-63-INT Coil on the P atient T able The HNC-63-INT Neurovascular Array Coil is designed for imaging of the brain, cervical spine, anterior neck and vasculature in the head & neck, to the aortic arch. Polarity (i.e., coil orientation) must be maintained to produce accepta ...

  • SMC Networks 500082 - page 10

    10 Rev 05 Adjust the patient so their shoulders are snug against the curved arch of the posterior coil housing. Positioning the P atient . Firmly hold the anterior coil housing and carefully place it on the posterior coil housing. As you lower the anterior coil housing, center and seat the RF connector pins while the flat mating surfaces meet. Secu ...

  • SMC Networks 500082 - page 11

    11 Rev 05 Synergy MultiConnect (SMC) The Synergy MultiConnect (SMC) shown here is a required option to use the HNC-63-INT coil. IMPORTANT: You must have the SMC device and Level 8.1.1 or higher software on your Intera system. Additionally, Release 8.1.1 software requires a Synergy MultiConnect patch to enable this coil to be recognized by your syst ...

  • SMC Networks 500082 - page 12

    12 Rev 05 Connecting the Cab le Attach the Synergy MultiConnect (SMC) holder to the same side of the support as the patient ineterface control unit (PICU) facing the magnet. In this example, the PICU is on the left side. Slide the holder toward the bore, even with the patients knees. For systems with PICU on left side of magnet, transfer SMC connec ...

  • SMC Networks 500082 - page 13

    13 Rev 05 Attach SMC connector to PICU. Example above is left side PICU. Attach SMC connector to PICU. Example above is right side PICU. Right Side Patient Interface Control Unit (PICU) Left Side Patient Interface Control Unit (PICU) ...

  • SMC Networks 500082 - page 14

    14 Rev 05 Quality Assurance Chapter 3 - Quality Assurance Remove the coil top and place the phantom positioner (part number 102485) and the phantom (part number 102690) as shown in the figures below. Replace the coil top. Connect the coil to the system as detailed in Chapter 2, "Connecting the cable." Landmark on the the phantom through t ...

  • SMC Networks 500082 - page 15

    15 Rev 05 Geometry Coil Selection 3IB-MRIDevic -channel Combination 2345 -connection d Multi Coil n o Homogeneity correction none FOV (mm) 400.00 RFOV (%) 100.00 Foldover suppression n o Matrix scan 2 5 6 -reconstruction 2 5 6 Scan percentage (%) 50.00 Stacks 3 -current A -type parallel -slices 3 -slice thickness (mm) 10.00 -slice gap user defined ...

  • SMC Networks 500082 - page 16

    16 Rev 05 Example of SURVEY images are below; note position of slices for the SNR check. Proceed to the SNR check on page 16. ...

  • SMC Networks 500082 - page 17

    17 Rev 05 Run an SNR CHECK scan using the parameters listed belo w . TIP: Sa v e this SNR CHECK scan for future QA scans . Position slices for the SNR CHECK as shown on page 15. Geometry Coil Selection 3IB-MRIDevic -channel Combination 2345 -connection d Homogeneity correction none FOV (mm) 400.00 RFOV (%) 100.00 Foldover suppression n o Matrix sca ...

  • SMC Networks 500082 - page 18

    18 Rev 05 Reconstr uct SNR CHECK image f or individual channels using the instr uctions pro vided below . 1. Before running the SNR CHECK scan, be sure to have the "Save Raw Data" option set to "Yes" under [Postproc]. 2. Do not delete the current scan from the [Scanlist]. 3. Under the [Scanlist] icon, select the [Add Rec] icon, ...

  • SMC Networks 500082 - page 19

    19 Rev 05 SNR Calulations 1. Select the reconstructed image from Channel 2, and display an eliptical ROI with a total pixel area of 1300 mm squared. Refer to Philips Gyroscan Intera Instructions Manual for ROI help. 2. Position ROI in the center of the phantom bottle, as shown below. 3. Bring up the Statistics for ROI, right-mouse select over the R ...

  • SMC Networks 500082 - page 20

    20 Rev 05 SNR Calulations Repeat SNR calculations as described on page 18 for the remaining reconstructed images from channels 3,4,5. Example ...

  • SMC Networks 500082 - page 21

    21 Rev 05 SNR Calulations Repeat SNR calculations as described on page 18 for the remaining reconstructed images from channels 3,4,5. Example ...

  • SMC Networks 500082 - page 22

    22 Rev 05 SNR Calulations Repeat SNR calculations as described on page 18 for the remaining reconstructed images from channels 3,4,5. Example ...

  • SMC Networks 500082 - page 23

    23 Rev 05 Chapter 4 - Scanning Set Up The neurovascular array coil contains both a head coil and a neck coil. The head coil (channel 2), neck coil (channels 3,4,5) or both head and neck coils (channels 2,3,4,5) can be active at any given time. To select the active coil(s), pick the proper channel(s) from the channel combination (ch. combination) me ...

  • SMC Networks 500082 - page 24

    24 Rev 05 Scanning Step by Step Scan Select: Software Release 8.1.1, Choose "3IB-MRIDEVIC", Software Releases above 8.1.1, choose "SMC-MRI". Select Channel combination(s): HEAD only: Select Ch. combination 2 NECK only: Select Ch. combination 3,4,5 HEAD & NECK: Select Ch. combination 2,3,4,5 Plan Slices Confirm Planning Geome ...

  • SMC Networks 500082 - page 25

    25 Rev 05 Chapter 5 - Scan Protocols Invivo Corporation recomends that you select imaging protocols that have been established by your radiologists. Additional protocols can be found within your Philips Application Guides. ...

  • SMC Networks 500082 - page 26

    26 Rev 05 Chapter 6 - Maintenance Cleaning The Neurov ascular Arra y coil and patient comfort pads may be cleaned b y wiping with a cloth dampened with a solution of 30% isopropyl alcohol and 70% tap w ater . Do not pour any cleaning solution directly on the coil! Let coil housing and pads dry before use . Storage The coil should be stored in an ai ...

  • SMC Networks 500082 - page 27

    27 Rev 05 ...

Manufacturer SMC Networks Category Scanner

Documents that we receive from a manufacturer of a SMC Networks 500082 can be divided into several groups. They are, among others:
- SMC Networks technical drawings
- 500082 manuals
- SMC Networks product data sheets
- information booklets
- or energy labels SMC Networks 500082
All of them are important, but the most important information from the point of view of use of the device are in the user manual SMC Networks 500082.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals SMC Networks 500082, service manual, brief instructions and user manuals SMC Networks 500082. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product SMC Networks 500082.

A complete manual for the device SMC Networks 500082, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use SMC Networks 500082 by users. Manuals are usually written by a technical writer, but in a language understandable to all users of SMC Networks 500082.

A complete SMC Networks manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual SMC Networks 500082 - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the SMC Networks 500082, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the SMC Networks 500082, that we can find in the current document
3. Tips how to use the basic functions of the device SMC Networks 500082 - which should help us in our first steps of using SMC Networks 500082
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with SMC Networks 500082
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of SMC Networks 500082 in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning SMC Networks 500082?

Use the form below

If you did not solve your problem by using a manual SMC Networks 500082, ask a question using the form below. If a user had a similar problem with SMC Networks 500082 it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (0)