    Computer-Based Instruments NI 5911 User Manual High-Speed Digitizer with FLEX ADC NI 5911 User Manual ™ June 2001 Edition Part Number 322150D-01 ...

  • National Instruments NI 5911 - page 2

    Support Worldwide Technical Support and Product Information ni.com National Instruments Corporate Headquarters 11500 North Mopac Expressway Austin, Texas 78759-3504 USA Tel: 512 794 0100 Worldwide Offices Australia 03 9879 5166, Austria 0662 45 79 90 0, Belgium 02 757 00 20, Brazil 011 284 5011, Canada (Calgary) 403 274 9391, Canada (Montreal) 514 ...

  • National Instruments NI 5911 - page 3

    Important Information Warranty The NI 5911 is warranted ag ainst defects i n materials an d workmanshi p for a peri od of one yea r from the da te of shipment , as eviden ced by receipts or other documenta tion. National Instruments will, at it s option, repair or replace equipment that proves to be defective during the warranty period. This warran ...

  • National Instruments NI 5911 - page 4

    Compliance FCC/Canada Radio Frequency Interference Compliance* Determining FCC Class The Federal Communications Commission (FCC) has rules to protect wireless communications from interference. The FCC places digital electronics into two classes. These classes are known as Class A (for use in industrial-commercial loc ations only) or Class B (for us ...

  • National Instruments NI 5911 - page 5

    Canadian Depart ment of Communications This Class B digital apparatus meets all requirements of the Canadian I nterference-Causing Equipment Regulations. Cet appareil num é rique de la classe B respecte toutes les exigences du R è gleme nt sur le mat é riel brouilleur du Canada. Compliance to EU Directives Readers in the European Union (EU) must ...

  • National Instruments NI 5911 - page 6

    Conventions The follo wing con v entions are used in this manual: » The » symbol leads you through nested menu items and dialog box options to a final action. The sequence File»Page Setup» Options directs you to pull down the File menu, selec t the Page Setup item, and select Options from the last dialog box. This icon denotes a note, which ale ...

  • National Instruments NI 5911 - page 7

    © National Instruments Corporatio n vii NI 5911 User Manual Contents Chapter 1 Taking Measurements with the NI 5911 Installing the NI 5911 ........... .............. ......................... ............... ......................... .......... 1-1 Connecting Signals ......... .......................... .............. ......................... ... ...

  • National Instruments NI 5911 - page 8

    Contents NI 5911 User Manu al viii ni.com RTSI Bus and Clock PFI .... .............. .......................... ........... .............. .......................... 2-11 PFI Lines ............ .............. ......................... ......................... ............ .............. ... 2-11 PFI Lines as Inputs .......... ........... ......... ...

  • National Instruments NI 5911 - page 9

    © National Instruments Corporatio n 1-1 NI 5911 User Manual 1 T aking Measurements with the NI 5911 Thank you for buyin g a National Instruments (NI) 5911 digitizer, featuring the FLEX AD C. This chapter provid es informatio n on installing, connecting signals to, a nd acquiring data from your N I 5911. Installing the NI 5911 There are two main st ...

  • National Instruments NI 5911 - page 10

    Chapter 1 T aking Measurements w ith the NI 5911 NI 5911 User Manual 1-2 ni.com Figure 1-1. NI 5911 Connectors Figure 1-2. 9-Pin Mini Cir cular DIN Connector 1 +5 Volts (Fused) 2G N D 3 Reserved 4R e s e r v e d 5R e s e r v e d 6P F I 2 7 Reserved 8 Reserved 9 Reserved CH0 PFI1 PFI2 (DIN) 4 3 5 6 7 8 9 1 2 ...

  • National Instruments NI 5911 - page 11

    Chapter 1 T ak ing Measuremen ts with the NI 5911 © National Instruments Corporatio n 1-3 NI 5911 User Manual Acquiring Data with Y our NI 5911 You can acquire data either programmatically — by writing an app lication f o ry o u rN I5 9 1 1 — or interactively with the Scope Soft Front Panel. Programmatically Controlling Y our NI 5911 To help y ...

  • National Instruments NI 5911 - page 12

    © National Instruments Corporatio n 2-1 NI 5911 User Manual 2 Hardware Overview This chapter includes an overview of the NI 5 911, explains the operation of each functional unit making up your NI 5911, and describes the signal connections. Figure 2-1 shows a block diagram of the NI 5911. Figure 2-1. NI 5911 Block D iagram Differential Programmable ...

  • National Instruments NI 5911 - page 13

    Chapter 2 Hardware Over view NI 5911 User Manual 2-2 ni.com Differential Input When measuring high dynamic range signals, ground n oise is often a problem. The PGIA of the NI 5911 allows you to make no ise-free signal measurements. The PGIA differential amplifier efficiently rejects any noise present on the ground sign al. Internal to the PGIA, the ...

  • National Instruments NI 5911 - page 14

    Chapter 2 Hardware Over view © National Instruments Corporatio n 2-3 NI 5911 User Manual Input Ranges To optimize the ADC resolution, you can select different gains for the PGIA. In this way, you can scale y our input signal to match the full inpu t range of the converter. The NI 5 911 PGIA offers seven different input ranges, fro m ±0.1 V to ±1 ...

  • National Instruments NI 5911 - page 15

    Chapter 2 Hardware Over view NI 5911 User Manual 2-4 ni.com if the de vice has 1 M Ω output impedance, your measured signal will be one-half the actual signal value. Input Bias The inputs of the PGIA typically draw an inpu t bias current of 1 nA at 25 ° C. Attaching a device with a very high source impedance can cause an offset voltage to be add ...

  • National Instruments NI 5911 - page 16

    Chapter 2 Hardware Over view © National Instruments Corporatio n 2-5 NI 5911 User Manual analysis, distortion analysis, and other measurements for which high resolution is crucial. Oscilloscope Mode The ADC converts at a constant rate of 10 0 MS/s, but you can choose to store only a fraction of these samples into m emory at a lower rate. This allo ...

  • National Instruments NI 5911 - page 17

    Chapter 2 Hardware Over view NI 5911 User Manual 2-6 ni.com Like an y o ther type of con verter that uses noise shaping to enhance resolution, the frequency response of the con verter is only flat to its maximum useful bandwidth. The NI 5911 h as a bandwidth of 4 MHz. Beyond this frequenc y , there is a span where the con verter acts resonant and w ...

  • National Instruments NI 5911 - page 18

    Chapter 2 Hardware Over view © National Instruments Corporatio n 2-7 NI 5911 User Manual Calibration The NI 5911 can be calibrated for very high accuracy and resolution due to an advanced calibration scheme. There are two different types of calibration: internal, or self, calibration and external calibration. A third option, internal restore, rest ...

  • National Instruments NI 5911 - page 19

    Chapter 2 Hardware Over view NI 5911 User Manual 2-8 ni.com • The linearity of the ADC is calibrat ed using an internal sinewa ve generator as reference. • The time-to-digital con verter used for RIS measurements is calibrated. Caution Do not apply high-amplit ude or high-frequency signals to the NI 5911 during internal calibration. For optimal ...

  • National Instruments NI 5911 - page 20

    Chapter 2 Hardware Over view © National Instruments Corporatio n 2-9 NI 5911 User Manual Figure 2-3. T rigger Sour ces Analog T rigger Circuit The analog trigger on the NI 5911 operates by c omparing the current analog input to an onb oard threshold voltage. Thi s threshold voltage is the trigger value, and can be set with in the current input ran ...

  • National Instruments NI 5911 - page 21

    Chapter 2 Hardware Over view NI 5911 User Manu al 2-1 0 ni.com T rigger Hold-Off Trigger hold-off is the minimum length of time (in seconds) from an accepted trigger to the start of the next record. In other words, when a trigger is accepted, the trigger counter is loaded with the desired hold-off time. After completing its current record, the digi ...

  • National Instruments NI 5911 - page 22

    Chapter 2 Hardware Over view © National Instruments Corporation 2-11 NI 5 911 User Manual Multiple-Record Acquisitions After the trigger has been received and the posttrigger samples have been stored, the NI 5911 can be configured to begin another acqui sition that is stored in another onboard memory record. This is a multiple-record acquisition . ...

  • National Instruments NI 5911 - page 23

    Chapter 2 Hardware Over view NI 5911 User Manu al 2-1 2 ni.com PFI Lines as Inputs You can select PFI1 or PFI2 as inputs for a trigger o r a reference clock. Please see the Synchronization section belo w for more information about the use of reference clocks in the NI 5911. PFI Lines as Outputs You can select PFI1 or PFI2 to output seve ral digital ...

  • National Instruments NI 5911 - page 24

    © National Instruments Corporatio n A- 1 NI 5911 User Manual A Specifications This appendix lists the sp ecifications of the NI 5911. These specifications are typical at 25 ° C unless otherwise stated. Acquisition System Bandwidth .............. ......................... ....... 100 MHz maximum, see T able 2-2, Av a il a ble S a mpl i ng Rates an ...

  • National Instruments NI 5911 - page 25

    Appendix A Specifications NI 5911 User Manual A-2 ni.com Sample onboard memory ........................ 4 MB or 16 MB Memory sample depth Vertical sensitivity (input ranges) 20 kS/s Flexible Resolution 20.5 Bi ts 10 kS/s Flexible Resolution 21 Bits *1 < n < 2 32 in oscilloscope mode Sampling Frequency Mode Sample Depth (4 MB) Sample Depth (16 ...

  • National Instruments NI 5911 - page 26

    Appendix A Specifications © National Instruments Corporatio n A- 3 NI 5911 User Manual Acquisition Characteristics Accuracy DC gain accuracy ................................... ±0.05% signal ±0.0001% fs for all input ranges at 1 MS/ s in flexible resolu tion mode DC offset accuracy ................................. ±0.1 mV ±0.01% f s for all i ...

  • National Instruments NI 5911 - page 27

    Appendix A Specifications NI 5911 User Manual A-4 ni.com Filtering Dynamic Range Noise (excluding input-referred noise) Sampling Frequency Filter Mode Bandwidth Ripple Alias Attenuation 100/ n * MS/s Oscilloscope 100 MHz ±3 dB N/A 12.5 MS/s Flex ible Resolution 3.75 MHz ±0.2 dB – 60 dB 5M S / s Flexible Resolution 2M H z ±0.1 dB – 70 dB 2.5 ...

  • National Instruments NI 5911 - page 28

    Appendix A Specifications © National Instruments Corporatio n A- 5 NI 5911 User Manual Distortion T imebase System Reference clock ....................... ........... .... 10 MHz Clock accuracy (as master) .................... 10 MHz ±50 ppm 1M S / s 400 kHz – 160 dBfs/ – 104 dBfs 500 kS/s 200 kHz – 160 dBfs/ – 107 dBfs 200 kS/s 80 kHz ? ...

  • National Instruments NI 5911 - page 29

    Appendix A Specifications NI 5911 User Manual A-6 ni.com Clock input tolerance (as slave)..............10 MHz ±100 ppm Clock jitter ........... ............... ....................<75 pS rms, independent of reference clock so urce Clock compatibility ................... ............TTL for both input and output Interpolator resoluti on (repetit ...

  • National Instruments NI 5911 - page 30

    Appendix A Specifications © National Instruments Corporatio n A- 7 NI 5911 User Manual Sensitivity ............................................... 170 steps in full-scale voltage range TRIG input range ......... .............. ............ ±5 V (without external attenuation) TRIG input impedance .................... ....... 1 M Ω ± 1% in paral ...

  • National Instruments NI 5911 - page 31

    Appendix A Specifications NI 5911 User Manual A-8 ni.com Operating Environment Note Multiple NI 5911s in the same co mputer may raise operating temperat ures beyond specification and gi ve rise to imprecise data. NI strong ly recommends leaving an empty PCI slot between multi ple NI 5911s or adding a fan. Ambient temperature .................. .... ...

  • National Instruments NI 5911 - page 32

    © National Instruments Corporatio n B- 1 NI 5911 User Manual B T echnical Support Resources W eb Support National Instruments W eb support is your first stop for help in solving installation, configuration, and application pro blems and questions. O nline problem-solving and di agnostic resources include frequ ently asked questions, knowledge base ...

  • National Instruments NI 5911 - page 33

    Appendix B T echnical Support R esources NI 5911 User Manual B-2 ni.com W orldwide Support National Instruments has of fices located around the world to help address your support needs. Y ou can access our branch of fice W eb sites from the W orldwide Of fices section of ni.com . Branch of fice W eb sites pro vide up-to-date contact information, su ...

  • National Instruments NI 5911 - page 34

    © National Instruments Corporatio n G-1 NI 59 11 User Manual Glossar y Pref ix Meanings V alue p- pico- 10 – 12 n- nano- 10 – 9 µ- micro- 10 – 6 m- milli- 10 – 3 k- kilo- 10 3 M- mega- 10 6 G- giga- 10 9 Symbols % percent + positiv e of, or pl us – negati ve of, or minus /p e r ° degree ± plus or minus Ω ohm A A amperes A/D analog t ...

  • National Instruments NI 5911 - page 35

    Glossary NI 5911 User Manual G-2 ni.com A C coupled the passing of a signal through a filter network that removes the DC component of the si gnal ADC analog-to-digital con verter — an electronic de vice, often an integrated circuit, that con verts an analog voltage to a digital number ADC resolution the resolution of the ADC , which is measured i ...

  • National Instruments NI 5911 - page 36

    Glossary © National Instruments Corporatio n G-3 NI 5911 User Manual clock hardware component that controls timing for reading from or writing to groups CMRR common-mode rejection ratio — a measure of an instrument ’ s ability to reject interference from a common-mode signal, usually e xpressed in decibels (dB) counter/timer a circuit that cou ...

  • National Instruments NI 5911 - page 37

    Glossary NI 5911 User Manual G-4 ni.com F filtering a type of signal conditioning that al lo w s you to filter unwanted signals from the signal you are trying to measure fs full-scale — total voltage in the input range. A ±10 V input range is 20 V fs G gain the factor by which a signal is amplif ied, sometimes expressed in decibels H hardware th ...

  • National Instruments NI 5911 - page 38

    Glossary © National Instruments Corporatio n G-5 NI 5911 User Manual interrupt lev e l the relativ e prior ity at which a device can interrupt ISA industry standard architecture K kk i l o — the standard metric prefix for 1,000, or 10 3 , used w ith units of measure such as volts, hertz, and meters kS 1,000 samples L LabVIEW laboratory virtual i ...

  • National Instruments NI 5911 - page 39

    Glossary NI 5911 User Manual G-6 ni.com N noise an undesirable electrical signal — noise comes from e xternal sources such as the A C power line, motors, generators, transformers, fluoresc ent lights, soldering irons, CR T displays, computers, electrical storms, welders, radio transmitters, and internal sources such as semi conductors, resistors, ...

  • National Instruments NI 5911 - page 40

    Glossary © National Instruments Corporatio n G-7 NI 5911 User Manual pretriggering the technique used on a de vice to keep a b uffer f illed with data, so that when the trigger conditions are met, the sample includes the data leading u p to the trigger con dition PXI PCI eXtensions for Instrumentation. PXI is an open specification that b uilds of ...

  • National Instruments NI 5911 - page 41

    Glossary NI 5911 User Manual G-8 ni.com sense in four-wire resistance the sense measures the v oltag e across the resistor being excited by the e xcitati on current settling time the amount of time required for a voltage to reach its f inal value within specified limits source impedance a parameter of signal sources that reflects curren t-driving a ...

  • National Instruments NI 5911 - page 42

    Glossary © National Instruments Corporatio n G-9 NI 5911 User Manual V V volts V AC volts alternating current VDC volts direct current V error v oltage error VI virtual instrument — (1) a combination of hardware and/or software elements, typically used with a PC, that has the functionality of a classic stand-alone instrument (2) a LabVIEW softwa ...

  • National Instruments NI 5911 - page 43

    © National Instruments Corporatio n I-1 NI 5911 User Manual Index A AC coupling, 2-4 accuracy characteristics, A-3 acquisition multiple record, 2-11 Scope Soft Front Panel, 1-3 acquisition characteristics specifications accuracy, A-3 common-mode characteristics, A-3 distortion, A-5 dynamic range, A-4 filtering, A-4 acquisition modes specifications ...

  • National Instruments NI 5911 - page 44

    Index NI 5911 User Manual I-2 ni.com G grounding considerations, 2-2 H hardware overview See also specifications acquisition syste m PFI lines, 2-11 triggering and arming , 2-8 block diagram of NI 591 1, 2-1 calibration, 2-7 differential programmable gain i nput amplifier (PGIA) AC coupling, 2-4 differential input, 2-2 grounding considerations, 2-2 ...

  • National Instruments NI 5911 - page 45

    Index © National Instruments Corporatio n I-3 NI 5911 User Manual O operating environment specifications, A-8 oscilloscope mode definition, 2-4 purpose and use, 2-5 Real-Time and RIS sampling methods, 2-5 output impedance, 2-3 P PFI lines as inputs, 2-12 as outputs, 2-12 overview, 2-11 PGIA. See differential programmable gain input amplifier ( PGI ...

Manufacturer National Instruments Category Graphics Tablet

Documents that we receive from a manufacturer of a National Instruments NI 5911 can be divided into several groups. They are, among others:
- National Instruments technical drawings
- NI 5911 manuals
- National Instruments product data sheets
- information booklets
- or energy labels National Instruments NI 5911
All of them are important, but the most important information from the point of view of use of the device are in the user manual National Instruments NI 5911.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals National Instruments NI 5911, service manual, brief instructions and user manuals National Instruments NI 5911. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product National Instruments NI 5911.

A complete manual for the device National Instruments NI 5911, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use National Instruments NI 5911 by users. Manuals are usually written by a technical writer, but in a language understandable to all users of National Instruments NI 5911.

A complete National Instruments manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual National Instruments NI 5911 - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the National Instruments NI 5911, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the National Instruments NI 5911, that we can find in the current document
3. Tips how to use the basic functions of the device National Instruments NI 5911 - which should help us in our first steps of using National Instruments NI 5911
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with National Instruments NI 5911
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of National Instruments NI 5911 in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning National Instruments NI 5911?

Use the form below

If you did not solve your problem by using a manual National Instruments NI 5911, ask a question using the form below. If a user had a similar problem with National Instruments NI 5911 it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (0)