  • IBM 6300 - page 1

    MODEL 6300 VISIBLE RANGE SPECTROPHOTOMETER OPERATING MANUAL 630 026 ...

  • IBM 6300 - page 2

    6300/REV B/12-96 SAFETY Please read this information carefully prior to installing or using this equipment. 1. The unit described in this manual is designed to be operated only by trained personnel. Any adjustments, maintenance and repair must be carried out as defined in this manual, by a person qualified to be aware of the hazards involved. 2. It ...

  • IBM 6300 - page 3

    6300/REV B/12-96 6300 SPECTROPHOTOMETER OPERATING MANUAL CONTENTS SECTION 1 INTRODUCTION Instrument Description 1.1 Instrument Specification 1.2 SECTION 2 INSTALLATION Unpacking 2.1 Installation 2.2 Displays 2.3 Controls 2.4 Outputs 2.5 SECTION 3 OPERATION Initial Set-Up 3.1 Sample Measurement 3.2 Good Practice Guidelines 3.3 SECTION 4 MAINTENANCE ...

  • IBM 6300 - page 4

    SECTION 1 INTRODUCTION 1.1 INSTRUMENT DESCRIPTION The Model 6300 is a microprocessor controlled Visible Range Spectrophotometer covering the wave- length range of 320 to 950nm with a 10nm bandwidth. The monochromator is of a modified Czerny Turner design, incorporating a stepper motor driven 1200 lines/mm holographic diffraction grating and featuri ...

  • IBM 6300 - page 5

    SECTION 2 INSTALLATION 2.1 UNPACKING Remove the Model 6300 from the packaging and ensure the following items are present: 1. Model 6300 Spectrophotometer 2. Mains Cable 3. Pack 100 disposable Cuvettes (060 084) 4. Optional Accessories (as ordered) Any shortages or damage should be reported immediately to the Manufacturer or your local Distributor. ...

  • IBM 6300 - page 6

    2.3 DISPLAYS 1. Primary display area - Transmission, Absorbance, Concentration 2. Primary display adjust annunciator 3. Secondary display area - Wavelength, Factor 4. Primary display units 5. Secondary display adjust annunciator 6. Operation with PC 7. Menu options - %T ABS CONC FACTOR UNITS 8. Menu pointers (for 7) 3 6300/REV B/12-96 ...

  • IBM 6300 - page 7

    2.4 CONTROLS 1. used to adjust values on the selected display 2. used to move horizontally between menu options 3. used to select the displayed menu option 4. initiates a calibration routine 5. Print key. Provides a printout of the current reading with an incremental sample number. When pressed for the first time after a calibration the print out w ...

  • IBM 6300 - page 8

    2.5 INPUTS/OUTPUTS Fig. 2.5.1 Rear Panel Layout 1. LAMP ACCESS PANEL This panel allows the user to gain access to the tungsten halogen lamp when replacement is necessary. NOTE: The Lamp Access Panel and all ventilation slots must not be covered or obstructed at any time. 2. ROCKER SWITCH On/Off switch for the unit. 3. POWER IN SOCKET IEC type conne ...

  • IBM 6300 - page 9

    SECTION 3 OPERATION 3.1 INITIAL SET-UP NOTE: If the unit is in continuous use it is recommended that it should be left switched on to obtain maximum lamp life. Do not open the sample chamber door during initialisation. Connect the unit to the correct mains supply and switch on. After switch on, the Model 6300 automatically re-aligns the monochromat ...

  • IBM 6300 - page 10

    7 The complimentary colour method of selecting wavelength may not be applicable in all situations, perhaps because the solution has no distinct colour, or the solution is of a complex nature and the absorbing species of interest is not the predominant colour. In these circumstances it will be necessary to scan across the spectrum to determine point ...

  • IBM 6300 - page 11

    8 Concentration Mode This mode is selected by moving the cursor to the CONC menu option using the LEFT or RIGHT arrow keys. The primary display will show the concentration, with ppm, mgl -1 , gl -1 , M, % or blank units. The wavelength can be adjusted using the UP and DOWN arrow keys. Pressing the CAL key places the 6300 in a concentration cal mode ...

  • IBM 6300 - page 12

    3.3 GOOD PRACTICE GUIDELINES 1. For optimum performance a calibration routine should be carried out at the beginning and end of every sample batch. 2. To ensure accurate results are obtained the sample area lid should be kept in the closed position during measurement. 3. The styrene cuvettes supplied with the unit are disposable (i.e; ideally they ...

  • IBM 6300 - page 13

    SECTION 4 MAINTENANCE 4.1 GENERAL The 6300 has been designed to give optimum performance with minimum maintenance. It is only necessary to keep the external surfaces clean and free from dust. The sample area should always be kept clean and any accidental spillage should be wiped away immediately. To give added protection when not in use, the unit s ...

  • IBM 6300 - page 14

    Fig. 4.2.1 Lamp Fitting 6. Close the lamp access panel and re-tighten the thumbscrew. NOTE: It is essential that only the specified replacement lamp should be used. Accuracy of optical alignment and performance cannot be guaranteed using alternative manufactured lamps. SECTION 5 OPTIONAL ACCESSORIES 5.1 OPTIONAL ACCESSORIES The following list of it ...

  • IBM 6300 - page 15

    SECTION 6 INTERFACING Serial Interface The Model 6300 has a bi-directional RS232 interface set to: 1200 baud 7 data bits odd parity 1 stop bit The 25 way D connector allows a standard one-to-one interconnection lead to be used, as supplied with the 40 column printer. A printout is initiated by pressing the PRINT key. If the sample munber is unity, ...

  • IBM 6300 - page 16

    ASCII Fxxxx.x<CR> Sets the concentration factor to xxxx.x. For example: F1000<CR> will set the factor to 1000. Note <CR> is an ASCII carriage return character. The last three commands provide an output which can readily be incorporated into most spreadsheet software packages. 6.2 RS232 Output The bi-directional RS232 interface is ...

  • IBM 6300 - page 17

    EC Declaration of Conformity Jenway Model 6300 Spectrophotometer complies with the following European Standards: EN 50081-1:1992 Electromagnetic compatibility - Generic emission standard EN 50082-1:1992 Electromagnetic compatibility - Generic immunity standard (Performance criterion B) EN 61010-1:1993 Safety requirements for electrical equipment fo ...

Manufacturer IBM Category Power Supply

Documents that we receive from a manufacturer of a IBM 6300 can be divided into several groups. They are, among others:
- IBM technical drawings
- 6300 manuals
- IBM product data sheets
- information booklets
- or energy labels IBM 6300
All of them are important, but the most important information from the point of view of use of the device are in the user manual IBM 6300.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals IBM 6300, service manual, brief instructions and user manuals IBM 6300. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product IBM 6300.

Similar manuals

A complete manual for the device IBM 6300, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use IBM 6300 by users. Manuals are usually written by a technical writer, but in a language understandable to all users of IBM 6300.

A complete IBM manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual IBM 6300 - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the IBM 6300, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the IBM 6300, that we can find in the current document
3. Tips how to use the basic functions of the device IBM 6300 - which should help us in our first steps of using IBM 6300
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with IBM 6300
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of IBM 6300 in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning IBM 6300?

Use the form below

If you did not solve your problem by using a manual IBM 6300, ask a question using the form below. If a user had a similar problem with IBM 6300 it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (0)