    VT1433B 8-Channel 196 kSamples/s Digitizer plus DSP User’s Guide Part Number 82-0068-000 Printed in U.S.A Print Date: July 30, 2004 © VXI Technology, Inc., 2004. All rights reserved. 2031 Main Street, Irvine, CA 92614-6509 U.S.A. bus ...

  • VXI VT1433B - page 2

    ii NOTICE The information contained in this document is subject to change without notice. VXI TECHNOLOGY MAKES NO WARRANTY OF ANY KIND WITH REGARD TO THIS MANUAL, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. VXI Technology shall not be liable for errors contained herein or direct, in ...

  • VXI VT1433B - page 3

    In This Book The VT1433B 8-Channel 196 kSamples/s Digitizer plus DSP is a C-size VXI module. “196 kSamples/s” refers to the maximum sample rate of 196608 samples per second. The VT1433B may contain either one or two 4-channel input assemblies so that the module may have a total of up to eight inputs. The module plugs into a single, C-size slot ...

  • VXI VT1433B - page 4

    VT1433B User's Guide iv ...

  • VXI VT1433B - page 5

    Table of Contents In This Book ..................................................... i i i Support Resources ................................................. i x Chapter 1. Installing the VT1433B Installing the VT1433B ............................................ 1 - 2 To inspect the VT1433B ........................................... 1 - 2 To ins ...

  • VXI VT1433B - page 6

    Synchronization in Multiple-mainframe Measurements .............. 3-14 Module Features ................................................ 3-15 Data Flow Diagram and FIFO Architecture ....................... 3-15 Base Sample Rates ........................................... 3-17 Measurement Process ............................................ 3-25 Mea ...

  • VXI VT1433B - page 7

    VT1433B Front Panel Description ................................... 5 - 5 Front panels for four or eight channels ............................ 5 - 5 Standard Front Panel .......................................... 5 - 6 VXI Backplane Connections ........................................ 5 - 8 Power Supplies and Ground ................................ ...

  • VXI VT1433B - page 8

    Chapter 10. Replacing Assemblies Replaceable Parts ............................................... 10-2 Ordering Information ......................................... 10-2 Direct Mail Order System ..................................... 10-2 CAGE Code Numbers ........................................ 10-3 Assemblies: without VT1433B-AYF or VT1433B-1D4 ...

  • VXI VT1433B - page 9

    Support Resources Support resources for this product are available on the Internet and at VXI Technology customer support centers. VXI Technology World Headquarters VXI Technology, Inc. 2031 Main Street Irvine, CA 92614-6509 Phone: (949) 955-1894 Fax: (949) 955-3041 VXI Technology Cleveland Instrument Division VXI Technology, Inc. 7525 Granger Road ...

  • VXI VT1433B - page 10

    x ...

  • VXI VT1433B - page 11

    1 Installing the VT1433B VT1433B User's Guide Installing the VT1433B 1-1 ...

  • VXI VT1433B - page 12

    Installing the VT1433B This chapter contains instructions for installing the VT1433B 8-Channel 196 kSamples/s Digitizer plus DSP Module and its drivers. This chapter also includes instructions for transporting and storing the module. To inspect the VT1433B The VT1433B 8-Channel 196 kSamples/s Digitizer plus DSP Module was carefully inspected both m ...

  • VXI VT1433B - page 13

    To install the VT1433B Caution To protect circuits from static discharge, observe anti-static techniques whenever handling the VT1433B 8-Channel 196 kSamples/s Digitizer plus DSP Module. 1 Set up the VXI mainframe. See the mainframe's installation guide for assistance. 2 Select a slot in the VXI mainframe for the VT1433B module. The VT1433B mo ...

  • VXI VT1433B - page 14

    4 Check the settings of the Boot Source and ROM Programming switches on the bottom of the module. Set switches 1 and 3 (BS1 and BS3) up and all the other switches down. VT1433B User's Guide Installing the VT1433B 1-4 Logical Address Boot Source & Programming 0 1 1 0 ...

  • VXI VT1433B - page 15

    5 Set the mainframe’s power switch to standby ( O I ). Caution Installing or removing the module with power on may damage components in the module. 6 Place the module’s card edges (top and bottom) into the module guides in the slot. 7 Slide the module into the mainframe until the module connects firmly with the backplane connectors. Make sure t ...

  • VXI VT1433B - page 16

    Install the host interface libraries After the hardware has been assembled, the next step in installing the VT1433B is to install the host interface libraries. Refer to the chapter titled “Getting Started With the VT1433B” to continue the installation process. To store the module Store the module in a clean, dry and static free environment. For ...

  • VXI VT1433B - page 17

    To transport the module • Package the module using the original factory packaging or packaging identical to the factory packaging. Containers and materials identical to those used in factory packaging are available through VXI Technology. • If returning the module to VXI Technology for service, attach a tag describing the following: q Type of s ...

  • VXI VT1433B - page 18

    VT1433B User's Guide Installing the VT1433B 1-8 ...

  • VXI VT1433B - page 19

    2 Getting Started With the VT1433B VT1433B User's Guide Getting Started With the VT1433B 2-1 ...

  • VXI VT1433B - page 20

    Introduction This chapter provides assistance in getting the VT1433B running and making simple measurements. It shows how to install the software libraries and how to run some of the example programs that are included. For more information see the other chapters in this book and the on-line function reference. (See “Where to get more information? ...

  • VXI VT1433B - page 21

    To install the VXI plug&play libraries System Requirements (Microsoft Windows) • An IBM compatible personal computer with either Microsoft Windows 95 or later or Microsoft Windows NT. (With either Windows OS, use the VXI plug&play library) • Additional hardware and software to connect the IBM compatible computer to a VXI mainframe. • ...

  • VXI VT1433B - page 22

    Getting Updates (Windows) The latest version of the VT1433B instrument drivers can be found on-line at www.vxitech.com. Getting Updates (HP-UX) For the latest HP-UX instrument drivers, please contact VXI Technology Customer Support Services. Contact information can be found in the Support section of the manual on page ix. VT1433B User's Guide ...

  • VXI VT1433B - page 23

    To install the Windows VXI plug&play drivers for the VT1433B (for Windows 95 and Windows NT) 1 Insert the VXI plug&play Drivers and Product Manuals CD into the CD-ROM drive. 2 Run the program: d:driversDAQ Driversdriver_ vxipnp_e1432_a_06_13.exe. (If the disc is in a driver other than "drive d:," replace "d:" with th ...

  • VXI VT1433B - page 24

    To install the HP-UX VXI plug&play drivers for the VT1433B (for HP-UX systems) 1 Log in as root. 2 Insert the VT1432A CD-ROM into the CD-ROM drive or obtain the latest VT1432A distribution. 3 Type swinstall. See the HP-UX Reference manual for information on the swinstall command. The VT1432A distribution is normally installed in the /opt/ vxipn ...

  • VXI VT1433B - page 25

    The VXI plug&play Soft Front Panel (SFP) Using the soft front panel If running the VT1432A/33B software in Microsoft Windows 95 or Windows NT, the Soft Front Panel (SFP) program can be used to interface with the VT1433B. The Soft Front Panel can be useful for checking the system to make sure that it is installed correctly and that all of its pa ...

  • VXI VT1433B - page 26

    The buttons on the right side of the SFP display are defined as follows: Meas This button opens the Measurement Control dialog box. It can set: q Measurement single/repeat q Mode block/continuous q Trigger auto/manual/input/source q Frequency span q Blocksize Input This button opens a dialog box where the VT1433B’s inputs can be setup. It can set ...

  • VXI VT1433B - page 27

    Display This button opens a dialog box which specifies how the data will be displayed. For each trace, an input channel (or OFF) and an output file can be specified. VXI This button opens a dialog box showing the modules installed in the VXI mainframe and indicating which are active and inactive. The “resource name” for each module is the inter ...

  • VXI VT1433B - page 28

    VEE example programs scope.vee This program displays four channels with time record and FFT for each channel. For the examples provided below, Agilent © VEE is used. The scope.vee program is located at VXIPNPHpe1432examplesvee on a Microsoft Windows system or at /usr/e1432/ vee-examples on an HP-UX system. To run scope.vee, first type: veetest ...

  • VXI VT1433B - page 29

    The program scope.vee starts, showing four channels, with time record and FFT for each channel. To start a measurement, click the Run button on the toolbar (triangle symbol). To pause, click on the Pause button (two vertical bars, next to the Run button). To stop the measurement, click the Stop button (square symbol). This screen is VEE’s panel v ...

  • VXI VT1433B - page 30

    To use and modify scope.vee, it is necessary to be familiar with using the Agilent VEE program. Refer to Agilent VEE documentation if necessary. In View Detail mode, click on Help on the menu bar for assistance in using Agilent VEE. VT1433B User's Guide Getting Started With the VT1433B 2-12 Figure2-4: Agilent VEE Help Text ...

  • VXI VT1433B - page 31

    In detail view there are boxes representing parts of the scope.vee program. For programs that are too large to be viewed all at one time, use the scroll bars at the bottom and left side of the screen to scroll the display. To see more detail, double-click on a box or click on the View Detail (chart symbol) button on the top bar of the box. Some of ...

  • VXI VT1433B - page 32

    Clicking on Alphnumeric in the Display menu sets up a box that specifies how to display the output of a function. Use Agilent VEE to look at the functions that make up the simple “scope.vee” program. This is an example of how the VT1433B can be programmed using Agilent VEE. Click on the Panel View button (sine wave symbol) to go to panel view. ...

  • VXI VT1433B - page 33

    VT1433B User's Guide Getting Started With the VT1433B 2-15 Figure 2-6: minimum.vee (scroll to see entire display) ...

  • VXI VT1433B - page 34

    Other Agilent VEE example programs There are several other example programs that can be examined in the same way as scope.vee. These programs are in the path VXIPNPHpe1432examplesvee on a Microsoft Windows system or /usr/e1432/ vee-examples on an HP-UX system. bsrcsine.vee (Burst SouRCe SINE) This program is similar to scope.vee. It displays ei ...

  • VXI VT1433B - page 35

    C-Language Host Interface Library example programs The VT1432A C-Language Host Interface Library comes with several example programs, which help demonstrate how to use the library. These example programs are found in the the VXIPNPHpe1432examples directory. The programs in this directory are all very small, so that they will be easily understood ...

  • VXI VT1433B - page 36

    Running a demo program: semascope.c To run this program, type:/opt/e1432/demo/ semascope. This program displays the time records for 32 channels (when hooked up to two VT1433B modules with 16 channels each). The channel that is active for changing the display is highlighted. To exit, double-click the horizontal bar symbol in the upper left corner o ...

  • VXI VT1433B - page 37

    Visual Basic example programs VEE programs and the VXI plug&play Library can be used on both UNIX and PC systems. In addition, the PC can use Visual Basic. Visual Basic example programs are at Hpe1432examplesvb on a Microsoft Windows system. VT1433B User's Guide Getting Started With the VT1433B 2-19 ...

  • VXI VT1433B - page 38

    VT1433B User's Guide Getting Started With the VT1433B 2-20 ...

  • VXI VT1433B - page 39

    3 Using the VT1432A 3-1 ...

  • VXI VT1433B - page 40

    Introduction This chapter shows how to use the VT1433B using the VXI plug&play Host Interface Library. The Host Interface Library for the VT1433B is a set of functions that allow the user to program the register-based VT1433B at a higher level than register reads and writes. The library allows groups of VT1433Bs to be set up and programmed as i ...

  • VXI VT1433B - page 41

    What is VXI plug&play ? Agilent uses VXI plug&play technology in the VT1433B. This section outlines some of the details of VXI plug&play technology. Overview The fundamental idea behind VXI plug&play is to provide VXI users with a level of standardization across different vendors well beyond what the VXI standard specifications spel ...

  • VXI VT1433B - page 42

    The actual instrument driver itself is a DLL (Dynamic Linked Library) created from: q A set of source (.C) files. q A set of header (.H) files, used for compiling the file as well as to describe the driver’s calls to any program using the driver. q A standard driver library (.LIB) file, to provide the standard functionality all the drivers would ...

  • VXI VT1433B - page 43

    The Soft Front Panel (SFP) The Soft Front Panel is a stand-alone Windows application, built on top of the VXI plug&play driver DLL; it is used for instrument evaluation and debugging and as a demo. It is not a programmable interface to the instrument, nor can it be used to generate code. The soft front panel also accesses the same Windows Help ...

  • VXI VT1433B - page 44

    Header and Library Files In the Windows environment, the following files are in the directory VxipnpWinXXHpe1432 hpe1432.fp The “FP” file used by VEE and CVI hpe1432.hlp Windows help file hpe1432.kb Knowledge base file hpe1432.bas header for Visual Basic hpe1432.exe Soft front panel program Binhpe1432_32.dll The VXI plug&play driver Inc ...

  • VXI VT1433B - page 45

    Channels and groups This section gives some information about using channels and groups. For more detailed information see the VT1433B help text. Channel Groups In the VT1433B VXI plug&play driver, a channel group is the basic unit of hardware control. To control any channel, it must first be assigned to a group with the hpe1432_createChannelGr ...

  • VXI VT1433B - page 46

    Creating a Channel Group The function hpe1432_ createChannelGroup creates and initializes a channel group. A channel group allows commands to be issued to several VT1433B channels at once, simplifying system setup. Channel groups can be overlapped. The state of an individual VT1433B channel that is in more than one channel group is determined by th ...

  • VXI VT1433B - page 47

    Multiple-module/mainframe Measurements Grouping of Channels/Modules The interface library for the VT1433B is designed to allow programming of several channels from one or several distinct modules, as if they were one entity. Each VT1433B module has up to eight channels. The library may control up to a maximum of 255 VT1433B modules (8160 channels). ...

  • VXI VT1433B - page 48

    All modules except the “system module” need to be set to use the VXI TTL trigger lines as the clock source. Use hpe1432_ setClockSource for this. Set the “system module” to output the clock. Use hpe1432_setClockMaster for this. After this is done, all system sync pulses come from the “system module” and drive the measurement state machi ...

  • VXI VT1433B - page 49

    Managing Multiple-mainframe Measurements In a single-mainframe measurement, the VT1433B communicates with other VT1433Bs through the TTLTRG lines. However, when using the VXI-MXI bus extender modules, the TTLTRG lines, which carry the group synchronization pulse and sample clock, are extended only in one direction. This unidirectional signal connec ...

  • VXI VT1433B - page 50

    In the example above, Mainframe A contains the Slot 0 Controller for a multiple mainframe system. Mainframe A is connected to Mainframe B with a VXI-MXI interface, Agilent/HP E1482B. To successfully manage this multiple mainframe environment, use the following guidelines. q Locate modules with logical addresses less than 128 in Mainframe A. q Locat ...

  • VXI VT1433B - page 51

    VT1433B User's Guide Using the VT1433B 3-13 MXI-2 Extender Interface MXI-2 Extender Interface MXI Bus Cable VXI Mainframe A VXI Mainframe B VXI Mainframe C VT1433B VT1433B VT1433B bus Fail Cal Acs Trigger ExSamp 8-CHANNEL 196 kSa/s DIGITIZER +DSP Chan 5-8 ExTrig Chan 1-4 bus Fail Cal Acs Trigger ExSamp 8-CHANNEL 196 kSa/s DIGITIZER +DSP Chan 5 ...

  • VXI VT1433B - page 52

    Synchronization in Multiple-mainframe Measurements A TTL Trigger line between VT1433Bs making group measurements keeps all modules synchronized. This is an open-collector line where each module holds the one designated as the SYNC line low until the module is ready to advance to the next measurement state. Another TTL Trigger line is designated to ...

  • VXI VT1433B - page 53

    Module Features Data Flow Diagram and FIFO Architecture The illustration on the next page shows data flow in the VT1433B. In this example there are four 4-channel input assemblies for a total of sixteen input channels. The data for all channels is sent to the FIFO. The FIFO is divided into sections, one for each channel. The data moves through a ci ...

  • VXI VT1433B - page 54

    VT1433B User's Guide Using the VT1433B 3-16 ch 1 Input 1 Input 2 ch 8 S tatic RAM FIFO FIFO (DRAM) circular buffer to VME Bus or Local Bus trigger ch 1 ch 2 ch 3 ch 4 ch 5 VME Bus Local Bus 96002 trigger 96002 host port Local Bus FIFO Figure 3-5: Data flow and FIFO architecture ...

  • VXI VT1433B - page 55

    Base Sample Rates Baseband Measurement Spans The table on the following page shows the measurement spans available for base sample rates, for baseband measurements. “Fs” is the sample frequency or sample rate. The value for zero divide-by-two steps and no divide-by-five step is the top measurement span corresponding to the sample rate. This is ...

  • VXI VT1433B - page 56

    Table of Baseband Measurement Spans (part 1 of 6) All values are in Hertz (Hz). sample frequency (Fs) —> 48000 50000 51200 number of ÷2 steps with ÷5 without ÷5 with ÷5 without ÷5 with ÷5 without ÷5 0 3750.000000 18750.000000* 3906.250000 19531.250000* 4000.000000 20000.000000* 1 1875.000000 9375.000000 1953.125000 9765.625000 2000.0000 ...

  • VXI VT1433B - page 57

    Table of Baseband Measurement Spans (part 2 of 6) All values are in Hertz (Hz). sample frequency (Fs) —> 64000 65536 51200 number of ÷2 steps with ÷5 without ÷5 with ÷5 without ÷5 with ÷5 without ÷5 0 5000.000000 25000.000000* 5120.000000 25600.000000* 6000.000000 30000.000000* 1 25000.000000 12500.000000 2560.000000 12800.000000 3000.0 ...

  • VXI VT1433B - page 58

    Table of Baseband Measurement Spans (part 3 of 6) All values are in Hertz (Hz). sample frequency (Fs) —> 80000 81920 96000 number of ÷2 steps with ÷5 without ÷5 with ÷5 without ÷5 with ÷5 without ÷5 0 6250.000000 31250.000000* 6400.000000 32000.000000* 7500.000000 37500.000000* 1 3125.000000 15625.000000 3200.000000 16000.000000 3750.00 ...

  • VXI VT1433B - page 59

    Table of Baseband Measurement Spans (part 4 of 6) All values are in Hertz (Hz). sample frequency (Fs) —> 10000 102400 128000 number of ÷2 steps with ÷5 without ÷5 with ÷5 without ÷5 with ÷5 without ÷5 0 7812.500000 39062.500000* 8000.000000 40000.000000* 10000.000000 50000.000000* 1 3906.250000 19531.250000 2000.000000 20000.000000 5000 ...

  • VXI VT1433B - page 60

    Table of Baseband Measurement Spans (part 5 of 6) All values are in Hertz (Hz). sample frequency (Fs) —> 133333 153600 156250 number of ÷2 steps with ÷5 without ÷5 with ÷5 without ÷5 with ÷5 without ÷5 0 10416.666667 52083.333333* 12000.000000 60000.000000* 12207.031250 61035.156250* 1 5208.333333 26041.666667 6000.000000 30000.000000 6 ...

  • VXI VT1433B - page 61

    Table of Baseband Measurement Spans (part 6 of 6) All values are in Hertz (Hz). sample frequency (Fs) —> 163840 192000 196608 number of ÷2 steps with ÷5 without ÷5 with ÷5 without ÷5 with ÷5 without ÷5 0 12800.000000 64000.000000* 15000.000000 75000.000000* 15360.000000 76800.000000* 1 6400.000000 32000.000000* 7500.000000 37500.000000 ...

  • VXI VT1433B - page 62

    VT1433B sample frequencies The following is a list of all sample frequencies (in Hz) available on the VT1433B, including those not listed in the preceding table. 48000.0 49152.0 50000.0 51200.0 52400.852878 61440.0 62500.0 64000.0 65536.0 66666.666667 76800.0 78125.0 80000.0 81920.0 69000.0 98304.0 10000.0 102400.0 123880.0 125000.0 128000.0 133333 ...

  • VXI VT1433B - page 63

    Measurement Process Measurement Setup and Control When the VT1433B makes a measurement, the measurement itself consists of two phases: the measurement initialization and the measurement loop. Each of these phases consists of several states, through which the measurement progresses. The transition from one state to the next is tied to a transition i ...

  • VXI VT1433B - page 64

    Parameter Settings Many parameters are channel-dependent, meaning that each channel can be set independently of the others in the module. Other parameters are module-dependent; changing a module-dependent parameter for a channel will change it for all channels on that module. For example, changing blocksize, a module-dependent parameter, for input ...

  • VXI VT1433B - page 65

    In the BOOTING state, the digital processors of the module load their parameters and their program. Once done, the module releases the Sync/Trigger line and moves to the BOOTED state. The VT1433B stays in the BOOTED state until it sees a high-to-low transition of the Sync/Trigger line (that is, all the VT1433Bs in the system have booted). In the SE ...

  • VXI VT1433B - page 66

    Upon entering the ARM state the VT1433B starts saving new data in its FIFO. It remains in the ARM state until the Sync/Trigger line goes high. If the VT1433B is programmed with a pre-trigger delay, it collects enough data samples to satisfy this pre-trigger delay and then releases the Sync/Trigger line. If no pre-trigger delay has been programmed, ...

  • VXI VT1433B - page 67

    Arm and Trigger This section explains some terminology relating the the “Arm” and “Trigger” steps in the measurement loop. As an example a measurement might be set up to arm at a certain RPM level and then subsequently trigger at an external event corresponding to top dead center (TDC). The settings would be: q Arm: RPM Step Arm q Trigger: ...

  • VXI VT1433B - page 68

    VT1433B Triggering The following is a short discussion of triggering for the VT1433B. Triggering is defined as the transition from the armed state to the triggered state. This transition is caused by a low going edge on a TTL trigger line. The function hpe1432_ getTtltrgLines selects which of the eight TTL trigger lines is to be used. The low-going ...

  • VXI VT1433B - page 69

    Trigger Level To set the trigger level, use hpe1432_ setTriggerMode to select “level” or “bound” mode; and use hpe1432_setTriggerLevel twice to set both the upper and lower trigger levels. The difference between the upper and lower trigger levels must be at least 10% of full scale (and 10% is usually the best amount). Also use hpe1432setTri ...

  • VXI VT1433B - page 70

    Data Transfer Modes The VT1433B can be programmed to use either of two data transfer modes: overlap block mode and continuous mode. Block mode will be discussed first. Block Mode (Agilent/HP E1431A) The VT1433B’s overlap block mode is similar the block mode which is used in some Agilent instruments such as the Agilent/HP E1431A. In block mode, th ...

  • VXI VT1433B - page 71

    Limit on Queuing of Data In overlap block mode, a number of trigger events may be queued up before the host reads the data for those triggers. The host may get further and further behind the data acquisition. However, if the host gets far enough behind that the FIFO fills up, data acquisition must momentarily stop and wait for data to get transferr ...

  • VXI VT1433B - page 72

    VT1433B Interrupt Behavior Interrupt Setup For an example of interrupt handling see the program event.c in the examples directory. The VT1433B VXI module can be programmed to interrupt a host computer using the VME interrupt lines. VME provides seven such lines. Using hpe1432_setInterruptPriority , the VT1433B module can be set up to use any one of ...

  • VXI VT1433B - page 73

    Once it has done this interrupt, the module will not do any more VME interrupts until re-enabled with hpe1432_ reenableInterrupt. Normally, the last thing a host computer’s interrupt handler should do is call hpe1432_ reenableInterrupt. Events that would have caused an interrupt, but which are blocked because hpe1432_reenableInterrupt has not yet ...

  • VXI VT1433B - page 74

    Host Interrupt Handling When the VT1433B asserts the VME interrupt line, the program will cause the specified interrupt handler to get called. Typically the interrupt handler routine will read data from the module and then re-enable VT1433B interrupts with hpe1432_reenableInterrupt . The call to hpe1432_ reenableInterrupt must be done unless the ho ...

  • VXI VT1433B - page 75

    Some parameters, such as clock frequency or data transfer mode, apply globally to a module. When a channel ID is used to change a parameter that applies to a whole module, the channel ID is used to determine which module. The parameter is then changed for that module. Starting and stopping a measurement is somewhat like setting a global parameter. ...

  • VXI VT1433B - page 76

    New features of the VT1433B software Span and center frequency Decimation spans are now allowed for all legal sample rates with the VT1433B. Zoom mode Zoom mode is available for all legal sample rates with the VT1433B. The zoom span will be allowed to go as high as ctock-freq/2.56 (the maximum non-zoom span). In addition, the maximum center frequen ...

  • VXI VT1433B - page 77

    Where to get more information There is more information available about the VT1432A. This section shows how to access it and print it, if desired. The Function Reference for VXI plug&play On a PC, the VT1432A Function Reference is in Microsoft Help text. Select the Help icon in the “VXIPNP” folder. Refer to Microsoft Windows documentation ( ...

  • VXI VT1433B - page 78

    VT1433B User's Guide Using the VT1433B 3-40 ...

  • VXI VT1433B - page 79

    4 The Host Interface Library 4-1 ...

  • VXI VT1433B - page 80

    Introduction The Host Interface Library for the VT1433B is a set of functions that allows the register-based VT1433B to be programed at a higher level than register reads and writes. The library allows groups of VT1433Bs to be set up and programmed as if they were one entity. The VT1433B uses the same software as the VT1432A 16 Channel 51.2 kSample ...

  • VXI VT1433B - page 81

    Header and Library Files The /opt/e1432/lib directory contains several versions of the VT1432A Host Interface library: lib1432.a A normal HP-UX archive library, used by host programs wanting to talk to VT1433B hardware. lib1432.sl An HP-UX shared library, used by host programs wanting to talk to VT1433B hardware. This and the above archive library ...

  • VXI VT1433B - page 82

    Parameter Information Description of VT1433B Parameters Some parameters, such as range or coupling, apply to specific channels. When a channel ID is given to a function that sets a channel-specific parameter, only that channel is set to the new value. Some parameters, such as clock frequency or data transfer mode, apply globally to a module. When a ...

  • VXI VT1433B - page 83

    Parameter Lists This section shows which parameters are global parameters, which are channel-specific and what types of channels the channel-specific parameters apply to. Default values are shown for all of these parameters. In addition, each parameter is categorized as “abort”, “wait”, “immediate” or “glitch” depending on the behav ...

  • VXI VT1433B - page 84

    Parameter Default Value Changes decimation_output Single Pass Wait decimation_oversample Off Wait decimation_undersamp 1 Wait delta_order 0.1 Wait fifo_size 0 (Use All DRAM) Wait filter_settling_time 64 samples Wait internal_debug 0x100 Immediate interrupt_mask 0 Immediate interrupt_priority None Immediate lbus_mode Pipe Immediate lbus_reset Off (N ...

  • VXI VT1433B - page 85

    196.608 kHz 4-channel Input Parameters Parameter Default Value Changes active Off Abort anti_alias_digital(*) On Abort auto_range_mode Up/Down Immediate calc_data Time Wait coupling DC Glitch enable On Immediate filter_freq 200 kHz Immediate input_high Normal Glitch input_low Floating Glitch input_mode(*) Volt Glitch range 10 V Glitch range_charge ...

  • VXI VT1433B - page 86

    Option 1D4 Single-channel Source Parameters Parameter Default Value Changes active Off Abort amp_scale 1.0 Immediate anti_alias_digital On Wait duty_cycle 0.5 Immediate filter_freq 25.6 kHz Wait ramp_rate 1 s Wait range 0.041567 volt Immediate sine_freq 1000 Hz Immediate sine_phase 0 Degrees Immediate source_blocksize 0 (Use Input Blocksize ) Wait ...

  • VXI VT1433B - page 87

    Option AYF Tachometer Parameters Parameter Default Value Changes active Off Abort input_high Normal Immediate pre_arm_rpm 600 RPM Immediate rpm_high 6000 RPM Immediate rpm_interval 25 RPM Immediate rpm_low 600 RPM Immediate rpm_smoothing 0 Immediate tach_decimate 0 Immediate tach_holdoff 10 µ s Immediate tach_max_time 30 s Immediate tach_ppr 1 Imm ...

  • VXI VT1433B - page 88

    Channel and Group IDs Most functions in the VT1432A C-Language Host Interface Library take an ID parameter which specifies what channel or group of channels the function should apply to. The ID can either be a channel ID or a group ID. If a group ID is used, then the function is applied to each channel contained in the group. Channel IDs Channel ID ...

  • VXI VT1433B - page 89

    Group IDs Group IDs are logical IDs which are created by a call to e1432_create_channel_group. This function can be called multiple times to create multiple groups and each group can contain any combination of channels, including mixtures of different types of channels. The channel groups can overlap as well. This function creates and initializes a ...

  • VXI VT1433B - page 90

    Multiple-module/Mainframe Measurements Grouping of Channels/Modules The interface library for the VT1433B is designed to allow programming of several channels from one or several distinct modules, as if they were one entity. Each VT1433B module has up to 16 channels. The library may control up to a maximum of 255 VT1433B modules (8160 channels). Wh ...

  • VXI VT1433B - page 91

    All modules need to be set to use the shared sync line rather than the default setting of internal sync. This can be done with the e1432_set_multi_sync function, using the group ID. One module of the set of modules needs to be set to output the sync pulse. The module with the lowest VXI logical address is called the “system module” and assigned ...

  • VXI VT1433B - page 92

    Managing Multiple-mainframe Measurements In a single mainframe measurement, the VT1433B communicates with other VT1433Bs through the TTLTRG lines. However, when using the VXI-MXI bus extender modules, the TTLTRG lines, which carry the group synchronization pulse and sample clock, are extended only in one direction. This unidirectional signal connec ...

  • VXI VT1433B - page 93

    In the example above, Mainframe A contains the Slot 0 Controller for a multiple mainframe system. Mainframe A is connected to Mainframe B with a VXI-MXI interface, Agilent/HP E1482B. To successfully manage this multiple mainframe environment, use the following guidelines. q Locate modules with logical addresses less than 128 in Mainframe A. q Locat ...

  • VXI VT1433B - page 94

    VT1433B User's Guide The Host Interface Library 4-16 MXI-2 Extender Interface MXI-2 Extender Interface MXI Bus Cable VXI Mainframe A VXI Mainframe B VXI Mainframe C VT1433B VT1433B VT1433B bus Fail Cal Acs Trigger ExSamp 8-CHANNEL 196 kSa/s DIGITIZER +DSP Chan 5-8 ExTrig Chan 1-4 bus Fail Cal Acs Trigger ExSamp 8-CHANNEL 196 kSa/s DIGITIZER +D ...

  • VXI VT1433B - page 95

    Synchronization in Multiple-mainframe Measurements A TTL Trigger line between VT1433Bs making group measurements keeps all modules synchronized. This is an open-collector line where each module holds the one designated as the SYNC line low until the module is ready to advance to the next measurement state. Another TTL Trigger line is designated to ...

  • VXI VT1433B - page 96

    Measurement Process Measurement Setup and Control When the VT1433B makes a measurement, the measurement itself consists of two phases: the measurement initialization and the measurement loop. Each of these phases consists of several states, through which the measurement progresses. The transition from one state to the next is tied to a transition i ...

  • VXI VT1433B - page 97

    Parameter Settings Many parameters are channel-dependent, meaning that each channel can be set independently of the others in the module. Other parameters are module-dependent; changing a module-dependent parameter for a channel will change it for all channels on that module. For example, changing blocksize, a module-dependent parameter, for input ...

  • VXI VT1433B - page 98

    In the BOOTING state, the digital processors of the module load their parameters and their program. Once done, the module releases the Sync/Trigger line and moves to the BOOTED state. The VT1433B stays in the BOOTED state until it sees a high-to-low transition of the Sync/Trigger line (that is, all the VT1433Bs in the system have booted). In the SE ...

  • VXI VT1433B - page 99

    Upon entering the ARM state the VT1433B starts saving new data in its FIFO. It remains in the ARM state until the Sync/Trigger line goes high. If the VT1433B is programmed with a pre-trigger delay, it collects enough data samples to satisfy this pre-trigger delay and then releases the Sync/Trigger line. If no pre-trigger delay has been programmed, ...

  • VXI VT1433B - page 100

    Arm and Trigger This section explains some terminology relating the the “Arm” and “Trigger” steps in the measurement loop. As an example a measurement might be set up to arm at a certain RPM level and then subsequently trigger at an external event corresponding to top dead center (TDC). The settings would be: q Arm: RPM Step Arm q Trigger: ...

  • VXI VT1433B - page 101

    VT1433B Triggering The following is a short discussion of triggering for the VT1433B. Triggering is defined as the transition from the armed state to the triggered state. This transition is caused by a low going edge on a TTL trigger line. Which one of the eight TTL trigger line is chosen by e1432_get_ttltrg_lines(). The low-going transition of the ...

  • VXI VT1433B - page 102

    Data Transfer Modes The VT1433B can be programmed to use either of two data transfer modes: overlap block mode and continuous mode. Block mode will be discussed first. Block Mode (Agilent/HP E1431A) In block mode, the input hardware acquires one block after getting an arm and trigger. It does not allow the system to trigger until it is ready to pro ...

  • VXI VT1433B - page 103

    Limit on Queuing of Data In overlap block mode, a number of trigger events may be queued up before the host reads the data for those triggers. The host may get further and further behind the data acquisition. However, if the host gets far enough behind that the FIFO fills up, data acquisition must momentarily stop and wait for data to get transferr ...

  • VXI VT1433B - page 104

    VT1433B Interrupt Behavior Interrupt Setup The VT1433B VXI module can be programmed to interrupt a host computer using the VME interrupt lines. VME provides seven such lines and the VT1433B module can be told to use any one of them (see e1432_set_interrupt_priority). The VT1432A can interrupt the host computer in response to different events. A mas ...

  • VXI VT1433B - page 105

    VT1433B Interrupt Handling To make the VT1433B module do the interrupt, both a mask and a VME Interrupt line must be specified, by calling e1432_set_interrupt_mask and e1432_set_interrupt_priority respectively. Once the mask and line have been set and an interrupt occurs, the cause of the interrupt can be obtained by reading the E1432_IRQ_STATUS_RE ...

  • VXI VT1433B - page 106

    Host Interrupt Handling When the VT1433B asserts the VME interrupt line, SICL will cause the specified interrupt handler to get called. Typically the interrupt handler routine will read data from the module and then re-enable VT1433B interrupts with e1432_reenable_interrupt. The call to e1432_reenable_interrupt must be done unless the host is not i ...

  • VXI VT1433B - page 107

    VT1433B Parameters Some parameters, such as range or coupling, apply to specific channels. When a channel ID is given to a function that sets a channel-specific parameter, only that channel is set to the new value. Some parameters, such as clock frequency or data transfer mode, apply globally to a module. When a channel ID is used to change a param ...

  • VXI VT1433B - page 108

    VT1433B User's Guide The Host Interface Library 4-30 ...

  • VXI VT1433B - page 109

    5 Module Description VT1433B User's Guide Module Description 5-1 ...

  • VXI VT1433B - page 110

    Module Features The VT1433B 8-Channel 196 kSamples/s Digitizer plus DSP is a VXI C-sized, scalable input module. The VT1433B may contain one or two 4-channel input assemblies so that the module may have a total of up to eight inputs. The following is a list of some of the features of the VT1433B. See “Specifications” for more detailed informati ...

  • VXI VT1433B - page 111

    Block Diagram For block diagrams of the Arbitrary Source and the Tachometer, see the chapters on the Arbitrary Source option and the Tachometer option. VT1433B User's Guide Module Description 5-3 LBUS/FIFO (optional) Local bus Bus connector 96002 micro- processor SRAM 512 kB DRAM 32 MB Input 1 Input 2 (optional) DMA/ memory control T achometer ...

  • VXI VT1433B - page 112

    VT1433B User's Guide Module Description 5-4 Input Circuit + – Differential Amp Input Circuit Range Selection Input Circuit Input Circuit Differential Amp Differential Amp Differential Amp Range Selection Range Selection Anti-Alias Filter Range Selection Anti-Alias Filter Anti-Alias Filter Anti-Alias Filter Delta- Sigma ADC Delta- Sigma ADC D ...

  • VXI VT1433B - page 113

    VT1433B Front Panel Description Front panels for four or eight channels The VT1433B may have any of several front panels depending on options and number of input channels. The following illustration shows front panels for four or eight channels. VT1433B User's Guide Module Description 5-5 b b u u s s Fail Fail Shut Shut Acs Acs T rigger Trigge ...

  • VXI VT1433B - page 114

    Standard Front Panel This is the front panel for a standard VT1433B (this example has eight inputs). The LEDs and connectors are described on the next page. If the VT1433B has an Arbitrary Source (VT1433B-1D4) or a Tachometer (VT1433B-AYF) its front panel will be different. See the chapter on the Arbitrary Source or the chapter on the Tachometer fo ...

  • VXI VT1433B - page 115

    Status LEDs q Fail: This is the standard VXI “Failed” indicator. It lights briefly when powering up and normally goes out after a few seconds. If it stays on it indicates a hardware failure in the module. q Acs: This is the standard VXI “Access” indicator. When it is on, it indicates that another device on the bus is contacting the module, ...

  • VXI VT1433B - page 116

    VXI Backplane Connections Power Supplies and Ground The VT1433B conforms to the VME and VXI specifications for pin assignment. The current drawn from each supply is given in the specifications chapter. Data Transfer Bus The VT1433B conforms to the VME and VXI specifications for pin assignment and protocol. A16, A24, D16 and D32 data transfers are s ...

  • VXI VT1433B - page 117

    The Local Bus (VT1433B-UGV) The VXI specification includes a 12-wire Local Bus between adjacent module slots. Using the Local Bus, VXI Technology has defined a standard byte-wide ECL protocol which can transfer data from left to right at up to 15.7 Mbytes /s using VT1433B. If equipped with VT1433B-UGV, the VT1433B can be programmed to output its da ...

  • VXI VT1433B - page 118

    The VT1433B VXI Device Address Space The VXI system architecture defines two types of address space. A16 space consists of 64 kBytes and A24 consists of 16 MBytes . The VT1433B has a 32-bit port through which it has access to the A16 and A24 space. It can also use D32 to send and receive data though the port. Or it can use the port for 16-bit data ...

  • VXI VT1433B - page 119

    The mapping of the fixed and movable windows is illustrated as follows: Address FFFFF 16 8000 0 16 Movable DSP Bus Window Movable 7FFFF 16 3000 0 16 Fixed DSP Bus Window Fixed 2FFFF 16 2000 0 16 Send/Receive Data Registers 1FFFF 16 0004 F 16 Fixed DSP Bus Window 0003 F 16 0000 0 16 VXI Bus A16 Registers For more information, see “The A24 Register ...

  • VXI VT1433B - page 120

    List of A16 Registers The following lists the A16 registers. For more information see “The A16 Registers” in the chapter titled Register Definitions. Address Read Write 3E 16 Parameter 7 Register 3C 16 3A 16 Parameter 6 Register 38 16 36 16 Parameter 5 Register 34 16 32 16 Parameter 4 Register 30 16 2E 16 Parameter 3 Register 2C 16 2A 16 Parame ...

  • VXI VT1433B - page 121

    Trigger Lines (TTLTRG) TTLTRG consist of eight TTL lines on the VXI backplane on connector P2. They are available to provide synchronization between devices. VXI devices can use the TTLTRG lines for simple communication with other devices. For example, a device can wait for a line to go high before taking an action or it can assert a line as a sign ...

  • VXI VT1433B - page 122

    Providing an External Clock The VT1433B can be programed to accept an external word rate clock from the Sample 0 line on the VXI Bus. The digital filters are still functional, providing a range of effective word rates. All sampling is done simultaneously and is not multiplexed. To connect an External Sample Clock, use the External Sample SMB connec ...

  • VXI VT1433B - page 123

    Calibration Description The Cal connector on the front panel of the standard VT1433B can be configured (in software) as either an input or an output. It can be set to any of four settings: q DC - The VT1433B outputs a DC calibration signal from the millivolt range up to 15 volts. q AC - The VT1433B outputs a signal from an Arbitrary Source option ( ...

  • VXI VT1433B - page 124

    VT1433B User's Guide Module Description 5-16 ...

  • VXI VT1433B - page 125

    6 The Arbitrary Source Option (VT1433B-1D4) VT1433B User's Guide The Arbitrary Source Option 6-1 ...

  • VXI VT1433B - page 126

    Arbitrary Source Description An arbitrary source can be included with the VT1433B 8-Channel 196 kSamples/s Digitizer plus DSP as VT1433B-1D4. (It cannot be installed with a Tachometer Option, VT1433B-AYF.) The Arbitrary Source Option can supply arbitrary or sine signals under control of measurement software. Trigger The Arbitrary Source can be used ...

  • VXI VT1433B - page 127

    Block Diagram VT1433B User's Guide The Arbitrary Source Option 6-3 56002 128k x 24 RAM Analog registers Gate array control Digital interpolation filter 8 96002 “B” Bus or A24 VXI Signal DAC Range DAC 25.6 kHz 6.4 kHz COLA AMP Summer input Shutdown circuitry OU T Figure 6-1: Arbitrary source option block diagram ...

  • VXI VT1433B - page 128

    The Arbitrary Source Option Front Panel The VT1433B with the Arbitrary Source Option may have four or eight input channels. The following illustration shows a front panel for eight channels. The LEDs and connectors are described on the next page. VT1433B User's Guide The Arbitrary Source Option 6-4 bus Fail Shut Acs Trigger COLA 8-CHANNEL 196 ...

  • VXI VT1433B - page 129

    LEDs and Connectors for the Arbitrary Source Option Status LEDs q Fail: This is the standard VXI “Failed” indicator. It lights briefly when powering up and normally goes out after a few seconds. If it stays on it indicates a hardware failure in the module. q Acs: This is the standard VXI “Access” indicator. When it is on, it indicates that ...

  • VXI VT1433B - page 130

    VT1433B User's Guide The Arbitrary Source Option 6-6 ...

  • VXI VT1433B - page 131

    7 The Tachometer Option (VT1433B-AYF) VT1433B User's Guide The Tachometer Option 7-1 ...

  • VXI VT1433B - page 132

    Tachometer Description A tachometer input can be included with the VT1433B 8-Channel 196 kSamples/s Digitizer plus DSP as VT1433B-AYF. (It cannot be installed with an Arbitrary Source Option, VT1433B-1D4.) The Tachometer Option is a two channel tachometer input used to capture the contents of a freerun counter whenever an external input crosses a p ...

  • VXI VT1433B - page 133

    Input Count Division The tachometer can be programmed to divide the input signal. For example if a signal is coming in at 100 counts per second, the tachometer can be set to look at only every 10th count for a result of 10 counts per second. Holdoff Time The tachometer can be programmed to wait for a specified period of time between counts that it ...

  • VXI VT1433B - page 134

    The Tachometer Option Front Panel The VT1433B with the Tachometer Option may have four or eight input channels. The following illustration shows a front panel for eight channels. The LEDs and connectors are described on the next page. VT1433B User's Guide The Tachometer Option 7-4 bus Fail T ach 2 Acs Trigger T ach 1 8-CHANNEL 196 kSa/s DIGITI ...

  • VXI VT1433B - page 135

    LEDs and Connectors for the Tachometer Option Status LEDs q Fail: This is the standard VXI “Failed” indicator. It lights briefly when powering up and normally goes out after a few seconds. If it stays on it indicates a hardware failure in the module. q Acs: This is the standard VXI “Access” indicator. When it is on, it indicates that anothe ...

  • VXI VT1433B - page 136

    VT1433B User's Guide The Tachometer Option 7-6 ...

  • VXI VT1433B - page 137

    8 Break Out Boxes VT1433B User's Guide Break Out Boxes 8-1 ...

  • VXI VT1433B - page 138

    Introduction A Break Out Box connects the VT1432A or VT1433B to a set of connectors to receive input signals. Several types of Break Out Boxes are available. This chapter covers: q VT3240A Voltage Break Out Box q VT3241A ICP ® Break Out Box Other Break Out Boxes include the VT3242A Charge Break Out Box and the VT3243A Microphone Break Out Box. See ...

  • VXI VT1433B - page 139

    The VT3240A and VT3241A Break Out Boxes Each of the Break Out Boxes described in this section has eight BNC connectors for input. They each have two cables which connect to the sub-miniature “D” connectors on the front panel of the VT1432A/33B. Each of the two cables carries four channels. VT1433B User's Guide Break Out Boxes 8-3 VOL T AGE ...

  • VXI VT1433B - page 140

    VT3240AVoltage-type Break Out Box In this type of Break Out Box the signal is sent straight through to the sub-miniature “D” connectors on the VT1432A/33B. VT3241A ICP ® -type Break Out Box Each of the eight connectors in this type of Break Out Box is connected to an independent, floating current source. These are intended to power integrated- ...

  • VXI VT1433B - page 141

    Break Out Box Cables Making a Custom Break Out Box Cable A cable to connect the Break Out Box with the VT1432A/33B is supplied with the each of the Break Out Boxes described in this chapter. However, this section is included for those users who may want to make their own connecting cable. The drawing on this page shows the AMP part numbers for the ...

  • VXI VT1433B - page 142

    Pin definitions for input connector definition pin # pin # definition RFI GND/Cable Shield 26 13 - Diff 1 +24V Power 25 12 +Diff 1 GND Return for ± 24V 24 11 RFI GND/Drain Shield 1 -24 Power 23 10 RFI GND/Drain Shield 2 RFI GND 22 9 - Diff 2 I2C SCL 21 8 + Diff 2 CAL HIGH 20 7 CAL LOW BoB_EN 19 6 - Diff 3 RFI GND 18 5 + Diff 3 I2C SDA 17 4 RFI GND ...

  • VXI VT1433B - page 143

    Recommendations on wiring for the VT1432A/33B 4-Channel Input Connector Allowed Connections Differential Input Channels Connect at VT1432A/33B end of cabling and at DUT Recommended: shielded twisted pair 1 + Diff 4 2 - Diff 4 5 + Diff 3 6 - Diff 3 8 + Diff 2 9 - Diff 2 12 + Diff 1 13 - Diff 1 Input Channel Shielding Connect at VT1432A/33B end of ca ...

  • VXI VT1433B - page 144

    Dis-allowed Connections Do NOT connect these pins on VT1432A/33B end of cabling. These signals and supplies are provided for VXI Technology specified break out boxes and are unspecified for other usage Do not use: 15 I2C_EN 17 I2C_SDA 21 I2C_SCL 16 RFI GND/I2C Shield 18 RFI GND/I2C Shield 22 RFI GND/I2C Shield 19 BOB_EN 7 CAL_LOW 20 CAL_HIGH 23 -24 ...

  • VXI VT1433B - page 145

    9 Troubleshooting the VT1433B VT1433B User's Guide Troubleshooting the VT1433B 9-1 ...

  • VXI VT1433B - page 146

    Diagnostics The following describes a limited diagnostic program for the VT1432A, VT1433B and VT1434A. It is to be run from an HP-UX host. The program is called “hostdiag.” It can be found with the VT1432A Host Interface Software Library at location /usr/e1432/bin. location: /usr/e1432/bin Usage: hostdiag [-hPsuvV] [-f file] [-L laddr] [-S slot ...

  • VXI VT1433B - page 147

    -u Display usage message. -v Specifies the verbose printing. Normally, hostdiag does not print anything unless an error is found. With this option, hostdiag prints status messages as it operates. This option also enables additional diagnostic information which is not generally useful. -V Print version info. Hostdiag returns 0 upon success or return ...

  • VXI VT1433B - page 148

    VT1433B User's Guide Troubleshooting the VT1433B 9-4 ...

  • VXI VT1433B - page 149

    10 Replacing Assemblies VT1433B User's Guide Replacing Assemblies 10-1 ...

  • VXI VT1433B - page 150

    Replaceable Parts For information on upgrading the module or replacing parts, contact a VXI Technology sales and service office. See the inside back cover of this guide for a list of office locations and address. Replacement parts are listed in the following tables: q Assemblies: without VT1433B-AYF or VT1433B-1D4 q Assemblies: with VT1433B-AYF q A ...

  • VXI VT1433B - page 151

    CAGE Code Numbers The following table provides the name and address for the manufacturers’ CAGE code numbers (Mfr Code) listed in the replaceable parts tables. Mfr Code Mfr Name Address 03LB1 VXI Technology, Inc. Irvine, CA 92614 U.S.A. 30817 Laird Technologies Delaware Water Gap, PA 18327 U.S.A. 83486 Elco Industries Inc. Rockford, IL 61101 U.S. ...

  • VXI VT1433B - page 152

    Assemblies: without VT1433B-AYF or VT1433B-1D4 VT1433B User's Guide Replacing Assemblies 10-4 ...

  • VXI VT1433B - page 153

    Ref Des VXI Part Number Qty Description Mfr Code Mfr Part Number A2 E1433-66534 2 PC ASSY-INPUT 03LB1 E1433-66534 A4 E1432-66504 1 PC ASSY-LED 03LB1 E1432-66504 A10 E1433-66510 1 PC ASSY-MAIN OPT UGV 03LB1 E1433-66510 A11 E1433-66511 1 PC ASSY-MAIN 03LB1 E1433-66511 A22 1818-5622 1 ICM DRAM, SIMM, 8x32 03LB1 1818-5622 MP001 E1433-00615 1 SHTF CVR-B ...

  • VXI VT1433B - page 154

    Assemblies: with VT1433B-AYF VT1433B User's Guide Replacing Assemblies 10-6 A10/A11 A22 ...

  • VXI VT1433B - page 155

    Ref Des VTI Part Number Qty Description Mfr Code Mfr Part Number A2 E1433-66534 2 PC ASSY-INPUT 03LB1 E1433-66534 A4 E1432-66504 1 PC ASSY-LED 03LB1 E1432-66504 A5 E1432-66505 1 PC ASSY-OPT AYF 03LB1 E1432-66505 A10 E1433-66510 1 PC ASSY—MAIN OPT UGV 03LB1 E1433-66510 A11 E1433-66511 1 PC ASSY-MAIN 03LB1 E1433-66511 A22 1818-5622 1 ICM DRAM, SIMM ...

  • VXI VT1433B - page 156

    Assemblies: with VT1433B-1D4 VT1433B User's Guide Replacing Assemblies 10-8 ...

  • VXI VT1433B - page 157

    Ref Des VTI Part Number Qty Description Mfr Code Mfr Part Number A2 E1433-66534 2 PC ASSY 03LB1 E1433-66534 A4 E1432-66504 1 PC ASSY-LED 03LB1 E1432-66504 A10 E1433-66510 1 PC ASSY-MAIN OPT UGV 03LB1 E1433-66510 A11 E1433-66511 1 PC ASSY-MAIN 03LB1 E1432-66511 A22 1818-5622 1 ICM DRAM, SIMM, 8x32 03LB1 1818-5622 A41 E1432-66541 1 PC ASSY-OPT 1D4 03 ...

  • VXI VT1433B - page 158

    Cables: without VT1433B-AYF or VT1433B-1D4 Ref Des VTI Part Number Qty Description Mfr Code Mfr Part Number W1 8120-6767 1 CBL-ASM CXL, 290MM 03LB1 8120-6767 W2 8120-6765 2 CBL-ASM CXL, 255 MM 03LB1 8120-6765 W4 8120-6762 1 CBL-FLEX, 5-COND, 225 MML 03LB1 8120-6762 VT1433B User's Guide Replacing Assemblies 10-10 ...

  • VXI VT1433B - page 159

    Cables: with VT1433B-AYF Ref Des VTI Part Number Qty Description Mfr Code Mfr Part Number W2 8120-6765 1 CBL-ASM CXL, 255MM 03LB1 8120-6765 W3 8120-6766 2 CBL-ASM CXL, 03LB1 8120-6766 W4 8120-6762 1 CBL-FLEX, 5-COND, 225MML 03LB1 8120-6762 VT1433B User's Guide Replacing Assemblies 10-11 ...

  • VXI VT1433B - page 160

    Cables: with VT1433B-1D4 Ref Des VTI Part Number Qty Description Mfr Code Mfr Part Number W2 8120-6765 1 CBL-ASM CXL, 255MM 03LB1 8120-6765 W3 8120-6766 2 CBL-ASM CXL, 03LB1 8120-6766 W4 8120-6762 1 CBL-FLEX, 5-COND, 225MML 03LB1 8120-6762 VT1433B User's Guide Replacing Assemblies 10-12 ...

  • VXI VT1433B - page 161

    Front Panel Ref Des VTI Part Number Qty Description Mfr Code Mfr Part Number MP200 E1433-00207 1 PNL-FRT, STANDARD 03LB1 E1433-00207 MP201 E1433-00208 1 PNL-FRT, OPT 1DL 03LB1 E1433-00208 MP203 E1432-44301 1 LBL-FRT PNL SMB’S, STD 03LB1 E1432-44301 MP204 E1432-44302 1 LBL-FRT PNL SMB’S, OPT 1D4 03LB1 E1432-44302 MP205 E1432-44303 1 LBL-FRT PNL ...

  • VXI VT1433B - page 162

    To remove the top cover 1 Remove the five long screws using a T-10 Torx driver and remove the three short screws using a T-8 Torx driver. Lift cover off. VT1433B User's Guide Replacing Assemblies 10-14 ...

  • VXI VT1433B - page 163

    To remove the front panel 1 Remove top cover, see “To remove the top cover.” Gently disconnect cables from the printed circuit assemblies. Using a T-8 Torx driver, remove the two screws that attach the handles to the assembly. Pull out the handles making sure not to lose the two spacers. 2 Using a T-8 Torx driver, remove the screw that attaches ...

  • VXI VT1433B - page 164

    3 Remove the nuts that fasten the cables and assembly to the front panel. Using a 1/4-inch nut driver. 4 Remove ribbon cable from the A4 assembly, by pulling back the latch on the connector and and removing cable. Be sure to note the orientation of the cable. VT1433B User's Guide Replacing Assemblies 10-16 ...

  • VXI VT1433B - page 165

    5 To replace the front panel with another that does not have its own side brackets, remove the brackets from the old front panel using a T-8 Torx driver. Be sure to note the positioning of the brackets, alignment is critical. 6 To replace the front panel with another that does not have the label already attached, remove the tape backing and place i ...

  • VXI VT1433B - page 166

    To remove the input assemblies 1 Remove top cover, see “To remove the top cover.” Remove the front panel, see steps 1 and 2 in “To remove the front panel.” Note that the following steps are showing illustrations of an VT1433B with a standard configuration (two input assemblies). If the VT1433B has option 1DL (one input assembly) , the follo ...

  • VXI VT1433B - page 167

    3 Remove the top assembly by gently pulling it forward, releasing it from the connectors. 4 Remove the remaining input assembly. VT1433B User's Guide Replacing Assemblies 10-19 ...

  • VXI VT1433B - page 168

    To remove the VT1433B-AYF assembly 1 Remove the top cover, see “To remove the top cover.” Disconnect the two cables leading to the A5 assembly and move cables aside. 2 Using a T-10 Torx driver, remove the three screws that attach the assembly to the VT1433B and lift the assembly off. VT1433B User's Guide Replacing Assemblies 10-20 ...

  • VXI VT1433B - page 169

    To remove the VT1433B-1D4 assembly 1 Remove the top cover, see “To remove the top cover.” Disconnect the three cables leading to the A41 assembly and move cables aside. 2 Using a T-10 Torx driver, remove the three screws that attach the assembly to the VT1433B and lift the assembly off. VT1433B User's Guide Replacing Assemblies 10-21 ...

  • VXI VT1433B - page 170

    To remove the A22 assembly 1 Remove the top cover, see “To remove the top cover.” Gently push the silver tabs outward and tilt the A22 assembly forward releasing it from the connector. VT1433B User's Guide Replacing Assemblies 10-22 ...

  • VXI VT1433B - page 171

    To remove the A10/A11 assembly 1 Remove top cover and input assemblies. See “To remove the top cover,” and “To remove the input assemblies.” 2A If the module does NOT have VT1433B-AYF or VT1433B-1D4 do the following: Remove the five standoffs using a 1/4-inch nut driver and remove the three screws using a T-8 Torx driver. VT1433B User' ...

  • VXI VT1433B - page 172

    2B If the module has VT1433B-AYF do the following: Remove the VT1433B-AYF assembly, see “To remove the VT1433B-AYF assembly.” Remove the five long and the three short standoffs using a 1/4-inch nut driver . 2C If the module has VT1433B-1D4 do the following: Remove the 1D4 option assembly, see “To remove the VT1433B-1D4 assembly.” Remove the ...

  • VXI VT1433B - page 173

    11 Backdating VT1433B User's Guide Backdating 11-1 ...

  • VXI VT1433B - page 174

    Backdating This chapter documents modules that differ from those currently being produced. With the information provided in this chapter, this guide can be modified so that it applies to any earlier version or configuration of the module. VT1433B User's Guide Backdating 11-2 ...

  • VXI VT1433B - page 175

    Appendix A Register Definitions VT1433B User's Guide Register Definitions A-1 ...

  • VXI VT1433B - page 176

    The VT1433B VXI Registers The VT1433B 8-Channel 196 kSamples/s Digitizer plus DSP is a register-based VXI device. Unlike message-based devices which use higher-level programming using ASCII characters, register-based devices are programmed at a very low level using binary information. The greatest advantage of this is speed. Register-based devices ...

  • VXI VT1433B - page 177

    Address Read Write 3E 16 Parameter 7 Register 3C 16 3A 16 Parameter 6 Register 38 16 36 16 Parameter 5 Register 34 16 32 16 Parameter 4 Register 30 16 2E 16 Parameter 3 Register 2C 16 2A 16 Parameter 2 Register 28 16 26 16 Parameter 1 Register 24 16 22 16 Query Response Register Command Register 20 16 1E 16 FIFO Count 1C 16 1A 16 Send Data Receive ...

  • VXI VT1433B - page 178

    The A24 Registers The following A24 registers are accessible at the base address defined by the device’s offset Register. The registers at offsets 0 to E 16 are not accessible using longword (D32) accesses. The registers at offsets 10 16 to FFFFF 16 may be accessed by any of the of the D08(EO), D16 or D32 modes. FFFFF 16 8000 0 16 Movable DSP Bus ...

  • VXI VT1433B - page 179

    The VXI Bus Registers are defined as follows: q Id Register: A read of this 16 bit register provides information about the device ’s configuration. Its value is always CFFF16 as defined in the following table. Bit 15-14 13-12 11-0 Contents 11 (Register Based Device) 00 (A16/A24) 111111111111 (Agilent’s ID) q Logical Address Register: A write to ...

  • VXI VT1433B - page 180

    Block Ready: A one (1) indicates that there is a block of data available to be read from the Send Data registers. A zero (0) indicates that less than a full block is available. Data Ready: A one (1) indicates that there is at least one word (32 bits) of data available in the Send Data register. A zero (0) indicates that there is not valid data in t ...

  • VXI VT1433B - page 181

    q Control Register: A write to this register causes specific actions to be executed by the device. The actions are described in the following table. Bit 15 14-2 1 0 Contents A24/A32 Enable Unused Sysfail Inhibit Reset A24/A32 Enable: A one (1) in this field enables access to the device’s A24 VME Bus registers. A zero (0) disables such access. Sys ...

  • VXI VT1433B - page 182

    q Page Map Register: This read/write register defines the internal location of the movable window into the device ’s DSP bus. (This 512 kB window begins at 512 kB into the device ’s A24 registers.) The eight least significant bits of the Page Map register are the page number. These bits are mapped to the internal DSP bus address lines as follow ...

  • VXI VT1433B - page 183

    q IRQ Status Register: This read-only register indicates the reason for asserting the VME Bus interrupt. The format of the data is identical to that of the Status/ID word returned by an interrupt acknowledge ( IACK) cycle. It differs from the IACK cycle in that the IACK cycle will clear the status bits and cause the de-assertion of the IRQ line. Th ...

  • VXI VT1433B - page 184

    q Parameter 1-7 Registers: These are 32-bit RAM locations used to pass parameters along with commands to the device or query responses from the device. See the following section regarding D16/D08 access of 32-bit registers and the communication protocol. 32-bit Registers Several of the A16 registers (and all other 24-bit registers) are implemented ...

  • VXI VT1433B - page 185

    Command/Response Protocol The Command/Response protocol uses the following resources: q Command/Query Response register implemented as a general purpose RAM q Three parameter registers implemented as a general purpose RAM q Additional A24 accessible RAM contiguous with the parameter registers q The Command Ready, Query Response Ready, Err* and Done ...

  • VXI VT1433B - page 186

    In order to avoid contention and/or invalid data reads, there are certain rules that must be observed: 5 A controller must not write to any of the RAM registers when Command/Parameter Ready is false. 6 The DSP must not write to any of the RAM registers when either Command/Parameter Ready or Query Response Ready is true. 7 A controller must not read ...

  • VXI VT1433B - page 187

    Wait for Done This is the procedure to wait for command completion and check for error. 20 Wait for Command/Parameter Ready true. 21 Wait for Done true. 22 I fE r r *=0, handle error. Complex Sequences A robust procedure for sending a query and reading the response would look like this: 23 Send Command. 24 Wait for Done. 25 If no error then Read Re ...

  • VXI VT1433B - page 188

    DSP Bus Registers There are two 32-bit registers in the DSP bus address space. The VXI FPGA does not assert TA* when these registers are accessed. 200A 16 DSP Command Register 200B 16 Boot Register Note that these registers appear multiple times in the memory map, since only the address lines A31-30, A17-13, A9-8 and A3-0 are used for decoding. The ...

  • VXI VT1433B - page 189

    q DSP Command Register: This register is used to assert VXI interrupts and toggle various status register bits. Many of the bits in this register are grouped into related Clock and Value pairs. This allow the bits to be modified independently with single register writes. In order to change an output value, the Clock bit must be written as a one (1) ...

  • VXI VT1433B - page 190

    VT1433B User's Guide Register Definitions A-16 ...

  • VXI VT1433B - page 191

    Glossary A16 registers Address space using 16 address lines. The VXI definition gives each VXI module 64 bytes of A16 registers. A24 registers Address space using 24 address lines. VXI modules can configure how much A24 address space they use. Agilent VEE An Agilent program for graphical programming. See VEE . arbitrary source A signal source capab ...

  • VXI VT1433B - page 192

    C-Library (interface library) A library of functions, written in C language, which can be used to operate the VT1432A and VT1433B. C-size One of several possible sizes for VXI modules. The VT1432A and VT1433B are C-size modules. channel-dependent commands Commands that are channel-dependent change a parameter for each channel independently. COLA Co ...

  • VXI VT1433B - page 193

    DSP Digital Signal Processing. DTB arbitration bus The VT1432A does not use the arbitration bus. The arbitration bus is part of the VXI specification and is used by some modules to request bus control. ECL Emitter-Collector Logic, a standard for electrical signals. Engineering Unit (EU) A scale factor used to convert the output of a transducer (in ...

  • VXI VT1433B - page 194

    kSamples/s Kilosamples per second. LED Light Emitting Diode. Local Bus A high-speed port that is defined as a standard byte-wide ECL protocol which can transfer measurement data at up to 2.62 MSamples per second from left to right on the VXI backplane. logical address The VXI logical address identifies where each module is located in the memory map ...

  • VXI VT1433B - page 195

    register-based VXI device Register-based devices communicate with the VXI Bus by way of registers. They must be programmed with low-level binary commands but they can communicate faster than message-based devices. The VT1432A and VT1433B are register-based VXI devices. registers Memory locations in the hardware of a VXI module which can be used to ...

  • VXI VT1433B - page 196

    Soft Front Panel (SFP) A VXI plug&play program which provides and easy-to-use interface for the VT1432A. It can be used in Microsoft Windows 95 or later or Windows NT. SRAM Static Random Access Memory. summer A circuit that outputs the sum of two input signals. sync/trigger line A TTL line on the VXI back plane, used for synchronization or trig ...

  • VXI VT1433B - page 197

    VME Bus An industry-standard bus on the VXI backplane for module control, setup and measurement data transfers. For measurement data transfers, the Local Bus offers higher transfer rates. VXI VME Extensions for Instrumentation, a standard specification for instrument systems. VXI plug&play A set of standards which provides VXI users with a leve ...

  • VXI VT1433B - page 198

    VT1433B User's Guide Glossary G-8 ...

  • VXI VT1433B - page 199

    Index Numerics 32-bit registers, A-10 writing, A-10 4-channel input, 4-7, 4-10, 7-2 8-channel input, 5-7, 6-5, 7-5 8-channel input (break out box), 8-3 A A16 address space, 5-10 A16 registers, 5-12, A-2 A24 address space, 5-10 A24 registers, 5-11, A-4 A32 address space, 5-10 A-bus, 5-10 AC/DC coupling, 5-2 access LED, 5-7, 6-5, 7-5 acs LED, 5-7, 6- ...

  • VXI VT1433B - page 200

    Shut, 6-5 Tach1, 7-5 Tach2, 7-5 connectors SMB, 5-7, 6-5, 7-5 constant output level amplifier, 6-2, 6-5 continuous mode, 3-32, 3-33, 4-24, 4-25 control measurement, 3-25, 4-18 control register, A-7 controller protocol examples, A-12 count division, 7-3 count register, A-9 coupling, 5-2 covers part numbers, 10-5, 10-7, 10-9 removing, 10-14 create gr ...

  • VXI VT1433B - page 201

    VXIplug&play , 3-39 Windows, 3-5 holdoff time, 7-3 host interface libraries installing, 2-3, 2-5, 2-6 host interface library, 4-2 Host Interface Library, 5-10, A-2 hpe1432_32.dll, 3-6 HP-UX 10.2, 2-3 HP-UX 9.05, 2-3 I icon, 3-6 ICP, 5-2, 8-4 IDLE state, 3-26, 3-27, 3-28, 4-19, 4-20, 4-21 incoming inspection, 1-2 initialization, 3-7 initiation, ...

  • VXI VT1433B - page 202

    tachometer, 4-9 types, 3-36, 4-4, 4-29 part numbers assemblies, 10-5, 10-7, 10-9 cables, 10-10, 10-11, 10-12 front panel, 10-13 port control register, A-7 power supplies, 5-8 pre-arm, 3-29, 4-22 pre-trigger delay, 5-2 priority interrupt bus, 5-8 programs example, 2-10 Visual Basic, 2-19 protocol command/response, A-11 controller, A-12 DSP, A-13 VXI ...

  • VXI VT1433B - page 203

    SETTLING, 3-26, 3-27, 4-19, 4-20 TESTED, 3-26, 3-27, 3-28, 4-19, 4-20, 4-21 TRIGGER, 3-27, 3-28, 4-20, 4-21 static RAM, 5-10 status LEDs, 5-7, 6-5, 7-5 status register, A-5 summer, 6-2 Support, ix Support Resources, ix sync/trigger line, 3-25, 3-26, 3-27, 3-28, 4-18, 4-19, 4-20, 4-21, 5-13 synchronization multiple-mainframe, 3-14, 4-17 TTLTRG, 5-13 ...

Manufacturer VXI Category Stereo Receiver

Documents that we receive from a manufacturer of a VXI VT1433B can be divided into several groups. They are, among others:
- VXI technical drawings
- VT1433B manuals
- VXI product data sheets
- information booklets
- or energy labels VXI VT1433B
All of them are important, but the most important information from the point of view of use of the device are in the user manual VXI VT1433B.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals VXI VT1433B, service manual, brief instructions and user manuals VXI VT1433B. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product VXI VT1433B.

Similar manuals

A complete manual for the device VXI VT1433B, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use VXI VT1433B by users. Manuals are usually written by a technical writer, but in a language understandable to all users of VXI VT1433B.

A complete VXI manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual VXI VT1433B - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the VXI VT1433B, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the VXI VT1433B, that we can find in the current document
3. Tips how to use the basic functions of the device VXI VT1433B - which should help us in our first steps of using VXI VT1433B
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with VXI VT1433B
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of VXI VT1433B in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning VXI VT1433B?

Use the form below

If you did not solve your problem by using a manual VXI VT1433B, ask a question using the form below. If a user had a similar problem with VXI VT1433B it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (0)