Инструкция обслуживания ResMed Flow Generator VPAP Adapt SV

42 страниц 1.26 mb
Скачать

Перейти на страницу of 42

Summary
  • ResMed Flow Generator VPAP Adapt SV - page 1

    1 VP AP™ A dapt S V User Guide English 228234-Eng/3 08 10 228234 VPAP Adapt SV User Guide USA Eng .book Page 1 Mo nday, October 13, 2008 2:15 PM ...

  • ResMed Flow Generator VPAP Adapt SV - page 2

    2 Cont ents Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 VP AP™ A dapt S V 3 User/o wner responsibilit y 3 Definitions 3 Medical Info rmation 3 Quic k Setup Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 VP AP Adapt SV S ystem . . . . . ...

  • ResMed Flow Generator VPAP Adapt SV - page 3

    3 Introduction English Introduction VPAP™ Adapt SV The VP AP Adapt S V flow generato r has been designed to be re liable, comfortable and easy to use. You can observe your treatment and mak e adjustments to settings using a keypad and LCD screen. ! CAUTION USA federal law restricts this device to sale by or on the order of a physician. User/Owner ...

  • ResMed Flow Generator VPAP Adapt SV - page 4

    4 • impaired abilit y to clear secretions • pneumothorax or pn eumomediasti num • recent cranial trauma or surger y • c hronic hypov entilation. Belo w are g ener al war nings and caut ions. Fur ther specific war nings, cautions and not es appear next t o the relev ant instr uctions in the manual. ! WARNINGS • The VP AP Adapt SV is NO T a ...

  • ResMed Flow Generator VPAP Adapt SV - page 5

    5 Quick Setup Guide English The following side effects may arise during a course of therapy with the VP AP A dapt S V: • drying of the nose, m outh, or throat • bloating • ear or sinus discomfort • ey e irritation • skin rashes • c hest discomfort. Quick Setup Guide 1 Attach the air tubing and pressure sensor tube to the VP AP A dapt S ...

  • ResMed Flow Generator VPAP Adapt SV - page 6

    6 7 Remove your mask to stop treatment. If SmartStop is enabl ed, treatment will stop automatically. If th e SmartStop function has been disabled, press the Start/St op key once. VPAP Adapt SV System VPAP Adapt SV Components The VP AP Adapt S V flow g enerator is supplied with air t ubing and a p o wer cord. Masks are supplied separately , as you a ...

  • ResMed Flow Generator VPAP Adapt SV - page 7

    7 VP AP Adapt SV System English Figure 2: Rear of VPAP Adapt SV See page 14 to connect a ResLink™. See page 30 to replace the air filter. Figure 3: Air tubing fully ass embled All necessary air tubing components are packed w ith the VP AP A dapt SV: • a 6’6" (2 m) length of air tubing • a pressure sensor t ube (to measure pressure at t ...

  • ResMed Flow Generator VPAP Adapt SV - page 8

    8 Assembling the VPAP Adapt SV System Complete the following steps to assemb le the VP AP Adapt S V system. Make sure the area around the unit is clean (dus t free) and clear o f bedding, clo thes and any other potential blockag es. To add a hum idifier see page 10. ! WARNING Use only the ai r tubing system supp lied with the VPAP Adapt SV system. ...

  • ResMed Flow Generator VPAP Adapt SV - page 9

    9 VP AP Adapt SV System English 2 Connect air t ubing Connect the a ir tube fir mly ont o t he air outlet at the front of the unit . Figure 4: Front view of VPAP Adapt SV T o attac h the pr essur e sensor t ube to the unit, s crew the Luer connector ont o the soc ket on the side of the unit. Figure 5: Twist ing the Luer lock connect ors together ! ...

  • ResMed Flow Generator VPAP Adapt SV - page 10

    10 Humidifiers A humidifier may be required if you are experiencing dryness of the nose , throat, or mouth. The followi ng ResMed humidifiers are compatib le with the VP AP A dapt S V: Figure 7: Humidifi ers compatible with the VPAP Adap t SV When usin g a humid ifier , stop the VP AP Adapt S V by using the Start/Stop key or by removing the mask to ...

  • ResMed Flow Generator VPAP Adapt SV - page 11

    11 VP AP Adapt SV System English HumidAire 2i™ The HumidAire 2i provides heated hu midification for the VPAP Adapt SV. To attach the HumidAire 2i, remove the VP AP A dapt S V front cover and attach the humidifier docking st ation and water chamber. No other accessories are required for use. Please refer to the HumidAi re 2i User’s Man ual for d ...

  • ResMed Flow Generator VPAP Adapt SV - page 12

    12 HumidAire™ You will need a medium size 21" (52 cm) air tube to connect the VPAP Adapt SV unit to the HumidAire™ (see Figure 7). 1 Make sure both the HumidAire and the VP AP Adapt S V are turned off. Fill the HumidAire with water as described in the humidifier manual. Place the filled water chamber inside th e HumidAire. 2 Gently det ac ...

  • ResMed Flow Generator VPAP Adapt SV - page 13

    13 VP AP Adapt SV System English 5 Connect the mask system to the free end of the long air tubing. The final assembly should look like this: 6 Switch on the HumidAire and the VP AP A dapt SV . ! WARNING Make sure that the powe r cords and plugs are in good condition and the equipment is not damaged. 7 Perform the Air Circuit Learn function (see pag ...

  • ResMed Flow Generator VPAP Adapt SV - page 14

    14 ResLink™ The Re sLink is a device that records da ta on to a SmartMedia™ card during your therapy. The card can be returned to your clinician so th at they can observe your progress and adjust flow generator parameters if nece ssary. The ResLink connects to the back of VP AP A dapt S V as shown. For full details on usin g the Re sLink, pleas ...

  • ResMed Flow Generator VPAP Adapt SV - page 15

    15 VP AP Adapt SV System English Using DC Power to R un the VPAP Adapt SV The VPAP Adapt SV can be run on DC in case of power failure. ResMed supplies separately a converter that allows a 12 V or 24V batte ry to be connected to the DC input of the VPAP Adapt SV. Wh en using a DC power supply, the p ower is turned on and off via the switch on the co ...

  • ResMed Flow Generator VPAP Adapt SV - page 16

    16 Using the VPAP Adapt SV ! WARNING Before starting therapy with a new mask, select the correct mask type in the Settings menu (page 21). When adding or removing a new component, such as a mask, humi difier, or anti-bacterial fi lter, perfo rm the Air Circuit Learn function (page 21). Starting Treatment When the system is fully assembled press the ...

  • ResMed Flow Generator VPAP Adapt SV - page 17

    17 Using the VP AP Adapt SV English Control Panel Functions The control panel of the VP AP A dap t S V has a keypad that allows you to: • start or stop treatment • adjust set tings on the flow generator • mute or stop alarms • vie w treatment dat a. Figure 13: The VP AP Adapt SV control pane l LCD screen Displays information about fl ow gen ...

  • ResMed Flow Generator VPAP Adapt SV - page 18

    18 Left key (gree n) • Enters a menu. • Confirms and applie s set tings. • P erfor ms the function ind icated b y the text a bov e it in the LCD (eg, ’enter’ , ’c hange’). Right key (red) • P erforms the function indicated by the te xt abov e it in the LCD (eg, ’e xit’). • Cancels operations. • Extended hold: exits to the to ...

  • ResMed Flow Generator VPAP Adapt SV - page 19

    19 Using the VP AP Adapt SV English Example You wish to change the setting for the mask you are using. Figure 14: Usi ng the keys to adju st settings MASK: FULL F ACE change  exit LEARN CIRCUIT? yes  exit SMARTST OP: OFF change  exit SOUND LEVEL: LOW change  exit (ASV MODE) VP AP Adapt SV menu SETTINGS enter  exit RESUL TS enter  ...

  • ResMed Flow Generator VPAP Adapt SV - page 20

    20 Menus The options available in the Settings menu will de pend on the therapy you are receiving (either ASV or CPAP ), as set by your clinician. Figure 15: Navigating the Patient men u. SETTINGS enter  exit VP AP ADAPT SV menu RESUL TS enter  exit SERVICING enter  exit RAMP: 20min change  exit MASK: FULL F ACE change  exit LEARN CI ...

  • ResMed Flow Generator VPAP Adapt SV - page 21

    21 Using the VP AP Adapt SV English Settings Menu You will only see the options relevant to your therapy. In CPAP (Continuous Positive Airway Pressure) mode, the clinician sets a pressure that will be constant throughout treatment. In ASV mode, the VP AP A dapt SV continually measures your breathing and regulat es the air pressure accordingly. Mask ...

  • ResMed Flow Generator VPAP Adapt SV - page 22

    22 7 . When the function is complete, a screen will appear sa ying that the circuit has been learned successfu lly . T he next menu item is then display ed. Commence treatment as described in “St arting T reatment” on page 1 6. If there is a problem with the circuit a screen saying ’Invalid circu it’ will be displayed inste ad of the progre ...

  • ResMed Flow Generator VPAP Adapt SV - page 23

    23 Using the VP AP Adapt SV English Mask-fit Gives a rating of how well the mask was fitting during your last session on the VP AP A dapt SV. The stars indicate how good the fit was (more stars = be tter fit). Average Pressure Provides info rmation on the average air pre ssure during yo ur most recent session on the VP AP A dapt S V . Usage Tells y ...

  • ResMed Flow Generator VPAP Adapt SV - page 24

    24 Servicing Menu This menu di splays the following information: • SN—the f low generator’ s serial number • PCB—the flow generator’ s PCB num ber (PCB—printed circuit board) • SW—the version of sof t ware currently inst alled on the flow generator . Treatment Screens There are th ree Treatment screen s displayed i n the VPAP Adap ...

  • ResMed Flow Generator VPAP Adapt SV - page 25

    25 Alarms English Alarms The VP AP Adapt SV unit is fitte d with alarms to alert you to changes that will affect your treatment. Th e most common reason for an alarm to s ound is because the system has not been properly assembl ed. Check that the air tubing and pressure sensor tube have be en proper ly attache d to the flow generator and mask (and ...

  • ResMed Flow Generator VPAP Adapt SV - page 26

    26 Will clear when: • the low pressure condition is fixed • treatment is stopped by pressing the Star t/St op key . Power Fail Alarm Trigge red by: • pow e r failure • mac hine is disconnected or switc hed of f while delivering treatment. If a power fail alarm sounds: • the flow generator stops deliv ering air pressure. Will stop: • whe ...

  • ResMed Flow Generator VPAP Adapt SV - page 27

    27 Alarms English High Leak Alert Triggered by: • high mask leak (greater than 30 L/min) f or more than 20 seconds • the mask being remo ved while SmartStop is disabled. If a high leak alert sounds: • adjust the mask to minimiz e leak • c hec k that the air tube and sensor t ube are connecte d properly . Will stop when: • the mask leak ha ...

  • ResMed Flow Generator VPAP Adapt SV - page 28

    28 Will stop when: • bloc kage is remov ed. Cleaning and Maintenance You should regul arly carry out th e clea ning and maintenance described in this manual. ! CAUTION Do not wash the pre ssure senso r tube. If fluid enters the pressure sensor tube allow it to dry comple tely by hang ing it in a clean place out of direct sunlight. If th e pressur ...

  • ResMed Flow Generator VPAP Adapt SV - page 29

    29 Cleaning and Maintenance English Figure 19: Air tubin g assembly removed for cleaning ! CAUTION • Do not use bleac h, c hlorine-, alcohol-, or aromatic-b ased solutions (including all scented oils), moisturisi ng or antibacterial soap s to clean the cushion, mask, air tubing, or the VP AP Ad apt S V . These solutions may cause hardening and re ...

  • ResMed Flow Generator VPAP Adapt SV - page 30

    30 Air Filter Inspect the air filter every month and check if it is blocked by dirt or contains holes. The filter should be replace d every six months, or more often if the flow generator is operating in a dusty environment. ! WARNING Do not wash the air fil ter once it has become soil ed. The air filter is not washable or reusab le. Fitting th e f ...

  • ResMed Flow Generator VPAP Adapt SV - page 31

    31 T roubleshooting English Applicable ResMed warranty details are pr ovi ded with the device at the time of original supply. Of course, as with all electrical devices, if any irregularit y becomes apparent, you should exercise caution and have the device inspected by an authorized ResMed service center. If you feel the uni t is not performin g pr ...

  • ResMed Flow Generator VPAP Adapt SV - page 32

    32 Glossary Alarms When more than one alarm is activ ated at the one time, the highest priorit y alarm will be displa yed. Backlight The bac klight t urns of f if no but ton has been pressed for two minutes. It can also be kept p ermanently on by selecting On in the B ac klight screen of the Options menu. bpm breaths per mi nute L/min liters per mi ...

  • ResMed Flow Generator VPAP Adapt SV - page 33

    33 T e chnical Specifications English SmartStop A f eat ure that stops the VP AP Adapt S V automatically when y ou remo ve t he mask. Y ou can t urn SmartStop on or of f in the Set tings menu. Tidal volume The v olume of air , in liters, breathed in or out in one breath. Technical Specifications Dimensions (H x W x D) 5.6" x 9.8" x 11.5&q ...

  • ResMed Flow Generator VPAP Adapt SV - page 34

    34 Alarm specifica tions • High maximum mask pr essur e alar m Stops air deliv er y at 30 c m H 2 O >700 msec. • High mean mask pressur e alar m Stops air deliver y when mean mask pressure e xceeds 22 cm H 2 O ( aver a g e d ov er approx 1 minute). • Low mask pr essure alar m Mask pressure is less than 3 cm H 2 O f or longer than 5 seconds ...

  • ResMed Flow Generator VPAP Adapt SV - page 35

    35 T e chnical Specifications English Display ed values Maximum low impeda nce flow at stated pressures Glossary of symbols Read the relevant operating manual before using Type CF equip ment AC switch Class II Drip proof Va l u e Ra ng e Accuracy Display Re s o l u t io n Pro ximal cuff P ressure -3 to 40 cm H 2 O 0.5 cm H 2 O 0.1 cm H 2 O Mass flo ...

  • ResMed Flow Generator VPAP Adapt SV - page 36

    36 Guidance and Manufacturer’s De claratio n — Electr omagne tic Emissions and Immunity Guidance and manuf acturer’ s declaration – electr omagnetic emi ssions The VP AP A dapt SV is intend ed for use in the electromagnet ic environment specified belo w . The customer or the user of the VP AP A dapt SV sh ould assure that it is used in suc ...

  • ResMed Flow Generator VPAP Adapt SV - page 37

    37 T e chnical Specifications English (Continued next page) Guidance and manuf acturer’ s declaration – electr omagnetic immu nity The VP AP A dapt SV is intend ed for use in the electromagnet ic environment specified belo w . The customer or the user of the VP AP A dapt SV sh ould assure that it is used in suc h an environment. Immunity t est ...

  • ResMed Flow Generator VPAP Adapt SV - page 38

    38 Guidance and ma nuf actur er ’ s de claration – electromagnetic immunity (continued) The VP AP A dapt SV is intended for use in the electromagnetic en vironment specified belo w . The customer or the user of the VP AP Adapt S V should assure that it is used in suc h an environment. Immunity tes t IEC6060 1 -1 - 2 test lev e l Complianc e lev ...

  • ResMed Flow Generator VPAP Adapt SV - page 39

    39 T e chnical Specifications English Recommended separation distances between portable and mobile RF communic ations equipment and the VP AP A dapt SV The VP AP A dapt SV is in tended for use in an en vironment in whic h radiated RF disturbances are controlle d. The custom er or the user of the VP AP Adapt SV can help pre vent electromagne tic int ...

  • ResMed Flow Generator VPAP Adapt SV - page 40

    40 Limited Warranty ResMed warrants that your ResMed product shall be free from defe cts in material and workmanship for the period sp ecified below from the date of purchas e by the initi al consumer. This warranty is n ot transferabl e. Not e: Some models are not avai lable in all regions. If the product fails under conditions o f normal use, Res ...

  • ResMed Flow Generator VPAP Adapt SV - page 41

    41 Index English Index A air circuit learn function 21 air filter 30 hyp oallergenic 30 air pressure monitoring 23 air t ubing at t ac hing to mask 9 cleaning 28 connecti ng 9 air t ubing sy stem 6 , 7 , 9 alarm flow bloc k ed 27 high leak 27 high pressure 26 lo w mask pressure 25 po wer f ail 26 alarm LED 18 anti-bacterial fil ter 14 B bac klight ...

  • ResMed Flow Generator VPAP Adapt SV - page 42

    42 standby mode 16 star ting treatment 16 stopping treatm ent 16 Supplemental o xygen 15 S ystem Error 7 message 32 T T reatment screens 24 U user/o wner responsibility 3 V VP AP A dapt SV usage 23 VP AP A dapt S V Components 6 228234 VPAP Adapt SV User Guide USA Eng .book Page 42 M onday, October 13, 2008 2:15 PM ...

Производитель ResMed Категория Portable Generator

Документы, которые мы получаем от производителя устройства ResMed Flow Generator VPAP Adapt SV мы можем разделить на несколько групп. Это в частности:
- технические чертежи ResMed
- инструкции обслуживания Flow Generator VPAP Adapt SV
- паспорта изделия ResMed
- информационные брошюры
- энергетические этикетки ResMed Flow Generator VPAP Adapt SV
Все из них важны, однако самую важную информацию с точки зрения пользователя мы найдем в инструкции обслуживания ResMed Flow Generator VPAP Adapt SV.

Группа документов, определяемая как инструкции обслуживания, делится также на более подробные типы, такие как: Инструкции монтажа ResMed Flow Generator VPAP Adapt SV, инструкции обслуживания, короткие инструкции или инструкции пользователя ResMed Flow Generator VPAP Adapt SV. В зависимости от потребностей, Вам необходимо поискать требуемый документ. На нашем сайте Вы можете просмотреть самую популярную инструкцию использования изделия ResMed Flow Generator VPAP Adapt SV.

Похожие инструкции обслуживания

Полная инструкция обслуживания устройства ResMed Flow Generator VPAP Adapt SV, как должна выглядеть?
Инструкция обслуживания, определяемая также как пособие пользователя, или просто "руководство" - это технический документ, цель которого заключается в использовании ResMed Flow Generator VPAP Adapt SV пользователями. Инструкции пишет, как правило технический писатель, языком, доступным для всех пользователей ResMed Flow Generator VPAP Adapt SV.

Полная инструкция обслуживания ResMed, должна заключать несколько основных элементов. Часть из них менее важная, как например: обложка / титульный лист или авторские страницы. Однако остальная часть, должна дать нам важную с точки зрения пользователя информацию.

1. Вступление и рекомендации, как пользоваться инструкцией ResMed Flow Generator VPAP Adapt SV - В начале каждой инструкции, необходимо найти указания, как пользоваться данным пособием. Здесь должна находится информация, касающаяся местонахождения содержания ResMed Flow Generator VPAP Adapt SV, FAQ и самых распространенных проблем - то есть мест, которые чаще всего ищут пользователи в каждой инструкции обслуживания
2. Содержание - индекс всех советов, касающихся ResMed Flow Generator VPAP Adapt SV, которое найдем в данном документе
3. Советы по использованию основных функций устройства ResMed Flow Generator VPAP Adapt SV - которые должны облегчить нам первые шаги во время использования ResMed Flow Generator VPAP Adapt SV
4. Troubleshooting - систематизированный ряд действия, который поможет нам диагностировать а в дальнейшем очередность решения важнейших проблем ResMed Flow Generator VPAP Adapt SV
5. FAQ - чаще всего задаваемые вопросы
6. Контактные данные Информация о том, где искать контактные данные производителя / сервисного центра ResMed Flow Generator VPAP Adapt SV в данной стране, если самостоятельно не получится решить проблему.

У вас вопрос, касающийся ResMed Flow Generator VPAP Adapt SV?

Воспользуйтесь формуляром, находящимся ниже

Если с помощью найденной инструкции Вы не решили свою проблему с ResMed Flow Generator VPAP Adapt SV, задайте вопрос, заполнив следующий формуляр. Если у какого то из пользователей была похожая проблема с ResMed Flow Generator VPAP Adapt SV со всей вероятностью он захочет поделиться методом ее решения.

Перепишите текст с картинки

Комментарии (0)