Bedienungsanleitung Omron Healthcare HEM-907

26 Seiten 0.61 mb
Download

Zur Seite of 26

Summary
  • Omron Healthcare HEM-907 - page 1

    1 INS TR U CTION MAN U AL OMR ON Digital Automatic Blood P ressure Mon itor Model HEM - 9 0 7 HID E P - SE T 1 00 AU T O 1 40 1 80 220 260 280 MOD E A VG. M AN U . CHE CK SI N G LE O N / OF F A V G . / 1 s t / 2 n d / 3 r d D E F L A T I O N STA RT STO P mm Hg SYS mmHg DIA BPM PU L SE A RT . INDEX M A X Tha nk yo u very much for pu rchas ing the OM ...

  • Omron Healthcare HEM-907 - page 2

    2 Intended Us e Medical P urpose This i s a medical de vice that meas ures the bra chial blood pressu re in a non-inv asiv e mann er . Intended Us er Legally c er ti ed : such as d octor , nu rse and med ic al e xpert. Pa tien t Popu lation It is us ed on adult p atien ts only . En vir onment The inst rument is used in physicians ’ of  ce s ...

  • Omron Healthcare HEM-907 - page 3

    3 E x emptions OM RON will not be ar any responsibilitie s on t he following mat ters. When a p roblem or d amage occu rs c au sed by th e main ten ance an d/or r epai r c on- du cted b y a person oth er tha n OM RON or th e deale r speci ed b y OMR ON The prob lem or da mage o f O MRON pr oduct caused b y the p roduct of o ther ma nu- fact urer ...

  • Omron Healthcare HEM-907 - page 4

    4 Names and F unctions of the P ar ts HIDE P - SET 10 0 AUTO 14 0 18 0 220 260 280 MODE A VG. MANU. CHECK SINGLE ON/OFF A V G . / 1 s t / 2 n d / 3 r d D E F L A T I O N START ST OP mmHg SYS mmHg DIA BPM PULSE AC DC8V 1 . L CD display 2. HID E ( non- display ) Button 3. DC jack 4 . P- SET (press ure setting ) V olume 5. MO DE Selector 1 0 . ST OP B ...

  • Omron Healthcare HEM-907 - page 5

    5 Ba t tery lev el Displays for the usable level. Displays for the l ow level. Displays for the unusable level. Names and F unctions of the P ar ts Displa y HIDE BPM mmHg PULSE SYS mmHg DIA AC Systolic blood pressure Dias tolic b lood pressure Con ten ts when the A V G. Mode is selected A VG. : Mean value 1 st : First me asurement 2nd : Sec ond mea ...

  • Omron Healthcare HEM-907 - page 6

    6 Names and F unctions of the P ar ts Explanation of F unct ions ( 1 ) P ressure sett ing function P - SET In ation c an be set by AUT O (automatic set ting ) or th e target value. A UTO ( automatic setting) : In the SIN GLE (single measurement ), A VG. ( av erage ), and MA NU. ( auscult ation ) Mo des, the monitor est imates the systolic blo od ...

  • Omron Healthcare HEM-907 - page 7

    7 Names and F unctions of the P ar ts ( 4) Set ting mo de ch ange func ti on Y o u can set th e number of times to m easure, the waiting ti me until the st ar t of measurem ent, and the inter val bet ween measurements. It em s to se t Set va lu e F 1 Number of measurem ents 2 times, 3 t imes F2 W aitin g time until t he star t of m easurement 0 sec ...

  • Omron Healthcare HEM-907 - page 8

    8 Notes on S afet y The war ning signs and the sample ic ons show n here are listed for you to use t he product s afely and cor rect ly as well as to prevent the risk and the damag e to you and others f rom happening. The ic ons and meanings are as follow. Warning sign Contents Warning Indicates mat ter s in which deat h or sever bodily damage may ...

  • Omron Healthcare HEM-907 - page 9

    9 Notes on S afet y Caution When you are n ot going to use t he unit for a long p eri od of time, be sure to remove the AC adapter from the electric outlet. Y ou may suf fer elect ric shoc k, or electr ic leak or  re may arise beca use of deter iorated insulation. Be sure to remove the AC adapter from the elec tr i c outlet w hen installing, rem ...

  • Omron Healthcare HEM-907 - page 10

    10 Notes on S afet y Caution Be sure to use the p ower supply of 23 0 V AC. It may cause  re or electr ic shock. Do not share an elect ric outl et with other unit or elec tric applianc e. It may c ause  re or electric leak. Do not pull the p ower co rd when pulling out the AC adapter fro m the elect ric out let. The power c ord will be disc o ...

  • Omron Healthcare HEM-907 - page 11

    11 Notes on S afet y Maintenance Be sure to inspec t the unit on regular basis. Check ac curac y on a 1 year inter val. If the unit has not be en used for a while, be sure to c on rm t hat the unit oper ates nor mally and safely before us e Durabilit y The durabilit y of thi s unit is  ve y ear s. (Ar m cuf f is a c onsumable. ) Me asur es t ...

  • Omron Healthcare HEM-907 - page 12

    12 Components of the P roduct Main unit Access ories HIDE P - S E T 10 0 A U TO 14 0 1 80 2 2 0 260 280 MODE AVG . MA N U . C HE C K S IN G L E ON/OFF A V G . / 1 s t / 2 n d / 3 r d D E F L A T I O N START STOP Medium size ar m cuf f ( with built- in air bag ) Model : HE M -9CMC Ar t.no. : 492864 0 - 0 Applic able ar m circumferenc e: 22 to 32 cm ...

  • Omron Healthcare HEM-907 - page 13

    13 Sel ec t th e ar m cu f f ac cor din g to t he ar m c ircu mf ere nce o f th e pat ie nt . Ar m circumferenc e Name of the ar m cuf f 1 7 - 22 cm 22 - 32 c m 32 - 42 cm Small size adult arm c uf f ( option ) Adult arm c uf f Larg e size adult arm cu f f ( opti on ) Be sure to use the ar m cuf f suitable for the size of the pat ient ’ s arm. If ...

  • Omron Healthcare HEM-907 - page 14

    14 Ho w to Apply the Ar m C uf f (c o n t .) R A N G E ART . RA N G E A R T. INDE X 22 ~ 32cm ( 9" ~ 13" ) R EF HEM - CR 19 Adju st t he l evel of t he a rm c uf f t o th e leve l of th e he ar t. Keep the level of the arm c uf f at the s ame level as the hear t dur ing the measure - ment. Ho w to use the P ow er So urce (A C ad apt er ex ...

  • Omron Healthcare HEM-907 - page 15

    15 Ho w to Measure B lood P ress ure Warning If any abnormal mat ter oc curs dur ing the me a- surement such as t he in ation does not sto p , remove the arm c uf f or pull the air tub e from t he main unit.  Y ou may suf fer per ipheral neuropathy . Requests from OMRON  Be sure that the patient should n ot touch the unit.  Be sure to m ...

  • Omron Healthcare HEM-907 - page 16

    16 Ho w to Meas ure Blood P ress ure (in SI NGLE M ode) Wrap t h e ar m cuf f ov e r th e patient’s ar m. 1. Set t he M OD E Sel ec to r to “SINGLE” . 3. ON/OFF A V G . / 1 s t / 2 n d / 3 r d D E F L A T I O N START STOP HIDE P - SET 100 AUTO 140 18 0 220 260 280 MODE AVG. MANU. CHECK SINGLE Push t he ST ART B ut to n to st ar t the meas ure ...

  • Omron Healthcare HEM-907 - page 17

    17 Wrap t h e ar m cuf f ov e r th e patient’s ar m. 1. Set t he M OD E Sel ec to r to “AV G . ”. 3. ON/OFF A V G . / 1 s t / 2 n d / 3 r d D E F L A T I O N START STOP HIDE P - SET 100 AUTO 140 18 0 220 260 280 MODE AVG. MANU. CHECK SINGLE Push t he ST ART B ut to n to st ar t the meas urement. Set the unit to th e star t t ime of the  rs ...

  • Omron Healthcare HEM-907 - page 18

    18 Ho w to Meas ure Blood P ress ure (in MA NU. Mode) Wrap t h e ar m cuf f ov e r th e patient’s ar m. 1. Push t he O N/OFF (p owe r) But t on to t ur n on t he pow er . 3. ON/OFF A V G . / 1 s t / 2 n d / 3 r d D E F L A T I O N START STOP HIDE P - SET 100 AUTO 140 18 0 220 260 280 MODE AVG. MANU. CHECK SINGLE Set t he P- S ET (pr es sure s et- ...

  • Omron Healthcare HEM-907 - page 19

    19 Ho w to Meas ure Blood P ress ure (in CHECK M od e) Disp lay of pre ssure c an b e con r me d by the CH ECK Mode (yea rly p roc edur e ). Wha t you nee d to pre pare ( 1 ) Well- adjusted reference pre ssure monitor (inc luding rubber ball), (2 ) T -shaped tub e, (3 ) T wo r ubber tubes, and (4 ) something in th e cylindr ical shape to w rap t ...

  • Omron Healthcare HEM-907 - page 20

    20 Installation and R eplacement of Ba t t er y Pa ck Rem ove the b at t er y c over on t he b ack o f th e main unit. Installation: T o install the bat ter y pack, c onnec t the bat ter y pack c onnec tor to the c onnecto r in the bat ter y cover . Replacement: Remove the batter y pac k fro m the co nnector and replac e wit h a new one. In sta ll ...

  • Omron Healthcare HEM-907 - page 21

    21 Cle an th e bl ood p re ssur e mon it or ,t he cu f f an d th e ca bl es usi ng a cl ot h da mp ene d wi th wa te r ,dilu te d disinfectant alcohol or diluted mild detergent. T o disinfect, wipe t he unit with a s of t clot h squeezed well af ter moistened w ith the foll owing disinfect ant. Benzalkonium c hlor ide 0.01 - 0.2 w/ v% Chlor hexizin ...

  • Omron Healthcare HEM-907 - page 22

    22 L ist o f Error C odes Erro r cod e Cause How to co rre ct Er1 In at ion erro r When the pre ssure does n ot ex ce ed 1 5 mmHg af ter oper ating a pump for 15 se conds from t he star t of in ati on When th e in ation does not re ach the set pressure of th e arm cu f f within the sp eci ed time after starting the in ation • • I ...

  • Omron Healthcare HEM-907 - page 23

    23 If a t roubl e occur s whil e using th e unit , pl ease che ck th e followin g. Proble m Wha t to inspe ct How to co rre ct The unit in ates to abnor mally high (low) pressure. Is the ar m cuf f w rapped c or rec tly ? Wrap the arm c uf f c orrec tly , and measure again. Is the patient m oving his/ her arm or b ody during in ati on ? Ask t ...

  • Omron Healthcare HEM-907 - page 24

    24 Speci  cati on s Name : OM RON Di gital Automat ic Bloo d Pressure M onitor H EM - 907 Model : HEM- 907 - E / HEM- 907 - E7 Display : Digita l display Measurement : Osc illometr ic method Measurement Range : Pressure; 0 to 29 9 mmHg Pulse rate: 30 to 199 beat s/min Acc uracy : Pressure; Within +/- 3 mmHg Pulse rate; Within +/- 5% of reading I ...

  • Omron Healthcare HEM-907 - page 25

    25 Manufacturer OMRON HEAL THC ARE Co., L td. 53, K unots ubo , T era do - cho , Muk o , K y ot o, 6 1 7 - 00 02 JAP A N EU- representative OMRON HEAL THCARE EUR OPE B. V . Scorpiu s 33, 2 1 3 2 LR Hoo fddo r p , THE NETHE RL A NDS w w w .omr o n - h e al th c a re.c om Production facilit y OMR ON HEAL THC ARE Co. , Ltd. Mie , JAP A N Subs idiar y ...

  • Omron Healthcare HEM-907 - page 26

    26 Englis h Im po r t ant i nf or ma ti on r eg ar di ng E le ct ro M ag net ic C om pa ti bi li ty ( E MC ) Wi th t he in cre as ed nu mbe r of el ec tr oni c dev ic es su ch as P C. s and m obi le ( c ellular) telephones, medic al devices in use may be susc eptible to electromagnetic interference from other de vices. Electromagnetic interfer- enc ...

Produzent Omron Healthcare Kategorie Blood Pressure Monitor

Dokumente, die wir vom Produzenten des Geräts Omron Healthcare HEM-907 erhalten, können wir in mehrere Gruppen teilen. Unteranderem in:
- technische Zeichnungen Omron Healthcare
- Bedienungsanleitungen HEM-907
- Produktkarten Omron Healthcare
- Informationsbroschüren
- oder Energieetiketten Omron Healthcare HEM-907
Jede von ihnen ist wichtig, jedoch finden wir die wichtigsten Informationen für den Nutzer des Geräts in der Bedienungsanleitung Omron Healthcare HEM-907.

Die Dokumentengruppe, die als Bedienungsanleitungen bezeichnet wird, wird ebenfalls in detaillierte Arten geteilt, solche wie: Montageanleitungen Omron Healthcare HEM-907, Wartungsanleitungen, Kurzanleitungen oder Benutzeranleitungen Omron Healthcare HEM-907. Abhängig vom Bedarf, sollten Sie das Dokument finden, das Sie brauchen. In unserem Service können Sie sich die populärste Bedienungsanleitung des Produkts Omron Healthcare HEM-907 ansehen.

Ähnliche Bedienungsanleitungen

Die komplette Bedienungsanleitung des Geräts Omron Healthcare HEM-907, wie sollte sie aussehen?
Die Bedienungsanleitung, auch bezeichnet als Benutzerhandbuch, oder einfach nur „Anleitung”, ist ein technisches Dokument, das dem Benutzer bei der Nutzung von Omron Healthcare HEM-907 hilfreich sein soll. Die Bedienungsanleitungen werden in der Regel von technischen Schriftstellern geschrieben, aber in einer Sprache, die für alle Nutzer von Omron Healthcare HEM-907 verständlich ist.

Eine gänzliche Bedienungsanleitung von Omron Healthcare sollte einige Grundelemente enthalten. Ein Teil von ihnen ist nicht so wichtig, wie z.B.: die Titelseite oder Autorenseiten. Die restlichen von ihnen jedoch, sollten Informationen liefern, die für den Nutzer von enormer Wichtigkeit sind.

1. Einführung und Hinweise, wie man sich in einer Bedienungsanleitung von Omron Healthcare HEM-907 bewegt - Am Anfang jeder Bedienungsanleitung sollten wir Hinweise bezüglich der Nutzungsart eines bestimmten Ratgebers finden. In ihr sollten sich Informationen über die Lokalisierung des Inhaltsverzeichnisses von Omron Healthcare HEM-907 befinden, FAQ oder über oft auftretende Probleme – also Stellen, die von den Benutzern in jeder Bedienungsanleitung am meisten gesucht werden
2. Inhaltsverzeichnis - Index aller Ratschläge bezüglich Omron Healthcare HEM-907, die wir im aktuellen Dokument finden
3. Ratschläge zur Nutzung der Grundfunktionen des Geräts Omron Healthcare HEM-907 - die uns die ersten Schritte während der Nutzung von Omron Healthcare HEM-907 erleichtern sollten
4. Troubleshooting - geordneter Tätigkeitslauf, der uns bei der Diagnose und als nächstes bei der Lösung wichtiger Probleme mit Omron Healthcare HEM-907 hilft
5. FAQ - häufig gestellte Fragen
6. Kontaktdaten Informationen darüber, wo man Kontakt zum Produzenten / Service von Omron Healthcare HEM-907 im bestimmten Land suchen kann, wenn es nicht gelingt, das Problem selbst zu lösen.

Haben Sie eine Frage bezüglich Omron Healthcare HEM-907?

Nutzen Sie das untere Formular

Wenn Sie mit Hilfe der gefundenen Bedienungsanleitung Ihr Problem mit Omron Healthcare HEM-907 nicht gelöst haben, stellen Sie eine Frage, indem Sie das untere Formular nutzen. Wenn einer der Nutzer ein ähnliches Problem mit Omron Healthcare HEM-907 hatte, ist es möglich, dass er mit Ihnen die Lösung teilen möchte.

Text vom Bild übertragen

Kommentare (0)