Bedienungsanleitung Fluke 9140

52 Seiten 0.51 mb
Download

Zur Seite of 52

Summary
  • Fluke 9140 - page 1

    9140 Dry-well Calibrator User’ s Guide Rev. 5B1702 Hart Scientific ...

  • Fluke 9140 - page 2

    Rev. 5B1702 Fluke Corporation, Hart Scientific Division 799 E. Utah Valley Drive • American Fork, UT 84003-9775 • USA Phone: +1.801.763.1600 • Telefax: +1.801.763.1010 E-mail: support@hartscientific.com www.hartscientific.com Subject to change without notice. • Copyright © 2005 • Printed in USA ...

  • Fluke 9140 - page 3

    T able of Contents 1 Bef ore Y ou Start . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1.1 Symbols Used . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1.2 Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . 2 1.2.1 W ARNINGS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 1.2.2 CA UTIONS ...

  • Fluke 9140 - page 4

    7.2 T emperature Set-point . . . . . . . . . . . . . . . . . . . . . . . . 21 7.2.1 Programmable Set-points . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 7.2.2 Set-point V alue . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 7.2.3 T emperature Scale Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 7 ...

  • Fluke 9140 - page 5

    10 Calibration Procedur e . . . . . . . . . . . . . . . . . . . . . . 43 10.1 Calibration Points . . . . . . . . . . . . . . . . . . . . . . . . . . 43 10.2 Calibration Procedure . . . . . . . . . . . . . . . . . . . . . . . . 43 10.2.1 Compute DEL T A: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43 10.2.2 Compute R 0 & ALPH ...

  • Fluke 9140 - page 6

    iv Figures Figure 1 9140 Back Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Figure 2 9140 Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Figure 3 Inserts av ailable for the 9140 block assembly . . . . . . . . . . . . . . 15 Figure 4 Controller Operation Flowchart . . . . . . . . . . . . . . . . . . . . ...

  • Fluke 9140 - page 7

    v T ables T able 1 International Electrical Symbols . . . . . . . . . . . . . . . . . . . . . 1 T able 2 9140 controller communications commands . . . . . . . . . . . . . . . 36 T able 3 9140 controller communications commands continued . . . . . . . . . 37 ...

  • Fluke 9140 - page 8

    1 Bef ore Y ou Start 1.1 Symbols Used T able 1 lists the symbol s used on the instrument or i n this manual and the meaning of each symbol. Symbol Description AC ( Alternating Current) AC- DC Batter y Complies with European Uni on directiv es DC Double Insulated Electric Shock Fuse PE Ground Hot Surf ace (Bur n Hazard) Read the User’ s Manual (Im ...

  • Fluke 9140 - page 9

    Symbol Description Canadian Standards Associati on O VERV OL T AGE (Installat ion) CA TEGORY II, P ollution Deg ree 2 per IEC1010-1 re - f ers to the le vel of Impulse Wit hstand V oltage protect ion provided. Equipment of O VERV OL T AGE CA TEGOR Y II is energy-consuming equipment to be suppli ed from the fix ed installat ion. Examples include hou ...

  • Fluke 9140 - page 10

    ing power such as storage in a low humidity temperature chamber operat - ing at 50°C for 4 hours or more. • DO NO T use this instrument for any application other than calibration work. The instrument was designed for temperature calibration. Any other use of the instrument may cause unknown hazards to the user. • Completely unattended operatio ...

  • Fluke 9140 - page 11

    ELECTRICAL SHOCK • These guidelines must be followed to ensure that the safety mechanisms in this instrument will operate properly . This instrument must be plugged into a 115 V A C, 60Hz (230 V A C, 50Hz optional), AC only electric outlet. The power cord of the instrument is equipped with a three-pronged grounding plug for your protection ag ain ...

  • Fluke 9140 - page 12

    • The instrument and any thermometer probes used with it are sensitive in - struments that can be easily damaged. Always handle these devices with care. DO NO T allo w them to be dropped, struck, stressed, or o v erheated. • The Factory Reset Sequence (see Section 12.1, Troubleshooting) should be performed only by authorized personnel if no oth ...

  • Fluke 9140 - page 13

    Science Park Eindhov en 5108 5692 EC Son NETHERLANDS Phone: +31-402-675300 T elefax: +31-402-675321 E-mail: ServiceDesk@fluke.nl Fluke Int'l Corporation Service Center - Instrimpex Room 2301 Sciteck T ower 22 Jianguomenwai Dajie Chao Y ang District Beijing 100004, PRC CHIN A Phone: +86-10-6-512-3436 T elefax: +86-10-6-512-3437 E-mail: xingye.h ...

  • Fluke 9140 - page 14

    2 Intr oduction The Hart Scientif ic 9140 Mid-Range Field Cal ibrator may be used as a portable instrument or bench top temperat ure calibrator for calibrating t hermocouple and R TD temperature probes. The 9140 is small enough t o use in the f ield, and accurate enough to use in the lab . The instrument features: • Rapid heating and cooling • ...

  • Fluke 9140 - page 15

    3 Specifications and En vir onmental Conditions 3.1 Specifications The following table lists the specifications for this instru ment. Accuracy speci - fications are applicable for a one-year calibration interval. In line with nor mal prudent metrology practices, Hart Scientif ic recommends a short-cycle interv al of six months for ne w units during ...

  • Fluke 9140 - page 16

    operated in an e xcessi vel y dusty or dirty en vironment. Maintenance and clean - ing recommendations can be found in the M aintenance Section of this manual. The instrument operates safel y under the follo wing conditions: • temperature range: 5–50°C (41–122°F) • ambient relati ve humidity: 15–50% • pressure: 75kPa–106kPa • main ...

  • Fluke 9140 - page 17

    4 Quic k Start 4.1 Unpac king Unpack the dry-well carefully and inspect it for a ny damage that may ha ve oc - curred during shipment. If there i s shipping damage, notify the carrier immediately . V erify that the follo wing components are present: • 9140 Dry-well • Insert • Insert Remov al T ool • Power Cord • Serial Cable • User&apos ...

  • Fluke 9140 - page 18

    dry-well will t urn on and begin to heat to the pre viously programmed tempera - ture set-point. The front panel LED display will indi cate the actual dry-well temperature. 4.4 Setting the T emperature Section 7.2 explains in det ail ho w to set the temperat ure set-point on the cali - brator using the front panel ke y s. The procedure is summarize ...

  • Fluke 9140 - page 19

    5 P ar ts and Contr ols The user should become familiar with the d ry-well calibrator and its parts: 5.1 Rear P anel Figure 1 on page 13. Pow er Co rd - At the rear of the calibrato r is t he remov able power cord inlet that plugs into an IEC grounded socket . Po wer Switch - The po wer switch is l ocated on the po wer entry module (PEM). The PEM a ...

  • Fluke 9140 - page 20

    around the two corners of the calibrator are pro vided for airflo w . The area around the calibrator must be kept clear to allow adequate ventilation. The air - flow is directed upw ard and as a result, can be e xtremely hot. 5.2 Fr ont P anel Figure 2 on page 14. Controller Display - The digital display is an important part of the temperature cont ...

  • Fluke 9140 - page 21

    EXIT – Used to e xit a function and t o skip to the ne xt function. An y changes made to the displayed v alue are ignored. 5.3 Constant T emperature Block Assembly Figure 3 on page 15. 5.3.1 Constant T emperature Block The “Block” is made of aluminum and pro vides a relati vely constant and accu - rate temperature en vironment in wh ich the s ...

  • Fluke 9140 - page 22

    • Insert D (comparison block): 2 each at 3 mm, 4 mm, and 6 mm 9140 Dry-well Calibrator User’s Guide 16 ...

  • Fluke 9140 - page 23

    6 General Operation 6.1 Calibrator Set-Up Place the calibrator on a flat surface with at least 6 inches of free space around the instrument. Overhead clearance is required. DO NO T place under a cabinet or other structure.Plug the po wer cord into a grounded mai ns outlet. Observ e that the nominal v oltage corresponds t o that indicated on the bac ...

  • Fluke 9140 - page 24

    • Using the same straight slot screwdri ver , move the heater switch to dis - play “230V”. See the rear panel drawing in Figure 1 on page 13. NO TE: If the heater switc h and the fuse holder do not both r ead 230V when complete, the unit will either not heat or only heat at a fraction of its capacity . If not done pr operly , the unit could b ...

  • Fluke 9140 - page 25

    Using the same hole for the reference thermometer and the test probe may have better results. This ho wev er requires switching the probes which tak es more time. One must allo w a few minutes after inserti ng the probes for the tempera - ture to stabilize before making measurements. Because of temperature varia - tions along the l ength of the wel ...

  • Fluke 9140 - page 26

    7 Contr oller Operation This chapter discusses in detail how to operate the dry-well temperature con - troller using the front control panel. Using the front panel k ey-swit ches and LED display the user may monitor the well temperature, set the t emperature set-point in de grees C or F , monitor the heater output po wer , adjust the control - ler ...

  • Fluke 9140 - page 27

    9140 Dry-well Calibrator User’s Guide 22 Figure 4 Controller Operation Flowchar t ...

  • Fluke 9140 - page 28

    100.0 C W ell temperatur e in de gr ees Celsius S Access set-point memory 1. 100. Set-point memory 1, 100°C currently used T o change to another set-point memory press “UP” or “ DO WN”. 4. 300. New set-point memory 4, 300°C Press “SET” to accept the ne w selection and access the set-point value. S Accept selected set-point memory 7.2. ...

  • Fluke 9140 - page 29

    Un= F New units selected 7.3 Scan The scan rate can be set and enabl ed so that when the set-point is changed the dry-well heats or cools at a specified ra te (degrees per minute) until it reaches the ne w set-point. W ith the scan disabled the dry-well heats or cools at the maximum possible rate. 7.3.1 Scan Contr ol The scan is controlle d with th ...

  • Fluke 9140 - page 30

    7.4 T emperature Display Hold The 9140 has a display hol d function which allo ws action of an external swi tch to freeze the displayed temperature and stop the set-point from scanning. This is useful for testing thermal swit ches and cutouts. This section e xplains the functions a vai lable for operating the temperature hol d feature. An example f ...

  • Fluke 9140 - page 31

    sistor . The calib rator measures the volta ge at the red terminal and interprets +5V as open and 0V as closed. 7.4.4 Switch T est Example This section describes a p ossible application for the temperatu re hold feature and how t he instrument is set up and operated. Suppose you hav e a thermal switch which is supposed to open a t about 75°C and c ...

  • Fluke 9140 - page 32

    is. W ith good control stability the percent heating power should not fluctuate more than ± 1% within one minute. The heater power display is accessed i n the secondary menu. Press “SET” and “EXIT” simultaneously and release. The heater power will be displayed as a percentage of full po wer . 100.0 C W ell temperatur e S + E Access heater ...

  • Fluke 9140 - page 33

    The temperature stability of the well and response time depend on the width of the proportional band. If the band is too wide the well temperat ure will de viate excessi vely from the set -point due to v arying external condit ions. This is be - cause the power output chang es v ery little with temperature and the controller cannot respond very wel ...

  • Fluke 9140 - page 34

    7.6 Contr oller Configuration The controller has a number of conf iguration and operati ng options and calibra - tion parameters which are programmable via the front panel. These are ac - cessed from the secondary menu after the proportional band funct ion by pressing “SET”. Pressing “SET” again enters the fi rst of three sets of configu - ...

  • Fluke 9140 - page 35

    7.6.1.2 ALPHA This probe parameter refers to the a verage sensit ivi ty of the probe between 0 and 100°C. The value of this parameter is set at the f actory for best instrument accuracy . 7.6.1.3 DEL T A This probe parameter characterizes the curvature of the resistance-temperatu re relationship of the sensor . The v alue of this parameter is set ...

  • Fluke 9140 - page 36

    4800 b New baud rate Press “SET” to set the baud rate to the ne w va lue or “EXIT” to abort the opera - tion and skip to the next parameter in the menu. 7.8.1 Sample P eriod The sample period is the ne xt parameter in the serial interf ace parameter menu. The sample period is the time period in seconds between temperature measure - ments tr ...

  • Fluke 9140 - page 37

    7.8.1.2 Linefeed The final parameter in the serial interface menu is the linefeed mode. This pa - rameter enables (on) or disables (of f) transmission of a linefeed character (LF , ASCII 10) after transmission of an y carri age-return. The linefeed parameter is indicated by , LF Serial linefeed parameter Press “SET” to access the linefeed param ...

  • Fluke 9140 - page 38

    8 Digital Comm unication Interface The dry-well calibrator is capable of communicating with and being controlled by other equipment through the digital serial interface. W ith a digital interface the instrument may be connected to a comput er or other equipment. Th is allo ws the user to set the set-point temperatur e, monitor the temperature, and ...

  • Fluke 9140 - page 39

    rial interf ace parameters menu are the B A UD rate, the sample rate, the duple x mode, and the linefeed parameter . 8.1.2.1 Baud Rate The baud rate is the first parameter in the menu. The display will prompt with the baud rate parameter by showing “ BAUd ”. Press “SET” to choose to set the baud rate. The current baud rate v alue will then ...

  • Fluke 9140 - page 40

    face commands are discussed in Section . All commands are ASCII character strings terminated with a carriage -return character (CR, ASCII 13). 8.2 Interface Commands The various commands for accessing the ca librator functions via the digital in - terfaces are listed in this section (see Ta b l e 2 ). These commands are used with the RS-232 serial ...

  • Fluke 9140 - page 41

    9140 Dry-well Calibrator User’s Guide 36 Command Description Command Format Command Example Returned Returned Example Acceptable Values Display Temperature Read current set-point s[etpoint] s set: 9999.99 {C or F} set: 150.00 C Set current set-point t o n s[etpoint]=n s=350 Instrument Range Read temperature t[emperature] t t: 9999.9 {C or F} t: 5 ...

  • Fluke 9140 - page 42

    37 8 Digital Communication Interface Interface Commands Command Description Command Format Command Example Returned Returned Example Acceptable Values Set serial linefeed mode: lf[eed]=on/of[f] ON or OFF Set serial li nefeed mode to on lf[eed]=on lf=on Set serial linefeed mode to off lf[eed]=of[f] lf=of Miscellaneous Other Commands Read firmware ve ...

  • Fluke 9140 - page 43

    9 T est Pr obe Calibration For optimum accurac y and stability , allow the calibrator to warm up for 10 minutes after power -up and then allo w adequate stabilization time after reach - ing the set-point temperature. Af ter completing operation of the calibrator, al - low t he well to cool by setting the temperature t o 100°C for one-half hour bef ...

  • Fluke 9140 - page 44

    2. Place the probe to be calibrated, the unit under test ( UUT), in another well. 3. W ith the reference inserted into one well and the probe under test in - serted into a second well, make measurements of each. 4. Swap the locations of the reference probe and probe under test. Allo w plenty of time for thermal settling. 5. Make another set of meas ...

  • Fluke 9140 - page 45

    W ARNING: DO NOT r emove inserts when heating or when the unit is hot. 9.2.3 Stabilization and Accuracy The stabilization time o f the dry-well calibrator w ill depend on the cond itions and temperatures inv olved. T ypically the test well will be stab le to 0.1°C within 5 minutes of reaching the set-point temp erature as indicated by the display ...

  • Fluke 9140 - page 46

    timate stability will be achie ved 10 t o 20 minutes after reaching the set temperature. Inserting a cold probe in to a well will require another period of stabilizin g de - pending on the magnitude of the disturb ance and the required accuracy . For ex - ample, inserting a .25 in ch diameter room temperature prob e into a slee ve at 300°C will ta ...

  • Fluke 9140 - page 47

    10 Calibration Pr ocedure Sometimes the user may want to calibrate the d ry-well to improv e the tempera - ture set-point accuracy . Calibration is done by adjusting the controller probe calibration constants R0 , ALPHA , and DEL T A so th at the t emperature of the dry-well as measured with a st andard th ermometer agrees more closely with the set ...

  • Fluke 9140 - page 48

    C TT TT = ⎡ ⎣ ⎢ ⎤ ⎦ ⎥ − ⎡ ⎣ ⎢ ⎤ ⎦ ⎥ − ⎡ ⎣ ⎢ ⎤ ⎦ ⎥ − ⎡ ⎣ ⎢ ⎤ ⎦ 33 22 100 1 100 100 1 100 ⎥ D TT TT = ⎡ ⎣ ⎢ ⎤ ⎦ ⎥ − ⎡ ⎣ ⎢ ⎤ ⎦ ⎥ − ⎡ ⎣ ⎢ ⎤ ⎦ ⎥ − ⎡ ⎣ ⎢ ⎤ ⎦ 22 11 100 1 100 100 1 100 ⎥ ER R =− 32 FR R =− 21 delta AF BE DE CF = − − T 1-3 - ...

  • Fluke 9140 - page 49

    b . Press SET then use the UP or DOWN keys until the correct numerical setting is displayed. Press SET to accept the new v alue. c. Repeat step b . for ALPHA and DEL T A. 10.2.3 Accuracy & Repeatability 1. Check the accuracy of the dry-well at v arious points ov er the calibration range. 2. If dry-well does not pass specification at all set -po ...

  • Fluke 9140 - page 50

    11 Maintenance • The calibration instrument has been designed with the utmost care. Ease of operation and simplicity of maintenance have been a central theme in the product development. Therefore, with proper care the instrument should require very little maintenance. A void operating the instrument in an oily , wet, dirty , or dusty environment. ...

  • Fluke 9140 - page 51

    12 T r oubleshooting If problems arise while operati ng the 9140, this section provides s ome sugges - tions that may help you sol ve the problem. A wiri ng diagram is also i ncluded. 12.1 T roub leshooting Pr oblems, P ossible Causes, and Solutions In the ev ent that the instrument appears to function abnormally , this section may help to fi nd an ...

  • Fluke 9140 - page 52

    Problem Possible Causes and Solutions The display shows an error code Controller probl em. The error messages signif y the following pr oblems with the controller. Err 1 - a RAM error Err 2 - a NVRAM error Err 3 - a Structure error Err 4 - an ADC setup e rror Err 5 - an ADC ready error Err 6 – a defective contr ol sensor Err 7 – a heater error ...

Produzent Fluke Kategorie Home Safety Product

Dokumente, die wir vom Produzenten des Geräts Fluke 9140 erhalten, können wir in mehrere Gruppen teilen. Unteranderem in:
- technische Zeichnungen Fluke
- Bedienungsanleitungen 9140
- Produktkarten Fluke
- Informationsbroschüren
- oder Energieetiketten Fluke 9140
Jede von ihnen ist wichtig, jedoch finden wir die wichtigsten Informationen für den Nutzer des Geräts in der Bedienungsanleitung Fluke 9140.

Die Dokumentengruppe, die als Bedienungsanleitungen bezeichnet wird, wird ebenfalls in detaillierte Arten geteilt, solche wie: Montageanleitungen Fluke 9140, Wartungsanleitungen, Kurzanleitungen oder Benutzeranleitungen Fluke 9140. Abhängig vom Bedarf, sollten Sie das Dokument finden, das Sie brauchen. In unserem Service können Sie sich die populärste Bedienungsanleitung des Produkts Fluke 9140 ansehen.

Ähnliche Bedienungsanleitungen

Die komplette Bedienungsanleitung des Geräts Fluke 9140, wie sollte sie aussehen?
Die Bedienungsanleitung, auch bezeichnet als Benutzerhandbuch, oder einfach nur „Anleitung”, ist ein technisches Dokument, das dem Benutzer bei der Nutzung von Fluke 9140 hilfreich sein soll. Die Bedienungsanleitungen werden in der Regel von technischen Schriftstellern geschrieben, aber in einer Sprache, die für alle Nutzer von Fluke 9140 verständlich ist.

Eine gänzliche Bedienungsanleitung von Fluke sollte einige Grundelemente enthalten. Ein Teil von ihnen ist nicht so wichtig, wie z.B.: die Titelseite oder Autorenseiten. Die restlichen von ihnen jedoch, sollten Informationen liefern, die für den Nutzer von enormer Wichtigkeit sind.

1. Einführung und Hinweise, wie man sich in einer Bedienungsanleitung von Fluke 9140 bewegt - Am Anfang jeder Bedienungsanleitung sollten wir Hinweise bezüglich der Nutzungsart eines bestimmten Ratgebers finden. In ihr sollten sich Informationen über die Lokalisierung des Inhaltsverzeichnisses von Fluke 9140 befinden, FAQ oder über oft auftretende Probleme – also Stellen, die von den Benutzern in jeder Bedienungsanleitung am meisten gesucht werden
2. Inhaltsverzeichnis - Index aller Ratschläge bezüglich Fluke 9140, die wir im aktuellen Dokument finden
3. Ratschläge zur Nutzung der Grundfunktionen des Geräts Fluke 9140 - die uns die ersten Schritte während der Nutzung von Fluke 9140 erleichtern sollten
4. Troubleshooting - geordneter Tätigkeitslauf, der uns bei der Diagnose und als nächstes bei der Lösung wichtiger Probleme mit Fluke 9140 hilft
5. FAQ - häufig gestellte Fragen
6. Kontaktdaten Informationen darüber, wo man Kontakt zum Produzenten / Service von Fluke 9140 im bestimmten Land suchen kann, wenn es nicht gelingt, das Problem selbst zu lösen.

Haben Sie eine Frage bezüglich Fluke 9140?

Nutzen Sie das untere Formular

Wenn Sie mit Hilfe der gefundenen Bedienungsanleitung Ihr Problem mit Fluke 9140 nicht gelöst haben, stellen Sie eine Frage, indem Sie das untere Formular nutzen. Wenn einer der Nutzer ein ähnliches Problem mit Fluke 9140 hatte, ist es möglich, dass er mit Ihnen die Lösung teilen möchte.

Text vom Bild übertragen

Kommentare (0)