Bedienungsanleitung Altek a 830-KP

25 Seiten 1.03 mb
Download

Zur Seite of 25

Summary
  • Altek a 830-KP - page 1

    Pr ocess Calibrator T echChek™ 830-KP ...

  • Altek a 830-KP - page 2

    1 Y our New T echChek 830-KP Lighten your load … take the T echChek 830-KP to every site. It ’ s like bringing a cartload of test equipment from the shop to the control room or the fi eld. The T echChek 830-KP sources and reads DC like a milliamp or voltage calibrator , simulates and measures T/Cs & R TDs like a temperature calibrator , ge ...

  • Altek a 830-KP - page 3

    3 GENERAL TURN-ON Each time you turn on the T echChek 830-KP it runs through a self check then returns to the most recently selected settings. SOURCE - The three QUIK-CHEK outputs will be the same as previously stored. Each time a different function is selected, the associated three QUIK-CHEK outputs will be recalled. READ - The 830-KP is ready to ...

  • Altek a 830-KP - page 4

    5 GENERAL CONFIGURING TEMPERA TURE SCALES The thermocouple and RTD ranges may be con fi gured for full time use of ° C, full time use of ° F or selectable ° C and ° F operation. This con fi guration is part of the DEF AUL T SETTINGS below . AUTO-OFF T echChek 830-KP can be set up to turn itself off after 30 minutes of inactivity . The interna ...

  • Altek a 830-KP - page 5

    GENERAL SELECTING FUNCTIONS Press a function key to choose among mA, V , T/C, Ω , RTD, FREQ and PRESSURE. SELECTING RANGES Press the RANGE/TYPE pushbutton to select the desired range and scale. SOURCE RANGES Milliamp: Source: mA, %mA (% 4-20), DP% (% Differential Pressure) 2-Wire Simulator: mA, %mA (% 4-20), DP% (% Dif ferential Pressure) VDC: mV ...

  • Altek a 830-KP - page 6

    9 CALIBRA TE MILLIAMP INPUTS mA, mA % (Percent of 4 to 20 mA), DP% (DP Flow) Choose this function to provide an output from 0.00 to 24.00 milliamps. The compliance voltage is a nominal 24 VDC to provide the driving power to your milliamp receivers. 1) Disconnect one or both input wires from the device to be calibrated. 2) Press the MILLIAMP/V push- ...

  • Altek a 830-KP - page 7

    11 CALIBRA TE VOL T AGE INPUTS V, m V Choose this function to provide an output from 0.00 mV to 1 10.00 mV and from 0.00 to 10.25 VDC. Current compliance up to 20 mA to provide the driving power to your voltage receivers. 1) Disconnect one or both input wires from the device to be calibrated 2) Press the MILLIAMP/V push-button until V or mV appear ...

  • Altek a 830-KP - page 8

    13 CONTROLLER (TYPICAL) ADDITIONAL 4-20 mA OR 1-5V INSTRUMENTS RECORDER COMPUTER LOGGER ETC, 1-5V DC 250 Ω T ypical 2-Wire T ransmitter +IN- REF +OUT- CHECK 1-5 VOL T INPUTS WITHOUT DISCONNECTING WIRES OPERA TING INSTRUCTIONS 12 CHECK 1-5 VOL T INPUTS WITHOUT DISCONNECTING WIRES Most 1-5 V olt receivers in 4-20mA loops have a 250 Ohm resistor acr ...

  • Altek a 830-KP - page 9

    15 CALIBRA TE RESIST ANCE INPUTS Choose this function to simulate a resistance into a variety of instruments. 1) Disconnect one or both input wires from the device to be calibrated 2) Press the T/C/RTD/ Ω pushbutton until the Ω symbol appears on the display . 3) Press the DISPLA Y/SOURCE/READ push-button until SOURCE appears on the display 4) C ...

  • Altek a 830-KP - page 10

    17 OPERA TING INSTRUCTIONS Instrument with R TD Input Controller T emperature T ransmitter T emperature Indicator T emperature T rip or Alarm Receiver Input Ohms or 2-Wire R TD Controller T ransmitter Computer Three Wire RTD Connection T wo Wire RTD Connection CALIBRA TE RTD INPUTS Four Wire RTD Connection SENSOR } + COMP - COMP Jumper Jumper 16 CA ...

  • Altek a 830-KP - page 11

    19 READ FUNCTIONS Select read by pressing the DISPLA Y/SOURCE/READ pushbutton until the word READ appears on the LCD display . The READ functions measure the desired signal. Multiple scales are available for some functions. MIN/MAX T o read the Maximum or Minimum INPUT since READ mode was entered, simply press the MAX or MIN push-button. The value ...

  • Altek a 830-KP - page 12

    21 READ DC and AC VOL T AGE OUTPUTS mV , V , Vhi and V ac Choose this function to measure from 0.00 to 1 10.00 millivolts (mV), 0.00 to 10.25 DC V olts (V). Use the high voltage connection to read from 0.0 to 200.0 VDC (Vhi). See the next page for V olts AC. 1) Press the MILLIAMP/V push-button until mV , V , Vhi or V AC appear on the display 3) Pre ...

  • Altek a 830-KP - page 13

    23 22 MEASURE THERMOCOUPLE SENSORS Choose this function to read a thermocouple. The input of the T echChek 830-KP is automatically cold junction compensated. 1) Disconnect the thermocouple from any instrument. 2) Press the T/C/RTD/ Ω pushbutton until the word TYPE & any T/C appear on the display . 3) Press the DISPLA Y/SOURCE/READ push-button ...

  • Altek a 830-KP - page 14

    25 CHECK CONTINUITY Choose this function to check continuity . A tone will sound and a sound symbol will appear on the display when the resistance between the leads is less than 100 Ohms. 1) Plug the leads into the T echChek 830-KP as shown below . 2) Press the T/C/RTD/ Ω pushbutton until the Ω symbol appears on the display . 3) Press the DISPL ...

  • Altek a 830-KP - page 15

    27 BLA CK BLA CK RED 3 2 1 R TD Element OPERA TING INSTRUCTIONS MEASURE RTD SENSORS 26 MEASURE RTD SENSORS Choose this function to read an RTD. Three wires must be use for both 2 and three wire RTDs. 1) Disconnect the RTD from the instrument being calibrated. 2) Press the T/C/RTD/ Ω pushbutton until any R TD type (Pt, Ni or Cu) appears on the dis ...

  • Altek a 830-KP - page 16

    29 Frequency Output Signal Flowmeter Flow Sensor V ariable Speed Drive Controller T ransmitter Frequency Output Signal Flow Meter Line Frequency V ariable Speed Drives Connection for signals with amplitudes below 10.25V RMS Connection for signals with amplitudes to 250 V RMS COUNT FREQUENCIES OPERA TING INSTRUCTIONS 28 COUNT FREQUENCIES Choose this ...

  • Altek a 830-KP - page 17

    31 +IN- REF T ypical 2-Wire T ransmitter OPERA TING INSTRUCTIONS CALIBRA TE 2-WIRE TRANSMITTERS (T/C / mA Shown) CALIBRA TE 4-WIRE TRANSMITTERS +IN- REF T ypical 4-Wire T ransmitter Loop Power Supply 24 VDC 30 OPERA TING INSTRUCTIONS CALIBRA TE TRANSMITTERS ANY SOURCE FUNCTION AND READ mA, READ %, READ DP% Choose this function to supply the input s ...

  • Altek a 830-KP - page 18

    33 Pressure Source PROCESS CALIBRA TOR QUIK-CHEK READ PRESSURE Choose this function to measure pressure with a ModPak 91 module holder and one of the many QuikCal 90 pressure modules. 1) Select the proper module for the pressure range of the process 2) Place the QuikCal 90 Module in the ModPak 91 holder 3) Plug the ModPak cable into the mating conn ...

  • Altek a 830-KP - page 19

    35 PROCESS CALIBRATOR QUIK-CHEK 34 OPERA TING INSTRUCTIONS CALIBRA TE PRESSURE TRANSMITTERS READ PRESSURE AND P-XMTR mA, P-XMTR %, P-XMTR DP% Choose this function to simultaneously supply power to a 2-Wire pressure transmitter while supplying the input signal to the transmitter and displaying the 4-20 mA output of the transmitter . 1) Disconnect bo ...

  • Altek a 830-KP - page 20

    37 SPECIFICA TIONS POWER & MEASURE 2-WIRE TRANSMITTERS RANGES & ACCURACY : Same as for MILLIAMP SOURCE OUTPUT CURRENT : up to 24.00 mA TYPICAL DRIVE CAP ABILITY : 1200 Ohms @ 20.00 mA COMPLIANCE VOL T AGE: nominal 25 VDC @ 20 mA COMMON MODE ERROR: 0.01% Full Scale/Common Mode V olt 2-WIRE TRANSMITTER SIMULA TO R RANGES: 1.00 to 24.00 mA; -1 ...

  • Altek a 830-KP - page 21

    39 SPECIFICA TIONS SOURCE RTD & OHMS RTD TYPES: Pt 100 Ω for 1.3850 (DIN/IEC 751 & New JIS), 1.3902 (Burns), 1.3926 (US Lab) & 1.3916 (Old JIS 1604C-1981) Ni 120 Ω , Cu 10 Ω & Cu 50 Ω RTD RESOLUTION: 1 ° C or 1 ° F RANGE OHMS: 0.0 to 400.0 Ohms ACCURACY : ± 0.05% of Full Scale + 0.075 mV/mA Excitation Current ACCURACY OHM ...

  • Altek a 830-KP - page 22

    41 THERMOCOUPLE SPECIFICA TIONS T/C TYPE N R S ° C ° F RANGE ACCURACY RANGE ACCURACY 300 to 1300 ± 2.1 572 to 2372 ± 3.0 100 to 299 ±2.3 212 to 571 ± 3.3 -50 to 99 ± 2.6 -58 to 21 1 ± 4.0 -200 to -49 ± 5.0 -328 to -57 ± 8.2 1750 to 1768 ± 4.2 3182 to 3214 ± 6.9 950 to 1749 ± 4.0 1742 to 3181 ± 6.5 650 to 949 ±4.4 1202 to 1741 ± 7.2 ...

  • Altek a 830-KP - page 23

    43 QuikCal 90 MODULE SPECIFICA TIONS OPERA TING TEMPERA TURE: -10 ° C TO 50 ° C (13 ° F TO 122 ° F) STORAGE TEMPERA TURE: -40 ° C T O 85 ° C (-40 ° F TO 185 ° F) WEIGHT : 0.4 kg (14.5 oz) TEMPERA TURE EFFECT : None (Compensated over full range) CONNECTION: 1/8 ” NPT FEMALE MEDIA COMP A TIBILITY : Any liquid or gas compatible with 316 stai ...

  • Altek a 830-KP - page 24

    44 W ARRANTY Altek products are warranted to be free from defects in material and workmanship (excluding fuses, batteries and leads) for a period of three years from the date of shipment. W arranty repairs can be obtained by returning the equipment prepaid to our factory . Products will be replaced, repaired, or adjusted at our option. Altek gives ...

  • Altek a 830-KP - page 25

    PN 1800470 Rev A April 2002 ©2002 Altek Industries, Inc. Speci fi cations subject to change without notice. All rights reserved. Printed in U.S.A. ...

Produzent Altek Kategorie Scale

Dokumente, die wir vom Produzenten des Geräts Altek a 830-KP erhalten, können wir in mehrere Gruppen teilen. Unteranderem in:
- technische Zeichnungen Altek
- Bedienungsanleitungen a 830-KP
- Produktkarten Altek
- Informationsbroschüren
- oder Energieetiketten Altek a 830-KP
Jede von ihnen ist wichtig, jedoch finden wir die wichtigsten Informationen für den Nutzer des Geräts in der Bedienungsanleitung Altek a 830-KP.

Die Dokumentengruppe, die als Bedienungsanleitungen bezeichnet wird, wird ebenfalls in detaillierte Arten geteilt, solche wie: Montageanleitungen Altek a 830-KP, Wartungsanleitungen, Kurzanleitungen oder Benutzeranleitungen Altek a 830-KP. Abhängig vom Bedarf, sollten Sie das Dokument finden, das Sie brauchen. In unserem Service können Sie sich die populärste Bedienungsanleitung des Produkts Altek a 830-KP ansehen.

Ähnliche Bedienungsanleitungen

Die komplette Bedienungsanleitung des Geräts Altek a 830-KP, wie sollte sie aussehen?
Die Bedienungsanleitung, auch bezeichnet als Benutzerhandbuch, oder einfach nur „Anleitung”, ist ein technisches Dokument, das dem Benutzer bei der Nutzung von Altek a 830-KP hilfreich sein soll. Die Bedienungsanleitungen werden in der Regel von technischen Schriftstellern geschrieben, aber in einer Sprache, die für alle Nutzer von Altek a 830-KP verständlich ist.

Eine gänzliche Bedienungsanleitung von Altek sollte einige Grundelemente enthalten. Ein Teil von ihnen ist nicht so wichtig, wie z.B.: die Titelseite oder Autorenseiten. Die restlichen von ihnen jedoch, sollten Informationen liefern, die für den Nutzer von enormer Wichtigkeit sind.

1. Einführung und Hinweise, wie man sich in einer Bedienungsanleitung von Altek a 830-KP bewegt - Am Anfang jeder Bedienungsanleitung sollten wir Hinweise bezüglich der Nutzungsart eines bestimmten Ratgebers finden. In ihr sollten sich Informationen über die Lokalisierung des Inhaltsverzeichnisses von Altek a 830-KP befinden, FAQ oder über oft auftretende Probleme – also Stellen, die von den Benutzern in jeder Bedienungsanleitung am meisten gesucht werden
2. Inhaltsverzeichnis - Index aller Ratschläge bezüglich Altek a 830-KP, die wir im aktuellen Dokument finden
3. Ratschläge zur Nutzung der Grundfunktionen des Geräts Altek a 830-KP - die uns die ersten Schritte während der Nutzung von Altek a 830-KP erleichtern sollten
4. Troubleshooting - geordneter Tätigkeitslauf, der uns bei der Diagnose und als nächstes bei der Lösung wichtiger Probleme mit Altek a 830-KP hilft
5. FAQ - häufig gestellte Fragen
6. Kontaktdaten Informationen darüber, wo man Kontakt zum Produzenten / Service von Altek a 830-KP im bestimmten Land suchen kann, wenn es nicht gelingt, das Problem selbst zu lösen.

Haben Sie eine Frage bezüglich Altek a 830-KP?

Nutzen Sie das untere Formular

Wenn Sie mit Hilfe der gefundenen Bedienungsanleitung Ihr Problem mit Altek a 830-KP nicht gelöst haben, stellen Sie eine Frage, indem Sie das untere Formular nutzen. Wenn einer der Nutzer ein ähnliches Problem mit Altek a 830-KP hatte, ist es möglich, dass er mit Ihnen die Lösung teilen möchte.

Text vom Bild übertragen

Kommentare (0)