Bedienungsanleitung Agilent Technologies 89441A

92 Seiten 1.55 mb
Download

Zur Seite of 92

Summary
  • Agilent Technologies 89441A - page 1

    Agilent Technologies 89441A Getting Started Guide Agilent Technologies Part Number 89441-90076 For instruments with firmware version A.08.00 Printed in U.S .A. Print Date: June 2000 © Agilent Technologies 1994, 1995, 2000. All rights reserved. 8600 Soper Hill Road Ev erett, Washington 98205-1298 U.S.A. This software and documentati on is based in ...

  • Agilent Technologies 89441A - page 2

    The Analyzer at a Glance 4 5 7 3 9 1 8 2 6 3 16 14 11 15 18 13 10 12 17 ii ...

  • Agilent Technologies 89441A - page 3

    Front Pane l 1 -A softkey’s function changes as different menus are displayed. Its current function is determined by the vide o label to its l eft, on the analyzer’s screen. 2 -The analyzer’s screen is divided into two main areas. The menu area, a narrow column at the screen ’s right edge, displays softkey labels. The data area, the remaini ...

  • Agilent Technologies 89441A - page 4

    This page left in tentionally blank. iv ...

  • Agilent Technologies 89441A - page 5

    Saftey Summary The following general safet y precautions must be observed during all phases of operation of this instrument. Failure t o comply with these precautions or with specific warnings else where in this manual viol ates safety standards of design, manufacture , and intended use of the instrument. Agilent Technologies, Inc. assumes no liabi ...

  • Agilent Technologies 89441A - page 6

    FUSES Only fuses with the required rated current, voltage, and specif ied type (normal blow, time delay, etc.) should be used. Do not use repaired fuses or short-circuited fuse holders. T o do so could cause a shock or fire hazard. DO NOT OPERATE I N AN EXPLOSIVE ATMO SPHERE Do not operate the instrument in the presence of flammab le gases or fumes ...

  • Agilent Technologies 89441A - page 7

    Safet y Symbo ls Warning, risk of electric shock Caution, refer to accompanyin g documents Alternating current Both direct and alternati ng current Earth (ground) termina l Protective earth (ground) terminal Frame o r chas sis ter minal Terminal is at earth potential. Standby (supply) . Units with this symbol are not completely d isconnected from a ...

  • Agilent Technologies 89441A - page 8

    Notation Conventions Before you use this b ook, it is important to understa nd the types of keys o n the front panel of the analyzer and how the y are denoted in this book. Hardkeys Hard keys are fron t-panel b uttons who se function s are always the same . Hardkeys have a label printed directly on t he key. In this book, t hey are printed like thi ...

  • Agilent Technologies 89441A - page 9

    In This Book This book, “Agilent Technologies 89441A Getting Started Guide”, is designed to help you become comfortable with the Agilent 89441A Vector Signal Analyzers. It provides step-by step examples of how to use this analyzer to perform tasks which you have probably performed with other analyzers. By performing these tasks you will become ...

  • Agilent Technologies 89441A - page 10

    This page left intentionally blank x ...

  • Agilent Technologies 89441A - page 11

    Table of Contents 1 Using Online Help To lea rn about onl ine he lp 1-2 To display help for hardkeys and softkeys 1-3 To display a related help topic 1-4 To sele ct a top ic from the h elp in dex 1-5 2 Making Simple Noise Me asurements To measure random noise 2-2 To measure band power 2-3 To measure signal to noise r atios 2-4 To measure ad jacent- ...

  • Agilent Technologies 89441A - page 12

    6 General Ta sks To set up peripherals. 6-2 To print or plot scre en contents 6-3 To save data with an internal or RAM disk 6-4 To recall data with an internal or RAM disk 6-5 To format a disk 6-6 To create a math f unction 6-7 To use a math function 6-8 To display a sum mary of in strume nt param eters 6-9 Inspection and Installation 7 Preparing t ...

  • Agilent Technologies 89441A - page 13

    1 Using Online Help You can learn about your analyzer from onli ne help which is built right int o the instrument and is available to you any time you use the analyzer. This section shows you how to use online help to learn about specif ic keys or topics. You can use online help in conjunction wi th other documentation to l earn about your analyzer ...

  • Agilent Technologies 89441A - page 14

    To learn about online h elp 1 Enter the online help system: Press [ Help ]. 2 Display online help for the [ Help ] hardkey: Press [ 5 ] on the numeric keypad. 3 Use the knob or the up-arrow or down-arrow keys to move through the pages. 4 Quit online help: Press [ Help ]. or Press [ 0 ] on the keypad. Take a few moments to read the help overview. It ...

  • Agilent Technologies 89441A - page 15

    To display help for hardkeys and s oftkeys This example displays topics related to triggerin g. 1 Enter the online help system: Press [ Help ]. 2 Display help for a hardkey: Press [ Trigger ]. 3 Use the knob or the up and down arrow keys to page through the topic. 4 Select a softkey topic: Press [ trigger ty pe ], [ IF channel 1 ]. 5 Quit online he ...

  • Agilent Technologies 89441A - page 16

    To display a related help topic This example displays topics related to sav ing and recalling. 1 Enter the online help system: Press [ Help ]. 2 Display help for a hardkey: Press [ Save/Recall ]. 3 Scroll with the knob to highlight the Math topic. 4 Select that topic: Press [ 4 ]. 5 Return to previous topics: Press [ 7 ]. 6 Quit online help: Press ...

  • Agilent Technologies 89441A - page 17

    To select a topic f rom the help index 1 Enter the online help system: Press [ Help ]. 2 Display the index: Press [ 1 ]. 3 Turn the knob to select the topic you want help on or for faster paging press and hold the up-arrow or down-arrow keys t hen use the knob to select a topic. 4 Display the topic: Press [ 4 ]. 5 Quit online help: Press [ Help ]. ...

  • Agilent Technologies 89441A - page 18

    ...

  • Agilent Technologies 89441A - page 19

    2 Making Simple Noise Measurements This chapter shows you how to make typical noise measurements. In this example, we will be making random noise, band power noise, and signal to noise measurements. 2-1 ...

  • Agilent Technologies 89441A - page 20

    To measure random noise 1 Initialize the analyzer: Press [ Preset ]. 2 Select a power spectral density measurement: Press [ Measurement Data ], [ PSD ]( s e l e c t ch1 with a 2-channel analyzer). 3 Turn on averaging: Press [ Average ], [ average on ]. 4 Start an averaged measurement: Press [ Meas Restart ]. 5 Use the knob to move the marker along ...

  • Agilent Technologies 89441A - page 21

    To measure band power 1 Initialize the analyzer: Press [ Pres et ]. 2 Turn on averaging: Press [ Average ], [ average on ]. 3 Start an averaged measurement: Press [ Meas Restart ]. 4 Turn on the band power markers: Press [ Marker Function ], [ band power markers ], [ band pwr mkr on ], [ band power ] Press [ ResBW/Window ], [ detector ], [ sample ] ...

  • Agilent Technologies 89441A - page 22

    To measure signal to n oise ratios 1 Select the baseband receiver mode and initialize the analyzer: Press [ Instrument Mode ][ receiver ][ RF section (0-10 MHz) ]. Press [ Pres et ]. 2 Supply a signal from the internal source: Connect th e SOURCE outp ut to the INP UT with a BNC cab le. Press [ Source ], [ source on ], [ sine freq ], 5, [ MHz ] 3 P ...

  • Agilent Technologies 89441A - page 23

    7 Change to a normalized noise measurement: Toggle to [ power ratio C/No ] The display shou ld appear a s below. The grid lin es have been turned off to highlight the band power markers. The carrier-to-noise and carrier-to-normalized-noise marker m easurements require that the standard (di amond shaped) marker be on the signal peak as a reference. ...

  • Agilent Technologies 89441A - page 24

    ...

  • Agilent Technologies 89441A - page 25

    3 Using Gating to Characterize a Burst Signal This chapter uses the time gating feature to analy ze a multi-burst signal which is provided on the Signals Disk which accompanies the analyzer’s Operator’s Guide . Time gating allow s you to isolate a p ortion of a tim e record for further viewing and analysis. For more details on time gating conce ...

  • Agilent Technologies 89441A - page 26

    To Use Tim e Gatin g First we’ll look at the spectrum of the signal and see that t hree components exist. Then we’ll look at the time display of the burst signal a nd analyze each burst separately to determine which spectral comp onents exist in each burst. 1 Select the baseband receiver mode and initialize the analyzer: Press [ Instrument Mode ...

  • Agilent Technologies 89441A - page 27

    5 Configure the display and the measurement: Press [ Display ], [ 2g r i d s ], [ more display setu p ], [ grids off ]. Press [ B ], [ Measurement Data ], [ main time ] (toggle to ch1 on a 2-channel analyzer). Press [ Ref Lvl/Scale ], [ Yp e rd i v ], 50, [ mV ]. Press [ Trigger ], [ trigger type ], [ internal source ]. Press [ Time ], [ main leng ...

  • Agilent Technologies 89441A - page 28

    ...

  • Agilent Technologies 89441A - page 29

    4 Measuring Relative Phase This section shows you how to make typical rel ative phase measurements on modulated carrier signals. In this example, you measure the phase of sidebands on AM a nd PM sign als relative to th e carrier. The test signal s are provided on the Signals Disk which accom panies the analyzer’s Operator’s Guide . 4-1 ...

  • Agilent Technologies 89441A - page 30

    To measure the relative phase of an AM signal 1 Select the baseband receiver mode and initialize the analyzer: Press [ Instrument Mode ][ receiver ][ RF section (0-10 MHz) ]. Press [ Pres et ]. 2 Load AM and PM signals from the Signals Disk int o registers and play the AM signal through the source: Insert the Signals Disk in the internal dis k driv ...

  • Agilent Technologies 89441A - page 31

    6 Activate two traces: Press [ Shift ], [ A ]( t w oA c t i v eT r a c eL E D sa r en o wt u r n e do n ) 7 Turn on marke r coupling and zero t he offset marker on the carrier: Press [ Marker ], [ couple mkrs on ], Press [ Shift ], [ Marker ] to place the marker on the carr ier peak, Press [ Shift ], [ Marker ⇒ ] to zero the offset marker. 8 Use ...

  • Agilent Technologies 89441A - page 32

    To measure the relative phase of an PM signal Continue from “To measure the relative phase of an AM signal.” 1 Replace the arbitrary source AM signal with the PM signal in register D2: Press [ Source ], [ source type ], [ arb data reg ], [ D2 ]. 2 Start a single sweep: Press [ Pause|Single ]. 3 Zero the offset marker on the carrier: Press [ Shi ...

  • Agilent Technologies 89441A - page 33

    5 Characterizing a Filter This section shows you how to make a typical network measurement . In this example, we will be charact erizing a 4.5 MHz bandpass fil ter. 5-1 ...

  • Agilent Technologies 89441A - page 34

    To set up a fr equency respons e measurement Note : This measurement can only be performed with a 2-channel anal yzer—you must have opt ion AY7. You must use the source output and the channel 1 and channel 2 inputs on the IF section for network measurements. 1 Using a BNC “T” adapter or power s plitter and BNC cables, connect the analyzer’s ...

  • Agilent Technologies 89441A - page 35

    4 Configure the source and measureme nt for a frequency r esponse measure ment: Press [ Source ], [ source on ], Press [ source type ], [ periodic chirp ], Press [ Return ], (bottom softkey) Press [ level ], .5, [ Vrms]. Press [ Res Bw/Window ], [ rbw mode arb ], Press [ main window ], [ uniform ]. Press [ Range ], [ channel both ], [ ch* single ra ...

  • Agilent Technologies 89441A - page 36

    To use the absolute marker Continue from “To set up a frequency response measurement .” 1 Move the marker to the largest part of the frequency response trace: Press [ Marker ⇒ ], [ marker to peak ]. or Press [ Shift ], [ Marker ] 2 Move the marker with the knob to view the absolute gain/loss of this particular filter network at different freq ...

  • Agilent Technologies 89441A - page 37

    To use the relative marker Continue from “To set up a frequency response measurement ” or from “Using t h ea b s o l u t em a r k e r . ” 1 Move the marker to the largest part of the frequency response trace if it is not already there: Press [ Shift ], [ Marker ]. 2 Establish the reference point for the relative (offset) marker: Press [ Mar ...

  • Agilent Technologies 89441A - page 38

    To use the search marker Complete “To set up a frequency response measurement” or continue f rom one of the previous m arker measurements. 1 Move the marker to the largest part of the frequency response trace if it is not already there: Press [ Shift ], [ Marker ]. 2 Activate and zero the offset marker if it is not already activated: Press [ Sh ...

  • Agilent Technologies 89441A - page 39

    To display phase Complete “To set up a frequency response measurement” or continue f rom one of the previous m arker measurements. 1 Display a second trace: Press [ Display ], [ 2g r i d s ]. 2 Activate the second trace and define it as a frequency response measurement: Press [ B ], [ Measurement Data ], [ frequency response ]. 3 Specify phase ...

  • Agilent Technologies 89441A - page 40

    To display coherence Complete “To set up a frequency response measurement” or continue f rom one of the previous m easurements. 1 Display a second trace: Press [ Display ], [ 2g r i d s ]. 2 Activate the second trace and select a coherence measurement: Press [ B ], [ Data Format ], [ magnitud e linear ], [ Measurement Data ], [ more choices ], ...

  • Agilent Technologies 89441A - page 41

    6 General Tasks This chapter shows you how to perform various common tasks. These include setting up and using peripherals and defining and using math functions. 6-1 ...

  • Agilent Technologies 89441A - page 42

    To set up per ipherals. You may connect peripherals to three ports—one GPIB port, one serial port, and one parallel port. GPIB peripherals may include printers, plo tters, and external disk drives. Supported serial devices are plotte rs and printers. Certain printers are parallel devices. 1 Connect the ports of your peripheral and analyzer with t ...

  • Agilent Technologies 89441A - page 43

    To print or plot screen content s 1 Set up your printer or plotter if you haven’t alrea dy done so. 2 Select the output format and device type: Press [ Plot/Print ], [ output fmt ] and select the desired forma t. Press [ device defaults ] and select a devic e if you want other than the default. 3 Select the type of output port: Press [ Plot/Print ...

  • Agilent Technologies 89441A - page 44

    To save data with an internal or RAM disk You may save trace data, instrument states, trace math functions, instrum ent BASIC programs, and time-capture buffers. 1 Sel ect t he de faul t di sk: Press [ Save/Recall ], [ default disk ] Press [ nonvolatile RAM disk ], [ volatile RA M dis k ]o r [ internal disk ] 2 Pres s [ Return ]. 3 Press the softke ...

  • Agilent Technologies 89441A - page 45

    To recall data with an internal or RAM disk You may recall trac e data, instrument sta tes, trace math functions, instrument BASIC programs, and time-capture buffers. 1 Sel ect t he de faul t di sk: Press [ Save/Recall ], [ default disk ] Press [ nonvolatile RAM disk ], [ volatile RA M dis k ]o r [ internal disk ] 2 Pres s [ Return ]. 3 To easily r ...

  • Agilent Technologies 89441A - page 46

    To format a disk 1 Select the disk drive you want to format: Press [ Disk Utility ], [ def ault disk ]. Press the softkey corresponding to the disk drive you want to format. 2 Pres s [ Return ], [ format disk ]. Select appropriate parameters for your disk drive ( disk type, interleave e tc.). 3 Pres s [ perform format ], [ proceed ]. You may format ...

  • Agilent Technologies 89441A - page 47

    To create a math function In this section you learn how to create a math function w hich inverts a signal. 1 Initialize the analyzer: Press [ Preset ] 2 Define a constant: Press [ Math ], [ define constant ], [ d efine K1 ] Press [ real part ], 1, [ enter ], [ imag part ], 0, [ enter ]. 3 Define a math function: Press [ Math ], [ define F 1 ] Press ...

  • Agilent Technologies 89441A - page 48

    To use a math function In this section you learn how to apply a a math fun ction to a signal. This task assumes that you have compl eted “To create a math f unction.” 1 Initialize the analyzer: Press [ Preset ] 2 Provide an averaged signal from the internal source: Press [ Source ], [ source on ], [ Average ], [ average on ]. 3 Apply the inv er ...

  • Agilent Technologies 89441A - page 49

    To display a summary of ins trument parameters 1 Pres s [ View State ]. 2 Pres s [ measurement state ]o r[ in put/source state ]. These summaries reflect the current states of important measurement, input, and source param eters. You may use these summaries to: quickly check the curre nt setup document the setup (The list can be print ed or plotted ...

  • Agilent Technologies 89441A - page 50

    ...

  • Agilent Technologies 89441A - page 51

    7 Preparing the Analyzer for Use 7-1 ...

  • Agilent Technologies 89441A - page 52

    Preparing the Analyzer for Use This chap ter contains instru ctions for inspecting an d installing th e analyzer. This chapter also includes instructions for cleaning the screen, transporti ng and stori ng the analy zer. Power Requirements The analyzer can operate from a single-phase ac power source supplying voltages as shown in the table. With al ...

  • Agilent Technologies 89441A - page 53

    Power Cable and G rounding Requirements On the GPIB connector, pin 12 and pins 18 through 24 are tied to chassis ground and the GPIB cable shield. The instrument frame, chassis , covers, and all exposed metal surfaces including the conne ctors’ outer shell are connected to ch assis ground. However, if channel 2 in t he IF section i s not instal l ...

  • Agilent Technologies 89441A - page 54

    *The number shown for the plug is the indu stry identifi er for the plug onl y, the number shown for the c able is an Agilent pa rt number for a complet e cable including th e plug. **UL listed for use in the United States of America. Warning The power cable plug must be inserted into an outlet provided with a protective earth terminal. Defeating t ...

  • Agilent Technologies 89441A - page 55

    To do the incoming inspection The analyzer was carefully inspected both mechanically and electr ically before shipment. It should be free of marks or scratches, and it should meet its published specif ications upon rece ipt. 1 Inspect the analyzer for physical damage incurred in transit. If the analyzer was damaged in transit, do the following: Sav ...

  • Agilent Technologies 89441A - page 56

    6 Set the RF section’s rear pa nel and front panel power switches to on. Press the ‘ ’ l ‘’ symbol end of the rocker-switches located on the lower right of the rear panel and on the lower left of the front panel . The RF section provides standby power for the high precision frequency reference . The rear-panel line switch interrupts all p ...

  • Agilent Technologies 89441A - page 57

    To connect the sections Do NOT use the IF section’s EXT REF O UT connector or optional OVEN REF OUT connector as an external reference output. 1 Attach the IF section to the RF section. If the hardwa re is not insta lled, follow the instr uction s supplied w ith the Rear Panel Lock Foot Kit. If the ha rdware is a lready instal led, slide the IF s ...

  • Agilent Technologies 89441A - page 58

    5 Connect the IF section’s SOURCE connector to the RF section’s IN connector using the supplied 8.5-inch BNC-to-BNC cable. 6 Connect the IF section’s CHANNEL 1 connector to the RF section’s OUT connector using the supplied 8.5-inch BNC-to-BNC cable. Preparing the Analy zer for Use 7-8 ...

  • Agilent Technologies 89441A - page 59

    To install the analyzer The analyzer is shipped wi th plastic feet i n place, ready for use a s a portable bench analyzer. The plastic feet are shaped to make full-width modular instruments self-alig n when they are stacked. Install the analyzer to allow free circulation of cooling air. Cooling air enters the analyzer t hrough the rear panel and ex ...

  • Agilent Technologies 89441A - page 60

    To change the IF section’s line-voltage s witch The line-voltage selector switch is set at the factory to match the most commonly used line voltage in the cou ntry of destination . 1 Unplug the power cord from the IF section (the section with ‘’Agilent 89431A’’ silk screened on the lower right rear panel). 2 Slide the line voltage selecto ...

  • Agilent Technologies 89441A - page 61

    To change the RF section’s line-voltage sw itch The line-voltage selector switch is set at the factory to match the most commonly used line voltage in the cou ntry of destination . 1 Unplug the power cord from the RF section (the section with “Agilent 89431A” silk screened on its lower left rear panel). 2 Using a small screw driver, pry open ...

  • Agilent Technologies 89441A - page 62

    To change the I F section’s fuse T h ef u s ei si n s t a l l e da tt h ef a c t o r yt om a t c ht h em o s tc o m m o n l yu s e dl i n e voltage in the count ry of destination. 1 Unplug the power cord from the IF section (the section with ‘’Agilent 89410A’’ silk screened on its lower right rear panel). 2 Using a small screw driver, pre ...

  • Agilent Technologies 89441A - page 63

    To change the RF section’s fuse T h ef u s ei si n s t a l l e da tt h ef a c t o r yt om a t c ht h em o s tc o m m o n l yu s e dl i n e voltage in the count ry of destination. 1 Unplug the power cord from the RF section (the section with “Agilent 89431A” silk screened on its lower left rear panel). 2 Using a small screw driver, pry open th ...

  • Agilent Technologies 89441A - page 64

    To connect the analyzer to a LAN Analyzers with opti on UFG, 4 megabyte extended RAM and addit ional I/O, have a ThinLAN and AUI (attachment unit interface) port for connecting the analyzer to the LAN (loc al area network). 1 Set the power switch to off ( O ). 2 Connect the ThinLAN BNC cable to the ThinLAN port or the appropriate media access unit ...

  • Agilent Technologies 89441A - page 65

    To connect the analyzer to a serial device The IF section’s Serial 1 port is a 9-pin, EIA-574 port that can interface with a printer or plott er. The total allowabl e transmission path len gth is 15 meters. Connect the IF section’s SERIAL 1 port to a printer or plotter using a 9-pin female to 2 5-pin RS-232-C cable. Part Number Cable Descriptio ...

  • Agilent Technologies 89441A - page 66

    To connect the analyzer to an GPIB device The analyzer is compatibl e with the General Purpose Interface B us (GPIB). Total allowable transmission p ath length is 2 meters times the nu mber of devices or 20 meters, whichever is less. Operating distances ca n be extended using an GPIB Extender. Analyzers with opti on UFG, 4 megabytes extended RAM an ...

  • Agilent Technologies 89441A - page 67

    To connect the optional keyboard The analyzer may be connected to an optional external keybo ard. The keyboard remains a ctive even when the analyzer is not in alpha entry mode .T h i s means that you can operate t he analyzer using the external key board rather than the front panel. Pressing the appropriate keyboard key does the same thing as pres ...

  • Agilent Technologies 89441A - page 68

    4 To configure your analyzer for a keyboard other than U.S. English, press [ System Utili ty ][ keyboard type ]. Then press the appropriate softkey to select the language. Configuring your analyze r to use a keyboard other t han U.S. English onl y ensures that the analyzer recognizes the proper keys for that pa rticular keyboard. Configuring your a ...

  • Agilent Technologies 89441A - page 69

    To clean the screen The analyzer screen is c overed with a plastic d iffuser screen (th is is not removable by the operat or). U nder normal operating conditions, the onl y cleaning required will be an occasional dusting. However, if a foreign material adheres itself to the screen, do the following: 1 Set the IF sect ion’s power sw itch to off ( ...

  • Agilent Technologies 89441A - page 70

    To transport the analyzer Disconnect the IF section from the RF section a nd package each section using the original factory packaging or packaging identical to the factory packaging. Containers and materials identical to those used in factory p ackaging are available through Hewlett-Packard o ffices. If returning the analyzer to Hewlett-Packard fo ...

  • Agilent Technologies 89441A - page 71

    If the IF section will not power up q Check that the powe r cord is connected to the IF section and to a live power source. q Check that the front-panel swit ch is on ( l ). q Check tha t the vo ltage sel ector sw itch is set p roperly. See ‘’To change the IF section’s line-vol tage switch’’ on page 7-10. q Check that the fuse i s good. S ...

  • Agilent Technologies 89441A - page 72

    If the RF section will not power up q Check that the powe r cord is connected to the RF section and to a live power source. q Check that the RF section’s rear panel and front panel power switches are on ( l ). q Check tha t the vo ltage sel ector sw itch is set p roperly. See ‘’To change the RF secti on’s line-voltage switch’’ on page 7 ...

  • Agilent Technologies 89441A - page 73

    If the analyzer’s stop frequency is 10 MHz q Check that the RF section’s fan is running. If the fan is not r unning, see ‘’If t he RF section will not power up. ’’ q Check that the Serial 2 port on the IF section and on the RF section are connected together. q Press [ Instrument Mode ] and check that the receiver softkey displays ‘’ ...

  • Agilent Technologies 89441A - page 74

    ...

  • Agilent Technologies 89441A - page 75

    Index ! 16QAM demodulation, examp le OP 8-1 2-channels digital demod OP 6-1 2 video demod OP 7-15 32QAM signal, example OP 9-2 A A,B,C,D LEDs HT ac line voltage GS 7-2 adja cent- chann el pow er GS 2-6 air circulation GS 7-9 aliasing digital demod OP 17-11 video demod OP 18-11 alpha e ntry, usin g HT alpha, sett ing HT AM demodulation algori thm OP ...

  • Agilent Technologies 89441A - page 76

    bandwidth See information bandwidth See noise equivalent bandwidth See reso lutio n band width See windo w bandw idth bandwidth coupling OP 14-4 baseband measurements OP 14-3, HT baseband signals digital demod OP 6-12 video demod OP 7-15 BASIC HT about IBASIC HT available utilities HT creating a program with keystrokes HT debugging a program HT del ...

  • Agilent Technologies 89441A - page 77

    ch1 + j*ch2 receiver HT changing numeric parameters HT channel 1 See input channels channel 1 and 2 connector s HT channel 2 See input channels circulation , air GS 7-9 cleaning the screen GS 7-19 clear source trip softkey HT coherence GS 5-8, HT color index softkey HT color, adjust ing HT connecting frequency reference GS 7-7 IF section to RF sect ...

  • Agilent Technologies 89441A - page 78

    demodulation , digital (continued) constellat ion points, viewin g OP 8-5 DQPSK HT error-vector magn itude (in symbol table) HT error-vect or magnitude t race HT example OP 8-1 eye diagram HT filtering OP 17-16, HT filter s, use r define d OP 9- 6 formattin g display s OP 17-7 I/Q meas ured, displ aying HT I/Q origin o ffset (in symbol table) HT I/ ...

  • Agilent Technologies 89441A - page 79

    detection OP 14-12, HT about detec tion HT normal detector HT positive peak detect or HT sample detector HT deviation (F SK) HT diagnostics softkey HT digital demodulation See demodulation, digital disk drive setting up G S 6-2 disk drives HT about disk drives HT comparison H T copying files HT deleting files HT displa ying co ntents o f HT externa ...

  • Agilent Technologies 89441A - page 80

    filters (continu ed) user defined OP 17-18 video OP 14-3 firmware version, displaying HT FM demodulation algori thm OP 15-12 e x a m p l e O P1 - 6 ,O P3 - 4 FM demodulation , using HT fonts HT forma tting a di sk GS 6 -6 fourier tra nsform OP 13-7 French softkey HT freq counter softkey HT frequency HT about frequen cy para meters HT center frequen ...

  • Agilent Technologies 89441A - page 81

    GPIB address: external disk drive HT GPIB address: external receiver HT GPIB ad dress : plotte r HT GPIB ad dress : printe r HT GPIB connector GS 7-16 GPIB controllers HT keys that have GPIB comm ands HT LAN use OP 10-6 system controller HT H half LED HT description o f HT u s i n gt os e tt h er a n g e H T hardkeys, defined HT Help hardkey HT hel ...

  • Agilent Technologies 89441A - page 82

    ports, selecting HT subnet mask, setting HT Telnet, about HT X-Windows with OP 1 0-7 X11 display, configuring HT X11 display, control HT X11 display, update rate HT X11 IP add ress, s etting HT leakage OP 14-6 learning produ cts (manuals) HT LEDs: front panel HT level softkey HT line voltage required G S 7-2 RF selector switch GS 7-11 linear averag ...

  • Agilent Technologies 89441A - page 83

    measured filter, selecting HT measured signal video demod OP 18-10 measured signal, digital demod OP 17-9 measurement calculati ons, disabling HT measurement data, abou t HT MEASUREMENT Group hardkeys HT measurement points, defined HT measurement resolution OP 14-11 measurement speed, enhancing HT measurement state GS 6-9, HT recalling HT saving HT ...

  • Agilent Technologies 89441A - page 84

    relative, example GS 4-1 wrap G S 5-7 phase error See demodulation, digital phase noise, analysis example OP 2-1 phase, displaying HT PHS (PHP) H T plotter setting up G S 6-2 plotter interface GS 7-15 plotting GS 6-3, HT aborting H T about plotting HT changing plotter pen assignments HT data and file formats HT date and time HT default pen assignme ...

  • Agilent Technologies 89441A - page 85

    range har dkey HT RBW See reso lutio n band width real-time bandwidth HT rear pa nel tou r HT recalling HT about recalling HT autostate file HT BASIC programs HT capture-buffer cont ents HT math funct ions HT measurement sta te HT recalling data GS 6-5 spectrogram displays OP 5-16 trace GS 3 -2, HT waterfall displays OP 5-16 receiver H T about rece ...

  • Agilent Technologies 89441A - page 86

    block diagram HT block diagrams OP 12-3 limitation s OP 14-11 overview OP 12-2 scale at markers softkey HT scaling HT auto scaling HT x-axis HT x-axis scaling: linear vs. log HT X-axis, example OP 8-7 y-axis scaling H T scan time See sweep time screen, cleaning GS 7-19 search leng th HT digital demod OP 6-8 video demod OP 7-13 search length , in di ...

  • Agilent Technologies 89441A - page 87

    pausing HT printing/p lotting, formats HT printing/plotti ng, overview HT real-time indi cator HT recalling HT saving HT threshol d, setti ng HT trace buffe rs, removin g HT trace, selecting HT turning o n HT u s i n gi nm a t hf u n c t i o n s H T spectrum analyzers comparisons OP 13-9 types of OP 13 -4 spectrum measurements HT speed, enhancing H ...

  • Agilent Technologies 89441A - page 88

    time data OP 14-15 time data softk ey HT time domain overview OP 13-2 time gati ng HT about time gat ing HT displaying gate time-record HT gate delay HT gate length HT selecting a window HT turning o n/off H T See also gating time points, digital & video demodulation HT time reco rd OP 13- 7, OP 14- 16, HT about th e time recor d HT applying a ...

  • Agilent Technologies 89441A - page 89

    digital demod OP 6-12 video demod OP 7-15 U U.K. English softkey HT U.S. English soft key HT UFG (LAN and 4 MB memory option) HT UG7 (advanced LAN optio n) HT UNCAL, definition of HT units: x-axis HT units: y-axis HT user defined filters OP 17-18 V vector averaging See time averaging vector diagram, exa mple OP 6-5, OP 7-12 vector measurement s OP ...

  • Agilent Technologies 89441A - page 90

    ...

  • Agilent Technologies 89441A - page 91

    Agilent 89400-Series Documen tation Roadmap If you are thin king about.. . And you want to... Then rea d the analyz er’s... Unpacking and i nstall ing the analyzer Install the an alyzer, or do operation verification or performance verificatio n tests Installatio n and Verification Guide Getting started Make your first measurements with your new a ...

  • Agilent Technologies 89441A - page 92

    Need Assistance? If you need assistance, c ontact your nearest Agilent Technologies Sales and Service Office listed in the Agilent Cat alog. You can also find a list of local service representatives on the Web at: http://www .agilent -tech.c om/serv ices/Engli sh/ind ex.html or contact your nearest regional office listed below. If you are cont acti ...

Produzent Agilent Technologies Kategorie Stereo System

Dokumente, die wir vom Produzenten des Geräts Agilent Technologies 89441A erhalten, können wir in mehrere Gruppen teilen. Unteranderem in:
- technische Zeichnungen Agilent Technologies
- Bedienungsanleitungen 89441A
- Produktkarten Agilent Technologies
- Informationsbroschüren
- oder Energieetiketten Agilent Technologies 89441A
Jede von ihnen ist wichtig, jedoch finden wir die wichtigsten Informationen für den Nutzer des Geräts in der Bedienungsanleitung Agilent Technologies 89441A.

Die Dokumentengruppe, die als Bedienungsanleitungen bezeichnet wird, wird ebenfalls in detaillierte Arten geteilt, solche wie: Montageanleitungen Agilent Technologies 89441A, Wartungsanleitungen, Kurzanleitungen oder Benutzeranleitungen Agilent Technologies 89441A. Abhängig vom Bedarf, sollten Sie das Dokument finden, das Sie brauchen. In unserem Service können Sie sich die populärste Bedienungsanleitung des Produkts Agilent Technologies 89441A ansehen.

Ähnliche Bedienungsanleitungen

Die komplette Bedienungsanleitung des Geräts Agilent Technologies 89441A, wie sollte sie aussehen?
Die Bedienungsanleitung, auch bezeichnet als Benutzerhandbuch, oder einfach nur „Anleitung”, ist ein technisches Dokument, das dem Benutzer bei der Nutzung von Agilent Technologies 89441A hilfreich sein soll. Die Bedienungsanleitungen werden in der Regel von technischen Schriftstellern geschrieben, aber in einer Sprache, die für alle Nutzer von Agilent Technologies 89441A verständlich ist.

Eine gänzliche Bedienungsanleitung von Agilent Technologies sollte einige Grundelemente enthalten. Ein Teil von ihnen ist nicht so wichtig, wie z.B.: die Titelseite oder Autorenseiten. Die restlichen von ihnen jedoch, sollten Informationen liefern, die für den Nutzer von enormer Wichtigkeit sind.

1. Einführung und Hinweise, wie man sich in einer Bedienungsanleitung von Agilent Technologies 89441A bewegt - Am Anfang jeder Bedienungsanleitung sollten wir Hinweise bezüglich der Nutzungsart eines bestimmten Ratgebers finden. In ihr sollten sich Informationen über die Lokalisierung des Inhaltsverzeichnisses von Agilent Technologies 89441A befinden, FAQ oder über oft auftretende Probleme – also Stellen, die von den Benutzern in jeder Bedienungsanleitung am meisten gesucht werden
2. Inhaltsverzeichnis - Index aller Ratschläge bezüglich Agilent Technologies 89441A, die wir im aktuellen Dokument finden
3. Ratschläge zur Nutzung der Grundfunktionen des Geräts Agilent Technologies 89441A - die uns die ersten Schritte während der Nutzung von Agilent Technologies 89441A erleichtern sollten
4. Troubleshooting - geordneter Tätigkeitslauf, der uns bei der Diagnose und als nächstes bei der Lösung wichtiger Probleme mit Agilent Technologies 89441A hilft
5. FAQ - häufig gestellte Fragen
6. Kontaktdaten Informationen darüber, wo man Kontakt zum Produzenten / Service von Agilent Technologies 89441A im bestimmten Land suchen kann, wenn es nicht gelingt, das Problem selbst zu lösen.

Haben Sie eine Frage bezüglich Agilent Technologies 89441A?

Nutzen Sie das untere Formular

Wenn Sie mit Hilfe der gefundenen Bedienungsanleitung Ihr Problem mit Agilent Technologies 89441A nicht gelöst haben, stellen Sie eine Frage, indem Sie das untere Formular nutzen. Wenn einer der Nutzer ein ähnliches Problem mit Agilent Technologies 89441A hatte, ist es möglich, dass er mit Ihnen die Lösung teilen möchte.

Text vom Bild übertragen

Kommentare (0)