    IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: 61-310 This Service Manual provides the following information: ● Precautions and safety information ● Specifications ● Basic maintenance (cleaning, replacing the battery and fuses) ● Performance test procedures ● Calibration and calibration adjustment procedures Form Number: TM61310 Revision: ...

  • IDEAL INDUSTRIES 61-310 - page 2

    Form Number TM61310 Rev 2 Aug 2007 T ABLE OF CONTENTS Page Introduction 1 Precautions and Safety Information 1 Symbols 1 Safety 2 S pecifications 3 General S pecifications 3 Measurement Characteristics 4 V olt age S pecifications 4 Current S pecifications 4 Resistance S pecifications 5 Diode Check, Continuity , T emperature S pecifications 5 Auto P ...

  • IDEAL INDUSTRIES 61-310 - page 3

    Form Number TM61310 Rev 2 Aug 2007 Page 1 Introduction W arning T o av ures Th y . oid shock or injury , do not perform the verification test s or calibration proced described in the manual unless you are qualified to do so. e information provided in this document is for the use of qualified personnel onl Caution The 61-310 series cont ain part s t ...

  • IDEAL INDUSTRIES 61-310 - page 4

    Form Number TM61310 Rev 2 Aug 2007 Page 2 SAFETY Re uct or any For operating instructions, see the 61-310 Di git al Multimeter Instruction Manual. view the following safety precautions to av oid injury and prevent damage to this prod products connected to it. T o avoid potentia l hazards, use the produc t only as specified. CAUTION: These statement ...

  • IDEAL INDUSTRIES 61-310 - page 5

    Form Number TM61310 Rev 2 Aug 2007 All S ta ry ind Ge Page 3 SPECIFICA TIONS specifications are warranted unle ss noted typical and apply to the 61-310. ted accuracies are at 23°C±5°C at less than 70% relative humidity and without the batte icator displayed. neral specifications Characteristics Description Display count 3 ½ digit liquid crystal ...

  • IDEAL INDUSTRIES 61-310 - page 6

    Form Number TM61310 Rev 2 Aug 2007 Page 4 Me Ac (1) asurement Characteristics curacy is ±(% reading + number of digi ts) at 23°C ± 5°C, less than 80% R.H. DC V olts Range Resolution Accuracy Over v oltage protection 200.0mV 100 μ V 2.000V 1mV 20.00V 10mV 200.0V 100mV 0.5% + 2 600V 1V 0.8% + 2 600V rms Inp (2) ut Impedance: 10M Ω AC V olts Ra ...

  • IDEAL INDUSTRIES 61-310 - page 7

    Form Number TM61310 Rev 2 Aug 2007 Page 5 (4) Resist ance Range Resolution Accuracy Over v oltage protection 200.0 Ω * 1 0.1 Ω 0.8% + 4 2.000K Ω 1 Ω 20.00K Ω 10 Ω 200.0K Ω 100 Ω 2.000M Ω 1K Ω 0.8% + 2 20.00M Ω * 2 10K Ω 3.0% + 3 250V rms Op * 1 < (5) en circuit V oltage: -1.5V approx. 5 digit of reading rolling. * 2 ± 2% ...

  • IDEAL INDUSTRIES 61-310 - page 8

    Form Number TM61310 Rev 2 Aug 2007 Page 6 Physical and Env ironmental Characteristics Characteristics Description Dimensions (H×W×D) 150mm(H) ×76mm (W) ×38mm(D) (with holster) 5.9” (H) x 3.0”(W) x1.5”(D) Weight (with battery& holster) 0.219Kg (7.1 oz.) Environment al characteristics Description T emperature operating range 0 to +40°C ...

  • IDEAL INDUSTRIES 61-310 - page 9

    Form Number TM61310 Rev 2 Aug 2007 Page 7 Re Required equipm le B. If the recommended models are not available, equipment with equivalent specifications may be used. Repa performed only b T able E Characteristics ended Model quired Equipment ent is listed in T ab irs or servicing should be y qualified personnel. B. Required Equipment quipment Requi ...

  • IDEAL INDUSTRIES 61-310 - page 10

    Form Number TM61310 Rev 2 Aug 2007 Page 8 Basic Maintenance W arning avoid shock, remove the test leads and any input signals T o before opening the case or rep Op lacing the battery or fuses. ening the Meter Case Caution ircuit, alwa ys place the unco T o avoi onal short c vered met protec . Whe uit connect 1. Disconnect test lead ary switc d remo ...

  • IDEAL INDUSTRIES 61-310 - page 11

    Form Number TM61310 Rev 2 Aug 2007 Page 9 Replacing Fuses W arning To avoid electrical shock, remove the test leads and any input signals before replacing battery or fuses. the amage or injury, INSTALL ONLY quick a cting fuses with the following Amp/Volt F-310 Fuse: 250mA 250V FAST BLOW Fu 3. Remove the two screws under the battery cover . case unt ...

  • IDEAL INDUSTRIES 61-310 - page 12

    Form Number TM61310 Rev 2 Aug 2007 Page 10 Perf The following performance test acc Acc an operating temperature of 18°C to 28°C and a maximum of 80% relative humidity . T o perform the following tests, it is not necessary to open the case, no adjustments are nec the required connecti ons, apply the designated inputs, determine if the reading on t ...

  • IDEAL INDUSTRIES 61-310 - page 13

    Form Number TM61310 Rev 2 Aug 2007 Page 11 e Test: Table 2: AC Voltag Step Range Input Frequency Reading 1 200V 2.0V 60Hz 1.0 to 3.0 2 200V 2.0V 400Hz 1.0 to 3.0 3 200V 190.0V 60Hz 186.7 to 193.3 4 200V 190.0V 400Hz 186.7 to 193.3 5 600V 600V 60Hz 583.0 to 617.0 6 600V 600V 400Hz 583.0 to 617.0 T esting the DC milliamp 1 s 1-3 in T able 3. meter r ...

  • IDEAL INDUSTRIES 61-310 - page 14

    Form Number TM61310 Rev 2 Aug 2007 Page 12 T able 4:  Resistance T est: Step Range Source Reading 1 *200 OHM 2.0 Ω 1.6 to 2.4 2 *200 OHM 190.0 Ω 188.3 to 191.7 3 2K OHM 1.900K 883 to 1 Ω 1. .917 4 20K OHM 19.00K 18.83 to 19.17 Ω 5 200K OHM 190.0K 188.3 to 191.7 Ω 7 20M OHM 19.00M 18.40 to 19.60 Ω *Lead resistance on the 200 ohm range ...

  • IDEAL INDUSTRIES 61-310 - page 15

    Form Number TM61310 Rev 2 Aug 2007 Page 13 Cali T o Recalibrate your meter: It is recommended that the Multimeter be ca ch year . 1. Perform calibration at an a rature of 23°C±2°C a nd a relative humidity of <70% Dis nect the te s and turn th ff. Rem s from the front terminals. 2. Position the meter face down. Remove the battery cover screws ...

Produzent IDEAL INDUSTRIES Kategorie Digital Photo Keychain

Dokumente, die wir vom Produzenten des Geräts IDEAL INDUSTRIES 61-310 erhalten, können wir in mehrere Gruppen teilen. Unteranderem in:
- technische Zeichnungen IDEAL INDUSTRIES
- Bedienungsanleitungen 61-310
- Produktkarten IDEAL INDUSTRIES
- Informationsbroschüren
- oder Energieetiketten IDEAL INDUSTRIES 61-310
Jede von ihnen ist wichtig, jedoch finden wir die wichtigsten Informationen für den Nutzer des Geräts in der Bedienungsanleitung IDEAL INDUSTRIES 61-310.

Die Dokumentengruppe, die als Bedienungsanleitungen bezeichnet wird, wird ebenfalls in detaillierte Arten geteilt, solche wie: Montageanleitungen IDEAL INDUSTRIES 61-310, Wartungsanleitungen, Kurzanleitungen oder Benutzeranleitungen IDEAL INDUSTRIES 61-310. Abhängig vom Bedarf, sollten Sie das Dokument finden, das Sie brauchen. In unserem Service können Sie sich die populärste Bedienungsanleitung des Produkts IDEAL INDUSTRIES 61-310 ansehen.

Ähnliche Bedienungsanleitungen

Die komplette Bedienungsanleitung des Geräts IDEAL INDUSTRIES 61-310, wie sollte sie aussehen?
Die Bedienungsanleitung, auch bezeichnet als Benutzerhandbuch, oder einfach nur „Anleitung”, ist ein technisches Dokument, das dem Benutzer bei der Nutzung von IDEAL INDUSTRIES 61-310 hilfreich sein soll. Die Bedienungsanleitungen werden in der Regel von technischen Schriftstellern geschrieben, aber in einer Sprache, die für alle Nutzer von IDEAL INDUSTRIES 61-310 verständlich ist.

Eine gänzliche Bedienungsanleitung von IDEAL INDUSTRIES sollte einige Grundelemente enthalten. Ein Teil von ihnen ist nicht so wichtig, wie z.B.: die Titelseite oder Autorenseiten. Die restlichen von ihnen jedoch, sollten Informationen liefern, die für den Nutzer von enormer Wichtigkeit sind.

1. Einführung und Hinweise, wie man sich in einer Bedienungsanleitung von IDEAL INDUSTRIES 61-310 bewegt - Am Anfang jeder Bedienungsanleitung sollten wir Hinweise bezüglich der Nutzungsart eines bestimmten Ratgebers finden. In ihr sollten sich Informationen über die Lokalisierung des Inhaltsverzeichnisses von IDEAL INDUSTRIES 61-310 befinden, FAQ oder über oft auftretende Probleme – also Stellen, die von den Benutzern in jeder Bedienungsanleitung am meisten gesucht werden
2. Inhaltsverzeichnis - Index aller Ratschläge bezüglich IDEAL INDUSTRIES 61-310, die wir im aktuellen Dokument finden
3. Ratschläge zur Nutzung der Grundfunktionen des Geräts IDEAL INDUSTRIES 61-310 - die uns die ersten Schritte während der Nutzung von IDEAL INDUSTRIES 61-310 erleichtern sollten
4. Troubleshooting - geordneter Tätigkeitslauf, der uns bei der Diagnose und als nächstes bei der Lösung wichtiger Probleme mit IDEAL INDUSTRIES 61-310 hilft
5. FAQ - häufig gestellte Fragen
6. Kontaktdaten Informationen darüber, wo man Kontakt zum Produzenten / Service von IDEAL INDUSTRIES 61-310 im bestimmten Land suchen kann, wenn es nicht gelingt, das Problem selbst zu lösen.

Haben Sie eine Frage bezüglich IDEAL INDUSTRIES 61-310?

Nutzen Sie das untere Formular

Wenn Sie mit Hilfe der gefundenen Bedienungsanleitung Ihr Problem mit IDEAL INDUSTRIES 61-310 nicht gelöst haben, stellen Sie eine Frage, indem Sie das untere Formular nutzen. Wenn einer der Nutzer ein ähnliches Problem mit IDEAL INDUSTRIES 61-310 hatte, ist es möglich, dass er mit Ihnen die Lösung teilen möchte.