    Trace Oxygen Analyzer Teledyne Analytical Instruments i OPERA TING INSTRUCTIONS FOR Model 3190Z T race Oxygen Analyzer DANGER Toxic gases and or flammable liquids may be present in this monitoring system. Personal protective equipment may be requi red when servicing this inst rume nt. Hazardous voltages exist on certain co m ponents internally whic ...

  • Teledyne 3190Z - page 2

    Model 3190Z Teledyne Analytical Instruments ii Copyright © 2009 Teledyne Instruments/ Analytical Instruments All Rights Reserved. No part of this manual may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any other language or computer language in whole or in part, in any form or by any means, whether it b ...

  • Teledyne 3190Z - page 3

    Trace Oxygen Analyzer Teledyne Analytical Instruments iii Safety Messages Your safety and the safety of others is very important. We have provided many important safety m essages in this manual. Please read these messages carefully. A safety message alerts you to pot ential hazards that could hurt you or others. Each safety message is a ssociated w ...

  • Teledyne 3190Z - page 4

    Model 3190Z Teledyne Analytical Instruments iv CAUTION: THE ANALYZER SHOULD ONLY BE US ED FOR THE PURPOSE AND IN TH E MA NNER DESCRIBE D IN THIS MANUAL. IF YOU USE THE ANALYZER IN A MANNER OTHER THAN THAT FOR WHICH IT WAS INTEND ED, UNPREDICTABLE BEHAVI OR COULD RESULT POSSIBLY ACCOMPANIED WITH HAZARDO US CONSEQUENCE S. This manual provides informa ...

  • Teledyne 3190Z - page 5

    Trace Oxygen Analyzer Teledyne Analytical Instruments v T able of Content s Safety Me ssages .......................................................................... iii Introducti on ................................................................................... 1 1.1 Overview 1 1.2 Main Features of the Analyzer 1 1.3 Front Panel Description ...

  • Teledyne 3190Z - page 6

    Model 3190Z Teledyne Analytical Instruments vi 4.3.2 LO Range 19 4.4 Setting the Alarm Setpoints 19 4.4.1 HI Alarm 19 4.4.2 LO Alarm 19 4.4.3 Sensor Fail Alarm 20 4.5 Selecting a Fixed Range or Autoranging 20 4.6 Calibration 20 4.7 LED Display on StartUp 21 Maintena nce ............................................................................... ...

  • Teledyne 3190Z - page 7

    Trace Oxygen Analyzer Teledyne Analytical Instruments vii DANGER COMBUSTIBLE GAS USAGE WARNING This is a general purpose instrument designed for use in a non-hazardous area. It is th e customer's responsibility to ensure safety especially when combustible gases are being analyzed since the potential of gas leaks always exist. The customer shou ...

  • Teledyne 3190Z - page 8

    Model 3190Z Teledyne Analytical Instruments viii (This page intentionally left blank) ...

  • Teledyne 3190Z - page 9

    Trace Oxygen Analyzer Introduction Teledyne Analytical Instruments 1 Introduction 1.1 Overview The Teledyne Electronic Technologi es Analytical Instruments (TETAI) Model 3190Z is a microprocessor-based trace oxygen analyzer for real-time measurem ent of trace am ounts of oxygen in inert gases, or in a wide variety of gas mixtures. It features simpl ...

  • Teledyne 3190Z - page 10

    Introduction Model 3190Z Teledyne Analytical Instruments 2 • Compact and rugged control unit with flush-panel case. Designed for indoor use. Fr ont panel NEMA-4 rated. • External probe 1.3 Front Panel Description All controls and displays except the power switch are accessible from the front panel. See Figure 1-1. The front panel has seven push ...

  • Teledyne 3190Z - page 11

    Trace Oxygen Analyzer Introduction Teledyne Analytical Instruments 3 Data Entry Keys: Two pushbutton membrane switches are used to manually change measurem ent parameters of the instrument as they are displayed on the LED meter readout:  Up Arrow Increment values of parameters upwards as they are displayed on the LED readout. • Down Arrow Incr ...

  • Teledyne 3190Z - page 12

    Introduction Model 3190Z Teledyne Analytical Instruments 4 Figure 1-2 Rear Panel (AC and DC versions)  Power Connection AC version: 100–240 VAC, 50/60 Hz. The connector housing includes the fuse holder and the power switch. DC version: Requires 10 - 36 VDC. Fuse Holder: Replacing the fuse is described in Chapter 5 - Maintenance . ...

  • Teledyne 3190Z - page 13

    Trace Oxygen Analyzer Introduction Teledyne Analytical Instruments 5 I/O Power Switch: Turns the instrument power ON ( I ) or OFF ( O ). • Analog Outputs 0–10 VDC concentration output. 0–10 VDC Range ID (or optional over-range) output. 4–20 mADC concentration output, negative ground. • Alarm Connections HI Alarm, LO Alarm, and Sensor Fail ...

  • Teledyne 3190Z - page 14

    Introduction Model 3190Z Teledyne Analytical Instruments 6 (This page intentionally left blank) ...

  • Teledyne 3190Z - page 15

    Trace Oxygen Analyzer Operational Theory Teledyne Analytical Instruments 7 Operational Theory 2.1 Introduction The analyzer is composed of three subsystems: 1. Analysis unit with zirc onium oxide sensor(s) 2. Control unit with signal pro cessing, display and controls 3. Interface unit, supplying si gnals between 1 and 2. The analysis unit is design ...

  • Teledyne 3190Z - page 16

    Operational Theory Model 3190Z Teledyne Analytical Instruments 8 The sensor is designed for l ong term stability with an operating life of 5 years. This unit features one or two sens ors, and can be custom-designed for a specific range of oxygen analysis. 2.2.2 The Effect of Pressure In order to state the amount of oxygen present in the sample as a ...

  • Teledyne 3190Z - page 17

    Trace Oxygen Analyzer Operational Theory Teledyne Analytical Instruments 9 2.3.2 Signal Processing Figure 2-4 is a block diagram of the signal processing electronics described below. Figure 2-4: Block Diagram of the Signal Processing Electronics In the presence of oxygen, the sensor(s) produce a current. A current to voltage amplifier (I–E AMPL) ...

  • Teledyne 3190Z - page 18

    Operational Theory Model 3190Z Teledyne Analytical Instruments 10 RS-232 (optional) output. The same digi tal information is also sent to a 8-bit digital to analog converter ( DAC) that produces the 0-1 0 VDC analog concentration signal and th e 0-10 VDC analog range ID output. A current to voltage converter (E –I CONV) produces the 4-20 m ADC an ...

  • Teledyne 3190Z - page 19

    Trace Oxygen Analyzer Installation Teledyne Analytical Instruments 11 Inst allation Installation of the analyzer includes: 1. Unpacking the system. 2. Mounting the control unit a nd external sample block. 3. Making the electrical connections. 4. Making the gas connections. 5. Testing the installation. CAUTION: READ THIS CHAPTER IN ITS ENTIRETY BEFO ...

  • Teledyne 3190Z - page 20

    Installation Model 3190Z Teledyne Analytical Instruments 12 3.2 Location and Mounting 3.2.1 Control Unit Installation The 3190Z control unit is designe d to be panel-mounted in a general purpose, indoor area, away fr om moisture and the elem ents. The unit should be installed at view ing level in a sheltered area. CAUTION: FOR THE DC POWERED VER SI ...

  • Teledyne 3190Z - page 21

    Trace Oxygen Analyzer Installation Teledyne Analytical Instruments 13 standard AC power cord is supplied with the control unit. Insert the female plug end of the power cord into the power cord receptacle. The universal power supply allows direct connection to any 100- 240VAC, 50/60 Hz power source. The fuse block, to the right of the power cord rec ...

  • Teledyne 3190Z - page 22

    Installation Model 3190Z Teledyne Analytical Instruments 14 Figure 3-2 Rear Panel Electrical Connectors for AC and DC Units Analog Outputs: There are three DC output signal connectors with screw terminals on the p anel. There are two wires per output with the polarity noted. See Figure 3-3. The outputs are: 0–10 V Concentration: Voltage is propor ...

  • Teledyne 3190Z - page 23

    Trace Oxygen Analyzer Installation Teledyne Analytical Instruments 15 Alarm Relays: The three alarm circuit c onnectors are screw term inals for making connections to internal al arm relay contacts. There is one set of contacts for each type of alarm. Contacts are F orm C, with normally open and normally closed contact co nnections capable of switc ...

  • Teledyne 3190Z - page 24

    Installation Model 3190Z Teledyne Analytical Instruments 16 External Probe: The receptacle for the analys is unit cable is located in the lower left-hand corner of the rear panel. The 6-pin Minifit ™ connector is keyed to fit only one wa y into the receptacle; do not force it. The other end of the cable consists of four separate wires. These shou ...

  • Teledyne 3190Z - page 25

    Trace Oxygen Analyzer Operation Teledyne Analytical Instruments 17 Operation 4.1 Introduction Once the analyzer has been mount ed, the gas lines connected, and the electrical connections made, the an alyzer can be configured for your application. This involves setting th e system param eters as follows: • Defining the user-selec table analysis ra ...

  • Teledyne 3190Z - page 26

    Operation Model 3190Z Teledyne Analytical Instruments 18 4.2 Using the Function a nd Dat a Entry Buttons When one of the function buttons is being pressed, the analyzer is in the Setup mode or the Calibration mode. The four setup function butt ons on the analyzer are: • SET HI ALARM • SET LO ALARM • SET HI RANGE • SET LO RANGE The Calibrati ...

  • Teledyne 3190Z - page 27

    Trace Oxygen Analyzer Operation Teledyne Analytical Instruments 19 4.3.1 HI Range Setting the HI Range full-scale value defines the LEAST sensitiv e analysis range to be used. To set the HI Range: 1. Press the SET HI RANGE function button once. 2. Immediately (within 5 seconds) press either the Δ or ∇ button to raise or lower the displayed value ...

  • Teledyne 3190Z - page 28

    Operation Model 3190Z Teledyne Analytical Instruments 20 1. Press the SET LO ALARM function button once. 2. Within 5 seconds press either the Δ or ∇ button to raise or lower the displayed value as require d, until the disp lay reads the desired concentration. 4.4.3 Sensor Fail Alarm Not active in this product. 4.5 Selecting a Fixed Range or Auto ...

  • Teledyne 3190Z - page 29

    Trace Oxygen Analyzer Operation Teledyne Analytical Instruments 21 3. Immediately (within 5 seconds) press either the Δ or ∇ button until the display is stable and disp lays the oxygen concentration of the applied calibration gas. Once s et, the unit will enter Analyze Mode after approximately three seconds. The unit is now calibrated. If the ou ...

  • Teledyne 3190Z - page 30

    Operation Model 3190Z Teledyne Analytical Instruments 22 (This page intentionally left blank) ...

  • Teledyne 3190Z - page 31

    Trace Oxygen Analyzer Maintenance Teledyne Analytical Instruments 23 Maintenance Aside from normal cleaning and checking for leaks at the gas connections, the Model 3190Z shoul d not require any maintenance beyond replacement of zirconium oxid e senso rs, and perhaps a blown fuse. Routine maintenance include s occasional recalibration, as described ...

  • Teledyne 3190Z - page 32

    Maintenance Model 3190Z Teledyne Analytical Instruments 24 3. Insert a small flat-blade screwdriv er into the slot in the receptacle wall nearest the fuse and gently pr y open the fuse receptacle. The fuse holder will slide out. The fuse in use is visible in the clip. To open the spare fuse compartment, push on one end until it slides out. 4. Remov ...

  • Teledyne 3190Z - page 33

    Trace Oxygen Analyzer Appendix Teledyne Analytical Instruments 25 Appendix A.1 S pecifications Ranges : Two user selectable ranges can be set between 0-10 ppm and 0-1,000 ppm oxygen. Default ranges are 0-100 ppm and 0-1,000 ppm oxygen Signal Output : Voltage: 0–10 VDC, negative ground Current: 4-20 mA, negative ground Range ID: 0-10 VDC Display : ...

  • Teledyne 3190Z - page 34

    Appendix Model 3190Z Teledyne Analytical Instruments 26 A.2 S p are Part s List 1* C-64586A PC Board, Power Supply (AC-version) 1* C-64586B PC Board, Power Supply (DC-version) 1 C-76491C PCB Board, Main 1 C-76490C Interface Unit 2* F-1130 Fuse (AC), ½A, 250 VAC IEC Type T, 5x20mm 1* F-51 Fuse (DC), ½A, 125 VDC Micro-fuse 1 S-1439 Sensor, 0-1,000 ...

  • Teledyne 3190Z - page 35

    Trace Oxygen Analyzer Appendix Teledyne Analytical Instruments 27 A.3 Drawing List D-76533 Outline/Interconnection Diagram A.4 Miscellaneous The symbol: ~ is used on the re ar panel of the m odel 3190Z to signify volts alternating current (VAC). ...

Produzent Teledyne Kategorie Respiratory Product

Dokumente, die wir vom Produzenten des Geräts Teledyne 3190Z erhalten, können wir in mehrere Gruppen teilen. Unteranderem in:
- technische Zeichnungen Teledyne
- Bedienungsanleitungen 3190Z
- Produktkarten Teledyne
- Informationsbroschüren
- oder Energieetiketten Teledyne 3190Z
Jede von ihnen ist wichtig, jedoch finden wir die wichtigsten Informationen für den Nutzer des Geräts in der Bedienungsanleitung Teledyne 3190Z.

Die Dokumentengruppe, die als Bedienungsanleitungen bezeichnet wird, wird ebenfalls in detaillierte Arten geteilt, solche wie: Montageanleitungen Teledyne 3190Z, Wartungsanleitungen, Kurzanleitungen oder Benutzeranleitungen Teledyne 3190Z. Abhängig vom Bedarf, sollten Sie das Dokument finden, das Sie brauchen. In unserem Service können Sie sich die populärste Bedienungsanleitung des Produkts Teledyne 3190Z ansehen.

Ähnliche Bedienungsanleitungen

Die komplette Bedienungsanleitung des Geräts Teledyne 3190Z, wie sollte sie aussehen?
Die Bedienungsanleitung, auch bezeichnet als Benutzerhandbuch, oder einfach nur „Anleitung”, ist ein technisches Dokument, das dem Benutzer bei der Nutzung von Teledyne 3190Z hilfreich sein soll. Die Bedienungsanleitungen werden in der Regel von technischen Schriftstellern geschrieben, aber in einer Sprache, die für alle Nutzer von Teledyne 3190Z verständlich ist.

Eine gänzliche Bedienungsanleitung von Teledyne sollte einige Grundelemente enthalten. Ein Teil von ihnen ist nicht so wichtig, wie z.B.: die Titelseite oder Autorenseiten. Die restlichen von ihnen jedoch, sollten Informationen liefern, die für den Nutzer von enormer Wichtigkeit sind.

1. Einführung und Hinweise, wie man sich in einer Bedienungsanleitung von Teledyne 3190Z bewegt - Am Anfang jeder Bedienungsanleitung sollten wir Hinweise bezüglich der Nutzungsart eines bestimmten Ratgebers finden. In ihr sollten sich Informationen über die Lokalisierung des Inhaltsverzeichnisses von Teledyne 3190Z befinden, FAQ oder über oft auftretende Probleme – also Stellen, die von den Benutzern in jeder Bedienungsanleitung am meisten gesucht werden
2. Inhaltsverzeichnis - Index aller Ratschläge bezüglich Teledyne 3190Z, die wir im aktuellen Dokument finden
3. Ratschläge zur Nutzung der Grundfunktionen des Geräts Teledyne 3190Z - die uns die ersten Schritte während der Nutzung von Teledyne 3190Z erleichtern sollten
4. Troubleshooting - geordneter Tätigkeitslauf, der uns bei der Diagnose und als nächstes bei der Lösung wichtiger Probleme mit Teledyne 3190Z hilft
5. FAQ - häufig gestellte Fragen
6. Kontaktdaten Informationen darüber, wo man Kontakt zum Produzenten / Service von Teledyne 3190Z im bestimmten Land suchen kann, wenn es nicht gelingt, das Problem selbst zu lösen.

Haben Sie eine Frage bezüglich Teledyne 3190Z?

Nutzen Sie das untere Formular

Wenn Sie mit Hilfe der gefundenen Bedienungsanleitung Ihr Problem mit Teledyne 3190Z nicht gelöst haben, stellen Sie eine Frage, indem Sie das untere Formular nutzen. Wenn einer der Nutzer ein ähnliches Problem mit Teledyne 3190Z hatte, ist es möglich, dass er mit Ihnen die Lösung teilen möchte.

Text vom Bild übertragen

Kommentare (0)