    Teledyne Analytical Instruments OPERA TING INSTRUCTIONS FOR Model 9060L Percent Oxygen Analyzer DANGER Toxic gases and or flammable liquids may be present in this monitoring system. Personal protective equipment may be requi red when servicing this inst rume nt. Hazardous voltages exist on certain co m ponents internally which may persist for a tim ...

  • Teledyne Percent Oxygen Analyzer - page 2

    Model 9060L Teledyne Analytical Instruments ii Copyright © 2005 Teledyne Instruments/ Analytical Instruments All Rights Reserved. No part of this manual may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any other language or computer language in whole or in part, in any form or by any means, whether it b ...

  • Teledyne Percent Oxygen Analyzer - page 3

    Percent Oxygen Analyzer Teledyne Analytical Instruments iii S pecific Model Information Instrument Serial Number: _______________________ Instrument Range: _______________ Calibrated for: _______________ Background Gas: _______________ Zero Gas: _______________ Span Gas: _______________ ...

  • Teledyne Percent Oxygen Analyzer - page 4

    Model 9060L Teledyne Analytical Instruments iv Safety Messages Your safety and the safety of others is very important. We have provided many important safety m essages in this manual. Please read these messages carefully. A safety message alerts you to pot ential hazards that could hurt you or others. Each safety message is a ssociated with a safet ...

  • Teledyne Percent Oxygen Analyzer - page 5

    Percent Oxygen Analyzer Teledyne Analytical Instruments v IF YOU USE THE ANALYZER IN A MANNER OTHER THAN THAT FOR WHICH IT WAS INTEND ED, UNPREDICTABLE BEHAVI OR COULD RESULT POSSIBLY ACCOMPANIED WITH HAZARDO US CONSEQUENCES. This manual provides informati on designed to guide you through the installation, calibration operati on and maintenance of ...

  • Teledyne Percent Oxygen Analyzer - page 6

    Model 9060L Teledyne Analytical Instruments vi T able of Content s List of Fi gures ............................................................................ viii List of Tables ................................................................................ ix Introducti on ....................................................................... ...

  • Teledyne Percent Oxygen Analyzer - page 7

    Percent Oxygen Analyzer Teledyne Analytical Instruments vii 3.4.1 Oxygen Pr obe Connections 28 3.5 Installation Checklist 29 Operation ..................................................................................... 31 4.1 Introduction 31 4.2 Using the Function and Data Entry Buttons 32 4.3 Setting the Analysis Ranges 32 4.3.1 HI Range 33 4.3. ...

  • Teledyne Percent Oxygen Analyzer - page 8

    Model 9060L Teledyne Analytical Instruments viii List of Figures Figure 1-1: Front P anel ................................................................ 12 Figure 1-2 Rear Panel ................................................................... 14 Figure 2-1: Model 9060L Oxygen Probe ....................................... 16 Figure 2-2: Model 9 ...

  • Teledyne Percent Oxygen Analyzer - page 9

    Percent Oxygen Analyzer Teledyne Analytical Instruments ix List of T ables Table 2-1: Pr obe Opti ons .............................................................. 17 ...

  • Teledyne Percent Oxygen Analyzer - page 10

    Model 9060L Teledyne Analytical Instruments x DANGER COMBUSTIBLE GAS USAGE WARNING This is a general purpose instrument designed for use in a non-hazardous area. It is th e customer's responsibility to ensure safety especially when combustible gases are being analyzed since the potential of gas leaks always exist. The customer should ensure th ...

  • Teledyne Percent Oxygen Analyzer - page 11

    Percent Oxygen Analyzer Introduction Teledyne Analytical Instruments 11 Introduction 1.1 Overview The Teledyne Instruments/ Analyt ical Instruments (TI/AI) Model 9060L is a microprocessor-based per cent oxygen analyzer for real-time measurement of the percent of oxygen in inert gases, or in a wide variety of gas mixtures. It features simp le operat ...

  • Teledyne Percent Oxygen Analyzer - page 12

    Introduction Model 9060L Teledyne Analytical Instruments 12 • Two concentration alarms w ith adjustable setpoints. • A failure alarm. • Three analog outputs: two for measurement (0–10VDC, and negative ground 4–20 mA DC) and one for range identification (0-10VDC). • Compact and rugged control unit with flush-panel case. Designed for indo ...

  • Teledyne Percent Oxygen Analyzer - page 13

    Percent Oxygen Analyzer Introduction Teledyne Analytical Instruments 13 • Set Alarm 2 Sets the concentration at which alarm 2 activates upon dropping below the threshold (LO alarm). • Set HI Range Set the high analysis range for the instrument (up to 0-100%). • Set LO Range Set the low analysis range for the instrument (down to 0-1%). • Spa ...

  • Teledyne Percent Oxygen Analyzer - page 14

    Introduction Model 9060L Teledyne Analytical Instruments 14 Figure 1-2 Rear Panel • Power Connection 110 VAC at 50/60 Hz. The connector housing includes the fuse holder and the power switch. (220VAC is optional via setting an internal switch). Fuse Holder: Replacing the fuse is described in Chapter 5, Maintenance. I/O Power Switch: Tur ns the ins ...

  • Teledyne Percent Oxygen Analyzer - page 15

    Percent Oxygen Analyzer Operational Theory Teledyne Analytical Instruments 15 Operational Theory 2.1 Introduction The analyzer is composed of two subsystems: 1. Oxygen probe 2. Control unit with signal pro cessing, display and controls The oxygen probe can be heated or unheated. It delivers the sample gas to a state-of-the-art zirconium se nsor whi ...

  • Teledyne Percent Oxygen Analyzer - page 16

    Operational Theory Model 9060L Teledyne Analytical Instruments 16 All Teledyne oxygen probe or sensors are desi gned and manufactured to exacting standards of performance and reliabil ity and are the result of extensive re search and development. Teledyne Analytical Instruments provides worldwide applica tion and after sales support for its oxygen ...

  • Teledyne Percent Oxygen Analyzer - page 17

    Percent Oxygen Analyzer Operational Theory Teledyne Analytical Instruments 17 Probes of 1000 mm (40”) normally have sufficient length for any installation. Customers requiring pr obes longer than 1500 mm (59”) are supplied with a flow guide tube that uses the flue velocity to pull flue gas through a filter at the sensing tip and exhaust it near ...

  • Teledyne Percent Oxygen Analyzer - page 18

    Operational Theory Model 9060L Teledyne Analytical Instruments 18 2.2.2 The Zirconia Sensor Input to the analyzer is obtained from a solid electrolyte oxygen probe containing a zirconia element and thermocouple. The probe is designed to be inserted into a boile r or furnace exit gas flue or similar process. The 9060LEX sensor is designed to be inst ...

  • Teledyne Percent Oxygen Analyzer - page 19

    Percent Oxygen Analyzer Operational Theory Teledyne Analytical Instruments 19 require a heater. When the outside a nd inside of the sensor are exposed to different oxygen partial pressures, an EMF (E) is developed which obeys the Nernst equation: ⎟ ⎟ ⎠ ⎞ ⎜ ⎜ ⎝ ⎛ = Outside ) (PO Inside ) (PO log 4F RT ts) E(millivol 2 2 e where T is ...

  • Teledyne Percent Oxygen Analyzer - page 20

    Operational Theory Model 9060L Teledyne Analytical Instruments 20 THE PROBE/SENSOR MUST ALWAYS BE CONNECTED TO EART H. Power to the heater is supplied d irectly from the m ain power source, and the temperature is ini tially controlled a bove 700°C (1300°F) after power up. 2.3 Applications Not at Atmospheric Pressure For high and low pressure appl ...

  • Teledyne Percent Oxygen Analyzer - page 21

    Percent Oxygen Analyzer Operational Theory Teledyne Analytical Instruments 21 located on the Main PCB, which is mounted vertically, just behind the front panel of the control unit. 2.5.2 Signal Processing Figure 2-3 is a block diagram of the signal processing electronics described below. Figure 2-3: Signal Proc essing Block Diagram In the presence ...

  • Teledyne Percent Oxygen Analyzer - page 22

    Operational Theory Model 9060L Teledyne Analytical Instruments 22 The digital concentration signal al ong with input from the front panel buttons (KEYBOARD) is pro cessed by the micro-controller, and appropriate output signals are directed to the display and alarm relays. The same digital inform ation is also sent to a 12-bit digital to analog conv ...

  • Teledyne Percent Oxygen Analyzer - page 23

    Percent Oxygen Analyzer Installation Teledyne Analytical Instruments 23 Inst allation Installation of the analyzer includes: 1. Unpacking the system. 2. Mounting the control unit. 3. Installing the probe. 4. Making electrical connections. 5. Testing the installation. CAUTION: READ THIS CHAPTER IN ITS ENTIRETY BEFORE INSTALLING THE SYSTEM. FOR INDOO ...

  • Teledyne Percent Oxygen Analyzer - page 24

    Installation Model 9060L Teledyne Analytical Instruments 24 Note: Mount the probe as close to the combustion source as possible. This reduces the lag time inherent in the sampling process and results in better control. The probe has a typical response time of less than four seconds, so most of the delay time is associa ted with the transit time of ...

  • Teledyne Percent Oxygen Analyzer - page 25

    Percent Oxygen Analyzer Installation Teledyne Analytical Instruments 25 probe or install a longer probe. This is normally caused by stratification of the flue gas. Figure 3-1: Oxygen Probe Mounting 3.4 Electrical Connections Figure 3-1 shows the Model 9060L rear panel. For detailed pinouts, see the wiring/interconnection drawings in the Drawings se ...

  • Teledyne Percent Oxygen Analyzer - page 26

    Installation Model 9060L Teledyne Analytical Instruments 26 Figure 3-2: Model 9060L Rear panel 3.4.2 Analog Outputs There are three DC output signal connectors with screw terminals on the panel. There are two wires pe r output with the polarity noted. See Figure 3-3. The outputs are: 0–10V % Range: Voltage rises with increasing oxygen concentrati ...

  • Teledyne Percent Oxygen Analyzer - page 27

    Percent Oxygen Analyzer Installation Teledyne Analytical Instruments 27 The alarm relay circuits are designed for failsafe operation, meaning the relays are energized dur ing normal operation. If power fails the relays de-energize (alarm s activated). The contact connections are indica ted diagram matically on the rear panel as Normally Closed, Com ...

  • Teledyne Percent Oxygen Analyzer - page 28

    Installation Model 9060L Teledyne Analytical Instruments 28 3.4.1 Oxygen Pr obe Connections The receptacle for the probe cable is located in the lower left-hand corner of the rear panel. The 6-pin polarized connector is keyed to fit only one way into the receptacle. Do not force it in. The other end of the cable connects to the probe assem bly . Se ...

  • Teledyne Percent Oxygen Analyzer - page 29

    Percent Oxygen Analyzer Installation Teledyne Analytical Instruments 29 Figure 3-5: Connections to Probe 3.5 Inst allation Checklist Prior to powering up the unit make sure you have: • Installed the unit correc tly • Checked the probe and welded probe socket for leaks Once the above checks have been made, you can connect the power source. Wait ...

  • Teledyne Percent Oxygen Analyzer - page 30

    Installation Model 9060L Teledyne Analytical Instruments 30 ...

  • Teledyne Percent Oxygen Analyzer - page 31

    Percent Oxygen Analyzer Operation Teledyne Analytical Instruments 31 Operation 4.1 Introduction Once the analyzer has been mounted , the probe installed, and the electrical connections made, the an alyzer can be configured for your application. This involves se tting the system parameters: • Defining the user select able analysis ranges. • Sett ...

  • Teledyne Percent Oxygen Analyzer - page 32

    Operation Model 9060L Teledyne Analytical Instruments 32 4.2 Using the Function a nd Dat a Entry Buttons When no buttons on the analyzer ar e being pressed, the instrument is in the analyze mode. It is mon itoring the percent of oxygen in the sample gas that is flowing through the remote probe. When one of the function buttons is being pressed, the ...

  • Teledyne Percent Oxygen Analyzer - page 33

    Percent Oxygen Analyzer Operation Teledyne Analytical Instruments 33 range full scale value, and it switches back to the LO range when the oxygen concentration reaches 85% of the LO range full scale value Note: The HI Range setpoint MUST be set at a higher concentration percentage than the LO Range setpoint. 4.3.1 HI Range Setting the HI Range full ...

  • Teledyne Percent Oxygen Analyzer - page 34

    Operation Model 9060L Teledyne Analytical Instruments 34 until the display reads the desired per cent concentration. If within 5 seconds no key is pressed, the instrum ent will return to the sample mode and display oxygen concentration. After setting the valu e wait for the un it to time out of this mode (approximately 5 seconds) and return to disp ...

  • Teledyne Percent Oxygen Analyzer - page 35

    Percent Oxygen Analyzer Operation Teledyne Analytical Instruments 35 2. Immediately (within 5 seconds) press either the Δ or ∇ button until Auto, Lo, or Hi displays on the LCD , as desired. After about three seconds, the an alyzer resumes monitoring in the selected range mode. 4.6 Oxygen Calibration There are two types of calibration one is air ...

  • Teledyne Percent Oxygen Analyzer - page 36

    Operation Model 9060L Teledyne Analytical Instruments 36 for five minutes before timing- out. During this five minute interval, the LED will conti nue to track gas through the analyzer. 2. Flow span gas of a known oxygen concentration (between 1 and 16% oxygen) into the analyzer. Th e analyzer will begin to track the span gas. Note: The alarms will ...

  • Teledyne Percent Oxygen Analyzer - page 37

    Percent Oxygen Analyzer Operation Teledyne Analytical Instruments 37 4.6.2 Air Calibration To perform air calibration, make sure you have air flowing through the same inlet you use for span calibration at a flow of 1- 5 liters per minute. Wait for readings to stabilize. Press both the SET ALARM 1 and SET ALARM 2 simultaneously. The LED display will ...

  • Teledyne Percent Oxygen Analyzer - page 38

    Operation Model 9060L Teledyne Analytical Instruments 38 Within the time the tem perature is being displayed, one can use the UP and DOWN arrow keys to adj ust the temperature, thus recalib rating the temperature readout. WARNING: TEMPERATURE READI NG IS CALIBRATED AT THE FACTORY. DO NOT ATTEMPT RECALIBRATION OF THE PROBE TEMPERATURE. ERROR OF A FE ...

  • Teledyne Percent Oxygen Analyzer - page 39

    Percent Oxygen Analyzer Operation Teledyne Analytical Instruments 39 output of the sensor higher than the top scale of the ADC. It may also occur if the thermocouple should open. -03- This message is displayed when the input voltage to the ADC has fallen below the low er negative limit of the ADC. ...

  • Teledyne Percent Oxygen Analyzer - page 40

    Operation Model 9060L Teledyne Analytical Instruments 40 ...

  • Teledyne Percent Oxygen Analyzer - page 41

    Percent Oxygen Analyzer Maintenance Teledyne Analytical Instruments 41 Maintenance Aside from normal cleaning and ch ecking for leaks at the probe connection, the Model 9060L shoul d not require any maintenance beyond replacement of a blown fuse. Routine maintenance includes occasional recalibration, as described in Chapter 4, Operation . 5.1 Repla ...

  • Teledyne Percent Oxygen Analyzer - page 42

    Maintenance Model 9060L Teledyne Analytical Instruments 42 3. Insert a small flat-blade s crewdrive r into the slot in the receptacle wall nearest the fuse and gently pr y open the fuse recep tacle. The fuse holder will slide out. There are two f uses in use and are visible in the clip. 4. Remove the bad fuse and replace it with a 5x20mm 2A, 250 VA ...

  • Teledyne Percent Oxygen Analyzer - page 43

    Percent Oxygen Analyzer Appendix Teledyne Analytical Instruments 43 Appendix A.1 S pecifications Ranges : 0-3 % and 0-10 % oxygen (Standard Ranges), and 0-25 % (nominal) Calibration Range. User selectable % Ranges can be set between 1% and 100 % oxygen . Signal Output : Voltage: 0–10 VDC, negative ground Current: 4-20 mA, negative ground Range ID ...

  • Teledyne Percent Oxygen Analyzer - page 44

    Appendix Model 9060L Teledyne Analytical Instruments 44 A. 2 Probe Sp ecifications Application: Combustion flue below 700°C (1290°F), above 700°C (1290°F) 900°C (1650°F) — no contaminants. With contaminants see notes 1 &2 Temperature Range: 0-900°C 800-1400°C 800-1200°C (32-1650°F) (1470-2550° F) (1470-2190°F) Length: 250-1000 mm ...

  • Teledyne Percent Oxygen Analyzer - page 45

    Percent Oxygen Analyzer Appendix Teledyne Analytical Instruments 45 Notes: 1. Care must be taken to avoid contact with explosive or inflammable gases with 9060L heated probes when hot. 2. Please contact factory for corrosive s other than sulphur or zinc. We can provide test materials to try in your atmosphere. A.3 S p are Part s List QTY. P/N DESCR ...

  • Teledyne Percent Oxygen Analyzer - page 46

    Appendix Model 9060L Teledyne Analytical Instruments 46 A.4 Reference Drawing D-77712 Outline Diagram D77713 Control Unit Final Assembly C77925 Interconnection Diagram A.5 Probe or Sensor EMF T ables ZIRCONIA OXYGEN SENSO R OUTPUT (mV) PROBE TYPE 9060L, SENSOR TYPE 9060LEX % OXYGEN mV at 720°C (1320°F) 21.0 0.00 20.5 0.46 20.0 0.99 19.5 1.53 19.0 ...

  • Teledyne Percent Oxygen Analyzer - page 47

    Percent Oxygen Analyzer Appendix Teledyne Analytical Instruments 47 11.0 13.78 10.5 14.78 10.0 15.82 9.5 16.92 9.0 18.08 8.5 19.30 8.0 20.60 7.5 21.98 7.0 23.45 6.5 25.04 6.0 26.75 5.5 28.61 5.0 30.65 4.5 32.90 4.0 35.42 3.5 38.28 3.0 41.58 2.5 45.48 2.0 50.25 1.5 56.41 1.0 65.08 0.5 79.91 0.2 99.51 'K’ TC mV 29.212 at 720 °C (1320 °F) The ...

  • Teledyne Percent Oxygen Analyzer - page 48

    Appendix Model 9060L Teledyne Analytical Instruments 48 ...

  • Teledyne Percent Oxygen Analyzer - page 49

    Percent Oxygen Analyzer Index Teledyne Analytical Instruments 49 Index accuracy, 43 address, 45 air calibration, 35 , 37 air purge, 16 alaram delay, 36 alarm, 43 concentration, 12 HI, 27 LO, 27 sensor fail, 27, 34 sensor failure, 12 setpoint, 33 threshold, 1 2, 27 Alarm 1, 27 Alarm 2, 27 alarm connection, 26 alarm connecti ons, 14 alarm delay , 37 ...

  • Teledyne Percent Oxygen Analyzer - page 50

    Index Model 9060L Teledyne Analytical Instruments 50 input connector, 13 insertion length, 17 instasllation checklist, 29 instrument ran ge, 11 internal power switch, 20, 25 lagging e xtension, 17 leak check, 29 LED, 13 LED flashing, 35 LO range, 33 main PCB, 21 maintenance, 41 manuals, a dditional, v Measurement Parameters Readout, 13 membrane swi ...

  • Teledyne Percent Oxygen Analyzer - page 51

    Percent Oxygen Analyzer Index Teledyne Analytical Instruments 51 tables listing, ix technician symbol, iv temperature, 44 temperature co ntroller, 18 temperature maximum , 24 temperature too low, 38 thermal shoc k, 15, 24 thermocouple, 18 thermocouple t ype, 44 time out, 3 4 time-out , 37 t-Lo, 38 transit time, 24 U length, 24 Up Arrow, 13 voltage ...

Produzent Teledyne Kategorie Oxygen Equipment

Dokumente, die wir vom Produzenten des Geräts Teledyne Percent Oxygen Analyzer erhalten, können wir in mehrere Gruppen teilen. Unteranderem in:
- technische Zeichnungen Teledyne
- Bedienungsanleitungen Percent Oxygen Analyzer
- Produktkarten Teledyne
- Informationsbroschüren
- oder Energieetiketten Teledyne Percent Oxygen Analyzer
Jede von ihnen ist wichtig, jedoch finden wir die wichtigsten Informationen für den Nutzer des Geräts in der Bedienungsanleitung Teledyne Percent Oxygen Analyzer.

Die Dokumentengruppe, die als Bedienungsanleitungen bezeichnet wird, wird ebenfalls in detaillierte Arten geteilt, solche wie: Montageanleitungen Teledyne Percent Oxygen Analyzer, Wartungsanleitungen, Kurzanleitungen oder Benutzeranleitungen Teledyne Percent Oxygen Analyzer. Abhängig vom Bedarf, sollten Sie das Dokument finden, das Sie brauchen. In unserem Service können Sie sich die populärste Bedienungsanleitung des Produkts Teledyne Percent Oxygen Analyzer ansehen.

Ähnliche Bedienungsanleitungen

Die komplette Bedienungsanleitung des Geräts Teledyne Percent Oxygen Analyzer, wie sollte sie aussehen?
Die Bedienungsanleitung, auch bezeichnet als Benutzerhandbuch, oder einfach nur „Anleitung”, ist ein technisches Dokument, das dem Benutzer bei der Nutzung von Teledyne Percent Oxygen Analyzer hilfreich sein soll. Die Bedienungsanleitungen werden in der Regel von technischen Schriftstellern geschrieben, aber in einer Sprache, die für alle Nutzer von Teledyne Percent Oxygen Analyzer verständlich ist.

Eine gänzliche Bedienungsanleitung von Teledyne sollte einige Grundelemente enthalten. Ein Teil von ihnen ist nicht so wichtig, wie z.B.: die Titelseite oder Autorenseiten. Die restlichen von ihnen jedoch, sollten Informationen liefern, die für den Nutzer von enormer Wichtigkeit sind.

1. Einführung und Hinweise, wie man sich in einer Bedienungsanleitung von Teledyne Percent Oxygen Analyzer bewegt - Am Anfang jeder Bedienungsanleitung sollten wir Hinweise bezüglich der Nutzungsart eines bestimmten Ratgebers finden. In ihr sollten sich Informationen über die Lokalisierung des Inhaltsverzeichnisses von Teledyne Percent Oxygen Analyzer befinden, FAQ oder über oft auftretende Probleme – also Stellen, die von den Benutzern in jeder Bedienungsanleitung am meisten gesucht werden
2. Inhaltsverzeichnis - Index aller Ratschläge bezüglich Teledyne Percent Oxygen Analyzer, die wir im aktuellen Dokument finden
3. Ratschläge zur Nutzung der Grundfunktionen des Geräts Teledyne Percent Oxygen Analyzer - die uns die ersten Schritte während der Nutzung von Teledyne Percent Oxygen Analyzer erleichtern sollten
4. Troubleshooting - geordneter Tätigkeitslauf, der uns bei der Diagnose und als nächstes bei der Lösung wichtiger Probleme mit Teledyne Percent Oxygen Analyzer hilft
5. FAQ - häufig gestellte Fragen
6. Kontaktdaten Informationen darüber, wo man Kontakt zum Produzenten / Service von Teledyne Percent Oxygen Analyzer im bestimmten Land suchen kann, wenn es nicht gelingt, das Problem selbst zu lösen.

Haben Sie eine Frage bezüglich Teledyne Percent Oxygen Analyzer?

Nutzen Sie das untere Formular

Wenn Sie mit Hilfe der gefundenen Bedienungsanleitung Ihr Problem mit Teledyne Percent Oxygen Analyzer nicht gelöst haben, stellen Sie eine Frage, indem Sie das untere Formular nutzen. Wenn einer der Nutzer ein ähnliches Problem mit Teledyne Percent Oxygen Analyzer hatte, ist es möglich, dass er mit Ihnen die Lösung teilen möchte.