    Teledyne Analytical Instruments HIGHL Y TO XIC AND OR FLAMMABLE LIQUIDS OR GASES MA Y BE PRESENT IN THIS MONIT ORING SYSTEM. PERSONAL PRO TECTIVE EQUIPMENT MA Y BE REQUIRED WHEN SER VICING THIS SYSTEM. HAZARDOUS V OL T AGES EXIST ON CER T AIN COMPONENTS INTERNALL Y WHICH MA Y PERSIST FOR A TIME EVEN AFTER THE PO WER IS TURNED OFF AND DISCONNECTED . ...

  • Teledyne ECO#02-0227 - page 2

    INSTRUCTION MANUAL FOR TRACE OXYGEN ANALYZER MODEL 311XL TELEDYNE ANALYTICAL INSTRUMENTS 16830 CHESTNUT STREET CITY OF INDUSTRY, CA 91749 TELEPHONE: (888) 789-8168 (626) 934-1500 (626) 961-9221 FAX: (626) 961-2538 (626) 934-1651 Web: www.teledyne-ai.com ...

  • Teledyne ECO#02-0227 - page 3

    WARNING The sensor(s) used in this instrument uses electrolytes which contain substances that are extremely harmful if touched, swallowed, or i nhaled. Avoid contact w ith ANY fluid or powder in or around the unit. What may appear to be plain water could contain one of these toxic substances. SKIN CONTACT: Flush contact area with cold water for sev ...

  • Teledyne ECO#02-0227 - page 4

    TABLE OF CONTENTS 311XL SERIES 1. INTRODUCTION 1 1.1 Description 1 1.2 Method of Analysis 1 1.3 Outstanding Features 1 1.3.1 Micro-Fuel Cell 1 1.3.2 Reliable Calibration 2 1.3.3 Integral Power Supply 2 1.3.4 Accuracy and Response 3 1.3.5 Compact Packaging 3 2. SUPPORTING EQUIPMENT AND SERVICES 4 2.1 Sampling Equipment 4 2.2 Power Service 4 3 . O P ...

  • Teledyne ECO#02-0227 - page 5

    311XL SERIES TRACE OXYGEN ANALYZER 1. INTRODUCTION 1.1 Description : The Teledyne Analytical Instruments (TAI) Model 311XL and 311TCXL are portable, intrinsically safe trace oxygen analyzer which can be operated without an external power source. These analyzers w ill be referred to as Model 311XL or 311XL Series analyzers in this manual. The instru ...

  • Teledyne ECO#02-0227 - page 6

    1.3.2 Reliable Calibration . Although this instrument can be calibrated with air, the use of a calibration gas gives better results on the lower PPM ranges. If the instrument is exposed to high concentrations of oxygen (percent levels), it will take a significant amount of time to purge the oxygen from the sensor a nd analyzer for use in the low PP ...

  • Teledyne ECO#02-0227 - page 7

    A current limiting resistor (R1 and R2) is potted into the end of each battery. This assures that under no circumstances can more than 25 m illiamp eres (100 m illiwatts) be switched or drawn from either battery supply. This means that the Model 311XL m ay be used in explosive atmospheres where arcs of 100 milliwatts or less can be tolerated. This ...

  • Teledyne ECO#02-0227 - page 8

    2. SUPPORTING EQUIPMENT AND SERVICES 2.1 Sampling Equipment . The customer must provide a m eans of controlling the pressure and flowrate of the sample gas. For positive pressure applications, TAI suggests a simple throttle valve installed in the sample line between the sample point and the analyzer. The flowrate should be limited to between 0.5 an ...

  • Teledyne ECO#02-0227 - page 9

    3.1 Introduction . The Model 311XL is supplied completely assembled and ready for instant use. The Micro-Fuel Cell must be installed prior use. When a sensor is first installed, the instrument should be purged for 12-16 hours with zero gas (an inert gas such as nitrogen or argon with less than 0.1 ppm oxygen). This will remove any trace levels of o ...

  • Teledyne ECO#02-0227 - page 10

    Span gas: An inert gas (nitrogen or argon) with a known trace amount of oxygen. The oxygen content should be 70 – 90% of the full scale on the range of interest (working or primary range). Note, if this instrument is exposed to high concentrations of oxygen (percent levels), it will take a significant amount of time to purge the oxygen from the s ...

  • Teledyne ECO#02-0227 - page 11

    2) Establish a flowrate in the sample line of from 0.2 to 5 liters/m in. Allow the sample to flow long enough to allow a stable low reading to be achieved. When disconnecting the instrument, reverse the procedure: source f itting first, and then vent fitting. The objective of the connection -- disconnection procedure is to obviate the possibility o ...

  • Teledyne ECO#02-0227 - page 12

    that an overnight recharge be accomplished ev ery three (3) weeks of continuous use. To recharge the batteries, place the range switch in the “OFF” position and connect the power cord to a convenient outlet. NOTE: The amber charge lamp (on the rear of the case) will be lit during charging process. The integral charging circuit will automaticall ...

  • Teledyne ECO#02-0227 - page 13

    The spare cell should be carefully stored in an ar ea that is not subject to large variations in ambient temperature (75 Deg. F nom inal), and in such a way as to obviate any possibility of damage. Under no circumstances, disturb the integr ity of the cell package until the cell is to be actually used. No tools are required to replace the cell in t ...

  • Teledyne ECO#02-0227 - page 14

    5. TRANSDUCER AND TEMPERATURE COMPENSATION The Micro-Fuel Cell has an inherent positive temper ature coeff icient, the effects of which have been minimized through the im plementation of a thermistor com pensation circuit. Internal electronic calibration is accomplished by TAI. However, should there by any doubt, the following procedure can be used ...

  • Teledyne ECO#02-0227 - page 15

    -11- 7. PRODUCT SPECIFICATION DATA RANGES: 0-2, 10, 100, 1000 PPM Oxygen and Cal range for air calibration ACCURACY: ...

  • Teledyne ECO#02-0227 - page 16

    Accuracy at 25 o C and constant pressure after thermal equilibrium has been reached. Basic accuracy +/- 2% of full scale for all ranges except 0-2 PPM range. Accuracy for 0-2 PPM range (+/- 5% of full scale). Accuracy over the temperature range 0-35C at constant pressure after thermal equilibrium has been reached. Basic accuracy +/- 5% of scale for ...

  • Teledyne ECO#02-0227 - page 17

    1 F-51 MICRO-FUSE, 0.5 AMP., FAST-BLOW (FOR 100-115 VAC) 1 C6689 -B2CXL MICRO-FUEL CELL, CLASS B-2CXL 2 B-9905 BATTERY 1 L-79 LAMP 1 M70 METER. 1 B-70283 DIAL-RANGE OF 0-2, 0-10 ppm O2 & CAL. (USE WITH M70 METER) 1 A36289 CALIBRATION KIT 1 A205 OP-AMP A3 (LF441) 1 A246 OP-AMP A1 (MAX 430) A MINIMUM CHARGE IS APPLICABLE TO SPARE PARTS ORDERS. IM ...

  • Teledyne ECO#02-0227 - page 18

    C-70246 SCHEMATIC D-70268 OUTLINE C-14627 W IRING DIAGRAM (110 VAC) C-10470 W IRING DIAGRAM (220VAC) DRAWING LIST MODEL 311TCXL C-70248 SCHEMATIC D-70270 OUTLINE C-41661 W IRING DIAGRAM (110 VAC) C-37938 W IRING DIAGRAM (220VAC) NOTE: The MSDS on this material is available upon request through the Teledyne Environmental Health and Safety Coordinato ...

Produzent Teledyne Kategorie TV Converter Box

Dokumente, die wir vom Produzenten des Geräts Teledyne ECO#02-0227 erhalten, können wir in mehrere Gruppen teilen. Unteranderem in:
- technische Zeichnungen Teledyne
- Bedienungsanleitungen ECO#02-0227
- Produktkarten Teledyne
- Informationsbroschüren
- oder Energieetiketten Teledyne ECO#02-0227
Jede von ihnen ist wichtig, jedoch finden wir die wichtigsten Informationen für den Nutzer des Geräts in der Bedienungsanleitung Teledyne ECO#02-0227.

Die Dokumentengruppe, die als Bedienungsanleitungen bezeichnet wird, wird ebenfalls in detaillierte Arten geteilt, solche wie: Montageanleitungen Teledyne ECO#02-0227, Wartungsanleitungen, Kurzanleitungen oder Benutzeranleitungen Teledyne ECO#02-0227. Abhängig vom Bedarf, sollten Sie das Dokument finden, das Sie brauchen. In unserem Service können Sie sich die populärste Bedienungsanleitung des Produkts Teledyne ECO#02-0227 ansehen.

Ähnliche Bedienungsanleitungen

Die komplette Bedienungsanleitung des Geräts Teledyne ECO#02-0227, wie sollte sie aussehen?
Die Bedienungsanleitung, auch bezeichnet als Benutzerhandbuch, oder einfach nur „Anleitung”, ist ein technisches Dokument, das dem Benutzer bei der Nutzung von Teledyne ECO#02-0227 hilfreich sein soll. Die Bedienungsanleitungen werden in der Regel von technischen Schriftstellern geschrieben, aber in einer Sprache, die für alle Nutzer von Teledyne ECO#02-0227 verständlich ist.

Eine gänzliche Bedienungsanleitung von Teledyne sollte einige Grundelemente enthalten. Ein Teil von ihnen ist nicht so wichtig, wie z.B.: die Titelseite oder Autorenseiten. Die restlichen von ihnen jedoch, sollten Informationen liefern, die für den Nutzer von enormer Wichtigkeit sind.

1. Einführung und Hinweise, wie man sich in einer Bedienungsanleitung von Teledyne ECO#02-0227 bewegt - Am Anfang jeder Bedienungsanleitung sollten wir Hinweise bezüglich der Nutzungsart eines bestimmten Ratgebers finden. In ihr sollten sich Informationen über die Lokalisierung des Inhaltsverzeichnisses von Teledyne ECO#02-0227 befinden, FAQ oder über oft auftretende Probleme – also Stellen, die von den Benutzern in jeder Bedienungsanleitung am meisten gesucht werden
2. Inhaltsverzeichnis - Index aller Ratschläge bezüglich Teledyne ECO#02-0227, die wir im aktuellen Dokument finden
3. Ratschläge zur Nutzung der Grundfunktionen des Geräts Teledyne ECO#02-0227 - die uns die ersten Schritte während der Nutzung von Teledyne ECO#02-0227 erleichtern sollten
4. Troubleshooting - geordneter Tätigkeitslauf, der uns bei der Diagnose und als nächstes bei der Lösung wichtiger Probleme mit Teledyne ECO#02-0227 hilft
5. FAQ - häufig gestellte Fragen
6. Kontaktdaten Informationen darüber, wo man Kontakt zum Produzenten / Service von Teledyne ECO#02-0227 im bestimmten Land suchen kann, wenn es nicht gelingt, das Problem selbst zu lösen.

Haben Sie eine Frage bezüglich Teledyne ECO#02-0227?

Nutzen Sie das untere Formular

Wenn Sie mit Hilfe der gefundenen Bedienungsanleitung Ihr Problem mit Teledyne ECO#02-0227 nicht gelöst haben, stellen Sie eine Frage, indem Sie das untere Formular nutzen. Wenn einer der Nutzer ein ähnliches Problem mit Teledyne ECO#02-0227 hatte, ist es möglich, dass er mit Ihnen die Lösung teilen möchte.

Text vom Bild übertragen

Kommentare (0)