    - i - INSTRUCTION MANUAL HFM-I-401 AND HFM-I-405 INDUSTRIAL FLOW METERS TELEDYNE HASTINGS INSTRUMENTS ISO 9001 CER TIFIED ...

  • Teledyne HFM-I-401 - page 2

    401-405 SERIES - ii - Manual Print History The print history shown below lists the p rinting dates of all revisions and ad denda created for this manual. The revision level letter increases alpha betically as the manual unde rgoes subsequent updates. Addenda, which are released between re visions, contain i mportant change informati on that the use ...

  • Teledyne HFM-I-401 - page 3

    401-405 SERIES - iii - Qui ck Sta rt Inst ruc tion s General Information Connect dry, clean gas and ensure connections are leak free. Connect Cable for power an d analog signal output. Check that electrical connections are correct. (See diagrams below) Replace front cover and cable feed-through ensuring gasket is seated and fasteners are secure. Di ...

  • Teledyne HFM-I-401 - page 4

    401-405 SERIES - iv - ...

  • Teledyne HFM-I-401 - page 5

    401-405 SERIES - v - ...

  • Teledyne HFM-I-401 - page 6

    401-405 SERIES - vi - CAUTION This instrument is available with multiple pin-outs. Ensure electrical connections are correct. CAUTION The 400-I series flow meters are designed for IEC Installation/Over voltage Category II – single phase receptacle connected loads. NOTE The Hastings 400 Series flow meters are designed for INDOOR and OUTDOOR operat ...

  • Teledyne HFM-I-401 - page 7

    401-405 SERIES - vii - Table of Contents GENERAL INFO RMATION ............................................................................................................ .........................1 1. G ENERAL I NFORMATION ............................................................................................................................... ...

  • Teledyne HFM-I-401 - page 8

    401-405 SERIES - viii - 5. W ARRANTY ...................................................................................................................................................... 23 5.1. W ARRANTY R EPAIR P OLICY ......................................................................................................................... 23 5.2 ...

  • Teledyne HFM-I-401 - page 9

    401-405 SERIES - 1 - 1. General Infor mation 1. General Information 1.1. Overview 1.1.1. 400 Seri es Family The Hastings 400 Series is a family of flow instrument s which is specifically designed to meet the needs of the industrial gas flow market. The “I” family in the 400 Series features an IP-65 enclosure which allows the use of the instrume ...

  • Teledyne HFM-I-401 - page 10

    401-405 SERIES - 2 - • A digitally reported status of alarms an d warnings such as overflow/underflow • A flow totalizer to track the amount of ga s added to a system • A digitizing channel for an auxiliary analog signal • An internal curve fitting routine for “fine tuning” the base calibration • An alternate calibration set of 8 diff ...

  • Teledyne HFM-I-401 - page 11

    401-405 SERIES - 3 - Mechanical Fittings Standard: 1/2" Swagelok Standard: 1" Swagel ok Optional: ½" VCO®, ½" VCR® , ¾” Sw agelok, Optional: 1" VCO®,1" VCR®, ¾” Sw agelok, , 10mm Swagelok, 3/8" male NPT, ½” male NPT 1" male NPT, ¾” male NPT, 1 5/16"-12 straight 12mm Swagelok, ¾"-1 ...

  • Teledyne HFM-I-401 - page 12

    401-405 SERIES - 4 - 2. Installation 2. Installation CAUTION Many of the functions described in this section require removing the enclosure front plate. Care must be taken when rei nstalling this plate to ensure that the sealing gasket is properly positioned and the fasteners are secure to maintai n an IP65 compliant seal. 2.1. Receiving Inspection ...

  • Teledyne HFM-I-401 - page 13

    401-405 SERIES - 5 - 2.4. Mounting the Electronics Remotely CAUTION In order to maintain the integrit y of the Electrostatic Discharge immunity both parts of the remote mounted versio n of the HFM- I-400 instrument must be screwed to a well grounded struc ture. The ferrite that is shipped with the instrument must be installed on the cable next to t ...

  • Teledyne HFM-I-401 - page 14

    401-405 SERIES - 6 - There are two possible connection me thods to the ana log terminal strip. The standard me thod is by inserting a ca ble through the supplied cable gland with an external jacket that meets the specifications of the following caut ion note and tighten ing down the cable gland nut securely to se al against the cable jacket. There ...

  • Teledyne HFM-I-401 - page 15

    401-405 SERIES - 7 - the range of 5-28 Vdc from a source external to the flow meter. The loop supply c an be the same supply as that for the instrument power or it can be an isolated loop supply. Figure 2-3 shows a typical setup using the same supply . This method requires a jumper from pin 2 to pin 4 on the terminal strip while connecting pin 3 to ...

  • Teledyne HFM-I-401 - page 16

    401-405 SERIES - 8 - Figure 2-4 Wiring diagram showing the curre nt loop powered by an external supply Fi g ure 2-3 Wirin g dia g ram showin g the current loo p su pp l y p owered b y the instrument su pp l y ...

  • Teledyne HFM-I-401 - page 17

    401-405 SERIES - 9 - 2.1.1.2. Voltage out put If the flow meter is configured for a voltage output, the signal will be available as a positive potential on pin 4 relative to pin 3 of the terminal strip. Since these pins are galvanically isolated, the signal cannot be read by an indicator between pin 4 and pin 1 of the terminal strip. Pin 3 must b e ...

  • Teledyne HFM-I-401 - page 18

    401-405 SERIES - 10 - half or full duplex; and jumper 5 is enabled when a ha rdware override of the baud rate (forcing it to 9600) is desired. These functions are summarized in Figure 2-6 . 2.7.2. RS-485 If RS485 is specified on the order, the flow meter is set to the default values: a ddress 61, unterminate d Tx and Rx lines. While the default add ...

  • Teledyne HFM-I-401 - page 19

    401-405 SERIES - 11 - Since the alarms act as switches they do not produce a voltage or current signal. However, the y can be used to generate a voltage si gnal on an Alarm Out line. This is done by connecting a suitable pull-up resistor between an external voltage supply and the desired alarm line while connecting Alar m Common to the common of th ...

  • Teledyne HFM-I-401 - page 20

    401-405 SERIES - 12 - 2.2. Rotary Gas Selector The Hastings 400 Series flow meters can have up to eight different calibrations stored internally. These are referred to as gas records. These records are use d to select different gases, but they can al so be useful in other ways; for instance reporting the flow in an al ternate range, flow uni t or r ...

  • Teledyne HFM-I-401 - page 21

    401-405 SERIES - 13 - Example 2- Changing the active gas record Selecting the active gas record is accomplished in one of two ways: 1. Hardware setting 2. Software setting Hardware: The hardware setting is selected by accessing a rota ry encoder on the upper PC board in the electronics enclosure. When set to a number po sitio n from 0 to 7 it activ ...

  • Teledyne HFM-I-401 - page 22

    401-405 SERIES - 14 - ...

  • Teledyne HFM-I-401 - page 23

    401-405 SERIES - 15 - 3. Operation 3. Operation The Hastings 400 Series flow meters are designed for operation with clean dry gas and in specified environmental conditions (See Section 1.2). The prop erly installed meter measures and reports the mass flow as an analog signal and, depending on the configuration and set up , as a digital response. Ot ...

  • Teledyne HFM-I-401 - page 24

    401-405 SERIES - 16 - 3.3.1. Digitally Reported Flow Output The flow rate can be read digitally by sending an ascii “F” command (preceded by the address for RS- 485). The instrument will respond with an ascii representation of the num erical value of the flow rate in the units of flow specified on the nameplate label. Example: A meter with RS-2 ...

  • Teledyne HFM-I-401 - page 25

    401-405 SERIES - 17 - 3.4.2. Adjusting Zero The pre-conditions required for a zero check must als o be followed when making a zero adjust ment. The zero adjustment is a di gitally controlled “r eset” type operation. Wh en commanded, the meter initiates an internal routine that performs the fo llowing seque nce: measure th e current flow reading ...

  • Teledyne HFM-I-401 - page 26

    401-405 SERIES - 18 - 3.4.2.3.6. Reverse Flow If the flow through the flow meter reverses and flow begins to enter the exit of the flow meter and leave through the entrance of the flow meter the flow meter will measure this flow and report it digitally with reduced accuracy. The analog output will also indicate this by either generating a negative ...

  • Teledyne HFM-I-401 - page 27

    401-405 SERIES - 19 - 3.7.1. Zero Shift The zero offset can occur as the result of natural convection flow through the sensor tube if the instrument is not mounted in a level orientation wi th flow horizontal. This natural convection effect causes a zero shift proportional to the system pressu re . The overall effect is more pronounced for gases wi ...

  • Teledyne HFM-I-401 - page 28

    401-405 SERIES - 20 - 3.10. Flow Totalization The Hastings 400 Series flow meters are capable of pr oviding a value for the “total amount of gas” that has passed through the flow meter since t he last time the totalization function wa s reset. This value can be used to determine for example, the amount of gas used to fill a chamber or drawn fro ...

  • Teledyne HFM-I-401 - page 29

    401-405 SERIES - 21 - 4. P ar ts and Accessor ies 4. Parts & accessories These are parts and acces sories that are available b y separate order from Teledyne Hastin gs Instruments. 4.1. Power Pod – Power & Display units THPS-100 Singel Channel Power Supply The Teledyne Hastings Inst rume nts microprocessor based Power Pod -100 Thermal Mas ...

  • Teledyne HFM-I-401 - page 30

    401-405 SERIES - 22 - 4.1. Fittings Fittings Hastings # HFM-I-401 1/2" Swagelok Fittings 41-03-086 1/2" VCO Fittings 41-03-119 1/2" VCR Fittings 41-03-090 3/4" Swagelok Fitting 41-03-152 10 mm Swagelok 41-03-153 3/8" Male NPT 41-03-154 1/2" Male NPT 41-03-155 12 mm Swagelok 41-03-160 3/4-16 SAE/MS Straight Female (no f ...

  • Teledyne HFM-I-401 - page 31

    401-405 SERIES - 23 - 5. WA R R A N T Y 5. Warranty 5.1. Warranty Repair Policy Hastings Instruments warrants this product for a period of one year from the date of shipment to be free from defects in material and workmanship. This w arr anty does not apply to de fects or failures resulting from unauthorized modification, misuse or mishan dling of ...

  • Teledyne HFM-I-401 - page 32

    401-405 SERIES - 24 - 6. Appendices 6. Appendices 6.1. Appendix 1- Volumetric versus Mass Flow Mass flow measures just what it says, the mass or weight of the gas flowing through the ins trument. Mass flow (or weight per unit time) units are given in pounds per hour (lb/ hour), kilograms per sec (kg/sec) etc. When your specifica tions state units o ...

  • Teledyne HFM-I-401 - page 33

    401-405 SERIES - 25 - 6.2. Appendix 2 - Gas Conversion Factors The gas correction factors (GCF’s) presented in this m anual were obtained by one of four methods. The following table summarizes the different methods for determi ning GCF’s and will help identify for which gases the highest degree of accuracy may be achieved when applying a correc ...

  • Teledyne HFM-I-401 - page 34

    401-405 SERIES - 26 - Some meters, specifically the high flow meters, are ca librated in air. The flow readings must then be corrected twice. Convert once from air to nitro gen, th en from nitro gen to the gas that will be measured with the meter. In this case, multiply the reading ti mes the ratio of the process gas’ GCF to the GCF of the calibr ...

  • Teledyne HFM-I-401 - page 35

    401-405 SERIES - 27 - 37 Cyclopropane C 3 H 6 0.4562 4 1.720 1.4440 38 Deuterium H 2 2 1.0003 4 0.165 0.3102 39 Diborane B 2 H 6 0.5063 5 1.131 1.0486 40 Dibromodifluoromethane CBr2F2 0.3590 4 8.576 5.2998 41 Dichlorofluoromethane CHCl 2 F 0.4481 4 4.207 3.2249 42 Dichloromethane CH 2 Cl 2 0.5322 4 3.472 3.0592 43 Dichloropropane C 3 H 6 Cl 2 0.269 ...

  • Teledyne HFM-I-401 - page 36

    401-405 SERIES - 28 - 96 Isobutane C 4 H 10 0.2725 2 2.376 1.6912 97 Isobutanol C 4 H 10 O 0.2391 4 3.030 1.9228 98 Isobutene C 4 H 8 0.2984 4 2.293 1.6663 99 Isopentane C 5 H 12 0.2175 4 2.949 1.8975 100 Isopropyl Alcohol C 3 H 8 O 0.2931 4 2.456 1.7335 101 Isoxazole C 3 H 3 NO 0.4333 4 2.823 2.1501 102 Ketene C 2 H 2 O 0.5732 4 1.718 1.5127 103 K ...

  • Teledyne HFM-I-401 - page 37

    401-405 SERIES - 29 - 157 R143A C 2 H 3 F 3 0.3394 4 3.435 2.2533 158 R152A C 2 H 4 F 2 0.3877 4 2.700 1.9753 159 R218 C 3 F 8 0.1818 4 7.685 3.0998 160 R1416 C 2 H 3 Cl 2 F 0.3047 4 4.780 2.7342 161 Radon Rn 1.4043 4 9.074 1.0000 162 Sec-butanol C 4 H 10 O 0.2327 4 3.030 1.9213 163 Silane SiH 4 0.6809 5 1.313 1.1934 164 Silicone Tetrafluoride SiF ...

  • Teledyne HFM-I-401 - page 38

    401-405 SERIES - 30 - HFM-I-405 Flow meter ...

  • Teledyne HFM-I-401 - page 39

    401-405 SERIES - 31 - HFM-I-401 Flow Meter ...

Produzent Teledyne Kategorie Automobile Parts

Dokumente, die wir vom Produzenten des Geräts Teledyne HFM-I-401 erhalten, können wir in mehrere Gruppen teilen. Unteranderem in:
- technische Zeichnungen Teledyne
- Bedienungsanleitungen HFM-I-401
- Produktkarten Teledyne
- Informationsbroschüren
- oder Energieetiketten Teledyne HFM-I-401
Jede von ihnen ist wichtig, jedoch finden wir die wichtigsten Informationen für den Nutzer des Geräts in der Bedienungsanleitung Teledyne HFM-I-401.

Die Dokumentengruppe, die als Bedienungsanleitungen bezeichnet wird, wird ebenfalls in detaillierte Arten geteilt, solche wie: Montageanleitungen Teledyne HFM-I-401, Wartungsanleitungen, Kurzanleitungen oder Benutzeranleitungen Teledyne HFM-I-401. Abhängig vom Bedarf, sollten Sie das Dokument finden, das Sie brauchen. In unserem Service können Sie sich die populärste Bedienungsanleitung des Produkts Teledyne HFM-I-401 ansehen.

Ähnliche Bedienungsanleitungen

Die komplette Bedienungsanleitung des Geräts Teledyne HFM-I-401, wie sollte sie aussehen?
Die Bedienungsanleitung, auch bezeichnet als Benutzerhandbuch, oder einfach nur „Anleitung”, ist ein technisches Dokument, das dem Benutzer bei der Nutzung von Teledyne HFM-I-401 hilfreich sein soll. Die Bedienungsanleitungen werden in der Regel von technischen Schriftstellern geschrieben, aber in einer Sprache, die für alle Nutzer von Teledyne HFM-I-401 verständlich ist.

Eine gänzliche Bedienungsanleitung von Teledyne sollte einige Grundelemente enthalten. Ein Teil von ihnen ist nicht so wichtig, wie z.B.: die Titelseite oder Autorenseiten. Die restlichen von ihnen jedoch, sollten Informationen liefern, die für den Nutzer von enormer Wichtigkeit sind.

1. Einführung und Hinweise, wie man sich in einer Bedienungsanleitung von Teledyne HFM-I-401 bewegt - Am Anfang jeder Bedienungsanleitung sollten wir Hinweise bezüglich der Nutzungsart eines bestimmten Ratgebers finden. In ihr sollten sich Informationen über die Lokalisierung des Inhaltsverzeichnisses von Teledyne HFM-I-401 befinden, FAQ oder über oft auftretende Probleme – also Stellen, die von den Benutzern in jeder Bedienungsanleitung am meisten gesucht werden
2. Inhaltsverzeichnis - Index aller Ratschläge bezüglich Teledyne HFM-I-401, die wir im aktuellen Dokument finden
3. Ratschläge zur Nutzung der Grundfunktionen des Geräts Teledyne HFM-I-401 - die uns die ersten Schritte während der Nutzung von Teledyne HFM-I-401 erleichtern sollten
4. Troubleshooting - geordneter Tätigkeitslauf, der uns bei der Diagnose und als nächstes bei der Lösung wichtiger Probleme mit Teledyne HFM-I-401 hilft
5. FAQ - häufig gestellte Fragen
6. Kontaktdaten Informationen darüber, wo man Kontakt zum Produzenten / Service von Teledyne HFM-I-401 im bestimmten Land suchen kann, wenn es nicht gelingt, das Problem selbst zu lösen.

Haben Sie eine Frage bezüglich Teledyne HFM-I-401?

Nutzen Sie das untere Formular

Wenn Sie mit Hilfe der gefundenen Bedienungsanleitung Ihr Problem mit Teledyne HFM-I-401 nicht gelöst haben, stellen Sie eine Frage, indem Sie das untere Formular nutzen. Wenn einer der Nutzer ein ähnliches Problem mit Teledyne HFM-I-401 hatte, ist es möglich, dass er mit Ihnen die Lösung teilen möchte.

Text vom Bild übertragen

Kommentare (0)