    Portable Oxygen Analyzer Teledyne Analytical Instruments i OPERA TING & SER VICE INSTRUCTIONS FOR AX300-I P ORTABLE O XYGEN A NALYZER 0086 TYPE B EQUIPMENT: Equipment providing a particular degree of protection against electric shock, particularly regarding— • Allowable LEAKAGE CURRENT • Reliability of the protective earth connection (if ...

  • Teledyne AX300-I - page 2

    AX300-I Teledyne Analytical Instruments ii Copyright © 2002 Teledyne Analytical Instruments All Rights Reserved No part of t his ma nual may be re produced, t ransmitt ed, transcribe d, stored in a retrieval system, or translated into any other language or computer language in whole or in part, i n any form or by any means, whe ther it be electr o ...

  • Teledyne AX300-I - page 3

    Portable Oxygen Analyzer Teledyne Analytical Instruments iii About This Manual The AX300-I Operator's Manual pr ovides both introductory and detailed information for configuri ng and operating these instruments. The manual takes you from the tim e you unpack the instrument until you complete the first gas analysis. The bulk of the m anual cont ...

  • Teledyne AX300-I - page 4

    AX300-I Teledyne Analytical Instruments iv Safety Messages Your safety and the safety of ot hers are very important. Please carefully read the following safety m essages. Safety message are indented to aler t the user of potential hazards. Each safety message is associated with a safety alert symbol. These symbols are found in the manual and on the ...

  • Teledyne AX300-I - page 5

    Portable Oxygen Analyzer Teledyne Analytical Instruments v T able of Content s Safety Me ssages .......................................................................... iv List of Fi gures.............................................................................. vii List of Ta bles ............................................................. ...

  • Teledyne AX300-I - page 6

    AX300-I Teledyne Analytical Instruments vi 2.3.5 Anesthetic Gases 25 2.3.5.1 Gases That Induce Reading Error 25 2.3.5.2 Care After Use in Nitrous Oxide 27 2.3.6 Cleaning 28 2.4 Do’s and Don’ts 28 Service Ma nual ............................................................................ 31 3.1 General Service Information 31 3.2 Overall Mainten ...

  • Teledyne AX300-I - page 7

    Portable Oxygen Analyzer Teledyne Analytical Instruments vii List of Figures Figure 1-1: AX300 Front View ......................................................... 9 Figure 2-1: Installing the R17MED Sensor .................................... 15 Figure 2-2: Sensor Cable Connection to Analyzer ....................... 16 Figure 2-3: Mounting the Sen ...

  • Teledyne AX300-I - page 8

    AX300-I Teledyne Analytical Instruments viii List of T ables Table 2-1: Oxygen Reading Error in a Mixture of Anesthetic Gas . 27 Table 3-1: Oxygen Reading Error in a Mixture of Anesthetic Gas . 35 Table 3-2 Troubl eshooting ............................................................ 36 Table 3-3 E rror C odes ...................................... ...

  • Teledyne AX300-I - page 9

    Portable Oxygen Analyzer Introduction Teledyne Analytical Instruments 9 Introduction The Teledyne Analytical Instru ments AX300-I Oxygen Analyzer here after referred to as AX300-I is a portable instrument that provides fast and accurate oxygen analysis. These instruments are designed to analyzer up to 100% oxygen concentr ation in medical gas mixtu ...

  • Teledyne AX300-I - page 10

    Introduction AX300-I Teledyne Analytical Instruments 10 The instrument is powered by th ree AA alkaline batteries and is designed to operate for 2000 hour s in Non-alarm State. Oxygen analysis is linear across the single range of 0-100% using Teledyne’s class R17MED oxygen sens or. A unique sensor failure alarm is incorporated which warns the us ...

  • Teledyne AX300-I - page 11

    Portable Oxygen Analyzer Introduction Teledyne Analytical Instruments 11 • Microprocessor controlled • Sensor fail/disconnect alarm indicator (audible and visual) • Alarm silence button • 2000 operating hours from 3 AA alkaline batteries • Battery status indicator • Stand for upright tabletop deployment • Hardware for pole clamping an ...

  • Teledyne AX300-I - page 12

    Introduction AX300-I Teledyne Analytical Instruments 12 1.4 Applications The AX300-I analyzers is intended for spot checking concentration of oxygen in a gas mixtures used in medical applications such as Anesthesia, Respiratory therapy and is intended for adult, pediatric and Neonatal populations. These instruments may be used in verifying oxygen c ...

  • Teledyne AX300-I - page 13

    Portable Oxygen Analyzer Introduction Teledyne Analytical Instruments 13 the sensor. The diverter, packaged separately when shipped, is necessary when the tee adapter is used to sample gas flowing through a tube. The diverter is not necessary and should not be used when the sensor is placed directly in a chambe r, or when the sensor is used in conf ...

  • Teledyne AX300-I - page 14

    Operation AX300-I Teledyne Analytical Instruments 14 Operation Note: Upon receipt, INSPECT THE ENTIRE UNIT FOR DAMAGE. Check the unit and a ll included accessories for broken or loose parts. If damaged, DO NOT USE. Notify the shipper, and consult Teledyne Analytical Instruments. Note: This equipment is internally powered using 3 AA batteries. CAUTI ...

  • Teledyne AX300-I - page 15

    Portable Oxygen Analyzer Operation Teledyne Analytical Instruments 15 2.1.1 Sensor Installation or Replacement Note: The R17MED oxygen sensor must be installed before the oxygen analyzer/analyzer can be operated 1. Remove the new sensor from its protective bag. Inspect the sensor for damage or electrolyte leakage. If the sensor is damaged, obtain a ...

  • Teledyne AX300-I - page 16

    Operation AX300-I Teledyne Analytical Instruments 16 Figure 2-2: Sensor Cable Connection to Analyzer Note: When the AX300-I instrument is used for diffusion sampling (i.e., incubators, tent s, etc.), the plastic flow diverter must be removed from the R17MED sensor. If the sensor is used in breathing circ uits, etc, the diverter must be used as show ...

  • Teledyne AX300-I - page 17

    Portable Oxygen Analyzer Operation Teledyne Analytical Instruments 17 2.1.2. 1 V-M OUNT A DAPTER I NSTALLATION The V-Mount Adapter consists of a matching plastic plate with integral V-grooves that attach to the rear of the instrument. To install V-Mount Adapter remove battery compartment door by prying up the hinged latch at the bott om of the cove ...

  • Teledyne AX300-I - page 18

    Operation AX300-I Teledyne Analytical Instruments 18 1. Turn the unit off (if it is on). 2. Hold the instrument face down in the palm of your hand. Use a coin to pry up the latch that secures the battery compartment door. Remove the battery compartm ent door. CAUTION: IMPROPER INSTALLATION OF T HE BATTERIES MAY RESULT IN DAMAGE T O THE UNIT AND BAT ...

  • Teledyne AX300-I - page 19

    Portable Oxygen Analyzer Operation Teledyne Analytical Instruments 19 keys except the CAL and ON/OFF ( I/O ) keys are inoperable until a successful calibration is achieved. 5. Note the battery display located below the oxygen readout display. It continuously shows a bar graph of the remaining life to the batteries. In the case of fresh batteries, i ...

  • Teledyne AX300-I - page 20

    Operation AX300-I Teledyne Analytical Instruments 20 5. Wait about 20 seconds to insure the sample line is completely purged with the calibration gas. Press the CAL key. The LCD will count down from 9 to 0. During this tim e the microprocessor is measuring th e sensor output to determine the gas concentration an d selects the calibration range i.e. ...

  • Teledyne AX300-I - page 21

    Portable Oxygen Analyzer Operation Teledyne Analytical Instruments 21 2.1.5 Output 0-1 VDC or RS232 The AX300-I Portable Oxygen Analyzer provides signal outputs for use with recorders and computers. These ins truments are supplied standard with a 0-1 VDC output. An optional 0-1 VDC Interface Cable (P/N B-75554) is available from Teledyne for this p ...

  • Teledyne AX300-I - page 22

    Operation AX300-I Teledyne Analytical Instruments 22 1. Remove the batteries and remove the five screws that hold the case together. 2. Remove the rear case section leavi ng the PCB in the front half of the case. 3. Remove jumper at position JP3 and reinstall it at position JP7. 4. Replace the rear cover and secure in place with five screws. Instal ...

  • Teledyne AX300-I - page 23

    Portable Oxygen Analyzer Operation Teledyne Analytical Instruments 23 BACKPRESSURE OR RESTRI CTION TO FLOW. ERRORS IN READINGS WILL RES ULT IF THIS IS NOT FOLLOWED. THE OXYGEN SENSOR IS A NON-STERILE DEVICE AND SHOULD BE USED IN CONJ UCTION WITH AN ANTIBACTERIAL FILTER. ALWAYS INSTAL L THE SENSOR ON THE INSPIRED LI NE AHEAD OF FILTERS, HUMIDIFIERS ...

  • Teledyne AX300-I - page 24

    Operation AX300-I Teledyne Analytical Instruments 24 2.3 Gas Sampling 2.3.1 Humidity Humidity does not directly affect the accuracy of the sensor's measurement. However, when a nebulizer or other device is used to increase moisture levels in gas m ixtures, the moisture actually dilutes the mixture. This dilution effect decreases the oxyge n co ...

  • Teledyne AX300-I - page 25

    Portable Oxygen Analyzer Operation Teledyne Analytical Instruments 25 It is recommended that the unit be calib rated prior to each use or every 8 hours. When the sensor is connected to a ventilator circuit, th e alternating “breathing” pressure cycles generated by the ventilator will be sensed as an increase in the oxygen pe rcentage (especiall ...

  • Teledyne AX300-I - page 26

    Operation AX300-I Teledyne Analytical Instruments 26 magnitude of this error will de pend upon the lev el of oxygen and the duration of exposure. The anesthetic agents listed in the following table (Halothane, Enflurane, Isoflurane, Sevoflurane, a nd Desflurane) were vaporized into a stream of 30% oxygen / 70% nitrous oxide, and the resulting drops ...

  • Teledyne AX300-I - page 27

    Portable Oxygen Analyzer Operation Teledyne Analytical Instruments 27 Table 2-1: Oxygen Reading Error in a Mixture of Anesthetic Gas Gas or Vapor Level (Balance: Mixture of 30% O 2 / 70% N 2 O, except where noted) Gas or Vapor Test Level Oxygen Reading Error Helium 50%, balance O 2 0% Nitrous Oxide 80%, balance O 2 0% Carbon Dioxide 10%, balance O ...

  • Teledyne AX300-I - page 28

    Operation AX300-I Teledyne Analytical Instruments 28 oxygen reading continues to drop after each use in nitrous oxide the sensor should be removed from service. If the sensor can no longer be calibrated or if there is any sign of electrolyte le akage, the sensor should be disposed of in accordance with local regulatio ns and the Material Safety Dat ...

  • Teledyne AX300-I - page 29

    Portable Oxygen Analyzer Operation Teledyne Analytical Instruments 29 • Recalibrate after replacing the senso r. • Use properly installed alkaline batteries only. • Make sure the R17MED sens or is properly attached. • Remove the plastic flow diverter only when using the tee adapter. • Remove and save the plastic flow diverter when using t ...

  • Teledyne AX300-I - page 30

    Operation AX300-I Teledyne Analytical Instruments 30 BLANK ...

  • Teledyne AX300-I - page 31

    Portable Oxygen Analyzer Service Manual Teledyne Analytical Instruments 31 Service Manual 3.1 General Service Information The Teledyne Model AX300-I portable oxygen analyzer is designed to be robust yet c ompact in size. In or de r to a chieve maximu m reliability in a microproce ssor-bas ed instrument, a single PC board is used wh ich relies exclu ...

  • Teledyne AX300-I - page 32

    Service Manual AX300-I Teledyne Analytical Instruments 32 Excessive alarm activation will wear down the battery f aster than usual. The minimum detectable change in battery voltage corresponds to an increment of about 50 hours, meaning that the battery voltage reading may not change for several hours at a time. If the analyzer is not used for a per ...

  • Teledyne AX300-I - page 33

    Portable Oxygen Analyzer Service Manual Teledyne Analytical Instruments 33 3.6 Gas Sampling 3.6.1 Humidity Humidity does not directly affect the accuracy of the senso r's measurem ent. However, when a nebulizer or other device is used to increase moisture levels in gas mixtures, the moisture actually dilu tes the mixture. This dilution effect ...

  • Teledyne AX300-I - page 34

    Service Manual AX300-I Teledyne Analytical Instruments 34 When the sensor is connected to a ventilator circuit, th e alternating “breathing” pressure cycles generated by the ventilator will be sensed as an increase in the oxygen percentage (especially if the sensor is fast enough to sense the changes, as is the R17MED). In reality, the percenta ...

  • Teledyne AX300-I - page 35

    Portable Oxygen Analyzer Service Manual Teledyne Analytical Instruments 35 oxygen / 70% nitrous oxide, and the resulting drops in oxygen level after an exposure of approximately two hours were noted. Exposures in excess of tw o hours may produce slightly greater errors. The errors listed are typical for all oxygen se nsors such as t he R17MED. Expo ...

  • Teledyne AX300-I - page 36

    Service Manual AX300-I Teledyne Analytical Instruments 36 3.6.5.2 C ARE A FTER U SE IN N ITROUS O XIDE CAUTION: THE R17MED SENSOR SHOULD NOT BE LEFT IN NITROUS OXIDE MIXTURES ANY LONGER THAN ABSOLUTELY NECESSARY. After exposure to nitrous oxide mixtures , the sensor should be left in 100% oxygen overnight (e.g., left in a breathing circuit that has ...

  • Teledyne AX300-I - page 37

    Portable Oxygen Analyzer Service Manual Teledyne Analytical Instruments 37 adapter to prevent back filling. O2 flow rate should not exceed 5 l/min. C) Oxygen concentration at the sensor is significantly higher than 21%. Take the instrument to a well- ventilated area and check the reading again. E) Try calibrating with a known good sensor; if this f ...

  • Teledyne AX300-I - page 38

    Service Manual AX300-I Teledyne Analytical Instruments 38 concentra tion of O2 is in excess of 40% , calibration with 100% is recommended. C) If humidified gas is used to ventilate the patient, water vapor actually dilutes the gas. See Appendix: Humidity, Temperature. D) If a blender is used, check its calibration. See Appendix: Discrepancy in Read ...

  • Teledyne AX300-I - page 39

    Portable Oxygen Analyzer Service Manual Teledyne Analytical Instruments 39 Note: In the event that none of these pr ocedures produce desired results, remove the batteries and return the unit to Teledyne for repair. 3.8 W atchdog The AX300-I is equipped with a watchdog ci rcuit that continuous ly analyzers the electronics for proper operati on. If t ...

  • Teledyne AX300-I - page 40

    Service Manual AX300-I Teledyne Analytical Instruments 40 6 (5) 6 (There is no indication of which key is stuck) A stuck key has been detected. The second digit example (5) on the display shows which key is stuck: 1 - Batt 2 - Cal 5 - Silence 8 - Key Lock 7(2) 7 (There is no indication of the type of DAC failure) A failure has occurred during the a ...

  • Teledyne AX300-I - page 41

    Portable Oxygen Analyzer Appendix Teledyne Analytical Instruments 41 a problem with your instrument to Tele dyne Customer Service. If an error code is indicated on your analyzer record the numbe r and report it to the Customer Service Department at the address below. 3.10 Return Authorization for Service For any service beyond sensor and battery re ...

  • Teledyne AX300-I - page 42

    Appendix AX300-I Teledyne Analytical Instruments 42 Expected Life: 36 months in air. (10 months when continuously exposed to 100% oxygen) Dimensions: 2.5" W × 1.25" D × 4.5" H (66 mm x 33- mm × 111.5 mm) Sensor Cable: Retracted: 2 ft / Extended: 10 ft. Storage Temp. 10-30 °C (continuous), 5-50 °C (Intermittent) Operating Temp: 0 ...

  • Teledyne AX300-I - page 43

    Portable Oxygen Analyzer Appendix Teledyne Analytical Instruments 43 A.2 S p are Part s List QTY PART NO DESCRIPTION 1 C43690-R17MED Micro-Fuel Cell R17MED with flow diverter P/N A50057 3 B99 “AA” size alkaline battery 1 B69934 Cable assembly 1 A268 Tee adapter (22 mm) 1 B-74543-B Front panel assembly (AX300-I) 1 D-74459 Back panel 1 C-74461 Ba ...

  • Teledyne AX300-I - page 44

    Appendix AX300-I Teledyne Analytical Instruments 44 the instrument for which the parts are intended. Orders should be sent to: TELEDYNE Analytical Instruments 16830 Chestnut Street City of Industry, CA 91749-1580 Phone (626) 934-1500, Fax (626) 961-2538 Web: www.teledyne-ai.com or your local representative. ...

  • Teledyne AX300-I - page 45

    Portable Oxygen Analyzer Appendix Teledyne Analytical Instruments 45 BLANK ...

  • Teledyne AX300-I - page 46

    Appendix AX300-I Teledyne Analytical Instruments 46 Index accessories, 46 accuracy, 45 address, 43, 47 ALARM SILENCE bu tton, 10, 13, 15 amplifi cation, 13 analog output , 22 analysis range, 10, 45 anesthetic gas, 27, 37 anode, 12 audio alarm , 13 bar graph indi cator, 2 0 batteries, 10, 19 battery compartment, 19 battery installation, 19 battery s ...

  • Teledyne AX300-I - page 47

    Portable Oxygen Analyzer Index Teledyne Analytical Instruments 47 thermal tracking error, 26, 35 thermistor, 2 5, 35 troublesho oting, 38 type B equipm ent, i ventilator, 26, 36 warranty, ii watchdog circuit, 41 water vapor, 2 2 website addres s, 43, 47 ...

Produzent Teledyne Kategorie Oxygen Equipment

Dokumente, die wir vom Produzenten des Geräts Teledyne AX300-I erhalten, können wir in mehrere Gruppen teilen. Unteranderem in:
- technische Zeichnungen Teledyne
- Bedienungsanleitungen AX300-I
- Produktkarten Teledyne
- Informationsbroschüren
- oder Energieetiketten Teledyne AX300-I
Jede von ihnen ist wichtig, jedoch finden wir die wichtigsten Informationen für den Nutzer des Geräts in der Bedienungsanleitung Teledyne AX300-I.

Die Dokumentengruppe, die als Bedienungsanleitungen bezeichnet wird, wird ebenfalls in detaillierte Arten geteilt, solche wie: Montageanleitungen Teledyne AX300-I, Wartungsanleitungen, Kurzanleitungen oder Benutzeranleitungen Teledyne AX300-I. Abhängig vom Bedarf, sollten Sie das Dokument finden, das Sie brauchen. In unserem Service können Sie sich die populärste Bedienungsanleitung des Produkts Teledyne AX300-I ansehen.

Ähnliche Bedienungsanleitungen

Die komplette Bedienungsanleitung des Geräts Teledyne AX300-I, wie sollte sie aussehen?
Die Bedienungsanleitung, auch bezeichnet als Benutzerhandbuch, oder einfach nur „Anleitung”, ist ein technisches Dokument, das dem Benutzer bei der Nutzung von Teledyne AX300-I hilfreich sein soll. Die Bedienungsanleitungen werden in der Regel von technischen Schriftstellern geschrieben, aber in einer Sprache, die für alle Nutzer von Teledyne AX300-I verständlich ist.

Eine gänzliche Bedienungsanleitung von Teledyne sollte einige Grundelemente enthalten. Ein Teil von ihnen ist nicht so wichtig, wie z.B.: die Titelseite oder Autorenseiten. Die restlichen von ihnen jedoch, sollten Informationen liefern, die für den Nutzer von enormer Wichtigkeit sind.

1. Einführung und Hinweise, wie man sich in einer Bedienungsanleitung von Teledyne AX300-I bewegt - Am Anfang jeder Bedienungsanleitung sollten wir Hinweise bezüglich der Nutzungsart eines bestimmten Ratgebers finden. In ihr sollten sich Informationen über die Lokalisierung des Inhaltsverzeichnisses von Teledyne AX300-I befinden, FAQ oder über oft auftretende Probleme – also Stellen, die von den Benutzern in jeder Bedienungsanleitung am meisten gesucht werden
2. Inhaltsverzeichnis - Index aller Ratschläge bezüglich Teledyne AX300-I, die wir im aktuellen Dokument finden
3. Ratschläge zur Nutzung der Grundfunktionen des Geräts Teledyne AX300-I - die uns die ersten Schritte während der Nutzung von Teledyne AX300-I erleichtern sollten
4. Troubleshooting - geordneter Tätigkeitslauf, der uns bei der Diagnose und als nächstes bei der Lösung wichtiger Probleme mit Teledyne AX300-I hilft
5. FAQ - häufig gestellte Fragen
6. Kontaktdaten Informationen darüber, wo man Kontakt zum Produzenten / Service von Teledyne AX300-I im bestimmten Land suchen kann, wenn es nicht gelingt, das Problem selbst zu lösen.

Haben Sie eine Frage bezüglich Teledyne AX300-I?

Nutzen Sie das untere Formular

Wenn Sie mit Hilfe der gefundenen Bedienungsanleitung Ihr Problem mit Teledyne AX300-I nicht gelöst haben, stellen Sie eine Frage, indem Sie das untere Formular nutzen. Wenn einer der Nutzer ein ähnliches Problem mit Teledyne AX300-I hatte, ist es möglich, dass er mit Ihnen die Lösung teilen möchte.

Text vom Bild übertragen

Kommentare (0)