Manual de instrucciones Evolution Technologies OM-900M

32 de páginas 1.18 mb
Descargar

Pasar a la página of 32

Summary
  • Evolution Technologies OM-900M - page 1

    PRODUCT INFORMATION AND INSTRUCTIONS OM-900M ...

  • Evolution Technologies OM-900M - page 2

    TABLE OF CONTENTS General Information ............................................. 3 Important Safety Rules and Precautions  Operating theUnit ......................................... 4-6  Accessories/Repairs ......................................... 6 Introduction Purpose .................................................. 7 User Q ...

  • Evolution Technologies OM-900M - page 3

    Care and Maintenance Hygienic Preparation ....................................... 17 Intervals ................................................. 17 Procedure ................................................ 18 Patient Change ............................................ 18 Functional Check .......................................... 19 Dispo ...

  • Evolution Technologies OM-900M - page 4

    This manual provides the information necessary to operate the EVOLUTION TM MOTION electronic oxygen conserver with built-in regulator in accordance with a physician’s prescription. The EVOLUTION TM MOTION conserver can be used with any CGA 870 post-valve cylinder (see Fig. A) at home or away from home to provide your specific oxygen requirements. ...

  • Evolution Technologies OM-900M - page 5

    IMPORTANT SAFETY RULES & PRECAUTIONS Read this instruction manual carefully to ensure your complete understanding before operating your EVOLUTION TM MOTION electronic conserver. This manual is part of the unit and must be available at all times. Use the unit for the designated purpose only (see “Purpose” on page 7). For your own safety and ...

  • Evolution Technologies OM-900M - page 6

    IMPORTANT SAFETY RULES & PRECAUTIONS WARNING! (Cont.) Please remember that this is for your own safety! • Keep all parts free of oil and grease. Hydrocarbon compounds such as oil, grease, petroleum-based products, cleaning agents containing alcohol, hand cream or adhesive bandages can cause explosive reactions if they come into contact with h ...

  • Evolution Technologies OM-900M - page 7

    CAUTION! (Cont.) • Oxygen conserving systems only work reliably upon sufficiently strong inhalation. Therefore, please observe the following: • Do not use the EVOLUTION TM MOTION at night or while sleeping; • Do not use the EVOLUTION TM MOTION for babies or children; • Do not use the EVOLUTION TM MOTION if you only breathe through your mout ...

  • Evolution Technologies OM-900M - page 8

    INTRODUCTION PURPOSE The EVOLUTION TM MOTION electronic conserver is designed for use as part of a portable ambulatory oxygen system for the purpose of facilitating long-term oxygen therapy (LTOT). It provides mobile LTOT patients with an extended operating time of their mobile system. The EVOLUTION TM MOTION conserver includes a combination of a l ...

  • Evolution Technologies OM-900M - page 9

    INTRODUCTION 8 FUNCTION The EVOLUTION TM MOTION OM-900M oxygen therapeutic device, which combines a regulator, motion sensor, and an oxygen conserver, is designed for use as part of an ambulatory oxygen system. When we breathe, approximately one-third of the time is spent inhaling, and two-third exhaling. As a result, oxygen delivered by continuous ...

  • Evolution Technologies OM-900M - page 10

    DESCRIPTION OF PARTS & CONTROLS FIGURE B EVOLUTION TM MOTION Conserver Top View Oxygen Pressure Gauge Cylinder Adjustment Handle Selector Knob Low Battery Indicator Light FIGURE D EVOLUTION TM MOTION Conserver Bottom View 9 Oxygen Supply Outlet OXYGEN INLET CONNECTION Seal Washer FIGURE C EVOLUTION TM MOTION Conserver Side View Battery ...

  • Evolution Technologies OM-900M - page 11

    • Low Battery Indicator Light: The EVOLUTION TM MOTION conserver incorporates a visual indicator light that alerts you when a battery change is needed. The following signal will be displayed as your batteries become depleted: Blinking red light - Low battery energy level. Replace batteries. • Cylinder Adjustment Handle: This is used to attach t ...

  • Evolution Technologies OM-900M - page 12

    11 ASSEMBLY AND USE • Make certain that your hands are free of oil, grease, and other contaminants. • Inspect the unit to insure that it has a manufacturer-specified seal washer in good working condition attached to the inlet nozzle. • Secure the cylinder in an upright position. • Inspect the post valve of the cylinder and the EVOLUTION TM ...

  • Evolution Technologies OM-900M - page 13

    12 INSTALLING ALKALINE BATTERIES The EVOLUTION TM MOTION conserver uses advanced technology that prolongs battery life through efficient power use. With normal use of four (4) hours per day at any setting, your batteries should last a minimum of one (1) year. Two (2) AA alkaline batteries are packaged in the box with your EVOLUTION TM MOTION conser ...

  • Evolution Technologies OM-900M - page 14

    ASSEMBLY AND USE MONITORING BATTERY ENERGY LEVEL: The EVOLUTION TM MOTION conserver is equipped with a low battery indicator light that alerts you when a battery change is needed. The following signal will be displayed as your batteries become depleted: Blinking red light – Low battery energy level. Replace batteries. Disposal of Batteries: Do no ...

  • Evolution Technologies OM-900M - page 15

    OPERATING INSTRUCTIONS: STEP 1: Make sure the EVOLUTION TM MOTION conserver is set to the “ PULSE ” position. Battery saving technology has eliminated the need for an “OFF” position. The unit will go to “Sleep Mode” after three minutes of non-use. The unit will turn on again when an inspiration is detected or the Selector Button is pres ...

  • Evolution Technologies OM-900M - page 16

    15 ASSEMBLY AND USE OPERATING INSTRUCTIONS (Cont.): The Active setting should be used during exercise or activity and the Rest setting should be used when inactive or at rest. The Rest and Active settings should be selected in accordance with a physician’s prescription or with a physician’s approval. The motion sensor will select Rest or Active ...

  • Evolution Technologies OM-900M - page 17

    M2 36 liters 3.0 1.5 1.0 0.8 0.6 0.5 0.3 M4(A) 113 liters 9.4 4.7 3.1 2.4 1.9 1.6 0.9 M6(B) 164 liters 13.7 6.8 4.6 3.4 2.7 2.3 1.4 ML6 171 liters 14.3 7.1 4.8 3.6 2.9 2.4 1.4 M7 198 liters 16.5 8.3 5.5 4.1 3.3 2.8 1.7 M9(C) 246 liters 20.5 10.3 6.8 5.1 4.1 3.4 2.1 D 425 liters 35.4 17.7 1.8 8.9 7.1 5.9 3.5 E 680 liters 56.7 28.3 18.9 14.2 11.3 9.4 ...

  • Evolution Technologies OM-900M - page 18

    17 CARE AND MAINTENANCE The EVOLUTION TM MOTION conserver is designed for a long and accurate life; however, as with any electronic device, prudent care is required. The unit should be kept clean and free from moisture and dust, as well as extreme temperature. Do not expose the unit to water, such as when bathing or swimming. It is advisable to kee ...

  • Evolution Technologies OM-900M - page 19

    18 CARE AND MAINTENANCE PROCEDURE Carry out hygienic preparation of the unit and accessories as described on page 16. We recommend using a 0.5% TERRALIN solution for wipe disinfection. Follow the instructions enclosed with the disinfectant. WARNING! • Take special care that no liquids enter the unit, as this may cause damage. • You should under ...

  • Evolution Technologies OM-900M - page 20

    19 CARE AND MAINTENANCE FUNCTIONAL CHECK CHECKING FOR LEAKS 1. Close the valve on the oxygen cylinder. 2. Depressurize the EVOLUTION TM MOTION by inhaling several times using the nasal cannula. The gauge indicator should drop to zero. 3. Check that all screwed unions and tube connections are tight. If necessary, tighten them by hand. * NOTE: Do ...

  • Evolution Technologies OM-900M - page 21

    STANDARD PRODUCT ORDER NUMBER DESCRIPTION OM-900M EVOLUTION TM MOTION with CGA 870 Connection SPARE PARTS ORDER NUMBER DESCRIPTION RP-3040 Black “S” Cylinder Adjustment Handle FR-870G Seal Washers (U.S. Only), bag of 10 ACCESSORIES ORDER NUMBER DESCRIPTION OP-150-800 3-in-1 carry bag, fits M4, M6, M7, and M9 cylinders OP-150T Horizontal carryin ...

  • Evolution Technologies OM-900M - page 22

    21 PROBLEM PROBABLE CAUSE SOLUTION Software needs to be reset Dead batteries. Batteries installed incorrectly (reversed). Dirty battery holder contacts. Cylinder valve is closed. Cylinder is empty. Oxygen cannula is blocked or kinked. Non-alkaline batteries are used. Batteries are faulty. Open the battery door and remove the batteries. Wait 10 seco ...

  • Evolution Technologies OM-900M - page 23

    PROBLEM PROBABLE CAUSE SOLUTION 22 The Active Mode setting must be equal to or higher than the Rest Mode setting The device may be interpreting the car’s motion as activity The battery may be dead or depleted The activity time was not long enough to activate the activity sensor Switching times between Rest and Active Modes are variable Select the ...

  • Evolution Technologies OM-900M - page 24

    23 CLASSIFICATIONS AND SPECIFICATIONS TECHNICAL DATA SPECIFICATIONS EVOLUTION TM MOTION Product class according to 93/42/EEC IIb Dimensions (L x H x W) 6.1” L (15.5 cm) x 2.5” H (6.4 cm) x 3.1” W (7.9 cm) Weight Approximately 14.9 ounces (422 grams) with batteries Input pressure 200 to 3000 PSI (13.8 bar to 206.8 bar) Temperature range • Op ...

  • Evolution Technologies OM-900M - page 25

    24 CLASSIFICATIONS AND SPECIFICATIONS SPECIFICATIONS (Cont.) Oxygen delivery at level: Liter Flow Equivalency 1 1 2 2 3 3 4 4 5 5 6 6 Continuous flow 2 LPM ±0.5 LPM Maximum breathing rate 40 breaths per minute Power supply • Batteries (2) x 1.5 V alkaline AA or LR6-type rechargeable batteries Low battery indicator light: • Blinking red light L ...

  • Evolution Technologies OM-900M - page 26

    25 CLASSIFICATIONS AND SPECIFICATIONS SPECIFICATIONS (Cont.) Classification according to EN 60601-1 • Degree of protection against electric shock Type BF • Degree of protection against harmful ingress of water IPX1 as per IEC 60529 Electromagnetic compatibility (EMC) according to EN 60601-1-2 • Radio interference suppression EN 55011 • Radi ...

  • Evolution Technologies OM-900M - page 27

    26 CLASSIFICATIONS AND SPECIFICATIONS SEPARATION DISTANCES Recommended separation distances between portable and mobile RF Communications equipment and the EVOLUTION TM MOTION. The EVOLUTION TM MOTION is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled. The customer or user of the EVOLUTION TM MOT ...

  • Evolution Technologies OM-900M - page 28

    27 CLASSIFICATIONS AND SPECIFICATIONS SYMBOLS KEY SYMBOL MEANING Warning, consult accompanying documents Refer to the instruction manual TYPE PLATE Year manufactured Degree of protection against electric shock: type BF unit Type of protection against electric shock: protection class II unit Do not dispose of the unit in the household waste SN Seria ...

  • Evolution Technologies OM-900M - page 29

    28 LIMITED WARRANTY The EVOLUTION TM MOTION oxygen conserver has been carefully manufactured and inspected and is warranted to be free from defects in workmanship and materials. Under this warranty, CHAD Therapeutics’ obligation shall be limited to the replacement or repair of any such units or parts that prove, by CHAD’s inspection, to be defe ...

  • Evolution Technologies OM-900M - page 30

    29 IMPORTANT INFORMATION TO RECORD Your Name: __________________________________________________________________ Date You Received Your Unit: ____________________________________________________ Prescribed Oxygen Flow Setting: • At Rest: ____________________________ • During Exercise: _____________________ Home Care Provider’s Name: _________ ...

  • Evolution Technologies OM-900M - page 31

    30 NOTES ...

  • Evolution Technologies OM-900M - page 32

    Parent Company Drive Medical Design & Manufacturing 99 Seaview Boulevard Port Washington, NY 11050 Toll-free: 877-224-0946 www.drivemedical.com 401 Leonard Blvd. N Lehigh Acres, FL 33971 T oll-free: 888-446-6862 www .chadtherapeutics.com MDSS Schiffgraben 41 30175 Hanover , Germany PM-00042/13/E Printed in the U.S.A. CHAD is a registered tradem ...

Fabricante Evolution Technologies Categoría Oxygen Equipment

Los documentos del dispositivo Evolution Technologies OM-900M que obtenemos del fabricante se pueden dividir en varios grupos. Entre ellos están::
- dibujos técnicos Evolution Technologies
- manuales de instrucciones OM-900M
- hojas de producto Evolution Technologies
- folletos informativos
- o etiquetas energéticas Evolution Technologies OM-900M
Todos son importantes, pero lo más importante desde el punto de vista del usuario de un dispositivo lo encontraremos en el manual de instrucciones Evolution Technologies OM-900M.

Un conjunto de documentos determinado como manual de instrucciones se divide también en tipos más detallados, tales como: instrucciones de montaje Evolution Technologies OM-900M, instrucciones de servicio, instrucciones cortas o instrucciones de usuario Evolution Technologies OM-900M. Dependiendo de la situación debes buscar el documento que necesitas. En nuestra página puedes consultar el manual de instrucciones más popular del producto Evolution Technologies OM-900M.

Manuales de instrucciones parecidos

El manual de instrucciones del dispositivo Evolution Technologies OM-900M, ¿cómo debería ser?
El manual de instrucciones, también determinado como el manual de usuario o simplemente instrucciones, es un documento técnico que tiene como objetivo ayudar a los usuarios a utilizar Evolution Technologies OM-900M. Las instrucciones normalmente las realiza un escritor técnico pero en un lenguaje comprensible para todos los usuarios de Evolution Technologies OM-900M.

El manual de instrucciones completo de Evolution Technologies debe contener unos elementos básicos. Una parte de ellos es menos importante, como por ejemplo: la portada / la página principal o las páginas de autor. Sin embargo, lo demás debe facilitarnos la información más importante desde el punto de vista del usuario.

1. Introducción y pistas sobre cómo orientarse con el manual Evolution Technologies OM-900M - Al principio de cada manual se deben encontrar las indicaciones acerca de la manera de usar un manual de instrucciones. Debe contener información sobre dónde encontrar el índice Evolution Technologies OM-900M, preguntas frecuentes o problemas más comunes – son los apartados más buscados por los usuarios de cada manual de instrucciones.
2. Índice - listados de todos los consejos acerca de Evolution Technologies OM-900M que encontraremos en el documento
3. Consejos de uso de las funciones básicas del dispositivo Evolution Technologies OM-900M - que deben facilitarnos los primeros pasos durante el uso de Evolution Technologies OM-900M
4. Troubleshooting - una secuencia sistematizada de acciones que nos ayudará a diagnosticar los problemas más importantes con Evolution Technologies OM-900M
5. FAQ - las preguntas frecuentes
6. Datos de contacto Información acerca de cómo encontrar los datos de contacto del fabricante / servicio de Evolution Technologies OM-900M en cada país si no somos capaces de solucionar el problema por nuestra cuenta.

¿Tienes una pregunta acerca de Evolution Technologies OM-900M?

Utiliza el formulario que se encuentra abajo

Si con la ayuda del manual de instrucciones encontrado no has podido solucionar tu problema con Evolution Technologies OM-900M, haz una pregunta utilizando el formulario que se encuentra abajo. Si alguno de los usuarios ha tenido un problema parecido con Evolution Technologies OM-900M probablemente compartirá una forma de solucionarlo.

Copia el texto de la imagen

Comentarios (0)