    MILLENNIUM Oxygen Gas Detector User Manual Models: MLP-A/AR/AD-ST1400-25 & (all available options) Part Number: MAN-0 060 Rev 04 August 2008 ...

  • Millennium Enterprises AD-ST1400-25 - page 2

    ...

  • Millennium Enterprises AD-ST1400-25 - page 3

    Net Safety Monitoring Inc I MPORTANT I NFOR MATION This ma nual is f or infor mational pu rposes only . Altho ugh every e ffor t has bee n made to ens ure the c orrec tness of t he infor mation, tec hnical inaccur acies may occur and pe riodic ch anges may be made without not ice. Net Saf ety Monitori ng Inc., assumes no r esponsibi lity for any er ...

  • Millennium Enterprises AD-ST1400-25 - page 4

    Net Safety Monitoring Inc T ABLE OF C ONTENTS Impo rta nt Inf orm ati on Warranty Contact Information Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 The Product ....................... ........................... .................. ........................... ....... 1 The Sensor ...... .... ...

  • Millennium Enterprises AD-ST1400-25 - page 5

    Net Safety Monitoring Inc Outputs ... ........................... .................. ........................... .................. .............. 18 Relays - Optional ........................ ............ ................ ............ ................. ........... ..... 18 Modbus ........... ........... ............ ................. ........... . ...

  • Millennium Enterprises AD-ST1400-25 - page 6

    Net Safety Monitoring Inc. MLP-A/AR/AD-ST1400-25 1 I NTR ODUCTI ON The Millenni um series is Net Sa fety's la test innov ation in a lin e of continuous ly evolvin g industria l gas detector s and sensors. The micr ocontrol ler based syst em provide s fast, accur ate and co ntinuous mon itoring of g ases in ex treme environ ments. T HE P RODUCT ...

  • Millennium Enterprises AD-ST1400-25 - page 7

    Net Safety Monitoring Inc MLP-A/AR/AD-ST1400-25 2 STEP 1 — P LAN L OCATE C ONTROLLER /S ENSOR Prior t o the inst allation p rocess, a location plan f or placi ng the Contr oller and Sensor shoul d be developed. Althoug h there are no absolute ru les for determi ning the qu antity an d locati on of a sensor or controller , the following points s h ...

  • Millennium Enterprises AD-ST1400-25 - page 8

    Net Safety Monitoring Inc. MLP-A/AR/AD-ST1400-25 3 Figur e 2: Dimensional Drawings STEP 2 — I NSTALL U NPACK Carefully r emove all components f rom the pack aging. Check components against th e enclosed pac king list and inspe ct all compone nts for obvious damage such as broken or loose pa rts. If you f ind any compon ents missing o r damaged, n ...

  • Millennium Enterprises AD-ST1400-25 - page 9

    Net Safety Monitoring Inc MLP-A/AR/AD-ST1400-25 4 Mount The Controll er is usually mounte d near eye-level an d be easily accessi ble for calibr ation and maintenanc e purposes . The Sensor s hould be plac ed where O2 may build up or where o ther gase s are li kely to bui ld up and di splace O2. Ensure al l device s are sec urely mo unted, taking i ...

  • Millennium Enterprises AD-ST1400-25 - page 10

    Net Safety Monitoring Inc. MLP-A/AR/AD-ST1400-25 5 Board Assembly There ar e three dif fer ent fixed boards, two optional relay boa rds and an o ptional Modb us board which mak e up the PCB Ass embly . Depending up on requir e ments, either a El ectromech anical or Solid S tate Re lay Board modu le can be u sed. Boards a re fiel d replaceabl e. Sim ...

  • Millennium Enterprises AD-ST1400-25 - page 11

    Net Safety Monitoring Inc MLP-A/AR/AD-ST1400-25 6 Figur e 5: Wiring—Controlle r and Senso r Note: If the 4-20 mA signal is not used, connect a jumper betwe en the 4-20 term inal and th e COM ter mina l on the T e rmin al Boar d. S ENSOR AND C ONTROLLER WA R N I N G : Power to the unit must be OFF before wiring. Note: The Sensor may be fact ory in ...

  • Millennium Enterprises AD-ST1400-25 - page 12

    Net Safety Monitoring Inc MLP-A/AR/AD-ST1400-25 7 S ENSOR S EPARATION Figur e 6: Wi ri ng — Sensor Separati on As the Sen sor must be l ocated in one location and the Con troller in a loca tion where it can be ea sily rea ched, it is often necessar y to “sepa rate” th e Controller and Sensor . The Sensor separat ion kit ( SEP) is composed o f ...

  • Millennium Enterprises AD-ST1400-25 - page 13

    Net Safety Monitoring Inc MLP-A/AR/AD-ST1400-25 8 N ON -I SOLATED /I SOLATED W IR ING T ermi nal C onnection Figur e 7: Non-isolat ed T erminal Conne ction Figur e 8: Isolated T erminal Conne ction Current Outpu t T o set an isolated current output, sim ply move the Ju mper (sho rting j ack) t o the isolated c urrent po sition ( refer t o Figure 9, ...

  • Millennium Enterprises AD-ST1400-25 - page 14

    Net Safety Monitoring Inc MLP-A/AR/AD-ST1400-25 9 R EMOTE R ESET The Mille nnium can be s et for r emote rese t. A normally ope n Push Button Switch must be connected bet ween the RST and the COM terminal s on the T erminal Boa rd and the Relay Option Latch S tatus mus t be set to Latching . Figur e 10: Remote Reset MODBUS T ERMINATION Devices a re ...

  • Millennium Enterprises AD-ST1400-25 - page 15

    Net Safety Monitoring Inc. MLP-A/AR/AD-ST1400-25 10 STEP 4 — O PERATE The Controll er faceplate c ontains most functi onal elements of the us er interface. Be low is a descripti on of that function ality . Figur e 12: Controll er Functional ity Status Red Slow Flash: Gas present Flast flash : Cal or Fault Solid: Overrange Pull Here Green Short Bl ...

  • Millennium Enterprises AD-ST1400-25 - page 16

    Net Safety Monitoring Inc MLP-A/AR/AD-ST1400-25 11 RT U R eg i s t e r s Register 40001 = % output (read only) Register 40002 = R TU Statu s (read only) Register 40 101 = Reset l atched r elays (writ e) Calibration Button The Calibr ation Button p rovide s access to t he Millennium’ s Main Menu , which in turn allows ca librati on and options to ...

  • Millennium Enterprises AD-ST1400-25 - page 17

    Net Safety Monitoring Inc MLP-A/AR/AD-ST1400-25 12 Magnetic R eed Sw itch The Magneti c Reed Switc h is provided to avoid opening the H ousing in a n environ ment where gas m ay be presen t. The Magnet ic Reed Switc h functions in the same manne r as the Cali bration Button but in a non- intrusi ve manner . Refer to Figure 12 , "Contr oller Fu ...

  • Millennium Enterprises AD-ST1400-25 - page 18

    Net Safety Monitoring Inc. MLP-A/AR/AD-ST1400-25 13 STEP 5 — C ALIBRATE C ALIBRATION P ROCEDUR E W ARNING: The calibra tion procedu re requir es about 5 minutes to c omplete. If gas is not ap plied at th e appropriat e time, a calib ration failure may occur (ref er to "Calibrat ion Failure" on page 14 for speci fic informati on). For ac ...

  • Millennium Enterprises AD-ST1400-25 - page 19

    Net Safety Monitoring Inc MLP-A/AR/AD-ST1400-25 14 Calibrat ion Fail ure If the calibrat ion proce dure fails , the di splay shows Fai l Cal , the Stat us LED alter nates Red an d Green fl ashes and t he analog o utput changes bac k and fo rth from 3.0 to 3.3 mA. The unit remains in a failed st ate unti l manually reset. Aft er the Manu al Rese t, ...

  • Millennium Enterprises AD-ST1400-25 - page 20

    Net Safety Monitoring Inc. MLP-A/AR/AD-ST1400-25 15 R ELAY O PTIONS Use to se t the ala rm level, co il status and latch status f or the Low an d High relay s (Fau lt Ala rm is fi xed). The followin g table descri bes the default set tings for the sen sor . Ta b l e 4 : Default Re lay Settings Setting Relay Options There ar e two setting s for Rela ...

  • Millennium Enterprises AD-ST1400-25 - page 21

    Net Safety Monitoring Inc MLP-A/AR/AD-ST1400-25 16 M ODBUS O PTIONS The digital RS- 485 Modbus R TU protocol is used. There are two Modbus options : Node Addres s and Baud Rat e. Node Address Each devi ce connect ed to the chai n must be as signed a u nique node a ddress (1 - 255). The last number in the add ress is selected firs t. St e p 1 : When ...

  • Millennium Enterprises AD-ST1400-25 - page 22

    Net Safety Monitoring Inc. MLP-A/AR/AD-ST1400-25 17 S ELECT D ISPLAY L ANGUAGE St e p 1 : After t he Restrict ed Menu the opt ion Select Disp lay Lan guage YES? displ ays. Press t he Calibra tion Button or use the Re ed Switch to select. The flas hing YES confir ms the select ion. St e p 2 : When the require d language displa ys ( Englis h , Espan ...

  • Millennium Enterprises AD-ST1400-25 - page 23

    Net Safety Monitoring Inc MLP-A/AR/AD-ST1400-25 18 O UTP UTS Relays - Optional Note: The fa ult rel ay output i s not use d to acti vate an aut omatic shu tdown procedur e. The faul t output indicate s a potent ial problem wi th the Controll er not an alarm condition. St andard ele ctro-mech anical r elay outpu ts have Form C SPDT con tacts ra ted ...

  • Millennium Enterprises AD-ST1400-25 - page 24

    Net Safety Monitoring Inc MLP-A/AR/AD-ST1400-25 19 H OW TO R ETURN E QUIPMENT A Materia l Return Author ization number is requi red in or der to re turn equipment . Please co ntact Net Safety Monit oring at (403) 219-0688 before return ing equipment or consu lt our Service Departmen t to possibly avoid return ing equipment. If you ar e required to ...

  • Millennium Enterprises AD-ST1400-25 - page 25

    Net Safety Monitoring Inc MLP-A/AR/AD-ST1400-25 20 Rotate PCB Assembly 1. Remove the Control ler ’ s Housing Cover . 2. T urn the power t o the dete ctor of f. 3. Unscrew both the knobs marked "Pull Here ". 4. Lift Contro ller fa ceplate f rom Housing a nd allow to hang from ribbon cable . 5. Unscrew the tw o metal standof fs. 6. Carefu ...

  • Millennium Enterprises AD-ST1400-25 - page 26

    Net Safety Monitoring Inc. MLP-A/AR/AD-ST1400-25 21 Appendix B: R ESISTAN CE T ABLE IN OHMS Note: Resis tance sho wn is one way . This figure should be doub led when det ermining clos ed loo p resistance . Distance (F eet) A WG #20 A WG #18 A WG #16 A WG #14 A WG #12 A WG #10 A WG #8 1 0 0 1 . 0 20 . 6 40 . 4 00 . 2 50 . 1 60 . 1 0 0 . 0 6 2 0 0 2 ...

  • Millennium Enterprises AD-ST1400-25 - page 27

    Net Safety Monitoring Inc. MLP-A/AR/AD-ST1400-25 22 Appendix C: S PECIFICA TIONS S ENSOR S PECIFICATIONS C ONTROLLER S PECIFICATION SENSOR ELEMENT Oxygen O2 (ST1400 ) Operati ng T emperatur e Range -20°C to + 50°C (-4F to +12 2F) % 5 2 o t 0 n o i t c e t e D f o e g n a R 5 9 T o t s d n o c e s 5 1 < e m i T e s n o p s e R r a e y / s s o l ...

  • Millennium Enterprises AD-ST1400-25 - page 28

    N et Safety Monito ring Inc., 2 721 Hopewe ll Place NE C algary , Alberta, Canada T1 Y 7J7 T elepho ne: (40 3) 219- 068 8 Fax: (403 ) 219- 0694 E -mail: netsafe@net-safety .com w ww .netsafety .com ...

Fabricante Millennium Enterprises Categoría Carbon Monoxide Alarm

Los documentos del dispositivo Millennium Enterprises AD-ST1400-25 que obtenemos del fabricante se pueden dividir en varios grupos. Entre ellos están::
- dibujos técnicos Millennium Enterprises
- manuales de instrucciones AD-ST1400-25
- hojas de producto Millennium Enterprises
- folletos informativos
- o etiquetas energéticas Millennium Enterprises AD-ST1400-25
Todos son importantes, pero lo más importante desde el punto de vista del usuario de un dispositivo lo encontraremos en el manual de instrucciones Millennium Enterprises AD-ST1400-25.

Un conjunto de documentos determinado como manual de instrucciones se divide también en tipos más detallados, tales como: instrucciones de montaje Millennium Enterprises AD-ST1400-25, instrucciones de servicio, instrucciones cortas o instrucciones de usuario Millennium Enterprises AD-ST1400-25. Dependiendo de la situación debes buscar el documento que necesitas. En nuestra página puedes consultar el manual de instrucciones más popular del producto Millennium Enterprises AD-ST1400-25.

Manuales de instrucciones parecidos

El manual de instrucciones del dispositivo Millennium Enterprises AD-ST1400-25, ¿cómo debería ser?
El manual de instrucciones, también determinado como el manual de usuario o simplemente instrucciones, es un documento técnico que tiene como objetivo ayudar a los usuarios a utilizar Millennium Enterprises AD-ST1400-25. Las instrucciones normalmente las realiza un escritor técnico pero en un lenguaje comprensible para todos los usuarios de Millennium Enterprises AD-ST1400-25.

El manual de instrucciones completo de Millennium Enterprises debe contener unos elementos básicos. Una parte de ellos es menos importante, como por ejemplo: la portada / la página principal o las páginas de autor. Sin embargo, lo demás debe facilitarnos la información más importante desde el punto de vista del usuario.

1. Introducción y pistas sobre cómo orientarse con el manual Millennium Enterprises AD-ST1400-25 - Al principio de cada manual se deben encontrar las indicaciones acerca de la manera de usar un manual de instrucciones. Debe contener información sobre dónde encontrar el índice Millennium Enterprises AD-ST1400-25, preguntas frecuentes o problemas más comunes – son los apartados más buscados por los usuarios de cada manual de instrucciones.
2. Índice - listados de todos los consejos acerca de Millennium Enterprises AD-ST1400-25 que encontraremos en el documento
3. Consejos de uso de las funciones básicas del dispositivo Millennium Enterprises AD-ST1400-25 - que deben facilitarnos los primeros pasos durante el uso de Millennium Enterprises AD-ST1400-25
4. Troubleshooting - una secuencia sistematizada de acciones que nos ayudará a diagnosticar los problemas más importantes con Millennium Enterprises AD-ST1400-25
5. FAQ - las preguntas frecuentes
6. Datos de contacto Información acerca de cómo encontrar los datos de contacto del fabricante / servicio de Millennium Enterprises AD-ST1400-25 en cada país si no somos capaces de solucionar el problema por nuestra cuenta.

¿Tienes una pregunta acerca de Millennium Enterprises AD-ST1400-25?

Utiliza el formulario que se encuentra abajo

Si con la ayuda del manual de instrucciones encontrado no has podido solucionar tu problema con Millennium Enterprises AD-ST1400-25, haz una pregunta utilizando el formulario que se encuentra abajo. Si alguno de los usuarios ha tenido un problema parecido con Millennium Enterprises AD-ST1400-25 probablemente compartirá una forma de solucionarlo.

Copia el texto de la imagen

Comentarios (0)