Manual de instrucciones National Instruments SCC-CO20

15 de páginas 0.34 mb
Descargar

Pasar a la página of 15

Summary
  • National Instruments SCC-CO20 - page 1

    USER GUIDE SCC-CO20 Isolated Current Output Module The SCC-CO20 is an isolated curren t output module that output s up to 20 mA. The output current lev el is proportional to the voltage applied to the module by an analog output channel of an E/M Series D A Q device. Conventions The following conventions are used in this guide: < > Angle brack ...

  • National Instruments SCC-CO20 - page 2

    SCC-CO20 Isolated Cu rrent Output Module User Guide 2 ni.com italic Italic text denotes variables, emphasis , a cross-reference, or an introduction to a key concept. Italic te xt also deno tes text that is a placeholder for a word or v alue that you must supply . monospace T e x t in this font denotes text or ch aracters that you should enter from ...

  • National Instruments SCC-CO20 - page 3

    © National Instruments Corpor ation 3 SCC-CO20 Isolated Curren t Output Module User Guide ❑ T o ols – 1/8 in. flathead scre wdri ver – Numbers 1 and 2 Phillips screwdriv er s – W ire insu lation stripper Note Software scaling of measurements is not supported on the Macintosh operating system. Refer to the SCC-CO20 Module Pin Assignments se ...

  • National Instruments SCC-CO20 - page 4

    SCC-CO20 Isolated Cu rrent Output Module User Guide 4 ni.com Figure 1. SCC-CO20 Sign al Connections to E/M Series DAQ Device Using the Isolated Loop Supply of the SCC-CO20 You can connect the SCC-CO20 to a load using either the isolated internal loop supply or an external supply. Th e internal loop supply outputs a voltage of 15 V with respect to t ...

  • National Instruments SCC-CO20 - page 5

    © National Instruments Corpor ation 5 SCC-CO20 Isolated Curren t Output Module User Guide Note If your load is g rounded and you are usin g the internal supply , connect the load ground to the current sink terminal at pin 2 on th e SCC-CO20. Do not connect the load ground to the isolated ground terminal because do ing so could defeat the module is ...

  • National Instruments SCC-CO20 - page 6

    SCC-CO20 Isolated Cu rrent Output Module User Guide 6 ni.com Figure 3. Connecting T wo SCC-CO20 Module s to a Higher-R esistance Load SCC-CO20 1 2 + – 15 V iso 3 SCC-CO20 1 2 + – 15 V iso 3 Load SUPPL Y SINK ISO GND SUPPL Y I I ...

  • National Instruments SCC-CO20 - page 7

    © National Instruments Corpor ation 7 SCC-CO20 Isolated Curren t Output Module User Guide Using an External Loop Supply In order for the current sink to oper ate properly with an external loop supply, the voltage must be within the range of 3 to 42 V. If the voltage is too high, overheating can occur and damage the module. Figure 4 shows four conf ...

  • National Instruments SCC-CO20 - page 8

    SCC-CO20 Isolated Cu rrent Output Module User Guide 8 ni.com Figur e 5. Connecting the SCC-CO20 to a Grounded External Loop Supply and a Grounded Load For information about how to conf igu re the SCC-CO20 modul e using NI-D A Qmx, refer to the SCC Quick Start Guide . Calibrating Gain and Offset Errors The SCC-CO20 is calibrated at the f actory befo ...

  • National Instruments SCC-CO20 - page 9

    © National Instruments Corpor ation 9 SCC-CO20 Isolated Curren t Output Module User Guide Note Normally you would adjust to 0 μ A in order to calibrate the of fset, but because the SCC-CO20 cannot source current, it is impossi ble to know when you ha ve adjusted past 0 μ A. Using the SCC-CO20 The SCC-CO20 is a 0 to 20 mA pr ogrammable current si ...

  • National Instruments SCC-CO20 - page 10

    SCC-CO20 Isolated Curr ent Output Module User Guide 10 ni.com Specifications These ratings are typical at 25 ° C unless otherw ise stated. Electrical Number of output channels ................. ....1 Output referencing .............. ................ ....Nonreferenced (floating) Input range .......................... ........... .........0 to 10 V ...

  • National Instruments SCC-CO20 - page 11

    © National Instruments Corporat ion 11 SCC-CO20 Isolated Curren t Output Module User Guide Physical Figure 6. SCC-CO20 Dimensions Weight .................... .............. ........... ....... 28 g (1.0 oz) I/O connectors ...................... .............. .... One 20-pin right-angle male connector One 3-pin screw terminal Field-wiring diameter ...

  • National Instruments SCC-CO20 - page 12

    SCC-CO20 Isolated Curr ent Output Module User Guide 12 ni.com Environmental Operating temperature .. .............. ............0 to 50 °C Storage temperature .............. .............. ....–20 to 70 °C Humidity .............. ............ .............. .........10 to 90% RH, noncondensing Maximum altitude ................. .............. . ...

  • National Instruments SCC-CO20 - page 13

    © National Instruments Corporat ion 13 SCC-CO20 Isolated Curren t Output Module User Guide Online Product Certification Refer to the product Declaration of Conformity (DoC) fo r additional regulatory compliance information. To obtain pro duct certifications an d the DoC for this product, visit ni.com/certification , search by model number or produ ...

  • National Instruments SCC-CO20 - page 14

    SCC-CO20 Isolated Curr ent Output Module User Guide 14 ni.com SCC-CO20 Module Pin Assignments Figure 7 shows the I/O connector pin s on the bottom of the modul e. Figure 7. SCC-CO20 Bottom View T able 1 lists the signal connect ion corresponding to each pin. 1P i n 1 2P i n 2 3 PWB Ke y 4 P i n 1 9 5P i n 2 0 Table 1. SCC-CO20 Pin Signal Connection ...

  • National Instruments SCC-CO20 - page 15

    National Instruments, NI, ni.com, and LabVIEW are tradem arks of National Instruments Corporation. Refer to t he Terms of Use section on ni.com/legal for more information a bout National Instruments trademarks. Other product and compan y names mentioned herein are trademarks or trade names of their respective companie s. For patents covering Natio ...

Fabricante National Instruments Categoría Network Card

Los documentos del dispositivo National Instruments SCC-CO20 que obtenemos del fabricante se pueden dividir en varios grupos. Entre ellos están::
- dibujos técnicos National Instruments
- manuales de instrucciones SCC-CO20
- hojas de producto National Instruments
- folletos informativos
- o etiquetas energéticas National Instruments SCC-CO20
Todos son importantes, pero lo más importante desde el punto de vista del usuario de un dispositivo lo encontraremos en el manual de instrucciones National Instruments SCC-CO20.

Un conjunto de documentos determinado como manual de instrucciones se divide también en tipos más detallados, tales como: instrucciones de montaje National Instruments SCC-CO20, instrucciones de servicio, instrucciones cortas o instrucciones de usuario National Instruments SCC-CO20. Dependiendo de la situación debes buscar el documento que necesitas. En nuestra página puedes consultar el manual de instrucciones más popular del producto National Instruments SCC-CO20.

Manuales de instrucciones parecidos

El manual de instrucciones del dispositivo National Instruments SCC-CO20, ¿cómo debería ser?
El manual de instrucciones, también determinado como el manual de usuario o simplemente instrucciones, es un documento técnico que tiene como objetivo ayudar a los usuarios a utilizar National Instruments SCC-CO20. Las instrucciones normalmente las realiza un escritor técnico pero en un lenguaje comprensible para todos los usuarios de National Instruments SCC-CO20.

El manual de instrucciones completo de National Instruments debe contener unos elementos básicos. Una parte de ellos es menos importante, como por ejemplo: la portada / la página principal o las páginas de autor. Sin embargo, lo demás debe facilitarnos la información más importante desde el punto de vista del usuario.

1. Introducción y pistas sobre cómo orientarse con el manual National Instruments SCC-CO20 - Al principio de cada manual se deben encontrar las indicaciones acerca de la manera de usar un manual de instrucciones. Debe contener información sobre dónde encontrar el índice National Instruments SCC-CO20, preguntas frecuentes o problemas más comunes – son los apartados más buscados por los usuarios de cada manual de instrucciones.
2. Índice - listados de todos los consejos acerca de National Instruments SCC-CO20 que encontraremos en el documento
3. Consejos de uso de las funciones básicas del dispositivo National Instruments SCC-CO20 - que deben facilitarnos los primeros pasos durante el uso de National Instruments SCC-CO20
4. Troubleshooting - una secuencia sistematizada de acciones que nos ayudará a diagnosticar los problemas más importantes con National Instruments SCC-CO20
5. FAQ - las preguntas frecuentes
6. Datos de contacto Información acerca de cómo encontrar los datos de contacto del fabricante / servicio de National Instruments SCC-CO20 en cada país si no somos capaces de solucionar el problema por nuestra cuenta.

¿Tienes una pregunta acerca de National Instruments SCC-CO20?

Utiliza el formulario que se encuentra abajo

Si con la ayuda del manual de instrucciones encontrado no has podido solucionar tu problema con National Instruments SCC-CO20, haz una pregunta utilizando el formulario que se encuentra abajo. Si alguno de los usuarios ha tenido un problema parecido con National Instruments SCC-CO20 probablemente compartirá una forma de solucionarlo.

Copia el texto de la imagen

Comentarios (0)