    CVI ™ , LabVIEW ™ , Natio nal Instru ments ™ ,n i . c o m ™ ,N I - D A Q ™ , and SCXI ™ are trademarks of National Instruments Corporation. Product a nd company names mentioned herein are tr ademarks or trade names of their respective companies. 370271A-01 © Copyright 2000 National Instruments Corp. All rights reserved. December 2000 C ...

  • National Instruments SCXI -1125 - page 2

    SCXI-1125 Calibration Procedure 2 ni.com Equipment and Other T est Requirements This section describes the equipment, software, documentation, and test conditions required for calibrat ing the SCXI-1125. T est Equipment Calibration requires a high-precisi on voltage source with at least 50 ppm accuracy, a multiranging 5 1/2 digit digital multimeter ...

  • National Instruments SCXI -1125 - page 3

    © National Instruments Corpor ation 3 SCXI-1125 Calibration Procedur e Documentation The NI-DAQ Function Refe rence Online Help fi le and NI-DAQ User Manual for PC Compa tibles c o n t a i nd e t a i l e di n f o r m a t i o no nu s i n gt h e NI-DAQ driver. The funct ion reference online h elp file includes detaile d information on the driver fun ...

  • National Instruments SCXI -1125 - page 4

    SCXI-1125 Calibration Procedure 4 ni.com Setting Up Y our Module Complete the following steps to set up yo ur module for testing : 1. Install the SCXI-1125 in sl ot 1 of the SCXI chassis. 2. Connect a 68-to-68-pin cable between the S CXI module and the E Series D A Q de vice installed in the host computer . 3. Configure the hardware with Measuremen ...

  • National Instruments SCXI -1125 - page 5

    © National Instruments Corpor ation 5 SCXI-1125 Calibration Procedur e 7. Connect the calibrator to the appro priate analog input, starting with channel 0. Refer to Figure 1 to determine which pins on the 96-pin front conn ector correspond to t he positiv e and negati ve inputs of the specified channel. Fo r example, the positive input for channel ...

  • National Instruments SCXI -1125 - page 6

    SCXI-1125 Calibration Procedure 6 ni.com 13. Call SCXI_Scale . Set the following parameters: • SCXIChassisID — The v alue assigned b y MAX • moduleSlot — 1 • channel — The same channel number from step 7 • SCXIgain — The same gain value from step 9 • TBgain — 1 • DAQboard — The device number for the D A Q device • DAQChann ...

  • National Instruments SCXI -1125 - page 7

    © National Instruments Corpor ation 7 SCXI-1125 Calibration Procedur e • calOp — ND_SELF_CALIBRATE • s etOfCalConstants — ND_USER_EEPROM_AREA • calRefVolts — 0.0 4. Call AI_Configure to put the E Series device in bipolar mode. Set t he following parameters: • deviceNumb er — The v alue assigned b y MAX • chan — 0 • inputMode ...

  • National Instruments SCXI -1125 - page 8

    SCXI-1125 Calibration Procedure 8 ni.com • filterMode — 1 • cutoffFrequency — 10,000 9. Connect the calibrator to the analog input you w ant to adjust. Refer to Figure 1 to determine the pins on the 96-pin f ront connector that correspond to the positiv e and neg ati ve inputs of the specif ied channel. For e xamp le, the positi ve input fo ...

  • National Instruments SCXI -1125 - page 9

    © National Instruments Corpor ation 9 SCXI-1125 Calibration Procedur e 16. Record binary1 with volt1 and binary2 with volt2 for later use. 17. Repeat steps 7 through 16 for the remaining gain v alues you want to adjust on this channel. 18. Repeat steps 6 through 17 for the remaining chan nels you want to adjust. Y ou hav e determined the new calib ...

  • National Instruments SCXI -1125 - page 10

    SCXI-1125 Calibration Procedure 10 ni.com V erifying Adjusted V alues After completing the adjustment s to your SCXI-1125 module, it is important that you verify the analog input operation by repeating t he steps listed in the Veri fying Gain and Offset Errors section of this docu ment. Reverifying after adjustments ensures that your SCXI-1125 mod ...

  • National Instruments SCXI -1125 - page 11

    © National Instruments Corpor ation 11 SCXI-1125 Calibration Procedur e Front and Rear Signal Connector Diagrams Figure 1. Front Signal Connector Pin Assignments CHSGND +5 V RSVD CH7+ CH6+ CH5+ CH4+ CH3+ CH2+ CH1+ NC NC NC NC RSVD CH0+ 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 AB RSVD CJTEMP RSVD RSVD ...

  • National Instruments SCXI -1125 - page 12

    SCXI-1125 Calibration Procedure 12 ni.com Figure 2. Rear Signal Connector Pin Assignments SERCLK DIGGND SLOT0SEL* DAQD*/A SERDATIN OUTREF MCH7+ MCH6+ MCH5+ MCH4+ MCH3+ MCH2+ MCH1+ MCH0+ AOGND SCANCLK SYNC SERDATOUT DIGGND MCH7 – MCH6 – MCH5 – MCH4 – MCH3 – MCH2 – MCH1 – MCH0 – AOGND 49 50 47 48 45 46 43 44 41 42 39 40 37 38 35 36 33 ...

  • National Instruments SCXI -1125 - page 13

    © National Instruments Corpor ation 13 SCXI-1125 Calibration Procedur e Specifications The following table contains the sp ecifications you should use w hen verifying and calibrating y our SCXI-1125 module. If you r device has been calibrated within the last year, the test point value should fall between the Upper Limit and Lower Limit valu es. Ta ...

  • National Instruments SCXI -1125 - page 14

    SCXI-1125 Calibration Procedure 14 ni.com 200 – 0.025 – 0.024750 – 0.02468642 – 0 .02481358 250 0.02 0.019800 0.01985126 0.019 74874 250 0 0.000000 0.000022 – 0.000022 250 – 0.02 – 0.019800 – 0.01974874 – 0.0 1985126 500 0.01 0.009900 0.00992663 0.009 87337 500 0 0.000000 0.000012 – 0.000012 500 – 0.01 – 0.009900 – 0.00987 ...

Fabricante National Instruments Categoría Car Stereo System

Los documentos del dispositivo National Instruments SCXI -1125 que obtenemos del fabricante se pueden dividir en varios grupos. Entre ellos están::
- dibujos técnicos National Instruments
- manuales de instrucciones SCXI -1125
- hojas de producto National Instruments
- folletos informativos
- o etiquetas energéticas National Instruments SCXI -1125
Todos son importantes, pero lo más importante desde el punto de vista del usuario de un dispositivo lo encontraremos en el manual de instrucciones National Instruments SCXI -1125.

Un conjunto de documentos determinado como manual de instrucciones se divide también en tipos más detallados, tales como: instrucciones de montaje National Instruments SCXI -1125, instrucciones de servicio, instrucciones cortas o instrucciones de usuario National Instruments SCXI -1125. Dependiendo de la situación debes buscar el documento que necesitas. En nuestra página puedes consultar el manual de instrucciones más popular del producto National Instruments SCXI -1125.

Manuales de instrucciones parecidos

El manual de instrucciones del dispositivo National Instruments SCXI -1125, ¿cómo debería ser?
El manual de instrucciones, también determinado como el manual de usuario o simplemente instrucciones, es un documento técnico que tiene como objetivo ayudar a los usuarios a utilizar National Instruments SCXI -1125. Las instrucciones normalmente las realiza un escritor técnico pero en un lenguaje comprensible para todos los usuarios de National Instruments SCXI -1125.

El manual de instrucciones completo de National Instruments debe contener unos elementos básicos. Una parte de ellos es menos importante, como por ejemplo: la portada / la página principal o las páginas de autor. Sin embargo, lo demás debe facilitarnos la información más importante desde el punto de vista del usuario.

1. Introducción y pistas sobre cómo orientarse con el manual National Instruments SCXI -1125 - Al principio de cada manual se deben encontrar las indicaciones acerca de la manera de usar un manual de instrucciones. Debe contener información sobre dónde encontrar el índice National Instruments SCXI -1125, preguntas frecuentes o problemas más comunes – son los apartados más buscados por los usuarios de cada manual de instrucciones.
2. Índice - listados de todos los consejos acerca de National Instruments SCXI -1125 que encontraremos en el documento
3. Consejos de uso de las funciones básicas del dispositivo National Instruments SCXI -1125 - que deben facilitarnos los primeros pasos durante el uso de National Instruments SCXI -1125
4. Troubleshooting - una secuencia sistematizada de acciones que nos ayudará a diagnosticar los problemas más importantes con National Instruments SCXI -1125
5. FAQ - las preguntas frecuentes
6. Datos de contacto Información acerca de cómo encontrar los datos de contacto del fabricante / servicio de National Instruments SCXI -1125 en cada país si no somos capaces de solucionar el problema por nuestra cuenta.

¿Tienes una pregunta acerca de National Instruments SCXI -1125?

Utiliza el formulario que se encuentra abajo

Si con la ayuda del manual de instrucciones encontrado no has podido solucionar tu problema con National Instruments SCXI -1125, haz una pregunta utilizando el formulario que se encuentra abajo. Si alguno de los usuarios ha tenido un problema parecido con National Instruments SCXI -1125 probablemente compartirá una forma de solucionarlo.

Copia el texto de la imagen

Comentarios (0)