    MW9076 Series Optical Time Domain Reflectometer 1.31/1.45/1.55/1.625 µm (SM), 0.85/1.3 µm (GI) T omorr o ws T echnolog y , T oda y ...

  • Anritsu MW9076 - page 2

    2 Highest P er formance, Functions, and Measuring Speed ■ 45 dB high dynamic range ■ 8 m short dead zone ■ Simple measurement of chr omatic dispersion fr om one end of optical fiber ■ Measurement in 10 s (Full-A uto mode), 0.15 s real-time s weep ■ A utomatic execution of functions suc h as wavelength/c hannel switc hing, file sa ving, pr ...

  • Anritsu MW9076 - page 3

    3 ...

  • Anritsu MW9076 - page 4

    4 ● High-Speed Measurement It takes only 10 seconds to measure and displa y the wav e- f or m and connection loss on one screen. Just one press of the Star t key is all that is needed to mak e measurement. ● Short Dead Zone Clearly measure up to near end by 8 m dead zone (back-scatter , SM unit) ● High Dynamic Range When using a wa velength o ...

  • Anritsu MW9076 - page 5

    5 ● Fresnel Reflection The f ar-end F resnel reflection can be measured for f our wa velengths (1310/1450/1550/1625 nm). ● Group Delay Characteristics The fitting f or mula suppor ts cubic or quintic Sellmeier , and polynomials can be applied to various types of fibers. ● Chromatic Dispersion Characteristics The zero and total dispersion can ...

  • Anritsu MW9076 - page 6

    6 Compact, Lightweight, and Easy to Operate O TDR (main frame) Lithium ion batter y mounted, with remaining power displa y Connector f or e xter nal CRT Displa y unit Measurement conditions can be changed at mea- surement screen Large easy-to-read 8.4 inch TFT -LCD Simultaneous displa y of measured results and e vent tab le Rotar y encoder f or eas ...

  • Anritsu MW9076 - page 7

    7 q Status-indicating LED w Function keys e Menu key r Star t key t Arrow ke y y Select key u Batter y pack i Optical power meter connector o Visible LD output connector !0 OTDR connector and light source connector f or optical loss measure- ment !1 Exter nal monitor (V GA) connector !2 PC card slot (two PC cards con- nectable) !33333 Exter nal key ...

  • Anritsu MW9076 - page 8

    8 High-Speed Measurement ● Full Automatic Mode Measurement results are displa yed by simply pressing the Star t key . All complicated settings of distance range, pulse width, attenuator , and mar ker can be automatically e xecuted. Measurement speed in this mode was significantly increased. When the wa velengths are set to ALL, w av elengths are ...

  • Anritsu MW9076 - page 9

    9 V arious Useful Functions ● Waveform Comparison Function Measured and sav ed data can be compared on the same screen. In addition, differences can be displa yed as a wa vef orm for simple observation of distance and le vel dif- f erences. This is useful f or checking aging changes or comparing sev eral fibers. ● Communication Light Check Func ...

  • Anritsu MW9076 - page 10

    10 ● Light Source, Power Meter Optical fiber loss can be measured using the optical powe r meter function and light source function. T w o types of optical power meters are supported: One is measurement range of –70 to +3 dBm (MW9076B/B1/C-02 option), the other is measurement range of –50 to +23 dBm (MW9076B/B1/C-03 option). ● VGA Output T ...

  • Anritsu MW9076 - page 11

    11 ● Both-End Measurement Function A new w av eform can be composed by a veraging data mea- sured at both ends of an optical fiber . ● Data T ransmission Function Data files recorded by the MW9076 series can be transf erred to a PC via the RS-232C por t. ● Emulation Function Measured wa vef orm data can be analyzed using a PC. MX907600A O TDR ...

  • Anritsu MW9076 - page 12

    ● Report Output The ev ent table of a specified file is analyzed and the printed automatically . It is also possible to print multiple wav eforms on one page. In par ticular , at both ends measurement, the measurement results f or both ends can be output automati- cally in a repor t. In addition, the repor t can be sav ed as the Excel file. ● C ...

  • Anritsu MW9076 - page 13

    13 Model MW9076B MW9076C MW9076B1 MW9076J MW9076K MW9076D1 W avelength 1310/1550 nm 1310/1550/1625 nm 1310/1550 nm 850 nm 850/1300 nm 1310/1450/1550/ ±25 nm ∗ 1 ±25 nm ∗ 1 ±25 nm ∗ 1 ±30 nm ±30 nm 1625 nm ±3 nm ∗ 1 Measurable optical fiber 10/125 µm single-mode optical fiber (ITU-T G.652) 62.5/125 µm GI fiber ∗ 2 Optical connector ...

  • Anritsu MW9076 - page 14

    14 Model MW9076B MW9076C MW9076B1 MW9076J MW9076K MW9076D1 Applicable optical fibers: SM optical fiber (ITU-T G.652) Optical connectors: Shared with O TDR (same por t) Light-emitting elements: FP-LD Center wav elength: 1310/1550 ±25 nm (MW9076B, CW , 25°C) 1310/1550/1625 ±25 nm (MW9076C, CW , 25°C) Spectrum width: ≤ 5/10 nm (MW9076B, CW , 25? ...

  • Anritsu MW9076 - page 15

    Model MW9076B MW9076C MW9076B1 MW9076J MW9076K MW9076D1 15 ∗ 1 At 25°C, pulse width: 1 µs ∗ 2 For GI fiber (core diameter: 62.5 µm ±3.0 nm, NA: 0.275 ±0.015, transmission loss: ≤ 3.2/0.9 dB/km (wav elength: 0.85/1.3 µm). At measurement of 50/125 µm GI fiber , the dynamic range drops by about 3.0 dB. ∗ 3 At 25°C, pulse width: SM 20 ? ...

  • Anritsu MW9076 - page 16

    16 ● MU250000A/A1/A4 Display Unit MU250000A Unit: 8.4 inch color , TFT -LCD (640 × 480 pixels , transparent type, with bac k light) Display MU250000A1 Unit: 7.2 inch color , STN-LCD (640 × 480 pixels , semi-transparent type, with bac k light on/off) MU250000A4 Unit: 7.8 inch color , STN-LCD (640 x 480 pixels , reflective type , with front light ...

  • Anritsu MW9076 - page 17

    Ordering Information 17 Model/order No . Name Remarks Optical Time Domain Reflectometer (main frame) MW9076B SMF 1.31/1.55 µm O TDR Requires Display Unit MW9076B1 SMF 1.31/1.55 µm OTDR Requires Display Unit MW9076C SMF 1.31/1.55/1.625 µm O TDR Requires Display Unit MW9076D1 SMF 1.31/1.45/1.55/1.625 µm O TDR Requires Display Unit MW9076J GIF 0.8 ...

  • Anritsu MW9076 - page 18

    Model/order No . Name Remarks JT512MA3-NT1 PC-A T A card (512 MB) J0057 Optical adapter FC type T o connect optical fiber cable with FC connector J0635 ∗ 3 Optical fiber cord With FC-PC at both ends (SM) B0442 Soft carrying case For MW9076B/B1/C/J/K, 440 (W) × 310 (H) × 110 (D) mm Z0435 Soft carr ying case For MW9076D1(MW9076B/B1/C + MU960001A/ ...

  • Anritsu MW9076 - page 19

    Selection Guide 19 Visible LD (Option 01) Display Units Main Frames Options MU250000A 8.4" TFT -LCD F or indoor , exceedingly easy to see MU250000A1 7.2" STN-LCD F or indoor , easy to see MU250000A4 7.8" STN-LCD F or outdoor , exceedingly easy to see under direct sunshine MW9076B1: High cost perf or mance F or SM fiber F or GI fiber ...

  • Anritsu MW9076 - page 20

    ANRITSU CORPORA TION 5-10-27, Minamiazabu, Minato-ku, T okyo 106-8570, Japan Phone: +81 - 3 - 3446 - 1111 T elex: J34372 F ax: +81 - 3 - 3442 - 0235 • • U.S.A. ANRITSU COMP ANY North American Region Headquar ters 1155 East Collins Blvd., Richardson, Tx 75081, U.S .A. T oll Free: 1 - 800 - ANRITSU (267 - 4878) Phone: +1 - 972 - 644 - 1777 F ax: ...

Fabricante Anritsu Categoría Telescope

Los documentos del dispositivo Anritsu MW9076 que obtenemos del fabricante se pueden dividir en varios grupos. Entre ellos están::
- dibujos técnicos Anritsu
- manuales de instrucciones MW9076
- hojas de producto Anritsu
- folletos informativos
- o etiquetas energéticas Anritsu MW9076
Todos son importantes, pero lo más importante desde el punto de vista del usuario de un dispositivo lo encontraremos en el manual de instrucciones Anritsu MW9076.

Un conjunto de documentos determinado como manual de instrucciones se divide también en tipos más detallados, tales como: instrucciones de montaje Anritsu MW9076, instrucciones de servicio, instrucciones cortas o instrucciones de usuario Anritsu MW9076. Dependiendo de la situación debes buscar el documento que necesitas. En nuestra página puedes consultar el manual de instrucciones más popular del producto Anritsu MW9076.

Manuales de instrucciones parecidos

El manual de instrucciones del dispositivo Anritsu MW9076, ¿cómo debería ser?
El manual de instrucciones, también determinado como el manual de usuario o simplemente instrucciones, es un documento técnico que tiene como objetivo ayudar a los usuarios a utilizar Anritsu MW9076. Las instrucciones normalmente las realiza un escritor técnico pero en un lenguaje comprensible para todos los usuarios de Anritsu MW9076.

El manual de instrucciones completo de Anritsu debe contener unos elementos básicos. Una parte de ellos es menos importante, como por ejemplo: la portada / la página principal o las páginas de autor. Sin embargo, lo demás debe facilitarnos la información más importante desde el punto de vista del usuario.

1. Introducción y pistas sobre cómo orientarse con el manual Anritsu MW9076 - Al principio de cada manual se deben encontrar las indicaciones acerca de la manera de usar un manual de instrucciones. Debe contener información sobre dónde encontrar el índice Anritsu MW9076, preguntas frecuentes o problemas más comunes – son los apartados más buscados por los usuarios de cada manual de instrucciones.
2. Índice - listados de todos los consejos acerca de Anritsu MW9076 que encontraremos en el documento
3. Consejos de uso de las funciones básicas del dispositivo Anritsu MW9076 - que deben facilitarnos los primeros pasos durante el uso de Anritsu MW9076
4. Troubleshooting - una secuencia sistematizada de acciones que nos ayudará a diagnosticar los problemas más importantes con Anritsu MW9076
5. FAQ - las preguntas frecuentes
6. Datos de contacto Información acerca de cómo encontrar los datos de contacto del fabricante / servicio de Anritsu MW9076 en cada país si no somos capaces de solucionar el problema por nuestra cuenta.

¿Tienes una pregunta acerca de Anritsu MW9076?

Utiliza el formulario que se encuentra abajo

Si con la ayuda del manual de instrucciones encontrado no has podido solucionar tu problema con Anritsu MW9076, haz una pregunta utilizando el formulario que se encuentra abajo. Si alguno de los usuarios ha tenido un problema parecido con Anritsu MW9076 probablemente compartirá una forma de solucionarlo.

Copia el texto de la imagen

Comentarios (0)