    OPERATING INSTRUCTIONS AND S PECIFICATIONS NI 9237 4-Channel, 24-Bit Half/Full- Bridge Analog Input Module ni.com/manuals Deutsch Français ...

  • National Instruments NI 9237 - page 2

    NI 9237 Operating I nstruction s and Specificati ons 2 ni.com This doc u ment describes ho w to u se the National Instr u ments 9 237 and incl u des specif ications and pin assignments for the NI 9237. Vi s i t ni.com/info and enter rd s oftwarever s ion to determine which software yo u need for the mod u les yo u are u sing. For information abo u ...

  • National Instruments NI 9237 - page 3

    © National Inst ruments Corp. 3 NI 9237 Operating Instructi ons and Specific ations Safety Guidelines Operate the NI 9237 only as described in these operating instr u ctions. Hot Surface This icon denotes that the component may be hot. T o u ching this component may res u lt in bodily inj u ry . Safety Guidelines fo r Hazardous Loca tions The NI 9 ...

  • National Instruments NI 9237 - page 4

    NI 9237 Operating I nstruction s and Specificati ons 4 ni.com Caution S u bstit u tion of components may impair s u itability for Class I, Division 2. Caution For Zone 2 applications, install the system in an enclos u re rated to at least IP 54 as defined b y IEC 60529 an d EN 60529 . Caution For Zone 2 applications, connected signals m u st be wit ...

  • National Instruments NI 9237 - page 5

    © National Inst ruments Corp. 5 NI 9237 Operating Instructi ons and Specific ations Special Conditions fo r Marine Appli cations Some mod u les are Lloyd’ s Re gister (LR) T ype Approved for marine applications. T o verify Llo yd’ s Register cert ification, visit ni.com/certification and search for the LR certificate, or look for the Lloyd’ ...

  • National Instruments NI 9237 - page 6

    NI 9237 Operating I nstruction s and Specificati ons 6 ni.com Connecting the NI 9237 The NI 9237 has fo u r RJ-50 receptacles that pro vide connections for fo u r half or f u ll bridges. Figure 1. NI 9237 Pin Assignm ents SC AI+ AI– RS+ RS– EX+ EX– T+ T– SC 1 2 3 4 5 6 7 8 9 10 Ch0 – Ch3 EX+ EX+ EX– EX– Ch 0 Ch 1 Ch 2 Ch 3 ...

  • National Instruments NI 9237 - page 7

    © National Inst ruments Corp. 7 NI 9237 Operating Instructi ons and Specific ations When yo u connect a half bridge to the NI 9237, yo u m u st connect the AI+, EX–, and RS– signals, b u t yo u do not connect the AI– signal. When yo u connect a f u ll bridge, yo u m u st connect the AI+, AI–, EX–, and RS– si gnals. Re fer to Fig u re 2 ...

  • National Instruments NI 9237 - page 8

    NI 9237 Operating I nstruction s and Specificati ons 8 ni.com The NI 9237 has a fo u r- terminal e xternal e xcitation v oltage so u rce connector . Y o u can u se the EX+ and EX– terminals on the connector to connect one e x ternal e xcitation voltag e so u rce to the mod u le. Y o u can u se the additional EX+ and EX– terminals on the connect ...

  • National Instruments NI 9237 - page 9

    © National Inst ruments Corp. 9 NI 9237 Operating Instructions and Specific ations Figure 2. Connecting a Half or Full Bridge to the NI 9237 EX+ EX– SC SC RS+ RS– T+ T– AI– AI+ TEDS NI 9237 ...

  • National Instruments NI 9237 - page 10

    NI 9237 Operating I nstructions and Specifications 10 ni.com Each channel on the NI 9237 has an independent 24-bit ADC and an inp u t amplifier that enable yo u to sample signals from all fo u r channels sim u ltaneo u sly . The NI 9237 is isolated from earth gro u nd. Ho wev er, the indi vid u al channels are not isolated from each other . The EX+ ...

  • National Instruments NI 9237 - page 11

    © National Inst ruments Corp. 11 NI 9237 Operating Instructions and Specific ations Wiring TEDS Cha nnels Ens u re that neither the TEDS da ta (T+) nor the TEDS ret u rn (T–) signal is tied in common to any AI signals on the NI 9237. The NI 9237 connects all the T– signals together internally . V isit ni.com/info and enter the info code rdted ...

  • National Instruments NI 9237 - page 12

    NI 9237 Operating I nstructions and Specifications 12 ni.com circ u it. Refer to Fig u re 3 for an ill u stration of ho w to connect remote sense wires to the NI 9237. Figure 3. Connecti ng Remote Sense Wires to the NI 9237 EX+ EX– RS+ RS– R bridge R lead R lead R bridge R bridge R bridge NI 9237 AI– AI+ ...

  • National Instruments NI 9237 - page 13

    © National Inst ruments Corp. 13 NI 9237 Operating Instructions and Specific ations The act u al bridge e xcitation voltage is smaller than the v oltage at the EX+ and EX– leads. If yo u do not u se remote sensing of the act u al bridge voltage, the res u lting gain error is: for half-bridge sensors and for f u ll-bridge sensors. If yo u connect ...

  • National Instruments NI 9237 - page 14

    NI 9237 Operating I nstructions and Specifications 14 ni.com sensors beca u se th ere may be signif icant resistance in the wiring to the activ e resistor in the bridge. The NI 9237 sh u nt calibrati on circ u itry consists of a precision resistor and a software-controlle d switch. Refer to the software help for information abo u t enabling the sh ...

  • National Instruments NI 9237 - page 15

    © National Inst ruments Corp. 15 NI 9237 Operating Instructions and Specific ations Excitation Voltages Altho u gh the sensor ind u stry does not recognize a single standard excita tion v oltage le vel, e xcitation v oltage le vels of between 2.5 V and 10 V are common. Y o u can program the NI 9237 to s u pply 2.5 V , 3.3 V , 5 V , or 10 V of exci ...

  • National Instruments NI 9237 - page 16

    NI 9237 Operating I nstructions and Specifications 16 ni.com The 150 mW limit allows yo u to po wer half and f u ll bridges as follows: •F o u r 350 Ω half bridges at 5.0 V •F o u r 350 Ω f u ll bridges at 3.3 V •F o u r 120 Ω half bridges at 2.5 V If yo u need an e xcitation voltage that ca u ses more than 150 mW to dissipate across all t ...

  • National Instruments NI 9237 - page 17

    © National Inst ruments Corp. 17 NI 9237 Operating Instructions and Specific ations or signals that ha ve been f iltered by at least the amo u nt of the stopband rejection. Passband The signals within the passband have freq u enc y-dependent gain or atten u ation. The small am o u nt of var iation in gain with respect to freq u ency is called the ...

  • National Instruments NI 9237 - page 18

    NI 9237 Operating I nstructions and Specifications 18 ni.com Figure 4. Typical Passband Flat ness for the NI 9237 Stopband The filter signif icantly atten u ates all signals abo ve the stop band freq u ency . The primary goal of the f ilter is to prev ent aliasing. Therefore, the stopband freq u ency scales precisel y with the data F requency/Data ...

  • National Instruments NI 9237 - page 19

    © National Inst ruments Corp. 19 NI 9237 Operating Instructions and Specific ations rate. The stopband rejection is the minim u m amo u nt of att en u ation applied by the filter to all signals with freq u en cies within the stopband. Alias-Free Bandwidth Any signal that appears in the alias-free bandwidth of the NI 9237 is not an aliased artifact ...

  • National Instruments NI 9237 - page 20

    NI 9237 Operating I nstructions and Specifications 20 ni.com The follo wing eq u ation provide s the a vailable da ta rates of the NI 9237: f s = where n is any inte ger from 1 to 31. Howe ver , the data rate m u st rema in within the appropriate data rate range. Refer to the Specifications sectio n for more information abo u t the data rate range. ...

  • National Instruments NI 9237 - page 21

    © National Inst ruments Corp. 21 NI 9237 Operating Instructions and Specific ations s u pport for sleep mode. If the chassis s u pports sleep mode, refer to the software help for information abo u t enabli ng slee p mode. V isit ni.com/info and enter c s erie s doc for information abo u t C Series doc u mentation. T ypically , when a system is in ...

  • National Instruments NI 9237 - page 22

    NI 9237 Operating I nstructions and Specifications 22 ni.com T ype of ADC ..... ........... ........... ......... . De lta-Sigma (w ith analog prefiltering) Sampling mode . ........... ........... .......... Sim u ltaneo u s Internal master timebase ( f M ) Freq u ency ..... ........... ........... ........ 12.8 MHz Acc u racy......... ........... ...

  • National Instruments NI 9237 - page 23

    © National Inst ruments Corp. 23 NI 9237 Operating Instructions and Specific ations Overv oltage protection between any two pins .......... ............. . ±30 V Acc u racy Gain drift ...... ........... ......... ............ ..... 10 ppm/°C max Offset drift 2.5 V excitation ............ ........... ... 0.6 μ V/V per °C 3.3 V excitation ...... ...

  • National Instruments NI 9237 - page 24

    NI 9237 Operating I nstructions and Specifications 24 ni.com Channel-to-channel ma tching (calibrated) Phase nonlineari ty f in = 0 to 1 kHz ....... ........... ......... . < 0.001° f in = 0 to 20 kHz.............. ........... . ±0. 1° Inp u t delay .......... ........... ........... ........ 38.4/ f s + 4.8 μ s Passband Freq u ency ..... .. ...

  • National Instruments NI 9237 - page 25

    © National Inst ruments Corp. 25 NI 9237 Operating Instructions and Specific ations Rejection at o versa mple rate 1 f s = 10 kS/s........ ............ ........... ... 60 dB @ 640 kHz f s = 50 kS/s........ ............ ........... ... 90 dB @ 3.2 MHz Common-mode vo ltage, all signals to earth gro u nd ......... ........ ±60 VDC CMRR Relati ve to ...

  • National Instruments NI 9237 - page 26

    NI 9237 Operating I nstructions and Specifications 26 ni.com Inp u t noise Excitation noise ........... ......... ........... . 0 .1 mV/V rms Crosstalk f in = 1 kHz ....... ........... ............ ..... 110 dB f in = 10 kHz .............. ......... .......... 100 dB Excitation Internal voltage ........ ........... ........ 2.5 V , 3.3 V , 5.0 V , ...

  • National Instruments NI 9237 - page 27

    © National Inst ruments Corp. 27 NI 9237 Operating Instructions and Specific ations Sh u nt calibration Resistance .............. ......... ............ 100 k Ω Resistor acc u racy 25°C .............. .................. ..... ±110 Ω –40 to 70°C ......... ............ ..... ±200 Ω MTBF .......... ........... ............ ........... ... 60 ...

  • National Instruments NI 9237 - page 28

    NI 9237 Operating I nstructions and Specifications 28 ni.com Physical Characteristics If yo u need to clean the mod u l e, wipe it with a dry to wel. W eight .......... ........... ............ .............. 152 g (5.4 oz) Safety Safety Voltages Connect only v oltages that are within the follo wing limits. Between any two pins ...... ............ ...

  • National Instruments NI 9237 - page 29

    © National Inst ruments Corp. 29 NI 9237 Operating Instructions and Specific ations meas u rements of voltages from spec ially protecte d secondary circ u its. S u ch v oltage meas u reme nts incl u de signal le vels, special eq u ipment, limited-ene rg y parts of eq u ipment , circ u its po wered by reg u lated lo w-voltage so u rces, and electro ...

  • National Instruments NI 9237 - page 30

    NI 9237 Operating I nstructions and Specifications 30 ni.com Safety Standards This prod u ct meets the req u irements of the following standards of safety for electrica l eq u ipment for meas u rement, control, and laboratory u se: • IEC 61010-1, EN 61010-1 • UL 61010-1, CSA 61010-1 Note For UL and other safety certif icat ions, refer to the pr ...

  • National Instruments NI 9237 - page 31

    © National Inst ruments Corp. 31 NI 9237 Operating Instructions and Specific ations Note For the standards applied to assess the EMC of this prod u ct, refer to the Online Product Certification section. Note For EMC compliance, operate this de vice with shielded cabling. CE Compliance This prod u ct meets the essent ial req u irem ents of applic a ...

  • National Instruments NI 9237 - page 32

    NI 9237 Operating I nstructions and Specifications 32 ni.com Shock and Vibration T o meet thes e specifications, yo u m u st panel mo u nt the system. Operating vibration Random (IEC 60068-2-64)......... 5 g rms , 10 to 500 Hz Sin u soidal (IEC 60068-2-6) ....... 5 g, 10 to 500 Hz Operating shock (IEC 60068-2-27)........... ........... ........ 30 ...

  • National Instruments NI 9237 - page 33

    © National Inst ruments Corp. 33 NI 9237 Operating Instructions and Specific ations Operating h u midity (IEC 60068-2-56)....... ........... ............ 10 to 90% RH, noncondensing Storage h u midity (IEC 60068-2-56)....... ........... ............ 5 to 95% RH, noncondensing Maxim u m altit u de.......... ......... .......... 2,000 m Poll u tion ...

  • National Instruments NI 9237 - page 34

    NI 9237 Operating I nstructions and Specifications 34 ni.com Waste Electrical and Electronic Equipment (WEEE) EU Customers At the end of the life c ycle, all prod u cts must be sent to a WEEE re c ycling center . For more information abo u t WEEE recycling ce nters and National Instr u ments WEEE initiati ves, visit ni.com/ environment/weee . Calib ...

  • National Instruments NI 9237 - page 35

    © National Inst ruments Corp. 35 NI 9237 Operating Instructions and Specific ations Where to Go for Support The National Instr u ments W eb site is yo u r complete reso u rce for technical s u pport. At ni.com/ s upport yo u have access to ev erything from tro u bleshooting and applic ation dev elopment self-help reso u rces to email and phone ass ...

  • National Instruments NI 9237 - page 36

    © 2006–2008 Nati onal Instruments Corp. All rights reserved. 374186C-01 Aug08 National Instruments, NI, n i .com, and LabVIEW are trademarks of National Instruments C orporation. Refer to the Terms of Use section on ni.com/legal for more information about National Instruments t rademarks. Other product and company names ment ioned herein are tra ...

Fabricante National Instruments Categoría Network Card

Los documentos del dispositivo National Instruments NI 9237 que obtenemos del fabricante se pueden dividir en varios grupos. Entre ellos están::
- dibujos técnicos National Instruments
- manuales de instrucciones NI 9237
- hojas de producto National Instruments
- folletos informativos
- o etiquetas energéticas National Instruments NI 9237
Todos son importantes, pero lo más importante desde el punto de vista del usuario de un dispositivo lo encontraremos en el manual de instrucciones National Instruments NI 9237.

Un conjunto de documentos determinado como manual de instrucciones se divide también en tipos más detallados, tales como: instrucciones de montaje National Instruments NI 9237, instrucciones de servicio, instrucciones cortas o instrucciones de usuario National Instruments NI 9237. Dependiendo de la situación debes buscar el documento que necesitas. En nuestra página puedes consultar el manual de instrucciones más popular del producto National Instruments NI 9237.

El manual de instrucciones del dispositivo National Instruments NI 9237, ¿cómo debería ser?
El manual de instrucciones, también determinado como el manual de usuario o simplemente instrucciones, es un documento técnico que tiene como objetivo ayudar a los usuarios a utilizar National Instruments NI 9237. Las instrucciones normalmente las realiza un escritor técnico pero en un lenguaje comprensible para todos los usuarios de National Instruments NI 9237.

El manual de instrucciones completo de National Instruments debe contener unos elementos básicos. Una parte de ellos es menos importante, como por ejemplo: la portada / la página principal o las páginas de autor. Sin embargo, lo demás debe facilitarnos la información más importante desde el punto de vista del usuario.

1. Introducción y pistas sobre cómo orientarse con el manual National Instruments NI 9237 - Al principio de cada manual se deben encontrar las indicaciones acerca de la manera de usar un manual de instrucciones. Debe contener información sobre dónde encontrar el índice National Instruments NI 9237, preguntas frecuentes o problemas más comunes – son los apartados más buscados por los usuarios de cada manual de instrucciones.
2. Índice - listados de todos los consejos acerca de National Instruments NI 9237 que encontraremos en el documento
3. Consejos de uso de las funciones básicas del dispositivo National Instruments NI 9237 - que deben facilitarnos los primeros pasos durante el uso de National Instruments NI 9237
4. Troubleshooting - una secuencia sistematizada de acciones que nos ayudará a diagnosticar los problemas más importantes con National Instruments NI 9237
5. FAQ - las preguntas frecuentes
6. Datos de contacto Información acerca de cómo encontrar los datos de contacto del fabricante / servicio de National Instruments NI 9237 en cada país si no somos capaces de solucionar el problema por nuestra cuenta.

¿Tienes una pregunta acerca de National Instruments NI 9237?

Utiliza el formulario que se encuentra abajo

Si con la ayuda del manual de instrucciones encontrado no has podido solucionar tu problema con National Instruments NI 9237, haz una pregunta utilizando el formulario que se encuentra abajo. Si alguno de los usuarios ha tenido un problema parecido con National Instruments NI 9237 probablemente compartirá una forma de solucionarlo.

Copia el texto de la imagen

Comentarios (0)