    USER MANUAL AIR-O XYGEN BLENDER (DIS S and NIST Connections) Model No. PM5200 Series PM5300 Series (shown) SAVE THESE INSTRUCTIONS CA UTION F ederal (USA) law restricts this device to sale by or on the order of a physician. 300 Held Drive T el: (+001) 610-262-6090 Northampton, P A 18067 USA F ax: (+001) 610-262-6080 www .precisionmedical.com ...

  • Precision Medical PM5200 - page 2

    1 Blender Air-Oxygen CONTENTS RECEIVING / INSPECTION ......................................................... 2 INTENDED USE .......................................................................... 2 READ ALL INSTRUCTIONS BEFORE USING ............................ 2 EXPLANA TION OF ABBREVIA TIONS ......................................... 2 SAFETY ...

  • Precision Medical PM5200 - page 3

    2 Blender Air-Oxygen RECEIVING / INSPECTION Remove the Precision Medical, Inc. Air -Oxygen Blender from the packaging and inspect for damage. If there is any damage, DO NOT USE and contact your Provider . INTENDED USE Precision Medical, Inc. Air -Oxygen Blender dispenses a continuous and precise blend of medical air and USP oxygen via outlet ports ...

  • Precision Medical PM5200 - page 4

    3 Blender Air-Oxygen SAFETY INFORMA TION - W ARNINGS AND CAUTIONS DANGER Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury . W ARNING Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury . CA UTION Indicates a potentially hazardous situatio ...

  • Precision Medical PM5200 - page 5

    4 Blender Air-Oxygen W ARNING An Oxygen Analyzer/Monitor must be used to verify oxygen concentration. Accuracy of oxygen concentration will be affected if bleed is not activated at flow settings below 15 lpm for the High Flow Blender , and 3 lpm for the Low Flow Blender . DO NOT obstruct the alarm. DO NO T use Blender when alarm is sounding. DO NO ...

  • Precision Medical PM5200 - page 6

    5 Blender Air-Oxygen SPECIFICA TIONS Model PM5200 High Flow PM5300 Low Flow Primary Outlet Flow Range 15 - 120 lpm 3 - 30 lpm W i t h b o t h s u p p l y p r e s s u r e s a t 5 0 p s i ( 3 . 4 b a r ) w i t h N O B L E E D Auxiliary Outlet Flow Range 2 - 100 lpm 0 - 30 lpm W i t h b o t h s u p p l y p r e s s u r e s a t 5 0 p s i ( 3 . 4 b a r ) ...

  • Precision Medical PM5200 - page 7

    6 Blender Air-Oxygen SPECIFICA TIONS continued Dimensions: (without fittings) Depth: 4.9 in (12.5 cm) Width: 2.3 in (5.7 cm) Height: 4.1 in (10.4 cm) W eight: 2.9 lbs (1.3 kg) Shipping W eight: 3.5 lbs (1.6 kg) Operating T emperature Range: 59°F to 104°F (15°C to 40°C) T ransport / Storage Requirements T emperature Range: -10°F to 140°F (-23 ...

  • Precision Medical PM5200 - page 8

    7 Blender Air-Oxygen DIAGRAMS CA UTION Missing or illegible labels must be replaced, contact Precision Medical, Inc. Depending on model, your fittings may differ from these diagrams. PM5300 Model B F A E PM5200 Model B G D C I H ...

  • Precision Medical PM5200 - page 9

    8 Blender Air-Oxygen COMPONENT DESCRIPTION ITEM DESCRIPTION A Oxygen Concentration Dial A d ia l u se d f or se le ct ing ox yg en c on ce ntr at ion s b et we e n 21%-100%. The FiO 2 scale is used for reference only . This Dial does not rotate 360°. The dial starts at 21% and ends at 100%. B Primary Outlet Port A ma le DISS oxy ge n fit tin g wi ...

  • Precision Medical PM5200 - page 10

    9 Blender Air-Oxygen COMPONENT DESCRIPTION ITEM DESCRIPTION F Air Inlet Fitting A male DISS or NIST air fitting with one way valve that is used to connect an air supply hose . G Alarm An audible alarm that sounds due to an excessive pressure drop or deletion of either gas supply . H Manifold Outlet (Optional) Manifold with 3 primary outlets. I Rea ...

  • Precision Medical PM5200 - page 11

    10 Blender Air-Oxygen ALARM TEST 1. The Air -Oxygen Blender is installed and the flowmeter is turned on. Disconnect or turn off the air supply line to the Air -Oxygen Blender . 2. The Blender should alarm with a loud whistle noise. The whistle indicates the alarm is operating correctly . 3. Reconnect and activate the air supply line to the Blender ...

  • Precision Medical PM5200 - page 12

    11 Blender Air-Oxygen 3. T o activate the bleed, turn the knurled collar until the bleed pin is engaged. Then slide the collar back until it contacts the cover . 4. Confirm the concentration of air/oxygen with an Oxygen Analyzer/Monitor . 5. Confirm the flow of air and/or oxygen mixture to the patient. 6. T urn “OFF” the Air and Oxygen suppl ...

  • Precision Medical PM5200 - page 13

    12 Blender Air-Oxygen 1. Disconnect all gas connections and equipment before cleaning. 2. Clean exterior surfaces with a cloth dampened with mild detergent and water . 3. Wipe dry with a clean cloth. MAINTENANCE The following maintenance on the Air -Oxygen Blender must be performed by a trained service technician: Every month verify proper function ...

  • Precision Medical PM5200 - page 14

    13 Blender Air-Oxygen TROUBLESHOOTING If the Air -Oxygen Blender fails to function, consult the T roubleshooting Guide below . If problem cannot be solved by using T roubleshooting Guide refer to the Air -Oxygen Blender Service Manual (P/N 504827) available on the Internet; www .precisionmedical.com or consult your Provider . Problem Probable Cause ...

  • Precision Medical PM5200 - page 15

    14 Blender Air-Oxygen LIMITED W ARRANTY AND LIMIT A TION OF LIABILITY Precision Medical, Inc. warrants that the Blender , (the Product), will be free of defects in workmanship and/or material for the following period: T wo (2) years from shipment Should any failure to conform to this warranty appear within the applicable period, Precision Medical, ...

  • Precision Medical PM5200 - page 16

    DECLARA TION OF CONFORMITY Manufacturer: Precision Medical, Inc. 300 Held Drive, Northampton, P A 18067, USA CONT ACT : Quality Manager Phone: 610-262-6090 Authorized European Representative: Emergo Europe Molenstraat 15 2513 BH, The Hague The Netherlands Product: Gas Mixers for Medical Use Model(s): PM5200EN, PM5200MEN, PM5300EN, PM5300MEN MDD Cla ...

Fabricante Precision Medical Categoría Respiratory Product

Los documentos del dispositivo Precision Medical PM5200 que obtenemos del fabricante se pueden dividir en varios grupos. Entre ellos están::
- dibujos técnicos Precision Medical
- manuales de instrucciones PM5200
- hojas de producto Precision Medical
- folletos informativos
- o etiquetas energéticas Precision Medical PM5200
Todos son importantes, pero lo más importante desde el punto de vista del usuario de un dispositivo lo encontraremos en el manual de instrucciones Precision Medical PM5200.

Un conjunto de documentos determinado como manual de instrucciones se divide también en tipos más detallados, tales como: instrucciones de montaje Precision Medical PM5200, instrucciones de servicio, instrucciones cortas o instrucciones de usuario Precision Medical PM5200. Dependiendo de la situación debes buscar el documento que necesitas. En nuestra página puedes consultar el manual de instrucciones más popular del producto Precision Medical PM5200.

Manuales de instrucciones parecidos

El manual de instrucciones del dispositivo Precision Medical PM5200, ¿cómo debería ser?
El manual de instrucciones, también determinado como el manual de usuario o simplemente instrucciones, es un documento técnico que tiene como objetivo ayudar a los usuarios a utilizar Precision Medical PM5200. Las instrucciones normalmente las realiza un escritor técnico pero en un lenguaje comprensible para todos los usuarios de Precision Medical PM5200.

El manual de instrucciones completo de Precision Medical debe contener unos elementos básicos. Una parte de ellos es menos importante, como por ejemplo: la portada / la página principal o las páginas de autor. Sin embargo, lo demás debe facilitarnos la información más importante desde el punto de vista del usuario.

1. Introducción y pistas sobre cómo orientarse con el manual Precision Medical PM5200 - Al principio de cada manual se deben encontrar las indicaciones acerca de la manera de usar un manual de instrucciones. Debe contener información sobre dónde encontrar el índice Precision Medical PM5200, preguntas frecuentes o problemas más comunes – son los apartados más buscados por los usuarios de cada manual de instrucciones.
2. Índice - listados de todos los consejos acerca de Precision Medical PM5200 que encontraremos en el documento
3. Consejos de uso de las funciones básicas del dispositivo Precision Medical PM5200 - que deben facilitarnos los primeros pasos durante el uso de Precision Medical PM5200
4. Troubleshooting - una secuencia sistematizada de acciones que nos ayudará a diagnosticar los problemas más importantes con Precision Medical PM5200
5. FAQ - las preguntas frecuentes
6. Datos de contacto Información acerca de cómo encontrar los datos de contacto del fabricante / servicio de Precision Medical PM5200 en cada país si no somos capaces de solucionar el problema por nuestra cuenta.

¿Tienes una pregunta acerca de Precision Medical PM5200?

Utiliza el formulario que se encuentra abajo

Si con la ayuda del manual de instrucciones encontrado no has podido solucionar tu problema con Precision Medical PM5200, haz una pregunta utilizando el formulario que se encuentra abajo. Si alguno de los usuarios ha tenido un problema parecido con Precision Medical PM5200 probablemente compartirá una forma de solucionarlo.

Copia el texto de la imagen

Comentarios (0)