    Dealer: This manual MUST be given to the end user. User: BEFORE using this product, read this manual and save for future reference. Operator’s Manual For more information regarding Invacare products, parts, and services, please visit www.invacare.com Perfecto 2 ™ Series Oxygen Concentrators without SensO 2 ®, Oxygen Concentrators with SensO 2 ...

  • Invacare IRC5PO2 - page 2

    WARNING Perfecto 2 ™ Series 2 Part No 114348 2  WARNING DO NOT OPERATE THIS EQUIPMENT WITHOUT FIRST READING AND UNDERSTAN DING THIS MANUAL. IF YOU ARE UN ABLE TO UNDERSTAND THE WARNINGS, CAUTIONS, AND INSTRUCTIONS, CONTACT INVACARE TECHNICAL SERVICES BEFORE ATTEMPTING TO USE THIS EQUIPMENT - OTHERWISE SERIOUS IN JURY OR PROPERTY DAMAGE MAY RES ...

  • Invacare IRC5PO2 - page 3

    TABLE OF CONTENTS Part No 1143482 3 Perfecto 2 ™ Series Table of Contents ACCESSORIES .................................................. 2 SPECIAL NOTES ............................................... 5 SECTION 1—GENERAL GUIDELINES .................. 6 Maintenance ......................... ...................................................... ...

  • Invacare IRC5PO2 - page 4

    Table of Contents TABLE OF CONTENTS Perfecto 2 ™ Series 4 Part No 114348 2 Indicators (If Your Unit Does Not Feature the O 2 Sensor) ... 24 SECTION 6—MAINTENANCE .......................... 25 Routine Maintenanc e ....................... ................................................. 25 SECTION 7—TROUBLES HOOTING GUIDE ......... 28 LIMITED ...

  • Invacare IRC5PO2 - page 5

    SPECIAL NOTES Part No 1143482 5 Perfecto 2 ™ Series SPECIAL NOTES Signal words are used in this manual and apply to hazards or unsafe practices which could re sult in personal injury or property damage. Refer to the following table for definitions of the signal words. NOTICE The information contained in th is document is subject to change without ...

  • Invacare IRC5PO2 - page 6

    SECTION 1—GENERAL GUIDELINES Perfecto 2 ™ Series 6 Part No 114348 2 SECTION 1—GENERAL GUIDELINES In order to ensure the safe installation, assembly and operation of the P erfecto 2 concentrator these instructions MUST be followed. CAUTION “Caution: Federal law restricts this device to sale or rental by or on order of a physician, or any oth ...

  • Invacare IRC5PO2 - page 7

    SECTION 1—GENERAL GUIDELINES Part No 1143482 7 Perfecto 2 ™ Series  DANGER The use of oxygen therapy requires that special care be taken to reduce the risk of fire. Any materials that will burn in air , and some that will no t, are easily ignited and burn rapidly in high concentrations of oxygen . For safety concerns, it is necessary that al ...

  • Invacare IRC5PO2 - page 8

    SECTION 1—GENERAL GUIDELINES Perfecto 2 ™ Series 8 Part No 114348 2  WARNING NEVER block the air openings of the product or place it on a soft surface, such as a bed or couch, where the air op ening may be blocked. Keep the openings free from lint, hair and the like. Fill humidifier with w ater to the lev el shown by the manufacturer . DO NO ...

  • Invacare IRC5PO2 - page 9

    SECTION 1—GENERAL GUIDELINES Part No 1143482 9 Perfecto 2 ™ Series  WARNING To Reduce The Risk Of Burns, Electrocution, Fire Or Injury To Persons. A void using while bathing. If co ntinuous usage is required by the physician’ s prescription, the concentrator MUST be located in another room at least seven feet from the bath. DO NOT come in ...

  • Invacare IRC5PO2 - page 10

    SECTION 1—GENERAL GUIDELINES Perfecto 2 ™ Series 10 Part No 1143482  WARNING Use of some administration acce ssories or certain humidifiers not specified for use with oxyge n concentrator may impair the performance. In certain circumstances oxygen therapy can be hazardous. Inv acare recommends that y ou seek medical advice before using this ...

  • Invacare IRC5PO2 - page 11

    SECTION 2—FEATURES Part No 1143482 11 Perfecto 2 ™ Series SECTION 2—FEATURES Oxygen Outlet Flowmeter Oxygen Purity Indicator Lights / Fault and Power Indicator Lights Power Switch Circuit Breaker Elapsed Time Meter Cabinet Filter REAR VIEW ACCESSORIES (NOT SHOWN): PreciseRX™ Pedi atric Humidifier/Flowmeter Accessory - IRCPF16 (Perfecto 2 5 ...

  • Invacare IRC5PO2 - page 12

    SECTION 3—HANDLING Perfecto 2 ™ Series 12 Part No 1143482 SECTION 3—HANDLING The concentrator should AL W A YS be kept in the upright position to prevent cabinet damage while being transported. The shipping container has been designed to assure maximum protection of the concentrator . If the concentrator is to be reshipped by common carrier , ...

  • Invacare IRC5PO2 - page 13

    SECTION 4—TYPICAL PRODUCT PARAMETERS Part No 1143482 13 Perfecto 2 ™ Series SECTION 4—TYPICAL PRODUCT PARAMETERS Electrical Requirements: 120 VAC ± 10% (132 VAC/10 8 VAC), 60 Hz Rated Current Input: Perfecto 2 5: 3 A Sound Level: Perfecto 2 5: 44 ± 2 dBA Average Altitude: Perfecto 2 5 - Up to 6 ,000 ft. (1828 meters) above sea level withou ...

  • Invacare IRC5PO2 - page 14

    SECTION 4—TYPICAL PRODUCT PARAMETERS Perfecto 2 ™ Series 14 Part No 1143482 *NOTE: Stated concentration levels achieved after initial warm-up period (approximately 30 minutes). Safety System: Current overload or li ne surge shutdown. High temperature compressor shutdown. High Pressure Alarm w/compressor shut- down. Low Pressure Alarm w/compress ...

  • Invacare IRC5PO2 - page 15

    SECTION 4—TYPICAL PRODUCT PARAMETERS Part No 1143482 15 Perfecto 2 ™ Series IRC5PO2 INDICATO RS IRC5P INDICATORS LABEL SYMBOL O 2 PURITY INDICATOR LIGHTS (LED) SYSTEM OKAY O 2 over 85% (± 2%) GREEN Indicator Light O 2 Between 73% (± 3%) to 85% (±2%) YELLOW Indicator light A. YELLOW Solid B. YELLOW Flashing Sensor Failure Call a qualif ied te ...

  • Invacare IRC5PO2 - page 16

    SECTION 5—OPERATING INSTRUCTIONS Perfecto 2 ™ Series 16 Part No 1143482 SECTION 5— OPERATING INSTRUCTIONS Introduction Y our oxygen concentrator is inte nded for individual use in the home. It is an electronically operated device that separates oxygen from room air . It prov ides high concentration of oxygen directly to you through a nasal ca ...

  • Invacare IRC5PO2 - page 17

    SECTION 5—OPERATING INSTRUCTIONS Part No 1143482 17 Perfecto 2 ™ Series Recommended Guidelines for Optimum Performance NOTE: Ensure that your concentrator is at least thr ee inches away from walls, draperies or furnitur e to assure su fficient air flow . Avoid deep pile carpets and heaters, radiators or hot air registers. Plug in Power Cord 1. ...

  • Invacare IRC5PO2 - page 18

    SECTION 5—OPERATING INSTRUCTIONS Perfecto 2 ™ Series 18 Part No 1143482 Connect Humidifier (If So Prescribed)  WARNING For optimum performance, Invacare recommends that each concentrator be on and running for a minimum of 30 minutes at a time. Shorter pe riods of operation may reduce maximum product life. DO NOT overfill humidifier. DO NOT r ...

  • Invacare IRC5PO2 - page 19

    SECTION 5—OPERATING INSTRUCTIONS Part No 1143482 19 Perfecto 2 ™ Series FIGURE 5.2 Humidifer Bottle Adapter 5. Attach the humidifier bottle adapter to the humidifier bottle by turning the wing nut on the humidifier bottle counterclockwise until it is securely fastened. See Detail “ A” in FIGURE 5.3. 6. Place the humidifier bottle/adapter as ...

  • Invacare IRC5PO2 - page 20

    SECTION 5—OPERATING INSTRUCTIONS Perfecto 2 ™ Series 20 Part No 1143482 FIGURE 5.3 Humidifier Compartmen t ...

  • Invacare IRC5PO2 - page 21

    SECTION 5—OPERATING INSTRUCTIONS Part No 1143482 21 Perfecto 2 ™ Series Power Switch NOTE: For this procedur e, refer to FIGURE 5.4. 1. Press pow er switch to on po sition. All the panel lights and the audible alarm will come on for one second, indicating that the unit is functionin g properly . After one second, only the GREEN system ok/power ...

  • Invacare IRC5PO2 - page 22

    SECTION 5—OPERATING INSTRUCTIONS Perfecto 2 ™ Series 22 Part No 1143482 2. If the flowrate on the flowme ter ev er falls below 0.5 L/min for P erfecto 2 5 for more than about one minute, the LOW FLOW alarm will be triggered. This is a rapid beeping of the audible alarm. Check your tubing or accessories for blocked or kinked tubing or a defectiv ...

  • Invacare IRC5PO2 - page 23

    SECTION 5—OPERATING INSTRUCTIONS Part No 1143482 23 Perfecto 2 ™ Series Initial Startup of the Concentrator NOTE: Concentrator may be used during the initial start warm-up time (approximately 30 min.) while waiting for the O 2 purity to reach maximum. When the unit is turned on, the GREEN light will come on (SYSTEM OK/O 2 greater than 85%). Aft ...

  • Invacare IRC5PO2 - page 24

    SECTION 5—OPERATING INSTRUCTIONS Perfecto 2 ™ Series 24 Part No 1143482 Indicators (If Your Unit Does Not Feature the O 2 Sensor) NOTE: For this procedur e, refer to FIGURE 5.7. IRC5P Model Indicator Light Explanation RED light ( ) - T otal Unit Shut-Down. Switch IMMEDIA TEL Y to a back-up ox ygen supply . Call supplier IMMEDIA TEL Y . GREEN li ...

  • Invacare IRC5PO2 - page 25

    SECTION 6—MAINTENANCE Part No 1143482 25 Perfecto 2 ™ Series SECTION 6—MAINTENANCE NOTE: The Invacare concentrators ar e specifically designed to minimize routine pr even tive mainte nance. In places with high dust or soot levels, maintenance may need to be performed mor e often. Only qualified personnel should p erform preventive maintenance ...

  • Invacare IRC5PO2 - page 26

    SECTION 6—MAINTENANCE Perfecto 2 ™ Series 26 Part No 1143482 FIGURE 6.1 Cleaning the Cabinet Filter Cabinet  WARNING UNPLUG the conc entrator when cleaning. To avoid electrical shock, DO NOT remove cabinet. 1. Clean the cabinet with a mild household cleaner and non-abrasive cloth or sponge. Humidifier NOTE: T o clean the oxygen humidifier , ...

  • Invacare IRC5PO2 - page 27

    SECTION 6—MAINTENANCE Part No 1143482 27 Perfecto 2 ™ Series PREVENTIVE MAINTENANCE RECORD Model No. IRC _______________________________ Serial No.____________________________________ On Each Inspection *NOTE: Refer to Preventive Ma inte nance se ction of Ser vice Manual. Record Date Of Service Record Elapsed Hours on Hour Meter Clean Cabinet F ...

  • Invacare IRC5PO2 - page 28

    SECTION 7—TROUBLESHOOTING GUIDE Perfecto 2 ™ Series 28 Part No 1143482 SECTION 7—TROUBLESHOOTING GUIDE SYMPTOM: PROBABLE CAUSE: SOLUTION: Alarm: Main Power Loss: Short beeps, long pause 1. Power cord not plugged in. 1. Insert plug in to outlet. Concentrator not operating, power switch On. 2. No power at outlet. 2 . Inspect house circuit break ...

  • Invacare IRC5PO2 - page 29

    SECTION 7—TROUBLESHOOTING GUIDE Part No 1143482 29 Perfecto 2 ™ Series YELLOW or RED light Illuminated. Alarm: Continuo us On RED light only. * Only applies to 5PO 2 models. 1. Low oxygen purity.* 1. Clean or Replace filters. 2. Kinked or blocked tubing, cannula or humidifier.* 2. Inspect for kinks or blockages. Correct, clean or replace item. ...

  • Invacare IRC5PO2 - page 30

    SECTION 7—TROUBLESHOOTING GUIDE Perfecto 2 ™ Series 30 Part No 1143482 Alarm: Rapid Beep....Beep... Beep...Beep LOW FLOW ALARM 1a. Kinked or blocked tubing, cannula or humidifier. 1b. Flowmeter set at 0.5 L/min for Perfecto 2 5. 1a. Inspect for kinks or blockages. Correct, clean or replace item. Once corrected, turn power Off for 60 seconds and ...

  • Invacare IRC5PO2 - page 31

    LIMITED WARRANTY Part No 1143482 31 Perfecto 2 ™ Series LIMITED WARRANTY NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURE D AFTER JULY 4, 1975. This warranty is extended only to th e original purchaser who purchases this product when new and unused from Invacare Corporation or a dealer. This ...

  • Invacare IRC5PO2 - page 32

    LIMITATIONS AND EXCLUSIONS: THE FOREGOING WARRANTY SHALL NOT APPLY TO PRODUCTS SUBJECTED TO NEGLIGENCE, ACCIDENT, IMPROPER OPERATION, MAINTENANCE OR STORAGE, SOOT OR SMOKE-FILLED ENVI RONMENTS, OR OTHER THAN NORMAL APPLICATION, USE OR SERVICE, OR TO PRODUCTS MODIFIED WITHOUT INVACARE CORPORATION'S EXPRESS WRITTEN CONSENT (INCLUDIN G, BUT NOT L ...

Fabricante Invacare Categoría Respiratory Product

Los documentos del dispositivo Invacare IRC5PO2 que obtenemos del fabricante se pueden dividir en varios grupos. Entre ellos están::
- dibujos técnicos Invacare
- manuales de instrucciones IRC5PO2
- hojas de producto Invacare
- folletos informativos
- o etiquetas energéticas Invacare IRC5PO2
Todos son importantes, pero lo más importante desde el punto de vista del usuario de un dispositivo lo encontraremos en el manual de instrucciones Invacare IRC5PO2.

Un conjunto de documentos determinado como manual de instrucciones se divide también en tipos más detallados, tales como: instrucciones de montaje Invacare IRC5PO2, instrucciones de servicio, instrucciones cortas o instrucciones de usuario Invacare IRC5PO2. Dependiendo de la situación debes buscar el documento que necesitas. En nuestra página puedes consultar el manual de instrucciones más popular del producto Invacare IRC5PO2.

Manuales de instrucciones parecidos

El manual de instrucciones del dispositivo Invacare IRC5PO2, ¿cómo debería ser?
El manual de instrucciones, también determinado como el manual de usuario o simplemente instrucciones, es un documento técnico que tiene como objetivo ayudar a los usuarios a utilizar Invacare IRC5PO2. Las instrucciones normalmente las realiza un escritor técnico pero en un lenguaje comprensible para todos los usuarios de Invacare IRC5PO2.

El manual de instrucciones completo de Invacare debe contener unos elementos básicos. Una parte de ellos es menos importante, como por ejemplo: la portada / la página principal o las páginas de autor. Sin embargo, lo demás debe facilitarnos la información más importante desde el punto de vista del usuario.

1. Introducción y pistas sobre cómo orientarse con el manual Invacare IRC5PO2 - Al principio de cada manual se deben encontrar las indicaciones acerca de la manera de usar un manual de instrucciones. Debe contener información sobre dónde encontrar el índice Invacare IRC5PO2, preguntas frecuentes o problemas más comunes – son los apartados más buscados por los usuarios de cada manual de instrucciones.
2. Índice - listados de todos los consejos acerca de Invacare IRC5PO2 que encontraremos en el documento
3. Consejos de uso de las funciones básicas del dispositivo Invacare IRC5PO2 - que deben facilitarnos los primeros pasos durante el uso de Invacare IRC5PO2
4. Troubleshooting - una secuencia sistematizada de acciones que nos ayudará a diagnosticar los problemas más importantes con Invacare IRC5PO2
5. FAQ - las preguntas frecuentes
6. Datos de contacto Información acerca de cómo encontrar los datos de contacto del fabricante / servicio de Invacare IRC5PO2 en cada país si no somos capaces de solucionar el problema por nuestra cuenta.

¿Tienes una pregunta acerca de Invacare IRC5PO2?

Utiliza el formulario que se encuentra abajo

Si con la ayuda del manual de instrucciones encontrado no has podido solucionar tu problema con Invacare IRC5PO2, haz una pregunta utilizando el formulario que se encuentra abajo. Si alguno de los usuarios ha tenido un problema parecido con Invacare IRC5PO2 probablemente compartirá una forma de solucionarlo.

Copia el texto de la imagen

Comentarios (0)