    OPERATING INSTRUCTIONS AND S PECIFICATIONS NI 9264 16-Channel, ± 10 V, 16-Bit Analog Voltage Outp ut Module ni.com/manuals Deutsch Français ...

  • National Instruments NI 9264 - page 2

    NI 9264 Operating I nstruction s and Specificati ons 2 ni.com This doc u ment describes ho w to u se the National Instr u ments 9 264 and incl u des specif ications and term inal assignments for the NI 9264. V isit ni.com/info and enter rd s oftwarever s ion to determine which software yo u need for the mod u les yo u are u sing. For informa tion a ...

  • National Instruments NI 9264 - page 3

    © National Inst ruments Corp. 3 NI 9264 Operating Instructi ons and Specific ations Safety Guidelines Operate the NI 9264 only as described in these operating instr u ctions. Hot Surface This icon denotes that the component may be hot. T o u ching this component may res u lt in bodily inj u ry . Safety Guidelines for Hazardous V oltages If hazardo ...

  • National Instruments NI 9264 - page 4

    NI 9264 Operating I nstruction s and Specificati ons 4 ni.com Caution When mod u le terminals are hazardo u s voltage LIVE (>42.4 V pk /60 VDC), yo u m u st ens u re that devices and circ u its connected to the mod u le are properly ins u lated from h u man contact. Y o u m u st u se the NI 9940 connector backshell kit to ens u re t hat the term ...

  • National Instruments NI 9264 - page 5

    © National Inst ruments Corp. 5 NI 9264 Operating Instructi ons and Specific ations Ex nA IIC T4 hazardo u s locations; and nonhazardo u s locations only . Follo w these g u idelines if yo u are installing the NI 9264 in a potentially explosi ve environm e nt. Not follo wing these g u idelines may res u lt in serio u s inj u ry or death. Caution D ...

  • National Instruments NI 9264 - page 6

    NI 9264 Operating I nstruction s and Specificati ons 6 ni.com Special Conditions for Hazard ous Locations Use in Europe This eq u ipment has been e val u ated as Ex nA IIC T4 eq u ipment u nder DEMK O Certificate No. 07 A TEX 0626664X. Each mod u le is marked II 3G and is s u itable for u se in Zone 2 hazardo u s locations. If yo u are u sing the N ...

  • National Instruments NI 9264 - page 7

    © National Inst ruments Corp. 7 NI 9264 Operating Instructi ons and Specific ations Connecting the NI 9264 The NI 9264 has a 36-terminal detachable spring-te rminal connector that provides connections for 16 analog o u tp u t channels. Figure 2. NI 9264 T erminal Assignments AO0 AO1 AO2 AO3 AO4 AO5 AO6 AO7 AO8 AO9 AO10 AO1 1 AO12 AO1 3 AO1 4 AO1 5 ...

  • National Instruments NI 9264 - page 8

    NI 9264 Operating I nstruction s and Specificati ons 8 ni.com Each channel of the NI 9264 has a terminal to which yo u can connect the positiv e lead of a load. Each cha nnel also has a common terminal, COM, and there are fo u r additional COM terminals at the bottom of the c onnector . All of the COM terminals are internally connected to the isola ...

  • National Instruments NI 9264 - page 9

    © National Inst ruments Corp. 9 NI 9264 Operating Instructi ons and Specific ations protection. Refer to Fig u re 4 for an ill u stration of the o u tp u t circ u itry for one channel of the NI 9264. Figure 4. Output Cir cuitr y for One Channel of the NI 9264 When the mod u le powers on, the channels o u tp u t the start u p voltage. Refer to the ...

  • National Instruments NI 9264 - page 10

    NI 9264 Operating I nstruction s and Specificati ons 10 ni.com Connecting Wires to the NI 9264 Connecto r Use a flathead scre wdriv er with a blade smaller than 2.3 × 1.0 mm (0.09 × 0.04 in.) to connect wires to the detachable spring-terminal connector . Insert the screwdri ve r into a spring clamp acti vation slot and press a wire into the corre ...

  • National Instruments NI 9264 - page 11

    © National Inst ruments Corp. 11 NI 9264 Operating Instructi ons and Specific ations Wiring for High-Vib ration Applications If an application is s u bject to high vibrat ion, National Instr u ments recommends that yo u u se the NI 9940 backshell kit to protect the connections. Re fer to Fig u re 1 for an ill u stration of the NI 9940 connector ba ...

  • National Instruments NI 9264 - page 12

    NI 9264 Operating I nstruction s and Specificati ons 12 ni.com Specifications The follo wing specifications a re typical for the range – 40 to 70 °C u nless otherwise noted. All v oltages are relat iv e to COM u nless otherwise noted. Output Characteristics N u mber of channels ................ .......... 16 an alog o u tp u t channels DAC r e s ...

  • National Instruments NI 9264 - page 13

    © National Inst ruments Corp. 13 NI 9264 Operating Instructi ons and Specific ations Power -do wn voltage .................... ..... 0 V 1 O u tp u t impedanc e ... ............ ........... ... 2.0 Ω Acc u racy Stability Gain drift .. ............. ........... .......... 6 ppm /°C Offset drift ........... ........... ........... . 80 μ V/°C 1 ...

  • National Instruments NI 9264 - page 14

    NI 9264 Operating I nstruction s and Specificati ons 14 ni.com Protection Overv oltage ............. ........... ........ ±27 V a t 25 °C Short-circ u it .... ........... ......... ........ Indefinitely Update time Noise ................................ ............ ..... 500 μ V rms Crosstalk .... ......... ........... ......... .......... 90 d ...

  • National Instruments NI 9264 - page 15

    © National Inst ruments Corp. 15 NI 9264 Operating Instructi ons and Specific ations Capacitiv e dri ve ............. ........... ........ 1,500 pF max Monotonicity......... .................. .......... 16 bits DNL .... ............ ........ ............ ......... ..... +1 LSB max INL (endpoint) ................. ............ ..... ±12 LSBs max ...

  • National Instruments NI 9264 - page 16

    NI 9264 Operating I nstruction s and Specificati ons 16 ni.com Physical Characteristics If yo u need to clean the mod u l e, wipe it with a dry to wel. Spring-terminal wiring..... ............ ..... 18 to 28 A WG copper cond u ctor wire with 7 mm (0.28 in.) of ins u lation stripped from the end W eight .......... ........... ............ .......... ...

  • National Instruments NI 9264 - page 17

    © National Inst ruments Corp. 17 NI 9264 Operating Instructi ons and Specific ations Meas u rement Cate gory II is for meas u rements performed on circ u its directly connect ed to the electrical dist rib u t ion system. This cate gory refers to lo cal-le vel electrical distr ib u tion, s u ch as that provided b y a standard w all o u tlet, for e ...

  • National Instruments NI 9264 - page 18

    NI 9264 Operating I nstruction s and Specificati ons 18 ni.com Hazardous Locations U.S. (UL) ... ........... ........... ............ ..... Class I, Division 2, Gro u ps A, B, C, D, T4; Class I, Zone 2, AEx nA IIC T4 Canada (C-UL) ....... ............ ........... ... Class I, Division 2, Gro u ps A, B, C, D, T4; Class I, Zone 2, Ex nA IIC T4 E u ro ...

  • National Instruments NI 9264 - page 19

    © National Inst ruments Corp. 19 NI 9264 Operating Instructi ons and Specific ations Operating h u midity (IEC 60068-2-56)....... ........... ............ 10 to 90% RH, noncondensing Storage h u midity (IEC 60068-2-56)....... ........... ............ 5 to 95% RH, noncondensing Maxim u m altit u de.......... ......... .......... 2,000 m Poll u tion ...

  • National Instruments NI 9264 - page 20

    NI 9264 Operating I nstruction s and Specificati ons 20 ni.com Electromagnetic Compatibility This prod u ct is designed to meet the req u ire ments of the follo wing standards of EMC for electrical eq u ipment for meas u remen t, control, and laboratory u se: • EN 61326 EMC req u irements; Ind u strial Imm u nity • EN 55011 Emissions; Gro u p 1 ...

  • National Instruments NI 9264 - page 21

    © National Inst ruments Corp. 21 NI 9264 Operating Instructi ons and Specific ations Note Refer to the Declaration of Co nformity (DoC) for this prod u ct for any additional reg u latory compliance information. T o obtain the DoC for this prod u ct, visit ni.com/certification , search by mod u le n u mber or prod u ct line, and click the appropria ...

  • National Instruments NI 9264 - page 22

    NI 9264 Operating I nstruction s and Specificati ons 22 ni.com Waste Electrical and Electronic Equipment (WEEE) EU Customers At t he end of their lif e cycl e, all prod u cts must be sent to a WEEE re c ycling center . For more information abo u t WEEE recycling ce nters and National Instr u ments WEEE initiati ves, visit ni.com/ environment/weee.h ...

  • National Instruments NI 9264 - page 23

    © National Inst ruments Corp. 23 NI 9264 Operating Instructi ons and Specific ations Where to Go for Support The National Instr u ments W eb site is yo u r comple te reso u rce for technical s u pport. At ni.com/ s upport yo u have access to ev erything from tro u bleshooting and applic ation dev elopment self-help reso u rces to email and phone a ...

  • National Instruments NI 9264 - page 24

    National Instruments, NI, n i.com, and LabVIEW are trademarks of National Instrument s Corporation. Refer to the Terms of Use section on ni.com/legal for more informati on about National Instruments t rademarks. Other product and company names ment ioned herein are tradema rks or trade names of their respec tive companies. For patents covering Nati ...

Fabricante National Instruments Categoría Network Card

Los documentos del dispositivo National Instruments NI 9264 que obtenemos del fabricante se pueden dividir en varios grupos. Entre ellos están::
- dibujos técnicos National Instruments
- manuales de instrucciones NI 9264
- hojas de producto National Instruments
- folletos informativos
- o etiquetas energéticas National Instruments NI 9264
Todos son importantes, pero lo más importante desde el punto de vista del usuario de un dispositivo lo encontraremos en el manual de instrucciones National Instruments NI 9264.

Un conjunto de documentos determinado como manual de instrucciones se divide también en tipos más detallados, tales como: instrucciones de montaje National Instruments NI 9264, instrucciones de servicio, instrucciones cortas o instrucciones de usuario National Instruments NI 9264. Dependiendo de la situación debes buscar el documento que necesitas. En nuestra página puedes consultar el manual de instrucciones más popular del producto National Instruments NI 9264.

Manuales de instrucciones parecidos

El manual de instrucciones del dispositivo National Instruments NI 9264, ¿cómo debería ser?
El manual de instrucciones, también determinado como el manual de usuario o simplemente instrucciones, es un documento técnico que tiene como objetivo ayudar a los usuarios a utilizar National Instruments NI 9264. Las instrucciones normalmente las realiza un escritor técnico pero en un lenguaje comprensible para todos los usuarios de National Instruments NI 9264.

El manual de instrucciones completo de National Instruments debe contener unos elementos básicos. Una parte de ellos es menos importante, como por ejemplo: la portada / la página principal o las páginas de autor. Sin embargo, lo demás debe facilitarnos la información más importante desde el punto de vista del usuario.

1. Introducción y pistas sobre cómo orientarse con el manual National Instruments NI 9264 - Al principio de cada manual se deben encontrar las indicaciones acerca de la manera de usar un manual de instrucciones. Debe contener información sobre dónde encontrar el índice National Instruments NI 9264, preguntas frecuentes o problemas más comunes – son los apartados más buscados por los usuarios de cada manual de instrucciones.
2. Índice - listados de todos los consejos acerca de National Instruments NI 9264 que encontraremos en el documento
3. Consejos de uso de las funciones básicas del dispositivo National Instruments NI 9264 - que deben facilitarnos los primeros pasos durante el uso de National Instruments NI 9264
4. Troubleshooting - una secuencia sistematizada de acciones que nos ayudará a diagnosticar los problemas más importantes con National Instruments NI 9264
5. FAQ - las preguntas frecuentes
6. Datos de contacto Información acerca de cómo encontrar los datos de contacto del fabricante / servicio de National Instruments NI 9264 en cada país si no somos capaces de solucionar el problema por nuestra cuenta.

¿Tienes una pregunta acerca de National Instruments NI 9264?

Utiliza el formulario que se encuentra abajo

Si con la ayuda del manual de instrucciones encontrado no has podido solucionar tu problema con National Instruments NI 9264, haz una pregunta utilizando el formulario que se encuentra abajo. Si alguno de los usuarios ha tenido un problema parecido con National Instruments NI 9264 probablemente compartirá una forma de solucionarlo.

Copia el texto de la imagen

Comentarios (0)