    E4H SERIES W A TER PURIFICA TION MACHINES OPERA TION AND MAINTENANCE MANUAL ...

  • GE E4H Series - page 2

    ...

  • GE E4H Series - page 3

    OPERA TION AND MAINTENANCE MANUAL FOR GE OSMONICS E4H SERIES W A TER PURIFICA TION MACHINES T ABLE OF CONTENTS Page 1.0 DESCRIPTION ......................................................................................................... 1 1.1 General Information and Principles of Operation ......................................... 1 1.2 Machine No ...

  • GE E4H Series - page 4

    Page 3.0 PREP ARA TION AND ST AR T -UP ............................................................................ 12 3.1 Pretreatment for W ater Purification ............................................................... 12 3.2 Start-Up .......................................................................................................... 12 ...

  • GE E4H Series - page 5

    Page LIST OF FIGURES 1 Normal V ersus Cross Flow Filtration .................................... 1 2 Membrane Element with Interconnectors.............................. 2 3 Cross Sectional V iew of Membrane Element ........................ 2 4 Principles of Operation .......................................................... 3 LIST OF T ABLES 1.1 F ...

  • GE E4H Series - page 6

    ...

  • GE E4H Series - page 7

    1.0 DESCRIPTION 1.1 General Information and Principles of Operation Y our E-Series reverse osmosis (RO) machine is a durable piece of equipment which, with proper care, will last for many years. These instructions give operating and maintenance details vital to the sustained performance of the machine. Reverse osmosis is the separation of one compo ...

  • GE E4H Series - page 8

    Figure 1.2 - Membrane Element with Interconnectors Figure 1.3 Cross Sectional V iew of Membrane Element 2 ...

  • GE E4H Series - page 9

    Some operating definitions are provided to help you further understand your machine: Permeate Rate [Product W ater Rate (Q p )] is the flow rate of purified water which has passed through the membrane element and out of the membrane element housing; expressed in gal/min (gpm) or gal/hr (gph) [in metric, liter/min (Lpm) or cubic meters/hour (m 3 /h) ...

  • GE E4H Series - page 10

    Flow Description - The feed water passes through a replaceable 5-micron cartridge pre-filter which removes bulk suspended solids. Filtered water then flows to the inlet con- trol valve. This solenoid-controlled diaphragm valve is wired to the ON/OFF switch and opens when the machine is turned ON, allowing water to flow to the pump inlet. When the m ...

  • GE E4H Series - page 11

    Example: E4H-21K/ECN, 230, 6, 50-75 • E4H indicates the machine series • H indicates horizontal membrane element housing configuration • 21K indicates the rated permeate flow in thousands of gallons per day @ 77ºF (25ºC),(i.e., 21K = 21,000 gallons per day) • ECN indicates Economy Model and DLX indicates the Deluxe Model • 230 indicates ...

  • GE E4H Series - page 12

    • Digital concentrate and permeate flow meters • Digital conductivity monitor , panel-mounted, for permeate quality monitor- ing • Autoflush System - programmable, automated high-velocity membrane ele- ment flushing for the longest membrane element life; set at the factory and adjustable in the field • Gauges, valves and rigid piping of sta ...

  • GE E4H Series - page 13

    • Alarms included: low inlet pressure, high amp draw , high / low pH 1.5 Specifications for E-Series Machines 1.5.1 Feed W ater Requirements T able 1.1 Feed W ater Requir ements 1.5.2 Permeate (Product W ater) Flow Rate Stated on the serial number label (assumes no permeate back pressure, 2000 mg/L TDS maximum feed concentration, and rated temper ...

  • GE E4H Series - page 14

    8 1.5.3 Concentrate Flow Rate Factory set as stated on serial number label Concentrate Outlet: 1-inch FNPT 1.5.4 T ypical Pure W ater Recovery: 50 - 75% 1.5.5 Final Operating Pressure Minimum: 200 psig (13.8 barg) Maximum: 235 psig (16.2 barg) 1.5.6 Pump Multi-stage centrifugal, approximate primary operating pressure of 190 psig (13.1 barg), exclud ...

  • GE E4H Series - page 15

    9 SAL TS CA TIONS Percent Maximum Percent Passage Concentration Name Symbol Rejection (A vg) Percent Sodium N a + 94-96 5 5-10 Calcium C a +2 96-98 3 * Magnesium M g +2 96-98 3 * P o t assium K +1 94-96 5 5-10 Iron F e +2 98-99 2 * Manganese M n +2 98-99 2 * Aluminum A l +3 9 9 + 1 10-20 Ammonium N H 4 +1 88-95 8 3 - 8 Copper C u +2 98-99 1 10-20 N ...

  • GE E4H Series - page 16

    10 2.0 INST ALLA TION 2.1 Mounting E4H machines are equipped with a stand alone frame, 61-inch (155-cm) H x 132-inch (335 cm) W x 34-inch (86 cm) D, which supports the machine. At least 45 inches (1 14 cm) of space should be allowed on each end of the membrane element housings for removal and loading of membrane elements. If 45-inches (1 14 cm) are ...

  • GE E4H Series - page 17

    2.2.4 Permeate Outlet Connection Install the CIP valve on the permeate outlet tee. The pure water (permeate) should be transported to the point of use via non-corroding-type tubing, pipe, or hose. Examples are: food-grade flexible nylon tubing, stainless steel tubing, or PVC hose. The permeate outlet is 1-inch FNPT . 2.3 Electrical The DLX and ECN ...

  • GE E4H Series - page 18

    3.0 PREP ARA TION AND ST AR T -UP 3.1 Pretreatment for W ater Purification All systems will operate most efficiently on filtered water with a pH of less than 6.5 and a Silt Density Index (SDI) of 5 or below . If the machine is operated on higher pH water , other forms of pretreatment may be necessary . A water analysis prior to start-up of the mach ...

  • GE E4H Series - page 19

    STEPS 1. Re-check the function and integrity of your pretreatment equipment. Ensure that your water softener , activated carbon filters and iron filters (where applicable) have been leak-checked, backwashed, and thoroughly rinsed for service before starting up your RO unit. 2. Attach the feed water pipe to the inlet of the machine. 3. Check for lea ...

  • GE E4H Series - page 20

    inlet stream to the RO pump. It is important to balance the operating pres- sure and the respective flows of these valves to ensure that your machine is operating correctly . It is also important to understand the relationship of these two valves, the pressure gauge, and your RO pump. The pump has a fixed amount of flow produced, and the valves are ...

  • GE E4H Series - page 21

    NOTE: Do not allow the pressure to exceed 250 psi (17.2 bar). If the pressure exceeds 250 psi (17.2 bar), open the concentrate flow control valve until the pressure gauge shows 250 psi (17.2 bar) or less. As the machine purges the air and fills with water , the pressure will grad- ually increase. Y ou should see water flowing through the permeate a ...

  • GE E4H Series - page 22

    Once the desired flow rate is achieved [250 psi (17.2 bar) operating pres- sure] no further valve adjustment is needed. The table below shows flow rates at 50%, 66% and 75% recovery for the E4H models. Use this table in adjusting flow rates. NOTE : Permeate flow rates are dependent upon temperature and con- ditions at your site. Contact your dealer ...

  • GE E4H Series - page 23

    23. A Daily Log Sheet (Section 10.0) which includes general operating con- ditions (pressures, flows, concentrations, pH, and pretreatment condi- tions), and routine or special maintenance (flushing or cleaning as need- ed) must be kept. This Daily Log Sheet will be required by GE Osmonics if a warranty question arises. 17 ...

  • GE E4H Series - page 24

    18 4.0 OPERA TION AND MAINTENANCE The operation and maintenance of your GE Osmonics E4H Machine is relatively simple but requires regular data recording and routine preventative maintenance. W e cannot emphasize too strongly the importance of filling out the daily log sheet during each operating shift. A data sheet was filled out upon start-up cont ...

  • GE E4H Series - page 25

    4.3 Flushing The machine should be flushed at least daily to remove sediment from membrane element surfaces. T o flush the unit: STEPS 1. Open the concentrate valve until the pressure gauge indicates the minimum pres- sure designated on the nameplate. This increases the flushing action on the mem- brane element. NOTE: If pressure will not decrease ...

  • GE E4H Series - page 26

    T able 4.4 Dry Chemical Cleaners CLEANING PROCEDURE STEPS 1. W ith the RO machine running, open the Clean-In-Place (CIP) permeate valve. After this valve has been opened, close the permeate service valve. Permeate water will flow into the CIP tank. Allow the water to run through the CIP tank and the CIP tank drain valve for a few minutes to ensure ...

  • GE E4H Series - page 27

    4. T urn the CIP ON/OFF switch to the ON position and recirculate the cleaning solu- tion through the machine. The cleaning solution should be recycled for approxi- mately 15 minutes or until the solution temperature reaches 85°F (29°C). If heat rise occurs too quickly , larger volumes of cleaning solution or the use of a heat exchanger will slow ...

  • GE E4H Series - page 28

    3. Remove the tubing connections on the inlets and outlets of the membrane element housings. 4. Open the concentrate valve. 5. Remove the drain plugs from all PVC manifolds. 6. Be sure the flow meters are drained by disconnecting the bottom fitting of each flow meter . 7. Allow the machine to drain for a minimum of eight hours or until the opened p ...

  • GE E4H Series - page 29

    7. Inspect the membrane element housing and clean as necessary to remove any con- taminants, obstructions, etc. 8. Apply a small amount of O-ring lubricant to all O-rings on the end caps, and the brine seal on the membrane element. 9. Insert the downstream end of the membrane element in the upstream end of the membrane element housing (i.e., load i ...

  • GE E4H Series - page 30

    2. Remove all the membrane elements from the membrane element housings in the direction of flow , where possible. If necessary , a membrane element can be removed against the direction of flow . A heavy duty pliers or channel lock pliers may be necessary to pull the old membrane element our of the membrane element housing. 3. T o reinstall replacem ...

  • GE E4H Series - page 31

    5.0 OPTIONAL ACCESSORIES 5.1 Level Controls Float switches, pressurized storage switches or other level controls should be wired into the control circuit line prior to the switch on the unit. The following ensures that the inlet valve, instruments, and pump are not powered when storage tanks are full: float switch assembly with cord, counterweight, ...

  • GE E4H Series - page 32

    6.0 TROUBLESHOOTING 26 SYMPTOM Low operating pressure Low permeate flow rate POSSIBLE CAUSES Insufficient feed water pressure or flow Clogged pre-filter cartridge High flow rates Dirty or fouled membrane elements Solenoid valve not opening Pump rotating backwards (three-phase power only) Insufficient electrical power Pump not operating correctly Lo ...

  • GE E4H Series - page 33

    27 SYMPTOM Low permeate flow rate (continued) Low concentrate flow rate, normal or higher than normal pressure Pressure does not drop when concentrate valve opened POSSIBLE CAUSES Dirty or fouled membrane elements Operating on cold water less than 55ºF (13ºC) Membrane elements installed backwards or damaged concentrate seal Flow meter inaccurate ...

  • GE E4H Series - page 34

    28 SYMPTOM High operating pressure Excessive pressure drop [over 50 psig (3.4 barg)] (high primary pressure - low final pressure) W ater flowing when machine is turned OFF Declining rejection (high permeate conductivity) POSSIBLE CAUSES Recycle or concentrate valve plugged Inaccurate pressure gauge Restricted or reduced permeate flow rate Restricte ...

  • GE E4H Series - page 35

    29 SYMPTOM Declining rejection (high permeate conductivity) (continued) Switch ON, unit not operating POSSIBLE CAUSES O-ring seal broken or damaged Change in incoming water quality Inaccurate conductivity monitor or fouled probe Pressurized storage switch or float switch has cut power to machine Thermal overload in motor has tripped No power to mac ...

  • GE E4H Series - page 36

    30 SYMPTOM Electrical machine shutdown POSSIBLE CAUSES Alarm condition has turned machine OFF Motor starter overloaded, heater tripped T iming relay defective/burned out REMEDIES Restart the machine by pushing the alarm bypass. Check all possible alarm conditions: inlet pressure or motor starter overload. T urn the switch OFF; rest the heater(s). C ...

  • GE E4H Series - page 37

    7.0 RETURN GOODS AUTHORIZA TION (RGA) PROCEDURE If you wish to return good for repair , warranty evaluation and/or credit, please have your original sales order or invoice available when you call GE Osmonics. Call (800) 848-1750 and ask to speak with Customer Service. A GE Osmonics Customer Service representative will provide instructions and a ret ...

  • GE E4H Series - page 38

    8.0 W ARRANTY W arranty T erms Subject to the terms and conditions set forth hereinafter , Seller (GE Osmonics, Inc. or any of its authorized subsidiaries) warrants to the original purchaser (hereafter the “Buyer”) that the products manufactured by Seller are free from defects in material and in work- manship for twelve (12) months from the W a ...

  • GE E4H Series - page 39

    workmanship. Required or replaced products or components shall be returned to the Buyer by the Seller , freight prepaid by Seller , via UPS ground or best way surface freight. In genuine emergency situation, Seller will (at Seller’ s sole option) forward replacement parts to Buyer without waiting for authorized return of the questionable part(s). ...

  • GE E4H Series - page 40

    34 T emperatur e Permeate Rate Concentrate Rate T otal Flow Rate Recovery Pre-Filter Pr essure Post-Filter Pressur e Primary Pressur e Final Pressur e Feed TDS Concentrate TDS A verage TDS Permeate TDS (manual) Permeate TDS (meter µ µ S) % Passage (Perm TDS / A vg TDS) Chlorine in Concentrate Low Pressur e Switch Setting Pump Model Number Pump Se ...

  • GE E4H Series - page 41

    35 ...

  • GE E4H Series - page 42

    36 © Copyright 2003 GE Osmonics, Inc. Printed in USA, P/N 1236825 Rev . A This is a template. Make copies as needed. *Symbols: Q - Flow Rate; C - Conductivity Reference the T roubleshooting Guide in your Instruction Manual where trends or differences are noted . DA TE AND TIME PRE-FILTER PRESS (psi or bar) POST -FIL TER PRESS (psi or bar) PRIMARY ...

  • GE E4H Series - page 43

    ...

  • GE E4H Series - page 44

    Corporate Headquarters 5951 Clearwater Drive Minnetonka, MN 55343-8995 USA (952) 933-2277 Phone (952) 933-0141 Fax Euro/Africa Operations 230 rue Robert Schuman Z A des Uselles F-77350 Le MØe sur Seine FRANCE +33 1 64 10 2000 Phone +33 1 64 10 3747 Fax Asia/Pacific Operations 1044/8 SOI 44/2 Sukhumvit Road Prakanong Bangkok 10110 THAILAND +66 2 38 ...

Fabricante GE Categoría Water System

Los documentos del dispositivo GE E4H Series que obtenemos del fabricante se pueden dividir en varios grupos. Entre ellos están::
- dibujos técnicos GE
- manuales de instrucciones E4H Series
- hojas de producto GE
- folletos informativos
- o etiquetas energéticas GE E4H Series
Todos son importantes, pero lo más importante desde el punto de vista del usuario de un dispositivo lo encontraremos en el manual de instrucciones GE E4H Series.

Un conjunto de documentos determinado como manual de instrucciones se divide también en tipos más detallados, tales como: instrucciones de montaje GE E4H Series, instrucciones de servicio, instrucciones cortas o instrucciones de usuario GE E4H Series. Dependiendo de la situación debes buscar el documento que necesitas. En nuestra página puedes consultar el manual de instrucciones más popular del producto GE E4H Series.

Manuales de instrucciones parecidos

El manual de instrucciones del dispositivo GE E4H Series, ¿cómo debería ser?
El manual de instrucciones, también determinado como el manual de usuario o simplemente instrucciones, es un documento técnico que tiene como objetivo ayudar a los usuarios a utilizar GE E4H Series. Las instrucciones normalmente las realiza un escritor técnico pero en un lenguaje comprensible para todos los usuarios de GE E4H Series.

El manual de instrucciones completo de GE debe contener unos elementos básicos. Una parte de ellos es menos importante, como por ejemplo: la portada / la página principal o las páginas de autor. Sin embargo, lo demás debe facilitarnos la información más importante desde el punto de vista del usuario.

1. Introducción y pistas sobre cómo orientarse con el manual GE E4H Series - Al principio de cada manual se deben encontrar las indicaciones acerca de la manera de usar un manual de instrucciones. Debe contener información sobre dónde encontrar el índice GE E4H Series, preguntas frecuentes o problemas más comunes – son los apartados más buscados por los usuarios de cada manual de instrucciones.
2. Índice - listados de todos los consejos acerca de GE E4H Series que encontraremos en el documento
3. Consejos de uso de las funciones básicas del dispositivo GE E4H Series - que deben facilitarnos los primeros pasos durante el uso de GE E4H Series
4. Troubleshooting - una secuencia sistematizada de acciones que nos ayudará a diagnosticar los problemas más importantes con GE E4H Series
5. FAQ - las preguntas frecuentes
6. Datos de contacto Información acerca de cómo encontrar los datos de contacto del fabricante / servicio de GE E4H Series en cada país si no somos capaces de solucionar el problema por nuestra cuenta.

¿Tienes una pregunta acerca de GE E4H Series?

Utiliza el formulario que se encuentra abajo

Si con la ayuda del manual de instrucciones encontrado no has podido solucionar tu problema con GE E4H Series, haz una pregunta utilizando el formulario que se encuentra abajo. Si alguno de los usuarios ha tenido un problema parecido con GE E4H Series probablemente compartirá una forma de solucionarlo.

Copia el texto de la imagen

Comentarios (0)