    Dealer: This manual must be given to the end user. User: Before using this concentrator, re ad this manual and save for future reference. Operator's Manual For more information regarding Invacare products, parts, and services, please visit www.invacare.com Platinum ™ Oxygen Concentrators Standard, with /SensO 2 ® and HomeFill ® Model IRC5L ...

  • Invacare IRC5LXO2AWQ - page 2

    WARNING Platinum ™ 2 Part No 113486 6  WARNING DO NOT USE THIS PRODUCT OR ANY AVAILABLE OPTIONAL EQUIPMENT WITHOUT FIRST COMPLETELY READING AND UN DERSTANDING THESE INSTRUCTIONS AND ANY ADDITIONAL INSTRUCTIONAL MATERIAL SUCH AS OWNER’S MANUALS, SERVICE MANUALS OR INSTRUCTION SHEETS SUPPLIED WITH THIS PRODUCT OR OPTIONAL EQUIPMENT. IF YOU ARE ...

  • Invacare IRC5LXO2AWQ - page 3

    TABLE OF CONTENTS Part No 1134866 3 Platinum ™ TABLE OF CONTENTS SPECIAL NOTES ............................................... 5 RECYCLING INFO RMATION ............................. 6 SECTION 1—GENERAL GUIDELINES ..................... 7 Maintenance ......................... ...................................................... ........... 9 Ra ...

  • Invacare IRC5LXO2AWQ - page 4

    TABLE OF CONTENTS Platinum ™ 4 Part No 113486 6 Routine Maintenanc e ....................... ................................................. 27 SECTION 7—TROUBLES HOOTING GUIDE ............ 31 LIMITED WARRANTY .................................... 33 ...

  • Invacare IRC5LXO2AWQ - page 5

    SPECIAL NOTES Part No 1134866 5 Platinum™ SPECIAL NOTES Signal  words  are  used  in  this  manual  and  apply  to  hazards  or  unsafe  practices  which  could  result  in  personal  injury  or  property  damage.  Refer  to  the  following  table  for  definit ...

  • Invacare IRC5LXO2AWQ - page 6

    RECYCLING INFORMATION Platinum™ 6 Part No 1134866 RECYCLING INFORMATION This  product  has  been  supplied  from  an  environmentally  aw are  manufacturer  that  complies  with  the  Wa s t e  Electrical  and  Electronic  Equipment  (WEEE)  Directive  2002/96/CE. This  pro duct ? ...

  • Invacare IRC5LXO2AWQ - page 7

    SECTION 1—GENERAL GUIDELINES Part No 1134866 7 Platinum™ SECTION 1—GENERAL GUIDELINES  WARNING SECTION 1 - GENERAL GU IDELINES contains important information for the safe operation and use of this product. In  order  to  ensure  the  safe  installation,  assembly  and  operation  of  the  Platinum  co ...

  • Invacare IRC5LXO2AWQ - page 8

    SECTION 1—GENERAL GUIDELINES Platinum™ 8 Part No 1134866  WARNINGS The  use  of  oxygen  therapy  requires  that  special  care  be  take n  to  reduce  the  risk  of  fire.  Any  materials  that  will  burn  in  air ,  and  some  that  will  not,  are  ...

  • Invacare IRC5LXO2AWQ - page 9

    SECTION 1—GENERAL GUIDELINES Part No 1134866 9 Platinum™ Fill  humidifier  with  wa t e r  to  the  lev el  shown  by  the  manufacturer .  DO  NOT  overfill. Keep  unit  at  least  7.5  cm  (3  in)  aw ay  from  w alls,  draperies,  furniture,  and  the  like. ...

  • Invacare IRC5LXO2AWQ - page 10

    SECTION 1—GENERAL GUIDELINES Platinum™ 10 Part No 1134866 DO  NOT  reach  for  prod uct  that  has  fallen  into  wat e r .  Unplug  IMMEDIA TEL Y . This  device  is  to  be  used  only  in  accordance  with  the  prescription  of  a  physician  and  this  Owner ...

  • Invacare IRC5LXO2AWQ - page 11

    SECTION 2—FEATURES Part No 1134866 11 Platinum™ SECTION 2—FEATURES Oxygen Purity and Fault Indicator Lights (SensO 2 model)/Fault and Power Indicator Lights (LX model) Circuit Breaker Power Switch Elapsed Time Meter (Hour Meter) Cabinet Filter (one of two shown) Flowmet er Oxygen Out let FRONT VIEW TOP VIEW ACCESSOR IES (NOT SHOWN): HomeFill ...

  • Invacare IRC5LXO2AWQ - page 12

    SECTION 3—SHIPPING AND HANDLING Platinum™ 12 Part No 1134866 SECTION 3—SHIPPING AND HANDLING NOTE:  For  this  procedure,  refe r  to  FIGURE  3.1  on  page 13. Before  yo u  install  and  operate  the  concentrator ,  please  refer  to  Platinum  Series  Owner ʹ s  Manual. ...

  • Invacare IRC5LXO2AWQ - page 13

    SECTION 3—SHIPPING AND HANDLING Part No 1134866 13 Platinum™ FIGURE 3.1 Unpacking ...

  • Invacare IRC5LXO2AWQ - page 14

    SECTION 4—TYPICAL PRODUCT PARAMETERS Platinum™ 14 Part No 1134866 SECTION 4—TYPICAL PRODUCT PARAMETERS Regulatory Listing: EN/IEC 60601-1 Alternating Current Type BF equipment Unit running Unit not running Attention - Consider Accompanying Documents DO NOT smoke Class II, Double Insulation Electrical Requirements: 230 VAC ± 10% (253 VAC/207 ...

  • Invacare IRC5LXO2AWQ - page 15

    SECTION 4—TYPICAL PRODUCT PARAMETERS Part No 1134866 15 Platinum™ *Oxygen Output Concentration Levels: All 5LXO2AWQ/5LXAW/ 5LXO2AW/9LXO2AWQ models * (Stated concentration levels achieved after in itial warm-up period (approximately 30 minutes). IRC5LXAW, IRC5 LXO2AW and IRC5LXO2AWQ 93% minimu m at 1 to 3 L/min. 91% minimu m at 4 L/min. 87% mini ...

  • Invacare IRC5LXO2AWQ - page 16

    SECTION 4—TYPICAL PRODUCT PARAMETERS Platinum™ 16 Part No 1134866 Change in maximum recommended flow when back pressure of 7k Pa is applied: 0.7 L/min. Filters: Cabinet (2), Outl et HEPA and Compressor Inlet. Safety System: Current overload or line surge shutdown. High temperature co mpressor shutdown. High Pressure Alarm w/compressor shutdown. ...

  • Invacare IRC5LXO2AWQ - page 17

    SECTION 4—TYPICAL PRODUCT PARAMETERS Part No 1134866 17 Platinum™ Regulatory Listing : IRC5LXAW/IR C5LXO2AW/ IRC5LXO2AWQ/IRC9LXO2AWQ ETL tested as complying with: EN55011 EN61000-3-2 EN61000-3-3 IEC 60601-1, A1, A2 IEC 60601-1-2 IRC5LXO2AW, IRC5LXO2AWQ, IRC9LXO2AWQ Models Only ISO8359 MDD CE marked models: IRC5LXO2AW, IRC5LXO2AWQ, IRC9LXO2AWQ E ...

  • Invacare IRC5LXO2AWQ - page 18

    SECTION 4—TYPICAL PRODUCT PARAMETERS Platinum™ 18 Part No 1134866 SENSO 2 INDICATO RS LX INDICATORS LABEL SYMBOL STATUS INDICATOR LIGHTS (LED) SYSTEM OKAY O 2 over 85% (± 2%) GREEN Indicator Light O 2 Between 73% (± 3%) to 85% (±2%) YELLOW Indicator light A. YELLOW Solid B. YELLOW Flashing Sensor Failure Call a qualif ied technician. SYSTEM ...

  • Invacare IRC5LXO2AWQ - page 19

    SECTION 5—OPERATING INSTRUCTIONS Part No 1134866 19 Platinum™ SECTION 5—OPERATING INSTRUCTIONS Introduction Yo u r  oxygen  concentrator  is  intended  for  individual  use  in  the  home.  It  is  an  electronically  operated  device  that  separates  oxygen  from  room  air . ...

  • Invacare IRC5LXO2AWQ - page 20

    SECTION 5—OPERATING INSTRUCTIONS Platinum™ 20 Part No 1134866 Recommended Guidelines for Optimum Performance NOTE:  Ensure  that  your  concentrator  is  at  least  7.5  cm  (3  in)  aw ay  from  walls,  drap erie s  or  furniture  to  assur e  sufficient  air  flow .  Av o i ...

  • Invacare IRC5LXO2AWQ - page 21

    SECTION 5—OPERATING INSTRUCTIONS Part No 1134866 21 Platinum™ Connect Humidifier (If So Prescribed) NOTE:  For  this  procedure,  refe r  to  FIGURE  5.1,FIGURE  5.2,  FIGURE  5.3  on  page 22  and  FIGURE  5.4  on  page 23. 1. Fill  humidifier  with  distilled  wa t e r  t ...

  • Invacare IRC5LXO2AWQ - page 22

    SECTION 5—OPERATING INSTRUCTIONS Platinum™ 22 Part No 1134866 3. Remove  the  filter  access  door  located  on  the  side  of  the  concentrator .  Insert  a  flathead  screwdriver  in  the  plate  groove  on  th e  top  edg e  of  the  filter  access  door  ...

  • Invacare IRC5LXO2AWQ - page 23

    SECTION 5—OPERATING INSTRUCTIONS Part No 1134866 23 Platinum™ 6. Attach  oxygen  tubing  from  the  humidifier  bottle  to  oxygen  outlet  connector  on  the  oxygen  concentrator . 7. Attach  the  cannula/patient  supply  tubing  to  the  humidifier  bottle  outlet. 8. After ...

  • Invacare IRC5LXO2AWQ - page 24

    SECTION 5—OPERATING INSTRUCTIONS Platinum™ 24 Part No 1134866 FIGURE 5.5 Power Switch Flowrate NOTE:  For  this  procedure,  refe r  to  FIGURE  5.6  on  page 25. 1. T urn  the  flowrate  knob  to  the  setting  prescribed  by  you r  physician  or  therapist.  WARNING DO NOT ...

  • Invacare IRC5LXO2AWQ - page 25

    SECTION 5—OPERATING INSTRUCTIONS Part No 1134866 25 Platinum™ NOTE:  The  use  of  some  accessories  such  as  the  pediatric  flowstand  and  the  HomeFill  compr essor  will  deactivate  the  Low  Flow  Alarm. FIGURE 5.6 Flowrate SensO 2 Oxygen Purity Indicators (LX O 2 Sensor Model ...

  • Invacare IRC5LXO2AWQ - page 26

    SECTION 5—OPERATING INSTRUCTIONS Platinum™ 26 Part No 1134866 FIGURE 5.7 SensO 2 Oxygen Purity Indicators (LX O 2 Sensor Model) GREEN  light  (O 2 ) ‐ Normal  Operation. YELLOW  light  () ‐ Call  supplier  IMMEDIA TEL Y .  Yo u  may  continue  to  use  the  concentrator  unless  in ...

  • Invacare IRC5LXO2AWQ - page 27

    SECTION 5—OPERATING INSTRUCTIONS Part No 1134866 27 Platinum™ FIGURE 5.8 Indicators (LX Model) Main Power Loss Alarm System The  alarm  system  contains  no  battery ‐ and  is  thereby  maintenance  free.  The  alarm  system  is  powered  by  a  capacitor  that  is  continuousl ...

  • Invacare IRC5LXO2AWQ - page 28

    SECTION 6—MAINTENANCE Platinum™ 28 Part No 1134866 SECTION 6—MAINTENANCE NOTE:  The  Invacare  concentrators  are  specifically  designed  to  minimize  rou ti n e  pr eventive  maintenance.  In  places  with  high  dust  or  soot  levels ,  maintenance  may  need  to  be ...

  • Invacare IRC5LXO2AWQ - page 29

    SECTION 6—MAINTENANCE Part No 1134866 29 Platinum™ 3. Dry  the  filters  thoroughly  before  reinstallation. FIGURE 6.1 Cleaning the Cabinet Filter Cabinet  WARNING Unplug the concentrator when cleaning. To avoid electrical shock, DO NOT remove cabinet. 1. Clean  the  cabinet  with  a  mild  household  cle ...

  • Invacare IRC5LXO2AWQ - page 30

    SECTION 6—MAINTENANCE Platinum™ 30 Part No 1134866 FIGURE 6.2 Preventive Maintenance Record ON EACH INSPECTION MODEL NO. I R C SERIAL NO. PREVENTIVE MAINTENANCE RECORD INVACARE PLATINUM OXYGEN CONCENTRATOR Record Date of Service Record Elapsed Hours On Hour Meter Clean Cabinet Filter(s) Check Prescribed L/min. Flow Rate EVERY 4,380 HOURS FOR IR ...

  • Invacare IRC5LXO2AWQ - page 31

    SECTION 7—TROUBLESHOOTING GUIDE Part No 1134866 31 Platinum™ SECTION 7—TROUBLESHOOTING GUIDE SYMPTOM: PROBABLE CAUSE: SOLUTION: Alarm: Main Power Loss: Short beeps, long pause 1. Power cord NOT plugged in. 1. Insert plug i nto outlet. Concentrator not operating, power switch On. 2. No power at outlet. 2 . Inspect house circuit breakers or fus ...

  • Invacare IRC5LXO2AWQ - page 32

    SECTION 7—TROUBLESHOOTING GUIDE Platinum™ 32 Part No 1134866 YELLOW or RED light illumin ated. Alarm: Continuous On red light only. * Only applies t o SensO 2 models. 1. Low oxygen p urity.* 2. Kinked or blocked tubing, cannula or humidifier.* 3. Flowmeter set at 0.5 L/min. or less.* 4. Unit overheating due to blocked air intake. 5. Insufficien ...

  • Invacare IRC5LXO2AWQ - page 33

    LIMITED WARRANTY Part No 1134866 33 Platinum™ LIMITED WARRANTY A vailable  At: Belgium  &  Luxemburg: Inv acare  nv ,  Autobaan  22,  B ‐ 8210  Loppem Te l :  (32)  (0)50  83  10  10,  Fax:  (32)  (0)50  83  10  11 belgium@inv acare.com Danmark:  Inv acare  A/S,  Sdr .  Ri ...

  • Invacare IRC5LXO2AWQ - page 34

    LIMITED WARRANTY Platinum™ 34 Part No 1134866 Italia: Inv acare  Mecc  San  s.r .l.,  Vi a  dei  Pini  62,  I ‐ 36016  Thiene  (VI) Te l :  (39)  0445  38  00  59,  Fax:  (39)  0445  38  00  34 italia@inv acare.com  Nederland: Inv acare  BV ,  Celsiusstraat  46,  NL ? ...

  • Invacare IRC5LXO2AWQ - page 35

    LIMITED WARRANTY Part No 1134866 35 Platinum™ Switzerland: Mobitec  Rehab  AG,  Benkenstrasse  260,  CH ‐ 4108  Wit t e rs w il Te l . :  (41)  (0)61  487  70  80,  Fax.:  (41)  (0)61  487  70  81 office@mobitec ‐ rehab.ch switzerland@inv acare.com United  Kingdom: Inv acare  Ltd,  S ...

  • Invacare IRC5LXO2AWQ - page 36

    Invacare Corporation www.invacare.com USA One Invacare Way Elyria, Ohio USA 44036-2125 440-329-6000 800-333-6900 Technical Serv ices 440-329-6593 800-832-4707 Manufacturer: Invacare Medical Equipment (Suzhou) Co., Ltd. Unit 2B, Export Processing Zone No. 200 Suhong Road SIP (Suzhou Industrial Park) Suzhou, Jiangsu, PRC 215021 Tel: 86-512-62586192 F ...

Fabricant Invacare Catégorie Oxygen Equipment

Les documents que nous recevons du fabricant de l'appareilInvacare IRC5LXO2AWQ peuvent être divisés en plusieurs groupes. Ceux-ci sont, entre autres:
- dessins techniques Invacare
- manuels d’utilisations IRC5LXO2AWQ
- fiches produit Invacare
- dépliants
- ou étiquettes-énergie Invacare IRC5LXO2AWQ
Tous sont importants, mais les informations les plus importantes du point de vue de l'utilisation de l'appareil se trouvent dans le manuel d’utilisation Invacare IRC5LXO2AWQ.

Un groupe de documents appelé manuels d’utilisation est également divisé en types plus spécifiques, tels que: Manuels d’installation Invacare IRC5LXO2AWQ, manuels d’entretien, brefs manuels ou manuels de l’utilisateur Invacare IRC5LXO2AWQ. Selon vos besoins, vous devriez chercher le document dont vous avez besoin. Sur notre site, vous pouvez voir le manuel le plus populaire d’utilisation du produit Invacare IRC5LXO2AWQ.

Manuels d’utilsiation similaires

Manuel d’utilisation complet de l’appareil Invacare IRC5LXO2AWQ, quelle devrait-elle être?
Le manuel d’utilisation, également appelé le mode d’emploi, ou tout simplement le manuel, est un document technique destiné à aider à utiliser Invacare IRC5LXO2AWQ par les utilisateurs. Des manuels sont généralement écrits par un rédacteur technique, mais dans un langage accessible à tous les utilisateurs Invacare IRC5LXO2AWQ.

Le manuel d’utilisation complet Invacare, devrait inclure plusieurs éléments de base. Certains d'entre eux sont moins importants, tels que: la couverture / page de titre ou pages d'auteur. Cependant, la partie restante, devrait nous fournir des informations importantes du point de vue de l'utilisateur.

1. Introduction et des conseils sur la façon d'utiliser le manuel Invacare IRC5LXO2AWQ - Au début de chaque manuel, nous devrions trouver des indices sur la façon d'utiliser le document. Il doit contenir des informations sur l'emplacement de la table des matières Invacare IRC5LXO2AWQ, FAQ ou des problèmes les plus fréquents - les points qui sont les plus souvent recherchés par les utilisateurs de chaque manuel
2. Table des matières - index de tous les conseils pour lInvacare IRC5LXO2AWQ qui peuvent être trouvés dans le document courant
3. Conseils sur la façon d'utiliser les fonctions de base de l’appareil Invacare IRC5LXO2AWQ - qui devraient nous aider dans les premières étapes lors de l'utilisation Invacare IRC5LXO2AWQ
4. Troubleshooting - séquence systématique des activités qui nous aideront à diagnostiquer et ensuite résoudre les principaux problèmes de Invacare IRC5LXO2AWQ
5. FAQ - questions fréquemment posées
6. Détails du contact Informations sur l'endroit où chercher le contact avec le fabricant / service Invacare IRC5LXO2AWQ dans un pays donné, si le problème ne peut être résolu par nous-mêmes.

Avez-vous une question à propos de Invacare IRC5LXO2AWQ?

Utiliser le formulaire ci-dessous

Si vous n’avez pas résolu votre problème avec Invacare IRC5LXO2AWQ, avec l'aide du manuel que vous avez trouvé, posez une question en utilisant le formulaire ci-dessous. Si un utilisateur a eu un problème similaire avec Invacare IRC5LXO2AWQ il est probable qu’il a envie de partager la façon de le résoudre.

Réécrire le texte de l'image

Commentaires (0)