Manuel d’utilisation Agilent Technologies 5964-8134

26 pages 0.15 mb
Télécharger

Aller à la page of 26

Summary
  • Agilent Technologies 5964-8134 - page 1

    ADDENDUM to the A gilent 6811B, 6812B, and 6813B User’ s Guide and Programming Guide for the Dual Pow er A naly zer Option 020/022 Agilent P/N 5964-8134 Printed in U.S.A. Microf iche No 5964-8135 May , 2000 ...

  • Agilent Technologies 5964-8134 - page 2

    2 Contents Agilent Dual Power Analyzer D escription 3 Rear Panel Connecti ons 5 Installation 5 Front Panel Programming 8 Additional Measurement Commands 10 Additional Sense Commands 12 Additional Frequency Modulation Comma nds 13 Additional Calibration Commands 14 Behavior Diff erences of Existing SCPI Commands 15 Programming Example 16 Using the P ...

  • Agilent Technologies 5964-8134 - page 3

    3 A gilent Dual Pow er A naly zer Description The Agilent Technologies D ual Power Analyz er Option 020 is available on the following ac sources: Agilent 6811B AC Power Source (375VA) Agilent 6812B AC Power Source (750VA) Agilent 6813B AC Power Source (1750VA) Agilent 6813B Option 019 AC Power Source (2000VA) Option 020 consists of an additional Po ...

  • Agilent Technologies 5964-8134 - page 4

    4 Front Panel Menus - Additions The Power Analy zer input uses the same front panel m enu comm ands that are used to measure the Main Output of the ac source. These are docum ented in the ac source User’ s Guide. The following m enu comm ands have been added to let you access the Pow er Analyzer input, control the output frequency modulation, and ...

  • Agilent Technologies 5964-8134 - page 5

    5 Rear Panel Connections Power Analyzer input. Connector plug is remo vable. L IN E R A T IN G HP-IB F L T I N H ++ - _ _ C O M _ _ C O M O 1 _ O 1 _ SENSE W A R N IN G 300 V A C M A X T O _ _ _ R S 2 3 2 L 2( N ) L 1 O P T I O N S C A U T IO N W A R N I N G TRIGGER O U T 1 I N 1 . . . . . . . . . LO HI LO HI V SENSE I SENSE Rear Panel, Overall Vie ...

  • Agilent Technologies 5964-8134 - page 6

    6 Measurement Cable Connections Voltage Sensing at a Location Other than the External Current Shunt You may also need to open up the cable connector if you need to voltag e sense at a point other than at the external current shunt. You m ust use your own sense cable to m ake the voltage sense connections. Please completely remov e the voltage sense ...

  • Agilent Technologies 5964-8134 - page 7

    7 External Shunt Line and Load Connections The Agilent supplied current shunt has both input (line) and output (load) connectors. The customer m ust supply the cables to connect the ac lines to and from the external current shunt. Agilent supplies one hooded cable cover for the input of the ac source and two hooded cable cov ers for the shunt. All ...

  • Agilent Technologies 5964-8134 - page 8

    8 Check the External Current Shunt Setting After you hav e completed your cable connections, you can turn on and check out the ac source as described in the Turn-On C heckout chapter in the ac source User’ s Guide. One additional step in checking out a unit with Option 020/022 installed, is to make sure that the correct shunt v alue has been ente ...

  • Agilent Technologies 5964-8134 - page 9

    9 Using the Front Panel Display w ith the Power Analy zer Input To select the measurement source: Action Display Press Me te r to return the display to Meter mode. Press Phase Select at any tim e to toggle between the Main Output and the Power Analy zer input. T he left-most digit of the front panel display will indicate either " φ 1" fo ...

  • Agilent Technologies 5964-8134 - page 10

    10 A dditional Measurement Commands The following additional m easurement com mands are provided with the A gilent Dual Power An alyzer Option 020/022. These m easurement comm ands are optimiz ed for use in specific applications such as evaluating uninterruptible power supplies. In this application the new com mands can be used to m easure transfer ...

  • Agilent Technologies 5964-8134 - page 11

    11 MEASure:TVOLt:A BSolute? FETCh:TVOLt: A BSolute? Returns the tim e at which the absolute value of the measured v oltage first exceeds level , with the following qualifications: The behavior of the sig nal before start_ time is ignored, and the absolute value of the voltag e must rem ain above level for at least the specified min_ pulse_ width . ...

  • Agilent Technologies 5964-8134 - page 12

    12 A dditional Sense Commands The following additional sense com mands are provided w ith the Agilent Dual Power A nalyzer Option 020/022. These comm ands let you select a measurem ent source and the frequency m easurement channel for the Power Analyz er input. SENSe:NSELect Selects the m easurement source that will return data when a query is sent ...

  • Agilent Technologies 5964-8134 - page 13

    13 A dditional Frequency Modulation Commands The following SCPI commands are used to prog ram the output frequency of the ac source. These comm ands apply only to the Main O utput of the ac source and do not affect the measurem ent capability of the unit. These comm ands are only provided with the A gilent Dual Power An alyzer Option 020/022. FM Tu ...

  • Agilent Technologies 5964-8134 - page 14

    14 A dditional Calibration Commands The following SCPI commands are used to calibrate the Power Analy zer input and external current shunt: CA Librate:VOLTage: EXTernal In itiates calibration of the external voltage m easurement. Command Syntax CALibrate:VOLTage:EXTernal Parameters None Examples CAL:VOLT:EXT CA Librate:CURRent: EXTernal Initiates c ...

  • Agilent Technologies 5964-8134 - page 15

    15 Behav ior Differences of Existing SCPI Commands This section documents the behav ior differences of existing SCPI com mands wh en the Power Analyzer input is selected or is active. MEA Sure | FETCh All standard ac source MEASure or FETCh functions are av ailable for the Main Output of the ac source as well as the Power Analyzer input. The SENSe: ...

  • Agilent Technologies 5964-8134 - page 16

    16 Programming Example UPS Transfer Time Measurement The following prog ramm ing example illustrates how to use the ac source w ith the Dual Power Analy zer Option 020/022 to m easure transfer time on a UPS. Transfer time is defined as the tim e it takes a UPS to go from online-operation to battery -backup operation when the ac line fails. This exa ...

  • Agilent Technologies 5964-8134 - page 17

    17 Using the Pow er A naly z er Input w ith the GUI The Agilent AC Source Graphical User Interface (G UI) is an easy to use soft front panel for the A gilent 6800-series AC Power Source/Analy zers. It is supplied with y our ac source. With it, you can also control the Power Analyz er input on the back of the unit. Instructions to install and run th ...

  • Agilent Technologies 5964-8134 - page 18

    18 If y ou have both the Main Output and the Power Analyzer input display ed, the Main window should display both wav eforms, as illustrated in the following screen shot. Making a Voltage Phase Difference M easurement The voltag e phase difference measures the phase difference between the output of the ac source and the output of a UPS after ac pow ...

  • Agilent Technologies 5964-8134 - page 19

    19 In the dialog box that appears on the screen, enter values for the indicated param eters. T he test uses these parameters to m ake the m easurement. For a detailed explanation of the Voltage Phase Difference parameters refer to the online help by pressing the F1 key . When you are ready to make the m easurement, press Measure . Performing a Tran ...

  • Agilent Technologies 5964-8134 - page 20

    20 Pow er A naly zer Input Specifications Table A-1 lists the specifications of the Agilent Dual Power Analy zer Option 020/022. Performance specifications are warranted over the am bient temperature rang e of 0 to 40 o C. Unless otherwise noted, specifications are for a sinewave w ith a resistive load at a frequency range of 45 H z to 1 kHz , with ...

  • Agilent Technologies 5964-8134 - page 21

    21 Table A -2. Suppl ement al Charact eristics Isolatio n t o Ground: 300 Vrms /425 Vdc Measurement Ranges AC Voltage: DC Voltage: Current: 0 − 300 Vrms ± 425 V ± 80 A p eak (high r ange) ± 8 A peak (l ow rang e) 13 A r ms m ax. continuous Average Measurement Resolution rm s Voltage: rm s Current: 10 mV 2 mA THD (for a fundamental amplitude ? ...

  • Agilent Technologies 5964-8134 - page 22

    22 Verification and Calibration for Option 020/022 This section includes verification and calibration procedures for the Agilent Dual Power Analy zer Option 020/022. Add these procedures to those described in Appendix B of the AC Source User’ s Guide. The verification procedures do not check all the operating parameters, but verify that the ac so ...

  • Agilent Technologies 5964-8134 - page 23

    23 Verification Procedure The following v erification procedures assume you understand how to operate the ac source from the front panel as explained in chapter 4. Perform the following tests for operation verification in the order indicated. 1. Turn-On Check out 2. Voltage Measurem ent Accuracy 3. Current Measurem ent Accuracy NOTE: The ac source ...

  • Agilent Technologies 5964-8134 - page 24

    24 Power Analy zer Voltage Measurement Accuracy This test verifies the accuracy of the Pow er Analyzer v oltage measurem ents. The Phase 2 annunciator on the left side of the display, m ust be on for the front panel to display Power Analy zer input measurem ents. Figure B- 1A shows the setup. Measure the ac output voltag e directly at the output te ...

  • Agilent Technologies 5964-8134 - page 25

    25 Calibration Procedure Table B-1 lists the equipm ent required for calibration. Figure B- 1 shows the test setup. T he following procedures assume y ou understand how to operate front panel keys (see chapter 4). WARNING LETHAL VOLTAGES . Ac sources can supply 425 V peak at their output. DEATH on contact may result if the output terminals or circu ...

  • Agilent Technologies 5964-8134 - page 26

    26 Calibrating and Entering Power Analy zer Voltage Calibration Values Action Display 4. Con nect an Ag ilent 3458A DVM (i n syn chronous ac v olts mode) t o the ac source output terminals. Short the Power Analy zer voltage terminals (V SENSE) to ensure n o dc offset error. 5. Press Shift Calibration , scroll to CAL VOLT EXT , and pres s Enter . CA ...

Fabricant Agilent Technologies Catégorie Turntable

Les documents que nous recevons du fabricant de l'appareilAgilent Technologies 5964-8134 peuvent être divisés en plusieurs groupes. Ceux-ci sont, entre autres:
- dessins techniques Agilent Technologies
- manuels d’utilisations 5964-8134
- fiches produit Agilent Technologies
- dépliants
- ou étiquettes-énergie Agilent Technologies 5964-8134
Tous sont importants, mais les informations les plus importantes du point de vue de l'utilisation de l'appareil se trouvent dans le manuel d’utilisation Agilent Technologies 5964-8134.

Un groupe de documents appelé manuels d’utilisation est également divisé en types plus spécifiques, tels que: Manuels d’installation Agilent Technologies 5964-8134, manuels d’entretien, brefs manuels ou manuels de l’utilisateur Agilent Technologies 5964-8134. Selon vos besoins, vous devriez chercher le document dont vous avez besoin. Sur notre site, vous pouvez voir le manuel le plus populaire d’utilisation du produit Agilent Technologies 5964-8134.

Manuels d’utilsiation similaires

Manuel d’utilisation complet de l’appareil Agilent Technologies 5964-8134, quelle devrait-elle être?
Le manuel d’utilisation, également appelé le mode d’emploi, ou tout simplement le manuel, est un document technique destiné à aider à utiliser Agilent Technologies 5964-8134 par les utilisateurs. Des manuels sont généralement écrits par un rédacteur technique, mais dans un langage accessible à tous les utilisateurs Agilent Technologies 5964-8134.

Le manuel d’utilisation complet Agilent Technologies, devrait inclure plusieurs éléments de base. Certains d'entre eux sont moins importants, tels que: la couverture / page de titre ou pages d'auteur. Cependant, la partie restante, devrait nous fournir des informations importantes du point de vue de l'utilisateur.

1. Introduction et des conseils sur la façon d'utiliser le manuel Agilent Technologies 5964-8134 - Au début de chaque manuel, nous devrions trouver des indices sur la façon d'utiliser le document. Il doit contenir des informations sur l'emplacement de la table des matières Agilent Technologies 5964-8134, FAQ ou des problèmes les plus fréquents - les points qui sont les plus souvent recherchés par les utilisateurs de chaque manuel
2. Table des matières - index de tous les conseils pour lAgilent Technologies 5964-8134 qui peuvent être trouvés dans le document courant
3. Conseils sur la façon d'utiliser les fonctions de base de l’appareil Agilent Technologies 5964-8134 - qui devraient nous aider dans les premières étapes lors de l'utilisation Agilent Technologies 5964-8134
4. Troubleshooting - séquence systématique des activités qui nous aideront à diagnostiquer et ensuite résoudre les principaux problèmes de Agilent Technologies 5964-8134
5. FAQ - questions fréquemment posées
6. Détails du contact Informations sur l'endroit où chercher le contact avec le fabricant / service Agilent Technologies 5964-8134 dans un pays donné, si le problème ne peut être résolu par nous-mêmes.

Avez-vous une question à propos de Agilent Technologies 5964-8134?

Utiliser le formulaire ci-dessous

Si vous n’avez pas résolu votre problème avec Agilent Technologies 5964-8134, avec l'aide du manuel que vous avez trouvé, posez une question en utilisant le formulaire ci-dessous. Si un utilisateur a eu un problème similaire avec Agilent Technologies 5964-8134 il est probable qu’il a envie de partager la façon de le résoudre.

Réécrire le texte de l'image

Commentaires (0)