    Instruction Sheet TUpH 396P and 396PVP Sensors LIQ_MAN_ABR_396P_396PVP July 2013 Submer sion/Inser tion pH/ORP Sensor s Specifications Sensor T ype: TUpH 396P , 396 PVP; GPLR hemi or at gl ass Range: *pH: 0–14 ORP: -1500 t o 1500 mv Percen t Linearity* pH Range GPLR hemi bulb GPLR flat bulb 0 – 2 pH 94% – 2 – 12 pH 97% 98% 12 – 13 pH 9 ...

  • Emerson Process Management 369P - page 2

    2 TUpH 396P and 396PVP Sensor s Instruc tion Sheet July 2013 LI Q_MAN_ABR_396P_396PVP S tora ge 1. It is rec ommen ded tha t electrodes be stored in their original shipp ing con tainers unti l needed. 2. Do not store a t te mpera tures be low -5 °C (23 °F). 3. Electrodes should be st ored with a prot ec tive cap c ontaining KCI solution. 4. For o ...

  • Emerson Process Management 369P - page 3

    3 Instruction Sheet TUpH 396P and 396PVP Sensors LIQ_MAN_ABR_396P_396PVP July 2013 ORP Calibration 1. After makin g an e lectr ical connection between the sensor and the instrume nt, ob tain a standar d solution of saturat ed quinh ydrone. This can also be m ade quite simply by adding a few cr ystals of quinhydrone t o e ither pH 4 or pH 7 bu ffe r ...

  • Emerson Process Management 369P - page 4

    4 TUpH 396P and 396PVP Sensor s Instruc tion Sheet July 2013 LI Q_MAN_ABR_396P_396PVP Figure 2. Flow Through and Insertion Installation Figure 3. Wiring for 81 Figure 4. Wiring for 1181 ...

  • Emerson Process Management 369P - page 5

    5 Instruction Sheet TUpH 396P and 396PVP Sensors LIQ_MAN_ABR_396P_396PVP July 2013 Figure 5. Wiring 81 through Remote Junction Box Figur e 6. Wiring 1181, 1050/1060, and 1003/1023 through Remote Junction Box ...

  • Emerson Process Management 369P - page 6

    6 TUpH 396P and 396PVP Sensor s Instruc tion Sheet July 2013 LI Q_MAN_ABR_396P_396PVP Figure 7. Wiring Details of 396P-01-50 for use with and without Junction Box (PN 22719-02) for 1181, 1050, and 1060. Figure 8. Wiring for 2081 Figure 9. Wiring for 3081, 4081 ...

  • Emerson Process Management 369P - page 7

    7 Instruction Sheet TUpH 396P and 396PVP Sensors LIQ_MAN_ABR_396P_396PVP July 2013 Figure 10. Wiring 2081 through Remote Junction Box Figure 11. Wiring 3081, 4081 through Remote Junction Box Figure 12. Wiring for 1054 A/B and 2054 ...

  • Emerson Process Management 369P - page 8

    8 TUpH 396P and 396PVP Sensor s Instruc tion Sheet July 2013 LI Q_MAN_ABR_396P_396PVP Figure 13. Wiring 1054 A/B and 2054 Through a Remote Junction Box ...

  • Emerson Process Management 369P - page 9

    9 Instruction Sheet TUpH 396P and 396PVP Sensors LIQ_MAN_ABR_396P_396PVP July 2013 Figure 14. Wiring 54 through Remote Junction Box Figure 15. Wiring for 54 ...

  • Emerson Process Management 369P - page 10

    10 TUpH 396P and 396PVP Sensor s Instruc tion Sheet July 2013 LI Q_MAN_ABR_396P_396PVP Figure 16. Wiring Details of 396P-01-54 for use with and without Junction Box (PN 22719-02) for 1054, 1054A, 1054B, 2054, 2081, and Extension Cabl e (9200254). ...

  • Emerson Process Management 369P - page 11

    11 Instruction Sheet TUpH 396P and 396PVP Sensors LIQ_MAN_ABR_396P_396PVP July 2013 Figure 17. Wiring of 396P-01-55 for use with 54, 81, 3081, and 4081. Figure 18. Junc tion Bo x (PN 23550-00) for 54, 3081, 4081, and 81. ...

  • Emerson Process Management 369P - page 12

    12 TUpH 396P and 396PVP Sensor s Instruc tion Sheet July 2013 LI Q_MAN_ABR_396P_396PVP Figure 19. Wiring Details of 396P-01 Gray Cable for use with 54e pH/ORP-01-09 Figure 20. Wiring Details of 396P-01 Blue Cable for use with 54e pH/ORP-01-09 If present, reove shrink slee ve from gra y reference wire before connecting wire to terminal. F or XMT and ...

  • Emerson Process Management 369P - page 13

    13 Instruction Sheet TUpH 396P and 396PVP Sensors LIQ_MAN_ABR_396P_396PVP July 2013 Figure 21. Wiring Details of 396P-02 Gray Cable for use with 54e pH/ORP-01-09 Figure 22. Wiring Details of 396P-02 Blue Cable for use with 54e pH/ORP-01-09 If present, reove shrink slee ve from gra y reference wire before connecting wire to terminal. For XMT and 105 ...

  • Emerson Process Management 369P - page 14

    14 TUpH 396P and 396PVP Sensor s Instruc tion Sheet July 2013 LI Q_MAN_ABR_396P_396PVP Figure 23. Wiring Details of 396PVP Gray Cable for use with 54e pH/ORP-01-09 Figure 24. Wiring Details of 396PVP Blue Ca ble for use with 54e pH/ORP -01-09 For XMT and 1056 termina te Inner Drain with a wire nut. VP8 Cable par t number 24281-xx shown. For XMT and ...

  • Emerson Process Management 369P - page 15

    15 Instruction Sheet TUpH 396P and 396PVP Sensors LIQ_MAN_ABR_396P_396PVP July 2013 Figure 25. Wiring Details of 396P-01 for use with 1055-01-10-22 Figure 26. Wiring Details of 396P-01 for use with 1055-01-10-22-32 ...

  • Emerson Process Management 369P - page 16

    16 TUpH 396P and 396PVP Sensor s Instruc tion Sheet July 2013 LI Q_MAN_ABR_396P_396PVP Figure 27. Wiring Details of 396P-02 for use with 1055-01-10-22-32 Figure 28. Wiring Details of 396PVP for use with 1055-01-10-22-32 ...

  • Emerson Process Management 369P - page 17

    17 Instruction Sheet TUpH 396P and 396PVP Sensors LIQ_MAN_ABR_396P_396PVP July 2013 Figure 29. Wiring Details of 396P-01 Gray Cable for use with XMT -P-HT -10 If present, remove shrink slee ve from gra y reference wire before connecting wire to terminal. For XMT and 1056 termina te Inner Drain with a wire nut. Figure 30. Wiring Details of 396P-01 B ...

  • Emerson Process Management 369P - page 18

    18 TUpH 396P and 396PVP Sensor s Instruc tion Sheet July 2013 LI Q_MAN_ABR_396P_396PVP Figure 31. Wiring Details of 396P-02 Gray Cable for use with XMT -P-HT -10 If present, remove shrink slee ve from gra y reference wire before connecting wire to terminal. F or XMT and 1056 terminate Inner Drain with a wire nut. Figure 32. Wiring Details of 396P-0 ...

  • Emerson Process Management 369P - page 19

    19 Instruction Sheet TUpH 396P and 396PVP Sensors LIQ_MAN_ABR_396P_396PVP July 2013 Figure 33. Wiring Details of 396PVP Gray Cable for use with XMT -P-HT -10 If present, remove shrink slee ve from gra y reference wire before connecting wire to terminal. F or XMT and 1056 terminate Inner Drain with a wire nut. Figure 34. Wiring Details of 396PVP Blu ...

  • Emerson Process Management 369P - page 20

    20 TUpH 396P and 396PVP Sensor s Instruc tion Sheet July 2013 LI Q_MAN_ABR_396P_396PVP Figure 35. Wiring through j -bo x for 1056 Figure 36. Wiring Details of 396P-01 Gray Cable for 1056/56 If present, remove shrink slee ve from gra y reference wire before connecting wire to terminal. F or XMT and 1056 terminate Inner Drain with a wire nut. ...

  • Emerson Process Management 369P - page 21

    21 Instruction Sheet TUpH 396P and 396PVP Sensors LIQ_MAN_ABR_396P_396PVP July 2013 Figure 37. Wiring Details of 396P-01 Blue Cable for 1056/56 If present, remove shrink slee ve from gra y reference wire before connecting wire to terminal. F or XMT and 1056 terminate Inner Drain with a wire nut. Figure 38. Wiring Details of 396P-02 Blue Cable for 1 ...

  • Emerson Process Management 369P - page 22

    22 TUpH 396P and 396PVP Sensor s Instruc tion Sheet July 2013 LI Q_MAN_ABR_396P_396PVP Figure 39. Wiring Details of 396P-02 Gray Cable for 1056/56 If present, remove shrink slee ve from gra y reference wire before connecting wire to terminal. F or XMT and 1056 terminate Inner Drain with a wire nut. Figure 40. Wiring Details of 396PVP Gray Cable for ...

  • Emerson Process Management 369P - page 23

    23 Instruction Sheet TUpH 396P and 396PVP Sensors LIQ_MAN_ABR_396P_396PVP July 2013 Figure 41. Wiring Details of 396PVP Blue Ca ble for 1056/56 For XMT and 1056 termina te inner Drain with a wire nut. VP8 Cable par t number 24281-xx shown Figure 42. pH/ORP Sensor Wiring for 1056/1057/56/396VP-70 (Blue Cable) ...

  • Emerson Process Management 369P - page 24

    24 TUpH 396P and 396PVP Sensor s Instruc tion Sheet July 2013 LI Q_MAN_ABR_396P_396PVP Figure 43. Wiring Details of 396P-01 Gray Cable for 1057 Figure 44. Wiring Details of 396P-01 Blue Cable for 1057 ...

  • Emerson Process Management 369P - page 25

    25 Instruction Sheet TUpH 396P and 396PVP Sensors LIQ_MAN_ABR_396P_396PVP July 2013 Figure 45. Wiring Details of 396P-02 Gray Cable for 1057 Figure 46. Wiring Details of 396P-02 Blue Cable for 1057 ...

  • Emerson Process Management 369P - page 26

    26 TUpH 396P and 396PVP Sensor s Instruc tion Sheet July 2013 LI Q_MAN_ABR_396P_396PVP Figure 47. Wiring Details of 396PVP Gray Cable for 1057 Figure 48. Wiring Details of 396PVP Blue Ca ble for 1057 ...

  • Emerson Process Management 369P - page 27

    27 Instruction Sheet TUpH 396P and 396PVP Sensors LIQ_MAN_ABR_396P_396PVP July 2013 Figure 49. pH/ORP Sensor Wiring for 1066/396P-01 (Gray Cable) Figure 50. pH/ORP Sensor Wiring for 1066/396P-01 (Blue Cable) ...

  • Emerson Process Management 369P - page 28

    28 TUpH 396P and 396PVP Sensor s Instruc tion Sheet July 2013 LI Q_MAN_ABR_396P_396PVP Figure 51. pH/ORP Sensor Wiring for 1066/396P-02 (Gray Cable) Figure 52. pH/ORP Sensor Wiring for 1066/396P-02 (Blue Cable) ...

  • Emerson Process Management 369P - page 29

    29 Instruction Sheet TUpH 396P and 396PVP Sensors LIQ_MAN_ABR_396P_396PVP July 2013 Figure 53. pH/ORP Sensor Wiring for 1066/396PVP (Gray Cable) Figure 54. pH/ORP Sensor Wiring for 1066/396PVP (Blue Cable) ...

  • Emerson Process Management 369P - page 30

    30 TUpH 396P and 396PVP Sensor s Instruc tion Sheet July 2013 LI Q_MAN_ABR_396P_396PVP Figure 55. pH/ORP Sensor Wiring for 1066/396PVP-70 (Blue Cable) Figure 56. Wiring Details of 396P-01 Gray Cable for 5081-P-HT ...

  • Emerson Process Management 369P - page 31

    31 Instruction Sheet TUpH 396P and 396PVP Sensors LIQ_MAN_ABR_396P_396PVP July 2013 Figure 57. Wiring Details of 396P-01 Blue Cable for 5081-P-HT Figure 58. Wiring Details of 396P-02 Gray Cable for 5081-P-HT ...

  • Emerson Process Management 369P - page 32

    32 TUpH 396P and 396PVP Sensor s Instruc tion Sheet July 2013 LI Q_MAN_ABR_396P_396PVP Figure 59. Wiring Details of 396P-02 Blue Cable for 5081-P-HT Figure 60. Wiring Details of 396PVP Gray Cable for 5081-P-HT ...

  • Emerson Process Management 369P - page 33

    33 Instruction Sheet TUpH 396P and 396PVP Sensors LIQ_MAN_ABR_396P_396PVP July 2013 Figure 61. Wiring Details of 396PVP Blue Ca ble for 5081-P -HT Figure 62. pH/ORP Sensor Wiring for 5081/398RVP -70 (Blue Cable) ...

  • Emerson Process Management 369P - page 34

    34 TUpH 396P and 396PVP Sensor s Instruc tion Sheet July 2013 LI Q_MAN_ABR_396P_396PVP Figure 63. Wiring Details of 396P-01 Gray Cable for use with 6081 Figure 64. Wiring Details of 396P-01 Blue Cable for use with 6081 ...

  • Emerson Process Management 369P - page 35

    35 Instruction Sheet TUpH 396P and 396PVP Sensors LIQ_MAN_ABR_396P_396PVP July 2013 Figure 66. Wiring Details of 396P-02 Blue Cable for use with 6081 Figure 65. Wiring Details of 396P-02 Gray Cable for use with 6081 ...

  • Emerson Process Management 369P - page 36

    36 TUpH 396P and 396PVP Sensor s Instruc tion Sheet July 2013 LI Q_MAN_ABR_396P_396PVP Figure 67. Wiring Details of 396PVP Gray Cable for use with 6081 Figure 68. Wiring Details of 396PVP Blue Ca ble for use with 6081 ...

  • Emerson Process Management 369P - page 37

    37 Instruction Sheet TUpH 396P and 396PVP Sensors LIQ_MAN_ABR_396P_396PVP July 2013 Figure 69. pH/ORP Sensor Wiring for 6081/3400HT VP-70 (Blue Cable) ...

  • Emerson Process Management 369P - page 38

    38 TUpH 396P and 396PVP Sensor s Instruc tion Sheet July 2013 LI Q_MAN_ABR_396P_396PVP Part Number 1 7007 16 ...

  • Emerson Process Management 369P - page 39

    39 Instruction Sheet TUpH 396P and 396PVP Sensors LIQ_MAN_ABR_396P_396PVP July 2013 Part Description 22698-00 Preamplie r plug-in for junction box, for 1003 22698-02 Preamplie r plug-in for junction box, for 1181, 1050 22698-03 Preamplie r plug-in for junction box f or 1054 A/B, 2054, 2081 22743-01 Pt100 preamplier f or 1181 22744-01 3K ...

  • Emerson Process Management 369P - page 40

    TUpH 396P and 396PVP Sensor s Instruc tion Sheet July 2013 LI Q_MAN_ABR_396P_396PVP R e v . J Credit Cards for U.S. Purchases Only. Emerson Proces s Manage ment 2400 Barranca Park wa y Ir vine , CA 92606 USA T el: (949) 757-8500 F ax: (949) 474-7250 rosemo untanalytical.com © Ro semount An aly tical Inc. 2013 ©2013 Ro semount An aly tical, Inc. A ...

Fabricant Emerson Process Management Catégorie Stud Sensor

Les documents que nous recevons du fabricant de l'appareilEmerson Process Management 369P peuvent être divisés en plusieurs groupes. Ceux-ci sont, entre autres:
- dessins techniques Emerson Process Management
- manuels d’utilisations 369P
- fiches produit Emerson Process Management
- dépliants
- ou étiquettes-énergie Emerson Process Management 369P
Tous sont importants, mais les informations les plus importantes du point de vue de l'utilisation de l'appareil se trouvent dans le manuel d’utilisation Emerson Process Management 369P.

Un groupe de documents appelé manuels d’utilisation est également divisé en types plus spécifiques, tels que: Manuels d’installation Emerson Process Management 369P, manuels d’entretien, brefs manuels ou manuels de l’utilisateur Emerson Process Management 369P. Selon vos besoins, vous devriez chercher le document dont vous avez besoin. Sur notre site, vous pouvez voir le manuel le plus populaire d’utilisation du produit Emerson Process Management 369P.

Manuel d’utilisation complet de l’appareil Emerson Process Management 369P, quelle devrait-elle être?
Le manuel d’utilisation, également appelé le mode d’emploi, ou tout simplement le manuel, est un document technique destiné à aider à utiliser Emerson Process Management 369P par les utilisateurs. Des manuels sont généralement écrits par un rédacteur technique, mais dans un langage accessible à tous les utilisateurs Emerson Process Management 369P.

Le manuel d’utilisation complet Emerson Process Management, devrait inclure plusieurs éléments de base. Certains d'entre eux sont moins importants, tels que: la couverture / page de titre ou pages d'auteur. Cependant, la partie restante, devrait nous fournir des informations importantes du point de vue de l'utilisateur.

1. Introduction et des conseils sur la façon d'utiliser le manuel Emerson Process Management 369P - Au début de chaque manuel, nous devrions trouver des indices sur la façon d'utiliser le document. Il doit contenir des informations sur l'emplacement de la table des matières Emerson Process Management 369P, FAQ ou des problèmes les plus fréquents - les points qui sont les plus souvent recherchés par les utilisateurs de chaque manuel
2. Table des matières - index de tous les conseils pour lEmerson Process Management 369P qui peuvent être trouvés dans le document courant
3. Conseils sur la façon d'utiliser les fonctions de base de l’appareil Emerson Process Management 369P - qui devraient nous aider dans les premières étapes lors de l'utilisation Emerson Process Management 369P
4. Troubleshooting - séquence systématique des activités qui nous aideront à diagnostiquer et ensuite résoudre les principaux problèmes de Emerson Process Management 369P
5. FAQ - questions fréquemment posées
6. Détails du contact Informations sur l'endroit où chercher le contact avec le fabricant / service Emerson Process Management 369P dans un pays donné, si le problème ne peut être résolu par nous-mêmes.

Avez-vous une question à propos de Emerson Process Management 369P?

Utiliser le formulaire ci-dessous

Si vous n’avez pas résolu votre problème avec Emerson Process Management 369P, avec l'aide du manuel que vous avez trouvé, posez une question en utilisant le formulaire ci-dessous. Si un utilisateur a eu un problème similaire avec Emerson Process Management 369P il est probable qu’il a envie de partager la façon de le résoudre.

Réécrire le texte de l'image

Commentaires (0)