Manuel d’utilisation Princeton 4411-0143

62 pages 7.18 mb
Télécharger

Aller à la page of 62

Summary
  • Princeton 4411-0143 - page 1

    so lane Manual ...

  • Princeton 4411-0143 - page 2

    4411 - 0143 Version 1.B May 8, 2013 www.princetoninstruments.com 15 Discovery Way, Acton, MA 10720 Phone: 978- 263 -3584 Fax: 978- 263 - 5086 ...

  • Princeton 4411-0143 - page 3

     Copyright 2012-2013 Princeton Instruments, a division of Roper Scientific, Inc. 3660 Quakerbridge Rd Trenton, NJ 08619 TEL: 800-874-9789 / 609-587-9797 FAX: 609-587-1970 www. princetoninstruments.com All rights reserved. No part of this publication may be reproduced b y any means without the written permission of Princeton Instruments, a divisi ...

  • Princeton 4411-0143 - page 4

    iii T abl e of Cont ent s Chapter 1 Is oPlane ® SCT 32 0 Description ............................................ 7 1.1 Introduction .......................................................................................................................7 1.2 IsoPlane SCT 320 Description .............................................................. ...

  • Princeton 4411-0143 - page 5

    iv IsoPlane ® SCT 320 Ma nual Version 1.B Append ix A Using the Mono chromator Con trol Software .................. 39 A.1 Introduction .................................................................................................................... 39 A.2 Install the Monochromator Control Software .............................................. ...

  • Princeton 4411-0143 - page 6

    Table of Conten ts v v Figure 2-18. Lift Turret up out of the Turret Compartment .........................................................27 Figure 2-19. Location of Hole on Turret Grating Assembly ........................................................27 Figure 2-20. Location of Post on Grating Spindle Assembly ................................... ...

  • Princeton 4411-0143 - page 7

    vi IsoPlane ® SCT 320 Ma nual Version 1.B This page intentionally left blank. ...

  • Princeton 4411-0143 - page 8

    7 Chapter 1 IsoPl ane ® SCT 320 Descript ion 1.1 Introduction This manual is intended to help you set up and operate your new IsoPlane SCT 320 spectrograph. Even if you are an experienced user of spec troscopic equipment, we suggest that you follow the manual (at least initially ) to insure proper setup and operation. If y ou have any questions ab ...

  • Princeton 4411-0143 - page 9

    8 IsoPlane ® SCT 320 Ma nual Version 1.B This page intentionally left blank. ...

  • Princeton 4411-0143 - page 10

    9 Chapter 2 IsoPl ane Set up 2.1 Unpacking and Inspecting Carefully unpack and examine the IsoPlane and any purchased accessories. Notes: 1. Report any dam age imm ediately to the carrier and to P rinceton I nstruments . 2. Save all packing material. 2.2 Initial Setup Procedure 1. Position the IsoPlane where it will be used. 2. Connect the power su ...

  • Princeton 4411-0143 - page 11

    10 IsoPlane ® SCT 320 Ma nual Version 1.B 2.2.1.3 SpectraSense ™ SpectraSense is a comprehensive spectral acquisition and da ta treatment package. It is designed to work exclusively with Princeton Instruments products and accessories including the SpectraHub Detector Interface. SpectraSense software will control and acquire data from both single ...

  • Princeton 4411-0143 - page 12

    Chapter 3 SCT 320 Operat ion 11 2.3.2 Term inal or RS-232 Port Sett ings A terminal or RS-232 computer port must be set up as follows: 9600 baud, 8 data bits, no parity, 1 start bit, 1 stop bit. 2.4 Power Supply with Input Cable Figure 2- 2. Power Supply Pin# Description 1 +5V 2 GND 3 GND 4 +24 Table 2- 2 . IsoP lane Power Connecto r Pin Arrangemen ...

  • Princeton 4411-0143 - page 13

    12 IsoPlane ® SCT 320 Ma nual Version 1.B Turret Compartment of the manual slit assembly has been provided in Appendix F. The general procedure for mounting an accessory to the slit is as follows: 1. Place the accessory directly ag ainst the face of the slit bod y . 2. Using four (4) 8-32 screws normally provided with the accessory, sec ure the ac ...

  • Princeton 4411-0143 - page 14

    Chapter 3 SCT 320 Operat ion 13 7. L oca te t he ve r ti c a l po s t on th e G ra ti ng Sp ind le As sem b ly (s ee F ig ur e 2 -6 ). 8. Po si ti on th e g ra tin g t u rr et on t he g ra ti ng sp ind le so the ho le i n the tu r re t a sse mb ly a li gn s w it h th e po st in the tu rre t a sse mb ly a s s how n i n F ig ure 2- 8. 9. Ma ke su re ...

  • Princeton 4411-0143 - page 15

    14 IsoPlane ® SCT 320 Ma nual Version 1.B Figure 2- 6. Location of P ost on Grating Spin dle Assembly Figure 2- 7. Lower Turret into the Turret Compartmen t Figure 2- 8. Turret Installed (Hole an d Post Align ed) Post Aligned Hole and Post ...

  • Princeton 4411-0143 - page 16

    Chapter 3 SCT 320 Operat ion 15 2.8 Ar ray Detector Mounting Instructions 2.8.1 Introduct ion The standard mounting plate for CCDs, FPAs, and diode arrays accommodates detec tors with a slotted flange or mounting holes aligne d on a 3.60” bolt circle pattern. There are six (6) equally spaced #10-32 tapped holes on a 3.60” bolt circle designed t ...

  • Princeton 4411-0143 - page 17

    16 IsoPlane ® SCT 320 Ma nual Version 1.B 2.8.2 Mounting an Array Detector to the IsoPlane The array detector mounting plate is mounted to the entrance port of the IsoPlane. 1. Remove the shipping cover from the mounting plate. 2. If the detector has a three-slotted flange (3.60” bolt circle): a. If the detector has a three-slotted flange, screw ...

  • Princeton 4411-0143 - page 18

    Chapter 3 SCT 320 Operat ion 17 2.8.3 Rotationa l Alignment This general procedure assumes that you ha ve alread y mounte d the array detector to the exit port of the spectrograph and that there is a light source at the entrance port. The basic actions are listed below. Refer to Chapter 3 of this manual for detailed LightField and WinSpec instructi ...

  • Princeton 4411-0143 - page 19

    18 IsoPlane ® SCT 320 Ma nual Version 1.B 2.8.4 Focusin g the Spectrom eter Figure 2- 10 . Micrometer Comp artment Location The basic actions for focusing the spectrometer are listed below. This general procedure assumes that you have already mounted the array detector to the exit port of the spectrograph and that there is a light source at the en ...

  • Princeton 4411-0143 - page 20

    Chapter 3 SCT 320 Operat ion 19 2.8.5 Height Adjus tment of Im age Plane Note: This proced ure should only be attempted by so meone experienced in aligning optical components and a ssemblies. 2.8.5.1 Overview D ur i ng t he m a nu fa c tu r in g p ro ces s, the I so Pl ane fo c al pl a ne is f ac tory -a lig ne d so th a t it is pr ec is e ly ce nt ...

  • Princeton 4411-0143 - page 21

    20 IsoPlane ® SCT 320 Ma nual Version 1.B 2. Adjustm ents to the Cap a nd Set Screws s hould be sm all, approximately 1/8 th -turn or less. 3. If the Cap Screw is adjusted 1/8 th -turn in a counter-cl ockwise direct ion, then the Set S crew should be adjusted 1/8 th -turn in a cl ockwise direct ion. 4. If the i nitial adjustm ent does not m ove th ...

  • Princeton 4411-0143 - page 22

    Chapter 3 SCT 320 Operat ion 21 2.10 Spectrometer Calibration Refer to the application software manual or the software help for information on calibrating the data pixels with wavelengths. 2.11 Slit Width Adjustments in the Model 716-1 Manual Bilateral Slit and the 716-MS Kinematic Slit The slit width of each bilateral slit assembly (716 type) is a ...

  • Princeton 4411-0143 - page 23

    22 IsoPlane ® SCT 320 Ma nual Version 1.B 2. 12 S li t W idt h A d ju st me nt s in the Mo d el 71 8- 1 Mo to rize d 3m m S li t The Model 718-1 motorized slit assembly is a self-calibrating, 10 µm to 3 mm adjustable bilateral slit. Stepping-motor-controlled adjustments are made in 1 µm increments over the full range. The Model 718-12 motorized ...

  • Princeton 4411-0143 - page 24

    Chapter 3 SCT 320 Operat ion 23 2.13.2 Re movin g an Intern al Shutter If you are planning to use a PI-MAX detector (or other detector without an external shutter control connector) and the IsoPlane has an internal shutter, you will e ither have to remove the shutter or control the shutter via a Princeton Instruments SHC-EXT external shutter contro ...

  • Princeton 4411-0143 - page 25

    24 IsoPlane ® SCT 320 Ma nual Version 1.B 2.14 Turret Removal CAUTION 1. The optical sur faces in the spec trograph are extr emely delicate and can be perm anently dam aged by contact with so lid objects as wel l as with most liquid s and aeroso ls. 2. DO NOT T OUCH optical surfaces w ith anything. Wear thin protectiv e powderless glov es to minim ...

  • Princeton 4411-0143 - page 26

    Chapter 3 SCT 320 Operat ion 25 Figure 2- 16 . Lift Turret up out of the Turret Compa rtment 6. I f y ou are n ot i ns ta lli ng /re mo vi ng a sh ut ter a ss e mb ly or i nsta ll in g a d iff ere nt tur r e t, r epl a ce a nd se cu re t he Tu rre t Com par tme nt co ve r. 2.15 Turret Interchange Re ad a nd fo ll ow the i nst ru c ti ons in th i s ...

  • Princeton 4411-0143 - page 27

    26 IsoPlane ® SCT 320 Ma nual Version 1.B 6. Us e e xt rem e c are so as no t t o to uc h t he g ra ti ng s ur fa ce s . Gra sp the Ce nte r T or qu e S c re w a nd li ft th e g r a ti ng tu r re t st ra ig ht up off th e g rat ing sp ind le a sse mbly ( se e F igur e 2- 18 ). I f th e tu rr et ca n no t b e l ift e d up , y ou may ne ed to lo ose ...

  • Princeton 4411-0143 - page 28

    Chapter 3 SCT 320 Operat ion 27 Figure 2- 18 . Lift Turret up out of the Turret Compa rtment Figure 2- 19 . Loca tion of Hole on Tu rret Grating Assembly Figure 2- 20 . Loca tion of Post on Grating Spindle Assemb ly Hole Post ...

  • Princeton 4411-0143 - page 29

    28 IsoPlane ® SCT 320 Ma nual Version 1.B Figure 2- 21 . Turret Installed (Hole and Post Aligned) Aligned Hole and Post ...

  • Princeton 4411-0143 - page 30

    29 Chapter 3 Oper a t ion 3.1 Introduction This chapter reviews how to enable the control of an IsoPlane and Princeton I nstruments detector by either the LightField (64-bit) or WinSpec (32-bit) software. This chapter also points out how to perform alignment using the software. Further information about operating the system can be found in the appr ...

  • Princeton 4411-0143 - page 31

    30 IsoPlane ® SCT 320 Ma nual Version 1.B 3.4 LightField Start LightField after making sure the detector and I soPlane are connected to the computer and are turned on. When LightField boots it will look for a vailable devices and will load their icons in the Available Devices panel. Before you can design a new experiment or run an existing one, th ...

  • Princeton 4411-0143 - page 32

    Chapter 3 SCT 320 Operat ion 31 Figure 3- 2. LightField E xperiment Settings stack an d Experiment Devices panel 3. Note that the Experiment Settings stac k on the left now displays several expanders. Because this is a new experiment, the default settings for the detector will be active. The Status bar (a t the bottom of the window) displays an ico ...

  • Princeton 4411-0143 - page 33

    32 IsoPlane ® SCT 320 Ma nual Version 1.B 3.4.2 Perform Alignment Because the detector is mounted directly to the mounting plate on the I soPlane, the rotational alignment and focusing operations are different from the way these operations are performed for an Acton SP Series spec trograph. The following information assumes that you are familiar w ...

  • Princeton 4411-0143 - page 34

    Chapter 3 SCT 320 Operat ion 33 9. Click on the peak you will be monitoring for the alignment. This will display the data cursor, which you can position at the top of the peak. Since the da ta cursor spans the ROIs (graphs), you can use the data cursor as your vertical reference. 10. Slowly rotate the camera until the peaks align in all of the ROI ...

  • Princeton 4411-0143 - page 35

    34 IsoPlane ® SCT 320 Ma nual Version 1.B 3.5 WinSpec Start WinSpec after making sure the detector and IsoPlane are connected to the computer a nd are turned on. WinSpec should find the detector and present a dialog for selecting the detector for your experiment. Before y ou can use the IsoPlane, you must tell WinSpec that it is the spectrograph y ...

  • Princeton 4411-0143 - page 36

    Chapter 3 SCT 320 Operat ion 35 2. On the Define Spectrograph dialog (Figure 3-6), click on the Install/Remove Spectrograph button. 3. On the Install/Remove Spectrograph dialog, scroll down the list of supported spectrographs, click on Acton SCT320 , click on the Install Selected Spectrograph button, and then click on OK to return to the Define Spe ...

  • Princeton 4411-0143 - page 37

    36 IsoPlane ® SCT 320 Ma nual Version 1.B 3.5.3 Perform Alignment The initial alignment described in this section assumes that you ha ve alread y connected and turned on the detector and spectrograph, have opened WinSpec, and identified the detec tor and spectrograph. After you have aligned the optics the first time, you may want to repeat the pro ...

  • Princeton 4411-0143 - page 38

    Chapter 3 SCT 320 Operat ion 37 3. Set the slit to 10 µm at a minimum. If necessary, adjust the Exposure Time to maintain optimum (near full-scale) signal intensity. 4. Wait until the detector locks at its default temperature. If the detector is LN-cooled, it should have been filled after the detector was powered ON. 5. Turn on Focus mode. 6. Adju ...

  • Princeton 4411-0143 - page 39

    38 IsoPlane ® SCT 320 Ma nual Version 1.B This page intentionally left blank. ...

  • Princeton 4411-0143 - page 40

    39 Appendix A U sing the Monochr omator C ont r ol Soft war e A.1 Introduction As has been previously mentioned in this manual, the Acton IsoPlane spectrograph can be controlled using Princeton I nstruments’ LightField, WinSpec, or SpectraSense software. It can also controlled from a computer using the supplied Monochromator Control software via ...

  • Princeton 4411-0143 - page 41

    40 IsoPlane ® SCT 320 Ma nual Version 1.B Par amete rs, if nee ded, p rec ede the comm and an d are separ ated fro m t he c ommand by a t le as t on e spa ce (e .g. , 5 46.0 7 GOTO ). F or RS - 23 2 o pe rat ion , t he p or t s e t- up is 9 600 ba ud , 8 da ta b its , 1 st o p bi t a nd no par ity . A co nve ni en t to ol f or try ing ou t t his m ...

  • Princeton 4411-0143 - page 42

    Appendix A Using the Monochrom at or Control So ftware 41 A.4.2 Gratin g Control Com mands GRATING Places specified grating in position to the wavelength of the wavelength on the p resent grating. Up to nine (9) gratings are allowed on three (3) turrets. This command takes a grating number from 1 - 9. IMPORTANT NOTE: This comm and assumes that the ...

  • Princeton 4411-0143 - page 43

    42 IsoPlane ® SCT 320 Ma nual Version 1.B A.4.3 Slit Widt h Control Com mands (for Motorized S lits only) SIDE-EXIT-SL IT Designates side exit slit to receive slit control commands. SIDE- ENT -SLIT Designates side entrance slit to receive slit control commands. MICRONS Sets the slit width for the designated slit in the range of 10 to 3000 microns ...

  • Princeton 4411-0143 - page 44

    43 Appendix B Spect ro graph to C omp uter Ca bl e Connect ion Figure B-1 . RS- 232 Cable Wiring Figure B-2 . IsoPlane Cab le Connectors FILTER Filter driver connection for Acton FA-2448 Filter Wheel. SHT TTL SHT TTL connection for TTL control of shutter. DSP DSP connection to connect two spectrographs as doubles. COM Com light flashes when communi ...

  • Princeton 4411-0143 - page 45

    44 IsoPlane ® SCT 320 Ma nual Version 1.B This page intentionally left blank. ...

  • Princeton 4411-0143 - page 46

    45 Appendix C Ac cessories List ing Slit Assembly SP - 716 -1 Bilaterally adjustable ENTRANC E slit assembly for IsoPlane monochromator or spectrograph. Micrometer adjustable widths from 10 µm to 3 mm. Supplied with interchangeable 4.0 mm and 14.0 mm high baffles. Triple Grating Turr et 300 - 483 Interchangeable triple grating turret for the IsoPl ...

  • Princeton 4411-0143 - page 47

    46 IsoPlane ® SCT 320 Ma nual Version 1.B Det ectors and Acces sories: All detectors include mounting flanges for I soPlane slit assemblies. PD - 438 Universal Photomultiplier Tube Housing for standard 1 1/8" side window photomultiplier tubes. Requires side window PMT, and photometer/power supply for operation. Accepts Acton tubes P1, P2 or P ...

  • Princeton 4411-0143 - page 48

    Appendix C Accessories Lis ting 47 Filter Wheels and Ord er Sorting Filter Ass emblies FA - 2448 Six Position Filter Wheel Assembly. Manual sample indexing, holds up to six samples, each 1.0” diameter; automated via SP Control Cable. Can be used with the IsoPlane. Requires FA-2448-F Filter Set or customer-supplied filters of 25 mm (1 inch) diamet ...

  • Princeton 4411-0143 - page 49

    48 IsoPlane ® SCT 320 Ma nual Version 1.B FC - 446 - 030 Im aging Fiber Adapter is designed specifically for our imaging spectrographs. The all-reflective design eliminates chromatic a berrations and the aspheric mirror cancels astigmatism, allowing precise imaging of fibers at the spectrograph entrance slit. Includes up/down, left/right screw ali ...

  • Princeton 4411-0143 - page 50

    49 Appendix D Gr at ings f or IsoPl ane Grooves/mm Blaze Wavelength Standard Part Number Mechanical Limit Notes: † Bl az e d H ol og r ap hic , 1 90- 450 n m , op tim um ra nge †† Op ti m um rang e f or t his gr ati ng i s f rom 45 0 n m t o th e u pp er me ch an ica l l imi t Standard Gratings are 68 X 68 mm. 50 600 nm I3-005-600 -P 150 500 ...

  • Princeton 4411-0143 - page 51

    50 IsoPlane ® SCT 320 Ma nual Version 1.B This page intentionally left blank. ...

  • Princeton 4411-0143 - page 52

    51 Appendix E Outl ine Dra win gs Figure E-1 . IsoPlane: 320 with Manual Slit Assembly Outline Drawing ...

  • Princeton 4411-0143 - page 53

    52 IsoPlane ® SCT 320 Ma nual Version 1.B Figure E-2 . IsoPlane: 320 with Motorized Slit Assemb ly Outline Dra wing ...

  • Princeton 4411-0143 - page 54

    53 Appendix F S t and ar d Sl it Assem bl y Dr a wing Figure F-1 . Standard Slit Assembly Drawing ...

  • Princeton 4411-0143 - page 55

    54 IsoPlane ® SCT 320 Ma nual Version 1.B This page intentionally left blank. ...

  • Princeton 4411-0143 - page 56

    55 Appendix G Decl ar at ion of Conf ormit y DECLARATION OF CO NFORMITY We, the undersigned Manufacturer: Princeton Instruments Address, City: 15 Discovery Way, Acto n, MA 01720 Country: USA Certify and declare under our sole responsibi lity that the following apparatus: IsoPlane SCT-320 with Power Supply Conforms to the emissions requirements of E ...

  • Princeton 4411-0143 - page 57

    56 IsoPlane ® SCT 320 Ma nual Version 1.B This page intentionally left blank. ...

  • Princeton 4411-0143 - page 58

    57 Appendix H Cert ificat ion and W arr ant y Certification Princeton Instruments certifies that this instrument was thoroughly tested and found to meet the specifications furnished by Princeton Instruments when it was shipped fr om the factory. Warranty Princeton Instruments instruments and accessories are warranted f or a period of one full y ea ...

  • Princeton 4411-0143 - page 59

    58 IsoPlane ® SCT 320 Ma nual Version 1.B This page intentionally left blank. ...

  • Princeton 4411-0143 - page 60

    59 Ind e x A-B Accessories listing of ....................................................................... 45 MicroSpec ................................................................... 46 mounting to slit assembly ................................ ............ 11 Additional commands ....................................................... 42 Ad ...

  • Princeton 4411-0143 - page 61

    60 IsoPlane ® SCT 320 Ma nual Version 1.B This page intentionally left blank. ...

  • Princeton 4411-0143 - page 62

    Contact Princeton Instruments for additional inf or mation. www .princetoninstruments.com | info@princetoninstruments.com USA TOLL-FREE +1.877.474.2286 | USA +1.609.587.9797 ...

Fabricant Princeton Catégorie Webcam

Les documents que nous recevons du fabricant de l'appareilPrinceton 4411-0143 peuvent être divisés en plusieurs groupes. Ceux-ci sont, entre autres:
- dessins techniques Princeton
- manuels d’utilisations 4411-0143
- fiches produit Princeton
- dépliants
- ou étiquettes-énergie Princeton 4411-0143
Tous sont importants, mais les informations les plus importantes du point de vue de l'utilisation de l'appareil se trouvent dans le manuel d’utilisation Princeton 4411-0143.

Un groupe de documents appelé manuels d’utilisation est également divisé en types plus spécifiques, tels que: Manuels d’installation Princeton 4411-0143, manuels d’entretien, brefs manuels ou manuels de l’utilisateur Princeton 4411-0143. Selon vos besoins, vous devriez chercher le document dont vous avez besoin. Sur notre site, vous pouvez voir le manuel le plus populaire d’utilisation du produit Princeton 4411-0143.

Manuels d’utilsiation similaires

Manuel d’utilisation complet de l’appareil Princeton 4411-0143, quelle devrait-elle être?
Le manuel d’utilisation, également appelé le mode d’emploi, ou tout simplement le manuel, est un document technique destiné à aider à utiliser Princeton 4411-0143 par les utilisateurs. Des manuels sont généralement écrits par un rédacteur technique, mais dans un langage accessible à tous les utilisateurs Princeton 4411-0143.

Le manuel d’utilisation complet Princeton, devrait inclure plusieurs éléments de base. Certains d'entre eux sont moins importants, tels que: la couverture / page de titre ou pages d'auteur. Cependant, la partie restante, devrait nous fournir des informations importantes du point de vue de l'utilisateur.

1. Introduction et des conseils sur la façon d'utiliser le manuel Princeton 4411-0143 - Au début de chaque manuel, nous devrions trouver des indices sur la façon d'utiliser le document. Il doit contenir des informations sur l'emplacement de la table des matières Princeton 4411-0143, FAQ ou des problèmes les plus fréquents - les points qui sont les plus souvent recherchés par les utilisateurs de chaque manuel
2. Table des matières - index de tous les conseils pour lPrinceton 4411-0143 qui peuvent être trouvés dans le document courant
3. Conseils sur la façon d'utiliser les fonctions de base de l’appareil Princeton 4411-0143 - qui devraient nous aider dans les premières étapes lors de l'utilisation Princeton 4411-0143
4. Troubleshooting - séquence systématique des activités qui nous aideront à diagnostiquer et ensuite résoudre les principaux problèmes de Princeton 4411-0143
5. FAQ - questions fréquemment posées
6. Détails du contact Informations sur l'endroit où chercher le contact avec le fabricant / service Princeton 4411-0143 dans un pays donné, si le problème ne peut être résolu par nous-mêmes.

Avez-vous une question à propos de Princeton 4411-0143?

Utiliser le formulaire ci-dessous

Si vous n’avez pas résolu votre problème avec Princeton 4411-0143, avec l'aide du manuel que vous avez trouvé, posez une question en utilisant le formulaire ci-dessous. Si un utilisateur a eu un problème similaire avec Princeton 4411-0143 il est probable qu’il a envie de partager la façon de le résoudre.

Réécrire le texte de l'image

Commentaires (0)