Manuel d’utilisation Agilent Technologies N5161A

38 pages 0.74 mb
Télécharger

Aller à la page of 38

Summary
  • Agilent Technologies N5161A - page 1

    Agilent T echnologies Agilent T echnologies N5161A/62A/81A/82A/ 83A MXG Signal Generators Installation Guide ...

  • Agilent Technologies N5161A - page 2

    Installati on Guide Notices © Agilent T e chnologies, Inc. 2006-2009 No part of this manual ma y be reproduc ed in any form or by any means (including elec- tronic storage and retrieval or translation into a foreign language) without p rior agree- ment and written consent from Agilent T echno logies, Inc. a s governed by Uni ted States and interna ...

  • Agilent Technologies N5161A - page 3

    Contents Agilent N5161A/62A/81A/82A/ 83A MXG Sign al Generators Installation Guide iii 1 Safety Information W arni ngs, Cautions, and Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 General Safety Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 ...

  • Agilent Technologies N5161A - page 4

    Contents Agilent N5161A/62A/81A /82A/83A MXG Signal Generators Installation Guide iv 4 Regulatory Information Cert ifica tion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 Assistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...

  • Agilent Technologies N5161A - page 5

    Agilent N5161A/62A/81A/82A /83A MXG Signal Generators Installation Guide v Documentation Overview Installa tion Guid e • Sa fe t y I n fo r ma t i on • Receiv ing the Instrument • Environmental & Electric al Requirements • Basic Setup • Accessori es • Operation V erification • Regu lato r y Infor mat ion User’s Guide • Signal ...

  • Agilent Technologies N5161A - page 6

    vi Agilent N5161A/62A/81A/82A/83A MXG Sig nal Generators Installation Guide SCPI R efe rence • SCPI Basics • Basic Funct ion Commands • LXI System Commands • System C ommands • Analog Modulation Commands • Arb Comm ands • Real- Time Comma nds • N5161A/62A/81A/82A SCPI Command Compatibi lity • N5183A SCPI Command Comp atibili ty Se ...

  • Agilent Technologies N5161A - page 7

    Agilent N5161A/62A/81A/82A /83A MXG Signal Generators Installation Guide 1 1 Safety Information • W arni ngs, Cautions, and Notes on page 1 • Gener al Sa fety Consi der ations on page 1 • Instr ume nt Mark ings on page 2 W arnings, Cautions, and Note s The documentation for this product uses the follo wing safety notations. Familiarize yourse ...

  • Agilent Technologies N5161A - page 8

    2 Agilent N5161A/ 62A/81A/82A/83A MXG Signal Generators Installation Guid e Safe ty I nfo rm atio n Instrument Marking s Instrumen t Mark ings The signal generator has t he follo wing markings. Familiarize yourself with each marking and its meaning before operating the sig nal generator . The ins tructi on manua l symbo l. The pr oduct is mar ked w ...

  • Agilent Technologies N5161A - page 9

    Agilent N5161A/62A/81A/82A /83A MXG Signal Generators Installation Guide 3 2 Getting Started NOTE For the N5161A/62A the soft ke y menus and features mentioned in this chapter are only a vailable thr ough the W eb- Enabled MX G or thr ough SCPI comma nds. Ref er to “Access ing the MXG (ATE) W eb- Enabled Page 1 ” on page 7 an d to the SCPI C om ...

  • Agilent Technologies N5161A - page 10

    4 Agilent N5161A/ 62A/81A/82A/83A MXG Signal Generators Installation Guid e Getting St a rted Checking the Shipm ent Checking the Shipment 1. Insp ect the ship ping container for damage. S i g n s o f d a m a g e c a n i n c l u d e a d e n t e d o r t o r n s h i p p i n g c o n t a i n e r o r c u s h i o n i n g m a t e r i a l t h a t indicates ...

  • Agilent Technologies N5161A - page 11

    Agilent N5161A/62A/81A/82A/83A M XG Signal Generators Installation Guide 5 Getting Started Meeting Environmen tal and Electrical Require ments Meeting Environmenta l and Electrical Require ments CAUTION T o av oid t he loss of dat a, GPIB settings, or current user instrument st ates t hat have not been permanently sav ed t o non- volatile memory , ...

  • Agilent Technologies N5161A - page 12

    6 Agilent N5161A/ 62A/81A/82A/83A MXG Signal Generators Installation Guid e Getting St a rted Meeting Environm ental and Electrical Requireme nts Connec ting th e AC Power Co rd This is a Safety Class 1 Product pro vided with a pr ot ective earth ground incorporated into th e po wer cord. The fr ont panel switch is only a stan dby sw itch; it is no ...

  • Agilent Technologies N5161A - page 13

    Agilent N5161A/62A/81A/82A/83A M XG Signal Generators Installation Guide 7 Getting Started Configuring the MXG A T E Configuring the MXG A TE NOTE For the N5161A/62A t he softkey menus and f eat ures mentioned in this guide are o nly a vailable t hrough the W eb- Enabled MXG or through SCPI commands. Ref er t o Accessin g the MXG (ATE) W eb- Enable ...

  • Agilent Technologies N5161A - page 14

    8 Agilent N5161A/ 62A/81A/82A/83A MXG Signal Generators Installation Guid e Getting St a rted Configuring the MXG A TE Figure 2-1 Web-Enabled MXG ATE 2 Note: If you do not see th is win dow , ch eck t o see i f the window is h idden behi nd you r br owse r window or you r web browse r settings ar e set to block pop- ups. T o use th is feature, you ...

  • Agilent Technologies N5161A - page 15

    Agilent N5161A/62A/81A/82A/83A M XG Signal Generators Installation Guide 9 Getting Started Configuring the Displa y (N5181A/82A/83A Only) Configuring the D isplay (N5181A/82A /83A Only) 1 Screen sav er settings are persistent st ates; t hey are unaffected by preset or a po wer cy cle. Use the ar ro w keys, numeric k eypad, or front panel knob to ad ...

  • Agilent Technologies N5161A - page 16

    10 Agilent N5161A/62A/81A/82A/83 A MXG Signal Generators Installation Guide Getting St a rted Configuring for Remote Control Configuring for Re mote Control 1 LAN Conf iguration Configur ing the LA N Interf ace 1 F or the N5161A/62A these sof tkey menus and features are only available through the W eb- En abled MXG A TE or thr ough SCPI com mands. ...

  • Agilent Technologies N5161A - page 17

    Agilent N5161A/62A/81A/82A/83A M XG Signal Generators Installation Guide 11 Getting Started Configuring for Remote Control Enabling LAN Service s: Browser , Sockets, and VXI-11 GPIB C onf igur ati on Use a brow ser to control the sign al generator Use a browser to view signal gene rato r file s. For det ails on a key , press Help and then the desir ...

  • Agilent Technologies N5161A - page 18

    12 Agilent N5161A/62A/81A/82A/83 A MXG Signal Generators Installation Guide Getting St a rted Ordering Accesso ries Ordering A ccessories Y ou can purcha se accessor ies or docume ntation at: http ://www.agi lent.com /find/mxg If you do not ha ve access to the In ter net, please contact your Agilent f ield engineer . See also, “Conta cting Agilen ...

  • Agilent Technologies N5161A - page 19

    Agilent N5161A/62A/81A/82A/83A M XG Signal Generators Installation Guide 13 Getting Started Ordering Accessories Prog ramm ing G uid e c N5180- 90005 n/ c Service Guide c N5180- 90006 n/c Docu men t Se t c : • Data Sheet • User’ s Gu ide • Prog ramm ing G uid e • SCPI Reference N5180- 90001 n/c Docu men t ati on CD- RO M c : PDF files : ? ...

  • Agilent Technologies N5161A - page 20

    14 Agilent N5161A/62A/81A/82A/83 A MXG Signal Generators Installation Guide Getting St a rted Prop er U se and Clea ning Proper Use and Cleaning The signal generator cover pr o tects against physical contact with int ernal assemblies that cont ain hazardous v oltages, but does not protect int ernal assemblies against cont act with liquids. T o av o ...

  • Agilent Technologies N5161A - page 21

    Agilent N5161A/62A/81A/82A /83A MXG Signal Generators Installation Guide 15 Operation V erification 3 Operation V erification NOTE For the N5161A/62A t he softkey menus and f eatures mentioned in this chapt er are only a vailable thr ough the W eb- Enabled MX G or thr ough SCPI comma nds. Ref er to “Access ing the MXG (ATE) W eb- Enabled Page 1 ? ...

  • Agilent Technologies N5161A - page 22

    16 Agilent N5161A/62A/81A /82A/83A MXG Signal Generators Installation Guide Operation V erification Running Self T est Running Self T est 1 Self T est is a series of internal tests of signal gene rat or f uncti ons. If t his t e st fail s, r efer to “Self T est Fai lure ” on pag e 17 for fur the r instr u ction s. Use the follo wing procedure t ...

  • Agilent Technologies N5161A - page 23

    Agilent N5161A/62A/81A/82A /83A MXG Signal Generators Installation Guide 17 Operation V erification Running Self T est • The current status of the self-test is: Failure. One or more tests have failed. System diagnostics indicate this test as the root failure: xxx If the si gnal generator f ails only one test, the tit le of the fai led test displa ...

  • Agilent Technologies N5161A - page 24

    18 Agilent N5161A/62A/81A /82A/83A MXG Signal Generators Installation Guide Operation V erification Freq uency R ange a nd Accur acy Ch eck Frequency Range and Ac curacy Check The frequency range is test ed by determining t he frequency accuracy relative to the timebase at t he freque ncy limit s of the s ignal generator . This test can be perf orm ...

  • Agilent Technologies N5161A - page 25

    19 Agilent N5161A/62A/81A/82A/83 A MXG Signal Generators Installation Guide Operation V erification Freq uency R ange a nd Accur acy Ch eck Frequenc y Count er Procedure (N5161A 1 /62A 1 /81A/8 2A) Te s t S et up 1. Connect t he equipment as show n. 2. Preset t he signal generato r: Press Preset . 3. T ur n m odul ation of f: Pr ess t he Mod On /Of ...

  • Agilent Technologies N5161A - page 26

    20 Agilent N5161A/62A/81A/82A/83 A MXG Signal Generators Installation Guide Operation V erification Freq uency R ange a nd Accur acy Ch eck Spectrum Analyz er Pro cedure (N5161A 1 /6 2A 1 /81A/82A/83 A) Te s t S et up 1. Connect t he equipment as show n. 2. V erify that th e spectrum analyzer is locked to the 10 MHz exter nal reference frequency. 3 ...

  • Agilent Technologies N5161A - page 27

    Agilent N5161A/62A/81A/82A /83A MXG Signal Generators Installation Guide 21 Operation V erification Check ing the O utput Pow er T roubleshooting Pr oblems with t he Frequency Ac curacy Ch eck • V erify the cables are co nnected correctly. • If you are us ing a frequency counter , verify that you are using the cor rect channel for the frequenci ...

  • Agilent Technologies N5161A - page 28

    22 Agilent N5161A/62A/81A/82A/83 A MXG Signal Generators Installation Guide Operation V erification Checking the Output Power N5161A 1 /81A T est Proce dure Te s t S et up 1. Zer o a nd ca libr ate the po we r se ns or to the po wer meter: 2. Connect t he equipment as show n: 3. Preset t he signal generato r: Press Preset . 4. T urn RF on: Press RF ...

  • Agilent Technologies N5161A - page 29

    23 Agilent N5161A/62A/81A/82A/83 A MXG Signal Generators Installation Guide Operation V erification Checking the Output Power N5162A 1 /82A T est Proce dure Te s t S et up 1. Zer o a nd ca libr ate the po we r se ns or to the po wer meter: 2. Connect t he equipment as show n below: 3. Preset t he signal generato r: Press Preset . 4. T urn RF on: Pr ...

  • Agilent Technologies N5161A - page 30

    24 Agilent N5161A/62A/81A/82A/83 A MXG Signal Generators Installation Guide Operation V erification Checking the Output Power With Modula tion 11. Pr eset the si gnal genera tor: Pres s Preset . 12. Select the factor y- supplied wavef orm SINE_TEST_WFM : a. Press Mode > Dual ARB > Select Waveform . b. H ighli ght the SINE _TEST_WFM wavef or m ...

  • Agilent Technologies N5161A - page 31

    25 Agilent N5161A/62A/81A/82A/83 A MXG Signal Generators Installation Guide Operation V erification Checking the Output Power N518 3A T est Proc edur e Te s t S et up 1. Zer o a nd ca libr ate the po we r se ns or to the po wer meter: 2. Connect t he equipment as show n: 3. Preset t he signal generato r: Press Preset . 4. T urn RF on: Press RF On/O ...

  • Agilent Technologies N5161A - page 32

    26 Agilent N5161A/62A/81A/82A/83 A MXG Signal Generators Installation Guide Operation V erification Checking the Output Power 40 GHz (Option 540) -- 7 -- 1 2 ± 2 a. Limit values are due to power meter uncertainty . b. For quest ions around measurem ent uncert ainty , refer to the Data Sheet . T able 3-8 Leveled Output Power Limi ts N5183A Output P ...

  • Agilent Technologies N5161A - page 33

    Agilent N5161A/62A/81A/82A /83A MXG Signal Generators Installation Guide 27 Operation V erification Check ing the O utput Pow er T roubl eshooting Problems with the Output Power Check • V erify th at you are using the appropriate pow e r sensor . • Normall y , po wer sens or calibration factors ar e automatically do wnloaded to the pow er meter ...

  • Agilent Technologies N5161A - page 34

    28 Agilent N5161A/62A/81A /82A/83A MXG Signal Generators Installation Guide Operation V erification Checking the Output Power ...

  • Agilent Technologies N5161A - page 35

    Agilent N5161A/62A/81A/82A /83A MXG Signal Generators Installation Guide 29 4 Regulatory Information Certification Agilent Technologies certifies t hat this product met its published specifications at th e time of shipment from the factor y. A gilent Technolo gies fu r ther cer tif ies that its calibration measurements are traceable to t he United ...

  • Agilent Technologies N5161A - page 36

    30 Agilent N5161A/62A/81A/82A/83 A MXG Signal Generators Installation Guide Regulatory Information Compliance with German Noise Requireme nts Compliance with German Noise Requirements This is to declare that th is instrument is in conformance with the Ger man Regulation on Noise Declaration for Machines (Laermangabe nach der Masc hinenlae rm rerord ...

  • Agilent Technologies N5161A - page 37

    Agilent N5161A/62A/81A/82A /83A MXG Signal Generators Installation Guide 31 Index A AC po wer cord, connecting , 6 symbol , 2 accessories , 12 address, GPIB , 11 Agil ent, c on tacti ng , 14 alcohol, c leaning with , 14 altitude requi rements , 5 Austral ian Comm unicati ons Author ity (C- tick) mark , 2 auto- IP , 10 B brightness adjustment , 9 C ...

  • Agilent Technologies N5161A - page 38

    32 Agilent N5161A/62A/81A/82A/83 A MXG Signal Generators Installation Guide Index sockets, enabling , 11 standby symbol , 2 T temperature requirements , 5 troubleshooting output power , 21 V ventil ation re quirement s , 5 verification, operation , 15 voltage r equirements , 5 VXI- 11, enabling , 11 W warning s , 1 web server , 11 weight, signal ge ...

Fabricant Agilent Technologies Catégorie Stereo System

Les documents que nous recevons du fabricant de l'appareilAgilent Technologies N5161A peuvent être divisés en plusieurs groupes. Ceux-ci sont, entre autres:
- dessins techniques Agilent Technologies
- manuels d’utilisations N5161A
- fiches produit Agilent Technologies
- dépliants
- ou étiquettes-énergie Agilent Technologies N5161A
Tous sont importants, mais les informations les plus importantes du point de vue de l'utilisation de l'appareil se trouvent dans le manuel d’utilisation Agilent Technologies N5161A.

Un groupe de documents appelé manuels d’utilisation est également divisé en types plus spécifiques, tels que: Manuels d’installation Agilent Technologies N5161A, manuels d’entretien, brefs manuels ou manuels de l’utilisateur Agilent Technologies N5161A. Selon vos besoins, vous devriez chercher le document dont vous avez besoin. Sur notre site, vous pouvez voir le manuel le plus populaire d’utilisation du produit Agilent Technologies N5161A.

Manuel d’utilisation complet de l’appareil Agilent Technologies N5161A, quelle devrait-elle être?
Le manuel d’utilisation, également appelé le mode d’emploi, ou tout simplement le manuel, est un document technique destiné à aider à utiliser Agilent Technologies N5161A par les utilisateurs. Des manuels sont généralement écrits par un rédacteur technique, mais dans un langage accessible à tous les utilisateurs Agilent Technologies N5161A.

Le manuel d’utilisation complet Agilent Technologies, devrait inclure plusieurs éléments de base. Certains d'entre eux sont moins importants, tels que: la couverture / page de titre ou pages d'auteur. Cependant, la partie restante, devrait nous fournir des informations importantes du point de vue de l'utilisateur.

1. Introduction et des conseils sur la façon d'utiliser le manuel Agilent Technologies N5161A - Au début de chaque manuel, nous devrions trouver des indices sur la façon d'utiliser le document. Il doit contenir des informations sur l'emplacement de la table des matières Agilent Technologies N5161A, FAQ ou des problèmes les plus fréquents - les points qui sont les plus souvent recherchés par les utilisateurs de chaque manuel
2. Table des matières - index de tous les conseils pour lAgilent Technologies N5161A qui peuvent être trouvés dans le document courant
3. Conseils sur la façon d'utiliser les fonctions de base de l’appareil Agilent Technologies N5161A - qui devraient nous aider dans les premières étapes lors de l'utilisation Agilent Technologies N5161A
4. Troubleshooting - séquence systématique des activités qui nous aideront à diagnostiquer et ensuite résoudre les principaux problèmes de Agilent Technologies N5161A
5. FAQ - questions fréquemment posées
6. Détails du contact Informations sur l'endroit où chercher le contact avec le fabricant / service Agilent Technologies N5161A dans un pays donné, si le problème ne peut être résolu par nous-mêmes.

Avez-vous une question à propos de Agilent Technologies N5161A?

Utiliser le formulaire ci-dessous

Si vous n’avez pas résolu votre problème avec Agilent Technologies N5161A, avec l'aide du manuel que vous avez trouvé, posez une question en utilisant le formulaire ci-dessous. Si un utilisateur a eu un problème similaire avec Agilent Technologies N5161A il est probable qu’il a envie de partager la façon de le résoudre.

Réécrire le texte de l'image