Manuel d’utilisation MaxTech OM-25ME

30 pages 0.19 mb
Télécharger

Aller à la page of 30

Summary
  • MaxTech OM-25ME - page 1

    OXYGEN MONIT OR MODEL OM-25ME (OM-25ME-L) USER’S GUIDE AND OPERA TING INSTRUCTIONS manufactured by: R213M05 Rev . H ® ® R213M05H.qxd 6/26/2006 9:04 AM Page 1 ...

  • MaxTech OM-25ME - page 2

    R213M05H.qxd 6/26/2006 9:04 AM Page 2 ...

  • MaxTech OM-25ME - page 3

    i Manufacturer: Maxtec, Inc European Representative Airox S.A. Parcd’activities Pau-Pyrenees L ’echangeur BP833 64008 P AU Cedex France R213M05H.qxd 6/26/2006 9:04 AM Page i ...

  • MaxTech OM-25ME - page 4

    ii PREF ACE This manual describes the function, operation and maintenance of the Maxtec Model OM-25ME and OM-25ME-L oxygen monitors. Notes are included where differences may occur between the two monitors. As members of Maxtec's MAXO 2  line of oxygen analyzers and monitors, the OM-25ME utilizes the Maxtec MAX-250E oxygen sensor and is engi ...

  • MaxTech OM-25ME - page 5

    iii T o avoid explosion, do not operate the Oxygen monitor in the presence of flammable anesthetics or in an atmosphere of explosive gases. Operating the oxygen monitor in flammable or explosive atmospheres may result in fire or explosion. Never allow an excess length of cable near the patient’ s head or neck, as such could result in strangulatio ...

  • MaxTech OM-25ME - page 6

    iv The OM-25ME-L™ oxygen monitor has been manufactured with a low alarm setting adjustable down to 15% This change is within the guidelines set forth in ASTM S pecification F1462-93. The OM-25ME-L™ is labeled to indicate this feature. Classification: Protection against electric shock: Internally powered equipment. Protection against water: Ordi ...

  • MaxTech OM-25ME - page 7

    v T ABLE OF CONTENTS 1. SYSTEM OVERVIEW ........................................... 1 1.1 Base Unit Description .............................. 1 1.2 Component s Description ........................ 2 1.3 MAX-250E Oxygen Sensor .......................... 6 2. SET -UP PROCEDURE ......................................... 7 2.1 Battery Inst allation ... ...

  • MaxTech OM-25ME - page 8

    1. SYSTEM OVERVIEW 1.1 Base Unit Description The MAXO 2  (Model OM-25ME) provides unparalleled performance and reliability , due to an advanced design that includes the following features and operational benefits. • Fast-responding, oxygen-specific, galvanic sensor that achieves 90% of final value in approximately 15 seconds at room temperatur ...

  • MaxTech OM-25ME - page 9

    2 1.2 Component s Description (please refer to p age 3) LCD Readout 3 1 / 2 -Digit Display- The 3 1 / 2 -digit liquid crystal display (LCD) provides direct readout of oxygen concentrations in the range of 0-100%. It also displays "CAL" when the calibration mode is entered. "%" Sign- In the calibration mode, the "%" sig ...

  • MaxTech OM-25ME - page 10

    LCD Readout Alarm LEDs Keyp ad External Probe A C B D 1 2 3 10 12 4 13 15 16 14 6 17 18 8 OM-25ME 11 9 5 7 3 R213M05H.qxd 6/26/2006 9:04 AM Page 2 ...

  • MaxTech OM-25ME - page 11

    4 Alarm LEDs Low Alarm LED- In a low alarm condition, the red "LOW ALARM" LED will flash each second, accompanied by the alarm beeper . High Alarm LED- In a high alarm condition, the red "HIGH ALARM" LED will flash each second, accompanied by the alarm beeper . Keyp ad ON/OFF Key- (after Calibration) This key is used to turn the ...

  • MaxTech OM-25ME - page 12

    D 13 14 15 16 17 18 ↑ ↑ and ↓ ↓ Keys- The ↑ and ↓ arrow keys are used in conjunction with the CALIBRA TE key to calibrate the unit or in conjunction with the LOW SET and HIGH SET keys to adjust the alarm settings. When used after pressing the CALIBRA TE key , pressing the ↑ or ↓ key will raise or lower the displayed oxygen value in ...

  • MaxTech OM-25ME - page 13

    6 CAUTION: The Maxtec MAX-250E oxygen sensor is a sealed device cont aining a mild acid electrolyte, lead (Pb), and lead acet ate. Lead and lead acetate are hazardous waste constituent s and should be disposed of properly , or returned to Maxtec for proper disposal or recovery . CAUTION: Do not use ethylene oxide for sterilization. Do not immerse t ...

  • MaxTech OM-25ME - page 14

    2. SET -UP PROCEDURE 2.1 Battery Inst allation All MAXO 2  units are powered by two, AA, alkaline batteries (2 x 1.5 V olts) and are shipped without the batteries installed. The battery compartment is accessible from the back side of the unit. T o install the batteries: 1) With the thumb, press down on the center of the battery compartment cover ...

  • MaxTech OM-25ME - page 15

    8 be performed when the geographic elevation at which the product is being used changes by more than 500 feet. The sensor is best calibrated while mounted in the industry standard, 15mm I.D. “T” adapter . As in normal operation, the oxygen sensor responds best when installed in a vertical position with the sensor facing down. In addition, calib ...

  • MaxTech OM-25ME - page 16

    6) Use the ↑ and ↓ arrow keys to adjust the displayed oxygen concentration to the level of the known concentration. Pressing the arrow keys changes the value in .1% increments. If the keys are held down for more than 1 second the display will scroll at a rate of .4% per second. Note: If 10 seconds elap se between key actuations, the system will ...

  • MaxTech OM-25ME - page 17

    10 T o use this function: 1) Place the external probe in room air . 2) Press the LOCKUNLOCK key to unlock the keypad. 3) Press and hold down the CALIBRA TE key . When the "%" sign start s to flash, press the ↓ arrow key to set the calibration value to 20.9%. 4) Release both the CALIBRA TE key and the ↓ key . The unit will automatical ...

  • MaxTech OM-25ME - page 18

    Pressure Effect Readings from the MAXO 2  Monitor are proportional to the partial pressure of oxygen. The p artial pressure of Oxygen (PO 2 ) is equal to the percentage of oxygen (%O 2 ) times the absolute pressure (AP) at which the sample enviroment is measured. (PO 2 =%O 2 x AP). Thus the readings are proportional to the concentration if the p ...

  • MaxTech OM-25ME - page 19

    12 3. OPERA TING INSTRUCTIONS 3.1 Alarm Setting Procedure 3.1.1 Low Alarm Setting T o adjust the low alarm setting: 1) Press the LOCK/UNLOCK key to unlock the keypad. 2) Press the LOW SET key . The "LOW" icon will start to flash. 3) Use the ↑ and ↓ arrow keys to set the low alarm to the desired value. Pressing the arrow keys changes t ...

  • MaxTech OM-25ME - page 20

    The high alarm value cannot be set closer than 1% from the low alarm value due to ASTM specifications. For example, if the low alarm is set at 50%, the system will not accept a high alarm setting less than 51%. Setting the high alarm to 100% turns off or deactivates the high alarm. 4) When the high alarm value is set, press the HIGH SET key again t ...

  • MaxTech OM-25ME - page 21

    14 4. SENSOR REMOV AL AND REPLACEMENT The OM-25ME is shipped with a new MAX-250E oxygen sensor installed. Although the sensor has a very long expected life, eventually the sensor will require replacement. Removing or installing a sensor , when necessary , is a very simple procedure. T o remove and install a new sensor: 1) Grasp the sensor in one ha ...

  • MaxTech OM-25ME - page 22

    5. PROBLEM SOL VING • If the "LOW BA T" icon is displayed on the LCD readout at any time, the batteries should be replaced as quickly as possible. • When the unit is in the power on mode and the LCD displays "000%," the sensor is not connected properly . The "A TT ACH SENSOR" icon will also appear on the display . ...

  • MaxTech OM-25ME - page 23

    16 7. SPECIFICA TIONS 7.1 Base Unit Specifications Measurement Range: 0.0-100% Resolution: 0.1% Accuracy and Linearity: ±1% of full scale at constant temperature, R.H. and pressure when calibrated at full scale. T otal Accuracy: ±3% Actual Oxygen Level over full operating temperature range. Response T ime: 90% of final value in approximately 15 s ...

  • MaxTech OM-25ME - page 24

    8. APPLICA TIONS 8.1 Exposure to Anesthetic Gases Because of the unique chemistry of the oxygen sensors provided with the MAXO 2  Monitor , there are no significant effects when exposed to commonly used anesthetic gases, however , the monitor is not designed for exposure to flammable gas mixtures. (See W ARNING page i) 8.2 Calibration T echnique ...

  • MaxTech OM-25ME - page 25

    18 8.3 Calibration Errors The MAXO 2  Monitors have a self test feature built into the software to detect faulty calibrations. During calibration, if the signal from the oxygen sensor is outside the limit s stored within the instrument’s memory , a flashing “CAL Er” is displayed. The error code is displayed to indicate that either the sens ...

  • MaxTech OM-25ME - page 26

    Additionally , gas streams of high humidity may tend to condense on the sensor . Condensation on the sensor may eventually affect performance. For this reason, it is recommended that the sensor be mounted in a vertical position, facing downward to prevent condensate from flowing onto the sensing surface. 19 R213M05H.qxd 6/26/2006 9:04 AM Page 18 ...

  • MaxTech OM-25ME - page 27

    20 9. SP ARE P ARTS AND ACCESSORIES Part Number Item R12503-002 MAX-250E Sensor R212P91 Battery Cover R103P16 Monitor/Analyzer External Cable R212P17 Keypad R212P10 LCD Display R212P30-001 PCBA Board Accessories R106P15 Concentrator Adapter for Sensor Probe R205P86 Monitor/Analyzer W all Mount Bracket R206P75 Monitor/Analyzer Pole Mount Clamp R213P ...

  • MaxTech OM-25ME - page 28

    10. W ARRANTY The MAXO 2  Monitor is designed for medical oxygen delivery equipment and systems. Under normal operating conditions, Maxtec warrants the MAXO 2  Monitor to be free from defects of workmanship or materials for a period of two (2) years from the date of shipment from Maxtec, provided that the unit is properly operated and maintai ...

  • MaxTech OM-25ME - page 29

    R213M05H.qxd 6/26/2006 9:04 AM Page 21 ...

  • MaxTech OM-25ME - page 30

    Maxtec, Inc. 6526 South Cottonwood S treet Salt Lake City , UT 84107 General T el: 801-266-5300 T oll Free Dial: 800-748-5355 F AX: 801-270-5590 Home Page: http://www .maxtecinc.com email: sales@maxtecinc.com R213M05H.qxd 6/26/2006 9:04 AM Page 22 ...

Fabricant MaxTech Catégorie Oxygen Equipment

Les documents que nous recevons du fabricant de l'appareilMaxTech OM-25ME peuvent être divisés en plusieurs groupes. Ceux-ci sont, entre autres:
- dessins techniques MaxTech
- manuels d’utilisations OM-25ME
- fiches produit MaxTech
- dépliants
- ou étiquettes-énergie MaxTech OM-25ME
Tous sont importants, mais les informations les plus importantes du point de vue de l'utilisation de l'appareil se trouvent dans le manuel d’utilisation MaxTech OM-25ME.

Un groupe de documents appelé manuels d’utilisation est également divisé en types plus spécifiques, tels que: Manuels d’installation MaxTech OM-25ME, manuels d’entretien, brefs manuels ou manuels de l’utilisateur MaxTech OM-25ME. Selon vos besoins, vous devriez chercher le document dont vous avez besoin. Sur notre site, vous pouvez voir le manuel le plus populaire d’utilisation du produit MaxTech OM-25ME.

Manuels d’utilsiation similaires

Manuel d’utilisation complet de l’appareil MaxTech OM-25ME, quelle devrait-elle être?
Le manuel d’utilisation, également appelé le mode d’emploi, ou tout simplement le manuel, est un document technique destiné à aider à utiliser MaxTech OM-25ME par les utilisateurs. Des manuels sont généralement écrits par un rédacteur technique, mais dans un langage accessible à tous les utilisateurs MaxTech OM-25ME.

Le manuel d’utilisation complet MaxTech, devrait inclure plusieurs éléments de base. Certains d'entre eux sont moins importants, tels que: la couverture / page de titre ou pages d'auteur. Cependant, la partie restante, devrait nous fournir des informations importantes du point de vue de l'utilisateur.

1. Introduction et des conseils sur la façon d'utiliser le manuel MaxTech OM-25ME - Au début de chaque manuel, nous devrions trouver des indices sur la façon d'utiliser le document. Il doit contenir des informations sur l'emplacement de la table des matières MaxTech OM-25ME, FAQ ou des problèmes les plus fréquents - les points qui sont les plus souvent recherchés par les utilisateurs de chaque manuel
2. Table des matières - index de tous les conseils pour lMaxTech OM-25ME qui peuvent être trouvés dans le document courant
3. Conseils sur la façon d'utiliser les fonctions de base de l’appareil MaxTech OM-25ME - qui devraient nous aider dans les premières étapes lors de l'utilisation MaxTech OM-25ME
4. Troubleshooting - séquence systématique des activités qui nous aideront à diagnostiquer et ensuite résoudre les principaux problèmes de MaxTech OM-25ME
5. FAQ - questions fréquemment posées
6. Détails du contact Informations sur l'endroit où chercher le contact avec le fabricant / service MaxTech OM-25ME dans un pays donné, si le problème ne peut être résolu par nous-mêmes.

Avez-vous une question à propos de MaxTech OM-25ME?

Utiliser le formulaire ci-dessous

Si vous n’avez pas résolu votre problème avec MaxTech OM-25ME, avec l'aide du manuel que vous avez trouvé, posez une question en utilisant le formulaire ci-dessous. Si un utilisateur a eu un problème similaire avec MaxTech OM-25ME il est probable qu’il a envie de partager la façon de le résoudre.

Réécrire le texte de l'image

Commentaires (0)