Manuel d’utilisation Invacare RPA600-1E

40 pages 1.33 mb
Télécharger

Aller à la page of 40

Summary
  • Invacare RPA600-1E - page 1

    Owner’s Operator and Maintenance Manual DEALER: This manual MUST be given to the user of the patient lift. USER: BEFORE using this patient lift, read this manual and save for future reference. For more information regarding Invacare products, parts, and services, please visit www.invacare.com Electric Por table P atient Lift - RP A600-1E ...

  • Invacare RPA600-1E - page 2

    SYMBOL LEGEND Electric Portable Pati ent Lift - RPA600-1E 2 Part No 1078988  WARNING DO NOT USE THIS PRODUCT OR ANY AVAILABLE OPTIONAL EQUIPMENT WITHOUT FIRST COMPLETELY READING AND UNDERSTANDING THESE IN STRUCTIONS AND ANY ADDITIONAL INSTRUCTIONAL MA TERIAL SUCH AS OWNER’ S MANUALS, SERVICE MANUAL S OR INSTRUCTION SHEETS SUPPLIED WITH THIS PR ...

  • Invacare RPA600-1E - page 3

    TABLE OF CONTENTS Part No 1078988 3 Electric Portable Patient Lift - RPA6001E TABLE OF CONTENTS SYMBOL LEGEND .............................................................................. 2 SPECIAL NOTES ................................................................................ 5 LABEL LOCATION ............................................... ...

  • Invacare RPA600-1E - page 4

    TABLE OF CONTENTS Electric Portable Pati ent Lift - RPA6001E 4 Part No 1078988 TABLE OF CONTENTS SECTION 4—LIFTING T HE PATIENT ................................................. 19 Preparing the Lift for Us e .................................................................... .................. ............... ..................19 Positioning th ...

  • Invacare RPA600-1E - page 5

    SPECIAL NOTES Part No 1078988 5 Electric Portable Patient Lift - RPA6001E SPECIAL NOTES Signal  words  are  use d  in  this  manual  and  apply  to  hazards  or  unsafe  practices  which  could  result  in  personal  injury  or  property  damage.  Refer  to  the  table ? ...

  • Invacare RPA600-1E - page 6

    LABEL LOCATION Electric Portable Pati ent Lift - RPA600-1E 6 Part No 1078988 LABEL LOCATION W ARNING Mast piv ot MUST be tight to ensure saf e use of your patient lift. Bolt MUST be check ed at least e v er y six (6) months in conjunction with periodic maintenance. See Owner's Manual. P/N 1130200 Rev . B W ARNING W ARNING ...

  • Invacare RPA600-1E - page 7

    PRODUCT PARAMETERS Part No 1078988 7 Electric Portable Patient Lift - RPA6001E PRODUCT PARAMETERS Patient Lift RPA600-1E *NOTE:  Va r i e s  depending  upon  load  and  str oke. Reliant Scale RLS6 204.5 kg Electric Low- Profile 204.5 Electric Standard 272.7 kg Electric Heavy-duty Height at Sling Hook-u p - MAX.: 188 cm 193 cm 177. ...

  • Invacare RPA600-1E - page 8

    SECTION 1—GENERAL GUIDELINES Electric Portable Pati ent Lift - RPA6001E 8 Part No 1078988 SECTION 1—GENERAL GUIDELINES  WARNING SECTION 1 - GENERAL GUID ELINES contains important information for the safe operation and use of this product. Check  all  parts  for  shipping  damage  before  using .  In  case  of ? ...

  • Invacare RPA600-1E - page 9

    SECTION 1—GENERAL GUIDELINES Part No 1078988 9 Electric Portable Patient Lift - RPA6001E Stand  Assist  Slings:  Before  lifting  the  patient,  make  sure  the  bottom  edge  of  the  stand  assist  sling  is  positioned  on  the  low er  back  of  the  patient  and ? ...

  • Invacare RPA600-1E - page 10

    SECTION 1—GENERAL GUIDELINES Electric Portable Pati ent Lift - RPA6001E 10 Part No 1078988 Before  lifting  a  patient  from  a  stationary  object  (wheelchair ,  commode  or  bed),  slightly  raise  the  patient  off  the  stationary  object  and  check  that  all  sling ? ...

  • Invacare RPA600-1E - page 11

    SECTION 2—ASSEMBLY Part No 1078988 11 Electric Portable Patient Lift - RPA600-1E SECTION 2—ASSEMBLY Assembling the Patient Lift Assembling the Mast to the Base NOTE:  The  mast  assembly  may  be  re mo v e d  from  the  ba se  for  storag e  or  transporting.  The  mast  assembly  MUST  be ...

  • Invacare RPA600-1E - page 12

    SECTION 2—ASSEMBLY Electric Portable Pati ent Lift - RPA600-1E 12 Part No 1078988 Assembling the Electric Actuator to the Boom NOTE:  For  this  pr ocedure,  re f e r  to  FIGURE 2.2. 1. Remove  the  shoulder  bolt,  was h er  and  nut  from  the  mounting  bracket  on  the  boom  assemb ...

  • Invacare RPA600-1E - page 13

    SECTION 2—ASSEMBLY Part No 1078988 13 Electric Portable Patient Lift - RPA600-1E Installing the Shifter Handle NOTE:  For  this  pr ocedure,  re f e r  to  FIGURE 2.3. 1. Line ‐ up  the  shifter  handle  threads  with  threaded  opening  in  the  base.  2. T urn  the  shifter  handle ? ...

  • Invacare RPA600-1E - page 14

    SECTION 3—OPERATION Electric Portable Pati ent Lift - RPA600-1E 14 Part No 1078988 SECTION 3—OPERATION Introduction  WARNING DO NOT attempt to transfer a patient without approval of the patient’s physician, nurse, or medical assistant. Thoroughly read the instruc tions in this owner’s manual, observe a trained team of ex perts performing ...

  • Invacare RPA600-1E - page 15

    SECTION 3—OPERATION Part No 1078988 15 Electric Portable Patient Lift - RPA600-1E FIGURE 3.1 Operating the Patient Lift DETAIL “B” - HAN D CONTROL DE TAILS DETAIL “C” - E MERGENCY STOP Down Button Up Button Hand Control Emergency Button Push in to stop , tur n clockwise to reset. Step Here to Lock Locking Lever DETAIL “A” - LOCKING/UN ...

  • Invacare RPA600-1E - page 16

    SECTION 3—OPERATION Electric Portable Pati ent Lift - RPA600-1E 16 Part No 1078988 Closing/Opening the Legs NOTE:  For  this  proced ure,  re fe r  to  FIGURE 3.2.  WARNING If the shifter handle is NOT positioned completely in to its mounting slot, DO NOT use the patient lift until the shifter handle is properly seate d and t ...

  • Invacare RPA600-1E - page 17

    SECTION 3—OPERATION Part No 1078988 17 Electric Portable Patient Lift - RPA600-1E Activating a Mechanical Emergency Release Primary Emergency Release NOTE:  For  this  proced ure,  re fe r  to  FIGURE 3.3. NOTE:  This  procedur e  will  bri ng  the  boom  down  or  lift  it  up  if  the  ...

  • Invacare RPA600-1E - page 18

    SECTION 3—OPERATION Electric Portable Pati ent Lift - RPA600-1E 18 Part No 1078988 Charging the Battery NOTE:  For  this  pr ocedure,  re f e r  to  FIGURE 3.5. NOTE:  Inva car e  reco m m en d s  the  battery  be  rec h a rge d  daily  to  pr olong  battery  life. NOTE:  An  audible  al ...

  • Invacare RPA600-1E - page 19

    SECTION 4—LIFTING THE PATIENT Part No 1078988 19 Electric Portable Patient Lift - RPA6001E SECTION 4—LIFTING THE PATIENT Preparing the Lift for Use NOTE:  Although  Invacare  reco m m e nd s  that  two  assistants  be  used  fo r  all  lifting  pr eparation,  transferring  from  and  transferring ...

  • Invacare RPA600-1E - page 20

    SECTION 4—LIFTING THE PATIENT Electric Portable Pati ent Lift - RPA6001E 20 Part No 1078988 Attaching a Sling to the Swivel Bar  WARNING The maximum weight capacity of the RPA600-1E patient lift is 600 lbs. DO NOT exceed the maximum weight capa city of the patient lift ot herwise, injury or damage may occur. Before using the slin g, read and u ...

  • Invacare RPA600-1E - page 21

    SECTION 4—LIFTING THE PATIENT Part No 1078988 21 Electric Portable Patient Lift - RPA6001E NOTE:  For  this  pr ocedure,  re f e r  to  FIGURE 4.2. NOTE:  When  the  patient  is  lifted  from  the  bed  or  the  floor  (with  the  patient’s  head  support ed  by  the  sl ...

  • Invacare RPA600-1E - page 22

    SECTION 5—TRANSFERRING THE PATIENT Electric Portable Pati ent Lift - RPA6001E 22 Part No 1078988 SECTION 5—TRANSFERRING THE PATIENT Introduction  WARNING DO NOT attempt any transfer of a patient without appr oval of the patient's physician, nurse, or medical assistant. DO NOT move the patient if the sling is not properly co nnected to t ...

  • Invacare RPA600-1E - page 23

    SECTION 5—TRANSFERRING THE PATIENT Part No 1078988 23 Electric Portable Patient Lift - RPA6001E Transferring to a Commode Chair NOTE:  For  this  proced ure,  re fe r  to  FIGURE 5.1. 1. Lift  the  patient  from  the  bed.  Refer  to  Lifting  the  Pat ie nt  on  page 19. 2. Press  the  ...

  • Invacare RPA600-1E - page 24

    SECTION 5—TRANSFERRING THE PATIENT Electric Portable Pati ent Lift - RPA6001E 24 Part No 1078988 Transferring to a Standard Commode NOTE:  The  Invacare  patient  lift  is  NOT  intended  as  a  transport  device.  Moving  a  person  suspended  in  a  sling  over  any  distance  is ...

  • Invacare RPA600-1E - page 25

    SECTION 5—TRANSFERRING THE PATIENT Part No 1078988 25 Electric Portable Patient Lift - RPA6001E 5. Use  the  straps  or  handles  on  the  side  and  the  back  of  the  sling  to  guide  the  patient’ s  hips  as  far  back  as  possible  into  the  seat  for  ...

  • Invacare RPA600-1E - page 26

    SECTION 6—TROUBLESHOOTING Electric Portable Pati ent Lift - RPA6001E 26 Part No 1078988 SECTION 6—TROUBLESHOOTING NOTE:  If  pr oblems  are  not  rem e d i ed  by  the  suggested  means,  please  contact  your  dealer  or  Invacare . SYMPTOMS FAULTS SOLUTION Patient Lift feels loose. Mast/Base joint ...

  • Invacare RPA600-1E - page 27

    SECTION 7—MAINTENANCE Part No 1078988 27 Electric Portable Patient Lift - RPA600-1E SECTION 7— MAINTENANCE Maintenance Safety Inspection Checklist *NOTE:  For  individual  home  use,  a  full  inspection  is  re qu i re d  prior  to  eac h  new  user . **NOTE:  Regular  cleaning  will  re ve ...

  • Invacare RPA600-1E - page 28

    SECTION 7—MAINTENANCE Electric Portable Pati ent Lift - RPA600-1E 28 Part No 1078988 The  Inv acare  Pa t i e n t  Lift  is  designed  to  provide  a  maximum  of  safe,  efficient  and  satisfactory  service  with  minimum  care  and  maintenance. All  parts  of  the  Inv ac ...

  • Invacare RPA600-1E - page 29

    SECTION 7—MAINTENANCE Part No 1078988 29 Electric Portable Patient Lift - RPA600-1E Detecting Wear and Damage It  is  important  to  inspect  all  stressed  parts,  such  as  slings,  spreader  bar  and  any  piv ot  for  slings  for  signs  of  cracking,  fraying,  deformation ...

  • Invacare RPA600-1E - page 30

    SECTION 7—MAINTENANCE Electric Portable Pati ent Lift - RPA600-1E 30 Part No 1078988 Checking and Tightening Mast Pivot Bolt NOTE:  For  this  pr ocedure,  ref er  to  FIGURE 7.3. 1. Lift  up  the  back  of  the  rubber  boot  and  slide  it  off  the  mast  along  the  boom. 2. Ch ...

  • Invacare RPA600-1E - page 31

    SECTION 7—MAINTENANCE Part No 1078988 31 Electric Portable Patient Lift - RPA600-1E FIGURE 7.4 Replacing the Swivel Bar Boom Boom Mounting Br acket Wa s h e r Shoulder Bolt Locknut Swivel Bar Swivel Bar Pin Nylon W asher s Pinch Guard ...

  • Invacare RPA600-1E - page 32

    SECTION 8—ACCESSORIES Electric Portable Pati ent Lift - RPA6001E 32 Part No 1078988 SECTION 8—ACCESSORIES Reliant Scale RLS6 The  Reliant  Scale  is  a  compact  precision  scale  system  designed  specifically  for  the  Inv acare  Pa t i en t  Lift  System. Removing the Swivel Bar NOTE:  For ...

  • Invacare RPA600-1E - page 33

    SECTION 8—ACCESSORIES Part No 1078988 33 Electric Portable Patient Lift - RPA6001E Installing the Reliant Scale NOTE:  For  this  proced ure,  re fe r  to  FIGURE 8.2.  WARNING Patient and sling MUST be removed from the lift during ALL installation proce- dures. 1. P osition  the  load  cell  assembly  of  t ...

  • Invacare RPA600-1E - page 34

    SECTION 8—ACCESSORIES Electric Portable Pati ent Lift - RPA6001E 34 Part No 1078988 Operating the Scale Keypad Functions NOTE:  For  this  proced ure,  re fe r  to  FIGURE 8.3. FIGURE 8.3 Operating the Scale - Keypad Functions KEY INDICATOR DISPLAYED INDICATOR LOCATION DEFINITION ON OFF OFF Center of Display Window Pressing this ...

  • Invacare RPA600-1E - page 35

    SECTION 8—ACCESSORIES Part No 1078988 35 Electric Portable Patient Lift - RPA6001E Weighing the Patient  WARNING The weight capacity is limite d to the lowest rated capac ity of any one of the compo- nents in use (e.g. Patient Li ft, Sling or Scale ). The patient's weight MUST not exceed the lowest rated ca pacity of any component. 1. Att ...

  • Invacare RPA600-1E - page 36

    SECTION 8—ACCESSORIES Electric Portable Pati ent Lift - RPA6001E 36 Part No 1078988 NOTE:  Stable  being  defined  as  the  weight  fluctuating  two  tenths  of  a  poun d.  For  example,  a  patient  weighing  one  hundred  pounds,  the  scale  will  fluctuate  bet ween ? ...

  • Invacare RPA600-1E - page 37

    SECTION 8—ACCESSORIES Part No 1078988 37 Electric Portable Patient Lift - RPA6001E 3. Press  the  CAL  button  located  on  the  PC  board.  The  CAL  switch  is  not  labeled  but  is  the  only  button  on  the  PC  board  located  in  the  lower  right  corn ...

  • Invacare RPA600-1E - page 38

    NOTES Electric Portable Pati ent Lift - RPA6001E 38 Part No 1078988 NOTES ...

  • Invacare RPA600-1E - page 39

    LIMITED WARRANTY Part No 1078988 39 Electric Portable Patient Lift - RPA600-1E LIMITED WARRANTY Available At: United  Kingdom: Inv acare  Ltd,  South  Road,  Bridgend  Industrial  Estate,  UK ‐ Bridgend  CF31  3PY Te l :  (44)  (0)1656  664  321,  Fax:  (44)  (0)1656  667  532 UK@inv acare. ...

  • Invacare RPA600-1E - page 40

    Invacare Corporation www.invacare.com USA One Invacare Way Elyria, Ohio USA 44036-2125 800-333-6900 Canada 570 Matheson Blvd. E Unit 8 Mississauga Ontario L4Z 4G4 Canada 800-668-5324 All rights reserved. Tradem arks are identified by ™ and ®. Al l trademarks are owned by or licens ed to Invacare Corporation unless otherwise noted. ©2008 Invacar ...

Fabricant Invacare Catégorie Fitness Equipment

Les documents que nous recevons du fabricant de l'appareilInvacare RPA600-1E peuvent être divisés en plusieurs groupes. Ceux-ci sont, entre autres:
- dessins techniques Invacare
- manuels d’utilisations RPA600-1E
- fiches produit Invacare
- dépliants
- ou étiquettes-énergie Invacare RPA600-1E
Tous sont importants, mais les informations les plus importantes du point de vue de l'utilisation de l'appareil se trouvent dans le manuel d’utilisation Invacare RPA600-1E.

Un groupe de documents appelé manuels d’utilisation est également divisé en types plus spécifiques, tels que: Manuels d’installation Invacare RPA600-1E, manuels d’entretien, brefs manuels ou manuels de l’utilisateur Invacare RPA600-1E. Selon vos besoins, vous devriez chercher le document dont vous avez besoin. Sur notre site, vous pouvez voir le manuel le plus populaire d’utilisation du produit Invacare RPA600-1E.

Manuel d’utilisation complet de l’appareil Invacare RPA600-1E, quelle devrait-elle être?
Le manuel d’utilisation, également appelé le mode d’emploi, ou tout simplement le manuel, est un document technique destiné à aider à utiliser Invacare RPA600-1E par les utilisateurs. Des manuels sont généralement écrits par un rédacteur technique, mais dans un langage accessible à tous les utilisateurs Invacare RPA600-1E.

Le manuel d’utilisation complet Invacare, devrait inclure plusieurs éléments de base. Certains d'entre eux sont moins importants, tels que: la couverture / page de titre ou pages d'auteur. Cependant, la partie restante, devrait nous fournir des informations importantes du point de vue de l'utilisateur.

1. Introduction et des conseils sur la façon d'utiliser le manuel Invacare RPA600-1E - Au début de chaque manuel, nous devrions trouver des indices sur la façon d'utiliser le document. Il doit contenir des informations sur l'emplacement de la table des matières Invacare RPA600-1E, FAQ ou des problèmes les plus fréquents - les points qui sont les plus souvent recherchés par les utilisateurs de chaque manuel
2. Table des matières - index de tous les conseils pour lInvacare RPA600-1E qui peuvent être trouvés dans le document courant
3. Conseils sur la façon d'utiliser les fonctions de base de l’appareil Invacare RPA600-1E - qui devraient nous aider dans les premières étapes lors de l'utilisation Invacare RPA600-1E
4. Troubleshooting - séquence systématique des activités qui nous aideront à diagnostiquer et ensuite résoudre les principaux problèmes de Invacare RPA600-1E
5. FAQ - questions fréquemment posées
6. Détails du contact Informations sur l'endroit où chercher le contact avec le fabricant / service Invacare RPA600-1E dans un pays donné, si le problème ne peut être résolu par nous-mêmes.

Avez-vous une question à propos de Invacare RPA600-1E?

Utiliser le formulaire ci-dessous

Si vous n’avez pas résolu votre problème avec Invacare RPA600-1E, avec l'aide du manuel que vous avez trouvé, posez une question en utilisant le formulaire ci-dessous. Si un utilisateur a eu un problème similaire avec Invacare RPA600-1E il est probable qu’il a envie de partager la façon de le résoudre.

Réécrire le texte de l'image

Commentaires (0)