    Model 228 Submersion/Inser tion T oroidal Sensor Instruction Sheet PN 51A-228/rev .P December 2010 SPECIFICA TIONS - SENSORS MODEL Wetted Materials T emperature Pressure CRN pressure* 228-02 PEEK (glass-filled) 248°F (120°C) 295 psig (2135 kPa abs) 220 psig (1618 kPa abs) 228-03 PEEK (glass-filled) 392°F (200°C) 295 psig (2135 kPa abs) 220 psig ...

  • Emerson 228 - page 2

    MODEL 228 INST ALLA TION INST ALLA TION Keep at least 1 in. (2.5 cm) between the sensor and pipe walls. If the clearance is too small, calibrate the sensor in place. Ensure that the sensor is completely submerged in liquid. Mounting the sensor in a vertical pipe run with the flow from bottom to top is best. If the sensor must be mounted in a horizo ...

  • Emerson 228 - page 3

    3 MODEL 228 WIRING FIGURE 6. Wiring 228-54 and 228-56 sensors to Model 54eC analyzer FIGURE 5. Wiring 228-54 and 228-56 sensors to Model 1056 and 56 analyzers ...

  • Emerson 228 - page 4

    MODEL 228 WIRING FIGURE 7. Wiring 228-54 and 228-56 sensors to Model Xmt-T p anel mount transmitter FIGURE 8. Wiring 228-54 and 228-56 sensors to Model Xmt-T pipe/wall mount transmitter 4 ...

  • Emerson 228 - page 5

    5 MODEL 228 WIRING FIGURE 9. Wiring 228-54 and 228-56 sensors to Model 1066 transmitter FIGURE 10. Wiring 228-54 and 228-56 sensors to Model 5081-T transmitter RCV B RCV A RSHLD DRV B 228-54 RTN SENSE RTD IN SHLD CLEAR WHITE GREEN BLACK GREEN BLACK WHITE DRV A DSHLD TB1 TB2 RCV B RCV A RSHLD DRV B 228-56 RTN SENSE RTD IN SHLD CLEAR WHITE GREEN BLAC ...

  • Emerson 228 - page 6

    MODEL 228 WIRING FIGURE1 1. Wiring sensors through a remote junction box Wire sensors point to point. For wiring at the analyzer end, refer to the appropriate analyzer wiring diagram. For interconnecting cable 23294-00, use the 2265-54 wiring diagram. For interconnecting cable 23294-04 and 23294-05, use the 226-56 wiring diagram. FIGURE 12. Remote ...

  • Emerson 228 - page 7

    7 MODEL 228 USE OF INSERTION - RETRACTION ASSEMBLIES INSERTION - RETRACTION ASSEMBLIES MANUAL RETRACTION ASSEMBL Y (PN 2331 1-01) INST ALLA TION 1. Loosen the collet nut and retract the sensor tube into the retraction chamber . See Figure 14. 2. Loosen the union nut and separate the retraction chamber from the assembly . 3. Install the retraction c ...

  • Emerson 228 - page 8

    MODEL 228 USE OF INSERTION - RETRACTION ASSEMBLIES FIGURE 14. Manual Retraction Assembly MILLIMETER INCH RETRACTION 1. Make certain that the system pressure is less than 35 psig (342 kPa abs). Push in on the sensor using the top of the junction box. Slowly loosen the collet nut. 2. When the collet nut is loose enough, slowly ease the sensor back so ...

  • Emerson 228 - page 9

    9 MODEL 228 USE OF INSERTION - RETRACTION ASSEMBLIES INST ALLA TION 1. Thread the sensor cable through the tube into the junction box. Screw the sensor into the tube. Hand tighten the sensor an additional half turn once the gas- ket is seated. See Figure 15. 2. T erminate the sensor wiring in the junction box. See Figure 13 for wiring details. 3. U ...

  • Emerson 228 - page 10

    MODEL 228 USE OF INSERTION - RETRACTION ASSEMBLIES 1. Retract the sensor into the retraction chamber and fully close the ball valve. 2. Drain the retraction chamber content s using the 1/8 in. flush ports. 3. (Mechanical) Mark the location of the nut housing cap and retraction collar on the sensor tube. Remove both socket head cup screws from the n ...

  • Emerson 228 - page 11

    11 MODEL 228 USE OF INSERTION - RETRACTION ASSEMBLIES FIGURE 16. Mechanical Retraction Assembly Replacement Part s ...

  • Emerson 228 - page 12

    FIGURE 17. Manual Retraction Assembly Replacement Part s MODEL 228 USE OF INSERTION - RETRACTION ASSEMBLIES 12 ...

  • Emerson 228 - page 13

    13 MODEL 228 CALIBRA TION CALIBRA TION The nominal cell constant of the Model 228 sensor is 2.7/cm. The error in the cell constant is about ± 10%, so con- ductivity readings made using the nominal cell constant will have an error of at least ± 10%. W all ef fects, discussed below , will likely make the error greater . For higher accuracy , the se ...

  • Emerson 228 - page 14

    MAINTENANCE Generally , the only maintenance required is to keep the opening of the sensor clear of deposits. Cleaning frequency is best determined by experience. MODEL 228 MAINTENANCE W ARNING Retraction chamber contents may be under pressure. CAUTION Be sure sensor has been cleaned of process liquid before handling. TROUBLESHOOTING PROBLEM PROBAB ...

  • Emerson 228 - page 15

    15 FIGURE 19. Resist ance check. Disconnect leads from analyzer before measuring resist ances. T ABLE 1. T emperature Resist ance 10 ° C 103.9 Ω 20 ° C 107.8 Ω 25 ° C 109.7 Ω 30 ° C 1 1 1.7 Ω 40 ° C 1 15.5 Ω 50 ° C 1 19.4 Ω ...

  • Emerson 228 - page 16

    Credit Cards for U.S. Purchases Only . The right people, the right answers, right now . ON-LINE ORDERING NOW A V AILABLE ON OUR WEB SITE http://www .raihome.com Emerson Process Management 2400 Barranca Parkway Irvine, CA 92606 USA T el: (949) 757-8500 Fax: (949) 474-7250 http://www .raihome.com © Rosemount Analytical Inc. 2010 8 ...

Fabricant Emerson Catégorie Stud Sensor

Les documents que nous recevons du fabricant de l'appareilEmerson 228 peuvent être divisés en plusieurs groupes. Ceux-ci sont, entre autres:
- dessins techniques Emerson
- manuels d’utilisations 228
- fiches produit Emerson
- dépliants
- ou étiquettes-énergie Emerson 228
Tous sont importants, mais les informations les plus importantes du point de vue de l'utilisation de l'appareil se trouvent dans le manuel d’utilisation Emerson 228.

Un groupe de documents appelé manuels d’utilisation est également divisé en types plus spécifiques, tels que: Manuels d’installation Emerson 228, manuels d’entretien, brefs manuels ou manuels de l’utilisateur Emerson 228. Selon vos besoins, vous devriez chercher le document dont vous avez besoin. Sur notre site, vous pouvez voir le manuel le plus populaire d’utilisation du produit Emerson 228.

Manuels d’utilsiation similaires

Manuel d’utilisation complet de l’appareil Emerson 228, quelle devrait-elle être?
Le manuel d’utilisation, également appelé le mode d’emploi, ou tout simplement le manuel, est un document technique destiné à aider à utiliser Emerson 228 par les utilisateurs. Des manuels sont généralement écrits par un rédacteur technique, mais dans un langage accessible à tous les utilisateurs Emerson 228.

Le manuel d’utilisation complet Emerson, devrait inclure plusieurs éléments de base. Certains d'entre eux sont moins importants, tels que: la couverture / page de titre ou pages d'auteur. Cependant, la partie restante, devrait nous fournir des informations importantes du point de vue de l'utilisateur.

1. Introduction et des conseils sur la façon d'utiliser le manuel Emerson 228 - Au début de chaque manuel, nous devrions trouver des indices sur la façon d'utiliser le document. Il doit contenir des informations sur l'emplacement de la table des matières Emerson 228, FAQ ou des problèmes les plus fréquents - les points qui sont les plus souvent recherchés par les utilisateurs de chaque manuel
2. Table des matières - index de tous les conseils pour lEmerson 228 qui peuvent être trouvés dans le document courant
3. Conseils sur la façon d'utiliser les fonctions de base de l’appareil Emerson 228 - qui devraient nous aider dans les premières étapes lors de l'utilisation Emerson 228
4. Troubleshooting - séquence systématique des activités qui nous aideront à diagnostiquer et ensuite résoudre les principaux problèmes de Emerson 228
5. FAQ - questions fréquemment posées
6. Détails du contact Informations sur l'endroit où chercher le contact avec le fabricant / service Emerson 228 dans un pays donné, si le problème ne peut être résolu par nous-mêmes.

Avez-vous une question à propos de Emerson 228?

Utiliser le formulaire ci-dessous

Si vous n’avez pas résolu votre problème avec Emerson 228, avec l'aide du manuel que vous avez trouvé, posez une question en utilisant le formulaire ci-dessous. Si un utilisateur a eu un problème similaire avec Emerson 228 il est probable qu’il a envie de partager la façon de le résoudre.

Réécrire le texte de l'image

Commentaires (0)