    Model 399VP Combina tion pH/ORP Sensor Instruction Manual PN 51-399VP/rev .B January 2005 ...

  • Emerson 399VP - page 2

    CAUTION SENSOR/PROCESS APPLICA TION COMP A TIBILITY The wetted sensor materials may not be comp atible with process composition and operating conditions. Application comp atibility is entirely the responsibili- ty of the user . DANGER HAZARDOUS AREA INST ALLA TION Installations near flammable liquids or in haz- ardous area locations must be careful ...

  • Emerson 399VP - page 3

    MODEL 399VP pH/ORP T ABLE OF CONTENTS MODEL 399VP pH/ORP SENSOR T ABLE OF CONTENTS SECTION TITLE P AGE 1.0 DESCRIPTION AND SPECIFICA TIONS .......................................................... 1 1.1 Features and Applications ................................................................................. 1 1.2 Physical S pecifications ....... ...

  • Emerson 399VP - page 4

    MODEL 399VP pH/ORP T ABLE OF CONTENTS MODEL 399VP pH/ORP SENSOR LIST OF FIGURES Figure No. Title Page 2-1 Sensor Dimensions ........................................................................................... 4 2-2 Model 399VP Shown in V arious Flow-Through Installations ............................. 5 2-3 Model 399VP Shown in Low Flow Cell ...

  • Emerson 399VP - page 5

    MODEL 399VP pH/ORP SENSOR SECTION 1.0 DESCRIPTION AND SPECIFICA TIONS SECTION 1.0. DESCRIPTION AND SPECIFICA TIONS 1.1 FEA TURES AND APPLICA TIONS 1.2 PHYSICAL SPECIFICA TIONS 1.3 ORDERING INFORMA TION 1 1.1 FEA TURES AND APPLICA TIONS The Rosemount Analytical Model 399VP Sensor meas- ures the pH or the Oxidation Reduction Potential (ORP) of aqueou ...

  • Emerson 399VP - page 6

    MODEL 399VP pH/ORP SENSOR SECTION 1.0 DESCRIPTION AND SPECIFICA TIONS 1.3 ORDERING INFORMA TION The Model 399VP pH/ORP Sensor is available with custom glass electrodes housed in a molded Tefzel body with 1 in. MNPT threads suit able for insertion, submersion or flow through installation. The sensor includes a general purpose hemi bulb pH electrode ...

  • Emerson 399VP - page 7

    MODEL 399VP pH/ORP SENSOR SECTION 2.0 INST ALLA TION SECTION 2.0. INST ALLA TION 2.1 UNP ACKING AND INSPECTION 2.2 MOUNTING 2.3 FLOW -THROUGH INST ALLA TIONS 2.4 INSERTION INST ALLA TIONS 2.5 SUBMERSION INST ALLA TIONS 2.1 UNP ACKING AND INSPECTION. Inspect the carton for any damage. If damage is detected, contact the carrier immediately . Inspect ...

  • Emerson 399VP - page 8

    MODEL 399VP pH/ORP SENSOR SECTION 2.0 INST ALLA TION 4 FIGURE 2-1. Sensor Dimensions INCH MILLIMETER ...

  • Emerson 399VP - page 9

    MODEL 399VP pH/ORP SENSOR SECTION 2.0 INST ALLA TION 2.3 FLOW-THROUGH INST ALLA TIONS. The Model 399VP Sensor has a 1 in. MNPT process connection at the front and back end of the sensor for mount- ing directly into a 1 in. or 1½ in. tee. See Figure 2-2 for installation configurations. It is recommended that shut-of f valves be provided for sensor ...

  • Emerson 399VP - page 10

    6 MODEL 399VP pH/ORP SENSOR SECTION 2.0 INST ALLA TION 2.4 INSERTION INST ALLA TIONS. The Model 399VP sensor has forward and rear facing 1 in. MNPT process connections for pipeline installations (see Figure 2-4 or 2-5) and flange connections. The Model 399VP is also suitable for side-of-tank inst allation. The Model 399VP must be mounted at least 1 ...

  • Emerson 399VP - page 11

    MODEL 399VP pH/ORP SENSOR SECTION 2.0 INST ALLA TION 2.5 SUBMERSION INST ALLA TIONS. The Model 399VP Sensor has a 1 in. MNPT process connection at the back of the sensor (see Figure 2-1). Utilizing a standard 1 in. union (such as PN 9330022), the sensor may be mounted to a 1 in. SCH 80 CPVC or PVDF st and- pipe. The sensor can also be mounted in th ...

  • Emerson 399VP - page 12

    8 MODEL 399VP pH/ORP SENSOR SECTION 2.0 INST ALLA TION FIGURE 2-6. SENSOR SHOWN IN HANDRAIL MOUNTING ASSEMBL Y (PN 1 1275-01) The Handrail Mounting Assembly can be used in applications where Model 399VP needs to be placed in a pond, tank, aer- ation basin, or open channel. Handrail Mounting Assembly includes all parts seen in drawing; sensor is sol ...

  • Emerson 399VP - page 13

    MODEL 399VP pH/ORP SENSOR SECTION 2.0 INST ALLA TION FIGURE 2-8. JUNCTION BOX (PN 22719-02) AND PIPE MOUNT UNION ACCESSOR Y (PN 9330022) FIGURE 2-7. JET SPRA Y CLEANER (PN 12707-00) USED WITH MODEL 399VP The Jet S pray Cleaner eliminates routine manual sensor maintenance by cleaning Model 399VP with water or compressed air . Flow through the cleane ...

  • Emerson 399VP - page 14

    10 MODEL 399VP pH/ORP SENSOR SECTION 3.0 WIRING SECTION 3.0. WIRING Figures 3-2 thru 3-23 provide the diagrams for wiring the 399VP sensor to various analyzer/transmitters. 1. If the cable needs to be extended, use a high quali- ty cable available from Rosemount Analytical. Refer to Figures 3-2 through 3-23 for the appropriate junc- tion box part n ...

  • Emerson 399VP - page 15

    11 MODEL 399VP pH/ORP SENSOR SECTION 3.0 WIRING FIGURE 3-1. Wire and Connector Pin Functions FIGURE 3-2. Model 81 Wiring FIGURE 3-3. Model 1 181 Wiring FIGURE 3-4. Model 54 Wiring through Remote Junction Box MODEL 399VP ...

  • Emerson 399VP - page 16

    12 MODEL 399VP pH/ORP SENSOR SECTION 3.0 WIRING FIGURE 3-6. Model 2081 Wiring FIGURE 3-8. Model 2081 Wiring through Remote Junction Box FIGURE 3-7. Model 1 181, 1050/1060, and 1003/1023 Wiring through Remote Junction Box FIGURE 3-5. Model 54 Wiring ...

  • Emerson 399VP - page 17

    13 MODEL 399VP pH/ORP SENSOR SECTION 3.0 WIRING FIGURE 3-9. Model 81 Wiring through Remote Junction Box FIGURE 3-10. Model 3081 & 4081 Wiring through Remote Junction Box FIGURE 3-12. Model 3081 & 4081 Wiring FIGURE 3-1 1. Model 1055-22-32 Wiring ...

  • Emerson 399VP - page 18

    14 MODEL 399VP pH/ORP SENSOR SECTION 3.0 WIRING FIGURE 3-13. Model 1054 Wiring FIGURE 3-14. Model 1054A/B and 2054 Wiring FIGURE 3-16. Model 1054A/B & 2054 Wiring through a Remote Junction Box FIGURE 3-15. Model 1054 Wiring through a Remote Junction Box ...

  • Emerson 399VP - page 19

    MODEL 399VP pH/ORP SENSOR SECTION 3.0 WIRING FIGURE 3-18. Model 2700 Wiring FIGURE 3-17. Model SCL-(P/Q) Wiring FIGURE 3-19. Model 54epH Wiring FIGURE 3-20. Model 1055-22-32 Wiring through Remote Junction Boxes 15 ...

  • Emerson 399VP - page 20

    16 MODEL 399VP pH/ORP SENSOR SECTION 3.0 WIRING FIGURE 3-21. Wiring Model 399VP-09, 400, 401, 402, 403, and 404 to Model 1055 (Pipe/W all Mount). DWG . NO. REV . 41055107 C ...

  • Emerson 399VP - page 21

    17 MODEL 399VP pH/ORP SENSOR SECTION 3.0 WIRING FIGURE 3-22. Wiring Model 399VP-09, 400, 401, 402, 403, and 404 to Model 1055 (Panel Mount). DWG . NO. REV . 41055108 C ...

  • Emerson 399VP - page 22

    18 MODEL 399VP pH/ORP SENSOR SECTION 3.0 WIRING FIGURE 3-23. Prep aration of Extension Cable (PN 9200273) This cable is shipped as a raw cable. Cable must be prepared, as shown above, before inst allation. T o avoid preparing raw cable, PN 23646-01 can be used. PN 23646-01 cable offers the wires already prepared for inst allation. ...

  • Emerson 399VP - page 23

    19 MODEL 399VP pH/ORP SENSOR SECTION 4.0 ST ART -UP AND CALIBRA TION SECTION 4.0. ST ART -UP AND CALIBRA TION 4.1 SENSOR PREP ARA TION Shake down the sensor to remove any air bubbles that may be present in the tip of the pH glass bulb. In most cases, the pH sensor can simply be installed as shipped, and readings with an accuracy of ± 0.6 pH may be ...

  • Emerson 399VP - page 24

    20 MODEL 399VP pH/ORP SENSOR SECTION 4.0 ST ART -UP AND CALIBRA TION 4.3 ORP CALIBRA TION Most industrial applications have a number of ORP reactions occurring in sequence or simult aneously . There can be several components that are oxidized or reduced by the reagent s that are used. Theoretically , the ORP poten- tial is absolute because it is th ...

  • Emerson 399VP - page 25

    21 MODEL 399VP pH/ORP SENSOR SECTION 5.0 MAINTENANCE SECTION 5.0. MAINTENANCE 5.1 GENERAL The Model 399VP sensor is a disposable type sensor and therefore requires minimum maintenance. The sensor should be kept clean and free of debris and sediment at all times. The frequency of cleaning, by wiping or brush- ing with a soft cloth or brush, is deter ...

  • Emerson 399VP - page 26

    22 MODEL 399VP pH/ORP SENSOR SECTION 5.0 MAINTENANCE 5.3 pH ELECTRODE CLEANING If the electrode is coated or dirty , clean as follows: 1. Remove the sensor from process. 2. Wipe the glass bulb with a soft, clean, lint free cloth or tissue. If this does not remove the dirt or coating, go to S tep 3 (detergents clean oil and grease; acids remove scal ...

  • Emerson 399VP - page 27

    23 MODEL 399VP pH/ORP SENSOR SECTION 6.0 TROUBLESHOOTING SECTION 6.0. TROUBLESHOOTING TROUBLESHOOTING . The table below lists common problems, causes and remedies typically encountered in process measurement. Problem Probable Cause Remedy Meter reads off scale (display T .C. element shorted. Check T .C. element as instructed reads overrange). in Se ...

  • Emerson 399VP - page 28

    MODEL 399VP pH/ORP SENSOR SECTION 7.0 RETURN OF MA TERIAL SECTION 7.0 RETURN OF MA TERIAL 7.1 GENERAL. T o expedite the repair and return of instrument s, proper communication between the customer and the factory is important. Before returning a product for rep air , call 1-949-757-8500 for a R e t urn Materials Authorization (RMA) number . 7.2 W A ...

  • Emerson 399VP - page 29

    FROM: RETURN BILL TO: _____________________________ _____________________________ _____________________________ _____________________________ _____________________________ _____________________________ _____________________________ _____________________________ _____________________________ CUSTOMER/USER MUST SUBMIT MA TERIAL SAFETY SHEET (MSDS) OR ...

  • Emerson 399VP - page 30

    A W orldwide Network of Sales and Service Emerson Process Management’s field sales of fices are your source for more information on the fill line of Rosemount Analytical products. Field sales personnel will work closely with you to supply technical dat a and application information. For more information, please contact your nearest Emerson Proces ...

  • Emerson 399VP - page 31

    W ARRANTY Seller warrants that the firmware will execute the programming instructions provided by Seller , and that the Goods manufactured or Services provided by Seller will be free from defects in materials or workmanship under normal use and care until the expira- tion of the applicable warranty period. Goods are warranted for twelve (12) months ...

  • Emerson 399VP - page 32

    Credit Cards for U.S. Purchases Only . The right people, the right answers, right now . ON-LINE ORDERING NOW A V AILABLE ON OUR WEB SITE http://www .raihome.com Specifications subject to change without notice. Emerson Process Management Liquid Division 2400 Barranca Parkway Irvine, CA 92606 USA T el: (949) 757-8500 Fax: (949) 474-7250 http://www .r ...

Fabricant Emerson Catégorie Stud Sensor

Les documents que nous recevons du fabricant de l'appareilEmerson 399VP peuvent être divisés en plusieurs groupes. Ceux-ci sont, entre autres:
- dessins techniques Emerson
- manuels d’utilisations 399VP
- fiches produit Emerson
- dépliants
- ou étiquettes-énergie Emerson 399VP
Tous sont importants, mais les informations les plus importantes du point de vue de l'utilisation de l'appareil se trouvent dans le manuel d’utilisation Emerson 399VP.

Un groupe de documents appelé manuels d’utilisation est également divisé en types plus spécifiques, tels que: Manuels d’installation Emerson 399VP, manuels d’entretien, brefs manuels ou manuels de l’utilisateur Emerson 399VP. Selon vos besoins, vous devriez chercher le document dont vous avez besoin. Sur notre site, vous pouvez voir le manuel le plus populaire d’utilisation du produit Emerson 399VP.

Manuels d’utilsiation similaires

Manuel d’utilisation complet de l’appareil Emerson 399VP, quelle devrait-elle être?
Le manuel d’utilisation, également appelé le mode d’emploi, ou tout simplement le manuel, est un document technique destiné à aider à utiliser Emerson 399VP par les utilisateurs. Des manuels sont généralement écrits par un rédacteur technique, mais dans un langage accessible à tous les utilisateurs Emerson 399VP.

Le manuel d’utilisation complet Emerson, devrait inclure plusieurs éléments de base. Certains d'entre eux sont moins importants, tels que: la couverture / page de titre ou pages d'auteur. Cependant, la partie restante, devrait nous fournir des informations importantes du point de vue de l'utilisateur.

1. Introduction et des conseils sur la façon d'utiliser le manuel Emerson 399VP - Au début de chaque manuel, nous devrions trouver des indices sur la façon d'utiliser le document. Il doit contenir des informations sur l'emplacement de la table des matières Emerson 399VP, FAQ ou des problèmes les plus fréquents - les points qui sont les plus souvent recherchés par les utilisateurs de chaque manuel
2. Table des matières - index de tous les conseils pour lEmerson 399VP qui peuvent être trouvés dans le document courant
3. Conseils sur la façon d'utiliser les fonctions de base de l’appareil Emerson 399VP - qui devraient nous aider dans les premières étapes lors de l'utilisation Emerson 399VP
4. Troubleshooting - séquence systématique des activités qui nous aideront à diagnostiquer et ensuite résoudre les principaux problèmes de Emerson 399VP
5. FAQ - questions fréquemment posées
6. Détails du contact Informations sur l'endroit où chercher le contact avec le fabricant / service Emerson 399VP dans un pays donné, si le problème ne peut être résolu par nous-mêmes.

Avez-vous une question à propos de Emerson 399VP?

Utiliser le formulaire ci-dessous

Si vous n’avez pas résolu votre problème avec Emerson 399VP, avec l'aide du manuel que vous avez trouvé, posez une question en utilisant le formulaire ci-dessous. Si un utilisateur a eu un problème similaire avec Emerson 399VP il est probable qu’il a envie de partager la façon de le résoudre.

Réécrire le texte de l'image

Commentaires (0)