Manuale per l’uso Emerson Combination pH/ORP Sensor

32 pagine 0.77 mb
Scarica

Vai alla pagina of 32

Summary
  • Emerson Combination pH/ORP Sensor - page 1

    Model 399VP Combina tion pH/ORP Sensor Instruction Manual PN 51-399VP/rev .B January 2005 ...

  • Emerson Combination pH/ORP Sensor - page 2

    CAUTION SENSOR/PROCESS APPLICA TION COMP A TIBILITY The wetted sensor materials may not be comp atible with process composition and operating conditions. Application comp atibility is entirely the responsibili- ty of the user . DANGER HAZARDOUS AREA INST ALLA TION Installations near flammable liquids or in haz- ardous area locations must be careful ...

  • Emerson Combination pH/ORP Sensor - page 3

    MODEL 399VP pH/ORP T ABLE OF CONTENTS MODEL 399VP pH/ORP SENSOR T ABLE OF CONTENTS SECTION TITLE P AGE 1.0 DESCRIPTION AND SPECIFICA TIONS .......................................................... 1 1.1 Features and Applications ................................................................................. 1 1.2 Physical S pecifications ....... ...

  • Emerson Combination pH/ORP Sensor - page 4

    MODEL 399VP pH/ORP T ABLE OF CONTENTS MODEL 399VP pH/ORP SENSOR LIST OF FIGURES Figure No. Title Page 2-1 Sensor Dimensions ........................................................................................... 4 2-2 Model 399VP Shown in V arious Flow-Through Installations ............................. 5 2-3 Model 399VP Shown in Low Flow Cell ...

  • Emerson Combination pH/ORP Sensor - page 5

    MODEL 399VP pH/ORP SENSOR SECTION 1.0 DESCRIPTION AND SPECIFICA TIONS SECTION 1.0. DESCRIPTION AND SPECIFICA TIONS 1.1 FEA TURES AND APPLICA TIONS 1.2 PHYSICAL SPECIFICA TIONS 1.3 ORDERING INFORMA TION 1 1.1 FEA TURES AND APPLICA TIONS The Rosemount Analytical Model 399VP Sensor meas- ures the pH or the Oxidation Reduction Potential (ORP) of aqueou ...

  • Emerson Combination pH/ORP Sensor - page 6

    MODEL 399VP pH/ORP SENSOR SECTION 1.0 DESCRIPTION AND SPECIFICA TIONS 1.3 ORDERING INFORMA TION The Model 399VP pH/ORP Sensor is available with custom glass electrodes housed in a molded Tefzel body with 1 in. MNPT threads suit able for insertion, submersion or flow through installation. The sensor includes a general purpose hemi bulb pH electrode ...

  • Emerson Combination pH/ORP Sensor - page 7

    MODEL 399VP pH/ORP SENSOR SECTION 2.0 INST ALLA TION SECTION 2.0. INST ALLA TION 2.1 UNP ACKING AND INSPECTION 2.2 MOUNTING 2.3 FLOW -THROUGH INST ALLA TIONS 2.4 INSERTION INST ALLA TIONS 2.5 SUBMERSION INST ALLA TIONS 2.1 UNP ACKING AND INSPECTION. Inspect the carton for any damage. If damage is detected, contact the carrier immediately . Inspect ...

  • Emerson Combination pH/ORP Sensor - page 8

    MODEL 399VP pH/ORP SENSOR SECTION 2.0 INST ALLA TION 4 FIGURE 2-1. Sensor Dimensions INCH MILLIMETER ...

  • Emerson Combination pH/ORP Sensor - page 9

    MODEL 399VP pH/ORP SENSOR SECTION 2.0 INST ALLA TION 2.3 FLOW-THROUGH INST ALLA TIONS. The Model 399VP Sensor has a 1 in. MNPT process connection at the front and back end of the sensor for mount- ing directly into a 1 in. or 1½ in. tee. See Figure 2-2 for installation configurations. It is recommended that shut-of f valves be provided for sensor ...

  • Emerson Combination pH/ORP Sensor - page 10

    6 MODEL 399VP pH/ORP SENSOR SECTION 2.0 INST ALLA TION 2.4 INSERTION INST ALLA TIONS. The Model 399VP sensor has forward and rear facing 1 in. MNPT process connections for pipeline installations (see Figure 2-4 or 2-5) and flange connections. The Model 399VP is also suitable for side-of-tank inst allation. The Model 399VP must be mounted at least 1 ...

  • Emerson Combination pH/ORP Sensor - page 11

    MODEL 399VP pH/ORP SENSOR SECTION 2.0 INST ALLA TION 2.5 SUBMERSION INST ALLA TIONS. The Model 399VP Sensor has a 1 in. MNPT process connection at the back of the sensor (see Figure 2-1). Utilizing a standard 1 in. union (such as PN 9330022), the sensor may be mounted to a 1 in. SCH 80 CPVC or PVDF st and- pipe. The sensor can also be mounted in th ...

  • Emerson Combination pH/ORP Sensor - page 12

    8 MODEL 399VP pH/ORP SENSOR SECTION 2.0 INST ALLA TION FIGURE 2-6. SENSOR SHOWN IN HANDRAIL MOUNTING ASSEMBL Y (PN 1 1275-01) The Handrail Mounting Assembly can be used in applications where Model 399VP needs to be placed in a pond, tank, aer- ation basin, or open channel. Handrail Mounting Assembly includes all parts seen in drawing; sensor is sol ...

  • Emerson Combination pH/ORP Sensor - page 13

    MODEL 399VP pH/ORP SENSOR SECTION 2.0 INST ALLA TION FIGURE 2-8. JUNCTION BOX (PN 22719-02) AND PIPE MOUNT UNION ACCESSOR Y (PN 9330022) FIGURE 2-7. JET SPRA Y CLEANER (PN 12707-00) USED WITH MODEL 399VP The Jet S pray Cleaner eliminates routine manual sensor maintenance by cleaning Model 399VP with water or compressed air . Flow through the cleane ...

  • Emerson Combination pH/ORP Sensor - page 14

    10 MODEL 399VP pH/ORP SENSOR SECTION 3.0 WIRING SECTION 3.0. WIRING Figures 3-2 thru 3-23 provide the diagrams for wiring the 399VP sensor to various analyzer/transmitters. 1. If the cable needs to be extended, use a high quali- ty cable available from Rosemount Analytical. Refer to Figures 3-2 through 3-23 for the appropriate junc- tion box part n ...

  • Emerson Combination pH/ORP Sensor - page 15

    11 MODEL 399VP pH/ORP SENSOR SECTION 3.0 WIRING FIGURE 3-1. Wire and Connector Pin Functions FIGURE 3-2. Model 81 Wiring FIGURE 3-3. Model 1 181 Wiring FIGURE 3-4. Model 54 Wiring through Remote Junction Box MODEL 399VP ...

  • Emerson Combination pH/ORP Sensor - page 16

    12 MODEL 399VP pH/ORP SENSOR SECTION 3.0 WIRING FIGURE 3-6. Model 2081 Wiring FIGURE 3-8. Model 2081 Wiring through Remote Junction Box FIGURE 3-7. Model 1 181, 1050/1060, and 1003/1023 Wiring through Remote Junction Box FIGURE 3-5. Model 54 Wiring ...

  • Emerson Combination pH/ORP Sensor - page 17

    13 MODEL 399VP pH/ORP SENSOR SECTION 3.0 WIRING FIGURE 3-9. Model 81 Wiring through Remote Junction Box FIGURE 3-10. Model 3081 & 4081 Wiring through Remote Junction Box FIGURE 3-12. Model 3081 & 4081 Wiring FIGURE 3-1 1. Model 1055-22-32 Wiring ...

  • Emerson Combination pH/ORP Sensor - page 18

    14 MODEL 399VP pH/ORP SENSOR SECTION 3.0 WIRING FIGURE 3-13. Model 1054 Wiring FIGURE 3-14. Model 1054A/B and 2054 Wiring FIGURE 3-16. Model 1054A/B & 2054 Wiring through a Remote Junction Box FIGURE 3-15. Model 1054 Wiring through a Remote Junction Box ...

  • Emerson Combination pH/ORP Sensor - page 19

    MODEL 399VP pH/ORP SENSOR SECTION 3.0 WIRING FIGURE 3-18. Model 2700 Wiring FIGURE 3-17. Model SCL-(P/Q) Wiring FIGURE 3-19. Model 54epH Wiring FIGURE 3-20. Model 1055-22-32 Wiring through Remote Junction Boxes 15 ...

  • Emerson Combination pH/ORP Sensor - page 20

    16 MODEL 399VP pH/ORP SENSOR SECTION 3.0 WIRING FIGURE 3-21. Wiring Model 399VP-09, 400, 401, 402, 403, and 404 to Model 1055 (Pipe/W all Mount). DWG . NO. REV . 41055107 C ...

  • Emerson Combination pH/ORP Sensor - page 21

    17 MODEL 399VP pH/ORP SENSOR SECTION 3.0 WIRING FIGURE 3-22. Wiring Model 399VP-09, 400, 401, 402, 403, and 404 to Model 1055 (Panel Mount). DWG . NO. REV . 41055108 C ...

  • Emerson Combination pH/ORP Sensor - page 22

    18 MODEL 399VP pH/ORP SENSOR SECTION 3.0 WIRING FIGURE 3-23. Prep aration of Extension Cable (PN 9200273) This cable is shipped as a raw cable. Cable must be prepared, as shown above, before inst allation. T o avoid preparing raw cable, PN 23646-01 can be used. PN 23646-01 cable offers the wires already prepared for inst allation. ...

  • Emerson Combination pH/ORP Sensor - page 23

    19 MODEL 399VP pH/ORP SENSOR SECTION 4.0 ST ART -UP AND CALIBRA TION SECTION 4.0. ST ART -UP AND CALIBRA TION 4.1 SENSOR PREP ARA TION Shake down the sensor to remove any air bubbles that may be present in the tip of the pH glass bulb. In most cases, the pH sensor can simply be installed as shipped, and readings with an accuracy of ± 0.6 pH may be ...

  • Emerson Combination pH/ORP Sensor - page 24

    20 MODEL 399VP pH/ORP SENSOR SECTION 4.0 ST ART -UP AND CALIBRA TION 4.3 ORP CALIBRA TION Most industrial applications have a number of ORP reactions occurring in sequence or simult aneously . There can be several components that are oxidized or reduced by the reagent s that are used. Theoretically , the ORP poten- tial is absolute because it is th ...

  • Emerson Combination pH/ORP Sensor - page 25

    21 MODEL 399VP pH/ORP SENSOR SECTION 5.0 MAINTENANCE SECTION 5.0. MAINTENANCE 5.1 GENERAL The Model 399VP sensor is a disposable type sensor and therefore requires minimum maintenance. The sensor should be kept clean and free of debris and sediment at all times. The frequency of cleaning, by wiping or brush- ing with a soft cloth or brush, is deter ...

  • Emerson Combination pH/ORP Sensor - page 26

    22 MODEL 399VP pH/ORP SENSOR SECTION 5.0 MAINTENANCE 5.3 pH ELECTRODE CLEANING If the electrode is coated or dirty , clean as follows: 1. Remove the sensor from process. 2. Wipe the glass bulb with a soft, clean, lint free cloth or tissue. If this does not remove the dirt or coating, go to S tep 3 (detergents clean oil and grease; acids remove scal ...

  • Emerson Combination pH/ORP Sensor - page 27

    23 MODEL 399VP pH/ORP SENSOR SECTION 6.0 TROUBLESHOOTING SECTION 6.0. TROUBLESHOOTING TROUBLESHOOTING . The table below lists common problems, causes and remedies typically encountered in process measurement. Problem Probable Cause Remedy Meter reads off scale (display T .C. element shorted. Check T .C. element as instructed reads overrange). in Se ...

  • Emerson Combination pH/ORP Sensor - page 28

    MODEL 399VP pH/ORP SENSOR SECTION 7.0 RETURN OF MA TERIAL SECTION 7.0 RETURN OF MA TERIAL 7.1 GENERAL. T o expedite the repair and return of instrument s, proper communication between the customer and the factory is important. Before returning a product for rep air , call 1-949-757-8500 for a R e t urn Materials Authorization (RMA) number . 7.2 W A ...

  • Emerson Combination pH/ORP Sensor - page 29

    FROM: RETURN BILL TO: _____________________________ _____________________________ _____________________________ _____________________________ _____________________________ _____________________________ _____________________________ _____________________________ _____________________________ CUSTOMER/USER MUST SUBMIT MA TERIAL SAFETY SHEET (MSDS) OR ...

  • Emerson Combination pH/ORP Sensor - page 30

    A W orldwide Network of Sales and Service Emerson Process Management’s field sales of fices are your source for more information on the fill line of Rosemount Analytical products. Field sales personnel will work closely with you to supply technical dat a and application information. For more information, please contact your nearest Emerson Proces ...

  • Emerson Combination pH/ORP Sensor - page 31

    W ARRANTY Seller warrants that the firmware will execute the programming instructions provided by Seller , and that the Goods manufactured or Services provided by Seller will be free from defects in materials or workmanship under normal use and care until the expira- tion of the applicable warranty period. Goods are warranted for twelve (12) months ...

  • Emerson Combination pH/ORP Sensor - page 32

    Credit Cards for U.S. Purchases Only . The right people, the right answers, right now . ON-LINE ORDERING NOW A V AILABLE ON OUR WEB SITE http://www .raihome.com Specifications subject to change without notice. Emerson Process Management Liquid Division 2400 Barranca Parkway Irvine, CA 92606 USA T el: (949) 757-8500 Fax: (949) 474-7250 http://www .r ...

Fabbricante Emerson Categoria Stud Sensor

I documenti che otteniamo dal fabbricante del dispositivoEmerson Combination pH/ORP Sensor possono essere suddivisi in diversi gruppi. Questi sono tra gli altri:
- disegni tecnici Emerson
- manuali per l’uso Combination pH/ORP Sensor
- schede prodotto Emerson
- opuscoli
- o etichette energetiche Emerson Combination pH/ORP Sensor
Tutti sono importanti, ma le informazioni più importanti dal punto di vista di utilizzo del dispositivo possono essere trovate nel manuale per l’uso Emerson Combination pH/ORP Sensor.

Un gruppo di documenti denominato manuali per l’uso, è anche suddiviso in tipi più specifici, come: Manuali di installazione Emerson Combination pH/ORP Sensor, manuali di manutenzione, brevi manuali o manuali utente Emerson Combination pH/ORP Sensor. A seconda delle esigenze, si dovrebbe cercare il documento necessario. Nel nostro sito web, puoi visualizzare il manuale più popolare per l'uso del prodotto Emerson Combination pH/ORP Sensor.

Manuali per l’uso simili

Manuale completo del dispositivo Emerson Combination pH/ORP Sensor, come deve essere?
Il manuale per l’uso, denominato anche istruzioni per l’uso, o semplicemente il manuale è un documento tecnico progettato per aiutare ad utilizzare Emerson Combination pH/ORP Sensor dagli utenti. I manuali sono di solito scritti da uno scrittore tecnico, ma in un linguaggio accessibile a tutti gli utenti Emerson Combination pH/ORP Sensor.

Il manuale per l’uso completo Emerson, dovrebbe includere vari elementi di base. Alcuni di loro sono meno importanti, come ad esempio: copertina / pagina del titolo o pagina di autore. Tuttavia, la parte rimanente, dovrebbe fornire informazioni importanti dal punto di vista dell'utente.

1. Introduzione e la guida su come utilizzare il manuale Emerson Combination pH/ORP Sensor - All'inizio di ogni manuale, dovremmo trovare indizi su come utilizzare il documento. Dovrebbe contenere informazioni sulla posizione dell’indice Emerson Combination pH/ORP Sensor, FAQ o i problemi più comuni : i punti che sono più spesso cercati dagli utenti di ogni manuale
2. Indice - elenco di tutti i suggerimenti per Emerson Combination pH/ORP Sensor che possono essere trovati nel documento corrente
3. Suggerimenti su come utilizzare le funzioni di base del dispositivo Emerson Combination pH/ORP Sensor - che ci dovrebbe aiutare nei primi passi durante l'uso Emerson Combination pH/ORP Sensor
4. Troubleshooting - sequenza sistematica delle attività che vi aiuterà a diagnosticare e, successivamente, risolvere i problemi più importanti del Emerson Combination pH/ORP Sensor
5. FAQ - domande più frequenti
6. Dettagli di contatto Informazioni su dove cercare il contatto con il fabbricante / servizio Emerson Combination pH/ORP Sensor in un dato paese, se il problema non può essere risolto da noi stessi.

Hai una domanda su Emerson Combination pH/ORP Sensor?

Utilizza il modulo sottostante

Se non hai risolto il problema con Emerson Combination pH/ORP Sensor con l'aiuto del manuale che hai trovato, fai una domanda utilizzando il modulo sottostante. Se un utente ha avuto un problema simile con Emerson Combination pH/ORP Sensor è probabile che voglia condividere il modo di risolverlo.

Riscrivi il testo dall’immagine

Commenti (0)