Manuale per l’uso Agilent Technologies 8753ET

28 pagine 3.37 mb
Scarica

Vai alla pagina of 28

Summary
  • Agilent Technologies 8753ET - page 1

    This document describes the performance and features of the following products: • Agilent 8753ES S-paramet er vector netw ork analyzer • Agilent 8753ES Option 011 vector network analyzer without test set • Agilent 8753ET T ransmission/ref lection vector netw ork analyzer F or more information about these analyzers, please see the follo wing d ...

  • Agilent Technologies 8753ET - page 2

    3 4 10 21 21 22 24 T able of Contents Definitions and T est Conditions System Performance Summaries Specifications and Characteristics Measurement Throughput Summary Options T est Set Specifications for Option 01 1 Accessories ...

  • Agilent Technologies 8753ET - page 3

    3 Specif ications describe t he instrument’s war ranted performance af ter a half-hour war m-up and ov er the temperature range of 25° ± 5° C, unless other- wise stat ed. Specifications for frequencies abo ve 3 GHz do not apply to instruments wit h Option 075 (75-ohm impedance). Supplemental charact eristics are typical but non- warranted per ...

  • Agilent Technologies 8753ET - page 4

    4 Agilent 8753ES (50-ohm systems) 7-mm test ports Configuration Network analyzer 8753ES Standard and Option 006 Calibration kit 85031B T est-port cables 11857D System dynamic range 30 kHz to 50 kHz 70 dB 1 50 kHz to 300 kHz 90 dB 1 300 kHz to 16 MHz 100 dB 16 MHz to 3 GHz 110 dB Option 014 108 dB 3 GHz to 6 GHz 105 dB Option 014 103 dB Measurement ...

  • Agilent Technologies 8753ET - page 5

    5 Agilent 8753ES (50-ohm systems) T ype-N test ports Configuration Network analyzer 8753ES Standard and Option 006 Calibration kit 85032B T est-port cables 11857D System dynamic range 30 kHz to 50 kHz 70 dB 1 50 kHz to 300 kHz 90 dB 1 300 kHz to 16 MHz 100 dB 16 MHz to 3 GHz 110 dB Option 014 108 dB 3 GHz to 6 GHz 105 dB Option 014 103 dB Measureme ...

  • Agilent Technologies 8753ET - page 6

    6 Agilent 8753ES (50-ohm systems) 3.5-mm test ports Configuration Network analyzer 8753ES Standard and Option 006 Calibration kit 85033D T est-port cables 11857D System dynamic range 30 kHz to 50 kHz 70 dB 1 50 kHz to 300 kHz 90 dB 1 300 kHz to 16 MHz 100 dB 16 MHz to 3 GHz 110 dB Option 014 108 dB 3 GHz to 6 GHz 105 dB Option 014 103 dB Measuremen ...

  • Agilent Technologies 8753ET - page 7

    Agilent 8753ES (75-Ohm Systems) T ype-N test ports Configuration Network analyzer 8753ES Option 075 Calibration kit 85036B T est-port cables 11857B System Dynamic Range 30 kHz to 50 kHz 68 dB 1 50 kHz to 300 kHz 90 dB 1 300 kHz to 16 MHz 96 dB 16 MHz to 3 GHz 106 dB Measurement Port Characteristics Frequency Range Corrected 30 kHz–300 kHz 2 300 k ...

  • Agilent Technologies 8753ET - page 8

    8 Agilent 8753ES (75-ohm systems) T ype-F test ports Configuration Network analyzer 8753E Option 075 Calibration kit 85039B T est-port cables 11857B System dynamic range 30 kHz to 50 kHz 68 dB 1 50 kHz to 300 kHz 90 dB 1 300 kHz to 16 MHz 96 dB 16 MHz to 3 GHz 106 dB Measurement Port Characteristics Data is shown for a T ype-F female reflection por ...

  • Agilent Technologies 8753ET - page 9

    9 1. One-port or enhanced-response calibration. 2. Enhanced-response or response-only calibration. 3. 2 dB less with Option 004. 4. Option 004 may degrade uncorrected source match as much as 2 dB. T ransmission uncertainty (enhanced-response calibration) T ransmission uncertainty (response calibration) Reflection uncertainty (one-port calibration) ...

  • Agilent Technologies 8753ET - page 10

    10 T est-Port Output Characteristics 7 Frequency characteristics Range 30 kHz to 3 GHz Option 006 30 kHz to 6 GHz Resolution 1 Hz Stability 9 ±7.5 ppm (0° to 55° C) ±3 ppm/year Option 1D5 ±0.05 ppm (0° to 55° C), ±0.5 ppm/year Accuracy ±10 ppm at 25° ±5° C Power range 2 -85 to +10 dBm Option 075/014 -85 to +8 dBm Power sweep range 25 dB ...

  • Agilent Technologies 8753ET - page 11

    11 External source mode 1 (CW time sweep only) Frequency range 300 kHz to 6 GHz R-channel input requirements (typical) Power level 0 to -25 dBm Spectral purity Maximum <-30 dBc spurious input Residual FM <20 kHz T ypical settling time 500 ms (automatic) 50 ms (manual) Frequency readout accuracy 0.1% (automatic) Input frequency accuracy requir ...

  • Agilent Technologies 8753ET - page 12

    12 T est Port Output Characteristics Frequency characteristics Range 300 kHz to 3 GHz Option 006 30 kHz to 6 GHz Resolution 1 Hz Stability typically ±7.5 ppm 0° to 55° C typically ±3 ppm/year Option 1D5 typically ±0.05 ppm 0° to 55° C typically ±0.5 ppm/year Accuracy ±10 ppm at 25° ± 5° C Power range -5 to +20 dBm Option 006 -5 to +18 d ...

  • Agilent Technologies 8753ET - page 13

    13 Frequency offset mode 3 Frequency range 300 kHz to 3 GHz Option 006 300 kHz to 6 GHz R-channel input requirements Power level 300 kHz to 3 GHz 0 to -35 dBm 3 GHz to 6 GHz 0 to -30 dBm LO spectral purity Maximum spurious input <-25 dBc Residual FM <20 kHz LO frequency accuracy ±1 MHz at nominal frequency External source mode 4 (CW time swe ...

  • Agilent Technologies 8753ET - page 14

    14 T est-Port Output Characteristics Frequency characteristics Range 300 kHz to 3 GHz Option 006 300 kHz to 6 GHz Resolution 1 Hz Stability 4 ±7.5 ppm (0° to 55° C) ±3 ppm/year Option 1D5 ±0.05 ppm (0° to 55° C) ±0.5 ppm/year Accuracy ±10 ppm at 25° ±5° C Power range -20 to +5 dBm -85 to +10 dBm (with Option 004) Resolution 0.01 dB Leve ...

  • Agilent Technologies 8753ET - page 15

    15 External source mode 1 (CW time sweep only) Frequency range 300 kHz to 6 GHz R-channel input requirements 2 Power level 0 to -25 dBm Spectral purity Maximum spurious input <-30 dBc Residual FM <20 kHz T ypical settling time 500 ms (automatic) 50 ms (manual) Frequency readout accuracy 2 0.1% (automatic) Input frequency accuracy requirement ...

  • Agilent Technologies 8753ET - page 16

    Measurement Number of display channels F our display channels a vailable. Number of measurement channels T w o primary and two auxiliary measurement channels available. Measurement parameters 8753ET: Ref lection, transmission, A, B, R, A/R, B/R, A/B. Conversion to im pedance or admitt ance. 8753ES: S11, S21, S12, S22, A, B, R, A/R, B/R, A/B. Conver ...

  • Agilent Technologies 8753ET - page 17

    17 Sweep trigger Set to either continuous, sing le, g roup sw eep, or external trigger . Set external trigger t o take a com- plete sweep or to measure individual points in a frequency , pow er or list sw eep. Power Set source po w er from -20 to +5 dBm f or the 8753ET 1 or from -85 to +10 dBm f or the 8753ES 2 . P o w er slope can be set in dBm/GH ...

  • Agilent Technologies 8753ET - page 18

    18 T race Functions Display data Displa y current measurement data, memor y dat a, or current measurement wit h measurement and memor y dat a simultaneously. T race math V ect or division or subtraction of cur rent linear measurement values and memor y dat a. Display annotations Star t/st op, center/span, or CW frequency , source lev el, scale/div ...

  • Agilent Technologies 8753ET - page 19

    19 Frequency offset on/off Sets the RF source to be swept at a fixed offset frequency abo ve the receiver as required in a sw ept RF/IF , fixed LO, mix er t est. Service menu Select the desired service test, service diagnostic, service or verification mode. T est Sequences Description Creat e, edit, sav e, or recall a series of front-panel k ey- s ...

  • Agilent Technologies 8753ET - page 20

    20 Front Panel Connectors 8753ES test ports (without Option 011) Connector type 7 mm, precision Impedance 50 ohms (nominal) 8753ES Option 011 test ports Connector type T ype-N female Impedance 50 ohms (nominal) 8753ES Option 075 test ports Connector type T ype-N female Impedance 75 ohms (nominal) 8753ET test ports Connector type T ype-N female Impe ...

  • Agilent Technologies 8753ET - page 21

    21 General Conditions RFI and EMI susceptibility: defined by VDE 0730, CISPR Publication 11, and FCC Class B Standards. ESD (electrostatic discharge): must be eliminated by use of static-safe w ork procedures and an anti- static bench mat. The f lexible r ubber ke ypad pro- tects ke y cont acts from dust, but the environment should be as dust-free ...

  • Agilent Technologies 8753ET - page 22

    22 Agilent 85046A/B S-Parameter T est Sets The 85046A/B S-parameter test sets provide the capability to measure ref lection and transmission characteristics (including S-parameters) of two- port devices in either direction wit h a sing le connection. The test sets are controlled from the 8753ES Option 011 and include a programmable step attenuat or ...

  • Agilent Technologies 8753ET - page 23

    23 Agilent 85047A S-Parameter T est Set The 85047A S-parameter test set provides the capability to simult aneously measure the ref lec- tion and transmission characteristics of two-port devices in either direction with a sing le connec- tion. This test set includes a frequency doubler that can be swit ched in by an 8753B/C Option 006 to measure 3 M ...

  • Agilent Technologies 8753ET - page 24

    24 Calibration Kits V ect or accuracy enhancement procedures require that t he systematic errors of the measurement sys- t em be characterized b y measuring kno wn devices (standards) on the syst em ov er the frequency range of interest. Return loss specif ications or typical values are pro vided where a vailable for the ter minations and adapters. ...

  • Agilent Technologies 8753ET - page 25

    25 Agilent 85038 Family of Calibration Kits The 85038A calibration kit contains open circuits, short circuits, and ter minations with bot h male and female 7-16 connectors. More economical v er- sions of this kit are available: the 85038M contains only male de vices, and the 85038F provides onl y female de vices (see f ollo wing t able). Each kit c ...

  • Agilent Technologies 8753ET - page 26

    26 V erification Kits Measuring kno wn devices other than t he st andards used in calibration is an easy way to v erify the proper operation of an 8753ES measurement sys- t em. Agilent of fers v erification kits which include devices, with data, for verifying the error -correct ed measurements of an 8753ES and 50-ohm test sets. 85029B 7-mm verifica ...

  • Agilent Technologies 8753ET - page 27

    27 Connectors 50-ohm type-N (f) to 75-ohm type-N (m) standard Option 004, 50-ohm type-N (m) to 75-ohm type-N (f) Dimensions 14-mm D x 70-mm L (0.56 in x 2.75 in) W eight Net 0.1 kg (0.316 lb) 50-Ohm Accessory Kits The 11853A 50-ohm type-N and the 11854A 50-ohm BNC accessory kits provide the RF com ponents generally required when using either the 85 ...

  • Agilent Technologies 8753ET - page 28

    Agilent 8347A RF Amplifier The 8347A RF amplifier delivers increased po wer acr oss a 300 kHz to 3 GHz frequency range. The 8347A pro vides lev eled output po w er without using an exter nal coupler and detect or , since these par ts are built-in. The exter nal ALC can be directly connected to the ext ernal AM input on the 8753ET/ES. This capabilit ...

Fabbricante Agilent Technologies Categoria Network Router

I documenti che otteniamo dal fabbricante del dispositivoAgilent Technologies 8753ET possono essere suddivisi in diversi gruppi. Questi sono tra gli altri:
- disegni tecnici Agilent Technologies
- manuali per l’uso 8753ET
- schede prodotto Agilent Technologies
- opuscoli
- o etichette energetiche Agilent Technologies 8753ET
Tutti sono importanti, ma le informazioni più importanti dal punto di vista di utilizzo del dispositivo possono essere trovate nel manuale per l’uso Agilent Technologies 8753ET.

Un gruppo di documenti denominato manuali per l’uso, è anche suddiviso in tipi più specifici, come: Manuali di installazione Agilent Technologies 8753ET, manuali di manutenzione, brevi manuali o manuali utente Agilent Technologies 8753ET. A seconda delle esigenze, si dovrebbe cercare il documento necessario. Nel nostro sito web, puoi visualizzare il manuale più popolare per l'uso del prodotto Agilent Technologies 8753ET.

Manuale completo del dispositivo Agilent Technologies 8753ET, come deve essere?
Il manuale per l’uso, denominato anche istruzioni per l’uso, o semplicemente il manuale è un documento tecnico progettato per aiutare ad utilizzare Agilent Technologies 8753ET dagli utenti. I manuali sono di solito scritti da uno scrittore tecnico, ma in un linguaggio accessibile a tutti gli utenti Agilent Technologies 8753ET.

Il manuale per l’uso completo Agilent Technologies, dovrebbe includere vari elementi di base. Alcuni di loro sono meno importanti, come ad esempio: copertina / pagina del titolo o pagina di autore. Tuttavia, la parte rimanente, dovrebbe fornire informazioni importanti dal punto di vista dell'utente.

1. Introduzione e la guida su come utilizzare il manuale Agilent Technologies 8753ET - All'inizio di ogni manuale, dovremmo trovare indizi su come utilizzare il documento. Dovrebbe contenere informazioni sulla posizione dell’indice Agilent Technologies 8753ET, FAQ o i problemi più comuni : i punti che sono più spesso cercati dagli utenti di ogni manuale
2. Indice - elenco di tutti i suggerimenti per Agilent Technologies 8753ET che possono essere trovati nel documento corrente
3. Suggerimenti su come utilizzare le funzioni di base del dispositivo Agilent Technologies 8753ET - che ci dovrebbe aiutare nei primi passi durante l'uso Agilent Technologies 8753ET
4. Troubleshooting - sequenza sistematica delle attività che vi aiuterà a diagnosticare e, successivamente, risolvere i problemi più importanti del Agilent Technologies 8753ET
5. FAQ - domande più frequenti
6. Dettagli di contatto Informazioni su dove cercare il contatto con il fabbricante / servizio Agilent Technologies 8753ET in un dato paese, se il problema non può essere risolto da noi stessi.

Hai una domanda su Agilent Technologies 8753ET?

Utilizza il modulo sottostante

Se non hai risolto il problema con Agilent Technologies 8753ET con l'aiuto del manuale che hai trovato, fai una domanda utilizzando il modulo sottostante. Se un utente ha avuto un problema simile con Agilent Technologies 8753ET è probabile che voglia condividere il modo di risolverlo.

Riscrivi il testo dall’immagine

Commenti (0)