Manuale per l’uso ResMed S8

20 pagine 0.21 mb
Scarica

Vai alla pagina of 20

Summary
  • ResMed S8 - page 1

    S8 Lightw eight™ II POSITIVE AIRWA Y PRESSURE DEVICE User Guide English ResMed Ltd (Manufacturer) 1 Elizabeth Macarthur Drive B ella Vista NSW 2153 A ustralia • ResMed Corp (US Designated Agent) 1 4040 Danielson Street P owa y CA 92064-6857 USA • ResMed (UK) Ltd (EU Authorized Representative) 96 Milton Park Abingdon Oxf ordshire OX1 4 4R Y UK ...

  • ResMed S8 - page 2

    i Contents Cont ents Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Contraindications 1 Adv erse Ef f ects 1 The S8 Lightweight II S ystem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Masks 2 Humidifiers 2 ResScan Data Card 2 Using the S8 Lightweight II . . . . . . . . . ...

  • ResMed S8 - page 3

    1 Introduction Introduction The S8 Lightweight II continuous posit ive airway pressure (CPAP) system is indicated for the treatment of obstructive sleep apnoe a (OSA) in patients > 30 kg. The S8 Lightweight II CPAP system is intended for home and hospital use. Contraindications The S8 Lightweight II is not a life sup port device and may stop ope ...

  • ResMed S8 - page 4

    2 The S8 Lightweight II System Your S8 Lightweight II comprises: • S8 Lightweight II positive air wa y pressure device • 2 m air tubing • P o wer cord • Carr y bag. Optional components include: • ResSca n™ Dat a Card • 3 m air tubing • DC-1 2 con verter . Masks For information on usi ng masks, see your mask manual. Recommended masks ...

  • ResMed S8 - page 5

    3 Using the S8 Lightweight II Using the S8 Lightweight II ! WARNING • Make sure the power cord and plug are in good conditio n and the equipment is not damaged. • Only Re sMed air tubing should be us ed with the device . A dif ferent type of air tubing may alter th e pressure you actually receive, reducing the effec ti veness of your treatment. ...

  • ResMed S8 - page 6

    4 Start Treatment 1 Make sure the power is on. The product name is displayed brie fl y on the LCD screen, then the standby (Ramp) screen appears. Th e key and LCD backl ights also turn on. 2 Lie down and arrange the air tubing so that it is free to move if you turn in your sleep. ! CAUTION Do not leave long lengths of air tubi ng around the top of ...

  • ResMed S8 - page 7

    5 Using the S8 Lightweight II The S8 Lightweight II keyp ad has the following keys: The keypad and LCD are equipped with backligh ts that come on when the device is powered. The LCD backlight turns off after two minutes of inactivity, and comes back on when you press a key. T he keypad backlight is always on while the S8 Lightweight II is on. Key F ...

  • ResMed S8 - page 8

    6 How to Change the Settings on your S8 Lightweight II You can change settings (such as Ramp Time) on the S8 Lightwei ght II by using the keypad and LCD. When the S8 Lightweight II i s in standby mode you can display a series of screens (a menu) on the LCD screen in order to view and change the settings for a particular function. Use the menu scree ...

  • ResMed S8 - page 9

    7 Using the S8 Lightweight II 1 . On the st andby (RAMP) screen, press . 2. Press until you see EPR LEVEL, then press (Change). 3. Press or until you see the setting that you require. 4. Press (Apply) to save the setting. How to change the language setting 1 . On the standby (Ramp) screen, press (Menu). 2. Press until you see LANG (Language ), then ...

  • ResMed S8 - page 10

    8 Using the Data Card to Collect data If your clinician needs to review your tr eatment, they will ask you to use the Data Card to copy data from your S8 Lightweight II, and to return th e card to them. The Data Card will be provided in a protective folder. Data that is copied to a Data Card is still stored and available on the S8 L ightweight II. ...

  • ResMed S8 - page 11

    9 Cleaning and Maintenance ! CAUTION • Do not hang the air t ubing in direct sunlight as it may harden over tim e and eventually crac k. • Do not use bleac h , c hlori ne, alcohol, or aromatic-based solutions , moisturising or antibacter ial soaps or scented oils to clean the air tubing or the S8 Lightweight I I. These solutions may caus e hard ...

  • ResMed S8 - page 12

    10 Servicing ! CAUTION • Do not at tempt to open the S8 Ligh tweight II case. There are no user serviceable par ts inside. • Inspection and repa ir should only be performed b y an authorised agent. Under no circumstances sh ould you attempt to serv ice or repair the S8 Lightweight II yourself. This device is intended to provide safe and reliabl ...

  • ResMed S8 - page 13

    11 T roubleshooting Anything special I should do before putting the mask on? Before wearing your mask, wash your face to remove excess facial oils. This will allow a better fit and prolong the life of the mask cushion. Do not use moisturiser when wearing a mask. My mouth feels dry during treatment—what can I do? You may experience dryness of the ...

  • ResMed S8 - page 14

    12 Insuf ficient air deli ver ed fr om S8 Lightw eight II The fo llow ing message i s display ed on the LCD af ter y ou try t o update set tings or copy data t o the D ata Card: Card Error Remove Card The fo llow ing message i s display ed on the LCD af ter y ou try t o update set tings or copy data t o the D ata Card: Settings Invalid Remove Card ...

  • ResMed S8 - page 15

    13 T echnical Specifications Technical Specifications . Note: The manufactu rer reserves the righ t to change these specif ications without no tice. Operating pressur e range 4 to 20 cm H 2 O Dimensions (H x W x D) 112 mm x 164 mm x 145 mm W eight 1.3 kg P ower Supply • Input ra nge for S8 L ight weight II with HumidAire 4i :1 00–240V , 50/60Hz ...

  • ResMed S8 - page 16

    14 Symbols which appear on the device At tention, consult accompan ying documents; Follow instru ctions for use; Drip proof; T ype CF equipment; Danger ous voltage; Class II equipment; Start/Stop; Manufacture r; Environmental infor mation WEEE 2002/96/EC is a European Directive that re quires the proper disposal o f electrical and electronic equipm ...

  • ResMed S8 - page 17

    15 General W arnings and Cautions • In the ev ent of pow er failure or mac hine malfunction, remove the mask. • The S8 Lightweight II can be set to deliver pressures up to 20 cm H 2 O. I n t h e unlik ely ev ent of certain fault cond itions, pressures up to 30 cm H 2 O are possible. • If o xygen is used with this de vice, the oxy gen flow mus ...

  • ResMed S8 - page 18

    16 Limited Warranty ResMed warrants that your ResMed product shall be free from defects in material and workmanship for the period specifie d below from the date of purchase by the initial co nsumer. Thi s warranty is not tr ansferable. Not e: Some models are not avai lable in all regions. If the product fails under conditions of normal use, ResMed ...

  • ResMed S8 - page 19

    17 Limited W arranty ...

  • ResMed S8 - page 20

    18 ...

Fabbricante ResMed Categoria Respiratory Product

I documenti che otteniamo dal fabbricante del dispositivoResMed S8 possono essere suddivisi in diversi gruppi. Questi sono tra gli altri:
- disegni tecnici ResMed
- manuali per l’uso S8
- schede prodotto ResMed
- opuscoli
- o etichette energetiche ResMed S8
Tutti sono importanti, ma le informazioni più importanti dal punto di vista di utilizzo del dispositivo possono essere trovate nel manuale per l’uso ResMed S8.

Un gruppo di documenti denominato manuali per l’uso, è anche suddiviso in tipi più specifici, come: Manuali di installazione ResMed S8, manuali di manutenzione, brevi manuali o manuali utente ResMed S8. A seconda delle esigenze, si dovrebbe cercare il documento necessario. Nel nostro sito web, puoi visualizzare il manuale più popolare per l'uso del prodotto ResMed S8.

Manuali per l’uso simili

Manuale completo del dispositivo ResMed S8, come deve essere?
Il manuale per l’uso, denominato anche istruzioni per l’uso, o semplicemente il manuale è un documento tecnico progettato per aiutare ad utilizzare ResMed S8 dagli utenti. I manuali sono di solito scritti da uno scrittore tecnico, ma in un linguaggio accessibile a tutti gli utenti ResMed S8.

Il manuale per l’uso completo ResMed, dovrebbe includere vari elementi di base. Alcuni di loro sono meno importanti, come ad esempio: copertina / pagina del titolo o pagina di autore. Tuttavia, la parte rimanente, dovrebbe fornire informazioni importanti dal punto di vista dell'utente.

1. Introduzione e la guida su come utilizzare il manuale ResMed S8 - All'inizio di ogni manuale, dovremmo trovare indizi su come utilizzare il documento. Dovrebbe contenere informazioni sulla posizione dell’indice ResMed S8, FAQ o i problemi più comuni : i punti che sono più spesso cercati dagli utenti di ogni manuale
2. Indice - elenco di tutti i suggerimenti per ResMed S8 che possono essere trovati nel documento corrente
3. Suggerimenti su come utilizzare le funzioni di base del dispositivo ResMed S8 - che ci dovrebbe aiutare nei primi passi durante l'uso ResMed S8
4. Troubleshooting - sequenza sistematica delle attività che vi aiuterà a diagnosticare e, successivamente, risolvere i problemi più importanti del ResMed S8
5. FAQ - domande più frequenti
6. Dettagli di contatto Informazioni su dove cercare il contatto con il fabbricante / servizio ResMed S8 in un dato paese, se il problema non può essere risolto da noi stessi.

Hai una domanda su ResMed S8?

Utilizza il modulo sottostante

Se non hai risolto il problema con ResMed S8 con l'aiuto del manuale che hai trovato, fai una domanda utilizzando il modulo sottostante. Se un utente ha avuto un problema simile con ResMed S8 è probabile che voglia condividere il modo di risolverlo.

Riscrivi il testo dall’immagine

Commenti (0)